ETMA150200UD-P5RP-IC [CUI]
AC/DC DESKTOP ADAPTER 15V 30W;型号: | ETMA150200UD-P5RP-IC |
厂家: | CUI INC |
描述: | AC/DC DESKTOP ADAPTER 15V 30W |
文件: | 总4页 (文件大小:362K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
Additional Resources: Product Page
|
3D Model
date 07/10/2015
page 1 of 4
SERIES: ETMA 30W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
FEATURES
• up to 30 W power
• universal input (90~264 Vac)
• single regulated output from 5~24 Vdc
• over voltage and short circuit protections
• full medical safety approvals
• level V efficiency
• custom designs available
output
voltage
output
current
max
output
power
max
ripple
efficiency
MODEL
and noise1
level
max
(Vdc)
5
(A)
(W)
(mVp-p)
ETMA050400UD
ETMA090300UD
ETMA120250UD
ETMA150200UD
ETMA180165UD
ETMA240125UD
4
3
20
27
30
30
30
30
50
V
V
V
V
V
V
9
90
12
15
18
24
2.5
2
120
150
180
240
1.65
1.25
1. at full load, 100 ~ 240 Vac input, 20 MHz bandwidth oscilloscope, each output terminated with a 10 µF aluminum electrolytic and 0.1 µF ceramic capacitors.
PART NUMBER KEY
ETMA050400UD - XX - IC - CXX
Base Number
Reserved for Custom
example of 5 Vdc, 4 A
Configurations
DC Plug Type:
Factory
See output plug options
on page 3 for more details
Designation
cui.com
Additional Resources: Product Page
|
3D Model
CUI Inc │ SERIES: ETMA 30W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 07/10/2015 │ page 2 of 4
INPUT
parameter
conditions/description
at 240 Vac
min
typ
max
264
63
units
Vac
Hz
A
voltage
90
47
frequency
input current
inrush current
no load power consumption
0.6
50
A
0.3
W
OUTPUT
parameter
conditions/description
min
typ
max
units
line regulation1
±1
%
5 Vdc output
9 Vdc output
12, 15 Vdc outputs
18, 24 Vdc outputs
±6
±4
±3
±2
%
%
%
%
load regulation2
voltage accuracy
hold-up time
±2
16
%
ms
at 115 Vac
switching frequency
temperature coefficient
70
kHz
%/°C
±0.05
Note:
1. measured from 100 ~ 240 Vac, full load
2. measured from 60% to full load and from 60 ~ 20% load (60% ±40% load)
PROTECTIONS
parameter
conditions/description
TVS component to clamp
continuous, auto restart
over voltage protection
short circuit protection
SAFETY & COMPLIANCE
parameter
isolation voltage
safety approvals
EMI/EMC
conditions/description
min
typ
max
units
input to output
5,656
Vdc
IEC 60601-1, EN 60601-1, UL 60601-1
EN 55011, EN 60601-1-2, EN 61000-3-(2,3)
as per MIL-HDBK-217F, 115 Vac, 25 °C
2011/65/EU
MTBF
300,000
hours
RoHS
ENVIRONMENTAL
parameter
conditions/description
min
0
typ
max
40
units
°C
operating temperature
storage temperature
humidity
-20
85
°C
non-condensing
93
%
cui.com
Additional Resources: Product Page
|
3D Model
CUI Inc │ SERIES: ETMA 30W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 07/10/2015 │ page 3 of 4
MECHANICAL
parameter
dimensions
input plug
weight
conditions/description
110 x 50 x 30 (4.331 x 1.969 x 1.181 inch)
IEC320 / C8
min
typ
max
units
mm
220
g
MECHANICAL DRAWING
units: mm
50 0.5
30 0.5
110 0.5
LED
DC OUTPUT PLUG OPTIONS / DC CORD
ń
ń
cord length
Standard PX
Right Angle PXR
A
B
C
Unit
mm
mm
MODEL NO.
CABLE GAUGE
18 AWG
CORD LENGTH
720 mm ±50
P5/P5R
P6/P6R
5.5
5.5
2.1
2.5
9.5
9.5
ETMA050400UD
ETMA090300UD
ETMA120250UD
ETMA150200UD
ETMA180165UD
ETMA240125UD
18 AWG
1,220 mm ±50
1,800 mm ±50
1,800 mm ±50
1,800 mm ±50
1,800 mm ±50
18 AWG
18 AWG
18 AWG
18 AWG
cui.com
Additional Resources: Product Page
|
3D Model
CUI Inc │ SERIES: ETMA 30W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 07/10/2015 │ page 4 of 4
REVISION HISTORY
rev.
description
date
1.0
initial release
12/02/2011
03/23/2012
08/21/2012
07/10/2015
07/09/2020
1.01
1.02
1.03
1.04
updated P7/P7R B dimension
V-Infinity branding removed
updated datasheet
company logo update
The revision history provided is for informational purposes only and is believed to be accurate.
Headquarters
20050 SW 112th Ave.
Tualatin, OR 97062
800.275.4899
Fax 503.612.2383
cui.com
techsupport@cui.com
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause undesired operation.
CUI offers a one (1) year limited warranty. Complete warranty information is listed on our website.
CUI reserves the right to make changes to the product at any time without notice. Information provided by CUI is believed to be accurate and reliable. However, no responsibility is
assumed by CUI for its use, nor for any infringements of patents or other rights of third parties which may result from its use.
CUI products are not authorized or warranted for use as critical components in equipment that requires an extremely high level of reliability. A critical