3290, 3290-10  
CLAMP ON AC/DC HiTESTER  
Field Measuring Instruments  
Current Measurement at DC and from 1 Hz AC,  
with Comprehensive Integrating Functions  
Choice of Three Sensors:  
20, 200 and 2000 A ranges  
Choice of Measurement Modes:  
DC, AC+DC, AC and PEAK  
Choice of Recording Outputs:  
RMS, waveform, frequency and integral value  
Choice of Response Times:  
switches in three steps  
Filter Function:  
fc = 550 Hz to suppress undesired harmonics  
Independent sensors for convenient portable measuring  
Functions  
3290-10  
* Measurement cable can be up to 100m(328.1ft) long  
Current integral measurement  
(obtain polarity-specific integrated DC values)  
100A  
Operating time/duty measurement  
φ 35mm  
(1.38")  
9691 Specifications  
Effect of conductor position: ±1.0% or less  
Effect of external magnetic field:  
equivalent to 0.5 A or less  
(in 400A/m external magnetic field)  
External dimensions:  
approx. 53W(2.09") × 129H(5.08") ×  
18D(0.71") mm  
Weight: approx. 230 g(8.1oz)  
200A  
φ 33mm  
(1.3")  
9692 Specifications  
Effect of conductor position: ±0.5% or less  
Effect of external magnetic field:  
equivalent to 0.7 A or less  
(in 400A/m external magnetic field)  
External dimensions:  
approx. 62W(2.44") × 167H(6.57") ×  
35D(1.38") mm  
Protected by dustproof cap!  
Weight: approx. 410 g(14.5oz)  
2000A  
55mm  
φ
9693 Specifications  
(2.17")  
Effect of conductor position: ±0.7% or less  
Effect of external magnetic field:  
equivalent to 2 A or less  
(in 400A/m external magnetic field)  
External dimensions:  
approx. 62W(2.44") × 196H(7.72") ×  
35D(1.38") mm  
Weight: approx. 500 g(17.6oz.)  
Sensor Common Specifications  
Cable length: 2m (6.56 ft) Maximum circuit voltage: 600V  
Applicable standards: Safety: EN61010-2-032:1995, CAT III 600V, Pollution Class 2,  
EMC: EN61326:1997+A1:1998+A2:2001  
1
Comprehensive Basic Current Measurement Functions of the 3290  
All functions needed for measuring DC and AC current and distortion waveforms in a single instrument  
Optional Model 9400 CARRYING  
CASE stores everything together  
10  
(common to 3290 and 3290- )  
Current range table  
Max. Display  
Sensor  
Used  
Frequency  
Range  
Rated  
Measurement  
Range  
Output Terminal  
Voltage  
Normal  
Peak  
Input  
Measurement Measurement  
25.00 A  
105.0 A  
25.00 A  
210.0 A  
250.0 A  
2100 A  
50.0 A  
150.0 A  
50.0 A  
300.0 A  
500 A  
100 mV/A  
10 mV/A  
100 mV/A  
10 mV/A  
10 mV/A  
1 mV/A  
20.00 A  
200.0 A  
20.00 A  
200.0 A  
200.0 A  
2000 A  
DC to  
10 kHz (-  
100 Arms  
150 Apeak  
9691  
9692  
9693  
3 dB)  
DC to  
20 kHz (-  
200 Arms  
300 Apeak  
3 dB)  
DC to  
15 kHz (-  
2000 Arms  
2840 Apeak  
3 dB)  
3000 A  
Features  
AC+DC mode measures true rms current of full- and half-wave rectified waveforms and inverter output waveforms from 1 Hz  
Peak mode measures the maximum waveform amplitude of in-rush current at device startup  
Maximum, Minimum and Mean value modes easily measure current variations in lines with severely fluctuating loads  
Simultaneous waveform and rms output enables rms and waveform or frequency fluctuations to be externally recorded together  
Selectable measurement response time enables setting the instrument response to suit changes in load current  
LPF (fc = 550 Hz) filters out unnecessary harmonic currents and noise  
Measures rms properly in the  
AC+DC Mode  
Clamp-On Current Sensor Temperature Characteristics  
Example of temperature sensitivity  
characteristics  
Example of zero-point temperature  
characteristic (Model 9693 SENSOR)  
Combined DC and AC waveform  
3
2
1
4.0  
0.0 A  
2.0  
0.0  
No.1  
No.2  
No.3  
No.4  
No.5  
9691  
Half-wave rectified waveform  
0
1  
2  
3  
9692  
9693  
-2.0  
-4.0  
0.0 A  
-40  
-20  
0
20  
40  
60  
80  
-40  
-20  
0
20  
40  
60  
80  
Full-wave rectified waveform  
Temperature [°C]  
Temperature [°C]  
* The Hall-effect current detection element in each sensor has a specific zero-point temperature characteristic that  
manifests as deviation (scatter)  
0.0 A  
Enhanced Low-Frequency Current Measurement Analysis Capabilities  
Measures current from 1 Hz, such as that used for inverter motor control.  
Inverter motor current and rms waveforms are recorded simultaneously. Simultaneous recording of control signals is the best method  
for evaluating and analyzing such devices during product development and maintenance.  
* Waveform and rms values can be recorded simultaneously (MON & REC Mode)  
Control Signal Recording  
Rms current recording (REC)  
Half-wave rectified waveform  
during operation (MON)  
Recording Current Consumption of a Hair Dryer  
Current  
Recording  
Data saved to PC Card at worksite.  
After returning to the  
office, data can be  
analyzed and stored  
on a PC.  
*Recording basic waveform component with  
3290 filter function active  
2
Comprehensive Current Integral Functions of the 3290-10  
Integral functions such as battery charging/discharging, integral AC current and repeat integral measurements are  
fully supported. In addition, the "Operating Time/Rate" function shows the operating state of loads.  
Features  
(positive, negative and total integral values)  
Polarity-specific integral function measures battery charging and discharging with one instrument  
(with settable threshold value)  
Operating time/Rate display function measures load operating status  
Polarity-specific peak current display function measures peak battery charge and discharge current with one instrument  
Repeat measurement and data storage function lets you easily conduct the same measurement up to 20 times and then save that  
data in the unit's memory.  
(can be used in combination with  
Time-limited mean value display function shows the mean value within a specific integration time  
repeat measurement)  
External DC power capability enables long-term measurements under battery power  
Current Integral Range Table  
* Integration interval can be up to 2000 hours  
Current  
Range  
Convenient Repeat Measurement Function  
Sensor Used  
Integral Range (5 ranges)  
10.000 Ah to 100.00 kAh  
Measurement Range  
Integral measurements during a specified interval can be  
repeated continuously up to 20 times.  
20 A  
0 to ± 210.0 kAh (9691)  
0 to ± 420.0 kAh (9692)  
9691/9692  
Integral and numerical measurement data values  
(maximum, minimum, mean and peak) plus operating  
time and duty are stored in internal memory. Saved data  
can be recalled and verified at any time.  
200 A  
100.00 Ah to 1000.0 kAh  
100.00 Ah to 1000.0 kAh  
1000.0 Ah to 10000 kAh  
200 A  
0 to ± 4200 kAh  
9693  
2000 A  
DC Load Current Measurement  
+
+ Peak Value  
Positive Integral  
Operating Time  
Current Threshold Setting  
0.0 A  
Operating  
Time E  
- Peak Value  
Operating  
Time A  
Operating Time C  
Negative Integral  
Total Measuring Time: Z=(A+B+C+D+E)  
Non-Operating Time B  
Non-Operating Time D  
Operating Time: A+C+E  
Rate (%): (A+C+E)/Z × 100  
-
Total Measuring Time (A+B+C+D+E)  
(Measurement may be repeated up to 20 times)  
Total integral value = positive integral +  
negative integral  
3290/3290- Function Comparison  
10  
-10  
Specification  
3290  
3290  
Measurement modes  
DC, AC+DC, AC (RMS/MEAN)  
DC, AC+DC, AC (RMS)  
(±peak value of waveform, in DC mode)  
Peak value display  
Maximum/minimum value display  
AC filter function  
Displays absolute value (of waveform peak)  
Displays maximum, minimum and mean values  
Displays polarities independently  
(in AC and AC+DC modes)  
Maximum and minimum values  
In AC and AC+DC modes, filter (fc = 550 Hz) ON/OFF  
Current  
Measurement  
(fc = 1 Hz)  
ON/OFF  
DC filter function  
During DC and AC+DC waveform output, filter (fc = 1 Hz) ON/OFF During DC waveform output, filter  
Separate AC/DC output function  
Output (connector 1)  
Output (connector 2)  
Timer settings  
Separate output of AC content and DC content  
-
Current waveform (2-V range)/rms current (2-V DC range) switchable  
(selectable +, -, total, 1-V DC/range)  
(for repeated measurement up to 20 times)  
(up to 2000 hours)  
Rms current/Low-battery warning switch  
Current integral value  
1 min. to 99 hours, 59 min  
Elapsed time display  
Mean value time limit display  
Output (connector 1)  
Output (connector 2)  
Rate  
Hours:minutes  
Integral  
Measurement  
-
Mean value = integral value / integration time  
Selectable current waveform or rms current  
(selectable +, -, total, 1-V DC/range)  
Current integral value  
Rate = operating time / total measurement time  
Rate  
Measurement  
-
(up to 2000 hours)  
Operating time measurement  
Range setting  
Displayed operating time: Hours:minutes  
Auto/manual ranging  
Auto ranging  
Frequency  
Measurement  
(1-V DC/range)  
Output (connector 1)  
Frequency value  
-
Output (connector 2)  
Data Storage  
Rms Current  
-
Peak, maximum, minimum, mean value, integral value, operating time,rate  
Bar Graph Display  
Display Refresh Rate  
When DC center is zero, bar graph can display 10X magnification  
Display of number of memorized data points  
10  
(3290- is 10 times/second)  
10  
* 3290- DC mode is once/second  
N: twice/second, F: 4 times/second  
, S: once/3 seconds  
F: 0.2 s, N: 0.8 s, S: 8 s switching  
Battery, AC adapter, external battery  
Measurement Response Time Switching  
Power Supply  
(8.4 to 15.6 V DC)  
Battery and AC adapter  
3
Model 3290 and 3290-10 Specifications (accuracy guaranteed at 23 ±5°C(73 ±9°F), 80% RH or less, non-condensating) Period of guaranteed accuracy: 1 year  
Combined accuracy with 9691 or 9692  
Measurement frequency f  
Mode  
Range Response time setting  
DC  
1 < f < 10Hz  
10 < f < 45Hz  
45 < f < 66Hz  
66Hz < f < 1kHz  
20.00 A  
±1.3%rdg.±10dgt.  
±1.3%rdg.±5dgt.  
DC  
200.0 A  
FAST  
±1.3%rdg.±12dgt.  
±2.3%rdg.±8dgt.  
(9691: up to 500Hz)  
±1.3%rdg.±8dgt.  
±1.3%rdg.±3dgt.  
20.00 A  
200.0 A  
NOMAL  
SLOW  
FAST  
±2.0%rdg.±8dgt.  
(when AC + DC, SLOW is set only)  
(when AC + DC is set only)  
±1.3%rdg.±8dgt.  
AC+DC  
AC RMS  
±1.3%rdg.±7dgt.  
±2.3%rdg.±3dgt.  
(9691: up to 500Hz)  
NOMAL  
SLOW  
±2.0%rdg.±3dgt.  
(when AC + DC, SLOW is set only)  
(when AC + DC is set only)  
±1.3%rdg.±3dgt.  
±1.3%rdg.±8dgt.  
±1.3%rdg.±3dgt.  
±2.3%rdg.±8dgt.  
±2.3%rdg.±3dgt.  
20.00 A  
200.0 A  
AC MEAN  
(only 3290)  
Combined accuracy with 9693  
Measurement frequency f  
10 < f < 45Hz  
Mode  
Range Response time setting  
DC  
1 < f < 10Hz  
45 < f < 66Hz  
66Hz < f < 1kHz  
±1.8%rdg.±10dgt.  
±1.8%rdg.±5dgt.  
200.0 A  
DC  
2000 A  
FAST  
±1.8%rdg.±12dgt.  
±1.3%rdg.±8dgt.  
±1.3%rdg.±3dgt.  
±2.3%rdg.±8dgt.  
±2.3%rdg.±3dgt.  
200.0 A  
2000 A  
NOMAL  
SLOW  
FAST  
±3.0%rdg.±8dgt.  
(when AC + DC, SLOW is set only)  
(when AC + DC is set only)  
±2.3%rdg.±8dgt.  
AC+DC  
AC RMS  
±1.8%rdg.±7dgt.  
NOMAL  
SLOW  
±3.0%rdg.±3dgt.  
(when AC + DC, SLOW is set only)  
(when AC + DC is set only)  
±2.3%rdg.±3dgt.  
±1.3%rdg.±8dgt.  
±1.3%rdg.±3dgt.  
±2.3%rdg.±8dgt.  
±2.3%rdg.±3dgt.  
200.0 A  
2000 A  
AC MEAN  
(only 3290)  
3290-10 Specifications  
Accuracy  
Frequency range (accuracy range) Maximum display  
Accuracy  
Integral value display  
Measurement time  
Current measurement accuracy ±1 dgt  
Not more than ±0.2 s deviation per hour (@23°C(73°F))  
Display accuracy ±2 mV  
10.00Hz (1.00 to 10.00Hz)  
100.0Hz (10.0 to 100.0Hz)  
1000Hz (100 to 1000Hz)  
12.50Hz  
125.0Hz  
1000Hz  
±0.3%rdg.±1dgt.  
±1.0%rdg.±1dgt.  
Output connector voltage  
Output rate:  
Provided functions Zero Adjustment (DC and AC+DC) Data  
Hold Setting Saving Function:Saves the state of settings Key Lock  
Battery Check Auto Power Off: about 10 minutes after last key  
operation, beeper warning available Beeper sound on/off  
(1) Rms current (REC_A): 2 V DC/range; (2) Current waveform (MON):  
2 V/range; (3) Frequency (REC_Hz)*1: 1 V DC/range (*1. 3290 only)  
Combined output connectors (OUT1 & OUT2*2):  
(1) Current waveform (MON / MON.FL*1) & rms current (REC);  
(2) Current waveform (MON / MON.FL*1) & low-battery detection (B.Lo)  
(3) Rms current (REC) & low-battery detection (B.Lo)  
Temperature characteristic: 0.1 × Specified Accuracy/°C (@0 to 40°C)  
(0.18 × Specified Accuracy/°F (@32 to 104°F))  
(4) Frequency (REC) & rms current (REC) (3290 only)  
*1. MON.FL is with fc = 1 Hz filter on,  
*2. OUT2 of the 3290-10 is total integral value (Ah), selected from positive, negative or total  
value  
Function display: 21-segment bar graph, over-range (O.L.), low-battery  
warning (B), Hold (HOLD), auto power off (APS)  
Measurement response time: (1) FAST (0.2 s) for 45+ Hz; (2) NORMAL  
(0.8 s) for 10+ Hz); (3) SLOW (8.0 s) for 1+ Hz)  
Operating temperature and humidity: 0 to 40°C (32 to 104°F) @80% RH or  
less (non-condensating) Storage temperature: -10 to 50°C (14 to 122°F)  
(non-condensating) Operating environment: up to 2000m ASL, indoors  
Power source: Four AA-size (LR6) alkaline batteries, or AC adapter  
Model 3290-10 also accepts 8.4 to 15.6 V external DC power Max.  
power consumption: 500 mVA (using batteries) Battery life: approx. 22h  
(continuous operation)  
Applicable safety standards EMC: EN61010-1:2001, EN61326:1997+A1:1998+  
A2:2001, EN61000-3-2:2000, EN61000-3-3:1995+A1:2001  
External dimensions, weight: approx. 155W (6.10") × 98H (3.86") × 47D  
(1.85") mm, approx. 545 g (19.2oz.)Supplied accessories: Carrying strap  
(1), AA-size (LR6) alkaline batteries (4)  
* Input: The time for analog output to stabilize after transition from 0 R 90% or from  
100 R 10%, setting AC (RMS) or AC+DC  
Display refresh rate: (1) NORMAL: twice per second; (2) SLOW: once  
per three seconds; (3) FAST: four times per second (ten times per  
second for 3290-10; (4) Bar graph display: 4 times per second  
Zero suppression: 5 counts  
Range switching: Auto/manual  
Crest factor: 2.5 or less, or the same or less than peak value of  
connected sensor  
Output impedance: 100 Ω or less  
Extension Cables Available by Special Order  
Options  
Connects from the standard 2m (6.52ft) sensor cable to the 3290 or 3290-10  
(available lengths are 5, 10, 20, 30, 50 and 100m (16.4, 32.8, 65.6, 98.4, 164.1 and  
328.1ft))  
9691  
9692  
9693  
CLAMP-ON AC/DC SENSOR (100 A)  
CLAMP-ON AC/DC SENSOR (200 A)  
CLAMP-ON AC/DC SENSOR (2000 A)  
9445-02 AC ADAPTER  
Measurement is not available with only the 3290 or 3290-10 CLAMP-ON AC/DC HiTESTER.  
A CLAMP-ON AC/DC SENSOR (Model 9691, 9692 or 9693) must also be purchased  
separately.  
9445-03 AC ADAPTER (EU)  
9094  
9400  
9199  
OUTPUT CABLE  
CARRYING CASE  
CONNECTOR ADAPTER (BNC-to-Banana [female])  
DISTRIBUTED BY  
HEAD OFFICE :  
Shanghai Representative Office :  
81 Koizumi, Ueda, Nagano, 386-1192, Japan  
1704 Shanghai Times Square Office  
TEL +81-268-28-0562 / FAX +81-268-28-0568 93 Huaihai Zhong Road  
E-mail: os-com@hioki.co.jp  
Shanghai, 200021, P.R.China  
TEL +86-21-6391-0090, 0092  
FAX +86-21-6391-0360  
E-mail: info@hioki.cn  
HIOKI USA CORPORATION :  
6 Corporate Drive, Cranbury, NJ 08512 USA  
TEL +1-609-409-9109 / FAX +1-609-409-9108  
E-mail: hioki@hiokiusa.com  
All information correct as of Oct. 5, 2004. All specifications are subject to change without notice.  
3290E3-4XB-05P Printed in Japan