MBC120-1048L [BEL]

AC-DC Regulated Power Supply Module,;
MBC120-1048L
型号: MBC120-1048L
厂家: BEL FUSE INC.    BEL FUSE INC.
描述:

AC-DC Regulated Power Supply Module,

文件: 总6页 (文件大小:775K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
The MBC120 Series of open frame medical power supplies feature a  
wide universal AC input range of 85 V 264 VAC, offering 120 W of  
output power in a compact footprint, with a variety of isolated single  
output voltages.  
The MBC series was designed to 3rd edition medical approvals and  
provides 2x MOPP (Means of Patient Protection) isolation for Class I  
and Class II installations.  
These power supplies are ideal for medical, telecom, datacom,  
industrial equipment and other applications.  
3 x 2 Inch Footprint  
120 Watts with Forced Air Cooling  
Approval to EN60601 3rd Edition  
Efficiencies up to 93%  
-40 To 70°C Operating Temperature  
Dual Fusing  
Thermal Shut-Down Feature  
>3.00 Million Hours, Telcordia -SR332-Issue 3  
Standby Power < 0.3 W  
Class II Option Available  
RoHS Compliant  
CE Marked  
Diagnostic  
Drug Pump  
Dialysis  
Home Health Care  
Monitoring  
Portable Equipment  
Asia-Pacific  
Europe, Middle East  
North America  
+86 755 298 85888  
+353 61 225 977  
+1 866 513 2839  
© 2015 Bel Power Solutions, Inc.  
BCD.00667_AA  
2
MBC120-1T12L  
MBC120-1012L  
MBC120-1T15L  
MBC120-1015L  
MBC120-1T24L  
MBC120-1024L  
MBC120-1T30L  
MBC120-1030L  
MBC120-1T48L  
MBC120-1048L  
MBC120-1T58L  
MBC120-1058L  
Screw Terminal  
Molex Header  
Screw Terminal  
Molex Header  
Screw Terminal  
Molex Header  
Screw Terminal  
Molex Header  
Screw Terminal  
Molex Header  
Screw Terminal  
Molex Header  
8.33 A  
6.66 A  
4.16 A  
3.33 A  
2.08 A  
1.72 A  
12 V  
15 V  
24 V  
30 V  
48 V  
58 V  
10.0 A  
8.0 A  
5.0 A  
4.0 A  
2.5 A  
2.07 A  
0.0 A  
0.0 A  
0.0 A  
0.0 A  
0.0 A  
0.0 A  
1%  
1%  
1%  
1%  
1%  
1%  
COVER-120-XBC  
metal cover kit accessory  
1
For Class II Option (without input Earth pin) add suffix: -2 (e.g.: MBC120-1012L-2).  
Ripple is peak to peak with 20 MHz bandwidth and 10 μF (Tantalum capacitor) in parallel with a 0.1 μF capacitor at rated line voltage  
2
and load ranges.  
Specifications are for nominal input voltage, 25°C unless otherwise stated.  
Input Voltage  
Input Frequency  
Universal (see derating under output power)  
85-264 VAC / 390 VDC3  
47-63 Hz  
115 VAC:  
230 VAC:  
1.2 A max.  
0.65 A max.  
Input Current  
No Load Power  
Typical  
< 0.3 W  
115 VAC:  
230 VAC:  
25 A  
45 A  
Inrush Current  
264 VAC:  
75 A  
Typical (N.A. For Class II Option)  
Touch current  
300 uA  
<100 uA  
Leakage Current  
Power Factor  
@ Full Load, Active PFC  
> 0.95  
Switching Frequency  
Typical  
60 KHz  
3
Functional, not approved.  
tech.support@psbel.com  
3
Forced cooling (with 200 LFM):  
Convection cooling (for input 100-264 VAC, de-rate  
linearly to 80 W @ 85 VAC):  
48 V, 58 V:  
24 V, 30 V:  
12 V, 15 V:  
120 W  
100 W  
Output Power  
Efficiency  
93%  
91%  
90%  
Hold-up Time  
Typical  
>10 ms  
+/-0.5%  
+/-1%  
Line Regulation  
Load Regulation  
25% step load change, at 0.1A/uS slew rate,  
50% duty cycle, 50 Hz = 4%  
Transient Response  
recovery time < 5 ms  
+/-3%  
Voltage Adjustment  
Rise Time  
Typical  
55 ms  
Set Point Tolerance  
Over Current Protection  
Over Voltage Protection  
Short Circuit Protection  
+/-1%  
> 110%  
110 to 140%  
Latch type (AC recycling required)  
Hiccup mode  
-40 to +70°C  
-40 to 0°C  
Operating Temperature  
Storage Temperature  
Cooling  
Startup guaranteed  
(derate linearly above 50°C to 70°C as per Fig 1)  
-40 to +85°C  
Forced: with 300 LFM (Refer to Mech. Drawing)  
Convection: for input 100-264 VAC  
120 W  
100 W  
(derate linearly to 80 W @ 85 VAC)  
Relative Humidity  
Altitude  
Noncondensing  
5% to 95%  
Operating:  
Nonoperating:  
16,000 ft  
40,000 ft.  
Reliability  
MTBF according to Telcordia SR332-Issue 3  
3.00 million hours  
Conducted Emissions  
Static Discharge  
EN55022-B, CISPR22-B, FCC PART15 B  
EN61000-4-2:  
Level-3  
Level-3  
Level-3  
RF Field Susceptibility  
Fast Transients/Bursts  
EN61000-4-3:  
EN61000-4-4:  
Radiated:  
Radiated with external core:  
Level A  
Level B  
Radiated Emissions  
(King core K5B RC 25x12x15-M in input cable with 5  
Turns)  
Surge Susceptibility  
Harmonic Current  
EN61000-4-5:  
Level-3  
Class D  
EN61000-3-2:  
tech.support@psbel.com  
4
Input to Output: (For medical applications)  
Input to GND: (Not Applicable For Class II Option)  
Output to GND: for type BF  
for type B (N/A For Class II Option)  
Primary to Secondary:  
Primary to Earth:  
Secondary to Earth:  
4000 VAC  
1500 VAC  
1500 VAC  
500 VAC  
2 MOPP  
1 MOPP  
1 MOPP  
Isolation Voltage  
Protection Level  
Approved to the latest edition of the following  
standards: CSA/UL60601-1, EN60601-1 and  
IEC60601-1.  
Safety Standard(s)  
Agency Approvals  
CE mark  
Nemko, cULus, CB  
Complies with LVD Directive  
Figure 1. Derating Curve  
Molex: 39357-0003  
Tyco-2-1776112-3  
Molex: 1722861103  
(Mating conn: Molex 1722561003)  
Molex: 39357-0004  
Tyco-2-1776112-4  
Molex: 1722861104  
(Mating conn: Molex 1722561004)  
Pin 1  
Pin 2  
Pin 3  
AC Line  
Not Fitted  
AC Neutral  
Screw Terminal (Option 1)  
Molex Header (Option 2)  
Screw Terminal (Option 1)  
Molex Header (Option 2)  
AC Input Connector  
DC Output Connector  
J1  
J2  
Pin 1, 2  
Pin 3, 4  
V1 -VE  
V1 +VE  
tech.support@psbel.com  
5
Weight  
150 g  
Dimensions  
76.2 x 50.8 x 30.1 mm (3 x 2 x 1.18 inch)  
Figure 2. Mechanical Drawing - Screw Terminal (Option 1)  
tech.support@psbel.com  
6
Figure 3. Mechanical Drawing - Molex Header (Option 2)  
NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support systems,  
equipment used in hazardous environments, or nuclear control systems.  
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the  
date manufactured. Specifications are subject to change without notice.  
tech.support@psbel.com  

相关型号:

MBC120-1058L

3 x 2 Inch Footprint
BEL

MBC120-1T12L

3 x 2 Inch Footprint
BEL

MBC120-1T15L

3 x 2 Inch Footprint
BEL

MBC120-1T24L

3 x 2 Inch Footprint
BEL

MBC120-1T30L

3 x 2 Inch Footprint
BEL

MBC120-1T48L

3 x 2 Inch Footprint
BEL

MBC120-1T58L

3 x 2 Inch Footprint
BEL

MBC1200

Single 24 VDC / 48 VDC output voltages available
BEL

MBC1200-1T24-PCF

Single 24 VDC / 48 VDC output voltages available
BEL

MBC1200-1T24-UCF

Single 24 VDC / 48 VDC output voltages available
BEL

MBC13706

925 MHz - 960 MHz RF/MICROWAVE NARROW BAND LOW POWER AMPLIFIER, PLASTIC, MICROPAK-8
MOTOROLA

MBC13720

Amplifier with Bypass Switch
MOTOROLA