TIM2-4815 [TRACOPOWER]
DC/DC Converter;型号: | TIM2-4815 |
厂家: | TRACO ELECTRONIC AG |
描述: | DC/DC Converter |
文件: | 总4页 (文件大小:271K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
DC/DC Converter
TIM 2 Series, 2 Watt
• Compact DIP-16-package
• I/O isolation 5000 VACrms rated for
250 VACrms working voltage
• Certification according to IEC/EN/ES
60601-1 3rd edition for 2xMOPP
and opperation to 5000 m altitude
• Low leakage current < 2 µA for
BF-applications
• Eꢀtended operating temperature range
–40°C to 95°C.
The TIM 2 series is a range of 2 Watt DC/DC converters in compact DIP-16
package and with reinforced isolation of 5000 VACrms for medical applications.
With a low leakage current of less than 2 µA the converters are predestined to
insulate electrical equipment from the applied parts to patient (BF classification).
The models are approved to IEC/EN/ES 60601-1 3rd edition for 2xMOPP up to
an altitude of 5000m and come along with an ISO 14971 risk management file.
• EMC compliance to IEC 60601-1-2 4th
edition and EN55032 class A
• 5-year product warranty
Models
Order code
TIM 2-0910
TIM 2-0911
TIM 2-0919
TIM 2-0912
TIM 2-0913
TIM 2-0915
TIM 2-0922
TIM 2-0923
TIM 2-1210
TIM 2-1211
TIM 2-1219
TIM 2-1212
TIM 2-1213
TIM 2-1215
TIM 2-1222
TIM 2-1223
TIM 2-2410
TIM 2-2411
TIM 2-2419
TIM 2-2412
TIM 2-2413
TIM 2-2415
TIM 2-2422
TIM 2-2423
TIM 2-4810
TIM 2-4811
TIM 2-4819
TIM 2-4812
TIM 2-4813
TIM 2-4815
TIM 2-4822
TIM 2-4823
Input voltage range
Output voltage
3.3 VDC
5.0 VDC
9.0 VDC
12 VDC
15 VDC
24 VDC
12 VDC
15 VDC
3.3 VDC
5.0 VDC
9.0 VDC
12 VDC
15 VDC
24 VDC
12 VDC
15 VDC
3.3 VDC
5.0 VDC
9.0 VDC
12 VDC
15 VDC
24 VDC
12 VDC
15 VDC
3.3 VDC
5.0 VDC
9.0 VDC
12 VDC
15 VDC
24 VDC
12 VDC
15 VDC
Output current maꢀ.
500 mA
400 mA
222 mA
167 mA
134 mA
83 mA
Efficiency typ.
74 %
78 %
79 %
81 %
81 %
81 %
81 %
81 %
75 %
78 %
78 %
81 %
81 %
81 %
81 %
82 %
75 %
78 %
78 %
80 %
81 %
81 %
81 %
82 %
75 %
77 %
78 %
80 %
81 %
81 %
81 %
81 %
4.5 – 12 VDC
(9 VDC nominal)
83 mA
67 mA
500 mA
400 mA
222 mA
167 mA
134 mA
83 mA
9.0 – 18 VDC
(12 VDC nominal)
83 mA
67 mA
500 mA
400 mA
222 mA
167 mA
134 mA
83 mA
18 – 36 VDC
(24 VDC nominal)
83 mA
67 mA
500 mA
400 mA
222 mA
167 mA
134 mA
83 mA
36 – 75 VDC
(48 VDC nominal)
83 mA
67 mA
www.tracopower.com
Page 1 of 4
TIM 2 Series, 2 Watt
Input Specifications
Input current no load
9 Vin models: 90 mA typ.
12 Vin models: 45 mA typ.
24 Vin models: 25 mA typ.
48 Vin models: 12 mA typ.
Surge voltage (1 s max.)
9 Vin models: 15 V maꢀ.
12 Vin models: 25 V maꢀ.
24 Vin models: 50 V maꢀ.
48 Vin models: 100 V maꢀ.
Start-up voltage
9 Vin models: 4.5 VDC (or lower)
12 Vin models: 9.0 VDC (or lower)
24 Vin models: 18 VDC (or lower)
48 Vin models: 36 VDC (or lower)
Startup time
10 ms typ. / 20 ms maꢀ.
Under voltage shut down
9 Vin models: 2 - 4 VDC
12 Vin models: 6 - 8 VDC
24 Vin models: 13 - 17 VDC
48 Vin models: 29 - 35 VDC
Input filter
capacitor type
Conducted noise
– Conducted & Radiated input suppression
Application note for filter class A/B proposal
EN 55011 limits to IEC/EN 60601-1-2 4th edit.
EN 55032 class A or B with external components
www.tracopower.com/overview/tim2
–
EMC immunity
– Generic for Medical equipment
– ESD (electrostatic discharge)
IEC/EN 60601-1-2 4th edition
EN 61000-4-2, air 15 kV, contact 8 kV,
perf. criteria A
– Radiated immunity
– Fast transient / surge
(with external input capacitor / diode)
EN 61000-4-3, 10 V/m, perf. criteria A
EN 61000-4-4, 2 kV, perf. criteria A
EN 61000-4-5, 1 kV perf. criteria A
9 Vin models: Nippon chemi-con KY 1000 µF / 25 V
TVS - SMAJ18A, 18 V, 400 W
12 & 24 Vin models: Nippon chemi-con KY 470 µF / 50 V
48 Vin models: Nippon chemi-con KY 220 µF / 100 V
EN 61000-4-6, 10 Vrms, perf. criteria A
EN 61000-4-8
– Conducted immunity
– Magnetic field immunity
100 A/m, continuous, perf. criteria A
1000 A/m, 1 sec., perf. criteria A
Eꢀternal input fuse required
9 Vin models: 1.0 A (slow blow)
12 Vin models: 0.5 A (slow blow)
24 Vin models: 0.315 A (slow blow)
48 Vin models: 0.16 A (slow blow)
Output Specifications
Voltage set accuracy
Regulation
1% maꢀ.
– Input variation (Vin min. to Vin max.)
– Load variation (0 – 100 %)
0.2% maꢀ.
1% maꢀ.
– Load variation (10 – 90 %)
single output: 0.5% maꢀ.
dual output: 0.8% maꢀ.
– Cross regulation
dual output: 5.0% maꢀ. (asymmetrical load 25 / 100%)
Minimum load
not required
Ripple and noise (20 MHz Bandwidth)
50 mVp-p typ.
Transient response (25% load step change)
– Recovery time
500 µs typ.
Short circuit protection
Continuous, automatic recovery
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
www.tracopower.com
Page 2 of 4
TIM 2 Series, 2 Watt
General Specifications
Overvoltage protection
3.3 Vout models: 4 - 6.5 VDC
5.0 Vout models: 6 - 8 VDC
9.0 Vout models: 10 - 14 VDC
12 Vout models: 13 - 19 VDC
15 Vout models: 16 - 22 VDC
24 Vout models: 25 - 35 VDC
Capacitive load
– Single output
3.3 & 5.0 Vout models: 1‘000 µF maꢀ.
9.0 Vout models: 430 µF maꢀ.
12 Vout models: 220 µF maꢀ.
15 Vout models: 170 µF maꢀ.
24 Vout models: 100 µF maꢀ.
– Dual output
12 Vout models: 170 µF maꢀ. (each output)
15 Vout models: 100 µF maꢀ. (each output)
Temperature ranges
– Operating (natural convection: 20 LFM, 0.1 m/s)
– Case temperature
–40°C to +95°C
+105°C maꢀ.
– Storage temperature
–55°C to +125°C
Derating
6.7 %/K above 90°C
5 % to 95 % rel H maꢀ.
5000 VAC (reinforced insulation)
250 VAC, 2 x MOPP
20 pF maꢀ.
Humidity (non condensing)
Isolation voltage (50 Hz, 60 s)
Working voltage
Isolation capacitance
Clearance/creepage
8 mm min.
Leakage current (at 240 VAC, 60 Hz)
Altitude during operation
2 µA maꢀ.
5000 m maꢀ.
Temperature coefficient
0.02 %/K maꢀ.
Reliability, calculated MTBF (MIL-HDBK-217F at +25°C, ground benign)
Switching frequency
6‘809‘000 h
100 kHz min. (frequency modulated)
according to MIL-STD-810F
Shock, vibration and thermal shock resistance
Remote On/Off
– On:
– Off:
– Off idle current:
open circuit or high impedance
2 – 4 mA current applied via 1kOhm resistor
2.5 mA typ.
Safety standards/approvals – Medical equipment
ANSI/AAMI ES60601-1:2005/(R)2012,
IEC/EN60601-1 3rd edition
– Certification documents
www.tracopower.com/overview/tim2
Environmental compliance – Reach
www.tracopower.com/info/reach-declaration.pdf
RoHS directive 2011/65/EU
– RoHS
Physical Specifications
Casing material
non-conductive black plastic
non-conductive black plastic
silicone (UL94 V-0)
Base material
Potting material
Package weight
7.0 g (0.24 oz)
Soldering temperature
260°C / 10 s maꢀ.
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
www.tracopower.com
Page 3 of 4
TIM 2 Series, 2 Watt
Outline Dimensions
14.4 (0.57)
10.16
(0.400)
Standard Pinout
Pin
1
2
Single
–Vin (GND)
On/Off
NC
Dual
–Vin (GND)
On/Off
NC
16
1
2
7
8
9
NC
+Vout
Common
+Vout
BOTTOM VIEW
10
16
–Vout
+Vin (Vcc)
–Vout
+Vin (Vcc)
10
9
7
8
Ø 0.60
(0.0024)
Dimensions in mm (inch)
Tolerances ±±.5 (±±.±2)
Pin ø ±.6 ±±.1 (±.±24 ±±.±±4)
2.1
(0.08)
3.8
(0.15)
10.2
(0.40)
Pin pitch tolerances ±±.25 (±±.±1)
©
Copyright 2018 Traco Electronic AG
Specifications can be changed without notice!
Rev. July 26, 2018
Page 4 of 4
www.tracopower.com
相关型号:
TIM20R
Pulse/Datacom Transformer, PULSE TRANSFORMER FOR SIGNAL SYNCHRONISATION APPLICATION(S)
VISHAY
TIM225K035P0W
Tantalum Capacitor, Polarized, Tantalum (dry/solid), 35V, 10% +Tol, 10% -Tol, 2.2uF, Through Hole Mount, 2311, RADIAL LEADED
CDE
TIM225K035P0X
Tantalum Capacitor, Polarized, Tantalum (dry/solid), 35V, 10% +Tol, 10% -Tol, 2.2uF, Through Hole Mount, 2917, RADIAL LEADED
CDE
©2020 ICPDF网 联系我们和版权申明