THM15-1211 [TRACOPOWER]

DC/DC Converter;
THM15-1211
型号: THM15-1211
厂家: TRACO ELECTRONIC AG    TRACO ELECTRONIC AG
描述:

DC/DC Converter

文件: 总4页 (文件大小:292K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
DC/DC Converter  
THM 15 Series, 15 Watt  
Wide 2:1 input voltage 15 W DC/DC  
converter in a 1.6 x 1 ” plastic case  
I/O isolation 5000 VACrms rated for  
250 VACrms working voltage  
Certification according to IEC/EN/ES  
60601-1 3rd edition for 2xMOPP  
Risk management process according to  
ISO 14971 including risk management file  
Acceptance criteria for electronic  
CB  
Scheme  
assemblies according to IPC-A-610 Level 3  
ES 60601-1  
IEC 60601-1  
Low leakage current < 2.5 µA  
The THM 15 series is a range of medical 15 Watt DC/DC converters in 1.6” x 1.0”  
plastic package and with wide 2:1 input voltage range. They provide a reinforced  
isolation system for 5000 VACrms isolation and a very low leakage current of  
less than 2.5 µA. The units are approved to IEC/EN/ES 60601-1 3rd edition for  
2 ꢀ MOPP (Means Of Patient Protection) and come along with an ISO 14971 risk  
management file. Design and production conform to the quality management  
system ISO 13485. With a high efficiency of up to 90% and highest grade com-  
ponents the converters can reliably operate in an ambient temperature range  
of –40°C up to +85°C. They constitute a reliable solution not only for medical  
equipment but also for demanding ranges of application such as transportation,  
control & measurement or IGBT drivers.  
Eꢀtended operating temperature range  
–40°C to 85°C.  
EMC compliance to IEC 60601-1-2 4th  
edition and EN55032 class A  
Operating up to 5000m altitude  
5 year product warranty  
Models  
Order code  
THM 15-1211  
THM 15-1212  
THM 15-1213  
THM 15-1215  
THM 15-1221  
THM 15-1222  
THM 15-1223  
THM 15-2411  
THM 15-2412  
THM 15-2413  
THM 15-2415  
THM 15-2421  
THM 15-2422  
THM 15-2423  
THM 15-4811  
THM 15-4812  
THM 15-4813  
THM 15-4815  
THM 15-4821  
THM 15-4822  
THM 15-4823  
Input voltage range  
Output voltage  
5.0 VDC  
12 VDC  
15 VDC  
24 VDC  
5 VDC  
Output current maꢀ.  
3000 mA  
1250 mA  
1000 mA  
625 mA  
Efficiency typ.  
87.0 %  
88.5 %  
88.5 %  
89.0 %  
86.0 %  
89.0 %  
89.0 %  
90.0 %  
90.0 %  
90.0 %  
90.0 %  
86.0 %  
90.0 %  
90.0 %  
89.5 %  
88.0 %  
88.0 %  
88.5 %  
86.0 %  
88.5 %  
88.0 %  
9.0 – 18 VDC  
(12 VDC nominal)  
1500 mA  
625 mA  
12 VDC  
15 VDC  
5.0 VDC  
12 VDC  
15 VDC  
24 VDC  
5 VDC  
500 mA  
3000 mA  
1250 mA  
1000 mA  
625 mA  
18 – 36 VDC  
(24 VDC nominal)  
1500 mA  
625 mA  
12 VDC  
15 VDC  
5.0 VDC  
12 VDC  
15 VDC  
24 VDC  
5 VDC  
500 mA  
3000 mA  
1250 mA  
1000 mA  
625 mA  
36 – 75 VDC  
(48 VDC nominal)  
1500 mA  
625 mA  
12 VDC  
15 VDC  
500 mA  
http://www.tracopower.com  
Page 1 of 4  
THM 15 Series, 15 Watt  
Input Specifications  
Input current no load  
12 Vin models: 18 mA typ.  
24 Vin models: 13 mA typ.  
48 Vin models: 10 mA typ.  
Surge voltage (3 s max.)  
12 Vin models: 25 V max.  
24 Vin models: 50 V max.  
48 Vin models: 100 V max.  
Start-up voltage  
12 Vin models: 9 VDC (or lower)  
24 Vin models: 18 VDC (or lower)  
48 Vin models: 36 VDC (or lower)  
Startup time  
60 ms max. (30 ms typ.)  
Under voltage shut down (lock-out circuit)  
12 Vin models: 7.8 - 8.6 VDC  
24 Vin models: 15.8 - 17.4 VDC  
48 Vin models: 32 - 34 VDC  
Input filter  
Pi-type  
Conducted noise  
– Conducted & Radiated input surpression  
EN 55011 limits to IEC 60601-1-2 4th editon  
EN55032 class A (internal filter)  
EN55032 class B with external components  
www.tracopower.com/overview/thm15  
– Filter proposal  
EMC immunity  
– Generic for Medical equipment  
– ESD (electrostatic discharge)  
IEC/EN 60601-1-2 4th edition  
EN 61000-4-2, air 15 kV, contact 8 kV,  
perf. criteria A  
– Radiated immunity  
– Fast transient / surge  
(with external input capacitor / diode)  
EN 61000-4-3, 10 V/m, perf. criteria A  
EN 61000-4-4, 2 kV, perf. criteria A  
EN 61000-4-5, 2 kV perf. criteria A  
12 Vin models: Nippon chemi-con KY 220 µF/ 100 V  
TVS - SMDJ36A, 36V, 3000 W)  
24 Vin models: Nippon chemi-con KY 220 µF/ 100 V  
TVS - SMDJ58A, 58V, 3000 W)  
48 Vin models: Nippon chemi-con KY 220 µF/ 100 V  
TVS - SMDJ120A, 120V, 3000 W)  
– Conducted immunity  
– Magnetic field immunity  
EN 61000-4-6, 10 Vrms, perf. criteria A  
EN 61000-4-8  
100 A/m, continuous, perf. criteria A  
1000 A/m, 1 sec., perf. criteria A  
External input fuse required  
(recommended values, slow blow type)  
12 Vin models: 3.15 A  
24 Vin models: 1.6 A  
48 Vin models: 0.8 A  
Output Specifications  
Voltage set accuracy  
1% max.  
Output voltage adjustment range  
(single output models only)  
5 & 12 VDC models:  
15 & 24 VDC models: –10 /+20%  
10 %  
Regulation  
– Input variation  
single output: 0.2% max.  
dual output: 0.5% max.  
single output: 0.2% max.  
dual output: 1.0% max.  
– Load variation 0 – 100 %  
– Cross regulation  
dual output: 5.0% max. (asymmetrical load 25/100%)  
Temperature coefficient  
Minimum load  
0.02 %/K typ.  
not required  
Ripple and noise  
(20 MHz Bandwidth)  
( )5.0 VDC models: 50 mVp-p typ. with cap. 10µF/25V X7R MLCC  
( )12 VDC models: 75 mVp-p typ. with cap. 10µF/25V X7R MLCC  
( )15 VDC models: 75 mVp-p typ. with cap. 10µF/25V X7R MLCC  
24 VDC models: 100 mVp-p typ. with cap. 4.7µF/50V X7R MLCC  
Transient response  
– Recovery time (25% load step change)  
250 µs typ.  
http://www.tracopower.com  
Page 2 of 4  
THM 15 Series, 15 Watt  
General Specifications  
Overload protection  
at 150 % typ. of Iout rated (hiccup mode)  
Short-circuit protection  
Overvoltage protection  
Continuous, automatic recovery  
( )5.0 VDC models: 6.2 VDC typ.  
( )12 VDC models: 15 VDC typ.  
( )15 VDC models: 20 VDC typ.  
24 VDC models: 30 VDC typ.  
Capacitive load  
–Single output  
5.0 VDC models: 3‘800 µF max.  
12 VDC models: 650 µF max.  
15 VDC models: 530 µF max.  
24 VDC models: 190 µF max.  
–Dual output  
5 VDC models: 1‘900 µF max. (each output)  
12 VDC models: 380 µF max. (each output)  
15 VDC models: 270 µF max. (each output)  
Temperature ranges  
– Operating  
– Case temperature  
– Storage temperature  
–40°C to +85°C  
+105°C max.  
–55°C to +125°C  
Derating  
2.86%/K above 65°C  
at 115°C typ.  
Overtemperature protection  
Thermal impedance  
15.30 K/W  
Humidity (non condensing)  
Isolation voltage (50 Hz, 60 s)  
Clearance/creepage  
5 % to 95 % rel H max.  
5000 VACrms, reinforced  
8 mm min.  
Leakage current (at 240VAC, 60Hz)  
Isolation capacitance (input/output)  
Altitude during operation  
2.5µA max.  
20 pF typ.  
5000 m  
Reliability, calculated MTBF (MIL-HDBK-217F at +25°C, ground benign)  
Switching frequency  
tbd  
250 kHz typ. (pulse width modulation)  
according to MIL-STD-810F  
Vibration and thermal shock resistance  
Safety standards/approvals – Medical equipment  
ANSI/AAMI ES 60601-1:2005/(R)2012,  
IEC/EN 60601-1 3rd edition  
– Certification documents  
www.tracopower.com/overview/thm15wi  
Environmental compliance – Reach  
www.tracopower.com/products/reach-declaration.pdf  
RoHS directive 2011/65/EU  
– RoHS  
Physical Specifications  
Casing material  
non-conductive plastic  
non-conductive plastic  
silicone (UL94 V-0 rated)  
24 g (0.85oz)  
Base material  
Potting material  
Package weight  
Soldering temperature  
max. 265°C / 10 s  
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.  
http://www.tracopower.com  
Page 3 of 4  
THM 15 Series, 15 Watt  
Outline Dimensions  
25.4 (1.00)  
Pinout  
10.2  
(0.40)  
5.6  
(0.22)  
5.08 5.08  
Pin  
1
2
Single  
+Vin (Vcc)  
–Vin (GND)  
+Vout  
Dual  
+Vin (Vcc)  
–Vin (GND)  
+Vout  
(0.20) (0.20)  
3
4
5
–Vout  
Trim  
Commom  
–Vout  
2
1
Dimensions in [mm], () = Inch  
Tolerances 0.5 ( 0.02)  
0.25 ( 0.01)  
5
4
3
Pin pich tolerances 0.25 ( 0.01)  
Pin ø 1.0 0.1 (0.04 0.004)  
2.54  
(0.10)  
10.16  
(0.40)  
10.16  
(0.40)  
©
Copyright 2018 Traco Electronic AG  
Specifications can be changed without notice!  
Rev. January 8, 2018  
Page 4 of 4  
www.tracopower.com  

相关型号:

THM15-1212

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

THM15-1213

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

THM15-1215

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

THM15-1221

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

THM15-1222

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

THM15-1223

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

THM15-2411

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

THM15-2411WI

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

THM15-2412

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

THM15-2412WI

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

THM15-2413

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

THM15-2413WI

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER