MDS-030AAC24 [DELTA]

Medical AC-DC Adapter;
MDS-030AAC24
型号: MDS-030AAC24
厂家: DELTA ELECTRONICS, INC.    DELTA ELECTRONICS, INC.
描述:

Medical AC-DC Adapter

文件: 总9页 (文件大小:1237K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
TECHNICAL DATASHEET  
Medical AC-DC Adapter  
MDS-030AAC Series / MDS-030AAC□  
030AAC  
Highlights & Features  
MDS-030AAC05, 12, 15, 24 Meet Efficiency Level VI  
MDS-030AAC07 Meet Efficiency Level V  
Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1  
Compliant with IEC 60601-1-2 4th Ed. Requirements  
Meets Limited Power Source (LPS) requirements  
Low touch current. Suitable for type BF applications  
Detachable AC plug with multiple country options  
IP22 ingress protection rating  
2 × MOPP isolation  
500K hours MTBF  
3 years warranty  
Safety Standards  
CB Certified for worldwide use  
Model Number:  
Unit Weight:  
MDS-030AAC☐  
150 grams  
Dimensions (W × L × H): 53.5 × 88.0 × 27.5 mm  
2.11 × 3.46 × 1.08 inch  
General Description  
The MDS series of external power supply comes with universal AC input at 85Vac to 264Vac. Other features include low touch current,  
risk management report available and the electric shock protection comply with 2 × MOPP. The MDS series is certified for EMC  
standards according to EN 55011 for industrial, scientific and medical (ISM) radio-frequency equipment and EN 55022 for Information  
Technology Equipment (ITE) radio-frequency equipment. In addition, only recognized Japanese capacitors are used.  
The MDS series come with both medical and ITE safety approvals including UL/cUL and IEC-60950-1 CB certification as a Limited  
Power Source (LPS), and CCC approval. Designs are fully compliant with RoHS Directive 2011/65/EU for environmental protection.  
Model Information  
Medical AC-DC Adapter  
Model Number  
MDS-030AAC05  
MDS-030AAC07  
MDS-030AAC12  
MDS-030AAC15  
MDS-030AAC24  
Input Voltage Range Efficiency Level  
Rated Output Voltage  
Rated Output Current  
85-264Vac  
Level VI  
Level V  
Level VI  
Level VI  
Level VI  
5Vdc  
3A  
7Vdc  
3A  
12Vdc  
15Vdc  
24Vdc  
2A  
2A  
1.25A  
Model Numbering  
MDS –  
030  
AAC  
Delta Medical  
Power Supply  
Max wattage in the  
product series  
Family Code  
Output Voltage (Single  
Output)  
Revision Code1  
Revision Code2  
DC plug type and output cable length  
Country duck-head type  
- 05 for 5V  
- 07 for 7V  
- 12 for 12V  
- 15 for 15V  
- 24 for 24V  
A2.1 x 5.5 x10mm DC plug  
B2.1 x 5.5 x12mm DC plug  
AChina ; BUnited States;  
CEuropean; DUnited Kingdom;  
EAustralia; GKorea; HIndia,  
J Argentina ; KBrazil;  
Both A & B are 1200mm cable length  
MSouth Africa  
* MDS-030AAC15 BB is not available.  
All parameters are specified at 25°C ambient unless otherwise indicated.  
www.DeltaPSU.com (May 2017, Rev.07)  
1
TECHNICAL DATASHEET  
Medical AC-DC Adapter  
MDS-030AAC Series / MDS-030AAC□  
Specifications  
MDS-030AAC12  
MDS-030AAC15  
MDS-030AAC24  
Model Number  
MDS-030AAC05  
MDS-030AAC07  
Input Ratings / Characteristics  
Nominal Input Voltage  
Input Voltage Range  
100-240Vac  
85-264Vac  
50-60Hz  
Nominal Input Frequency  
Input Frequency Range  
47-63Hz  
Input Current (max)  
115Vac 0.5A  
230Vac 0.3A  
0.6A  
0.8A  
0.8A  
0.8A  
0.4A  
0.6A  
0.6A  
0.6A  
Average Efficiency (min)  
Standby Power (max)  
115Vac  
230Vac  
115Vac  
230Vac  
81.4%  
0.1W  
81.4%  
86.21%  
87.0%  
87.0%  
0.1W  
0.1W  
0.1W  
0.1W  
Inrush Current (typ.)  
Touch Current (max)  
60A @ 115Vac and 100A @ 230Vac  
0.1mA @ 264Vac NC1), 0.3mA @ 264Vac SFC2)  
0.1mA @ 264Vac NC1), 0.5mA @ 264Vac SFC2)  
Output-PE (protective earth)  
leakage current for Type BF  
application (max)  
1)  
2)  
NC: Normal condition  
SFC: Single fault condition  
MDS-030AAC12  
MDS-030AAC15  
15Vdc  
MDS-030AAC24  
24Vdc  
Model Number  
MDS-030AAC05  
5Vdc  
MDS-030AAC07  
Output Ratings / Characteristics  
Nominal Output Voltage  
Output Total Regulation  
Output Current  
7Vdc  
12Vdc  
±6%  
2A  
3A  
3A  
2A  
1.25A  
30W  
Output Power  
15W  
±1%  
±5%  
21W  
24W  
30W  
Line Regulation  
Load Regulation  
Ripple & Noise (max)  
100mVpk-pk @  
Rated load  
100mVpk-pk @  
Rated load  
150mVpk-pk @  
Rated load  
200mVpk-pk @  
Rated load  
200mVpk-pk  
@ Rated load  
Start-up Time (max)  
Hold-up Time (min)  
3000ms @ 115Vac  
10ms @ 115Vac  
Dynamic Response (Overshoot & ±10% @ 50-100% load  
Undershoot O/P Voltage)  
Capacitive load (max)  
4700uF  
*Periodic and Random Deviation.  
Ripple & Noise measurement circuit with 20MHz BW  
All parameters are specified at 25°C ambient unless otherwise indicated.  
www.DeltaPSU.com (May 2017, Rev.07)  
2
TECHNICAL DATASHEET  
Medical AC-DC Adapter  
MDS-030AAC Series / MDS-030AAC□  
MDS-030AAC12  
MDS-030AAC15  
MDS-030AAC24  
Model Number  
MDS-030AAC05  
MDS-030AAC07  
Mechanical  
Case Chassis  
Case Cover  
PC  
PC  
Dimensions (W × L × H)  
Unit Weight  
53.5 × 88.0 × 27.5 mm (2.11 × 3.46 × 1.08 in)  
150 grams (5.3 ounces)  
Indicator  
NA  
Cooling System  
Terminal  
NA  
Input Detachable type AC plug  
Output Barrel type. Dimensions 2.1 × 5.5 × 10 mm (see page 5)  
Environment  
Surrounding Air  
Temperature  
Operating 0°C to +40°C  
Storage -40°C to +85°C  
Operating Humidity  
Operating Altitude  
5 - 95% RH (Non-Condensing)  
5,000 meters (16400 feet)  
Shock Test  
50G, 11ms, 3 shocks for each direction  
(Non-Operating)  
Vibration  
5-500Hz, 2.09Grms, 20 minute for each three axis  
(Operating)  
Protections  
Overvoltage (max)  
150%, Latch Mode  
Overload / Overcurrent (max)  
Over Temperature  
Short Circuit  
250% of rated load current, Hiccup Mode,  
(Non-Latching, Auto-Recovery)  
Hiccup Mode,  
(Non-Latching, Auto-Recovery)  
Hiccup Mode,  
(Non-Latching, Auto-Recovery)  
Degree of Protection  
IP22  
Protection Against Shock  
Class II  
Reliability Data  
MTBF (min)  
500K Hours based on Telecordia SR-332 (at 100Vac, Max. load and 25°C Ambient)  
All parameters are specified at 25°C ambient unless otherwise indicated.  
www.DeltaPSU.com (May 2017, Rev.07)  
3
TECHNICAL DATASHEET  
Medical AC-DC Adapter  
MDS-030AAC Series / MDS-030AAC□  
MDS-030AAC12  
MDS-030AAC15  
MDS-030AAC24  
Model Number  
MDS-030AAC05  
MDS-030AAC07  
Safety Standards / Directives  
Medical Safety  
IEC60601-1 Edition 3.1 (2012), EN60601-1 (2006) + A11 + A1 + A12, CAN/CSA-C22.2 NO.  
60601-1:14, ANSI/AAMI ES60601-1:2005/(R)2012  
UL60601-1 (File No. E356265)  
ITE Safety  
IEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-2008, GB17625.1-2003  
UL60950-1 (File No. E131881), LPS (Limited Power Source)  
MDD Directive 93/42/EEC  
CE  
Material and Parts  
RoHS Directive 2011/65/EU Compliant  
Galvanic Isolation  
Input to 4000 Vac  
Output  
EMC (Compliant with IEC 60601-1-2 4th Ed. Requirements)  
EMC / Emissions  
EN55011/EN55022,FCC Title 47:Class B  
Harmonic Current Emissions  
Immunity to  
IEC61000-3-2  
Voltage Flicker  
IEC61000-3-3  
Electrostatic Discharge  
IEC61000-4-2 Level 4 Criteria A1)  
Air Discharge: 15kV  
Contact Discharge: 8kV  
Criteria A1)  
Radiated Field  
IEC61000-4-3  
80MHz-2700MHz, 10V/m AM modulation  
385MHz-5785MHz, 28V/m Pulse mode and other  
modulation  
Electrical Fast Transient / Burst  
Surge  
IEC61000-4-4 Level 3 Criteria A1): 2kV  
IEC61000-4-5 Level 3 Criteria A1)  
Differential Mode3): 1kV  
Level 2 Criteria A1)  
IEC61000-4-6  
Conducted  
150kHz-80MHz, 3Vrms, 6Vrms at ISM bands and  
Amateur radio bands  
IEC61000-4-8 Criteria A1)  
Power Frequency Magnetic Fields  
Voltage Dips  
Magnetic field strength 30A/m  
Criteria A1)  
IEC61000-4-11  
0% UT, 0.5 cycle (10ms)  
, 0o/45o/90o/135o/180o/225o/270o/315o/360o  
Criteria B2)  
0% UT, 1 cycle (20ms), 0o  
Criteria B2)  
70% UT, 25 cycle (500ms), 0o  
Criteria B2)  
0% UT, 250 cycle (5000ms), 0o  
1) Criteria A: Normal performance within the specification limits  
2) Criteria B: Output out of regulation, or shuts down during test. Automatically restore to normal operation after test.  
3) Symmetrical: Differential mode (Line to line)  
Block Diagram  
All parameters are specified at 25°C ambient unless otherwise indicated.  
www.DeltaPSU.com (May 2017, Rev.07)  
4
TECHNICAL DATASHEET  
Medical AC-DC Adapter  
MDS-030AAC Series / MDS-030AAC□  
Dimensions  
W × H × H: 53.5 × 88.0 × 27.5 mm  
Note:  
The plug’s polarity is  
All parameters are specified at 25°C ambient unless otherwise indicated.  
www.DeltaPSU.com (May 2017, Rev.07)  
5
TECHNICAL DATASHEET  
Medical AC-DC Adapter  
MDS-030AAC Series / MDS-030AAC□  
Duck Head Type:  
CN type  
US/JP type  
EU type  
UK type  
AU  
KR type  
IN type  
AR type  
BZ type  
SA type  
All parameters are specified at 25°C ambient unless otherwise indicated.  
www.DeltaPSU.com (May 2017, Rev.07)  
6
TECHNICAL DATASHEET  
Medical AC-DC Adapter  
MDS-030AAC Series / MDS-030AAC□  
Functions  
Start-up Time  
The time required for the output voltage (Vo) to reach 90% of its set value, after the input AC voltage is applied.  
Rise Time  
The time required for the output voltage (Vo) to change from 10% to 90% of its steady state value.  
Hold-up Time  
Hold up time is the time when the AC input collapses and output voltage retains regulation for a certain period of time. The time  
required for the output to reach 95% of its set value, after the input voltage is removed.  
Graph illustrating the Start-up Time, Rise Time, and Hold-up Time  
Inrush Current  
Dynamic Response  
Inrush current is the input current that occurs when the input  
voltage is first applied. For AC input voltages, the maximum  
peak value of inrush current will occur during the first half cycle  
of the applied AC voltage. This peak value decreases  
exponentially during subsequent cycles of AC voltage.  
The power supply output voltage will remain within ±3% of its  
steady state value, when subjected to a dynamic load change  
from 50 to 100% of its rated current.  
All parameters are specified at 25°C ambient unless otherwise indicated.  
www.DeltaPSU.com (May 2017, Rev.07)  
7
TECHNICAL DATASHEET  
Medical AC-DC Adapter  
MDS-030AAC Series / MDS-030AAC□  
Overload & Over current Protections  
Overvoltage Protection  
The power supply’s Overload (OLP) and Over current (OCP)  
Protections will be activated when output current is between  
110% and 250% of IO (Max load). Upon such an occurrence, VO  
will start to drop. Once the power supply has reached its  
maximum power limit, the protection will be activated and the  
power supply will go into “Hiccup mode” (Auto-Recovery). The  
power supply will recover once the fault condition causing the  
OLP and OCP is removed and IO is back within the specified limit.  
The power supply’s overvoltage circuit will be activated when its  
internal feedback circuit fails. The output voltage shall not exceed  
its specifications defined on Page 3 under “Protections”. Power  
supply will latch off, and require removal/re-application of input  
AC voltage in order to restart.  
Over Temperature Protection  
As mentioned above, the power supply also has Over  
Temperature Protection (OTP). This is activated when the  
overload condition persists for an extended duration and the  
output current is below the overload trigger point but >100% load.  
In the event of a higher operating condition at 100% load, the  
power supply will run into OTP when the surrounding air  
temperature is higher than the operating temperature. When  
activated, the output voltage will go into hiccup mode until the  
input voltage is removed; then, reapplied, and the surrounding air  
temperature drops to its normal operating temperature.  
Additionally, if the IO is <250% but >110% for a prolong period of  
time (depending on the load), the Over Temperature Protection  
(OTP) will be activated due to high temperature on critical  
components. The power supply will then go into hiccup mode until  
the fault is removed; and, the input voltage is removed, then  
reapplied.  
Short Circuit Protection (Auto-Recovery)  
The power supply’s output OLP/OCP function also provides  
protection against short circuits. When a short circuit is applied,  
the output current will operate in “Hiccup mode”, as shown in the  
illustration in the OLP/OCP section on this page. The power  
supply will return to normal operation after the short circuit is  
removed.  
All parameters are specified at 25°C ambient unless otherwise indicated.  
www.DeltaPSU.com (May 2017, Rev.07)  
8
TECHNICAL DATASHEET  
Medical AC-DC Adapter  
MDS-030AAC Series / MDS-030AAC□  
Certificate  
All Delta Medical Power products conform to the European directive 2011/65/EU. RoHS is the abbreviation for  
“Restriction of the use of certain hazardous substances.  
Typically, the input current waveform is not sinusoidal due to the periodical peak charging of the input capacitor.  
In industrial environment, complying with EN 61000-3-2 is only necessary under special conditions. Complying  
to this standard can have some technical drawbacks, such as lower efficiency as well as some commercial  
aspects such as higher purchasing costs. Frequently, the user does not profit from fulfilling this standard,  
therefore, it is important to know whether it is mandatory to meet this standard for a specific application.  
In addition to a UL Total Certification Program (TCP) approved client laboratory for IEC60950 and IEC60065.  
Delta also has participated UL Client Test Data Program (CDTP) for IEC 60601.  
Meet Level V Efficiency Requirement for MDS-030AAC07  
Meet Level VI Efficiency Requirement for MDS-030AAC05, 12, 15, 24  
Meets Limited Power Source (LPS) requirement  
All parameters are specified at 25°C ambient unless otherwise indicated.  
www.DeltaPSU.com (May 2017, Rev.07)  
9

相关型号:

MDS-040APS12-B

MDS Medical AC-DC Open Frame
DELTA

MDS-040APS12B

Medical AC-DC Open Frame
DELTA

MDS-040APS15-B

MDS Medical AC-DC Open Frame
DELTA

MDS-040APS15B

Medical AC-DC Open Frame
DELTA

MDS-040APS18-B

MDS Medical AC-DC Open Frame
DELTA

MDS-040APS18B

Medical AC-DC Open Frame
DELTA

MDS-040APS24-B

MDS Medical AC-DC Open Frame
DELTA

MDS-040APS24B

MDS Medical AC-DC Open Frame
DELTA

MDS-060AAS15-B

MDS Medical AC-DC Adapter
DELTA

MDS-060AAS15B

Medical AC-DC Adapter
DELTA

MDS-060AAS19-B

MDS Medical AC-DC Adapter
DELTA

MDS-060AAS19B

Medical AC-DC Adapter
DELTA