MBC120 [BEL]
3 x 2 Inch Footprint;型号: | MBC120 |
厂家: | BEL FUSE INC. |
描述: | 3 x 2 Inch Footprint |
文件: | 总5页 (文件大小:731K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
The MBC120 Series of open frame medical power supplies feature a
wide universal AC input range of 85 V – 264 VAC, offering 120 W of
output power in a compact footprint, with a variety of isolated single
output voltages.
MBC series is designed and approved to the latest Medical
dards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I
ss II applications.
e power supplies are ideal for medical, telecom, datacom,
trial equipment and other applications.
3 x 2 Inch Footprint
•
•
•
•
•
•
•
120 Watts with Forced Air Cooling
Approved to EN/IEC 60601-1
Efficiencies up to 93%
-40 To 70°C Operating Temperature
Dual Fusing
Thermal Shut-Down Feature
>3.00 Million Hours, Telcordia -SR332-Issue 3
Standby Power < 0.3 W
Class II Option Available
RoHS Compliant
CE Marked
Diagnostic
Drug Pump
Dialysis
Home Health Care
Monitoring
•
•
•
•
•
•
Portable Equipment
2
MBC120 Series
MBC120-1T12L
MBC120-1012L
MBC120-1T15L
MBC120-1015L
MBC120-1T24L
MBC120-1024L
MBC120-1T30L
MBC120-1030L
MBC120-1T48L
MBC120-1048L
MBC120-1T58L
MBC120-1058L
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
8.33 A
6.66 A
4.16 A
3.33 A
2.08 A
1.72 A
12 V
15 V
24 V
30 V
48 V
58 V
10.0 A
8.0 A
5.0 A
4.0 A
2.5 A
2.07 A
0.0 A
0.0 A
0.0 A
0.0 A
0.0 A
0.0 A
1%
1%
1%
1%
1%
1%
COVER-120-XBC3
metal cover kit accessory
Specifications are for nominal input voltage, 25°C unless otherwise stated.
Input Voltage
Universal (see derating under output power)
85-264 VAC / 390 VDC4
47-63 Hz
Input Frequency
Input Current
115 VAC:
230 VAC:
1.2 A max.
0.65 A max.
No Load Power
Typical
< 0.3 W
115 VAC:
25 A
Inrush Current
230 VAC:
264 VAC:
45 A
75 A
Typical (N.A. For Class II Option)
Touch current
300 uA
<100 uA
Leakage Current
Power Factor
@ Full Load, Active PFC
Typical
> 0.95
Switching Frequency
60 KHz
1 For Class II Option (without input Earth pin) add suffix: -2 (e.g.: MBC120-1012L-2).
2 Ripple is peak to peak with 20 MHz bandwidth and 10 μF (Tantalum capacitor) in parallel with a 0.1 μF capacitor at rated line voltage
and load ranges.
3 When used in Cover Kit, de-rate output power to 70 % under all operating conditions
4 Functional, not approved.
tech.support@psbel.com
3
MBC120 Series
Forced cooling (with 200 LFM):
Convection cooling (for input 100-264 VAC, de-rate
linearly to 80 W @ 85 VAC):
48 V, 58 V:
24 V, 30 V:
12 V, 15 V:
120 W
100 W
Output Power
Efficiency
93%
91%
90%
Hold-up Time
Typical
>10 ms
+/-0.5%
+/-1%
Line Regulation
Load Regulation
25% step load change, at 0.1A/uS slew rate,
50% duty cycle, 50 Hz = 4%
Transient Response
recovery time < 5 ms
+/-3%
Voltage Adjustment
Rise Time
Typical
55 ms
Set Point Tolerance
Over Current Protection
Over Voltage Protection
Short Circuit Protection
+/-1%
> 110%
110 to 140%
Latch type (AC recycling required)
Hiccup mode
-40 to +70°C
-40 to 0°C
Operating Temperature5
Storage Temperature
Cooling
Startup guaranteed
(derate linearly above 50°C to 70°C as per Fig 1)
-40 to +85°C
Forced: with 300 LFM (Refer to Mech. Drawing)
Convection: for input 100-264 VAC
120 W
100 W
(derate linearly to 80 W @ 85 VAC)
Relative Humidity
Altitude
Noncondensing
5% to 95%
Operating:
Nonoperating:
16,000 ft
40,000 ft.
Reliability
MTBF according to Telcordia –SR332-Issue 3
3.00 million hours
Conducted Emissions
Radiated Emissions
EN 55011-B, CISPR22-B, FCC PART15-B
EN 55011 A;
with external core (King core K5B RC 25x12x15-M in input cable)
Pass
Pass
Level B
Input Current Harmonics
Voltage Fluctuation and Flicker
ESD Immunity
EN 61000-3-2
Class D
EN 61000-3-3
EN 61000-4-2
EN 61000-4-3
EN 61000-4-4
EN 61000-4-5
EN 61000-4-6
EN 61000-4-8
EN 61000-4-11
Pass
Level 4, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 4, Criterion A
Criterion B
Radiated Field Immunity
Electrical Fast Transient Immunity
Surge Immunity
Conducted Immunity
Magnetic Field Immunity
Voltage Dips, Interruptions
5 Output ripple can be more than 10% of the output voltage.
North America
Asia-Pacific
Europe, Middle East
+1 408 785 5200
+86 755 298 85888
+353 61 225 977
© 2017 Bel Power Solutions & Protection
BCD.00667_AC
4
MBC120 Series
Input to Output: (For medical applications)
Input to GND: (Not Applicable For Class II Option)
Output to GND: for type BF
for type B (N/A For Class II Option)
Primary to Secondary:
Primary to Earth:
Secondary to Earth:
4000 VAC
1500 VAC
1500 VAC
500 VAC
2 MOPP
1 MOPP
1 MOPP
Isolation Voltage
Protection Level
Approved to the latest edition of the following
standards: CSA/UL60601-1, EN60601-1 and IEC60601-
1.
Safety Standard(s)
Agency Approvals
CE mark
Nemko, cULus, CB
Complies with LVD Directive
Convection load: 100 W up to 50 °C
De-rate above 50 °C @ 2.5% per °C
Forced air cooled load: 120 W up to 50°C
De-rate above 50 °C @ 2.5% per °C
Figure 1. Derating Curve
Molex: 39357-0003
Tyco-2-1776112-3
Molex: 1722861103
(Mating conn: Molex 1722561003)
Molex: 39357-0004
Tyco-2-1776112-4
Molex: 1722861104
(Mating conn: Molex 1722561004)
Pin 1
Pin 2
Pin 3
AC Line
Not Fitted
AC Neutral
Screw Terminal (Option 1)
AC Input Connector
DC Output Connector
J1
J2
Molex Header (Option 2)
Screw Terminal (Option 1)
Molex Header (Option 2)
Pin 1, 2
Pin 3, 4
V1 -VE
V1 +VE
Weight
150 g
Dimensions
76.2 x 50.8 x 30.1 mm (3 x 2 x 1.18 inch)
tech.support@psbel.com
5
MBC120 Series
Figure 2. Mechanical Drawing - Screw Terminal (Option 1)
Figure 3. Mechanical Drawing - Molex Header (Option 2)
NOTES: In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following:
1
2
3
Stand off, used to mount PCB has OD of 5.4 mm max.
Screws, used to fix PCB on stand off, have head dia of 6.0 mm max.
Washer, if used, to have dia of 6.5 mm max.
NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support systems,
equipment used in hazardous environments, or nuclear control systems.
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the
date manufactured. Specifications are subject to change without notice.
North America
Asia-Pacific
Europe, Middle East
+1 408 785 5200
+86 755 298 85888
+353 61 225 977
© 2017 Bel Power Solutions & Protection
BCD.00667_AC
相关型号:
SI9130DB
5- and 3.3-V Step-Down Synchronous ConvertersWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9135LG-T1
SMBus Multi-Output Power-Supply ControllerWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9135LG-T1-E3
SMBus Multi-Output Power-Supply ControllerWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9135_11
SMBus Multi-Output Power-Supply ControllerWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9136_11
Multi-Output Power-Supply ControllerWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9130CG-T1-E3
Pin-Programmable Dual Controller - Portable PCsWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9130LG-T1-E3
Pin-Programmable Dual Controller - Portable PCsWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9130_11
Pin-Programmable Dual Controller - Portable PCsWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9137
Multi-Output, Sequence Selectable Power-Supply Controller for Mobile ApplicationsWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9137DB
Multi-Output, Sequence Selectable Power-Supply Controller for Mobile ApplicationsWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9137LG
Multi-Output, Sequence Selectable Power-Supply Controller for Mobile ApplicationsWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9122E
500-kHz Half-Bridge DC/DC Controller with Integrated Secondary Synchronous Rectification DriversWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
©2020 ICPDF网 联系我们和版权申明