Technical Data Sheet  
Calibration Kit  
RPC-1.85  
08CK020-150  
LRL Version  
Contents  
Device  
Calibration  
Optiona  
Part number  
Remarks  
QTY  
13 mm Air line plug/jack  
14.9 mm Air line plug/jack  
22.6 mm Air line plug/jack  
Short circuit plug  
08S101-K013  
08S101-K014  
08S101-K022  
08S12S-000S3  
08K12S-000S3  
03W008-000  
03W021-000  
L1  
1
1
1
1
1
1
1
FC  
FC  
FC  
FC  
FC  
L2  
L3  
Reflect standard  
Short circuit jack  
Reflect standard  
Combi wrench  
-
-
Torque wrench  
FC  
a.  
See “Declaration of calibration options” for explanation.  
LRL calibrations without lowband loads are band limited. There are many different possibilities to calibrate  
a VNA depending from the line length and the number of lines used during the calibration procedure. If  
two lines are used the third line can be used as verification standard. The following graphs show two  
possibilities using LRL with lowband loads. Without lowband loads the lower band cannot be calibrated.  
Frequency Range (2 Bands)  
Frequency Range (3 Bands)  
Lowband loads  
L1 and L2  
Lowband loads  
L1 and L2  
L1 and L3  
0
13  
70  
f in GHz  
0
2.6  
13  
70  
f in GHz  
Rosenberger Hochfrequenztechnik GmbH & Co. KG  
P.O.Box 1260 D-84526 Tittmoning Germany  
www.rosenberger.de  
Tel. : +49 8684 18-0  
Fax : +49 8684 18-499  
Email : info@rosenberger.de  
Page  
1 / 3  
Technical Data Sheet  
Calibration Kit  
RPC-1.85  
08CK020-150  
LRL Version  
Documentation  
This kit is delivered with  
USB-Stick  
Standard Definitions as data files for Vector Network Anaylzer Families PNA (Keysight/Agilent) and ZVA  
(Rohde&Schwarz). Calibration Certificate as PDF-file.  
Standard Definitions Cards  
Printed Standard Definitions that can be used on nearly all Vector Network Analyzers.  
Kit Info Card  
Handling precautions and information for installing Standard Definitions on a Vector Network Analyzer.  
Calibration Certificate  
Details see “Declaration of calibration options”  
Operating Manual  
Electrical specifications  
This specification covers electrical key values for the main items of the calibration kit. Specific datasheets  
are available for each component among the part number.  
Device  
Frequency  
Parameter  
Specification  
4 GHz  
to  
2.0°  
5.0°  
8.0°  
DC  
Shortsb  
(plug and jack)  
Error from  
Nominal Phase  
> 4 GHz to  
> 40 GHz to  
40 GHz  
70 GHz  
4 GHz  
40 GHz  
70 GHz  
34 dB  
26 dB  
24 dB  
0.5 GHz to  
> 4 GHz to  
> 40 GHz to  
Precision air linesc  
Return Loss  
b.  
c.  
The specifications for the shorts are given as allowed deviation from nominal model as defined in calibration certificate included  
with your kit.  
Calculated value without warranty. The return loss specification includes the connector interfaces. The minimum frequency of  
the air lines is based on calculations of the impedance change due to skin depth. Refer to the calibration certificate included with  
your kit for the exact dimensions of your precision air lines.  
Rosenberger Hochfrequenztechnik GmbH & Co. KG  
P.O.Box 1260 D-84526 Tittmoning Germany  
www.rosenberger.de  
Tel. : +49 8684 18-0  
Fax : +49 8684 18-499  
Email : info@rosenberger.de  
Page  
2 / 3  
Technical Data Sheet  
Calibration Kit  
RPC-1.85  
08CK020-150  
LRL Version  
Declaration of calibration options  
Factory Calibration  
Standard delivery for this calibration standard includes a Factory Calibration. The Calibration Certificate issued  
reports individual mechanical calibration results, traceable to national / international standards. Model based  
standard definitions are reported in an Agilent/Keysight, Rohde & Schwarz and Anritsu compatible VNA  
format.  
Accredited Calibration  
Not available.  
For further, more detailed information see application note AN001 on the Rosenberger homepage.  
Calibration interval  
Recommendation  
12 months  
Recommended accessories  
- Rosenberger Test Port Adaptor  
- Rosenberger Gauge Kit 08GK0KS-010  
- Rosenberger VNA Test cable kit and Microwave Cable Assemblies  
- Rosenberger Calibration Loads 08S150-C10S3 and 08K150-C10S3  
For further, more detailed information please visit our homepage www.rosenberger.com.  
While the information has been carefully compiled to the best of our knowledge, nothing is intended as representation or warranty on  
our part and no statement herein shall be construed as recommendation to infringe existing patents. In the effort to improve our  
products, we reserve the right to make changes judged to be necessary.  
Draft  
Date  
Approved  
Date  
Rev.  
b00  
Engineering change number  
16-0852  
Name  
Date  
Martin Moder  
04.03.15  
Herbert Babinger  
01.06.16  
S. Andorfer  
01.06.16  
Rosenberger Hochfrequenztechnik GmbH & Co. KG  
P.O.Box 1260 D-84526 Tittmoning Germany  
www.rosenberger.de  
Tel. : +49 8684 18-0  
Fax : +49 8684 18-499  
Email : info@rosenberger.de  
Page  
3 / 3