AC AUTOMATIC INSULATION/WITHSTANDING HiTESTER 3174
4
Withstanding Voltage Testing
Timers
Setting range
Operation
0.3 to 999 s
:
:
Test Voltage
When set to on: Display counts down from the set time after start.
When set to off: Display indicates time elapsed since start.
AC 0.2 to 5 kV (50/60 Hz), single-range output
Output voltage
:
:
PWM switching method (0 V start; voltage can be
changed while generating output)
Voltage output method
0.1 s (0.3 to 99.9 s) ±50 ms
1 s (100 to 999 s) ±0.5 s
Setting resolution/accuracy
:
Digital (setting resolution: 0.01 kV)
Voltage specification method
Output voltage accuracy
Maximum rated output
Transformer capacity
Voltmeter
:
:
:
:
:
Ramp Timers (Withstand voltage testing)
Setting range
±1.5% of setting ±20 V
AC 100 VA (5 kV/20 mA) continuous rating
100VA
0.1 to 99.9 s: Ramp-up and ramp-down can be set independently.
:
:
Ramp-up: The output voltage increases linearly from the
initial voltage to the test voltage over the ramp-up time.
Ramp-down: The output voltage decreases from the set
voltage to 0 V over the ramp-down time after the test time
elapses, and the display counts down from the set time.
*The actual ramp-up waveform varies with the load due to the analog response delay.
Operation
True effective value display
Digital meter: AC 0 kV to 5.00 kV
Accuracy: ±1.5% rdg (1,000 V or less, ±15 V)
Sine wave
0.1s
Setting resolution
:
Output waveform
:
:
15% or less (converges to setting within 1 s during
change from maximum rated load to no load)
Voltage change rate
Delay Timers (Insulation resistance testing)
Setting range
0.1 to 99.9 s
:
:
5% or less (tester impedance during measurement with 5
kV output under 40 MΩ load)
Distortion factor
Frequency
:
:
0.1s
Setting resolution
Decision Function
50/60Hz (±0.2%)
Window comparison method with upper and lower bound settings (digital specification)
Decision method
Decision results
:
:
Current Detection
Current measurement range
Measurement range
Indicated value
UPPER-FAIL: The measured current (measured insulation
value) exceeded the specified upper threshold.
0.01 mA to 20 mA (2 ranges)
10mA/20mA
:
:
:
:
PASS: The measured current (measured insulation value) fell within
the range defined by the specified upper and lower thresholds.
LOWER-FAIL: The measured current (measured
insulation value) was less than the specified lower bound.
UPPER LOWER-FAIL: A testing error occurred, for
example due to a failure to generate the set voltage.
True effective value display
0.00 mA to 10.00 mA, 0.01 mA (10 mA range)
0.0 mA to 20.0 mA, 0.1 mA (20 mA range)
±2% rdg ±0.05 mA (10 mA range)
Measurement ranges and
resolutions
Measurement accuracy
:
±2% rdg ±0.5 mA (20 mA range)
Display, buzzer, and EXT I/O signal output is generated according to each decision result.
Decision processing
Setting range
:
:
AC withstanding voltage: 0.1 to 20.0 mA (upper threshold), 0.1 to 19.9 mA (lower threshold)
DC insulation: 0.2 to 2,000 MΩ (500 V) or 1.0 to 2,000 MΩ (1,000 V) for both upper and lower thresholds
Insulation Resistance Measurement
AC withstanding voltage: 0.1 mA
DC insulation: 0.01 MΩ (0.2 to 2.00 MΩ), 0.1 MΩ (2.10 to
20.0 MΩ), 1 MΩ (21.0 to 200 MΩ), 10 MΩ (210 to 2,000 MΩ)
Setting resolution
:
Test Voltage
DC 500 V/1,000 V (positive polarity)
1 to 1.2 times rated voltage
Rated voltage
:
:
:
:
:
:
Unloaded voltage
Rated measurement current
Short-circuit current
Measurement range
Contact Check
1 to 1.2 mA
Detection method: Average value detection/effective value conversion
Accuracy: Setting ±50 V *Inaccuracy may increase when the waveform is distorted.
Voltmeter accuracy
:
4 to 5 mA (500 V), 2 to 3 mA (1,000 V)
0.2 to 2,000 MΩ (500 V), 0.5 to 2,000 MΩ (1,000 V)
Enables the contact check function (does not increase cycle time).
LOW: A break in the lead is determined to have occurred when the check current is not detected.
HIGH: Upper and lower thresholds for the check detection voltage can be set. A
break in the lead is determined to have occurred when the contact check measurement
voltage falls outside the range defined by the upper and lower thresholds.
Decision results
:
0.5 to 999 MΩ (500 V), 1 MΩ to 999 MΩ (1,000 V): ±4% rdg
1,000 MΩ to 2,000 MΩ: ±8% rdg
Guaranteed accuracy
ranges/accuracies
0.01 MΩ (0.20 MΩ to 19.9 MΩ)
0.1 MΩ (20.0 MΩ to 199.9 MΩ), 1 MΩ (200 MΩ to 2,000 MΩ)
Measurement resolution
:
:
Withstanding voltage testing: 0.20 kV to 5.0 kV (0.01 kV resolution; applies to both
upper and lower thresholds)Insulation resistance measurement: Upper threshold of
600 V and lower threshold of 500 V (during 500 V measurement); upper threshold of
1,200 V and lower threshold of 1,000 V (during 1,000 V measurement) (both fixed)
Voltage setting range
:
2 MΩ, 20 MΩ, 200 MΩ, 2,000 MΩ (500 V)
4 MΩ, 40 MΩ, 400 MΩ, 2,000 MΩ (1,000 V)
Measured resistance ranges
General Specifications
Model
Fluorescent display tube (digital display)
Display
:
:
:
:
:
:
AC AUTOMATIC INSULATION/WITHSTANDING HiTESTER 3174
AC AUTOMATIC INSULATION/WITHSTANDING HiTESTER 3174-01 (GP-IB Model)
Output voltage, detected current, insulation resistance
2 times per second, minimum
Monitor functions
Monitor period
0°C to 40°C, 80% RH or less (non-condensing)
-10°C to 50°C, 90% RH or less (non-condensing)
Operating temperature range
Storage temperature range
Options
23 ±5°C, 80% RH or less (non-condensing)
(With warm-up period of at least 10 min)
1 year
Temperature and humidity
range for guaranteed accuracy
ELECTRIC SAFETY TESTING SOFTWARE
SINGLE-HAND REMOTE CONTROL
TWO-HAND REMOTE CONTROL
HIGH-VOLTAGE TEST LEADS
WARNING LAMP (AC100V,0.1A)
RS-232C CABLE (9-pin to 9-pin/cross/1.8 m)
9267 (available soon)
Guaranteed accuracy term
Operating environment
Supply voltage
:
:
:
:
:
:
:
:
:
:
9613
9614
9615
9616
9637
Indoors at elevations of up to 2,000 m at a pollution level of 2
AC 100 to 240 V Designed to tolerate voltage fluctuations of ±10% of the rated supply voltage.
50Hz/60Hz
Power supply frequency
Withstanding voltage
Maximum rated power
Dimensions
Power supply to chassis: 1.39 kV at 10 mA for 15 s
200VA
RS-232C CABLE (9-pin to 25-pin/cross/1.8 m) 9638
GP-IB CONNECTING CABLE (2 m)
Approx. 320 (W) × 155 (H) × 395 (D) mm (excluding protruding parts)
Approx. 15 kg
9151-02
Weight
EMC: EN61326 Class A, EN61000-3-2, EN61000-3-3, Safety: EN61010-1
Applicable standards
Included accessories
HIGH-VOLTAGE TEST LEADS 9615 (1 each high-voltage
and return), power cord, disconnection prevention plate
HIOKI (Shanghai) Sales & Trading Co., Ltd. :
DISTRIBUTED BY
1904 Shanghai Times Square Office, 93 Huai Hai Zhong Road
Shanghai, P.R.China POSTCODE: 200021
TEL +86-21-6391-0090/0092 FAX +86-21-6391-0360
E-mail: info-sh@hioki.cn
Beijing Office :
HEAD OFFICE :
A-2602 Freetown, 58 Dong San Huan Nan Road
Beijing, P.R.China POSTCODE: 100022
TEL +86-10-5867-4080/4081 FAX +86-10-5867-4090
E-mail: info-bj@hioki.cn
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@hioki.co.jp
Guangzhou Office :
HIOKI USA CORPORATION :
Room 303, Profit Plaza, No.76, West Huangpu Road
Guangzhou, P.R.China POSTCODE: 510623
TEL +86-20-38392673/2676 FAX +86-20-38392679
E-mail: info-gz@hioki.cn
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: hioki@hiokiusa.com
All information correct as of Jun. 4, 2008. All specifications are subject to change without notice.
3174E2-86B