TPMBU30-106 [TOTAL-POWER]
30W Open Frame Type Medical Power Supplies;型号: | TPMBU30-106 |
厂家: | TOTAL POWER INTERNATIONAL |
描述: | 30W Open Frame Type Medical Power Supplies |
文件: | 总2页 (文件大小:259K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
Total Power International, Inc.
30W Open Frame Type
Medical Power Supplies
TPMBU30 SERIES
Features:
Wide Operating Voltage 90 to 264 VAC,47 to 63 Hz
Internal EMI filter
Single Output
Over Voltage and Over Load protection
Output Voltage Protection (Crowbar Design)
Size: 2"x4"
Class I
3 year warranty
Electrical Characteristics:
Safety Approvals Input Voltage Range
100~240VAC
Vin
Application:
Operate Voltage Range
90~264VAC
47~63Hz
- Blood Pressure system
- Portable medical devices
- ECG machine
Input Frequency
fin
Output Power Range
Output Voltage Range
Output Current Range
Input Current (Low Line)
Input Current (High Line)
See rating chart
See rating chart
See rating chart
Po
Vo
Io
Io=Full load, Vin=100VAC
Io=Full load, Vin=240VAC
Io=Full load, 25°C, Cool start, Vin=115VAC
Io=Full load, 25°C, Cool start, Vin=230VAC
Io=Full Load, Vin=230VAC
Io=Full Load
0.7A
Iil
0.3A
Iih
15A (max.)
30A (max.)
Low Line Inrush Current
High Line Inrush Current
Efficiency
Ir
Safety Approvals:
76.4~85%
1% (max.)
5% (max.)
Eff
Line Regulation
REG-i
REG-o
OVP
RoHS
COMPLIANT
EN 60601-1
IEC 60601-1
2002/95/EC
Load Regulation
Vin=230VAC
Over Voltage Protection
112~132%
110~150%
4mS (max.)
10mS (min.)
OCP Over Current Protection
Time of Transient Response
Hold-Up Time
Io=Full Load to Half Load, Vin=100VAC
Io=Full Load, Vin=110VAC
Io=Full Load, Vin=100VAC
Full Load, Vin=90VAC
Ttr
Th
Ts
Start Up Time
0.3~2S
Vp-p Ripple & Noise(Peak to Peak)
1% (max.)
0.1mA (max.)
Safety Ground Leakage Current
Ilk
Vin=240VAC/60Hz
All output
Temperature Coefficient
TC
0.04%/°C
Operating Altitude
Up to 3000m
See rating chart
5831VDC (min.)
OA
Pno
Vps
No-Load Power Consumption
Dielectric Withstanding Voltage
No load, Vin=230VAC
Primary to secondary
Dielectric Withstanding Voltage
for Primary to PE
Vpg
Ris
Primary to PE
2121VDC (min.)
Test Voltage=500VDC
(min.)
50M
Isolation Resistance
Recommend to be used on the metal chassis.
Environmental
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
See derating curve
-40~85°C
To
Ts
0~95%
Ho
0~95%
Hr
Operating Temperature at 25°C, Calculated per MIL-HDBK-217F
0.1M Hrs (min.)
MTBF
Pd
Derate linearly from 100% load at 50°C to 50% load at 70°C
Total Power International, Inc
Toll Free: 877-646-0900
www.total-power.com
Total Power International, Inc.
30W Open Frame Type
Medical Power Supplies
TPMBU30 SERIES
Output Voltage And Current Rating Chart ( Single Output ) :
Total
Regulation
Max.
Output Voltage
Output Current
Pno (max.)
Output Power
Model Number
5 VDC
7 VDC
6.00 A
4.28 A
3.33 A
2.50 A
2.00 A
1.66 A
1.25 A
1.00 A
0.83 A
5%
5%
5%
5%
5%
5%
3%
3%
3%
30W
30W
30W
30W
30W
30W
30W
30W
30W
0.5W
0.5W
0.5W
0.5W
0.5W
0.5W
0.5W
0.5W
0.5W
TPMBU30-102
TPMBU30-103
TPMBU30-104
TPMBU30-105
TPMBU30-106
TPMBU30-107
TPMBU30-108
MTPBU30-109
TPMBU30-110
9 VDC
12 VDC
15 VDC
18 VDC
24 VDC
30 VDC
36 VDC
PIN CHART
Mechanical Specifications:
PIN
1
2
3
4
5
6
MODEL
RTN
RTN
RTN
Vout
Vout
Vout
TPMBU30-1XX
Note:
1. Dimensions are shown in inches or mm.
2. Weight: 200gs approx.
3. Input connector mates with Molex housing
09-50-3031and Molex 2478 series crimp terminal.
4. Output connector mates with Molex housing
09-50-3061 and Molex 2478 series crimp terminal.
Derating Curve :
120%
1. Operating Temperature: 0 to 70°C
2. Derate linearly from 100% load at 50°C to 50%
load at 70°C
100%
80%
60%
40%
20%
0%
0
10
20
30
40
50
60
70
80
Temperature (deg c)
Total Power International, Inc
Toll Free: 877-646-0900
www.total-power.com