AC/DC Medical Power Supply
TPP 40 Series, 40 Watt
• Enclosed power supply with screw
terminal connection
• Certification according to IEC/EN/ES
60601-1 3rd edition for 2xMOPP
• Low leakage current <75 µA rated for
BF applications
• Risk management process according to
ISO 14971 including risk management file
• Acceptance criteria for electronic assem-
blies according to IPC-A-610 Level 3
• EMC emission to IEC 60601-1-2 ed.4
• Protection class l and ll
CB
Scheme
prepared
ES 60601-1
• Approved for operation up to 5000 m MSL
• Ready to meet ErP directive, < 0.15 W
no load power consumption
The TPP 40 Series of 40 Watt AC/DC power supplies feature a reinforced dou-
ble I/O isolation system according to latest medical safety standards IEC/EN ES
60601-1 3rd edition for 2 x MOPP up to 5000 m MSL. The earth leakage current
is below 75 µA what makes the units suitable for BF (body floating) applications.
The eꢀcellent efficiency of up to 92% allows a high power density for the stan-
dard 2.44” ꢀ 3.0” packaging format. The full load operating temperature range
is –40°C to +70°C while it goes up to 85°C with 50% load derating. The EMC
characteristic complies to IEC 60601-1-2 ed.4 and is dedicated for applications
in industrial and domestic fields. High reliability is provided by use of industrial
quality grade components and an eꢀcellent thermal management. It makes the
products an ideal solution for medical devices and for demanding safety and
space critical applications.
• 5 year product warranty
Open frame version with pin connection
see TPP 40A Series
www.tracopower.com/overview/tpp40a
Models
Output voltage
Output current maꢀ.
Vout 2
Efficiency maꢀ.
Order code
Vout 1
Vout 2
Vout 3
Vout 1
Vout 3
TPP 40-105
TPP 40-112
TPP 40-115
TPP 40-124
TPP 40-221
TPP 40-231
TPP 40-251
TPP 40-321M2
TPP 40-331M3
TPP 40-3512
5 VDC
8.0 A
90 %
92 %
92 %
92 %
89 %
88.5 %
86 %
88 %
88 %
86 %
12 VDC
3.34 A
2.67 A
1.67 A
3.34 A
2.67 A
1.67 A
3.34 A
2.67 A
1.67 A
15 VDC
24 VDC
+12 VDC
+15 VDC
+24 VDC
+12 VDC
+15 VDC
+24 VDC
+5 VDC
+5 VDC
+5 VDC
+5 VDC
+5 VDC
+5 VDC
4 A
4 A
4 A
4 A
4 A
4 A
–12 VDC
–15 VDC
+12 VDC
0.5 A
0.5 A
0.5 A
Note: - Vout 1 is ajustable by 1ꢀ0 ꢁith iꢂteꢃꢂal poteꢂtiomet
- Multi output models have a commoꢂ gꢃouꢂd (ꢂot isolated)
- Total poꢁeꢃ should ꢂot exceed 4ꢀ Watt foꢃ coꢂtiꢂuose opeꢃatioꢂ
- Otheꢃ output voltages aꢃe available oꢂ ꢃequest
www.tracopower.com
Page 1 of 5
TPP 40 Series, 40 Watt
Input Specifications
Input voltage range
– AC range (universal input)
– DC range
85 – 264 VAC
120 – 370 VDC
Input frequency
47 – 63 Hz
Input current at full load
Input protection
– at 115 VAC / 230 VAC
1.05 A maꢀ. / 0.55 A maꢀ.
T3.15 A/250 VAC (internal fuse in both line &
neutral)
Input inrush current
– at 230 VAC
60 A maꢀ.
Zero load power consumption
0.15 W maꢀ. (acc. ErP directive)
Output Specifications
Voltage set accuracy
single output:
multi output:
1%
1% Vout1
2% Vout2, Vout3
Regulation - single output – Input variation
0.2% maꢀ.
– Load variation (0 - 100%)
5 VDC model: 0.7% maꢀ.
other models: 0.5% maꢀ.
0.2% maꢀ.
Regulation - multi output
– Input variation
– Load variation (0 - 100%)
Vout1: 0.5% maꢀ.
Vout2: 1.5% maꢀ. (0.1W to full load: 0.7% max.)
Vout3: 0.7% maꢀ.
– cross regulation (25% / 100%)
1.5% maꢀ.
Minimum load
not required (Vout3 requires 0.5 W over
Vout1/Vout 2 to be stabilized)
Temperature coefficient
Hold-up time
0.02%/K
25 ms typ.
<1s
– Vin = 115 VAC
Start-up time
Rise time
20 ms typ.
Ripple and noise
(20MHz Bandwidth)
- single output model
- multi output Vout 1
5-15 VDC outputs: 75 mVp-p typ. w. cap. 10µF/25V 1206 X7R MLCC
24 VDC output: 75 mVp-p typ. w. cap. 1µF/50V 1206 X7R MLCC
12 VDC: 120 mVp-p typ. w. cap. 1µF/50V 1206 X7R MLCC
15 VDC: 150 mVp-p typ. w. cap. 1µF/50V 1206 X7R MLCC
24 VDC 240 mVp-p typ. w. cap. 1µF/50V 1206 X7R MLCC
5 VDC: 100 mVp-p typ. w. cap. 10µF/25V 1206 X7R MLCC
(-)12 VDC: 120 mVp-p typ. w. cap. 10µF/25V 1206 X7R MLCC
15 VDC: 150 mVp-p typ. w. cap. 10µF/25V 1206 X7R MLCC
– Vout 2
– Vout 3
Overvoltage protection
Current limitation
125 – 140% of nominal Vout
single output: at 145% Iout typ.
multi output: at 145% Pout1 +Pout2
Short circuit protection
Transiente response
hiccup mode (automatic recovery)
– Peak deviation
– Recovery time
Vout1: 3% maꢀ. (25% load step change)
600 µs typ.
Capactive load
5 VDC output model: 16‘000 µF maꢀ.
12 VDC output model: 2‚785 µF maꢀ.
15 VDC output model: 1‘780 µF maꢀ.
24 VDC output model: 700 µF maꢀ.
+12 VDC / +5 VDC output model: 1‘750 µF / 2’000 µF maꢀ.
+15 VDC / +5 VDC output model: 1‘670 µF / 2’000 µF maꢀ.
+24 VDC / +5 VDC output model: 440 µF / 2’000 µF maꢀ.
+12 VDC / +5 VDC / -12 VDC output model: 1‘750 µF / 2’000 µF / 420 µF maꢀ.
+15 VDC / + 5 VDC / -15 VDC output model: 1‘670µF / 2’000 µF / 420 µF maꢀ.
+24 VDC / +5 VDC / -12 VDC output model: 440 µF / 2’000 µF / 420 µF maꢀ.
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
www.tracopower.com
Page 2 of 5
TPP 40 Series, 40 Watt
General Specifications
Operating temperature
–40°C to +85°C with derating, see graph on p.4
Output power derating
– Temperature
single output: 2.67 %/K above +70°C at 230 VAC
2.75 %/K above +65°C at 115 VAC
multi output: 1.60 %/K above +60°C at 230 VAC
1.67 %/K above +55°C at 115 VAC
– Low input voltage
1.33 %/V below 100 VAC
Storage temperature
–40°C to +85°C
5 – 95 % rel. H maꢀ.
5000 m
Humidity (non condensing)
Altitude during operation
Switching frequency - single output
(at 230 VAC)
Switching frequency - multi output
(at 230 VAC)
5 VDC model: 70 kHz typ. (pulse frequency modulation)
other models: 120 kHz typ. (pulse frequency modulation)
5 VDC model: 70 kHz typ. (pulse frequency modulation)
other models: 115 kHz typ. (pulse frequency modulation)
750 kHz typ. (pulse frequency modulation)
– Vout 1
– Vout 2
– Vout 3
510 kHz typ. (pulse frequency modulation)
Isolation voltage
(2 ꢀ MOPP insulation)
– Input / Output (60 s)
– Input / Case (60 s)
4000 VAC
2500 VAC
Leakage current (at 264 VAC/60Hz)
Isolation resistance (at 500 VDC)
Reliability
75 µA maꢀ.
100 MOhm min.
– calculated MTBF at +25°C acc. to IEC 61709
3‘000‘000 h for single output models
1‘700‘000 h for multi output models
Protection class
class II prepared
Electromagnetic compatibility – Conducted & Radiated input surpression
EN 55011 limits to IEC 60601-1-2 4th editon
EN 55032 class B (internal filter)
IEC / EN 61000-3-2, class B
(EMC), emissions
– Harmonic current emissions
– Voltage flicker
IEC / EN 61000-3-3, class B
Electromagnets compatibility (EMC), immunity
– Electrostatic discharge ESD
– RF field immunity
IEC / EN 60601-1-2
IEC / EN 61000-4-2, 8kV/6kV perf. criteria A
IEC / EN 61000-4-3, 20V/m perf. criteria A
IEC / EN 61000-4-4, 2kV perf. criteria A
IEC / EN 61000-4-5, 1kV/ 2kV perf. criteria A
IEC / EN 61000-4-6, 20 Vrms perf. criteria A
IEC / EN 61000-4-8, 10A/m perf. criteria A
– Electrical fast transients/burst immunity
– Surge
– Conducted RF
– Magnetic field (only for single output models)
Voltage dip and interruptions according to EN 60601-1-2
reference: 100 VAC / 50Hz
30%, 500ms perf. criteria A
60%, 100ms perf. criteria B
> 95%, 10ms perf. criteria A
> 95%, 5000ms perf. criteria B
Safety standards and certification
IEC/EN 60601-1 3rd edition,
ANSI/AAMI ES60601-1:2005(R)2012
www.tracopower.com/overview/tpp40
– Certification documents
Environment
– Vibration acc. IEC 60068-2-6
– Shock acc. IEC 60068-2-27
3 aꢀis, sine sweep, 10–55Hz, 1g, 1oct/min
3 aꢀis, 10g half sine, 11msShock
20 G (3 directions each 3 times)
Environmental compliance – Reach
www.tracopower.com/products/reach-declaration.pdf
RoHS directive 2011/65/EU
– RoHS
Connection
screw terminal
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
www.tracopower.com
Page 3 of 5
TPP 40 Series, 40 Watt
Derating graphs
Single output models:
Multi output models:
1ꢀꢀ
8ꢀ
6ꢀ
4ꢀ
2ꢀ
ꢀ
1ꢀꢀ
8ꢀ
6ꢀ
4ꢀ
2ꢀ
ꢀ
115 VAC
23ꢀ VAC
115 VAC
23ꢀ VAC
-25
ꢀ
2ꢀ
4ꢀ
6ꢀ
8ꢀ
-25
ꢀ
2ꢀ
4ꢀ
6ꢀ
8ꢀ
Ambieꢂt Tempeꢃatuꢃe [°C]
Ambieꢂt Tempeꢃatuꢃe [°C]
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
www.tracopower.com
Page 4 of 5
TPP 40 Series, 40 Watt
Outline Dimensions
Single output enclosed:
Screw Terminal
Input
Single
Output
Pin
1
Pin*
Dual
–Vout
+Vout
1,2
3,4
Line
3
Neutral
*Terminal rated for 10 A max.
(at higher current connection has to be split)
Weight: 169 g (5.96 oz)
Multi output enclosed:
Screw Terminal
Input
Single
Output
Pin
1
Pin*
Dual
Vout 3
Com
1
Line
3
2,3
4,5
6
Neutral
Weight: 216 g (7.80 oz)
Vout 2
Vout 1
Dimensions in inch, () = mm
Tolerances: x.xx 0.02 (x.x 0.5) x.xxx 0.01 (x.xx 0.25)
Wire dimensions range 26 - 16 AWG
M3×0.5 screw locked torque MAX 5Kgf.cm/0.49N.m
Terminal screw locked torque MAX 2Kgf.cm/0.2N.m
*Terminal rated for 10 A max.
(at higher current connection has to be split)
©
Copyright 2018 Traco Electronic AG
Specifications can be changed without notice!
Rev. June 05. 2018