MQF500E-12S [ARCH]

AC-DC ITE & Medical Switching Power Supply;
MQF500E-12S
型号: MQF500E-12S
厂家: ARCH ELECTRONICS CORP.    ARCH ELECTRONICS CORP.
描述:

AC-DC ITE & Medical Switching Power Supply

文件: 总5页 (文件大小:294K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
VER : A_7  
update : 2018.06.06  
AC-DC ITE & Medical Switching Power Supply  
MQF500E SERIES  
500 Watts  
KEY FEATURES  
Enclosed Medical Switching Power Supply  
Remote ON/OFF Function  
Standby 5V@1A  
High Efficiency up to 92%  
With P.F.C. Function >0.94  
Ultra Compact Size: 5.5 x 3.25 x 2.48 Inches  
3-Year Product Warranty  
(except for 15S)  
ELECTRICAL SPECIFICATIONS  
All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated.  
Model No.  
MQF500E-12S  
500 W  
MQF500E-15S  
MQF500E-24S  
MQF500E-48S  
Max Output Wattage (W)  
Voltage  
90-264 VAC or 127-370 VDC  
47-63 Hz  
Frequency (Hz)  
Current (Full load)  
<6.3 A max. (115 VAC) / <3.15 A max. (230 VAC)  
< 40 A max. (115 VAC) / < 80 A max. (230 VAC)  
< 0.1 mA max. (Input-Output)  
Input  
Inrush Current (<2ms) (Clod Start)  
Leakage Current  
Power Factor (at 230 VAC)  
Voltage (V.DC.)  
PF>0.94 at Full Load  
12V  
15V  
24V  
20.8  
48V  
Voltage Accuracy  
±2%  
±4% Output Voltage  
41.5  
Voltage Adj. Range (V.DC)  
Current (A) (max.)  
33.3  
10.41  
Line Regulation (115-264 VAC)  
Load Regulation (10-100%) (typ.)  
Minimum Load  
±0.5%  
Output  
±1%  
3%  
Maximum Capacitive Load  
Ripple & Noise (typ.)  
5,000µF  
160mV  
3,750µF  
160mV  
89%  
2,500µF  
240mV  
91%  
1,250µF  
480mV  
92%  
Efficiency (at 230 VAC)  
Hold-up Time (at 115 VAC)  
Over Power Protection  
Over Voltage Protection  
Overt Temperature Protection  
89%  
8 ms min.  
Auto recovery  
Auto recovery  
Auto recovery  
Protection  
Isolation  
Protection level 1 (nominal)Continuous, Auto recovery  
Short Circuit Protection  
Protection level 2 (instantaneous high current)Latch  
Input-Output (V.AC)  
4000VAC or 5656VDC  
Input-PE  
(V.AC)  
2000V  
Output-PE .(V.AC)  
Operating Temperature  
Storage Temperature  
1500V  
-30°C…+70°C (with derating)  
-35°C…+85°C  
±0.03%/°C ( 0~50°C )  
Temperature Coefficient  
±0.06%/°C ( -30~0°C )  
Environment  
Altitude During Operation  
Humidity  
5000m  
95% RH  
Atmospheric Pressure  
MTBF  
56 kPa to 106 kPa  
>160,000 h @ 25°C (MIL-HDBK-217F)  
10~500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes.  
Vibration  
http://www.archcorp.com.tw  
TEL: +886-2-26989508 FAX: +886-2-26981319  
-1-  
VER : A_7  
update : 2018.06.06  
AC-DC ITE & Medical Switching Power Supply  
MQF500E SERIES  
500 Watts  
ELECTRICAL SPECIFICATIONS  
All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated.  
Model No.  
MQF500E-12S  
MQF500E-15S  
MQF500E-24S  
MQF500E-48S  
Dimension (L x W x H)  
Weight  
5.5 x 3.25 x 2.48 Inches ( 139.7 x 82.55 x 62.9 mm ) Tolerance 0.5 mm  
690 g  
Others: UL / IEC / EN 60601 3.1rd Edition & UL / IEC / EN 60950 AM2  
Physical  
Safety  
EMC  
Approval  
15S:  
UL / IEC / EN 60601 3.1rd Edition (In Progress)  
Conducted and Radiated EMI  
EMS  
EN55011 / conducted class B, Radiated Class A  
EN60601-1-2 4th edition  
NOTE  
1. Ripple & Noise are measured at 20MHz of bandwidth with ceramic 0.1uF & chemi-con KY 47uF parallel capacitor.  
A 30cm twisted pair of no.18 AWG copper wire is connected to a  
47uF and 0.1uF capacitor of proper polarity and voltage rating.  
The oscilloscope probe ground led should connect right to the  
ground ring of the probe and be as short as possible.  
The oscilloscope bandwidth should be at 20MHz and connected  
to AC ground.  
2. Hold-up Time measured at 90% Vout.  
3. Main Vout >3% Load, 12V (Aux) / 0.3A., 12V (Aux) need 0.1A Minimum Load, Auxiliary voltage output ground 10.2~13.3V  
4. Strongly recommend to conduct this test with DC Voltage. If customer wishes to test with AC Voltage,  
please disconnect all Y-Capacitors within Arch power supply.  
DERATING  
http://www.archcorp.com.tw  
TEL: +886-2-26989508 FAX: +886-2-26981319  
-2-  
VER : A_7  
update : 2018.06.06  
AC-DC ITE & Medical Switching Power Supply  
MQF500E SERIES  
500 Watts  
MECHANICAL DIMENSION  
( Top View )  
A=For fixture to din rail clip only  
B=For fixture to pcb/chassis only  
A=M3x0.5P  
B=M3x0.5P  
ASSEMBLY INSTRUCTIONS  
*U Case T=1.5mm  
Customer is advised to screw into the threads no more than 2.5mm  
Connector Pin (CN1) = Cherng Weei PHD2.0 - 2x4P  
PIN#  
A
Single  
PE  
Mating Housing  
Terminal  
PIN#  
C1  
C2  
C3  
C4  
C5  
C6  
C7  
C8  
Single  
-5VSB  
+5VSB  
GND  
DC OK  
-RC  
Mating Housing  
Terminal  
AC Input Connector Pin : Alex 9397-3  
1
2
3
AC IN (N)  
NO PIN  
Alex 9396-3  
or equivalent  
Alex 96T Series  
or equivalent  
Cherng Weei  
PHD2.0 - 2x4P  
or equivalent  
Cherng Weei  
PHD2.0 - 2x4P  
or equivalent  
AC IN (L)  
DC Output Connector Pin  
+RC  
4
5
+DC OUT  
-DC OUT  
M5 Pan HD screw in 2 positions  
Torque to 8 Ibs-in(90 cNm) max.  
-S  
+S  
Connector Pin (FAN) = Cherng Weei CX-W250-02  
PIN#  
F1  
Single  
+12V  
GND  
Mating Housing  
Cherng Weei  
CS-H250-02  
or equivalent  
Terminal  
Cherng Weei  
CS-T2501  
F2  
or equivalent  
http://www.archcorp.com.tw  
TEL: +886-2-26989508 FAX: +886-2-26981319  
-3-  
VER : A_7  
update : 2018.06.06  
AC-DC ITE & Medical Switching Power Supply  
MQF500E SERIES  
500 Watts  
FUNCTION DESCRIPITON of CN1  
Pin No.  
C1  
Function  
-5VSB  
Description  
This pin connects to the negative terminal(-V). Return for DC-OK and -RC signal output.  
Stand by voltage output ground 4.2~5.5V, referenced to pin C1(-5VSB).  
The maximum load current is 1A with Fan, 0.4A without Fan..  
C2  
+5VSB  
C3  
C4  
C5  
GND  
DC OK  
-RC  
This pin connects to the negative terminal(-V). Return for DC-OK and -RC signal output.  
DC-OK Signal is a DC output, referenced to pin C3(DC-OK GND).  
This pin connects to the negative terminal(-V). Return for DC-OK and -RC signal output.  
Turns the output on and off by electrical or dry contact between pin C5 (-RC), Short: Power OFF, Open: Power ON.  
The input voltage must be less than 1V in order to disable VOUT and greater than 3.3V (up to 5V) to enable it.  
Negative sensing. The -S signal should be connected to the negative terminal of the load. The -S and +S leads should be  
twisted in pair to minimize noise pick-up effect.  
Positive sensing. The +S signal should be connected to the positive terminal of the load. The +S and -S leads should be  
twisted in pair to minimize noise pick-up effect.  
C6  
C7  
C8  
+RC  
-S  
+S  
FUNCTION MANUAL & APPLICATION NOTE  
1. DC-OK Signal  
Between  
DC-OK and GND  
Output  
Status  
3.7~6V  
0~1V  
ON  
OFF  
2. Remote Control  
It can be turned ON/OFF by using the "Remote Control" function.  
Between  
+RC and -RC  
Output  
Status  
SW ON (Short)  
SW OFF (Open)  
OFF  
ON  
2. +S and -S Sense  
Shorter wiring to each unit is recommended, as well as twisting +S and -S in pairs, as shown below  
http://www.archcorp.com.tw  
TEL: +886-2-26989508 FAX: +886-2-26981319  
-4-  
VER : A_7  
update : 2018.06.06  
AC-DC ITE & Medical Switching Power Supply  
MQF500E SERIES  
500 Watts  
BLOCK DIAGRAM  
http://www.archcorp.com.tw  
TEL: +886-2-26989508 FAX: +886-2-26981319  
-5-  

相关型号:

MQF500E-15S

AC-DC ITE & Medical Switching Power Supply
ARCH

MQF500E-24S

AC-DC ITE & Medical Switching Power Supply
ARCH

MQF500E-48S

AC-DC ITE & Medical Switching Power Supply
ARCH

MQF500E_18

AC-DC ITE & Medical Switching Power Supply
ARCH

MQF500O

AC-DC ITE & Medical Switching Power Supply
ARCH

MQF500O-12S

AC-DC ITE & Medical Switching Power Supply
ARCH

MQF500O-15S

AC-DC ITE & Medical Switching Power Supply
ARCH

MQF500O-24S

AC-DC ITE & Medical Switching Power Supply
ARCH

MQF500O-48S

AC-DC ITE & Medical Switching Power Supply
ARCH

MQF500O_18

AC-DC ITE & Medical Switching Power Supply
ARCH

MQF500U

AC-DC ITE & Medical Switching Power Supply
ARCH

MQF500U-12S

AC-DC ITE & Medical Switching Power Supply
ARCH