TIM2-0922SM [TRACOPOWER]

DC/DC Converter;
TIM2-0922SM
型号: TIM2-0922SM
厂家: TRACO ELECTRONIC AG    TRACO ELECTRONIC AG
描述:

DC/DC Converter

文件: 总4页 (文件大小:266K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
DC/DC Converter  
TIM 2SM Series, 2 Watt  
Compact SMD-16-package  
I/O isolation 5000 VACrms rated for  
250 VACrms working voltage  
Certification according to IEC/EN/ES  
60601-1 3rd edition for 2ꢀMOPP  
and opperation to 5000 m altitude  
Low leakage current < 2 µA for  
BF-applications  
Extended operating temperature range  
–40°C to 95°C.  
The TIM 2SM series is a range of 2 Watt DC/DC converters in compact SMD  
package and with reinforced isolation of 5000 VACrms for medical applications.  
With a low leakage current of less than 2 µA the converters are predestined to  
insulate electrical equipment from the applied parts to patient (BF classification).  
The models are approved to IEC/EN/ES 60601-1 3rd edition for 2xMOPP up to  
an altitude of 5000m and come along with an ISO 14971 risk management file.  
EMC compliance to IEC 60601-1-2 4th  
edition and EN55032 class A  
5-year product warranty  
Models  
Order code  
Input voltage range  
Output voltage  
3.3 VDC  
5.0 VDC  
9.0 VDC  
12 VDC  
15 VDC  
24 VDC  
12 VDC  
15 VDC  
3.3 VDC  
5.0 VDC  
9.0 VDC  
12 VDC  
15 VDC  
24 VDC  
12 VDC  
15 VDC  
3.3 VDC  
5.0 VDC  
9.0 VDC  
12 VDC  
15 VDC  
24 VDC  
12 VDC  
15 VDC  
3.3 VDC  
5.0 VDC  
9.0 VDC  
12 VDC  
15 VDC  
24 VDC  
12 VDC  
15 VDC  
Output current max.  
500 mA  
400 mA  
222 mA  
167 mA  
134 mA  
83 mA  
Efficiency typ.  
74 %  
78 %  
79 %  
81 %  
81 %  
81 %  
81 %  
81 %  
75 %  
78 %  
78 %  
81 %  
81 %  
81 %  
81 %  
82 %  
75 %  
78 %  
78 %  
80 %  
81 %  
81 %  
81 %  
82 %  
75 %  
77 %  
78 %  
80 %  
81 %  
81 %  
81 %  
81 %  
TIM 2-0910SM  
TIM 2-0911SM  
TIM 2-0919SM  
TIM 2-0912SM  
TIM 2-0913SM  
TIM 2-0915SM  
TIM 2-0922SM  
TIM 2-0923SM  
TIM 2-1210SM  
TIM 2-1211SM  
TIM 2-1219SM  
TIM 2-1212SM  
TIM 2-1213SM  
TIM 2-1215SM  
TIM 2-1222SM  
TIM 2-1223SM  
TIM 2-2410SM  
TIM 2-2411SM  
TIM 2-2419SM  
TIM 2-2412SM  
TIM 2-2413SM  
TIM 2-2415SM  
TIM 2-2422SM  
TIM 2-2423SM  
TIM 2-4810SM  
TIM 2-4811SM  
TIM 2-4819SM  
TIM 2-4812SM  
TIM 2-4813SM  
TIM 2-4815SM  
TIM 2-4822SM  
TIM 2-4823SM  
4.5 – 12 VDC  
(9 VDC nominal)  
83 mA  
67 mA  
500 mA  
400 mA  
222 mA  
167 mA  
134 mA  
83 mA  
9.0 – 18 VDC  
(12 VDC nominal)  
83 mA  
67 mA  
500 mA  
400 mA  
222 mA  
167 mA  
134 mA  
83 mA  
18 – 36 VDC  
(24 VDC nominal)  
83 mA  
67 mA  
500 mA  
400 mA  
222 mA  
167 mA  
134 mA  
83 mA  
36 – 75 VDC  
(48 VDC nominal)  
83 mA  
67 mA  
www.tracopower.com  
Page 1 of 4  
TIM 2SM Series, 2 Watt  
Input Specifications  
Input current no load  
9 Vin models: 90 mA typ.  
12 Vin models: 45 mA typ.  
24 Vin models: 25 mA typ.  
48 Vin models: 12 mA typ.  
Surge voltage (1 s max.)  
9 Vin models: 15 V maꢀ.  
12 Vin models: 25 V maꢀ.  
24 Vin models: 50 V maꢀ.  
48 Vin models: 100 V maꢀ.  
Start-up voltage  
9 Vin models: 4.5 VDC (or lower)  
12 Vin models: 9.0 VDC (or lower)  
24 Vin models: 18 VDC (or lower)  
48 Vin models: 36 VDC (or lower)  
Startup time  
10 ms typ. / 20 ms maꢀ.  
Under voltage shut down  
9 Vin models: 2 - 4 VDC  
12 Vin models: 6 - 8 VDC  
24 Vin models: 13 - 17 VDC  
48 Vin models: 29 - 35 VDC  
Input filter  
capacitor type  
Conducted noise  
– Conducted & Radiated input suppression  
Application note for filter class A/B proposal  
EN 55011 limits to IEC/EN 60601-1-2 4th edit.  
EN 55032 class A or B with external components  
www.tracopower.com/overview/tim2sm  
EMC immunity  
– Generic for Medical equipment  
– ESD (electrostatic discharge)  
IEC/EN 60601-1-2 4th edition  
EN 61000-4-2, air 15 kV, contact 8 kV,  
perf. criteria A  
– Radiated immunity  
– Fast transient / surge  
(with external input capacitor / diode)  
EN 61000-4-3, 10 V/m, perf. criteria A  
EN 61000-4-4, 2 kV, perf. criteria A  
EN 61000-4-5, 1 kV perf. criteria A  
9 Vin models: Nippon chemi-con KY 1000 µF / 25 V  
TVS - SMAJ18A, 18 V, 400 W  
12 & 24 Vin models: Nippon chemi-con KY 470 µF / 50 V  
48 Vin models: Nippon chemi-con KY 220 µF / 100 V  
EN 61000-4-6, 10 Vrms, perf. criteria A  
EN 61000-4-8  
– Conducted immunity  
– Magnetic field immunity  
100 A/m, continuous, perf. criteria A  
1000 A/m, 1 sec., perf. criteria A  
Eꢀternal input fuse required  
9 Vin models: 1.0 A (slow blow)  
12 Vin models: 0.5 A (slow blow)  
24 Vin models: 0.315 A (slow blow)  
48 Vin models: 0.16 A (slow blow)  
Output Specifications  
Voltage set accuracy  
Regulation  
1% maꢀ.  
– Input variation (Vin min. to Vin max.)  
– Load variation (0 – 100 %)  
0.2% maꢀ.  
1% maꢀ.  
– Load variation (10 – 90 %)  
single output: 0.5% maꢀ.  
dual output: 0.8% maꢀ.  
– Cross regulation  
dual output: 5.0% maꢀ. (asymmetrical load 25 / 100%)  
Minimum load  
not required  
Ripple and noise (20 MHz Bandwidth)  
50 mVp-p typ.  
Transient response (25% load step change)  
– Recovery time  
500 µs typ.  
Short circuit protection  
Continuous, automatic recovery  
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.  
www.tracopower.com  
Page 2 of 4  
TIM 2SM Series, 2 Watt  
General Specifications  
Overvoltage protection  
3.3 Vout models: 4 - 6.5 VDC  
5.0 Vout models: 6 - 8 VDC  
9.0 Vout models: 10 - 14 VDC  
12 Vout models: 13 - 19 VDC  
15 Vout models: 16 - 22 VDC  
24 Vout models: 25 - 35 VDC  
Capacitive load  
– Single output  
3.3 & 5.0 Vout models: 1‘000 µF maꢀ.  
9.0 Vout models: 430 µF maꢀ.  
12 Vout models: 220 µF maꢀ.  
15 Vout models: 170 µF maꢀ.  
24 Vout models: 100 µF maꢀ.  
12 Vout models: 170 µF maꢀ. (each output)  
15 Vout models: 100 µF maꢀ. (each output)  
– Dual output  
Temperature ranges  
– Operating (natural convection: 20 LFM, 0.1 m/s)  
– Case temperature  
–40°C to +95°C  
+105°C maꢀ.  
– Storage temperature  
–55°C to +125°C  
Derating  
6.7 %/K above 90°C  
5 % to 95 % rel H maꢀ.  
IPC J-STD-033C Level 2a  
5000 VAC (reinforced insulation)  
250 VAC, 2 x MOPP  
20 pF maꢀ.  
Humidity (non condensing)  
Moisture sensitivity level (MSL)  
Isolation voltage (50 Hz, 60 s)  
Working voltage  
Isolation capacitance  
Clearance/creepage  
8 mm min.  
Leakage current (at 240 VAC, 60 Hz)  
Altitude during operation  
2 µA maꢀ.  
5000 m maꢀ.  
Temperature coefficient  
0.02 %/K maꢀ.  
Reliability, calculated MTBF (MIL-HDBK-217F at +25°C, ground benign)  
Switching frequency  
6‘809‘000 h  
100 kHz min. (frequency modulated)  
according to MIL-STD-810F  
Shock, vibration and thermal shock resistance  
Remote On/Off  
– On:  
– Off:  
– Off idle current:  
open circuit or high impedance  
2 – 4 mA current applied via 1kOhm resistor  
2.5 mA typ.  
Safety standards/approvals – Medical equipment  
ANSI/AAMI ES60601-1:2005/(R)2012,  
IEC/EN60601-1 3rd edition  
– Certification documents  
www.tracopower.com/overview/tim2sm  
Environmental compliance – Reach  
www.tracopower.com/info/reach-declaration.pdf  
RoHS directive 2011/65/EU  
– RoHS  
Physical Specifications  
Casing material  
non-conductive black plastic  
non-conductive black plastic  
silicone (UL94 V-0)  
Base material  
Potting material  
Package weight  
7.0 g (0.24 oz)  
Lead-free reflow solder process  
IPC J-STD-020E  
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.  
www.tracopower.com  
Page 3 of 4  
TIM 2SM Series, 2 Watt  
Outline Dimensions  
14.4 (0.57)  
1
2
16  
FRONT VIEW  
TOP VIEW  
7
8
10  
9
Standard Pinout  
18.1 (0.71)  
Pin  
1
2
Single  
–Vin (GND)  
On/Off  
NC  
Dual  
–Vin (GND)  
On/Off  
NC  
7
8
9
NC  
+Vout  
Common  
+Vout  
Dimensions in mm (inch)  
Tolerances ±±.5 (±±.±2)  
Pin dimension ±.5 ±±.1 (±.±2 ±±.±±4)  
Pin pitch tolerances ±±.25 (±±.±1)  
10  
16  
–Vout  
+Vin (Vcc)  
–Vout  
+Vin (Vcc)  
©
Copyright 2018 Traco Electronic AG  
Specifications can be changed without notice!  
Rev. July 26, 2018  
Page 4 of 4  
www.tracopower.com  

相关型号:

TIM2-0923

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

TIM2-0923SM

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

TIM2-1210

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

TIM2-1210SM

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

TIM2-1211

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

TIM2-1211SM

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

TIM2-1212

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

TIM2-1212SM

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

TIM2-1213

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

TIM2-1213SM

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

TIM2-1215

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER

TIM2-1215SM

DC/DC Converter

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TRACOPOWER