EMMA180167 [CUI]
MEDICAL AC-DC POWER SUPPLY;型号: | EMMA180167 |
厂家: | CUI INC |
描述: | MEDICAL AC-DC POWER SUPPLY |
文件: | 总4页 (文件大小:633K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
For more information, please visit the product page
|
For information on related products, view Dc Power Jacks
date 07/10/2015
page 1 of 4
SERIES: EMMA 30W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
FEATURES
• up to 30 W power
• universal input (90~264 Vac)
• interchangeable AC blades
• single regulated output from 5~24 Vdc
• over voltage and short circuit protections
• full medical safety approvals
• level V efficiency
ꢀEC
ꢁ0ꢁ01ꢂ1
EN
®
ꢁ0ꢁ01ꢂ1
output
voltage
output
current
max
output
power
max
ripple
efficiency
MODEL
and noise1
level
max
(Vdc)
5
(A)
(W)
(mVp-p)
EMMA050400
EMMA090300
EMMA120250
EMMA150200
EMMA180167
EMMA240125
4
3
20
27
30
30
30
30
50
90
V
V
V
V
V
V
9
12
15
18
24
2.5
2
120
150
180
240
1.67
1.25
1. at full load, 100 ~ 240 Vac input, 20 MHz bandwidth oscilloscope, each output terminated with a 10 µF aluminum electrolytic and 0.1 µF ceramic capacitors.
PART NUMBER KEY
EMMA120250 X - XX - IC - CXX
Base Number
Reserved for Custom
example of 12 Vdc, 2.5 A
DC Plug Type:
Configurations
Factory
See output plug options
on page 3 for more details
Designation
Blades:
“blank” = North American, European, United Kingdom,
Australian
N = North American blade included
E = European blade included
B = United Kingdom blade included
A = Australian blade included
K = No blades included
cui.com
For more information, please visit the product page
|
For information on related products, view Dc Power Jacks
CUI Inc │ SERIES: EMMA 30W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 07/10/2015 │ page 2 of 4
INPUT
parameter
conditions/description
at 240 V ac, cold start
conditions/description
min
90
typ
max
264
63
units
Vac
Hz
A
voltage
frequency
47
input current
inrush current
no load power consumption
0.8
100
0.3
A
W
OUTPUT
parameter
min
typ
max
units
line regulation1
±1
%
5 Vdc output
9 Vdc output
all other outputs
±6
±3
±2
%
%
%
load regulation2
voltage accuracy
hold-up time
±2
10
%
ms
at 115 Vac
switching frequency
temperature coefficient
70
kHz
%/°C
±0.05
Note:
1. measured from 100 ~ 240 Vac, full load
2. measured from 60% to full load and from 60 ~ 20% load (60% ±40% load)
PROTECTIONS
parameter
conditions/description
TVS component to clamp
continuous, auto restart
over voltage protection
short circuit protection
SAFETY & COMPLIANCE
parameter
conditions/description
input to output
min
typ
max
units
isolation voltage
5,656
Vdc
safety approvals
IEC 60601-1, EN 60601-1, UL 60601-1
EN 55011 Class B, FCC CRF47 Part 18 Class B, EN 60601-1-2, EN 61000-3-(2,3),
IEC 61000-4-(2,4,5,6,8,11)
EMI/EMC
leakage current
MTBF
0.1
mA
as per MIL-HDBK-217F, 115 Vac, 25 °C
2011/65/EU
200,000
hours
RoHS
ENVIRONMENTAL
parameter
conditions/description
min
0
typ
max
40
units
°C
operating temperature
storage temperature
humidity
-20
85
°C
non-condensing
93
%
cui.com
For more information, please visit the product page
|
For information on related products, view Dc Power Jacks
CUI Inc │ SERIES: EMMA 30W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 07/10/2015 │ page 3 of 4
MECHANICAL
parameter
dimensions
input plug
weight
conditions/description
min
typ
max
units
108.67 x 61.98 x 36.70 (4.278 x 2.440 x 1.445 inch)
interchangeable blades (US, Europe, UK, Australia)
mm
300
g
MECHANICAL DRAWING
units: mm [inches]
tolerance: ±0.5 [±0.02]
36.70 1.445
61.98 2.440
1.40 0.055
EMM-US
EMM-AU
108.67 4.278
EMM-EU
EMM-UK
DC OUTPUT PLUG OPTIONS / DC CORD
cord length
ń
ń
Standard PX
Right Angle PXR
A
B
C
Unit
mm
mm
MODEL NO.
EMMA050040
EMMA090300
EMMA120250
EMMA150200
EMMA180167
EMMA240125
CABLE GAUGE
18 AWG
CORD LENGTH
1,220 mm ±50
1,220 mm ±50
1,800 mm ±50
1,800 mm ±50
1,800 mm ±50
1,800 mm ±50
P5/P5R
5.5
5.5
2.1
2.5
9.5
9.5
P6/P6R
18 AWG
18 AWG
PXX X X
18 AWG
Plug Type
Plug Angle:
“Blank” = Standard “Blank” = N/A
R = Right Angle P = Center Positive
Plug Polarity:
18 AWG
18 AWG
N = Center Negative
cui.com
For more information, please visit the product page
|
For information on related products, view Dc Power Jacks
CUI Inc │ SERIES: EMMA 30W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 07/10/2015 │ page 4 of 4
REVISION HISTORY
rev.
description
date
1.0
initial release
12/16/2011
03/23/2012
08/21/2012
07/10/2015
1.01
1.02
1.03
updated P7/P7R B dimension
V-Infinity branding removed
updated datasheet
The revision history provided is for informational purposes only and is believed to be accurate.
Headquarters
20050 SW 112th Ave.
Tualatin, OR 97062
800.275.4899
Fax 503.612.2383
cui.com
techsupport@cui.com
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause undesired operation.
CUI offers a one (1) year limited warranty. Complete warranty information is listed on our website.
CUI reserves the right to make changes to the product at any time without notice. Information provided by CUI is believed to be accurate and reliable. However, no responsibility is
assumed by CUI for its use, nor for any infringements of patents or other rights of third parties which may result from its use.
CUI products are not authorized or warranted for use as critical components in equipment that requires an extremely high level of reliability. A critical
component is any component of a life support device or system whose failure to perform can be reasonably expected to cause the failure of the life support device or system, or to
affect its safety or effectiveness.
相关型号:
©2020 ICPDF网 联系我们和版权申明