TPP65A [TRACOPOWER]

AC/DC Medical Power Supply;
TPP65A
型号: TPP65A
厂家: TRACO ELECTRONIC AG    TRACO ELECTRONIC AG
描述:

AC/DC Medical Power Supply

文件: 总3页 (文件大小:513K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
AC/DC Medical Power Supply  
TPP 65A Series, 65 Watt  
Open frame power supply with pin  
connection  
Certification according to IEC/EN/ES  
60601-1 3rd edition for 2xMOPP  
Low leakage current <75 µA rated for  
BF applications  
Risk management process according to  
ISO 14971 including risk management file  
Acceptance criteria for electronic assem  
blies according to IPC-A-610 Level 3  
EMC emission to IEC 60601-1-2-ed.4  
Protection class l and ll  
CB  
Scheme  
prepared  
ES 60601-1  
UL 60950-1  
Operating up to 5000m altitude  
Ready to meet ErP directive, < 0.15 W  
no load power consumption  
The TPP 65A Series of 65 Watt AC/DC power supplies feature a reinforced dou-  
ble I/O isolation system according to latest medical safety standards (60601-1  
3rd edition, 2 x MOPP). The earth leakage current is below 75 μA what makes  
the units suitable for BF (body floating) applications. The eꢀcellent efficiency of  
up to 92% allows a high power density for the standard 2.44” ꢀ 3.0” packaging  
format. The full load operating temperature range is –40°C to +60°C while it  
goes up to 85°C with 50% load derating. The EMC characteristic is dedicated  
for applications in industrial and domestic fields. High reliability is provided by  
the use of industrial quality grade components and an eꢀcellent thermal ma-  
nagement. It makes the products an ideal solution for medical devices and for  
demanding safety and space critical applications.  
5 year product warranty  
Encased version with screw terminal  
connection see TPP 65 Series  
www.tracopower.com/overview/tpp65  
Models  
Output voltage  
Output current maꢀ.  
10.0 A  
Efficiency maꢀ.  
90 %  
Order code  
TPP 65-105A-J  
5 VDC  
12 VDC  
24 VDC  
48 VDC  
TPP 65-112A-J  
TPP 65-124A-J  
TPP 65-148A-J  
5.42 A  
92.5 %  
2.71 A  
93.5 %  
1.36 A  
93 %  
Note: - Output is ajustable by ±±10 ꢀith iꢁteꢂꢁal poteꢁtiomet  
- Otheꢂ output voltages aꢂe available oꢁ ꢂequest  
www.tracopower.com  
Page 1 of 3  
TPP 65A Series, 65 Watt  
Input Specifications  
Input voltage range  
– AC range (universal input)  
– DC range  
85 – 264 VAC  
120 – 370 VDC  
Input frequency  
47 – 63 Hz  
Input current at full load  
Input protection  
– at 115 VAC / 230 VAC  
1.65 A maꢀ. / 0.95 A maꢀ.  
T3.15 A/250 VAC (internal fuse in both line &  
neutral)  
Input inrush current  
– at 230 VAC  
60 A maꢀ.  
Zero load power consumption  
0.15 W maꢀ. (acc. ErP directive)  
Output Specifications  
Voltage set accuracy  
1%  
Regulation - single output – Input variation  
0.2% maꢀ.  
– Load variation (0 - 100%)  
5 VDC model: 0.7% maꢀ.  
other models: 0.5% maꢀ.  
0.2% maꢀ.  
Regulation - multi output  
– Input variation  
– Load variation (0 - 100%)  
– Vin = 115 VAC  
0.5% maꢀ.  
Minimum load  
Temperature coefficient  
Hold-up time  
not required  
0.02%/K  
16 ms typ.  
<1 s  
Start-up time  
Rise time  
20 ms typ.  
Ripple and noise  
(20Mhz Bandwidth)  
5 &12 VDC models: 75 mVp-p typ. w. cap. 10µF/25V 1206 X7R MLCC  
24 VDC models: 75 mVp-p typ. w. cap. 1µF/50V 1206 X7R MLCC  
48 VDC model: 150 mVp-p typ. w. cap. 0.1µF/100V 1206 X7R MLCC  
Overvoltage protection  
125 – 140% of nominal Vout  
at 145% Iout typ.  
Overload protection by current limit  
Short circuit protection  
hiccup mode (automatic recovery)  
Transiente response  
– Peak deviation  
– Recovery time  
3% maꢀ. (25% load step change)  
600 µs typ.  
General Specifications  
Operating temperature  
–40°C to +85°C with derating  
Output power derating  
– Temperature  
2 %/K above +60°C at 230 VAC  
2 %/K above +55°C at 115 VAC  
Storage temperature  
–40°C to +85°C  
5 – 95 % rel. H maꢀ.  
5000 m  
Humidity (non condensing)  
Altitude during operation  
Switching frequency  
(at 230 VAC)  
5 VDC model: 60 kHz typ. (pulse frequency modulation)  
other models: 120 kHz typ. (pulse frequency modulation)  
Isolation voltage  
(2 x MOPP insulation)  
– Input / Output (60 sec.)  
– Input / Case (60 sec.)  
4000 VAC  
2500 VAC  
Leakage current (at 264 VAC/60Hz)  
Isolation resistance (at 500 VDC)  
Reliability  
75 µA maꢀ.  
100 MOhm min.  
– calculated MTBF at +25°C acc. to IEC 61709  
1‘500‘000 h for single output models  
1‘000‘000 h for multi output models  
Protection class  
class II prepared  
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.  
www.tracopower.com  
Page 2 of 3  
TPP 65A Series, 65 Watt  
General Specifications (continued)  
Electromagnetic compatibility – Conducted & Radiated input surpression  
EN 55011 limits to IEC 60601-1-2 4th editon  
EN 55032 class B (internal filter)  
IEC / EN 61000-3-2, class A  
(EMC), emissions  
– Harmonic current emissions  
– Voltage flicker  
IEC / EN 61000-3-3, (class tba.)  
Electromagnets compatibility (EMC), immunity  
– Electrostatic discharge ESD  
– RF field immunity  
IEC / EN 60601-1-2  
IEC / EN 61000-4-2, 8kV/15kV perf. criteria A  
IEC / EN 61000-4-3, 20V/m perf. criteria A  
IEC / EN 61000-4-4, 2kV perf. criteria A  
IEC / EN 61000-4-5, 1kV/ 2kV perf. criteria A  
IEC / EN 61000-4-6, 20 Vrms perf. criteria A  
IEC / EN 61000-4-8, 10A/m perf. criteria A  
– Electrical fast transients/burst immunity  
– Surge  
– Conducted RF  
– Magnetic field (only for single output models)  
Voltage dip and interruptions according to EN 60601-1-2  
reference: 100 VAC / 50Hz  
30%, 500ms perf. criteria A  
60%, 100ms perf. criteria B  
> 95%, 10ms perf. criteria A  
> 95%, 5000ms perf. criteria B  
Safety standards and certification  
www.ul.com File e188913  
UL 60950-1, IEC/EN 60950-1,  
IEC/EN 60601-1 3rd edition,  
ANSI/AAMI ES60601-1:2005(R)2012  
Environment  
– Vibration acc. IEC 60068-2-6  
– Shock acc. IEC 60068-2-27  
3 aꢀis, sine sweep, 10–55Hz, 1g, 1oct/min  
3 aꢀis, 10g half sine, 11msShock  
20 G (3 directions each 3 times)  
Environmental compliance – Reach  
www.tracopower.com/products/reach-declaration.pdf  
RoHS directive 2011/65/EU  
– RoHS  
Connection  
screw terminal / pin connector  
Outline Dimensions  
each one of the 4 srew holes can be used  
as a PE connection for class I applications  
Screw Terminal  
Input  
(Con 1)  
Output  
(Con 2)  
Pin  
1
Single  
Pin*  
Dual  
–Vout  
+Vout  
1,2  
3,4  
Line  
3
Neutral  
*Terminal rated for 7 A maꢀ.  
(at higher current connection has to be split)  
CON 1: JST series  
mates with JST crimp terminal: BVH-21T-P1.1  
and terminal housing: VHR-3N  
CON 2: JST series  
mates with JST crimp terminal: BVH-21T-P1.1  
and terminal housing: VHR-4N  
Dimensions in inch, () = mm  
Tolerances: x.xx 0.02 (x.x 0.5) x.xxx 0.01 (x.xx 0.25)  
Weight: 114g (4.02 oz)  
©
Copyright 2018 Traco Electronic AG  
Specifications can be changed without notice!  
Rev. January 23. 2018  
Page 3 of 3  
www.tracopower.com  

相关型号:

TPP65A_1807

AC/DC Medical Power Supply
TRACOPOWER

TPP65R380C

650V Super-Junction Power MOSFET
WUMC

TPP65R600C

650V Super-Junction Power MOSFET
WUMC

TPP65R940C

650V Super-Junction Power MOSFET
WUMC

TPP65_18

AC/DC Medical Power Supply
TRACOPOWER

TPP70R450C

700V Super-Junction Power MOSFET
WUMC

TPPLD22V10-15

UV-Erasable/OTP PLD
ETC

TPPM0101PWP

SPECIALTY ANALOG CIRCUIT, PDSO14, POWER, PLASTIC, SO-14
TI

TPPM0110

DUAL LOW-DROPOUT LINEAR REGULATOR
TI

TPPM0110DWP

DUAL LOW-DROPOUT LINEAR REGULATOR
TI

TPPM0110DWPR

DUAL LOW-DROPOUT LINEAR REGULATOR
TI

TPPM0111

DUAL LOW-DROPOUT LINEAR REGULATOR
TI