6.6 to 15 Watts Medical Grade PowerAdapters
FEATURES
• Single 3.3V to 24V DC Outputs
• Interchangeable and Fixed AC Plugs
• USB Socket on 5V Output Models
• Approved to IEC/EN/UL 60601-1 Edition 3.1
• Meets IEC/EN 60601-1-2 4th Edition Medical EMC
• DOE/GEMS Level VI & ErP Stage II Compliant
• Black (Standard) and White (Optional) Colors
INPUT SPECIFICATIONS
GENERAL SPECIFICATIONS
Input Voltage Range ...............90-264 VAC
Input Frequency ......................47-63 Hz
Efficiency ................................78% min. @ full load
Hold-up Time..........................10 ms @ full load
° °
Input Current ...........................0.32A rms @ 115 VAC,
................................................0.19A rms @ 230 VAC
Inrush Current ........................60A @ 240 VAC, at 25 C cold start
Touch Current .........................100 µA max.
No Load Input Power. .............0.1W max.
................................................
Operating Temperature .........0 C to +40 C
°
°
Storage Temperature .............-20 C to +80 C
°
Operating Humidity ................20% to 80% non-condensing
Storage Humidity ....................10% to 90% non-condensing
Withstand Voltage ..................4,000 VAC, input-output
MTBF .......................................300 KHrs minimum at full load,
°
25 C ambient, calculated per
OUTPUT SPECIFICATIONS
Telecordia SR-332
Output Power Ratings ............See table
Line Regulation ......................±1% max.
Load Regulation .....................±5% max.
MODELS LIST
Ripple and Noise* ..................1% peak to peak max.
Over Voltage Protection..........Shut down
Output
Voltage
Maximum
Current
Maximum
Power
Product No.
TMW10-3.3-X
TMW10-5-X
TMW15-5-X
TMW10-9-X
TMW12-12-X
TMW12-15-X
TMW12-18-X
TMW12-24-X
Short Circuit Protection ..........Shut down; Auto-recovery
Over Current Protection .........Shutdown, Auto-recovery
3.3V
5V
2A
2A
6.6W
10W
15W
10.8W
12W
12W
12W
12W
°
Temperature Coefficient ........±0.04%/ C max.
Transient Response..............0.5 ms @ 50% load change
* Measured with 20MHz bandwidth at rated line voltage and output load ranges, with
a 10µF tantalum capacitor in parallel with a 0.1µF ceramic capacitor in parallel
across the output.
5V
3A
9V
1.2A
1A
STANDARDS & COMPLIANCES
12V
15V
18V
24V
IEC/EN 60601-1-2:2014 ....... EMC & Immunity Performance
EN 55011, CISPR 11, FCC ... Class B, conducted & radiated
EN 61000-3-2 ........................ Harmonic distortion, Class A
EN 61000-3-3 ........................ Line fluctuations & flicker
EN 61000-4-2 ........................ ESD, ±15 KV air and ±8 KV contact
EN 61000-4-3 ........................ Radiated immunity, 10V/m
EN 61000-4-4 ........................ Fast transient/burst, ±2 KV
EN 61000-4-5 ........................ Surge, ±1 KV diff., ±2 KV com.
EN 61000-4-6 ........................ Conducted immunity, 6 Vrms
EN 61000-4-8 ........................ Magnetic field immunity, 30A/m
EN 61000-4-11 ...................... Voltage dip immunity,
30% reduction for 500ms,
0.8A
0.67A
0.5A
Note: X = AC plug code. See the tables and drawings below for details.
Code X
Designation/Description
Fixed US Plug
Interchangeable Plug
US Plug
Part No.
TIPU
U
E
B
Fixed EU Plug
TIPE
EU Plug
Fixed UK Plug
60% reduction for 100ms,
TIPB
UK Plug
Interchangeable "Main Body"
without the AC Plug
>95% reduction for 10ms
I
TIPA
AU Plug
Safety Standards ................... UL/IEC/EN 60601-1 (Edition 3.1),
...............................................ANSI/AAMI ES 60601-1 (2012),
...............................................CSA C22.2 No. 60601-1 (2014);
IEC/EN 60950-1 (2nd Edition, CB)
Pending Agency Approvals.... UL, cUL, CE, CB, TUV/GS
Other Compliances .............. RoHS, DOE/GEMS Level VI,
...............................................ErP Stage II
Notes:
1.Specify the AC plug code when ordering, e.g. TMW10-5-U for 5V output with US
AC plug, TMW12-12-I for 12V output "main body" for interchangeable plug, etc.
2.For interchangeable type, at least one interchangeable plug must be ordered
together with the "main body" (X=I), e.g. TMW10-5-I for 5V output main body and
TIPU for US type interchangeable plug, etc.
3.Add suffix "P" to order units with a built-in USB socket, e.g. TMW10-5-UP for 5V
output with a fixed US plug and USB type A receptacle. The maximum current for
USB type A connector is 2.4A.
4.Standard color is black. Add suffix "W" for white version, e.g. TMW10-5-UPW.
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Building 13 • Santa Clara, California 95054, USA • 408-988-6616 • sales@trumpower.com • www.TRUMPower.com
TMW10/12/15 Series
6.6 to 15 Watts Medical Grade Power Adapters
MECHANICAL SPECIFICATIONS
Interchangeable Plug
INTERCHANGEABLE AC PLUG
TIPU
TIPE
TIPB
TIPA
Fixed US Plug
Fixed US Plug with DC Output Cable
Notes:
1. Unit: mm
2. Weight: 112g (0.38 lbs) approx.
3. Color: Black (standard) or White (optional)
4. Standard output connector: Built-in USB Type A receptacle for 5V output, or 2.1x5.5x9.5mm barrel plug on the cable for other outputs.
5. Contact TRUMPower for output connector options.
TMW10/12/15_2016-11-21
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Building 13 • Santa Clara, California 95054, USA • 408-988-6616 • sales@trumpower.com • www.TRUMPower.com
相关型号:
©2020 ICPDF网 联系我们和版权申明