AC/DC POWER HiTESTER 3334  
POWER HiTESTER 3333  
Power Measuring Instruments  
By Popular Demand  
Ideal for Meeting Energy Efficiency Standards  
High Accuracy & Cost Performance All in One Device  
Choose the 3334 for DC and Current/Power Integration Applications  
3334  
High Accuracy (±0.1% rdg. ±0.1% f.s. for 1 year)  
Exceeds the 0.5% accuracy benchmark stipulated by international standards.  
Extended Period of Guaranteed Accuracy of 3 Years  
Calibration expenses are reduced by a calibration interval six times that of our former models.  
Maximum Cost Performance  
All the necessary functions and accuracy requirements are  
provided in an easy-to-use, no-frills device.  
3333  
2
The HIOKI AC/DC POWER HiTESTER Solves All of  
3334 AC/DC POWER  
AlltheFeatures forDCandCurrent/PowerIntegrationMeasurements  
Complete Accuracy Over a Wide Input Range  
1.00mA  
0.150V  
30.00A  
300.0V  
All Measurements Within this Range Fully Guaranteed for Accuracy  
0.0000W  
9.000kW  
Current: 1mA to 30A, Voltage: 0.15V to 300V, Apparent Power: 0W to 9kW  
Measure AC or DC Loads  
Capture Inrush Current with the Peak  
Measurement Function  
With a DC to 5kHz frequency bandwidth, all AC and DC  
measurement and AC/DC elements such as half-wave rectified  
values can be tested reliably and accurately  
Measure for the Peak Value of Voltage and Current for Each Polarity  
Indepedently. Also measure the inrush current or surge current of electrical  
equipment.  
[
AC+DC Mode]: For half-wave rectified loads common in small  
household appliances such as hair dryers  
Measure simply by pressing the SHIFT+HOLD keys.  
[DC Mode]: For pure DC loads in batteries  
[AC Mode]: For loads in commercial power lines powering  
common household appliances  
inrush current  
Switch modes simply by pressing the DC/AC button on the p
Measure the inrush current when copiers  
and similar equipment are started  
Half-wave rectified  
0.0 A  
peak current  
All waves rectified  
Some hair dryers operate by switching  
the output by a half-wave rectified.  
0.0 A  
Measure the peak current of the standby  
power of home entertainment devices  
80% of Actual Size Shown  
Intuitive Setting Procedures and Easy-to-Understand Displays  
Both the 3333 and 3334 offer simple operating procedures and quick and easy-to-understand readings and alarm displays. Settings can be made  
for obtaining the average of captured data (AVG), VT ratio (conversion ratio), CT ratio, GP-IB address, integration time (from 1 minute to 10,000  
hours), and D/A Output Parameters. Information regarding the Power HiTESTER's currents status such as display hold, remote control settings, and  
key lock (to prevent unauthorized reconfigurations) can be viewed at a glance.  
*Easily test for over-consumption even when testing distorted waveforms that are commonly found in switching power supplies and similar devices by  
monitoring for the [PEAK OVER] alarm, simply by setting for the alarm to activate and the display to light up when the input exceeds the range.  
3
your Energy Consumption Testing Needs  
Meet Industrial Standard Requirements for Test Accuracy  
Measure for Consumed Power  
Also ideal for measuring the standby power and power consumption level of household appliances  
AC/DC Current and Power Integration  
Accurate Even for Waveforms  
with Large Crest Factors  
Even measure the discharge level of each individual polarity of batteries  
Reliably test waveforms with large crest factors (peak value with  
respect to the RMS value) such as pulsed systems  
*Highest effective peak voltage and peak current values on the 3334  
are 300% of the range. Accuracy is guaranteed for 1% to 100% of both  
ranges.  
For example, in the 100mA range  
where the  
300mA peak to peak  
RMS value= 1mA  
Peak value= 300mA,  
CF=300  
Measure the amount of solar gener-  
ated power and how much is being  
sold back to the power company  
Devices that are highly vulnerable to power fluctua-  
tions such as copiers and cycle-controlled equip-  
ment can also be measured for integrated power  
1mA rms  
Even waveforms such as this can be  
measured accurately with the 3334.  
Universal Power Supply  
Easy-to-connect Terminals  
Compatible to 100 - 240V AC Power Supplies  
Make a secure connection with the screw-type terminals  
*Use a No.3 Phillips screwdriver  
Actual Size  
Small Footprint  
Leading edge technology and innovations helped to  
minimize the space required, while still enabling DC  
measurements.  
*Standard rack mount kits also available upon special order.  
Rear Panel Design of Model 3334-01  
with built-in GP-IB Interface  
210  
Analog Output on All 4 Channels  
Voltage  
Current  
100  
15  
Simultaneously ouput the voltage, current  
and active power values (DC ±2 V f.s., data  
refreshed 5 times/second)  
Activepower  
Output the apparent power, power factor,  
or integrated current/active power over an  
additional 1 channel  
Integratedactivepower  
Data management with PC  
Waveform Output over 3 Channels  
Ask your distributor for more information  
regarding the freeware for processing your  
measurement data  
Instantaneous waveforms of  
the measured voltage, current  
and active power can be  
simultaneously output  
Waveformofamotorstartup  
Voltage  
Make full use of these interfaces  
to increase efficiency  
Current  
Output: 1 V f.s.  
Sampling speed: 74.4kHz  
RS-232C (3334)  
(at 50Hz: 1488 points/waveform)  
(at 60Hz: 1240 points/waveform)  
RS-232C, GP-IB (3334-01)  
Activepower  
4
Fully Answering the Needs for a High Accuracy, Long-lasting, and User-  
3333 AC POWER  
Accuracy That Can Only Be Realized with a Digital Display  
Multifunctional - 3 Meters In One  
Quick and Simple Settings  
Selectable simultaneous display of voltage, current, power  
(active or apparent) or power factor  
Display a: voltage, current or active power  
Display b: current, active power or apparent power  
Display c: active power, power factor, voltage or current  
Press the SET key to display settings  
Display a: AVG setting (number of samples to  
average)  
convenient for displaying severely  
fluctuating values  
Display b: VT (voltage transformer) ratio setting  
Display c: CT (current transformer) ratio setting  
To expand the measurement input  
level, VT and CT ratios can be set for  
transformer measurements. Primary  
winding input values can be read directly.  
Operation: Press the a, b or c key to switch between parameters  
Easy-to-see display  
Operation:Press NEXT or the a, b or c key to select a setting  
item  
Bright emissive numeric and unit symbol  
display  
Actual  
Size  
Check for Errors at a Glance  
Clearly visible instrument status  
Warning lights indicate out-of-range input to Model  
3333 when measuring voltage and current of distorted  
waveforms, such as in power switching circuits.  
Display hold, remote control and  
key-lock states are indicated separately.  
Easy range setting  
Convenient functions  
Select current range simply by pressing the button  
(accepts up to 30 AAC, including auto-ranging)  
No voltage range setting required (accepts up to  
300 VAC)  
1. Key Lock (SHIFT + KEY LOCK keys)  
Useful for avoiding inadvertent operations on production  
lines.  
2. Data Print (SHIFT + DATA OUT keys)  
Obtain hard copies of measurement data using the  
optional Printer Model 9442.  
5
Friendly Power Measuring Device for the Production and Inspection Lines  
Model 3333  
Measurement accuracy:  
±0.5% rdg. or better  
What are the advantages?  
Model3333fullyexceedstheaccuracyleveloftraditional  
analog meters that has an accuracy of only ±0.5% f.s.  
Period of guaranteed accuracy  
±0.5% f.s is assured for a full three years, reducing  
(Recommended calibration interval): 3 years calibration costs and production time losses  
Gone is the need to check for zero-position before  
Easy Operation  
measurement as you would on traditional analog meters  
Digital Display  
Quickly grasp the measurement data at a glance  
Facilitate reporting and data recording needs using your  
computer  
Data management on a PC  
Take care of a multitude of measurement needs with a  
single low-cost instrument  
Cost-Performance  
Universal Power Supply  
Easy-to-connect Terminals  
Screw-in terminal block affixes wires securely.  
Because bad wiring connections can present a fire hazard,  
the screw-type terminal block has been incorporated to  
ensure secure wiring.  
Compatible to 100 - 240VAC Power Supplies  
*BesuretouseaNo.3Phillipsscrewdriver(availableasanoptional  
accessory)toloosenandsecurethescrews.  
Actual Size  
Space-saving footprint  
Smaller installation space.  
The installed footprint of the POWER HiTESTER  
3333 is about 34% smaller than that of former models.  
This size reduction makes the instrument especially easy  
to install.  
*Rackmountsforvariousinstallationsalsoavailableon  
specialorder.Pleaseinquirefordetails.  
Rear Panel Design of Model 3333-01  
with built-in GP-IB Interface  
160  
Three-channel analog output  
100  
15  
Voltage, current and active power measurements  
are simultaneously output as +2 VDC f.s. levels  
(refreshed about five times per second).  
PC measurement and data management  
Connect to Printer Model 9442  
Use the optional Printer Model 9442 to print without  
concern for troublesome settings.  
RS-232C interface built-in  
Select Model 3333-01 for additional built-in GP-IB interface  
6
3334 and 3333 Specifications  
3334 (AC/DC)  
3333 (AC)  
General Specifications  
Measurable lines  
Single-phase, 2-wire (AC/DC)  
Single-phase, 2-wire (AC)  
Voltage, current, active power, apparent power, power  
factor  
Voltage, current, active power, apparent power, power factor,  
frequency, integrated current and active power, waveform  
peak (voltage and current)  
Measurement parameters  
Measurement method  
Sampling Frequency  
Measurement Range  
Simultaneous digital sampling of voltage and current, True RMS  
Approximately 74.4kHz  
Approximately 48kHz  
Switch between auto-range or manual  
Voltage 15.000/ 30.00/ 150.00/ 300.0V  
200.0V  
Current 100.00m/ 300.0m/ 1.0000/ 3.000/ 10.000/ 30.00  
A
50.00m/ 200.0m/ 500.0m/ 2.000/ 5.000/ 20.00A  
(refer to range composition table below)  
(refer to range composition table below)  
Power 1.5000W to 9.000kW  
DC, 45Hz to 5kHz  
10.000W to 4.000kW  
45Hz to 5kHz  
Frequency bandwidth  
Accuracy  
Guaranteed at 23°C±5, max. 80%rh, sine wave input, power factor=1, in-phase voltage =0V  
(accuracy specifications differ depending on usage period of 1 or 3 years)  
Warm-up time  
3 minutes  
10 minutes  
Period of guaranteed accuracy 3 years (better accuracy specifications available for 1-year period)  
Effective measurement range Voltage, current:1% to 100% (Power: 0% to 100%)  
Effect of power factor (at pf=0.5) Maximum ±0.4%±rdg. (45 to 66Hz)  
Temperature Coeffi cient Maximum ±0.03%f.s./ºC  
Voltage, current, power: 10% to 150%  
Valuesinthe()representtheeffectivemeasurementrange  
Measurementsbelow0.5%ofthevoltageorcurrentrangewillbezerosuppressed.  
3334  
Measurement ranges - Model  
Current  
100.00mA  
300.0mA  
1.0000A  
3.000A  
10.000A  
30.00A  
(1.00 to 100.00mA)  
(3.0 to 300.0mA)  
(0.0100 to 1.0000A)  
(0.030 to 3.000A)  
(0.100 to 10.000A)  
(0.30 to 30.00A)  
Voltage  
15.000V  
1.5000W  
4.500W  
15.000W  
45.00W  
150.00W  
450.0W  
(0.150to 15.000V)  
(0.0000 to1.5000W)  
(0.000to4.500W)  
(0.000 to 15.000W)  
(0.00 to 45.00W)  
(0.00 to 150.00W)  
(0.0 to 450.0W)  
30.00V  
3.000W  
9.000W  
30.00W  
90.00W  
300.0W  
900.0W  
(0.30 to 30.00V)  
(0.000 to 3.000W)  
(0.000 to 9.000W)  
(0.00 to 30.00W)  
(0.00 to 90.00W)  
(0.0 to 300.0W)  
(0.0 to 900.0W)  
150.00V  
15.000W  
45.00W  
150.00W  
450.0W  
1.5000kW  
4.500kW  
(1.50to 150.00V)  
(0.000 to 15.000W)  
(0.00 to 45.00W)  
(0.00 to 150.00W)  
(0.0 to 450.0W)  
(0.0000 to 1.5000kW) (0.000 to 4.500kW)  
300.0V  
30.00W  
90.00W  
300.0W  
900.0W  
3.000kW  
9.000kW  
(3.0to 300.0V)  
(0.00 to 30.00W)  
(0.00 to 90.00W)  
(0.0 to 300.0W)  
(0.0 to 900.0W)  
(0.000 to 3.000kW)  
(0.000 to 9.000kW)  
Valuesinthe()representtheeffectivemeasurementrange  
Measurementsbelow0.5%ofthevoltage,current,orpowerrangewillbezerosuppressed.  
Measurement ranges - Model 3333  
Current  
50.00mA  
200.0mA  
500.0mA  
2.000A  
5.000A  
20.00 A  
(5.00 to 75.00mA)  
(20.0 to 300.0mA)  
(50.0 to 750.0mA)  
(0.200 to 3.000A)  
(0.500 to 7.500A)  
(2.00 to 30.00A)  
Voltage  
200.0V  
10.000W  
40.00W  
100.00W  
400.0W  
1.0000kW  
4.000kW  
(20.0 to 300.0V)  
(1.000 to 15.000W)  
(4.00 to 15.000W)  
(10.00 to 15.00W)  
(40.0 to 600.0W)  
(0.1000 to 1.5000kW)  
(0.400 to 6.000kW)  
Measurement accuracy - Model 3334  
3334 and 3333 Arithmetic Expressions  
Measurement  
Formula  
Guaranteed  
Period  
Voltage, current and active power  
(at less than 50% of input range)  
Current and active power  
(at 50% to 100% of input range)  
Frequency  
Notes  
Parameters  
Apparent Power  
1 year  
3 years  
1 year  
±0.1%rdg.±0.2%f.s.  
±0.1%rdg.±0.35%f.s.  
±0.1%rdg.±0.1%f.s.  
S=U×I  
(S)  
DC  
⎥ ⎥  
= P/S  
Power Factor ( )  
λ
λ
±0.2%rdg.  
±0.3%rdg.  
±0.3%rdg.  
±0.45%rdg.  
±3.0%rdg.  
±4.5%rdg.  
(Sum of I from start of integration)  
(1 hour of data)  
45 Hz ≤ f ≤ 66 Hz  
66 Hz < f ≤ 1 kHz  
1 kHz < f ≤ 5 kHz  
Integrated Current  
3 years  
1 year  
±0.1%rdg.±0.2%f.s.  
±0.1%rdg.±0.2%f.s.  
±0.1%rdg.±0.35%f.s.  
±3.0%f.s.  
Integrated Active (Sum of P from start of integration)  
(1 hour of data)  
Accuracy not  
d e f i n e d f o r  
current input  
exceeding 20A  
Power  
3 years  
1 year  
*U=TestedVoltageValue,I=TestedCurrentValue,P=Tested  
ActivePowerValue  
3 years  
±4.5%f.s.  
Calculatingprecisionis ±1dgt.againsttheresultsobtained  
fromeachmeasuredvalue  
CurrentandactivepowerintegrationavailableonlyonModel  
3334.  
*Add±50µAtotheaccuracywhenmeasuringDCcurrent  
*Add(±50µAxvoltagevalue)totheaccuracywhenmeasuringDCactivepower  
Measurement accuracy - Model 3333  
Valuesinthe()indicateaccuracywheninputexceeds100%ofrange.  
Guaranteed  
Voltage, current and active power  
(input current 20 A or less)  
Current and active power  
(input current over 20 A)  
Frequency  
Notes  
Period  
(±0.2%rdg.)  
(±0.3%rdg.)  
1 year  
3 years  
1 year  
±0.1%rdg.±0.1%f.s.  
±0.1%rdg.±0.2%f.s.  
45 Hz ≤ f ≤ 66 Hz  
66 Hz < f ≤ 1 kHz  
1 kHz < f ≤ 5 kHz  
(±0.3%rdg.)  
±0.1%rdg.±0.2%f.s.  
±0.1%rdg.±0.35%f.s.  
Accuracy not  
d e f i n e d f o r  
current input  
exceeding 20A  
(±0.45%rdg.)  
3 years  
1 year  
(±3.0%rdg.)  
±3.0%f.s.  
±4.5%f.s.  
(±4.5%rdg.)  
3 years  
7
3334 (AC/DC)  
3333 (AC)  
Input  
Input impedance  
2.4 MΩ for voltage, 10 mΩ or better (50/60 Hz) for current  
300V, ±425Vpeak  
2.4 MΩ for voltage, 7 mΩ or better (50/60 Hz) for current  
300 Vrms, 425 Vpeak  
Maximum input voltage  
Maximum input current  
Maximum effective peak voltage  
Maximum effective peak current  
Max. rated voltage to earth  
Display  
30 A, ±42.5 Apeak  
30 Arms, 42.5 Apeak  
±300% of each voltage range, Within ±425Vpeak  
±300% of each current range, Within ±42.5Apeak  
Within 425Vpeak  
±300% of each current range, Within ±42.5Apeak  
(DC, 50/60Hz)  
(50/60Hz)  
300V  
300V  
voltage and current: 0.5% to 105% of range  
active power: 0% to 110.25% of range  
voltage and current: 1% to 152% of range  
active power: 0% to 231.04% of range  
Display indication range  
Displacement power factor  
Display refresh rate  
Response time  
0.000 to 1.000 (no polarity display)  
approx. 5 times per second  
within 0.5 s (time to rated accuracy after abrupt change in input [0 to 90% or 100 to 10% of range])  
Functions  
No.of displayed digits: Six digits  
Current Integration  
:
form 0.00000mAh, Polarity-independent integration and Sum value  
Active power Integration:  
Integration measurement  
form 0.00000mWh, Polarity-independent integration and Sum value  
Integration time: 1 min to 10000 h  
Measurement accuracy:  
measurement accuracy of active power ±1dgt.  
Maximum value of positive and negative waveform of voltage/  
current (up to 300% of full scale range)  
Wave peak measurement  
Rectification method  
Measurement accuracy: ±1.2%f.s. ("f.s." is 300% of each range)  
Switchable between AC+DC(True RMS), DC(simple  
average display) and AC(True RMS)  
AC(True RMS)  
Parameter output representation:  
Parameter output representation:  
voltage, current and active power (3 simultaneous channels)  
voltage, current and active power (3 simultaneous channels)  
Analog output  
(D/A output)  
D/A select an item from current integration, active power integration, Voltage output: +2 VDC f.s. for each range  
apparent power, power factor  
Output accuracy: ±0.5% f.s. + individual measurement accuracy  
Voltage output: 1 VDC f.s. for each range  
Output accuracy: ±0.5% f.s. + individual measurement accuracy  
Parameter output representation:  
voltage, current and active power (3 simultaneous channels)  
Voltage output: 1 VDC f.s. for each range  
Waveform output  
Output accuracy: ±1.0% f.s. + individual measurement accuracy  
Average function  
VT or CT ratio  
Simple averaging of specified number of samples: 1, 2, 5, 10, 25, 50 or 100  
VT ratios: 1, 2, 4, 10, 20, 30, 60, 100  
CT ratios: 1,2,3,4,5,6,8,10,12,15,16,20,24,25,30,40,50,60,  
75, 80,100,200,300,500,1000,2000,3000,5000,  
10000  
VT ratios: 1, 2, 4, 10, 20, 30, 60, 100  
CT ratios: 1,2,3,4,5,6,8,10,12,15,16,20,24,25,30,40,50,60,  
75,80,100  
RS-232C interface: included as standard, Asynchronous communication method: full-duplex; Baud rate: 9600 bps (fixed)  
GP-IB interface: Model 3334-01 only GP-IB interface: Model 3333-01 only  
IEEE-488.1 1987 compliant, IEEE-488.2 1987 reference IEEE-488.1 1987 compliant, IEEE-488.2 1987 reference  
External Interfaces  
Miscellaneous  
Display Hold (HOLD), Key Lock (KEYLOCK),  
Settings backup (preserves settings)  
Display Hold (HOLD), Maximum value hold, Peak value hold,  
Key Lock (KEYLOCK), Backup function (preserves settings,  
integration data)  
General Specifications  
EN61010-1:2001 Pollution Factor 2,  
Safety  
Measurement Category III (4000 V anticipated overvoltage)  
EN61326, EN61000-3-2, EN61000-3-3  
EMC  
Operating environment  
Storage environment  
Rated supply voltage  
Maximum rated power  
0 to 40 ºC, 80% RH or less, non-condensating  
-10 to 50 ºC, 80% RH or less, non-condensating  
100 to 240 VAC, 50/60 Hz  
20 VA  
210W × 100H × 245D mm (excluding feet and projections), 160W × 100H × 227D mm (excluding feet and projections),  
2.5 kg 1.9 kg  
Size and weigh  
8
Operate the Power HiTESTER from Your PC  
Data Management is as Easy as 1-2-3  
RS-232C (built-in with the 3334 and 3333)  
RS-232C, GP-IB (built-in with the 3334-01 and 3333-01)  
Free RS-232C application for both models available from your authorized HIOKI distributor only.  
Features and Functions  
Use of the software require a comprehensive understanding of the protocols and commands.  
Support for modifications to the software not available.  
1. Operate the Power HiTESTER's keys on the PC  
as you would on the actual unit  
2. Further process test data on spreadhseet  
software such as Excel  
3334  
3333  
AC/DC POWER HiTESTER 3334  
POWER HiTESTER (with GP-IB) 3334-01  
: Instruction Manual (1), Power cord (1))  
(Accessories  
POWER HiTESTER 3333  
POWER HiTESTER (with GP-IB) 3333-01  
: Instruction Manual (1), Power cord (1))  
(Accessories  
Options (Common to both Models 3334 and 3334-01  
)
RS-232C CABLE (9-pin to 9-pin, crossed cable/1.8m(0.07ft)) 9637  
RS-232C CABLE (9-pin to 25-pin, crossed cable/1.8m(0.07ft)) 9638  
GP-IB CONNECTOR CABLE (2m) 9151-02  
GP-IB CONNECTOR CABLE (4m) 9151-04  
No. 3 Phillips screwdriver  
RS-232C CABLE 9637  
GP-IB CONNECTOR CABLE 9151-02  
No. 3 Phillips screwdriver  
Print method : Thermal serial dot printing  
Paper width : 112 mm(4.41ft)  
Power supply : 9443-02/03 AC adapter, or supplied nickel-metal hydride battery  
Dimensions and weight : 160W(6.30")× 66.5H(2.62") × 17D(0.67") mm, 580g(20.5oz.)  
Option  
Printer (For the 3333 and 3333-01  
)
PRINTER 9442  
(
CONNECTION CABLE for printer 9442) 9444  
RECORDING PAPER 1196  
CONNECTION CABLE 9444  
AC ADAPTER 9443-02  
AC ADAPTER (for printer 9442 operation in Europe, except Switzerland) 9443-02  
AC ADAPTER (for printer 9442, for USA) 9443-03  
When purchasing the Printer 9442, make sure you also purchase the Connection cable 9444  
and AC adapter 9443-02/03 so that you can connect it to the 3333/3333-01  
.
HIOKI (Shanghai) Sales & Trading Co., Ltd. :  
DISTRIBUTED BY  
1904 Shanghai Times Square Office, 93 Huai Hai Zhong Road  
Shanghai, P.R.China POSTCODE: 200021  
TEL +86-21-6391-0090/0092 FAX +86-21-6391-0360  
E-mail: info-sh@hioki.cn  
Beijing Office :  
HEAD OFFICE :  
A-2602 Freetown, 58 Dong San Huan Nan Road  
Beijing, P.R.China POSTCODE: 100022  
TEL +86-10-5867-4080/4081 FAX +86-10-5867-4090  
E-mail: info-bj@hioki.cn  
81 Koizumi, Ueda, Nagano, 386-1192, Japan  
TEL +81-268-28-0562 / FAX +81-268-28-0568  
E-mail: os-com@hioki.co.jp  
Guangzhou Office :  
HIOKI USA CORPORATION :  
Room 303, Profit Plaza, No.76, West Huangpu Road  
Guangzhou, P.R.China POSTCODE: 510623  
TEL +86-20-38392673/2676 FAX +86-20-38392679  
E-mail: info-gz@hioki.cn  
6 Corporate Drive, Cranbury, NJ 08512 USA  
TEL +1-609-409-9109 / FAX +1-609-409-9108  
E-mail: hioki@hiokiusa.com  
All information correct as of Dec. 14, 2007. All specifications are subject to change without notice.  
3334E1-7ZM-07P Printed in Japan