3156 Specifications
ꢀꢀLeak current measurement
: Indication of a current value calculated based
the measured drop in voltage caused by simulated
resistance of the human body.
Measurement mode : Earth leakage current
/
/
Leakage current between
Leakage current between
/ Leakage current between
Measurement
system
enclosure and earth
enclosure and enclosure
Measurement of true effective value.
The measurement section: chassis-grounded and
floating.
enclosure and line / Patient leakage current I /
Patient leakage current II / Patient leakage current
III / Patient auxiliary current
: ∆ ∑ system (20 bits)
Target current
: DC, AC, AC+DC, AC peak
A/D conversion system
: 1 MΩ ±1% (single-ended input)
Allowable value
: 25 mA max. (DC/AC/AC+DC mode)
75 mA max. (ACpeak mode)
Measurement range : DC/AC/AC+DC mode:
50 µA/500 µA/5 mA/25 mA
AC peak mode:
Input resistance
measurement current
Excluding voltmeter section, simulated resistance
of the human body (current detection circuit)
: 200 pF or lower (between terminals T1 and T2)
(f = 100 kHz, with network circuit isolated)
Input capacity
:
200 pF or lower(between terminals T1/T2 and chassis)
500 µA/1 mA /10 mA/75 mA
Grounding capacity
: 60 dB or higher, at 60 Hz, 10 kHz
40 dB or higher, at 100 kHz, 1 MHz
Range switch
Trigger system
: Auto range/Hold range
CMRR
(between terminals T1/T2
and chassis)
: Manual: Automatic generation of internal trigger,
and free-run measurement
Automatic: measurement started by external
start signal
ꢀꢀNetwork (human simulated resistance)
ꢀꢀFor medical electrical equipment
: Basic measurement element: 1 kΩ
Filter: 10 kΩ + 15 nF
ꢀꢀFor IEC 60990 : Basic measurement element: 1.5 kΩ + 500 Ω
Filter 1: 10 kΩ + 22 nF
1. Started by pressing Start key on the
operation screen.
2. Asserted by the START terminal on
the EXT I/O connector.
3. Started by the :START interface
command.
Filter 2: 10 kΩ + (20 kΩ+ 6.2 nF) // 9.1 nF
ꢀꢀFor JIS
ꢀꢀFor UL
: Basic measurement element: 1 kΩ
Filter: 10 kΩ + 11.22 nF + 579 Ω
: Basic measurement element: 1.5 kΩ // 0.15 µF
Measurement speed
:
Trigger-system manual measurement: 100 ±1 ms
(indication of moving average of 16 measurements)
Trigger-system automatic measurement:
ꢀꢀGeneral-purpose 1: Basic measurement element : 1 kΩ
ꢀꢀGeneral-purpose 2: Basic measurement element : 2 kΩ
100 ±1 ms (min.) - measurement setting time
: Terminal T1, terminal T2 (with built-in fuse holder
Terminal T3 (110% voltage application terminal
Measurement
terminals
)
)
ꢀꢀFunctions
: Allowable value: Sets the upper-limit current value
Judgement: PASS measurement value ≤ upper-limit value
FAIL measurement value > upper-limit value
Processing: Indication, buzzer, judgement output from
EXT I/O
:
Equipped with a voltage output terminal (T3) that applies
110% power supply voltage between the functionally
insulated signal input/output section (or Type F applied
part) and ground.
Allowable value
judgement function
110% voltage
application function
Output ON/OFF selection
: •Current measurement function
Output impedance: 22.5 ±1 kΩ
Mode selection
function
Unit of current measurement: Auto/ mA, fixed
•Voltage measurement function
•Applied when positive phase (to input power supply voltage)
•Applied when negative phase (to input power supply voltage)
Automatic switching function (automatic measurement function)
Isolates the internal network for using the product as a
voltmeter between terminals T1 and T2.
Maximum measurement voltage: 25 V
: •Allowable value judgement
: Polarity check/VA check
Wiring check function
: Setting of measuring time.
Automatic measurement
function
Setting of delay (wait) time for changing setting conditions.
The power supply polarity and equipment status are
automatically switched during measurement.
: Use of T2 and internal contact/Use of T1 and T2
Beep sound setting
Save/load function
•Key input
•T3 (110% voltage application terminal) output
•Line voltage output from T2
Application line
selection function
: 30 panels for saving the following setting data
(measurement mode, network, equipment name, control
number, grounding class, applied part, measurement
range, filter, target current, allowable value setting,
malfunction condition setting, power supply polarity
switching, automatic measurement items, automatic
measuring time, measurement delay time)
: Saved content: Sample equipment information (equipment
name, control number), measurement data, date
Memory capacity: Data on up to 100 units
: Auto calendar, automatic leap-year adjustment, 24-
hour clock
:
Pre-check of current value between connection terminals to
prevent a ground fault. Only effective for leakage current
measurement between enclosure and line.
Ground fault
prevention function
: •Setting of malfunction mode for power line for
sample equipment.
Setting of single-
fault condition
1. Disconnection of one wire in power line (neutral side)
2. Disconnection of protective earth conductor
Data save function
Clock function
•Application of 110% voltage for simulated connection of
malfunctioning equipment.
Positive phase/negative phase
•Selection of application line for leakage current
measurement between enclosure and line.
: Positive/negative phase
Clock accuracy: Deviation of about 4 minutes a month
: SRAM (test condition data), RTC
Power line for sample
equipment (switching
power supply polarity)
Data backup
function
(Automatic switching possible when using automatic
measurement function)
Backup battery life: 4 years (reference value at 25ºC)
: Constant ON/Auto OFF
Backlight automatic
OFF function
: Setting range: 1 sec. to 5 min., in 1-sec. increments
Effective only in automatic measurement
: Setting range: 1 sec. to 30 min., in 1-sec. increments
•Wait time from the completion of measurement to
power supply disconnection
Setting of measuring
time
1 min. to 30 min., in 1-min. increments
: MEM (internal RAM)/KEY (6 × 6 matrix touch panel)/
Self-test function
Measurement delay
(setting) function
LCD (front LCD panel)/LED /Buzzer
: Japanese or English
Language setting
System reset
: Clears all data including measurement conditions and
measurement data.
•Wait time from switching power supply polarity to
the start of measurement
Clears all saved measurement data.
Clears all saved condition setting data including panels.
•Wait time for operations other than switching polarity
: Effective in all measurement modes
Maximum value hold
function