1620-3-N [ETC]
Signal Conditioner ; 信号调理\n型号: | 1620-3-N |
厂家: | ETC |
描述: | Signal Conditioner
|
文件: | 总6页 (文件大小:1500K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
Model 1620
Disposable Blood Pressure Sensor
AAMI Specification
þ Disposable Blood Pressure
þ Measurements
þ Kidney Dialysis Machines
þ Medical Instrumentation
Low Cost
Multiple Configurations
DESCRIPTION
The Model 1620 is a fully piezoresistive silicon pressure sensor
for use in invasive blood pressure monitoring. The sensor is
designed to be used with automated assembly equipment and
can be dropped directly into a customer’s disposable blood
pressure housing. The sensor is designed to meet the
requirements as described in the Association for the
Advancement of Medical Instrumentation (AAMI) specification
for Blood Pressure Transducers.
The pressure sensor consists of a pressure sensing element
mounted on a ceramic substrate. Thick-film resistors on the
ceramic substrate are laser-trimmed for compensation and
calibration.
A plastic cap is attached to the ceramic substrate to provide
an easy method of attachment to the customers assembly and
protection for the sensing element. A dielectric gel is placed
over the sensor to provide electrical and fluid isolation.
The Model 1620 pressure sensors are batch manufactured
in a 10x12 element array on a ceramic substrate (120 units per
substrate). The products are shipped in anti-static shipping
containers. Performance characteristics and packaging can be
easily tailored on a special order basis to meet the requirements
of specific customers.
FEATURES
þ Low Cost Disposable Design
þ Solid State Piezoresistive Sensor
þ Top Side Pressure Entry
þ Compatible with Automated Assembly
Equipment
þ Integral Dielectric Gel Barrier
þ Fully Tested and Compensated
STANDARD RANGE
-30 to 300
mmHg
1-48
Blood Pressure
Disposable Blood Pressure Sensor
PERFORMANCE SPECIFICATIONS
Model 1620
Supply Voltage: 6.0 VDC
Ambient Temperature: 25°C (Unless otherwise specified)
PARAMETERS
MIN
TYP
MAX
UNITS
NOTES
Operating Pressure Range
Overpressure
-30
300
mmHg
PSI
125
Zero Pressure Offset
Sensitivity
20
5.05
102.5
1
mmHg
4.95
97.5
5.0
µV/V/mmHg
mmHg
mmHg
% output
% output
Ω
Calibration
100
1
Linearity & Hysteresis (-30 to 100 mmHg)
Linearity & Hysteresis (>100 to 200 mmHg)
Linearity & Hysteresis (>200 to 300 mmHg)
Input Impedance
1
1.5
3200
315
1.05
10
1200
285
0.95
1
2
7
Output Impedance
Ω
Output Symmetry
ratio
Supply Voltage
6
5
VDC
Risk Current (at 120 VAC rms 60 Hz)
Warm-up Time
2
mA
seconds
Hz
Frequency Response
Offset Drift
1200
2
mmHg
%/°C
3
4
4
Thermal Span Shift
0.1
0.3
5
Thermal Offset Shift
Phase Shift (at 5 kHz)
Light Sensitivity (3000 Foot Candle)
Defibrillator Withstand (400 Joules)
Sterilization (ETO)
mmHg/°C
degrees
mmHg
discharges
cycles
1
5
5
6
3
Operating Temperature
Storage Temperature
Humidity
+10°C to +40°C
-25°C to +70°C
10 - 90% (non-condensing)
168 hours
Operating Product Life
Shelf Life
2 years
Dielectric Breakdown
Media Interface
10,000 VDC
Dielectric Gel
4.5 x 10-4 in3
0.5 grams
Volume Displacement
Weight
Notes
1. Output of sensor with no pressure applied and a 150kΩ resistor
5. One discharge per minute.
shorted across +VIN to +OUT.
6. Sterilization performed by customer.
2. For input impedance of 350Ω
5% select pad configuration 1.
7. Defined as common mode symmetry between signal output and either
excitation terminal.
3. Over an 8 hour time period and after warm-up.
4. Over operating temperature range (+10°C to +40°C).
Blood Pressure
1-49
Model 1620
Disposable Blood Pressure Sensor
CONNECTIONS AND DIMENSIONS
All dimensions are in inches [millimeters].
BODY STYLE N
*Dimensions indicated do not include flares.
BODY STYLE W
*Dimensions indicated do not include flares.
Notes
1. Materials Used:
SECTION A-A
Substrate: 96% Alumina
Transducer (Die): Silicon
Die Attach Adhesive: Room Temperature Vulcanizer
Lid Adhesive: Medical Grade UV Curing Adhesive
Conductor And Contact Pads: Palladium-Silver Alloy
Wire Bonds And Bond Pads: Gold
Resistors: Ruthenium-Based Thick Film Paste
Solder Dams: Green Glass
Protective Gel Lid: Rad-Stable Poly carbonate Resin
2. Miniscus of Gel:
Max dimension below surface A = .035" [0.89].
Max dimension above surface A = .001" [0.03].
3. All dimensions taken at maximum draft.
4. All unspecified fillets and radii are.015" [0.38].
5. All draft angles 1° maximum.
1-50
Blood Pressure
Disposable Blood Pressure Sensor
Model 1620
STANDARD PAD CONFIGURATIONS
All dimensions are in inches [millimeters].
PRELIMINARY
Note
1. Custom pad configurations not shown here may be special ordered. Call IC Sensors for details.
SHIPPING INFORMATION
Notes
1. Sensors are shipped as 120 UP snapstrates and must be singulated by the purchaser.
2. Each plate may include units that have failed visual or electrical parameters as well as good units. Bad units are identified with a dot on the backside of
the cell location.
3. Plates are shipped in dust free anti-static containers to prevent contamination of the gel surface.
SCHEMATIC DIAGRAM
ORDERING INFORMATION
1620 - 1 - N
Body Style (N = Narrow, W = Wide)
Pad Configuration (1, 2, 3, 4)
Model
April 2000
Blood Pressure
1-51
相关型号:
1620-3H-10A
Protection of electrical systems of passenger cars, trucks, buses, watercraft, extra low voltage wiring systems
ETAL
1620-3H-15A
Protection of electrical systems of passenger cars, trucks, buses, watercraft, extra low voltage wiring systems
ETAL
1620-3H-20A
Protection of electrical systems of passenger cars, trucks, buses, watercraft, extra low voltage wiring systems
ETAL
1620-3H-30A
Protection of electrical systems of passenger cars, trucks, buses, watercraft, extra low voltage wiring systems
ETAL
1620-3H-5A
Protection of electrical systems of passenger cars, trucks, buses, watercraft, extra low voltage wiring systems
ETAL
1620-3H-7.5A
Protection of electrical systems of passenger cars, trucks, buses, watercraft, extra low voltage wiring systems
ETAL
©2020 ICPDF网 联系我们和版权申明