TWM15 Series  
Medical Grade 15W Wall Mount PowerAdapters  
TRUMPower  
FEATURES  
• 90~264 VAC Universal Input  
• Class II Insulation  
• Standby Power Consumption < 0.3W  
• Interchangeable and Fixed Plugs Available  
• Meet CEC and Energy Star Efficiency Level V  
INPUT SPECIFICATIONS  
MODELS LIST  
Input Voltage Range ................. 90~264 VAC  
Input Frequency ........................ 47~63Hz  
Input Current ............................. 0.5A rms @ 115 VAC  
0.3A rms @ 230 VAC  
Inrush Current .......................... 30A@115 VAC or 60A@230 VAC  
(except 80A for 5V output) at 25°C  
Max.  
Output  
Active Efficency  
Output  
Ripple &  
Noise  
*Product No. Voltage Current  
Tol.  
±5%  
±5%  
±5%  
±5%  
±5%  
±5%  
@ 115 VAC @ 230 VAC  
TWM15-05-X  
TWM15-06-X  
TWM15-09-X  
TWM15-12-X  
TWM15-15-X  
TWM15-24-X  
5V  
6V  
3A  
2.5A  
70mV  
60mV  
78%  
80%  
82%  
82%  
83%  
85%  
77%  
78%  
80%  
80%  
80%  
81%  
9V  
1.67A  
1.25A  
1A  
90mV  
cold start  
Enclosure Leakage Current ..... 100 μA max. @ 264 VAC, 63Hz  
12V  
15V  
24V  
120mV  
150mV  
240mV  
OUTPUT SPECIFICATIONS  
0.625A  
Output Power ....................... 15W max.  
Output Voltage ..................... See table  
Output Current ..................... See table  
Ripple and Noise* ............... 1% p-p max., except 70 mVp-p  
on 5V with interchangeable plug  
Overvoltage Protection ........ Set at 116-230% of its nominal  
output voltage  
Overcurrent Protection ........ Protected to short-circuit conditions  
Transient Response........... Max. excursion of 4% or better on all  
models, recovering to 1% of final  
*X = AC plug code. See below tables and drawings for details.  
P/N  
IPB  
*Interchangable AC Plug  
British/UK Plug  
X
E
Fixed Type of AC Plug  
EU Plug  
IPE  
EU Plug  
U
USA Plug, Polarized  
IPU  
USA Plug (for 115V socket)  
USA Plug (for 230V socket)  
UN  
*I  
USA Plug, Non-Polarized  
*Interchangeable "main body" without AC plug  
IPUH  
*Must order with the interchangeable "main  
body" such as TWM15-05-I, etc.  
value within 500 μs after a 25%  
step load change  
STANDARDS & COMPLIANCES  
* Measured with 20MHz bandwidth at rated line voltage and output load ranges, with  
a 47 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor in parallel  
across the output.  
IEC 60601-1-2 ................... EMC & Immunity performance  
EN 55011, FCC, VCCI ....... Class B, conducted & radiated  
EN 61000-3-2 .................... Harmonic distortion, Class A & D  
EN 61000-3-3 .................... Line flicker  
GENERAL SPECIFICATIONS  
EN 61000-4-2 .................... ESD, ±8KV air and ±6KV contact  
EN 61000-4-3 .................... Radiated immunity, 3V/m  
EN 61000-4-4 .................... Fast transient/burst, ±2KV  
EN 61000-4-5 .................... Surge, ±1KV diff., ±2KV com.  
EN 61000-4-6 .................... Conducted immunity, 3 Vrms  
EN 61000-4-8 .................... Magnetic field immunity, 3A/m  
EN 61000-4-11 .................. Voltage dip immunity,  
30% reduction for 500 ms,  
60% reduction for 100 ms,  
>95% reduction for 10 ms  
Safety Standards ............... UL 60601-1, EN 60601-1  
CSA C22.2 No. 601.1  
AgencyApprovals .............. UL, cUL, TUV, CE (LVD)  
Other Compliances .......... RoHS, CEC & Energy Star Level V  
Efficiency ............................. See rating chart  
Hold-up Time....................... 8 ms min. @ 115 VAC  
Turn-on Delay Time............. 3 seconds max. @ 115 VAC  
Line Regulation................... ±0.5% max. at full load  
°
°
Operating Temperature ...... 0 C to +40 C  
Derating............................... Derate from 100% at +40 C linearly  
°
°
to 50% at +60 C  
°
°
Storage Temperature .......... -40 C to +85 C  
°
Temperature Coefficient ..... ±0.04%/ C max.  
Relative Humidity ................ 10% to 90% non-condensing  
Withstand Voltage ............... 4,000 VAC from input to output  
MTBF .................................... 300K hours minimum at full load,  
°
25 C ambient, calculated per  
MIL-HDBK-217F  
Tumbler Technologies + TRUMPower  
3350 Scott Blvd., Building 13 • Santa Clara, California 95054, USA 408-988-6616 sales@trumpower.com • www.TRUMPower.com  
TWM15 Series  
Medical Grade 15W Wall Mount PowerAdapters  
MECHANICAL SPECIFICATIONS  
Interchangeable Types  
Fixed Types  
USA/North American, Polarized  
USA/North American, Non-Polarized  
NOTES:  
1. Dimensions: inches [mm]  
2. Tolerance: 0.02 [0.5] maximum  
3. Weight: approx. 200 grams (0.44 lbs.)  
4. The DC connector shown on the drawings is just an example. Contact TRUMPower for connector options.  
5. The output cable is 20 AWG x 1600 mm, except 16 AWG x 1000 mm on 5V & 6V output models, and 18 AWG x 1200 mm on 9V output model,  
so as to comply with CEC and Energy Star efficiency level V requirements. Longer cables are available with efficiency reduced.  
#TWM15-2011-1117  
Tumbler Technologies + TRUMPower  
3350 Scott Blvd., Building 13 • Santa Clara, California 95054, USA 408-988-6616 sales@trumpower.com • www.TRUMPower.com  

相关型号:

TWM15-15-X

Medical Grade 15W Wall Mount Power Adapters
TRUMPOWER

TWM15-24-X

Medical Grade 15W Wall Mount Power Adapters
TRUMPOWER

TWM180

180W Medical Grade AC/DC Power Supplies
TRUMPOWER

TWM180-12

180W Medical Grade AC/DC Power Supplies
TRUMPOWER

TWM180-15

180W Medical Grade AC/DC Power Supplies
TRUMPOWER

TWM180-19

180W Medical Grade AC/DC Power Supplies
TRUMPOWER

TWM180-24

180W Medical Grade AC/DC Power Supplies
TRUMPOWER

TWM180-28

180W Medical Grade AC/DC Power Supplies
TRUMPOWER

TWM180-48

180W Medical Grade AC/DC Power Supplies
TRUMPOWER

TWM200

200W Medical Grade AC/DC Power Supplies
TRUMPOWER

TWM200-12

200W Medical Grade AC/DC Power Supplies
TRUMPOWER

TWM200-16

200W Medical Grade AC/DC Power Supplies
TRUMPOWER