3156 LEAK CURRENT HiTESTER  
Safety Standards  
Measuring Instruments  
The Long-Awaited Bench-type Model Is Here !  
Leakage Current Measurement Essential  
for Electrical Safety  
Because of the risks it poses, many standards stipulate leak  
current test methods, the performance of test equipment, limits  
for leak current, and other factors related to leak current. The  
bench-type 3156 LEAK CURRENT HiTESTER by itself, without  
any additional equipment, is capable of performing tests that  
comply with s tandards for a wide range of equipment, from  
ge ne ra l p urp os e e le c tric a l e q uip me nt to me d ic a l e le c tric a l  
equipment. The 3156 also is equipped with functions and an  
interface that support embedded test lines.  
http://www.hioki.co.jp/  
HIOKI company overview, new products, environmental considerations and  
other information are available on our website.  
1
Automatically compatible with Networks  
stipulated by IEC/UL/JIS standards  
In order to avoid dangers such as electric shock,  
electrical equipment is designed so that those  
portions that are likely to be touched by people  
Main Fe ature s  
are insulated from the power source. However,  
the insulation resistance is not infinite, so there is  
always some leak current present. Furthermore,  
the extent of the leak current changes over time  
as the insulation deteriorates. To determine their  
safety level, the 3156 LEAK CURRENT HiTESTER  
can prove to be highly valuable, both as a tester  
on the production line as well as a maintenance  
and inspection tool.  
Standard support for a variety of Networks  
The 3156 is engineered to automatically support the  
"Networks" (human simulated resistance) stipulated by  
the various international safety standards. Switch  
between Networks or use the filters using a PC.  
Measurement that complies with standards  
The 3156 by itself can measure leak current in compliance  
with IEC, UL, and JIS standards.  
Automatic measurement function  
RS-232C, GP-IB, and EXT I/O interfaces  
standard  
Because the 3156 is equipped with several standard  
interfaces, it is easy to set up for automatic inspection on  
a production line, etc. In addition, the RS-232C port can  
be used to connect the optional 9442 Printer for easy  
print-outs.  
The 3156 is capable of automatically measuring power  
supply polarity switching as well as the normal state/single  
failure state, and can display the maximum values. The  
measurement time and the wait time can be set by the user.  
Stores data for 100 units  
The 3156 can store data such as equipment name, control  
number, ground class, and measured values for up to 100  
pieces of equipment being tested.  
Save up to 30 sets of measuring conditions  
The 3156 can save and load up to 30 different sets of  
measuring conditions, and can switch between different  
sets quickly.  
Power supply separation  
The power supply for the 3156 is separate from the line  
power supply of the equipment being tested, preventing  
damage to equipment due to wrong supply voltage being  
input.  
Simple operation in an interactive format  
The 3156 is equipped with a touch panel that can  
be used to make settings simply by touching the  
items to be selected on the display, making  
operation even more simple.  
110% voltage application terminals  
The 110% voltage application terminals are used when testing  
medical equipment. It outputs on a 1:1 basis the voltage that is  
supplied to the line power supply for the equipment being  
tested. The polarity of the connectors can be switched.  
Measurement mode  
Leakage current between enclosure and line  
Leakage current between enclosure and earth  
Leakage current between enclosure and enclosure  
General  
equipment  
Breaker for equipment being tested  
Since the 3156 is a bench-type unit that is suited  
embedded test lines, it is equipped with a terminal  
board and breaker on the front panel, allowing  
test equipment to be easily connected to the 3156  
even when it is rack mounted.  
Earth leakage current  
Patient leakage current I  
Patient leakage current II  
Patient leakage current III  
Patient auxiliary current  
Medical  
equipment  
2
Based on the settings for the Network and  
the equipment being tested, the 3156 displays  
the appropriate meas urement mode. The  
state of the settable keys can be seen at a  
glance. All settings can be made by selecting  
the appropriate items or values.  
[Initial screen]  
Wiring check functions  
Support for proprietary networks  
The 3156 performs polarity  
checks and VA checks.  
When making measurements on a proprietary network that is  
not built into the 3156, the  
3156 can be used as a true  
effective value ammeter in the  
1MHz band.  
Browsing saved data  
The 3156 allows the user to  
browse all saved data. Saved  
data can also be printed out  
or downloaded to a computer.  
[System screen]  
Result of evaluation against tolerance setting  
Current measured value  
Maximum value display  
Displays the power supply failure status  
and the maximum value for the leak current  
as it fluctuates due to the operation of the  
equipment being tested.  
Data storage  
Measurement data: For up to 100 units  
Measuring conditions: Up to 30 sets  
Power supply polarity/equipment  
status/measured current  
Allowable value  
The upper limit for tolerance is set according  
to the standard. The set value can be changed  
freely, if necessary.  
[Measurement screen]  
Single failure setting  
The single failure status can be set for  
individual items in automatic testing, with  
continuous testing performed automatically  
only for the items for which the single failure  
setting was made. In addition, the wait time  
and measurement time can also be set.  
3
Leak current measurement  
for general equipment  
The ne twork (huma n s imula te d re s is ta nc e ) is  
specified separately in each standard, and when  
taking measurements a network that is suited to  
the standards is needed. The 3156 is pre-engineered  
with bas ic Networks that corres pond to each  
standard.  
IEC standard  
[ IEC 60990, others ]  
UL standard  
[ UL471, UL1310, UL1437, others ]  
JIS standard  
The standard numbers that are listed are only examples.  
[ JIS B8561, others ]  
The networks listed below can be used with all applicable  
standards.  
Ne twork type s  
For IEC60990s  
For JIS standards  
For UL standards  
0.22µF  
±1%  
1.5kΩ  
±0.1%  
1.5kΩ  
±0.1%  
0.15µF  
±2%  
1kΩ  
±0.1%  
500Ω  
±0.1%  
Basic measuring element  
Basic measuring element  
Basic measuring element  
General purpose 1  
10kΩ±0.1%  
0.22µF±1%  
11.22nF  
±1%  
1.5kΩ  
±0.1%  
1kΩ  
±0.1%  
10kΩ±0.1%  
1kΩ  
±0.1%  
579Ω  
±1%  
500Ω  
±0.1%  
0.022µF  
±1%  
Basic measuring element  
Filter on  
Filter 1 on  
General purpose 2  
0.22µF±1%  
1.5kΩ  
±0.1%  
10kΩ±0.1%  
20kΩ  
±0.1%  
500Ω  
±0.1%  
2kΩ  
±0.1%  
0.0091µF±1%  
0.0062µF±1%  
Basic measuring element  
Filter 2 on  
4
Leak current measurement  
for medical equipment  
Medical equipment leak current includes not only  
ground leak current and exterior leak current, but  
IEC standard  
a ls o p a tie nt le a k c urre nt (I to III) a nd p a tie nt  
measured current. The 3156 can measure all of  
these different types of leak current without any  
additional equipment.  
[ IEC 60601-1 ('88) + am1 ('91) + am2 ('95) ]  
UL standard  
[ UL 2601-1, others ]  
The 3156 does not have a built-in  
insulated transformer. When  
measuring  
medical equipment, use a step-up insulated transformer or other  
device to provide 110% of the rated voltage as the line power  
supply for the equipment being tested.  
JIS standard  
[ JIS T 0601-1, others ]  
The standard numbers that are listed are only examples. The  
Le ak curre nt type s and Ne twork type s  
For medical equipment  
Medical equipment leak current types  
Faulty medical equipment  
Faulty medical equipment  
1kΩ  
±0.1%  
Patient leakage  
current III  
Signal input section  
Medical equipment  
Patient leakage  
current II  
Basic measuring element  
10kΩ±0.1%  
Patient leakage  
current I  
Earth leakage current  
1kΩ  
±0.1%  
0.015µF  
±1%  
Enclosure leakage  
current  
Filter on  
Earth leakage current  
A current flowing from the power supply through the interior or surface of  
the insulation to the protective ground line.  
Enclosure leakage current  
A current flowing from a part of the enclosure which during normal use can be  
touched by an operator or patient either to ground or to another part of the  
enclosure, through an external conductor other than the protective ground.  
Patient leakage current  
A current flowing from the device through the patient to ground.  
Patient auxiliary current  
A current flowing between contacts which during  
normal use are connected through the patient, with  
no intended physiological effect.  
5
GP-IB  
RS-232C  
EXT I/O  
Separation of the power supply for the 3156 and  
the line for the equipment being tested  
The power supply voltage for the 3156 does not need to  
be changed, even if the supply voltage for the equipment  
being tested is different. The difference can be managed  
by controlling the line supply voltage for the equipment  
being tested. This helps prevent malfunctions due to the  
wrong voltage being input for the 3156 power supply.  
Main unit (3156)  
power supply  
Line power supply  
100 to 240V (50/60Hz)  
1500VA  
External control through EXT I/O  
The start of measurement and the loading of the measuring  
conditions can be controlled externally. The evaluation  
results, test signals, etc., can also be output externally,  
permitting the construction of an automated line.  
Contents of EXT I/O  
Outputs  
TEST  
: Continues to output a low signal during automatic measurement  
MEAS  
: Outputs the count when measuring multiple items during  
automatic measurement  
Input signal  
Active low signal  
PASS  
FAIL  
: Outputs the evaluation result "PASS" for individual items  
during automatic measurement  
Maximum  
applied voltage  
Voltage input to the EXT.DCV terminal  
: Outputs the evaluation result "FAIL" for individual items  
during automatic measurement  
Voltage input to the EXT.DCV  
terminal, or open  
High level  
Low level  
T-FAIL  
:
:
:
Continues output even if a FAIL occurs during automatic testing.  
Internal 5V DC output (not insulated from internal circuitry)  
Internal GND output (identical to rack grounded GND level)  
NT.DCV  
INT.GND  
0.3V DC or less  
Output signal  
Open collector output  
Inputs  
START  
Maximum load 24V DC (when not using the  
voltage  
EXT.DCV terminal)  
: Starts automatic measurement when low  
: Forcibly ends measurement when low  
Maximum  
output current  
60mA DC per signal  
(when low level)  
STOP  
LOAD (0 to 4): Loads a saved panel (30 panels)  
0ADJ Zero adjust signal (valid only in ground leak current mode)  
:
In addition to outputting evaluation results for each  
measurement item, the T-FAIL output is provided for  
continued output after one FAIL during automatic testing.  
EXT.DCV : External power supply input (5V to 24V DC)  
EXT.COM : External COM input  
External control through a PC  
Includes RS-232C interface and GP-IB interface as  
standard. These can be used to control the functions of  
the 3156 through a PC, and to collect measurement  
values.  
RS-232C  
GP-IB  
RS-232C interface  
Transfer method: Start-stop synchronization, full-duplex  
Transfer speed : 9600 fixed  
Data length  
Stop bit  
: 8 bits  
: 1  
Parity bit  
Delimiter  
Handshaking  
XON/XOFF  
Connector  
: None  
: CR+LF  
: None  
: Not used  
: D-sub 9-pin male  
GP-IB interface  
Compliant standards: IEEE-488.1 1987,  
Refer to IEEE-488.2 1987  
Connector  
: D-sub 24 pin female  
Data printout  
Data can be printed out on the optional 9442 Printer by  
using the RS-232C interface connector. This is useful  
for attaching inspection data, etc.  
Printout contents  
(The printed items can be selected from among the items listed below.)  
Measurement date  
Equipment name  
Control number  
Class (mounting section)  
Network  
Maximum value  
Evaluation result  
Measured current  
Printing method: Thermal serial dot printing  
Paper width/printing speed: 112 mm/52.5cps  
Power supply: 9443 AC Adapter, or NiMH  
battery provided (charge with the 9443; can print  
approximately 3000 characters with a full charge)  
(AC, DC, AC+DC, AC peak)  
Power supply polarity  
(normal phase, reverse phase)  
Measurement mode  
Filter setting  
Dimensions: Approx. 160(W)×66.5(H)×170 (D) mm  
Mass: Approx. 580g  
9442 Printer  
(Optional)  
Equipment status  
(normal, not grounded)  
Tolerance  
* Connecting the 9442 requires either the 9444 Connecting Cable or the AC adapter.  
6
3156 Specifications  
Leak current measurement  
: Indication of a current value calculated based  
the measured drop in voltage caused by simulated  
resistance of the human body.  
Measurement mode : Earth leakage current  
/
/
Leakage current between  
Leakage current between  
/ Leakage current between  
Measurement  
system  
enclosure and earth  
enclosure and enclosure  
Measurement of true effective value.  
The measurement section: chassis-grounded and  
floating.  
enclosure and line / Patient leakage current I /  
Patient leakage current II / Patient leakage current  
III / Patient auxiliary current  
: ∆ ∑ system (20 bits)  
Target current  
: DC, AC, AC+DC, AC peak  
A/D conversion system  
: 1 M±1% (single-ended input)  
Allowable value  
: 25 mA max. (DC/AC/AC+DC mode)  
75 mA max. (ACpeak mode)  
Measurement range : DC/AC/AC+DC mode:  
50 µA/500 µA/5 mA/25 mA  
AC peak mode:  
Input resistance  
measurement current  
Excluding voltmeter section, simulated resistance  
of the human body (current detection circuit)  
: 200 pF or lower (between terminals T1 and T2)  
(f = 100 kHz, with network circuit isolated)  
Input capacity  
:
200 pF or lower(between terminals T1/T2 and chassis)  
500 µA/1 mA /10 mA/75 mA  
Grounding capacity  
: 60 dB or higher, at 60 Hz, 10 kHz  
40 dB or higher, at 100 kHz, 1 MHz  
Range switch  
Trigger system  
: Auto range/Hold range  
CMRR  
(between terminals T1/T2  
and chassis)  
: Manual: Automatic generation of internal trigger,  
and free-run measurement  
Automatic: measurement started by external  
start signal  
Network (human simulated resistance)  
For medical electrical equipment  
: Basic measurement element: 1 kΩ  
Filter: 10 k+ 15 nF  
For IEC 60990 : Basic measurement element: 1.5 k+ 500 Ω  
Filter 1: 10 k+ 22 nF  
1. Started by pressing Start key on the  
operation screen.  
2. Asserted by the START terminal on  
the EXT I/O connector.  
3. Started by the :START interface  
command.  
Filter 2: 10 k+ (20 k+ 6.2 nF) // 9.1 nF  
For JIS  
For UL  
: Basic measurement element: 1 kΩ  
Filter: 10 k+ 11.22 nF + 579 Ω  
: Basic measurement element: 1.5 k// 0.15 µF  
Measurement speed  
:
Trigger-system manual measurement: 100 ±1 ms  
(indication of moving average of 16 measurements)  
Trigger-system automatic measurement:  
General-purpose 1: Basic measurement element : 1 kΩ  
General-purpose 2: Basic measurement element : 2 kΩ  
100 ±1 ms (min.) - measurement setting time  
: Terminal T1, terminal T2 (with built-in fuse holder  
Terminal T3 (110% voltage application terminal  
Measurement  
terminals  
)
)
Functions  
: Allowable value: Sets the upper-limit current value  
Judgement: PASS measurement value upper-limit value  
FAIL measurement value > upper-limit value  
Processing: Indication, buzzer, judgement output from  
EXT I/O  
:
Equipped with a voltage output terminal (T3) that applies  
110% power supply voltage between the functionally  
insulated signal input/output section (or Type F applied  
part) and ground.  
Allowable value  
judgement function  
110% voltage  
application function  
Output ON/OFF selection  
: •Current measurement function  
Output impedance: 22.5 ±1 k  
Mode selection  
function  
Unit of current measurement: Auto/ mA, fixed  
•Voltage measurement function  
Applied when positive phase (to input power supply voltage)  
•Applied when negative phase (to input power supply voltage)  
Automatic switching function (automatic measurement function)  
Isolates the internal network for using the product as a  
voltmeter between terminals T1 and T2.  
Maximum measurement voltage: 25 V  
: •Allowable value judgement  
: Polarity check/VA check  
Wiring check function  
: Setting of measuring time.  
Automatic measurement  
function  
Setting of delay (wait) time for changing setting conditions.  
The power supply polarity and equipment status are  
automatically switched during measurement.  
: Use of T2 and internal contact/Use of T1 and T2  
Beep sound setting  
Save/load function  
•Key input  
•T3 (110% voltage application terminal) output  
•Line voltage output from T2  
Application line  
selection function  
: 30 panels for saving the following setting data  
(measurement mode, network, equipment name, control  
number, grounding class, applied part, measurement  
range, filter, target current, allowable value setting,  
malfunction condition setting, power supply polarity  
switching, automatic measurement items, automatic  
measuring time, measurement delay time)  
: Saved content: Sample equipment information (equipment  
name, control number), measurement data, date  
Memory capacity: Data on up to 100 units  
: Auto calendar, automatic leap-year adjustment, 24-  
hour clock  
:
Pre-check of current value between connection terminals to  
prevent a ground fault. Only effective for leakage current  
measurement between enclosure and line.  
Ground fault  
prevention function  
: •Setting of malfunction mode for power line for  
sample equipment.  
Setting of single-  
fault condition  
1. Disconnection of one wire in power line (neutral side)  
2. Disconnection of protective earth conductor  
Data save function  
Clock function  
Application of 110% voltage for simulated connection of  
malfunctioning equipment.  
Positive phase/negative phase  
•Selection of application line for leakage current  
measurement between enclosure and line.  
: Positive/negative phase  
Clock accuracy: Deviation of about 4 minutes a month  
: SRAM (test condition data), RTC  
Power line for sample  
equipment (switching  
power supply polarity)  
Data backup  
function  
(Automatic switching possible when using automatic  
measurement function)  
Backup battery life: 4 years (reference value at 25ºC)  
: Constant ON/Auto OFF  
Backlight automatic  
OFF function  
: Setting range: 1 sec. to 5 min., in 1-sec. increments  
Effective only in automatic measurement  
: Setting range: 1 sec. to 30 min., in 1-sec. increments  
•Wait time from the completion of measurement to  
power supply disconnection  
Setting of measuring  
time  
1 min. to 30 min., in 1-min. increments  
: MEM (internal RAM)/KEY (6 × 6 matrix touch panel)/  
Self-test function  
Measurement delay  
(setting) function  
LCD (front LCD panel)/LED /Buzzer  
: Japanese or English  
Language setting  
System reset  
: Clears all data including measurement conditions and  
measurement data.  
•Wait time from switching power supply polarity to  
the start of measurement  
Clears all saved measurement data.  
Clears all saved condition setting data including panels.  
•Wait time for operations other than switching polarity  
: Effective in all measurement modes  
Maximum value hold  
function  
7
Accuracy (Current measurement )  
Operating temperature and humidity for guaranteed accuracy: 23 ± 5 ºC, 80%rh or lower (no condensation)  
Temperature coefficient: 0.1 x basic accuracy x (T-23) weighted --- operating temperature T (ºC)  
Warm-up time: 30 min.  
Value calculated based on voltage detected at terminals of Network having non-inductive resistance of 1 k  
(theoretical value).  
Measurement of DC  
Guaranteed Measurement  
accuracy range resolution  
Measurements in voltmeter mode conform to the following accuracy level.  
Range  
Accuracy  
1
Measurement of AC* , AC+DC  
25.00 mA 4 mA or more  
5.000 mA 400  
10 µA  
1 µA  
±(0.2%rdg.+3dgt.)  
µA or more  
Accuracy  
Guaranteed Measurement  
accuracy range resolution  
Range  
500.0 µA 40 µA or more 0.1 µA  
50.00 µA 4 µA or more 0.01 µA  
DC < f 100kHz  
±(2.0%rdg.+6dgt.)  
±2.0%f.s.  
100kHz < f 1MHz  
±1.0%f.s.  
25.00 mA 4 mA or more  
5.000 mA 400 µA or more  
10 µA  
1 µA  
±(2.0%rdg.+10dgt.)  
±2.0%f.s.  
Power supply voltage monitor accuracy  
500.0 µA 40 µA or more 0.1 µA  
50.00 µA 4 µA or more 0.01 µA  
Guaranteed Measurement  
accuracy range resolution  
Range  
Accuracy  
85 Vor more*3  
0.1 V  
300.0 V  
±(2.0%rdg.+10dgt.)  
2
Measurement of AC Peak*  
Current consumption monitor accuracy  
Accuracy  
Guaranteed Measurement  
Range  
(measurement method: average value response, effective value conversion)  
accuracy range resolution  
20Hz < f 1kHz  
1kHz  
< f 10kHz  
Guaranteed Measurement  
accuracy range resolution  
Range  
Accuracy  
75.0 mA  
8mA or more  
100 µA  
10 µA  
1 µA  
±(2.0%rdg.+2dgt.)  
±(5.0%rdg.+10dgt.)  
0.5Aor more*4  
16 A  
0.1 A  
±(2.0%rdg.+5dgt.)  
10.00 mA 0.8mA or more  
1.000 mA 100 µA or more  
*1: Weighted with frequency characteristic (fc = 4 Hz) of high-pass filter.  
*2: Cannot be set when network A , B and C(filter off) is selected.  
*3: "less than 80 V" is indicated for a value less than 80 V.  
*4: "less than 0.5 A" is indicated for a value less than 0.5 A.  
±
±
2.5  
4.0  
%
f.s.  
f.s.  
±
5.0%f.s.  
500.0 µA 40 µA or more 0.1 µA  
%
General Specifications  
Display section  
Operation section  
: EMC: EN61326:1997+A1:1998+A2:2001  
EN61000-3-2:2000  
: 320  
×
240 dot matrix LCD (with backlight)  
Applicable Standards  
: 6  
× 6 matrix touch panel  
Operating temperature  
and humidity  
Storage temperature  
and humidity  
Recommended  
Calibration Period  
Operating Environment  
Power source for  
main unit  
EN61000-3-3:1995+A1:2001  
Safety: EN61010-1:2001  
: 0 to 40 ºC , 80%rh or less (no condensation)  
: -10 to 50 ºC, 80%rh or less (no condensation)  
: 6 months  
EN61010-2-031:1994 Pollution Degree 2  
Terminals T1, T2: Measurement category II  
(Anticipated Transient Overvoltage: 2.5 kV)  
Terminal T3: Measurement category I  
(Anticipated Transient Overvoltage: 1.5 kV)  
: 16%f.s. at 3 V  
: Indoors, <2000 m ASL  
: 100, 120, 220, 240 VAC (default setting)  
Rated power source frequency: 50/60 Hz  
Rated power: 30 VA  
: 100 to 240 VAC  
Rated power source frequency: 50/60 Hz  
Rated power: 1,500 VA  
Effect of conducted  
radio-frequency  
(typical value for measurement in AC 500 mA range)  
electromagnetic field  
Power line for sample  
equipment and power  
output  
Accessories  
: 9170 TEST LEAD 2 sets, 9195 ENCLOSURE  
PROBE 1, 9399 CARRYING CASE 1, Alligator  
clips 3 (2 red, 1 black), AC power cord 2 (for  
main unit, for power line of sample equipment),  
Spare fuse 2 (for main unit power supply 250V  
T0.1AL, for measurement operation 250V  
T32mAL)  
P o we r o utp ut m a xim um  
allowable leakage current  
Withstand voltage  
: 25 mA  
:
[All power supply terminals] - [Protective earthing]  
1.35 kV AC (5 mA), 1 min.  
[All measurement terminals] - [All power supply terminals]  
2.30 kV AC (5 mA), 1 min  
[All measurement terminals] - [Control circuit]  
2.30 kV AC (5 mA), 1 min  
Dimensions  
Mass  
: Approx. 320W  
: Approx. 4.0kg  
× 110H × 263D mm  
3156 LEAK CURRENT HiTESTER  
Options  
9637 RS-232C CABLE (9-pin to 9-pin., crossing cable,1.8m)  
9638 RS-232C CABLE (9-pin to 25-pin., crossing cable,1.8m)  
9151-02 GP-IB CABLE (2 m)  
9195 ENCLOSURE PROBE  
(Standard accessory)  
9170 TEST LEAD  
9399 CARRYING CASE  
(Standard accessory)  
9686 CARRYING CASE  
(Option)  
(Standard accessory/2 set)  
9151-04 GP-IB CABLE (4 m)  
9442 PRINTER  
9444 CONNECTION CABLE (for printer)  
1196 RECORDING PAPER (for printer)  
9686 CARRYING CASE (with casters)  
9443-01 AC ADAPTER (for printer, for use in Japan)  
9443-02 AC ADAPTER (for printer, for use in EU)  
9443-03 AC ADAPTER (for printer, for use in U.S.)  
9267 SAFETY TEST DATA MEASUREMENT SOFTWARE  
DISTRIBUTED BY  
Shanghai Representative Office :  
1704 Shanghai Times Square Office  
93 Huaihai Zhong Road  
Shanghai, 200021, P.R.China  
TEL +86-21-6391-0090, 0092  
FAX +86-21-6391-0360  
HEAD OFFICE :  
81 Koizumi, Ueda, Nagano, 386-1192, Japan  
TEL +81-268-28-0562 / FAX +81-268-28-0568  
E-mail: os-com@hioki.co.jp  
HIOKI USA CORPORATION :  
6 Corporate Drive, Cranbury, NJ 08512 USA  
TEL +1-609-409-9109 / FAX +1-609-409-9108  
E-mail: hioki@hiokiusa.com  
E-mail: hioki-sh@81890.net  
All information correct as of Jun. 20, 2003. All specifications are subject to change without notice.  
3156E2-36M-05P Printed in Japan