MBC300-1T12G [BEL]
5.0 x 3.0 x 1.5 in (127.0 x 76.2 x 38.1 mm);型号: | MBC300-1T12G |
厂家: | BEL FUSE INC. |
描述: | 5.0 x 3.0 x 1.5 in (127.0 x 76.2 x 38.1 mm) |
文件: | 总7页 (文件大小:929K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
The MBC300 Series of open-frame medical power supplies, with its
wide universal 90-264 VAC input range, is available at 300 W of output
power and a variety of single output voltages.
The MBC series is designed and approved to the latest Medical
standards (EN/IEC 60601-1), providing 2 x MOPP (Means of Patient
Protection) isolation for Class I & Class II applications.
These medical power supplies are ideal for monitoring, home health
quipment as well as surgical devices.
5.0 x 3.0 x 1.5 in (127.0 x 76.2 x 38.1 mm)
Approved to EN/IEC 60601-1
2x MOPP
12 V fan output
5 V standby output
IEC Protection Class Options:
o
Class I: Earth pin J4 (no suffix)
Class II: No Earth pin (-2 suffix)
o
EMI Class B
•
•
•
•
Medical Safety Agency Approvals
RoHS Compliant
CE marked
Diagnostic
Drug Pump
Dialysis
Home Health Care
Monitoring
•
•
•
•
•
•
Imaging
2
MBC300 Series
MAX LOAD
CONVECTION
OUTPUT
VOLTAGE
MINIMUM
LOAD
RIPPLE &
NOISE 2
MODEL 1
CONNECTOR
POWER
300 LFM
40.0 A
MBC300-1T05G
MBC300-1T12G
MBC300-1T15G
MBC300-1T24G
MBC300-1T30G
MBC300-1T48G
Cover-300-XBC 3
Screw Terminal
Screw Terminal
Screw Terminal
Screw Terminal
Screw Terminal
Screw Terminal
5 VDC
12 VDC
15 VDC
24 VDC
30 VDC
48 VDC
28.0 A
16.67 A
13.33 A
8.33 A
6.67 A
4.17 A
0.0 A
0.0 A
0.0 A
0.0 A
0.0 A
0.0 A
2%
2%
2%
2%
2%
2%
200 W
300 W
300 W
325 W
325 W
325 W
25.0 A
20.0 A
13.54 A
10.83 A
6.77 A
Metal cover kit accessory
1
Class II means without input Earth pin. Add suffix “-2” at the end of the Model Number to order Class II version.
Ripple is peak to peak with 20 MHz bandwidth and 10 µF (Electrolytic capacitor) in parallel with a 0.1 µF capacitor at rated line voltage
and load ranges.
2
3
When used in Cover Kit, de-rate output power to 70 % under all operating conditions.
Specifications are for nominal input voltage, 25°C unless otherwise stated.
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
90 – 264 VAC / 120 – 390 VDC
47 to 63 Hz
Input Voltage
Universal
Input Frequency
120 VAC
230 VAC
3.2 A max
1.65 A max
Input Current
120 VAC
230 VAC
35 A max
65 A max
Inrush Current
< 125 μA
< 250 μA
80 kHz typical
35 to 250 kHz, 90 kHz typical
120 VAC
230 VAC
PFC converter (fixed)
Resonant converter (variable)
Leakage Current
Switching Frequency
tech.support@psbel.com
3
MBC300 Series
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
12 to 48 V
Output Voltage 4
Stand-by Output 5
Output Power 6, 7
5 VDC
Derate linearly to 80% from 90 VAC to 80 VAC input.
200 to 325 W
120 VAC
230 VAC
88% typical
92% typical
Efficiency
Hold Up Time
Power Factor
120 / 230 VAC
10 ms
120 VAC
230 VAC
0.98
0.95
Line Regulation
± 0.5 %
Load Regulation
± 2 %
50% to 100% load change, 50 Hz, 50% duty cycle, 0.1 A/μs,
Transient Response
Rise Time
< 10 %, recovery time < 5 ms
< 100 ms
Set Point Tolerance
Voltage Output Adjustment
Over Voltage Protection
Over Current Protection
Short Circuit Protection
Over Temperature Protection
± 1 %
± 3 %
Automatic recovery
110 to 150 %
110 to 150 %
Short term, automatic recovery
Automatic Recovery
110° C primary heat sink
5 V model 140 W max
12, 15, 24, 30 & 48 V models 200 W max
Convection
Cooling 8
5 V model 200 W max
12 &15 V models 300 W max
24, 30 & 48 V models 325 W max
With 300 LFM
4
Peak current rating on main output is 120% of max., lasting < 30 s with a maximum 10% duty cycle.
Standby output voltage tolerance including set point accuracy, line and load regulation is +/-10%. Ripple and noise is less than 5%.
Combined output power of main output, fan supply and standby supply shall not exceed max. power rating.
Derate output power linearly to 80% from 90 VAC to 80 VAC input.
5
6
7
8
Refer de-rating curves to determine output power over the entire operating temperature range
Conducted Emissions 9
Radiated Emissions
EN 55011-B, CISPR22-B, FCC PART15-B
EN 55011 B
Pass
Pass
Input Current Harmonics
Voltage Fluctuation and Flicker
ESD Immunity
EN 61000-3-2
Class D
EN 61000-3-3
Pass
EN 61000-4-2
Level 4, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 4, Criterion A
Criterion A & B
Radiated Field Immunity
Electrical Fast Transient Immunity
Surge Immunity
EN 61000-4-3
EN 61000-4-4
EN 61000-4-5
Conducted Immunity
EN 61000-4-6
Magnetic Field Immunity
Voltage Dips, Interruptions
EN 61000-4-8
EN 61000-4-11
9
Class II product meets Class A limit line for conducted emission.
North America
+1 408 785 5200
Asia-Pacific
+86 755 298 85888
Europe, Middle East
+353 61 225 977
© 2020 Bel Power Solutions & Protection
BCD.00290_AF
4
MBC300 Series
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
Input to Output (2 MOPP)
Input to Earth (1 MOPP)
Output to Earth (1 MOPP)
4000 VAC
1500 VAC
500 VAC
Isolation Voltage
EN 60601-1, IEC 60601-1 (ed.3), ANSI / AAMI ES 60601 - 1,
CSA C22.2 No. 60601-1
Safety Standards
Agency Approvals
CE mark
Nemko
Complies with LVD Directive
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
Refer to de-rating curves
-20 to 0°C start-up is guaranteed
-20 to 70°C
Operating Temperature
Storage Temperature
Relative Humidity
-40 to 85° C
95% Rh
Non Condensing
Operating:
Non-Operating:
10,000 ft.
40,000 ft.
Altitude
Reliability
MTBF according to Telcordia –SR332-issue 3
1.77 million hours
PARAMETER
Power Good 10
Remote On/ Off
Remote Sense
DESCRIPTION / CONDITION
TTL signal goes high after main output is within regulation band, delay is 0.1 to 0.3 s
To turn on PSU short remote pin to ground
Compensates for 200 mV drop
10
Power good signal cannot be used as a current source. Internal pull up resistor from PG signal to 5V is 10K. It is recommended to use
external transistor if intended to source current.
CONNECTOR
PIN
DESCRIPTION / CONDITION
MANUFACTURER / PN
Pin 1
Pin 2
AC LINE
AC NEUTRAL
Molex: 26-60-4030
Mating: 09-50-3031; Pins: 08-50-0106
AC Input Connector
J1
6-32 inches Screw Pan HD
Mating: Designed to accept Ring Tongue Terminal AMP : 8-31886-1,
wherein one 16 AWG (max) wire can be crimped.
Note: One Ring Tongue Terminal with 16 AWG is recommended for
current up to 11 A only.
Pin 1
Pin 2
RTN
V1
DC Output Connector
Signals & Aux Power
J2
Use multiple tongue terminals with wire for more current.
Pin 1
Pin 2
Pin 3
Pin 4
Pin 5
Pin 6
Pin 7
Pin 8
REMOTE ON/OFF
RTN
VFAN (+12 V/0.5 A)
-VE REMOTE SENSE
VSTBY (+5 V/2 A, +/-5%)
+VE REMOTE SENSE
RTN
Molex: 22-23-2081
Mating: 22-01-2087; Pins: 08-50-0113
J3 11
POWER GOOD
Molex: 19705-4301
Mating: 190030001
Earth (Spade Connector)
J4 12
11
PSU is supplied with J3 housing, pin-1 and pin-2 shorted to enable main output without remote on/off feature.
Class I models only.
12
tech.support@psbel.com
5
MBC300 Series
Forced air cooled load: 200 W up to 40°C
De-rate above 40 °C @ 1.67% per °C
Convection load: 140 W up to 40 °C
De-rate above 40 °C @ 1.67% per °C
Forced air cooled load: 300 W up to 50°C
De-rate above 50 °C @ 2.5% per °C
Convection load: 200 W up to 40 °C
De-rate between 40-50 °C @ 1% per °C
De-rate above 50 °C @ 2.5% per °C
Forced air cooled load: 325 W up to 50°C
De-rate above 50 °C @ 2.5% per °C
Convection load: 200 W up to 40 °C
De-rate between 40-50 °C @ 1% per °C
De-rate above 50 °C @ 2.5% per °C
Figure 1. Derating Curves
North America
+1 408 785 5200
Asia-Pacific
+86 755 298 85888
Europe, Middle East
+353 61 225 977
© 2020 Bel Power Solutions & Protection
BCD.00290_AF
6
MBC300 Series
PARAMETER
DESCRIPTION / CONDITION
450 g (0.99 lbs)
Weight
Dimensions
127.0 x 76.2 x 38.1 mm (5.0 x 3.0 x 1.5 inch)
Figure 2. Mechanical Drawing
tech.support@psbel.com
7
MBC300 Series
Figure 3. Mechanical Drawing with Cover Kit
NOTES: In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following:
1
2
3
Stand off, used to mount PCB has OD of 5.4 mm max.
Screws, used to fix PCB on stand off, have head dia of 6.0 mm max.
Washer, if used, to have dia of 6.5 mm max.
NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support systems,
equipment used in hazardous environments, or nuclear control systems.
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the
date manufactured. Specifications are subject to change without notice.
North America
Asia-Pacific
Europe, Middle East
+1 408 785 5200
+86 755 298 85888
+353 61 225 977
© 2020 Bel Power Solutions & Protection
BCD.00290_AF
相关型号:
©2020 ICPDF网 联系我们和版权申明