TPP100-112A-J [TRACOPOWER]

AC/DC Medical Power Supply;
TPP100-112A-J
型号: TPP100-112A-J
厂家: TRACO ELECTRONIC AG    TRACO ELECTRONIC AG
描述:

AC/DC Medical Power Supply

文件: 总4页 (文件大小:1087K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
AC/DC Medical Power Supply  
TPP 100A-J Series, 100 Watt  
Open frame 100 W power supply with  
JST connection in 2.0” x 3.0” package  
Certification according to IEC/EN/ES  
60601-1 3rd edition for 2 x MOPP  
Low leakage current <75 µA rated for  
BF applications  
Risk management process according to  
ISO 14971 incl. risk management file  
Acceptance criteria for electronic  
assemblies acc. to IPC-A-610 Level 3  
Active power factor correction >0.95  
Protection class l and ll prepared  
ES 60601-1 IEC 60601-1  
UL 60950-1 IEC 60950-1  
Operating up to 5000 m altitude  
The TPP 100A-J series of 100 Watt AC/DC open frame power supplies feature  
a reinforced double I/O isolation system according to latest medical safety  
standards (60601-1 3rd edition, 2 x MOPP). The earth leakage current is below  
75 µA which makes the units suitable for BF (body floating) applications. The  
excellent efficiency of up to 92% allows a high power density for the standard  
2.0” x 3.0” packaging format. The full load operating temperature range is  
–25°C to +55°C while it goes up to 80°C with 50% load derating. The EMC  
characteristic is dedicated for applications in industrial and medical fields. High  
reliability is provided by the use of industrial quality grade components and an  
excellent thermal management. It makes the products an ideal solution for  
medical devices and for demanding safety and space critical applications.  
Ready to meet ErP directive, <0.3 W no  
load power consumption  
5-year product warranty  
Models  
Order Code  
Output Power  
max.  
Output Voltage  
nom. (adjustable)  
Output Current  
max.  
Efficiency  
typ.  
TPP 100-112A-J  
TPP 100-115A-J  
TPP 100-124A-J  
TPP 100-128A-J  
TPP 100-136A-J  
TPP 100-148A-J  
12 VDC (10.8 - 13.2 VDC)  
15 VDC (13.5 - 16.5 VDC)  
24 VDC (21.6 - 26.4 VDC)  
28 VDC (25.2 - 30.8 VDC)  
36 VDC (32.4 - 39.6 VDC)  
48 VDC (43.2 - 52.8 VDC)  
8'340 mA  
6'670 mA  
4'170 mA  
3'580 mA  
2'780 mA  
2'090 mA  
91 %  
92 %  
92 %  
92 %  
91 %  
91 %  
100 W  
Page 1 / 4  
www.tracopower.com  
June 4, 2019  
TPP 100A-J Series, 100 Watt  
Input Specifications  
Input Voltage  
- AC Range  
- DC Range  
85 - 264 VAC (Full Range)  
120 - 370 VDC (Designed for, no certification)  
Input Frequency  
Input Current  
47 - 63 Hz  
550 mA max.  
1'150 mA max.  
300 mW max. (Ready to meet ErP directive)  
60 A max.  
0.95 min. (Active Power Factor Correction)  
0.95 min. (Active Power Factor Correction)  
- Full Load & Vin = 230 VAC  
- Full Load & Vin = 115 VAC  
- at no Load  
- at 230 VAC  
- at 230 VAC  
Power Consumption  
Input Inrush Current  
Power Factor  
- at 115 VAC  
Input Protection  
T 3.15 A / 250 VAC (Internal Fuse in L & N)  
Output Specifications  
Output Voltage Adjustment  
10% (By trim potentiometer)  
Output power must not exceed rated power!  
Voltage Set Accuracy  
Regulation  
1% max.  
0.2% max.  
0.5% max.  
- Input Variation (Vmin - Vmax)  
- Load Variation (0 - 100%)  
Ripple and Noise  
12 VDC model: 120 mVp-p typ. (with 10 µF X7R)  
(20 MHz Bandwidth)  
15 VDC model: 150 mVp-p typ. (with 10 µF X7R)  
24 VDC model: 160 mVp-p typ. (with 1 µF X7R)  
28 VDC model: 180 mVp-p typ. (with 1 µF X7R)  
36 VDC model: 190 mVp-p typ. (with 1 µF X7R)  
48 VDC model: 340 mVp-p typ. (with 0.1 µF X7R)  
12 VDC model: 6'950 µF max.  
15 VDC model: 4'450 µF max.  
24 VDC model: 1'750 µF max.  
28 VDC model: 1'280 µF max.  
36 VDC model: 770 µF max.  
48 VDC model: 430 µF max.  
Not required  
Capacitive Load  
Minimum Load  
Temperature Coefficient  
Hold-up Time  
0.02 %/K max.  
16 ms min.  
16 ms min.  
1'000 ms max.  
- at 230 VAC  
- at 115 VAC  
- at 230 VAC  
- at 115 VAC  
Start-up Time  
1'000 ms max.  
Short Circuit Protection  
Output Current Limitation  
Overvoltage Protection  
Transient Response  
Continuous, Automatic recovery  
115 - 150% of Iout max.  
115 - 135% of Vout nom.  
3% max. (50% to 75% Load Step)  
500 µs typ. (50% to 75% Load Step)  
- Response Deviation  
- Response Time  
Safety Specifications  
Safety Standards  
- IT / Multimedia Equipment  
IEC 60950-1  
EN 60950-1  
UL 60950-1  
- Medical Equipment  
EN 60601-1  
IEC 60601-1  
ANSI/AAMI ES 60601-1  
2 x MOPP (Means Of Patient Protection)  
www.tracopower.com/overview/tpp100a-j  
Class I Prepared: Connection to PE  
Class II Prepared: Reinforced Insulation  
PD 2  
- Certification Documents  
Protection Class  
Pollution Degree  
Over Voltage Category  
OVC II  
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.  
www.tracopower.com June 4, 2019  
Page 2 / 4  
TPP 100A-J Series, 100 Watt  
EMC Specifications  
EMC Emissions  
EN 60601-1-2 edition 4 (Medical Devices)  
EN 55011 class B (internal filter)  
EN 55032 class B (internal filter)  
FCC Part 15, class B  
- Conducted Emissions  
- Radiated Emissions  
FCC Part 18, class B  
EN 55011 class A (internal filter)  
EN 55032 class A (internal filter)  
FCC Part 15, class A  
FCC Part 18, class A  
- Harmonic Current Emissions  
- Voltage Fluctuations & Flicker  
EN 61000-3-2, class A  
EN 61000-3-2, class D  
EN 61000-3-3  
EMC Immunity  
EN 55024 (IT Equipment)  
EN 60601-1-2 edition 4 (Medical Devices)  
Air: EN 61000-4-2, 15 kV, perf. criteria A  
Contact: EN 61000-4-2, 8 kV, perf. criteria A  
EN 61000-4-3, 20 V/m, perf. criteria A  
EN 61000-4-4, 2 kV, perf. criteria A  
L to L: EN 61000-4-5, 1 kV, perf. criteria A  
L to PE: EN 61000-4-5, 2 kV, perf. criteria A  
EN 61000-4-6, 20 Vrms, perf. criteria A  
EN 61000-4-8, 10 A/m, perf. criteria A  
- Electrostatic Discharge  
- RF Electromagnetic Field  
- EFT (Burst)  
- Surge  
- Conducted RF Disturbances  
- PF Magnetic Field  
- Voltage Dips & Interruptions  
230 VAC / 50 Hz: EN 61000-4-11  
30%, 25 periods, perf. criteria A  
>95%, 0.5 periods, perf. criteria A  
>95%, 1 period, perf. criteria A  
>95%, 250 periods, perf. criteria B  
115 VAC / 60 Hz: EN 61000-4-11  
30%, 25 periods, perf. criteria A  
>95%, 0.5 periods, perf. criteria A  
>95%, 1 period, perf. criteria A  
>95%, 250 periods, perf. criteria B  
General Specifications  
Relative Humidity  
95% max. (non condensing)  
–25°C to +85°C  
–40°C to +85°C  
2 %/K above 55°C  
1.33 %/V below 100 VAC  
Natural convection (20 LFM)  
5'000 m max.  
45 - 75 kHz (PWM QR)  
Reinforced Insulation  
250 VAC  
5'657 VDC  
2'121 VDC  
2'121 VDC  
100 MOhm min.  
75 µA max.  
Temperature Ranges  
- Operating Temperature  
- Storage Temperature  
- High Temperature  
- Low Input Voltage  
Power Derating  
Cooling System  
Altitude During Operation  
Switching Frequency  
Insulation System  
Working Voltage (rated)  
Isolation Test Voltage  
- Input to Output, 60 s  
- Input to Case or PE, 60 s  
- Output to Case or PE, 60 s  
- Input to Output, 500 VDC  
- Touch Current  
Isolation Resistance  
Leakage Current  
(264 VAC)  
Reliability  
Environment  
- Calculated MTBF  
- Vibration  
- Mechanical Shock  
790'000 h (MIL-HDBK-217F, ground benign)  
IEC 60068-2-6  
IEC 60068-2-27  
JST  
Connection Type  
Weight  
156 g  
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.  
www.tracopower.com June 4, 2019  
Page 3 / 4  
TPP 100A-J Series, 100 Watt  
www.tracopower.com/info/reach-declaration.pdf  
www.tracopower.com/info/rohs-declaration.pdf  
Environmental Compliance - Reach  
- RoHS  
Supporting Documents  
Overview Link (for additional Documents)  
www.tracopower.com/overview/tpp100a-j  
Outline Dimensions  
Each one of the 4 screw holes can be used  
as a PE connection for CLASS I application.  
Pin connectors  
Input (CON1)  
Output (CON2)  
76.2 (3.00)  
Pin Function Pin* Function  
1
3
1,2  
3,4  
Line  
–Vout  
+Vout  
69.8 (2.748)  
Neutral  
*Terminal rated for 7 A max.  
(at higher current connection has to be split)  
CON1: JST series  
1
3
mates with JST crimp terminal: SVH-21T-P1.1  
and terminal housing: VHR-3N  
4
1
CON2: JST series  
mates with JST crimp terminal: SVH-21T-P1.1  
and terminal housing: VHR-4N  
Voltage adjustment  
TOP VIEW  
Dimension in mm, () = inch  
Tolerances: x.x 0.50 ( 0.02)  
x.xx 0.25 ( 0.01)  
2.5 (0.10) max.  
FRONT VIEW  
©
Copyright 2019 Traco Electronic AG  
Specifications can be changed without notice.  
Rev. June 4, 2019  
www.tracopower.com  
Page 4 / 4  

相关型号:

TPP100-115

AC/DC Medical Power Supply
TRACOPOWER

TPP100-115A-J

AC/DC Medical Power Supply
TRACOPOWER

TPP100-124

AC/DC Medical Power Supply
TRACOPOWER

TPP100-124A-J

AC/DC Medical Power Supply
TRACOPOWER

TPP100-128

AC/DC Medical Power Supply
TRACOPOWER

TPP100-128A-J

AC/DC Medical Power Supply
TRACOPOWER

TPP100-136

AC/DC Medical Power Supply
TRACOPOWER

TPP100-136A-J

AC/DC Medical Power Supply
TRACOPOWER

TPP100-148

AC/DC Medical Power Supply
TRACOPOWER

TPP100-148A-J

AC/DC Medical Power Supply
TRACOPOWER

TPP1000

Transistor
ALLEGRO

TPP1000

TRANSISTOR POWER TRANSISTOR, DIP-14, BIP General Purpose Power
VISHAY