06351543 [TESTO]
Multi-function VAC measuring instrument;型号: | 06351543 |
厂家: | Testo |
描述: | Multi-function VAC measuring instrument |
文件: | 总8页 (文件大小:1619K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
We measure it.
Multi-function VAC
measuring instrument
testo 480 – Cutting-edge
technology for professionals
Measurement of all VAC-relevant parameters: flow velocity,
temperature, humidity, pressure, light intensity, radiant heat,
degree of turbulence, CO2, PMV/PPD and WBGT index
°C
%RH
m/s
High-quality, digital probes and intelligent calibration
concept
Highly accurate, integrated differential pressure sensor
Fast and professional report creation via PC software
„EasyClimate“
hPa
Integrated, guided measurement programs:
ppm
CO2
– VAC grid measurement according to EN 12599
– PMV/PPD measurement according to ISO 7730
– degree of turbulence measurement according to EN 13779
– WBGT-measurement based on ISO 7243 / DIN 33403
Lux
With the testo 480, you record, analylze and document all
VAC-relevant parameters with only one instrument. The
multi-function VAC measuring instrument stands out above
all thanks to its accuracy and practice-oriented handling.
testo 480 supports assessors, consultants, technical
service providers or service technicians in the ventilation
and air conditioning field. Measurement tasks such as the
standardized adjustment of VAC systems in office,
residential and industrial buildings can be carried out
quickly and efficiently.
In addition to this, you test the relevant quality parameters
for your production and processing systems reliably and
precisely – thanks to the measuring instrument's
comprehensive probe range specially tailored to industrial
requirements.
The multi-function VAC measuring instrument is equipped
with intelligent, digital probes which are calibrated
independently of the hand-held instrument.
www.testo.com
testo 480
We measure it.
Technical data
Comfort measurement
testo 480
- High-end VAC measuring instrument testo 480 incl. PMV/PPD
measurement ±Part no. 0563 4800)
- Comfort probe for degree of turbulence measurement according to
EN 13779 ±Part no. 0628 0143)*
- Globe probe Ø 150mm, TC Type K, for measuring radiant heat ±Part
no. 0602 0743)
High-end VAC measuring instrument testo
480, including "EasyClimate" PC software,
power supply, USB cable and calibration
protocol
- IAQ probe for analyzing Indoor Air Quality, CO2, humidity,
temperature and absolute pressure measurement, incl. table tripod
±Part no. 0632 1543)*
- Lux probe for measuring light intensity ±Part no. 0635 0543)
- 2 x Plug-in head cable for digital probes ±Part no. 0430 0100)
- Tripod for workplace evaluation ±Part no. 0554 0743)
Part no. 0563 4800
- System case for comfort level measurement ±Part no. 0516 4801)
*Plug-in head cable required ±order no. 0430 0100)
General technical data
HVAC measurement
Probe connection
2 x TC Type K, 1 x differential pressure, 3 x
digital
- High-end VAC measuring instrument testo 480 incl. PMV/PPD
measurement ±Part no. 0563 4800)
- Vane measurement probe Ø 16 mm with telescope ±scaling max.
960 mm) and integrated measurement button ±Part no. 0635 9542)*
- Thermal flow velocity probe ±hot wire) Ø 10 mm, bendable by 90°
±200 mm) with telescope ±scaling max. 1100 mm) and integrated
measurement button ±Part no. 0635 1543)*
Interfaces
USB connection, SD card, mains unit,
infrared for fast printer
Operating temperature
Storage temperature
Power supply
0 to +40 °C
-20 to +60 °C
Rechargeable battery, plug-in mains unit for
long-term measurements and charging battery
- Humidity and temperature probe Ø 12 mm, highly accurate humidity
measurement with 1% accuracy ±Part no. 0636 9743)*
- Vane measurement probe Ø 100 mm, for measurements on
ventilation outlets ±Part no. 0635 9343)*
- Plug-in head cable for digital probes ±Part no. 0430 0100)
- System case for HVAC measurements ±Part no. 0516 4800)
Battery life
approx. 17 hours ±hand instrument without
probes, with 50 % display brightness)
Display
Colour graphic display
Memory
1.8 GB ±approx. 60.000.000 measurement
values)
*Plug-in head cable required ±order no. 0430 0100)
Technical data
Sensor type
Differential pressure, integrated
Absolute pressure,
integrated and external
Type K (NiCr-Ni)
Measuring range
-100 to +100 hPa
700 to 1100 hPa
± 3 hPa
-200 to +1370 °C
Accuracy
± 1 digit
± ±0.3 Pa +1% of m.v.) ±0 to +25 hPa)
± ±0.1 hPa + 1.5% of m.v.) ±+25.001 to +100 hPa)
± ±0.3 °C +0.1% of m.v.)
Resolution
0.001 hPa
0.1 hPa
0.1 °C
Sensor type
Measuring range
Resolution
Radiation temperature, globe
0 to +120 °C
Pt100
Vane, 16 mm
+0.6 to +50 m/s
0.1 m/s
-100 to +400 °C
0.01 °C
0.1 °C
Sensor type
Measuring range
Resolution
Vane, 100 mm
+0.1 to +15 m/s
0.01 m/s
Hot wire, Hot bulb
0 to +20 m/s
0.01 m/s
Comfort probe
0 to +5 m/s
0.01 m/s
Sensor type
Measuring range
Resolution
Testo humid. sensor, cap.
0 to 100 %RH
0.1 %RH
CO2
Lux
0 to 10000 ppm CO2
1 ppm CO2
0 to 100000 Lux
1 Lux
testo 480
We measure it.
Probes
Measuring
range
Accuracy ±± ꢀigit
Part no.
Probe type
Digital flow velocity probes
0.6 to 50 m/s
-10 to +70 °C
(0.2 m/s +1 % of m.v.) (0.6 to 40 m/s)
(0.2 m/s +2 % of m.v.) (40.1 to 50 m/s)
1.8 °C
Vane measurement probe Ø 16 mm
with telescope (scaling max. 960
mm) and integrated measurement
button*
0635 9542
0635 9552
0635 1543
0635 1024
0635 1050
0.6 to 50 m/s
-30 to +140 °C
(0.2 m/s +1 % of m.v.) (0.6 to 40 m/s)
(0.2 m/s +2 % of m.v.) (40.1 to 50 m/s)
(2.5 °C +0.8 % of m.v.)
High-temperature vane
measurement probe Ø 16 mm with
telescope (scaling max. 960 mm)
and integrated measurement
button*
0 to +20 m/s
-20 to +70 °C
0 to 100 %RH
+700 to +1100
hPa
(0.03 m/s +4% of m.v.)
0.5 °C
(1.8 %RH + 0.7% of m.v.)
3 hPa
Thermal flow velocity probe (hot
wire) Ø 10 mm, bendable by 90°
(200 mm) with telescope (scaling
max. 1100 mm) and integrated
measurement button*
0 to +20 m/s
-20 to +70 °C
(0.03 m/s +5% of m.v.)
0.5 °C
Thermal flow velocity probe (hot
wire) Ø 7.5 mm, with telescope
(max. 790 mm) and fixed plug-in
head cable
0 to +10 m/s
-20 to +70 °C
(0.03 m/s +5% of m.v.)
0.5 °C
Thermal flow velocity probe (robust
hot bulb) Ø 3 mm, with telescope,
(max. 860 mm) and fixed plug-in
head cable, for direction-
independent flow velocity
measurement
+0.1 to +15 m/s
0 to +60 °C
(0.1 m/s +1.5% of m.v.)
0.5 °C
Vane measurement probe Ø 100
mm, for measurements on
ventilation outlets*
0635 9343
0635 1048
0 to +5 m/s
0 to +50 °C
(0.02 m/s +5% of m.v.)
0.5 °C
Thermal flow velocity probe (hot
wire) Ø 10 mm, with telescope,
(max. 730 mm) and fixed plug-in
head cable, for the measurement of
air flow velocity in laboratory extrac-
tors according to EN 14175-3/-4
Digital comfort probes
0 to 100 %RH
-20 to +70 °C
(1.0 %RH + 0.7% of m.v.) 0 to 90 %RH
(1.4 %RH + 0.7% of m.v.) 90 to 100
Humidity and temperature probe Ø
12 mm, highly accurate humidity
measurement with 1% accuracy*
0636 9743
0632 1543
0628 0143
0602 0743
%RH
0.2 °C (+15 to +30 °C)
0.5 °C (remaining range)
0 to +50 °C
0 to 100 %RH
0 to +10000
ppm CO2
+700 to +1100
hPa
0.5 °C
IAQ probe for analyzing Indoor Air
Quality, CO2, humidity, temperature
and absolute pressure
(1.8 %RH + 0.7% of m.v.)
(75 ppm CO2 +3 % of m.v.)
0 to +5000 ppm CO2
(150 ppm CO2 +5 % of m.v.)
5001 to +10000 ppm CO2
3 hPa
measurement, incl. table tripod*
0 to +50 °C
0 to +5 m/s
+700 to +1100
hPa
0.5 °C
(0.03 m/s +4% of m.v.)
3 hPa
Comfort probe for degree of
turbulence measurement according
to EN 13779*
0 to +120 °C
Class 1
Globe probe Ø 150mm, TC Type K,
for measuring radiant heat
*Plug-in head cable required (order no. 0430 0100)
testo 480
We measure it.
Probes
Dimensions
Probe shaft/probe shaft tip
Measuring
range
Accuracy
t99
Part no.
Probe type
Digital comfort probes
0 to +100000
Lux
Class C according to DIN 5032-7;
f1 = 6% V-Lamda; f2 = 5% cos
Lux probe for measuring light
intensity
0635 0543
0635 8888
0 to +120 °C
+10 to +60 °C
+5 to +40 °C
Class 1
(0ꢀ25 °C +0ꢀ3% of mꢀvꢀ)
(0ꢀ25 °C +0ꢀ3% of mꢀvꢀ)
WBGT set (Wet Bulb Globe
Temperature) for the evaluation of
heat workplaces affected by heat,
based on ISO 7243 / DIN 33403-3,
consisting of globe, ambient
temperature and wet bulb
ID noꢀ
0699 6920/1
temperature probes, plug-in head
cables, tripod and case
Digital humidity probe
0 to 100 %RH
-20 to +180 °C
3 %RH (0 to 2 %RH)
Robust humidity probe
0636 9753
2 %RH (2ꢀ1 to 98 %RH)
3 %RH (98ꢀ1 to 100 %RH)
0ꢀ5 °C (-20 to 0 °C)
0ꢀ4 °C (0ꢀ1 to +50 °C)
0ꢀ5 °C (+50ꢀ1 to +180 °C)
Do not use in condensing amosphereꢀ For continuous use in high-humidity ranges
>80 %RH at ≤30 °C for > 12 h
>60 %RH at >30 °C for > 12 h
please refer to Testo customer service or contact us via the Testo websiteꢀ
Digital temperature probe
-100 to +400 °C
-200 to +300 °C
-80 to +300 °C
-100 to +265 °C
(0ꢀ15 °C + 0ꢀ2 % of mꢀvꢀ) (-100 to -0ꢀ01 °C)
(0ꢀ15 °C + 0ꢀ05 % of mꢀvꢀ) (0 to +100 °C)
(0ꢀ15 °C + 0ꢀ2 % of mꢀvꢀ)
Digital precision air probe, Pt100,
0614 0072
0614 0195
0614 0275
0614 0071
150 mm
Plug-in head cable required (Order
noꢀ 0430 0100)
(+100ꢀ01 to +350 °C)
(0ꢀ5 °C + 0ꢀ5 % of mꢀvꢀ)
(+350ꢀ01 to +400 °C)
Ø 9 mm
(2ꢀ5 °C + 0ꢀ8 % of mꢀvꢀ) (-40 to +300 °C)
Remaining range (-200 to -40ꢀ1 °C) is not
specified
Fast-reaction digital surface probe
150 mm
with sprung thermocouple Type K,
short-term up to +500 °C, Plug-in
Ø 10 mm
head cable required (Order noꢀ
0430 0100)
(0ꢀ3 °C (-80 to -40ꢀ001 °C)
(0ꢀ1 °C + 0ꢀ05 % of mꢀvꢀ) (-40 to -0ꢀ001 °C)
(0ꢀ05 °C (0 to +100 °C)
(0ꢀ05 °C + 0ꢀ05 % of mꢀvꢀ)
(+100ꢀ001 to +300 °C)
Highly precise digital
295 mm
immersion/penetration probe,
Pt100, Plug-in head cable required
Ø 4 mm
(Order noꢀ 0430 0100)
(0ꢀ30 °C + 0ꢀ3 % of mꢀvꢀ)
(-100 to -50ꢀ01 °C)
Flexible digital precision
penetration probe, PTFE cable
heat-proof to +300 °C, Plug-in
head cable required (Order noꢀ
0430 0100)
(0ꢀ15 °C + 0ꢀ2 % of mꢀvꢀ) (-50 to -0ꢀ01 °C)
(0ꢀ15 °C + 0ꢀ05 % of mꢀvꢀ) (0 to +100 °C)
(0ꢀ15 °C + 0ꢀ5 % of mꢀvꢀ)
(+100ꢀ01 to +265 °C)
Ø 4 mm
1000 mm
-100 to +400 °C
(0ꢀ15 °C + 0ꢀ2% of mꢀvꢀ) (-100 to -0ꢀ01 °C)
(0ꢀ15 °C + 0ꢀ05% of mꢀvꢀ) (0 to +100 °C)
(0ꢀ15 °C + 0ꢀ2% of mꢀvꢀ)
High-precision Pt100 immersion
and penetration probe, Plug-in
head cable required (order noꢀ 0430
0100)
0614 0073
200 mm
Ø 3 mm
(+100ꢀ01 to +350 °C)
(0ꢀ5 °C + 0ꢀ5% of mꢀvꢀ)
(+350ꢀ01 to +400 °C)
Special versions of the Pt100 probe on request ( eꢀgꢀ as a surface probe or air probe, probe shaft extended, strengthened)
Adapter cable for connecting analog Pt100 probe to testo 480
testo 480
We measure it.
Probes
Dimensions
Probe shaft/probe shaft tip
Measuring
range
Accuracy
t99
Part no.
Probe type
Analog temperature probe
-60 to +400 °C
-60 to +300 °C
Class 2 1)
25 s
Robust air probe, T/C Type K,
Fixed cable
0602 1793
0602 0393
115 mm
Ø 4 mm
Class 2 1)
3 s
Fast-action surface probe with
sprung thermocouple strip, also for
uneven surfaces, measurement
range short-term to +500°C, TC
Type K, Fixed cable
115 mm
Ø 5 mm
Ø 12 mm
0 to +300 °C
Class 2 1)
5 s
Fast-reaction paddle surface
probe, for measurements in
inaccessible places, e.g. narrow
apertures and slots, TC Type K,
Fixed cable
0602 0193
40 mm
145 mm
Ø 8 mm
-60 to +1000 °C
-60 to +300 °C
-50 to +250 °C
Class 1 1)
Class 2 1)
Class 2 1)
20 s
Efficient, waterproof surface probe
with small measurement head for
flat surfaces,
0602 0693
0602 0993
0602 2394
150 mm
Ø 2.5 mm
Ø 4 mm
TC Type K, Fixed cable
3 s
Fast-action surface probe with
sprung thermocouple strip, bent,
also for uneven surfaces,
measurement range short-term to
+500°C, TC Type K, Fixed cable
80 mm
Ø 5 mm
Ø 12 mm
3 s
Flat head surface probe with
telescopic handle max. 680 mm for
measurements at hard-to-access
points, TC Type K, Fixed cable 1.6
m (correspondingly shorter when
telescope extended)
985 5 mm
12 mm
Ø 25 mm
-50 to +170 °C
-50 to +400 °C
-60 to +400 °C
Class 2 1)
Class 2 1)
Class 2 1)
150 s
Magnetic probe, adhesive force
approx. 20 N, with magnets, for
measurements on metal surfaces,
TC Type K, Fixed cable
0602 4792
0602 4892
0602 1993
35 mm
Ø 20 mm
Magnetic probe, adhesive force
approx. 10 N, with magnets, for
higher temp., for measurements on
metal surfaces, TC Type K, Fixed
cable
75 mm
Ø 21 mm
30 s
Waterproof surface probe with
widened measurement tip for flat
surfaces, T/C Type K, Fixed cable
115 mm
Ø 5 mm
Ø 6 mm
1)
According to standard EN 60584-2, the accuracy of Class 1 refers to -40 to +1000 °C (Type K), Class 2 to -40 to +1200 °C (Type K), Class 3 to -200 to +40 °C (Type K).
A probe always corresponds to only one accuracy class.
Information on surface measurement:
• The response times t99 stated are measured on ground steel or aluminium plates at +60 °C.
• The stated accuracies are sensor accuracies.
• The accuracy in your application is dependent on the surface structure (roughness), material of the measurement object (heat capacity and heat transfer), as wel as sensor accuracy.
Testo creates a corresponding calibration certificate for the deviations of your measurement system in your application. For this purpose, Testo uses a surface test bench developed in
cooperation with the PTB (Physikalisch Technische Bundesanstalt).
testo 480
We measure it.
Probes
Dimensions
Probe shaft/probe shaft tip
Measuring
range
Accuracy
t99
Part no.
Probe type
Analog temperature probe
-50 to +120 °C
-60 to +130 °C
-60 to +130 °C
-50 to +100 °C
-60 to +1000 °C
-60 to +800 °C
Class 1 1)
Class 2 1)
Class 2 1)
Class 2 1)
Class 1 1)
Class 1 1)
Class 1 1)
Class 3 1)
Class 1 1)
Class 2 1)
Class 1 1)
90 s
5 s
5 s
5 s
2 s
3 s
5 s
5 s
4 s
7 s
1 s
Pipe wrap probe with velcro strip;
for temperature measurement on
pipes with diameter up to max. 120
mm; Tmax. +120 °C; TC Type K ,
Fixed cable
0628 0020
0602 4592
0602 0092
0602 4692
0602 0593
0602 2693
0602 5792
0602 5793
0602 5693
0602 1293
0602 0493
395 mm
20 mm
Pipe wrap probe for pipe diameter
5 to 65 mm, with exchangeable
measuring head. Meas. range
short-term up to +280°C, TC Type
K, Fixed cable
Spare meas. head for pipe wrap
probe, TC Type K
35 mm
15 mm
Clamp probe for measurements on
pipes, pipe diameter 15 to 25 mm
(max. 1"), meas. range short-term
up to +130°C, TC Type K, Fixed
cable
Efficient and fast-action immersion
probe, waterproof, TC Type K,
Fixed cable
Ø 1.5 mm
300 mm
Fast-action, waterproof
immersion/penetration probe, TC
Type K, Fixed cable
60 mm
14 mm
Ø 5 mm
Ø 1.5 mm
-200 to
+1000 °C
Immersion tip, flexible, TC Type K
Immersion tip, flexible, TC Type K
Ø 1.5 mm
Ø 1.5 mm
Ø 3 mm
500 mm
500 mm
1000 mm
-200 to +40 °C
-200 to
+1300 °C
Immersion measurement tip,
flexible, for measurements in
air/exhaust gases (not suitable for
measurements in smelters), TC
Type K
-60 to +400 °C
Waterproof immersion/penetration
probe, TC Type K, Fixed cable
114 mm
Ø 5 mm
50 mm
Ø 3.7 mm
-200 to
+1000 °C
Flexible, low-mass immersion
measurement tip, ideal for
Ø 0.25 mm
500 mm
measurements in small volumes
such as petri dishes, or for surface
measurements (e.g. attached with
adhesive tape), TC Type K, 2 m,
FEP insulated thermal wire,
temperature proof up to 200 °C,
oval wire with dimensions: 2.2 mm
x 1.4 mm
-60 to +400 °C
Class 2 1)
7 s
Waterproof food probe made of
stainless steel (IP65), TC Type K,
Fixed cable
0602 2292
125 mm
Ø 4 mm
30 mm
Ø 3.2 mm
1)
According to standard EN 60584-2, the accuracy of Class 1 refers to -40 to +1000 °C (Type K), Class 2 to -40 to +1200 °C (Type K), Class 3 to -200 to +40 °C (Type K).
A probe always corresponds to only one accuracy class.
testo 480
We measure it.
Probes
Dimensions
Probe shaft/probe shaft tip
Measuring
range
Accuracy
t99
Part no.
Probe type
Thermocouples
-50 to +400 °C
-50 to +400 °C
-50 to +250 °C
Class 2 1)
Class 2 1)
Class 2 1)
5 s
5 s
5 s
Thermocouple with TC adapter,
flexible, 800 mm long, fibre glass,
TC Type K
0602 0644
0602 0645
0602 0646
800 mm
Ø 1.5 mm
Thermocouple with TC adapter,
flexible, length 1500 mm ,
fibreglass, TC Type K
1500 mm
Ø 1.5 mm
Thermocouple with TC adapter,
flexible, 1500 mm long, PTFE, TC
Type K
1500 mm
Ø 1.5 mm
1)
According to standard EN 60584-2, the accuracy of Class 1 refers to -40 to +1000 °C (Type K), Class 2 to -40 to +1200 °C (Type K), Class 3 to -200 to +40 °C (Type K).
A probe always corresponds to only one accuracy class.
Pitot tubes
Pitot tube, 500 mm long, Ø 7
mm, stainless steel, for
Measuring range: 1 to 100 m/s
Operating temperature: 0 to +600 °C
Pitot tube factor: 1.0
0635 2045
0635 2145
0635 2345
500 mm
350 mm
1000 mm
Ø 7 mm
Ø 7 mm
Ø 7 mm
measuring flow velocity*
Pitot tube, 350 mm long, Ø 7
mm, stainless steel, for
measuring flow velocity*
Measuring range: 1 to 100 m/s
Operating temperature: 0 to +600 °C
Pitot tube factor: 1.0
Pitot tube, 1000 mm long,
stainless steel, for measuring
flow velocity*
Measuring range: 1 to 100 m/s
Operating temperature: 0 to +600 °C
Pitot tube factor: 1.0
Straight pitot tube with integrated
temperature measurement, incl.
connection hose, length 360 mm
Measuring range: 1 to 30 m/s
Operating temperature: 0 to +600 °C
Pitot tube factor: 0.67
0635 2043
0635 2143
0635 2243
360 mm
500 mm
1000 mm
Minimum immersion depth: 150 mm
Straight pitot tube with integrated
temperature measurement, incl.
connection hose, length 500 mm
Measuring range: 1 to 30 m/s
Operating temperature: 0 to +600 °C
Pitot tube factor: 0.67
Minimum immersion depth: 150 mm
Straight pitot tube with integrated
temperature measurement, incl.
connection hose, length 1000
mm
Measuring range: 1 to 30 m/s
Operating temperature: 0 to +600 °C
Pitot tube factor: 0.67
Minimum immersion depth: 150 mm
*Connection hose required (order no. 0554 0440) or (order no. 0554 0453)
testo 480
We measure it.
Accessories
Accessories for measuring instrument
Part no.
Telescope for digital probes, with ball joint and probe bracket, length 1.8
m. Use 5 m plug-in head cable (order no. 0430 0101).
0430 0946
Tripod for workplace evaluation With holders for hand-held instrument and probe. Can also be used as telescope extension
Plug-in head cable for digital probes
0554 0743
0430 0100
0430 0101
0554 0410
0554 0415
0563 4170
Plug-in head cable for digital probes, length 5 m
testovent 410, volume flow funnel, Ø 340 mm/330x330 mm, incl. case
testovent 415, volume flow funnel, Ø 210 mm/210x210 mm, incl. case
testovent 417 funnel set for plate outlets (Ø 200 mm) and funnel for ventilator (330 x 330 mm) for ingoing and
outgoing air
Flow straightener testovent 417
0554 4172
0554 0660
Control and adjustment set for Testo humidity probes, salt solution with 11.3% RH and 75.3% RH, incl. adapter for
Testo humidity probes
Connection hose; silicone; 5 m long; max. load 700 hPa (mbar)
0554 0440
0554 0453
Connection hose silicone-free for differential pressure measurement, length 5 m, load up to maximum 700 hPa, (mbar)
Transport and Protection
Soft case testo 480 incl. carrying strap
0516 0481
0516 4801
0516 4800
System case for comfort level measurement For instrument, probes and other accessories
System case for HVAC measurements, For instrument, probes and other accessories
Printer and Accessories
Testo fast printer IrDA with wireless infrared interface; 1 roll thermal paper; 4 AA batteries for printing out
measurements on site
0554 0549
0554 0568
Spare thermal paper for printer (6 rolls), permanent ink, measurement data documentation legible for up to 10 years
Calibration Certificates
ISO calibration certificate/temperature for air/immersion probes, calibration points -18°C; 0°C; +60°C
DAkkS calibration certificate/temperature; meas. instr. with air/immersion probe; calibration points -20 °C; 0 °C; +60 °C
ISO calibration certificate humidity; Calibration points 11.3 %RH and 75.3 %RH at +25 °C
DAkkS calibration certificate/humidity; electronic hygrometers; calibration points 11.3%RH and 75.3%RH at +25°C
ISO calibration certificate pressure; accuracy 0.1 to 0.6 (% of fsv), 5 points distributed over meas. range
ISO calibration certificate pressure; accuracy > 0.6 (% of fsv)
0520 0001
0520 0211
0520 0006
0520 0206
0520 0025
0520 0005
0520 0004
0520 0034
0520 0010
0520 0033
ISO calibration certificate velocity; hot wire, vane anemometer, Pitot tube; calibration points 1; 2; 5; 10 m/s
ISO calibration certificate velocity; hot wire, vane anemometer, Pitot tube; calibration points 5; 10; 15; 20 m/s
ISO calibration certificate/light; Calibration points 0; 500; 1000; 2000; 4000 Lux
ISO calibration certificate/CO2; CO2 probes; calibration points 0; 1000; 5000 ppm
More calibration certificates on request
www.testo.com
相关型号:
©2020 ICPDF网 联系我们和版权申明