3193 POWER HiTESTER  
Power measuring instruments  
Printer is optional unit  
DC/0.5Hz to 1MHz broad-band POWER HiTESTER measures up to 6 systems simultaneously.  
Wide Spectrum Power Meter for  
Comprehensive Device Assessment  
3194  
The 3193 POWER HiTESTER is a multi-function power meter  
for use with single phase power lines to 3-phase, 4-wire circuits.  
MOTOR / HARMONIC  
HiTESTER  
Accommodating up to 6 units, it is not only capable of measuring up to  
6 single phase systems, but can simultaneously measure the input and  
output of a 3-phase inverter and provide effective power measurements.  
Additionally, it supports harmonic analysis and flicker measurement  
(optional), features which are essential for overall device assessment.  
Standard features include a GP-IB/RS-232C interface and 3.5-inch  
3194 MOTOR / HARMONIC HiTESTER  
provides analysis of high-order harmonics  
up to the 3000th order. Additionally, with the  
optional 9603-01 EXTERNAL SIGNAL  
INPUT UNIT installed, the HiTESTER can  
directly measure torque and rotation speed, an  
essential feature for evaluating the performance  
of inverter motors. This makes it easy to  
construct measurement systems.  
floppy disk drive, making it easy to feed data to a personal computer  
for processing and analysis. This unit is ideal for those requiring greater  
efficiency in electrical device assessment.  
1
Broad coverage, high accuracy, and well-developed interface  
Complete with functions that answer  
all your power measurement needs.  
Features  
Wide range of measurement functions  
High Basic Accuracy of ±0.2%  
Capable of measuring voltage, current, active /reactive /apparent  
Measurements of even greater precision can be obtained using the  
power, power factor, phase, frequency, and current, and of integrating  
optional 9600 to 9602 input unit, which provides a basic accuracy  
power according to polarity, the 3193  
(
or 3194  
)
also provides wave peak  
of ±0.1% rdg,±0.1% f.s. (With the 9602, the accuracy of the clamp-on sensor is a factor  
affecting total accuracy during power measurement.)  
and efficiency measurements that are essential to device assessment.  
Can Measure Motor Output  
A Variety of Interfaces for Differing Needs  
Connecting to a PC  
With the optional 9603 EXTERNAL SIGNAL INPUT UNIT , the  
HiTESTER can take analog input in from torque and revolution  
measurements and use that information to calculate motor output.  
The 3.5 inch floppy disk drive and RS-232C / GP-IB interface,  
provided as standard features, make it possible to connect the power  
meter directly to a PC, allowing efficient  
Measurement for Minute Stand-by Power also Available  
measurement, management and analysis of data.  
(by special-order product)  
The 9600 and 9601 input units have 10-times improved  
current sensitivity, and currents starting from the 20.000mA  
range can be measured. (Please ask for further information.)  
Connecting to a Recorder  
With 8 selectable D/A outputs and voltage, current and power analog/  
monitor output (current and voltage only) as standard features, the  
HiTESTER allows recording of changes and  
transient fluctuations in waveforms using a  
recording unit.  
High Visibility Color LCD  
Featuring a wide viewing angle, the  
color LCD displays a variety of items  
simultaneously, making it ideal for  
quickly grasping power usage on the  
system being measured. Expanded  
display is possible for any four  
selected items.  
Enlarged Indication  
Connecting to a Printer  
Data can be output to the optional 9604 PRINTER UNIT  
.
Print type  
Paper width  
Main  
:
Thermal line dot printing  
: 72mm  
:
printing of items measured, hard copy output  
functions  
of displayed screens, printout of meter settings,  
printout of various times (such as interval  
time, timer time, and realtime control time).  
Printouts are performed either automatically,  
upon input of an external control signal, or  
synchronized with an integrator.  
Harmonic and Flicker Analysis  
Harmonic and flicker analysis are possible when using the optional  
9605 HARMONIC / FLICKER MEASUREMENTS UNIT(3193 only)  
.
2
Other Analysis Functions  
Integration According to Polarity  
Peak Measurements Function  
Positive, negative, and total  
current and power can be  
integrated simultaneously for  
all channels. This makes it  
possible to grasp the income  
and outflow of power at a  
glance.  
Voltage and current wave peaks  
can be measured. The Peak  
Hold function can be used to  
find peak values and effective  
maximums for motor rush  
current waves.  
peak  
value  
Provides Analog and D/A output.  
Wave Monitor Output  
With the voltage and current  
ranges, waveforms are output  
as 1V full scale values,  
allowing waveforms to be  
monitored using devices such  
as recorders or synchroscopes.  
Analog (voltage, current, and  
Input wave  
50kHz  
Pulse  
effective power) and D/A  
outputs (any selected eight  
items) are output as a 5V range  
full scale value. (Except for the  
1000V range), 100ms response  
time can be obtained by using  
the FAST setting.  
Input wave  
60Hz, sinwave  
Voltage  
Monitor Output  
Active power  
Analog output  
current  
Monitor Output  
Active power  
D/A output  
20μs  
100ms  
3 Types of Averaging Function  
3 Types of Built-in Low Pass Filters  
Selectable cutoff frequencies (500/5k/300kHz) allow extraction of the  
frequency component of fundamental inverter waveforms and provide  
data compatibility with previous instruments.  
Select from time average, moving average and exponential average.  
3ch Frequency Measurement Function  
With the frequency ranges, LPF and HPF can be used in combination, allowing  
measurement of fundamental waveforms and carrier waveforms of inverters.  
Efficiency Calculation Function  
Choose from Three Types of Calculation Algorithms  
Three selectable algorithms are provided for calculating apparent power  
and reactive power, providing compatibility with previous devices  
Three efficiency calculations can be obtained simultaneously from measured voltage values.  
Choose from a variety of input units according to application  
Simultaneous Measurement of Multiple Systems  
Three types of input units are available, including the 9602 clamp  
input meter, which can be used with current levels exceeding 50A  
under live circuit conditions, as well as the 9600 and 9601 which  
accept direct input of up to 1000V/50A.  
Since all units are mutually isolated, the primary and  
secondary sides of devices or disparate power lines can be  
measured simultaneously. Simultaneous measurement of single  
phase 6 wire or 3 phase 2 wire systems which previously  
required multiple units, can now be handled with one. What's  
more, measurements of all devices can be taken at the same  
instant, providing a powerful tool for integrated, all-round  
device assessment.  
9600 · · · DC/0.5Hz to 1MHz wide band  
9601 · · · 5Hz to 100kHz, for AC only  
9602 · · · DC/0.5Hz to 200kHz clamp input  
A variety of clamp on sensors are available, including models 9270 to 9272 for  
AC only, and models 9277 to 9279 for AC/DC.  
1ch  
2ch  
3ch  
4ch  
5ch  
6ch  
1 ø2 W  
1 ø 2W  
1 ø 2W  
1 ø 2W  
1 ø 2W  
1 ø 2W  
1 ø 2W  
1 ø 2W  
1 ø 2W  
1 ø 2W  
1 ø 3W / 3 ø 3W  
1 ø 3W / 3 ø 3W  
1 ø 3W / 3 ø 3W  
1 ø 2W  
1 ø 3W / 3 ø 3W  
1 ø 3W / 3 ø 3W  
1 ø 2W  
1 ø 3W / 3 ø 3W  
3V3A (3  
3V3A (3  
3V3A (3  
ø
ø
ø
3W) / 3 ø 4W  
1 ø 2W  
1 ø 3W / 3 ø 3W  
3V3A (3  
1 ø 2W  
1 ø 2W  
1 ø 2W  
3W) / 3 ø 4W  
3W) / 3 ø 4W  
ø
3W) / 3 ø 4W  
9600 Frequency characteristics (at 15W[15V1A] range, response slow)  
Effect of common mode voltage  
frequency (Hz)  
frequency (Hz)  
3
Can measure all items at same point in time  
The Power Analysis Station  
Applications of 3193  
Example of Application With Power Converter  
Measure mixed AC and DC components with a single unit.  
Example of Assessment Trial of EV ( Electric Vehicle)  
Separate charge/generation  
integration capability.  
Measurement under live  
circuit conditions.  
(Clamp input)  
1, 3, 4ch  
3ch  
AC U /I /P, Primary to secondary  
efficiency  
U /I /P  
S
eparate charge / generation integration ±Wh  
2, 5 ch  
1, 2 ch  
Monitoring of DC U /I /P after DC/AC  
3 phase after DC/AC, AC U /I /P /Wh  
3194 Performs Comprehensive Evaluation of 3-phase Inverter Motors  
Capable of measuring carrier frequencies on the secondary side of inverters. Also allows analysis to be synchronized with motor rotation.  
1, 2, 3 ch (Primary)  
V/A/W/PF,  
4, 5, 6 ch (Secondary)  
V/A/W/PF and Motor  
output, Efficiency,  
Harmonic analysis  
Harmonic analysis  
3194 Specifications  
(Harmonic Waveform Analysis Function)  
:Single-phase 2- and 3- wire, three-phase 3- and 4-wire  
:Up to 3 channels from channels 1 to 6  
Measurement lines  
No. of channels  
Measurement range  
:Fundamental frequency: 10 Hz to 4.5 kHz  
Harmonic Waveform Analysis Functions  
Measurement method  
A/D resolution  
Windowing type  
Crest factor  
:PLL or external clock  
:12 bits  
:Rectangular tiling (with gap between windows)  
:Up to 2.5 (voltage, current)  
Harmonic analysis to the 3000th order  
PLL source  
:U, I or external synchronization  
Ext. synchronization  
signal  
Measurement items  
:Input to a rear panel control terminal on the 3194 main unit  
Input level: 1V to 10 Vrms, Division function: 1/1 to 1/255  
:RMS voltage, RMS current, effective power value,  
frequency, ±U peak, ±I peak  
Harmonic wave  
measurement items  
Screen displays  
Wiring conversion  
Sort function  
Averaging function  
Frequency ranges :  
:Harmonic level, percentage and phase angle of harmonic  
wave, Total harmonic distortion (THD-F, THD-R)  
:List, graph, vector and waveform displays  
: Δ-Y voltage conversion, Y-Δ voltage conversion  
:Sorts according to decreasing order of analysis, displays up to 50th order  
:Index average with time constant of 1.5 sec  
9603-01 Specifications  
Torque, rotation speed, and motor power can be measured by inputting analog  
torque and r/m signals to the 9603-01 EXTERNAL INPUT UNIT. Further, an  
input terminal is provided for use with a strain gauge-type torque sensor.  
Fundamental Sampling  
Frequency speed (Hz)  
Window  
Width  
Analysis Order  
10 - 17.5 Hz f × 8192  
17.5 - 35 Hz f × 8192  
35 - 70 Hz f × 8192  
1 waveform 3000 ( 10kHz or less)  
1 waveform 3000 ( 10kHz or less)  
1 waveform 3000 (100kHz or less)  
9603-01 EXTERNAL INPUT UNIT  
Input channels  
: 2 channels  
Ch A  
Ch B  
: Strain gauge sensor or DC voltage input (BNC)  
: Pulse or DC voltage input (BNC)  
70 - 140 Hz f × 4096 2 waveforms 1500 (100kHz or less)  
PLL  
Strain gauge input specifications (dedicated connector)  
Applicable converter : Strain gauge type converter (bridge resistance 350Ω - 1.5kΩ)  
Measurement range : 1 mV/V / 1.5 mV/V / 2 mV/V  
Synchronization 140 - 280 Hz f × 2048 4 waveforms  
800 (100kHz or less)  
400 (100kHz or less)  
200 (100kHz or less)  
100 (100kHz or less)  
50 (100kHz or less)  
Ranges  
280 - 560 Hz f × 1024 8 waveforms  
Measurement accuracy : ±0.1%rdg. ±0.065%f.s.  
560 - 1120 Hz f × 512 16 waveforms  
1120 - 2240 Hz f × 256 32 waveforms  
2240 - 4500 Hz f × 128 64 waveforms  
Connector  
DC voltage input specifications (ch A / ch B common BNC connector)  
Input resistance : 200kΩ ±5% (differential)  
: PRC03-23A10-7F (manufactured by Tajimi)  
Measurement range : ±1.0000 / ±5.0000 / ±10.000V  
Measurement accuracy : ±0.1%rdg. ±0.1%f.s. (23˚C ±5˚C, not higher than 80%rh)  
Pulse input specifications (ch B, BNC connector)  
Frequency measurement : 1Hz to 100 kHz  
50 × 8192  
Fixed  
Fixed clock  
----  
2 waveforms 3000 (100kHz or less)  
Common specifications  
Note 1: Analysis order accuracy is restricted to the frequency in brackets.  
Note 2: With PLL synchronization in the range of 10 to 35 Hz, an anti-aliasing filter of about  
15 kHz is used, and with PLL synchronization in the range of 35 Hz to 4.5kHz, an  
anti-aliasing filter of about 120 kHz is used.  
Analog output  
: ±5V f.s. (Output accuracy : Measurement accuracy ±0.2%f.s)  
4
Applications of 3193  
Support up to Harmonic / Flicker analysis when using the optional 9605 HARMONIC / FLICKER MEASUREMENTS UNIT  
.
Graph Display of Harmonics  
Voltage, current and power  
can be analyzed and displayed  
by bargraphs of harmonic  
amplitude, content and phase  
angle.Voltage, current and power  
can be displayed simultaneously  
for a single channel, or a single  
parameter can be displayed  
simultaneously for each of three  
channels.  
List Display of Harmonics  
The harmonic list display  
shows the amplitude, pro-  
portion, phase angle and  
distortion for each harmonic  
of voltage, current and power.  
Displaying only proportion,  
or two parameters simul-  
taneously, such as amplitude  
and phase angle, is selectable.  
Vector Display of Harmonics  
The harmonic vector display  
shows the voltage, current and  
phase angle for each harmonic,  
making clear the voltage-  
current phase relationship.  
Waveform Display  
The waveform display shows  
one cycle of the voltage and  
current waveforms. RMS and  
peak values can be displayed  
along with voltage and current  
waveforms, or voltage and  
current waveforms for up to  
three channels can be displayed  
at the same time.  
Flicker Measurement Display  
Displays data during measurement  
in real-time. Display can also be  
switched to D measurement and  
"Pst" value.  
Monitor Display  
The relative "d" voltage change  
∆V/V and the instantaneous  
flicker value "S(t)" can be  
displayed in a time series,  
so past variations are clearly  
displayed.  
9605 Specifications (optional)  
:
Every 1 window (except during communications with other devices)  
Update rate  
: List, graph, vector and waveform  
: Harmonic levels; at 45 to 66 Hz  
Voltage/current: ±0.5%rdg.±0.05% f.s.  
Screen displays  
: Installs in the 3193 main unit  
Installation  
3
Accuracy*  
: Single-phase 2- and 3-wire, three-phase 4-wire  
: Up to 3 channels within channels 1 to 6, depending  
on 3193 wiring mode  
Measurement lines  
No. of channels  
Active power: ±1.0%rdg.±0.1% f.s.  
For 45 to 66-Hz fundamental, effective input is 0.1 to 110%  
of range, total error of harmonic current when PLL is locked  
is 5% of the limit value or within 0.2% of the rated current  
of the device under test (no limitation with 9602).  
: Floppy disk, RS-232C, GP-IB, printer  
Output functions  
1
Harmonic Waveform Analysis Functions (IEC61000-3-2* )  
:
Fundamental frequency: 1 to 440 Hz  
PLL system (5 to 440 Hz , external clock system (1 to 5 Hz)  
Measurement range  
2
)
Flicker Measurement Function (IEC61000-3-3  
*
)
: Up to 50th harmonic (with 1 to 250 Hz fundamental)  
: 16 cycles (for 40 to 70 Hz fundamental)  
: Rectangular tiling (no gap between or overlap of windows)  
: 512 points (for 40 to 70 Hz fundamental)  
: Up to 4 (current), and up to 3 (voltage)  
Orders analyzed  
Window width  
Windowing type  
Amount of data analyzed  
Crest factor  
:
:
Fundamental frequency; 45 to 66 Hz, PLL synchronization system  
dc (relative constant voltage change), d max (max. relative constant  
Measurement range  
Analysis items  
voltage change), d(t)200ms (relative voltage change per time), P0.1/  
P1s/ P3s/ P10s and 30s (cumulative probability), Pst (short-term flicker  
value), Plt (long-term flicker value)  
:
Harmonic level, percentage and phase angle of each order  
Measurement items  
of harmonic wave for each of voltage, current and power.  
Total up to 50th harmonic (of 40 to 70 Hz fundamental) for  
voltage, current and power. Total harmonic distortion for  
voltage and current (THD-F and THD-R)  
:
:
Measured value, CPF, Pst list, Monitor  
Screen displays  
3
Accuracy*  
Within ±5% rdg. of limit on the limit curve line (Pst = 1) specified  
for voltage fluctuation in IEC60868-0 (d measurement is the same)  
RMS voltage 0.5%rdg. 0.05%f.s. (45Hz~66Hz)  
*2. Decision function not included.  
Measurement of voltage, current, active power, peak voltage  
and peak current values of the fundamental  
±
±
*3. The reading accuracy of the input unit must combined with the analysis accuracy shown above. When  
used with a clamp sensor, accuracy and frequency characteristics of the clamp must be added to the  
analysis accuracy above.  
*1. The 9605 does not support limit values, so pass/fail decisions based on limit values, classification  
of special waveforms (A to D) and 2.5-minute measurement function for measuring transient harmonic  
currents are not supported.  
5
The 9600 and 9601 are also the products for stand-by power measurement with 10-times  
improved sensitivity for current ranges. Please ask for detailed specs.  
Optional Input Unit Specifications  
9600 AC/DC DIRECT INPUT UNIT  
9602 AC/DC CLAMP INPUT UNIT  
Voltage  
Current  
Active power  
Voltage  
Current  
Active power  
6.0000/15.000/30.000/  
200.00/500.00mA/  
Depends on  
6.0000/15.000/30.000/ 500.00mA to 500.00A  
Depends on  
Measurement  
range  
60.000/150.00/300.00/ 1. 0000/2.0000/5.0000/ combination of voltage 60.000/150.00/300.00/  
(
Depends on clamp-on combination of voltage  
600.00V/1.0000 kV  
10.000/20.000/50.000  
65Arms/100 A peak  
A
and current ranges  
600.00V  
sensor)  
and current ranges  
1000Vrms/1500 V peak  
650Vrms/850Vpeak (Depends on clamp-on sensor)  
Max.operating input(55Hz)  
Crest factor  
Lower of either (measured range 6) / measured value  
or maximum permissible rated peak / measured value  
Lower of either (measured range 6)/ measured value or  
maximum permissible rated peak/measured value  
2MΩ±5%  
1mΩ max.  
2MΩ±5%  
200kΩ±5%  
Input resistance  
(Accuracy assured at 23˚C±5˚C (73˚F ±9˚F) at 80% R.H., power factor = 1, sine wave input, in-phase voltage 0, after DMAG)  
Accuracy  
DC  
±0.1%rdg.±0.2%f.s.  
±0.5%rdg.±0.5%f.s.  
±0.2%rdg.±0.2%f.s.  
±0.1%rdg.±0.2%f.s.  
±0.1%rdg.±0.1%f.s.  
±0.1%rdg.±0.2%f.s.  
±0.3%rdg.±0.3%f.s.  
±0.1%rdg.±0.2%f.s.  
±0.5%rdg.±0.5%f.s.  
±0.2%rdg.±0.2%f.s.  
±0.1%rdg.±0.2%f.s.  
±0.1%rdg.±0.1%f.s.  
±0.1%rdg.±0.2%f.s.  
±0.5%rdg.±0.5%f.s.  
0.5 to 1Hz  
1
to 10Hz  
10 to 45Hz  
45 to 66Hz  
66Hz to 10kHz  
10 k to 50kHz  
Less than 5 A Greater than 5 A Less than 5 A Greater than 5 A  
±0.5%rdg.  
±0.5%f.s.  
±0.5%rdg.  
±0.5%f.s.  
±0.5%rdg.±0.5%f.s.  
±0.5%rdg.±0.5%f.s.  
±1.5%rdg.±0.5%f.s.  
±2.0%rdg.±1.0%f.s.  
±
2.5%f.s.  
±
5.0%f.s.  
±
0.5%rdg.  
±
0.5%f.s.  
±0.3%rdg.±0.5%f.s.  
50 k to 100kHz  
100k to 300kHz  
300k to 400kHz  
400k to 500kHz  
±0.5%rdg.  
±0.5%f.s.  
±1.0%rdg.  
±1.5%f.s.  
±5%f.s.  
±
10.0%f.s.  
±15%f.s  
.
(up to 200kHz)  
±30%f.s  
.
(up to 200kHz)  
Compatible  
9279  
±1.0%rdg.  
±0.5%f.s.  
±1.0%rdg.  
±2.5%f.s.  
9709  
Clamp (Optional)  
±2.0%rdg.  
±1.0%f.s.  
±2.0%rdg.  
±2.5%f.s.  
9270·9271  
9272  
9277/9278  
With the 9602, the accuracy of the clamp-on sensor is a factor affecting  
total accuracy during current measurement, as are phase and frequency  
characteristics. See page 7 for details on the clamp-on sensor specifications.  
±
±
15.0%f.s.  
30.0%f.s.  
±
10.0%f.s.  
15.0%f.s.  
±
10.0%f.s.  
15.0%f.s.  
500k to 700kHz  
700k to 1 MHz  
±
±
9601 AC DIRECT INPUT UNIT  
Assured Accuracy Range for Input Frequency of the 9600  
9601 and 9602 each have assured ranges for input frequency.  
Voltage  
Current  
Active power  
200.00/500.00mA/  
Depends on  
Measurement  
range  
60.000/150.00/300.00/6  
00.00V/1.0000 kV  
1. 0000/2.0000/5.0000/ combination of voltage  
10.000/20.000/50.000  
65Arms/100 A peak  
A
and current ranges  
1000Vrms/1500 V peak  
Max.operating input(55Hz)  
Crest factor  
Lower of either (measured range 6) / measured value  
or maximum permissible rated peak / measured value  
2MΩ±5%  
1mΩ max.  
Input resistance  
(Accuracy assured at 23˚C ±5˚C(73˚F ±9˚F) at 80% R.H., power factor = 1, sine wave input, in-phase voltage 0  
)
Accuracy  
±2.5%f.s.  
±1.0%f.s.  
5
to 10Hz  
10 to 20Hz  
20 to 45Hz  
45 to 66Hz  
66Hz to 5kHz  
5k to 10kHz  
10k to 20kHz  
20k to 50kHz  
50k to 100kHz  
±0.1%rdg.±0.2%f.s.  
±0.1%rdg.±0.1%f.s.  
±0.1%rdg.±0.2%f.s.  
±0.2%rdg.±0.4%f.s.  
±1.0%f.s.  
±2.5%f.s.  
Frequency (Hz)  
±10.0%f.s.  
Note 1: Assured accuracy ranges for different response settings are as follows: FAST (0.1 sec) to DC and greater than 50 Hz, MID (0.8 sec) to DC and greater than 10 Hz,  
SLOW (5.0 sec) to DC or greater than 0.5 Hz  
Note 2: Assured accuracy ranges for combined mode measurement are 10 Hz or greater for the AC mode, and DC only for the AC+DC mode or DC mode.  
(Indicated only for single phase, 2 wire and 3 phase, 3 wire (3V3A).Two additional calculation algorithms can be selected for apparent/reactive power)  
Calculation algorithm  
Voltage  
Active power Apparent power Reactive power  
Current  
Power factor  
s1 P1/S1  
Phase  
cos-1  
1ø2W  
λ1  
=
U
1
I
1
P
1
S
1
=
U
1
I  
1
ø
1
=
s
1
λ
1
S
U
M
3ø3W  
P
1·2·3  
+
+
+ +  
I I I  
3
P
1
·
2
·
3
=
P
1
+P  
2
ø
1·2·3=su cos-1  
λ
1·2·3  
U
1
U
2
U
3
1
2
3
su  
=
λ1  
·2·3  
=
=
I1·2·3  
U1·2·3  
(3V3A)  
S
1·2·3  
3
Note 1: The above calculation algorithm is for a single phase, 2 wire input to ch 1, and 3 phase, 3 wire input to ch 1/2/3 (3 voltage, 3 current).  
Note 2: The "s" before each power factor or phase operation indicates the lead or lag of current phase in relation to voltage. The "-" sign means current phase leads voltage  
and when there is no symbol, it lags. "su" is "-" when the sum of reactive power is negative and "+" (but unsigned) when it is positive.  
6
Basic specifications  
Measurement line  
: Single phase 2 wire, single phase 3 wire, 3 phase 3 wire  
(3V3A is possible), 3 phase 4 wire  
[Motor output (Pm) measured] 9603 (optional) external input unit required  
Measurement method  
:
Digital calculation from measured voltage or pulse signal  
Effective when 9603 ch A is torque (any of N/ m, mN/m, kN/m, kgf/m,  
:
When using 9600, 9601, 9602 (optional)  
Measurement item  
kgf/cm), and ch B is r/m  
Voltage, current, voltage/current peak, effective/reactive/apparent power,  
power factor, phase, frequency, current/power integration, load rate, efficiency.  
Display indication range  
Calculation algorithm  
:
:
0.1% to 130% of 9603 voltage range (polarity not indicated)  
Set to required format  
When using the 9603 (optional)  
Voltage, torque, r/min, frequency, motor output.  
When using the 9605 (optional)  
[Efficiency measurement]  
Calculable factors : Maximum of 3 formats  
Calculated items  
:
P for each input unit, Pm when combined with 9603  
(Items measured with 9605 not allowed)  
Harmonic, waveform, voltage fluctuation / flicker measurement function.  
Display indication range  
:
At the lowest range in the DC mode of Models 9600 & 9602: 0.2% to 130%  
At the lowest range in the AC+DC mode of Models 9600 & 9602: 0.5% to 130%  
At the 200mA range of Model 9601: 0.5% to 130%  
At 0.1% to 130% of all other ranges.  
All range is zero suppressed at less than lower % value.  
Valid input range for voltage, current, and power is 0.5% to 110%  
[D/A output]  
Number of channels : 8ch (12 bit D/A converter polarity + 11bit)  
Output impedance : 100Ω±5%  
Output items  
Output voltage : DC±5V/f.s.  
Output update rate : 16 times/sec  
: Outputs 8 arbitrarily selected items  
: 6.4 inch TFT color LCD (640 × 480 dot)  
Display  
Display resolution  
Rectification method  
:
:
99999 count (except with integration), 9999999 count (with integration)  
Switchable between RMS (true root mean square value) nd MEAN  
(average rectified RMS indication) When combined mode DC is  
selected, its not possible to switch between them.  
a
[FDD]  
.
: 3.5-inch, 2HD (1.2MB/1.44MB)  
: MS-DOS  
Compatible media  
Format  
Functions  
: 8 times/sec.  
Display update rate  
Combined mode  
:
Function for saving and loading settings, and saving  
measured values. Function for outputting measured values.  
Floppy disk formatting function, file renaming and deleting  
: DC, AC + DC, AC (AC only when used in combination  
with 9601 or 9602 + AC clamp-on sensor)  
: FAST (0.1 sec.), MID (0.8 sec.), SLOW (5.0 sec.)  
(Time required for stabilization to within ±1% when input is changed  
from 0% to 90%, or 100% to 10% of range.)  
: OFF / 500Hz / 5kHz / 300kHz (-3dB) For 9601, 5k  
/ 300kHz not available.  
Analog response time  
[Interface]  
GP-IB  
RS-232C  
:
:
Conforms to IEEE-488.1 1987, with reference to IEEE-488.2 1987  
Start-stop synchronous, with baud rate of 2400 or 9600 bits/sec  
Low pass filter  
: OFF / 200Hz (-3dB)  
: Voltage / current / active power  
DC ±5V f.s (1000V range is DC ±3.333 V f.s.)  
Monitor output : Voltage / current: 1Vrms f.s. (1000V range is 0.6667 Vrms f.s.)  
Polarity detection regulation filter  
[External Control]  
Analog output  
:
Integration start / stop control, Integration data reset, External  
A/D (For display update when power meter display is in hold mode)  
Functions  
,
Manual print control, FD save control  
[Voltage/ Current/ Active power measurement]  
Control signal level  
:
From 0 / 5V logic signal or short//open contact signal  
Measurement range : See Page 5 specifications for individual input units  
[Other functions]  
[Integration measurement]  
: PT/CT ratio Set range 0.0001 to 10000  
: Time average (set interval time, timer time, and average of  
realtime control time)  
Moving average (number of samples: 8/16/32/64)  
Exponential average (attenuation factor: 8/16/32/64)  
Scaling  
Averaging  
: 64 times/sec  
Number of measurements  
:
0 to±9999999 TAh / TWh (integration time up to 10,000 hours)  
Measurement range  
[Power factor/ Phase angle measurement]  
Multilingual display : Japanese/English screen display switching  
Interval control time (10 sec to 100 hours) in 10 sec increments  
When used in combination with FDD or printer, auto select increments  
Measurement range  
: -1.0000 (lead) to 0.0000 to 1.0000 (lag)  
-180˚(lead) to 0.00˚to 180.00˚(lag)  
Set time (all types)  
:
(
)
[Frequency measurement]  
Timer control time (1 min to 10000 hours) in 1 minute increments  
Realtime control time, 1 minute increments  
: Max. 3ch (selection of voltage or current for arbitrary channel)  
Number of channels  
Effective input range : 0.5 Hz to 2 MHz  
[Harmonic / Flicker measurement] 9605 (optional) required  
: Auto / 50Hz/ 500Hz / 50 kHz / 2 MHz  
Measurement range  
: See Page 4 specifications for 9605  
Measurement item  
[Wave peak measurement]  
Measurement items  
Effective Input Range  
:
:
Select either voltage or current for each unit(Shows absolute value of max.)  
Effective value of sine wave is within effective input permi-  
ssible in the range  
Measurement accuracy(23˚C±5˚C (73˚F ±9˚F), Less than 80% rh, warm up time greater than 1 hour, sine wave input, power factor = 1, in-phase voltage = 0)  
: ±1dgt for calculations of each measured value  
Max. 7 dgt with respect to values calculated from measure-  
ments of items substituted into algorithm  
: Within ±0.03% f.s/ ˚C  
: Within ±0.05% f.s  
: Per accuracy table on page 5  
Motor output  
Efficiency  
V, A, W  
Apparent / reactive  
:
±
:
±1 dgt. with respect to calculation from measured value (U, I, P)  
sum value is max.  
±
3dgt.  
power  
:
:
:
:
±
1dgt with respect to values calculated from measurements ( I, P)  
Integration  
Thermal coefficient  
Effect of in-phase  
voltage  
Effect of power factor  
Actual time  
D/A output  
Analog output  
Monitor output  
Max.  
Max.  
±
3dgt with respect to values calculated from measurements (U, I, P)  
3dgt with respect to values calculated from measurements (U, I, P)  
Power factor  
Phase angle  
Frequency  
±
(1000 Vrms, 50/60Hz, between shorted voltage input terminals and case)  
±
0.1% rdg.  
input between 10% to 130% of U/I range  
1% (at 0.5Hz to 1kHz) 2% (at 1kHz to 10kHz)  
(at 10kHz to 100kHz)  
±1 dgt. (0˚C to 40˚C(32˚F to 104˚F), for sine wave  
: ±0.15% f.s (power factor = 0)  
)
: ±25ppm ±1dgt. (0 to 40˚C (73˚F to 41˚F)  
: Display accuracy - ±0.2% f.s  
: Display accuracy - ±0.2% f.s  
)
:
±
,
±
, ±10%  
Wave peak  
: Display accuracy - ±0.2% f.s (less than 100kHz)  
Display accuracy - ±3dB (100k to 1MHz)  
General Specifications  
: Indoors, altitude to 2000 m  
: Safety  
EN61010-1: 2001  
: EMC  
Location for use  
Certifications  
:
Power meter, 0˚Cto 40˚C (32˚F to104˚F), rh below 80%(no condensation)  
˚C to 40˚C (41˚F to 104˚F), rh below 80%  
Ambient use humidity  
When using FDD / printer,  
5
:
-10˚C to 50˚C (14˚F to 122˚F)  
,
rh below 80% and no condensation  
EN61326  
EN61000-3-2: 2000  
EN61000-3-3: 1995+A1: 2001  
AC100V/120V/200V/230V (switched automatically), 50/60Hz  
: 150VA max.  
Approx 430 W 150 H 370 D mm, Approx13 kg  
(Approx 16.93ʺ(W ) 5.91ʺ (H)14.57ʺ (D), Approx 458.6 oz.)  
(in configuration including 9600 5 6ch, 9603, 9604)  
: 1997+A1: 1998+A2: 2001 class A  
Ambient storage humidity  
Insulation resistance  
: More than 100MΩ from DC500V  
Greater than 50M( at DC500V between U/I input terminal and  
unit case, between U/I input terminal and power supply plug  
(for the 9600/9601), between U input terminal and clamp input  
terminal, between U input terminal and unit case, and between  
U input terminal and power supply plug (for 9602)  
:
Power supply  
Maximum rated power  
Dimensions, mass  
:
:
AC5.55kV between U/I terminal and unit case, between U/I  
terminal and power supply plug (for 9600 and 9601), between U  
terminal and clamp input terminal, between U terminal and unit  
case, between U input terminal and power supply plug (for 9602)  
Withstand voltage  
(50/60 Hz, 1 minute)  
(Not including projections such as terminals, feet, and handles)  
: Power cord 1, ground adapter (3P to 2P) 1,  
connector 1  
Accessories  
7
CLAMP-ON SENSOR(Optional) Specifications  
To use the Clamp-On sensor, be sure to order the factory option 9602 AC/DC clamp input unit.  
AC  
AC / DC  
20A  
20 / 200A  
9270  
9271  
9272  
9277 / 9278  
9279  
3193 or 3194  
ø20mm  
ø20mm  
9290  
5Hz to 50kHz  
DC to 100kHz  
9277  
500A  
ø40mm  
DC to 20kHz  
200A  
ø20mm  
5Hz to 50kHz  
1500A  
ø40mm  
+
9602  
80mm-wide busbar  
20Hz to 4kHz  
CT ratio 10 : 1  
20A/200A  
500A  
9709  
ø
46mm  
,
5020mm busbar  
ø36mm  
5Hz to 10kHz  
DC to 100kHz  
Ordering information  
3193 POWER HiTESTER (main unit only)  
3194 MOTOR / HARMONIC HiTESTER (main unit only)  
Measurements cannot be taken with a 3193 POWER HiTESTER and 3194 MOTOR / HARMONIC  
HiTESTER unit only.  
A factory option input unit must be purchased.  
1ch  
2ch  
3ch  
) 1 ø 2W (  
4ch  
) 1 ø 2W (  
) 1 ø 2W (  
5ch  
) 1 ø 2W (  
6ch  
) 1 ø 2W (  
) 1 ø 2W (  
) 1 ø 2W (  
Notes on input unit selection  
1 ø 2W (  
) 1 ø 2W (  
)
)
)
Pattern A  
Pattern B  
Pattern C  
Pattern D  
Pattern E  
Pattern F  
Pattern G  
Use the same input unit for a particular  
measurement line.  
1 ø 3W / 3 ø 3W (  
1 ø 3W / 3 ø 3W (  
1 ø 3W / 3 ø 3W (  
×
×
×
2 ) 1 ø 2W (  
) 1 ø 2W (  
2 ) 1 ø 3W / 3 ø 3W (  
2 ) 1 ø 3W / 3 ø 3W (  
×
×
2 ) 1 ø 2W (  
Install units in succession starting  
from channel 1.  
2 ) 1 ø 3W / 3 ø 3W (  
) 1 ø 2W ( ) 1 ø 2W (  
×3 ) 1 ø 3W / 3 ø 3W ( 2 ) 1 ø 2W (  
×3 ) 3V3A (3 3W) / 3  
×2 )  
3V3A (3  
3V3A (3  
3V3A (3  
ø
ø
ø
3W) / 3  
3W) / 3  
3W) / 3  
ø
ø
ø
4W (  
4W (  
4W (  
×3 ) 1 ø 2W (  
)
)
For the 9603 or 9603-01, only one unit  
can be installed.  
×
ø
ø
4W (  
×3 )  
When the 9602 is selected, use the  
optional clamp-on sensor.  
(
) : 9600, 9601, 9602 can be selected.  
Options that can be installed at the factory  
(specify at time of order)  
Options  
*
*
*
9270  
9271  
9272  
9277  
9278  
9279  
9709  
9290  
9232  
CLAMP ON SENSOR (20A AC)  
9600  
9601  
9602  
9603  
AC/DC DIRECT INPUT UNIT  
AC DIRECT INPUT UNIT  
CLAMP ON SENSOR (200A AC)  
CLAMP ON SENSOR (20/200A AC)  
UNIVERSAL CLAMP ON CT (20A AC/DC)  
UNIVERSAL CLAMP ON CT (200A AC/DC)  
UNIVERSAL CLAMP ON CT (500A AC/DC)  
AC/DC CURRENT SENSOR (500A AC/DC)  
CLAMP ON ADAPTER (1500A AC, ratio10:1)  
RECORDING PAPER (10 m, 10 roll, For 9604)  
AC/DC CLAMP INPUT UNIT  
EXTERNAL SIGNAL INPUT UNIT (3193 only)  
9603-01 EXTERNAL SIGNAL INPUT UNIT (3194 only)  
*
*
9604  
9605  
PRINTER UNIT  
HARMONIC / FLICKER MEASUREMENTS UNIT  
(3193 only)  
*The voltage cable is not supplied. Contact your dealer when it is  
necessary to use the clip type leads.  
*No CE marking  
DISTRIBUTED BY  
HEAD OFFICE :  
Shanghai Representative Office :  
1704 Shanghai Times Square Office  
81 Koizumi, Ueda, Nagano, 386-1192, Japan  
TEL +81-268-28-0562 / FAX +81-268-28-0568  
93 Huaihai Zhong Road  
E-mail: os-com@hioki.co.jp  
Shanghai, 200021, P.R.China  
TEL +86-21-6391-0090/0092  
FAX +86-21-6391-0360  
E-mail: info@hioki.cn  
HIOKI USA CORPORATION :  
6 Corporate Drive, Cranbury, NJ 08512 USA  
TEL +1-609-409-9109 / FAX +1-609-409-9108  
E-mail: hioki@hiokiusa.com  
All information correct as of Feb.13, 2006. All specifications are subject to change without notice.  
3193E6-62M-04K Printed in Japan