60W Ultraminiature Medical Open Frame Switching Power Supplies  
ASM60 series  
Up to 60W Output Power  
Universal 90-264VAC Input  
Single and Dual Outputs from 5VDC to 48VDC  
High Efficiency up to 86%  
4000V Isolation  
1U Compatible Height  
2”W x 4”L x 1.2”H  
Model Number  
Output Voltage Max. O/P Amps Min. O/P Amps Efficiency @ 230VAC R&N P-P Current Limit  
SINGLE OUTPUT  
ASM60-5(A)*  
5VDC  
9
0
0
0
0
0
0
0
73%  
82%  
84%  
84%  
86%  
85%  
86%  
100mV  
100mV  
150mV  
150mV  
150mV  
100mV  
400mV  
112%  
135%  
162%  
162%  
162%  
125%  
162%  
ASM60-9(A)*  
9VDC  
6.66  
5
ASM60-12(A)*  
ASM60-15(A)*  
ASM60-24(A)*  
ASM60-36(A)*  
ASM60-48(A)*  
12VDC  
15VDC  
24VDC  
36VDC  
48VDC  
4
2.5  
1.66  
1.25  
DUAL OUTPUT  
ASM60D-1  
ASM60D-2  
ASM60D-3  
ASM60D-5  
(V1/V2)  
(I1/I2)  
5/3  
5/12VDC  
5/15VDC  
5/24VDC  
5/48VDC  
1/0.60  
1/0.48  
1/0.30  
1/0.15  
81%  
82%  
82%  
82%  
100/120mV 150%  
100/150mV 150%  
100/240mV 150%  
100/400mV 150%  
5/2.4  
5/1.5  
5/0.73  
*Note: The Single Outputs can be ordered with Molex 09-50-30x1 I/O connectors.. Simply add the suffix “A” to the end of the Model Number.  
For example, ASM60-5A  
Order On-Line!  
astrodyne.com  
300 Myles Standish Blvd., Taunton, MA 02780  
Ph: 1-800-823-8082 Fax: 508-823-8181  
WWW.ASTRODYNE.COM  
60W Ultraminiature Medical Open Frame Switching Power Supplies  
ASM60 series  
INPUT SPECIFICATIONS  
GENERAL SPECIFICATIONS  
Input Voltage Range  
Frequency Range  
Inrush Current, typ:  
90-264 VAC  
Input-Out Isolation  
Efficiency  
4000VAC  
47-63 Hz  
See Selection Chart  
65Khz, (fixed, typical)  
UL, TUV, CB  
30A @ 100VAC Input *  
60A @ 200VAC Input *  
<0.1mA  
Switching Frequency  
Safety  
Leakage Current  
PHYSICAL SPECIFICATIONS  
OUTPUT SPECIFICATIONS  
Size  
2” x 4” x 1.2”  
Voltage and Current  
Load Regulation (0%-FL)  
Single O/P:  
See Selection Chart  
50.8 x 101.6 x 30.48mm  
Open Frame  
Construction  
Weight  
± 1%  
7oz (198g)  
Dual O/P (V1/V2):  
± 3/5% max  
Line Regulation (LL-HL)  
Single O/P:  
ENVIRONMENTAL SPECIFICATIONS  
± 1% max  
Dual O/P (V1/V2):  
± 1/2% max  
Oper. Temperature  
Storage Temperature  
Relative Humidity  
EMC  
-10 to +50°C(FL) , See Derate  
Cross Regulation (Dual Outputs)  
-25 to +85°C *  
(100-20% L w. other O/P at 60%) ± 3/5% max  
0 to + 95%, non-cond *  
EN 60601-1-2:2001  
185,000 Hrs  
Preset Accuracy  
1%  
DC Voltage Adjust (typ)  
Temperature Coefficient  
Ripple/Noise, 20Mhz BW  
Over Voltage Protection  
Current Limit, Self Recovering  
Short Circuit Protection  
Hold Up Time (115V / 60Hz)  
± 6% of FS (Primary)  
± 0.05%/°C  
MTBF  
Mil Std 217, 25°C  
See Selection Chart  
Clamp, 118-130%  
See Selection Chart *  
Continuous, self-recovering *  
10 mS, min.  
OUTPUT DERATING CURVE  
100  
80  
60  
40  
20  
-10  
0
10  
20  
30  
40  
50  
60  
All specifications are typical at nominal input, full load, and 25°C unless other-  
wise noted  
Astrodyne products are not authorized or warranteed for use  
as critical components in life support systems, equipment used  
in hazardous environments, nuclear controls systems, or other  
mission-critical applications.  
* These are stress ratings. Exposure of the devices to any of these conditions  
may adversely affect long term reliability. Proper operation under conditions  
other than the standard operating conditions is neither warranteed nor implied.  
300 Myles Standish Blvd., Taunton, MA 02780  
WWW.ASTRODYNE.COM  
Ph: 1-800-823-8082  
Fax: 508-823-8181  
60W Ultraminiature Medical Open Frame Switching Power Supplies  
ASM60 series  
MECHANICAL DIMENSIONS WITH JST CONNECTORS - STANDARD  
Input Connector: JST B3P-VH  
Input Mating Connector: JST VHR-3N or  
Molex 51144 Series  
Output Connector: JST B4P-VH  
Output Mating Connector: JST VHR-4N  
or Molex 53520 & 53521 Series  
Use with JST # SVH-21T-P1.1 or Molex  
# 50539-8100 Crimp Terminals  
4.000  
101.60  
All 4 corner holes need to be bonded accordingly to the end  
product PE (Protective Earth) path.  
4X O3.17(O0.125)  
3.701  
94  
0.150  
3.80  
3
4
5
6
L
1
2
N
Pin # Single Outputs Dual Outputs  
1
2
3
4
5
6
AC Line  
AC Neutral  
+ Output  
GND  
AC Line  
AC Neutral  
+ Output 2  
GND  
GND  
GND  
+ Output  
+ Output 1  
300 Myles Standish Blvd., Taunton, MA 02780  
WWW.ASTRODYNE.COM  
Ph: 1-800-823-8082  
Fax: 508-823-8181  
60W Ultraminiature Medical Open Frame Switching Power Supplies  
ASM60 series  
MECHANICAL DIMENSIONS WITH MOLEꢀ CONNECTORS (“A” VERSIONS SINGLE OUTPUTS ONLY)  
Input Connector: Molex 09-50-3031  
Input Mating Connector: JST VHR-3N or  
Molex 51144 Series  
Output Connector: Molex 09-50-3041  
Output Mating Connector: JST VHR-4N  
or Molex 53520 & 53521 Series  
Use with JST # SVH-21T-P1.1 or Molex  
# 50539-8100 Crimp Terminals  
4.000  
101.60  
All 4 corner holes need to be bonded accordingly to the end  
product PE (Protective Earth) path.  
3.701  
94  
0.150  
3.80  
4X O3.17(O0.125)  
3
4
5
6
L
1
2
N
Pin # Single Outputs  
1
2
3
4
5
6
AC Line  
AC Neutral  
+ Output  
+ Output  
GND  
GND  
300 Myles Standish Blvd., Taunton, MA 02780  
WWW.ASTRODYNE.COM  
Ph: 1-800-823-8082  
Fax: 508-823-8181