3286-20  
CLAMP ON POWER HiTESTER  
Power measuring instruments  
All P o w e rfu l  
Effe c tive inp ut c urre nt le ve l  
1.00 A to 1000 A  
W
COS  
H
armonics  
VA var  
600.0kW max.(1ø)  
Up to 20th  
±1.000 (±90.0º)  
CAT III 600V  
http://www.hioki.co.jp/  
1
Functionality  
and Safety  
80 mm 20 mm  
busbar max.  
ø 55 mm  
Applications  
max.  
Use as a single-phase power meter  
1ø Power meter mode  
Voltage  
Effective power/ voltage/ current  
AC  
RMS  
V
600 Vrms max.  
Current  
This displays the result of  
vector calculations on  
single-phase power.  
AC  
PT/CT conversion ratio settings are not possible.  
RMS  
A
(effective value  
power factor)  
1000 Arms max.  
Power factor  
Apparent/ reactive power  
Power  
W
VA var  
600.0kW max.(1ø)  
Measure phase on a single-phase line  
1ø Power Factor meter mode  
The phase angle between the voltage waveform and the current  
waveform is found, and the power factor (cos ø) and reactivity  
(sin ø) are calculated and displayed.  
Phasedetection  
S
Normal/ Reverse/ Missing  
Power factor (cos ø)  
Powerfactor  
COS  
ø
±1.000 (±90.0º)  
Peak  
Reactivity (sin ø)  
Phase angle (ø)  
P
eak  
value  
1700 A pe ak max.  
Frequency  
The 3286-20 provides the following 2 types of power factor measurement. If  
the waveform includes harmonic distortion, there will be a difference between  
the two measurement values, and the method must be selected accordingly.  
Hz  
Power factor derived by phase discrimination (cos ø):  
1ø Power Factor meter and 3ø Power Factor meter mode  
From the phase information for the fundamental frequency component, cos ø is  
calculated. This is suitable for assessing the quality of a power supply.  
1000 Hz max.  
Harmonics  
Power factor (effective value power factor)  
derived from effective power ÷ apparent power: 1ø Power meter mode  
This is calculated from the power value including harmonic components  
(effective power ÷ apparent power). This is suitable for evaluating the  
performance of standalone load equipment and similar cases.  
H
armonics  
* It is not possible to calculate the effective power factor of a three-phase line.  
Up to 20th  
2
Packed into a Handheld Unit  
Helpful battery capacity check  
Whenever the unit is powered on, and  
while the record function is operating, the  
battery capacity can be checked to avoid  
battery exhaustion during measurement.  
Simple checking of three-phase lines  
3ø Power Factor meter mode  
For harmonic suppression  
Harmonic measurement function  
Harmonics effective value/  
total harmonic distortion  
Harmonics effective value/  
harmonic percentage  
Power factor (cos ø)  
Phase angle (ø) Reactivity (sin ø)  
Harmonic coefficients Total harmonic distortion Harmonic coefficients Harmonic percentage  
Ratio to case in which  
fundamental component is  
100%  
Ex. Fundamental  
component is 40.3 A  
Ex. THD-R is 65.9%  
Ex. 3rd order is 33.7 A  
The analysis values for coefficients for each of the harmonics  
from 1 to 20 of voltage or current can be displayed.  
Balanced three-phase power can also be displayed  
Effective power/ voltage/ current  
Apparent/ reactive  
power  
External output of data  
Output to PC  
Phase detection  
check  
(Optional 9636-01 required)  
Cord length:  
approx 2m  
The 3286-20's three-phase power measurement method calculates and  
displays the power values for a sine wave input at 50/60 Hz, assuming  
it is balanced and there is no distortion. Accurate measurement is not  
possible on a three-phase line if it is not balanced, for example when  
controlled by an inverter or thyristor.  
Since there is no integration function, it is not possible to measure total energy consumed (Wh).  
Data output connector RS-232C  
interface by optical insulating coupler  
Check power supply fluctuations  
1
OS:Windows 95, 98, NT4.0, 2000, Me*  
Connector: D-sub 9 pin  
Fetch interval: 6s/30s/1m/5m/10m.  
Buffer size: 32,700 max.  
Max. and min. value displays  
Using the record function, it is  
possible to check the amplitude of  
Care must be taken with regard to battery life  
fluctuations.  
*1. Windows is a registered trademark of Microsoft Corporation.  
Record elapsed time  
Example: 4 h 25 m  
Printer output to 9442 (optional)  
Printing method: Thermal serial  
dot matrix Paper width: 112 mm  
Printing speed: 52.5cps ●  
Power supply: 9443 AC  
ADAPTER or supplied nickel-  
hydride battery (capable of printing  
about 3000 lines on full charge from  
9443) Dimensions and mass:  
160W 66.5H 170D mm; 580 g  
The output is sent  
to the printer each  
time the 3286-20  
HOLD key is  
V
Waveform display is not available  
Care must be taken with regard to battery life  
Record peak value fluctuations  
pressed.  
(For about 1 sec.)  
Easily check peak value fluctuations.  
Please note that waveforms with amplitudes at or  
less than 250ms cannot be detected accurately  
9636  
9443-02  
RS-232C CABLE AC ADAPTER  
(For the EU)  
9443-03  
(For the America)  
+
0
Cord length  
approx 1.3m  
-
Waveform display is not available  
Not for use with  
Model 3286-20  
3
Basic specifications  
[Power factor/ Phase angle/ Reactivity measurement]  
:
Voltage, current, voltage/current peak, effective/ reactive /  
apparent power(Single-phase or 3-phase), power factor,  
Measurement items  
: Phase discrimination by phase detection (zero crossing)  
: 0.000 (lead) to 1.000 to 0.000 (lag)  
: 90.0˚(lead) to 0.0˚to 90.0˚(lag)  
Detection method  
Power factor (cos ø)  
Phase angle  
reactivity, phase angle, frequency, phase detection(3-  
phase), voltage/current harmonic levels(up to 20th  
ø55mm (2.16"), 80 mm 20 mm busbar max.  
LCD, digital (6000 counts)  
RMS (true root mean square value)  
NORMAL approx. 1 time/ sec, SLOW 1 time/ 3-sec  
at HARM meas. approx. 1 time/ 2-sec  
)
Measurable  
conductor diameter  
Display  
Rectification method  
Display update rate  
:
:
:
:
: 0.000 (lead) to 1.000 to 0.000 (lag)  
Reactivity (sin ø)  
[Frequency measurement] Effective in the voltage and current functions  
: 30.0 Hz to 100 Hz (at 100.0 Hz range)  
100 Hz to 1000Hz (at 1000 Hz range)  
Measurement range  
:
4.0 seconds or less (when input is changed from 0% to 90% of range.)  
: Voltage 10 Vrms-sine wave, Current 1 Arms-sine wave  
Analog response time  
Min. input level  
[Voltage/ Current/ Power measurement]  
[Wave peak measurement] Effective in the voltage and current functions  
: 150 (375 peak) / 300 (750 peak) / 600 (1020 peak)  
20 (50 peak) / 200 (500 peak) / 1000 (1700 peak)  
: Effective value of sine wave is within effective input  
permissible in the range and within circuit dynamic  
V
A
Measurement range  
AC Current  
20.00 A 200.0 A  
Range Table  
1000 A  
Effective Input Range  
150.0 V Single-phase  
3.000 kW 30.00 kW 150.0 kW  
6.000 kW 60.00 kW 300.0 kW  
6.000 kW 60.00 kW 300.0 kW  
*
3-phase  
[Harmonic measurement] Effective in the voltage and current functions  
(balanced load)  
: Level of each order, percentage of each order and total  
harmonic distortion (THD-F and THD-R)  
: Fundamental frequency 50 / 60 Hz  
: 1 cycle (50 / 60 Hz), Data points: 256 points  
: Rectangular  
Measurement items  
300.0 V Single phase  
*
3-phase  
Measurement range  
Window width  
Window type  
(balanced load) 6.000/12.00 kW 60.00/120.0 kW 600.0 kW  
12.00 kW 120.0 kW 600.0 kW  
600 V Single phase  
: Up to 20th  
Orders analyzed  
*
3-phase  
24.00 kW 240.0 kW 600.0/1200 kW  
(balanced load)  
[Other functions]  
*3-phase power is calculated and displayed on the basis of a balanced, 50/60 Hz, sine wave input.  
For apparent power and reactive power, the unit of watts in the above table is replaced by VA and var respectively.  
: Normal/ reverse/ missing (at 3-phase balanced load)  
Phase detection  
Record  
: 0.000 (lead) to 1.000 to 0.000 (lag); 1ø only  
: 1000 Arms cont.  
: 600 Vrms (insulated conductor)  
: Voltage: 10 V to 600 V, Current: 1 A to 1000A,  
: MAX. value and MIN. value (Effective in the voltage,  
current and effective / apparent power functions)  
: Displayed in % when the unit is powered on  
: Holds display  
Effective value P.F.  
Max. allowable current  
Max. usable circuit voltage  
Effective input range  
Battery capacity  
Data hold  
Power: 80 V to 600 V and 1 A to 1000 A  
: Approx. 10 minutes, buzzer sounds just before power is  
turned off, can be extended or cancelled  
Auto power off  
: Voltage: 0.6 Vrms, Current: 0.06 Arms  
Min. Display value  
Display indication range  
:
5 or less are zero-suppressed, and the upper limit is to 125%  
of the range setting (to 100% for the 1000 A range)  
2.5 or less (1000 A and 600 V range is 1.7 or less)  
: RS-232C interface by optical insulating coupler  
Data output  
(RMS value)  
:
Circuit dynamic  
Measurement accuracy(23 ˚C±5 ˚C (73˚F ±9˚F), Less than 80%rh., sine wave input, power factor = 1)  
: ±0.3 % rdg.±1 dgt. (at 100.0Hz range)  
Frequency  
[Voltage/ Current/ Power measurement]  
±1.0 % rdg.±1 dgt. (at 1000Hz range)  
: ±3.0 %rdg.±5 dgt. (45 Hz to 1 kHz)  
30 Hz to 45 Hz 45 Hz to 66 Hz 66 Hz to 1 kHz  
Wave peak  
Thermal coefficient  
:
Voltage and current: within ±0.1Accuracy/ ˚C (0 to 40˚C)  
±1.5 %rdg.±5 dgt. ±1.0 %rdg.±3 dgt. ±1.5%rdg.±5 dgt.  
Voltage  
Current  
Phase: within ±2  
˚ (0 to 40˚C)  
not rated  
±1.3 %rdg.±3 dgt. ±2.0 %rdg.±5 dgt.  
: Within 0.7 % in any direction from the center of sensor  
±
Conductor position  
External magneticfield  
Power factor influence  
: 400A/m corresponds to 1.00 A max.  
: ±10.0 % f.s. (cos ø=0.5)  
Accuracy guaranteed only for 50/60 Hz (COS ø=1)  
Single-phase: ±2.3 %rdg.±5 dgt.  
3-phase: ±3.0 %rdg.±10 dgt.(at balanced load)  
Power  
:
Harmonics  
Order  
1
Order  
9, 10  
Accuracy  
Accuracy  
±3.0 %rdg.±10 dgt.  
±3.5 %rdg.±10 dgt.  
±4.5 %rdg.±10 dgt.  
±5.0 %rdg.±10 dgt.  
±7.0 %rdg.±10 dgt.  
: ±3˚  
: ±±2 dgt.  
Phase angle  
Power factor (cos ø)  
2 to 6  
7, 8  
11 to 15  
16 to 20  
±10.0 %rdg.±10 dgt.  
General Specifications  
: Safety  
: Digital sampling method  
Applicable standards  
Measurement method  
Operating temperature  
EN61010-1:1992+A2:1995 CAT III600V  
EN61010-2-031:1996, EN61010-2-032:1995  
EN60529:1991 IP40  
: 0ºC to 40ºC, 80%rh or less, no condensation  
: -10ºC to 50ºC, no condensation  
Storage temperature  
Power supply  
(9V battery)  
:
6LR61 alkaline battery1  
(continuous operation max. 25 hours)  
: EMC  
6F22 manganese battery1  
(continuous operation max. 10 hours  
)
EN61326-1:1997+A1:1998+A2:2001  
: 5.55 kV AC between clamp and frame, between  
clamp and circuitry  
: Approx 100 W 287 H 39 D mm, Approx 650g  
(Approx 3.94(W )11.3(H)1.54(D), Approx 22.6 oz.)  
Withstand voltage  
(50/60 Hz, 1 minute)  
Dimensions, mass  
Options  
9635-01 VOLTAGE CORD  
9636-01 RS-232C PACKAGE  
3286-20 CLAMP ON POWER HiTESTER  
(Includes 9635 VOLTAGE CORD, 9245 CARRYING CASE, HAND STRAP)  
9635-01  
Cord length  
approx 3m  
9442  
9636  
PRINTER  
RS-232C CABLE (For 9442 printer)  
9443-02 AC ADAPTER (For 9442 printer, EU)  
9443-03 AC ADAPTER (For 9442 printer, America)  
1196  
RECORDING PAPER (For printer, 10 rolls)  
Cord length  
approx 3m  
When ordering the 9442 PRINTER, also order the 9636 RS-232C CABLE required for  
connection to the 3286-20, and 9443 AC ADAPTER.  
DISTRIBUTED BY  
HEAD OFFICE :  
Shanghai Representative Office :  
1704 Shanghai Times Square Office  
93 Huaihai Zhong Road  
Shanghai, 200021, P.R.China  
TEL +86-21-6391-0090, 0092  
FAX +86-21-6391-0360  
81 Koizumi, Ueda, Nagano, 386-1192, Japan  
TEL +81-268-28-0562 / FAX +81-268-28-0568  
E-mail: os-com@hioki.co.jp  
HIOKI USA CORPORATION :  
6 Corporate Drive, Cranbury, NJ 08512 USA  
TEL +1-609-409-9109 / FAX +1-609-409-9108  
E-mail: hioki@hiokiusa.com  
E-mail: info@hioki.cn  
All information correct as of Oct. 3, 2003. All specifications are subject to change without notice.  
328620E2-3XB-05K Printed in Japan