PA-1001A1AP1X [ETC]

SIGNAL CONDITIONER & CONVERTERS; 信号调理器和转换器
PA-1001A1AP1X
型号: PA-1001A1AP1X
厂家: ETC    ETC
描述:

SIGNAL CONDITIONER & CONVERTERS
信号调理器和转换器

转换器
文件: 总11页 (文件大小:189K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
INSTALATION, OPERATION & MAINTENANCE  
MANUAL  
FOR  
PA-1001A Series  
SIGNAL CONDITIONER  
& CONVERTORS  
PA1001A – 7/02  
Page 1 of 11  
SIGNAL CONDITIONER &  
CONVERTERS  
PA1001A Series  
INTRODUCTION:  
The PA1001A series preamplifiers/signal conditioners and converters are suitable for use  
with either magnetic or RF type pickup coils and allow convenient interfacing with process  
measuring devices and digital electronics.  
When used with Magnetic type pickups they accept low level signals and produces a pulse  
output, while providing rejection of unwanted noise and false signals.  
A RF type pickup reduces the pickup drag associated with conventional magnetic pickups,  
resulting in a significant increase in the usable range. The PA1001A series excite a RF type  
pickup with a 20KHz signal. Motion of ferrous material is sensed and the motion modulates  
the coil field and subsequent conditioning provides a pulse output signal.  
The output signal of the PA1001A series have a square pulse wave form, each pulse  
representative of an actuator (rotor blade, gear tooth, turbine blade, etc.) passing by the  
sensing area of the pickup.  
Four output signal options are available:  
i.  
TTL/CMOS compatible  
ii.  
Open collector  
iii.  
iv.  
0 to 10VDC square wave and  
A square wave that is tied to the input power  
.
The unit is powered by a user supplied 8 to 30 VDC filtered power supply. An on-board  
regulator provides the required regulation and noise rejection.  
Optional features include factory installed pulse scaling and field installable plug in modules  
for Analog output. Pulse scaling provides the user with scaling options of ÷2, ÷4, ÷8, ÷16,  
÷32, ÷64, or ÷128, which is useful when a remote device has a maximum count speed that  
is less than the pulse-generating device. Analog output provides a current and voltage  
outputs proportional to RPM, flow rate etc.  
PA1001A – 7/02  
Page 2 of 11  
CHARACTERISTICS AND SPECIFICATIONS:  
POWER INPUT:  
Voltage: 8-30 VDC  
Current Draw: 125mA @ 24VDC, with all options equipped.  
Reverse polarity protected  
PICKUP INPUT:  
RF Pickup:  
Modulation Carrier Frequency: 20KHz  
Input Frequency: 2 Hz to 5KHz, adjustable trigger level  
Trigger Sensitivity: 20 millivolt peak to peak (minimum)  
Carrier Adjustment: to allow precise tuning of carrier signal to physical environment (e.g. air gap,  
wall thickness)  
Maximum Air Gap: 0.160”  
Preferred Coil Inductance: 1.0 milliHenry +/- 10%  
Maximum Distance From Pickup To Preamplifier 100 feet  
Magnetic Pickup:  
Input protected, RF and band pass filtered  
Input Frequency: 0 to 10KHz, adjustable trigger level  
Trigger Sensitivity: 4 millivolt RMS (minimum)  
NOTE: At 4mV frequency range 10 Hz to 3KHz  
Input Impedance: 40K Ohms (nominal)  
Maximum Input Signal: 600 Volts RMS  
Maximum Distance From Pickup To Preamp: 100 feet  
PULSE OUTPUTS:  
TTL/CMOS (Standard); Fan-out of 5 TTL/CMOS loads  
Open Collector: Maximum OFF state voltage 30 VDC  
Maximum ON current 0.40 amps  
0 to 10 VDC square wave  
Square wave tied to input voltage  
Pulse scaling (option): ÷2, ÷4, ÷8, ÷16, ÷32, ÷64, or ÷128 user selectable.  
CURRENT OUTPUT OPTION:  
Accuracy +/- 0.05% of full-scale +/-200 PPM/°C.  
Available range 4 to 20mA.  
Output suitable for driving floating or grounded loads.  
Maximum loop impedance 500 ohms.  
Response time 0.3 seconds for 10 to 90 %.  
VOLTAGE OUTPUT OPTION:  
Accuracy +/- 0.05% of full-scale +/-200 PPM/°C.  
Available range 0 to 5VDC or 0 to 10VDC, user selectable.  
Impedance less than 10 ohms.  
Response time 0.3 seconds for 10 to 90 %.  
PA1001A – 7/02  
Page 3 of 11  
ENVIRONMENTAL:  
OPERATING: -40°C TO 85°C  
STORAGE: -40°C TO 150°C  
ENCLOSURES:  
Unit is intended for mounting in a conduit box and comes with 2 ½” on center mounting holes. An  
optional explosion proof or NEMA-4x enclosure is available upon request.  
APPROVALS:  
EN50081-1:1992 Generic Emission Standard  
EN50082-2:1995 Generic Immunity Requirements  
CE compliant to EMC Directive 89/336/EECfor uses in residential, commercial, light or heavy  
industrial environments.  
INSTALLATION AND WIRING DETAILS:  
In considering the interconnections between the PA1001A and the measurement system some  
attention must be given to anticipated noise sources and to the coupling of these noise sources to the  
interconnecting wiring.  
Noise signals may be coupled into the wiring between the pickup coil, the PA1001A, and the  
electronic measuring systems. In general, utilizing a shielded, twisted pair for the interconnection  
greatly reduces this coupling. The shield should be grounded on both ends of the cable.  
However, even with proper interconnecting cabling, cross talk with other signal lines or power lines  
may still occur and should be avoided. Physical isolation of the wiring reduces the chance of  
potential problems.  
The pickup coil should not be located more than 100 feet from the PA1001A preamplifier. It is  
recommended that the PA1001A be installed on or near the pickup coil to assure optimum  
performance.  
INSTALLATION:  
The Model PA1001A should be placed in a convenient location that maintains access to the unit  
should repairs or readjustment is required.  
Connections to the terminal block should be carefully dressed to avoid having bare wires extend pass  
the screw clamp on the terminal block. This is particularly important when the PC card is mounted  
within an explosion proof enclosure. Wires should be neatly dressed near bottom of enclosure to  
prevent problems when cover is installed.  
For compliance with EMI/RFI requirements the pre-amplifier must be installed in a metal enclosure  
such as a suitable NEMA 4 (e.g. GUAC) junction box. The enclosure must be properly grounded.  
Also a grounding strap is required from the enclosure to the power supply common (pin 7).  
The cable entry requires 360° grounding coverage to the enclosure case. This is obtained by using  
twisted pair with full braid cable and a cable fitting that provides a metal to metal cable clamping  
connection. The insulation of the cables should be stripped back to allow for connection to the  
terminal block and allow for the cable fitting to clamp onto the braided shield. Any open areas should  
then be covered using foil tape.  
In hazardous locations where metal conduit is used, 360° grounding are achieved. All cable shields  
should be terminated to the enclosure-grounding stud.  
PA1001A – 7/02  
Page 4 of 11  
Wiring For RF Pickup Coil:  
1. Connect a two conductor shielded cable from the pickup to Terminals 5 and 6. Connect the  
shield to Terminal 4 (see Figure 1). NOTE: It is also preferable that the pickup coil be grounded  
to the shield.  
2. Connect a power source, which must be within 8-30 VDC, to Terminals 7 and 8. The Plus  
voltage is connected to Terminal 8 and the Return to Terminal 7 (see Figure 1).  
3. Connect the Pulse output wire to Terminal 3, and connect the Pulse common to Terminal 4 (see  
Figure 1).  
4. Set Jumper HD2 (see Figure 3) to the RF position.  
5. Configure Jumper HD1 (see Figure 3) to the desired pulse output as follows:  
Pulse Output Type  
Open Collector  
HD1 Jumper Configuration  
JP2  
Square Wave to Vin  
TTL/CMOS  
JP1 and JP2  
JP1, JP2 and JP3  
JP1, JP2 and JP4  
0-10 VDC Square Wave  
6. Install pickup (e.g. in meter) and provide power to PA1001A. Use a digital voltmeter set to DC  
voltage and attach the positive lead to Terminal 6 and the common lead to Terminal 2. Adjust the  
Carrier Adjustment (R2) (see Figure 3) to a reading of 3.00 VDC to fine-tune the pickup for air  
gap, wall thickness, etc.  
7. With power still supplied to PA1001A determine if pulse output is being generated in the absence  
of motion. Adjust the RF Trigger (R24) (see Figure 3) clock wise until the false pulsation has  
stopped. If pulsation continues when adjusted fully clock wise, check that all shields have been  
properly terminated and grounded.  
PA1001A – 7/02  
Page 5 of 11  
Wiring For Magnetic Pickup Coil:  
1. Connect a two conductor shielded cable from the pickup to Terminals 1 and 2. Connect the  
shield to Terminal 4 (see Figure 2). NOTE: It is also preferable that the pickup coil be grounded  
to the shield.  
2. Connect a power source, which must be within 8-30 VDC, to Terminals 7 and 8. The Plus  
voltage is connected to Terminal 8 and the Return to Terminal 7(see Figure 2).  
3. Connect the Pulse output wire to Terminal 3, and connect the Pulse common to Terminal 4 (see  
Figure 2).  
4. Set Jumper HD2 (see Figure 3) to the MAG position.  
5. Configure Jumper HD1 (see Figure 3) to the desired pulse output as follows:  
Pulse Output Type  
Open Collector  
HD1 Jumper Configuration  
JP2  
Square Wave to Vin  
TTL/CMOS  
JP1 and JP2  
JP1, JP2 and JP3  
JP1, JP2 and JP4  
0-10 VDC Square Wave  
5. Install pickup (e.g. in meter) and provide power to PA1001A. Determine if pulse output  
is being generated in the absence of motion. Adjust the MAG Trigger (R19) (see Figure  
3) clock wise until the false pulsation has stopped.  
PA1001A – 7/02  
Page 6 of 11  
Analog Output Option:  
1. For current output, wire two conductor, shielded cable to terminals 2 (common) and 3 (current  
output).  
2. For voltage output, wire two conductor, shielded cable to terminals 2 (common) and 1 (voltage  
output).  
PA1001A – 7/02  
Page 7 of 11  
CONTROLS AND ADJUSTMENTS:  
STANDARD PRE-AMPLIFIER ADJUSTMENTS:  
CARRIER ADJUSTMENT:  
A twenty turn control used to precisely tune the specific sensing  
depth, wall thickness, material type, air gap,  
etc. of the RF pickup coil.  
RF TRIGGER:  
A twenty turn control used to set the trigger level of the  
RF coil input above the ambient noise level.  
MAG TRIGGER:  
COIL SELECTION:  
A twenty turn control used to set the trigger level of the  
Magnetic coil input above the ambient noise level.  
A two-position jumper that is used to select the desired  
pickup coils to be used as the input.  
Input Coil Type  
RF  
HD2 Setting  
RF  
Magnetic  
MAG  
LIN\RAW PULSE:  
A two-position jumper that is used to select the desired  
Pulse train.  
Pulse Type  
JP1 Setting  
RAW  
Raw pulse train  
Linearized pulse train  
(NOT AVAILABLE YET)  
LIN  
PULSE SELECTION:  
A four-position jumper that is used to select the  
Desired type of pulse output.  
Pulse Output Type  
Open Collector  
HD1 Jumper Settings  
JP2  
Square Wave to Vin  
TTL/CMOS  
JP1 and JP2  
JP1, JP2 and JP3  
JP1, JP2 and JP4  
0-10 VDC Square Wave  
PULSE SCALING ADJUSTMENTS:  
PULSE SCALER:  
A 8 position jumper header located on PA1001 circuit board which is  
used in programming the desired pulse output scaling (Division of 1,  
2, 4, 8, 16, 32, 64, or 128). Factory set with division of 1 as standard.  
PA1001A – 7/02  
Page 8 of 11  
ANALOG OUTPUT ADJUSTMENTS (PCA-121):  
20mA ADJ  
A twenty turn adjustment which is used to set current output to the  
desired span corresponding to the equivalent full scale range. (i.e.  
20 mA at 100 RPM).  
4mA ADJ  
VFS ADJ  
A twenty-turn adjustment that is used to set the current output signal  
to the desired zero value. (i.e. 4mA at zero RPM).  
A twenty-turn adjustment that is used to set the voltage output to the  
desired span corresponding to the equivalent full scale range. (i.e.  
5VDC or 10 VDC at 100 RPM).  
VO ADJ  
A twenty-turn adjustment that is used to set the voltage output signal  
to the desired zero value. (i.e. 0 VDC at zero RPM).  
COARSE RANGE ADJ  
A dual in line DIP switch which is located on the PCA-121 board,  
which is used to perform a coarse range adjustment for the input  
frequency range.  
VFS SELECTION  
A two position jumper (JP1) that is used to select the desired voltage  
output range of either 0 to 5VDC or 0 to 10VDC.  
CALIBRATION PROCEDURE:  
In order to perform field calibration and troubleshoot the PA1001A, the following equipments are  
recommended.  
MANUFACTURER  
Fluke  
PART NUMBER  
8060A  
DESCRIPTION  
True RMS Multimeter  
Digital Function Generator  
Multi Function Counter  
Adjustment Tool  
Topward  
Viz  
8112  
WD-755  
8-TOOO  
Spectrol  
Table - 1  
(Coarse Range Adjustment):  
Frequency Range  
Switch  
Position  
50 to 100Hz  
1
2
3
4
5
6
7
8
100 to 200Hz  
200 to 400Hz  
400 to 800Hz  
800 to 1600Hz  
1600 to 3200Hz  
3200 to 6400Hz  
6400 to 12800Hz  
PA1001A – 7/02  
Page 9 of 11  
FOR CURRENT OUTPUT:  
Connect a digital milliampmeter or equivalent, across the current output terminals.  
On the PCA-121 select the “Coarse Range Adj” that corresponds to the maximum input  
frequency. Refer to Table - 1 for selection options.  
Adjust “4mA ADJ” control for desired “Zero” current (i.e. 4mA).  
Inject the full-scale test frequency while adjusting “20mA ADJ” for current equal to SET (SPAN).  
Iterate above steps until no change is observed.  
FOR VOLTAGE OUTPUT:  
NOTE: For proper adjustment of the voltage-output, First  
Perform calibration of the current output.  
Connect a digital voltmeter across the voltage output terminals.  
On the PCA-121 select the “Coarse Range Adj” that corresponds to the maximum input  
frequency. Refer to Table -1 for selection options.  
Adjust “VO ADJ” for 0.00 volts).  
Inject the test frequency while adjusting “VFS ADJ” for voltage equal to SET (SPAN).  
Iterate above steps until no change is observed.  
Control and Adjustment Detail for the PCA-121  
PA1001A – 7/02  
Page 10 of 11  
Ordering Information  
PA-1001A  
IA  
B
1
w
Series:  
Options:  
1 = Pulse Output  
1A = Pulse and Analog output  
1PS = With pulse scaling  
Mounting:  
C = Explosion Proof Enclosure  
B = Nema-4X Enclosure  
P = Panel Mount  
Operating Voltage:  
1 = 24 VDC (8 to 30 VDC)  
2 = 110 VAC 50/60 Hz  
3 = 220 VAC 50/60 Hz  
Electrical Interface  
W = Mounted Integral to Flowmeter  
X = 10 feet cable attached to mating connector  
Y = 25 feet cable attached to mating connector  
Z = Customer Specific Length (250 feet max.)  
Ordering Information For  
PCA-121 For Digi-Pulse Pickup  
PCA-121  
Series:  
B
1
w
Mounting:  
C = Explosion Proof Enclosure  
B = Nema-4X Enclosure  
P = Panel Mount  
Operating Voltage:  
1 = 24 VDC (8 to 30 VDC)  
2 = 110 VAC 50/60 Hz  
3 = 220 VAC 50/60 Hz  
PA1001A – 7/02  
Page 11 of 11  

相关型号:

SI9130DB

5- and 3.3-V Step-Down Synchronous Converters

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY

SI9135LG-T1

SMBus Multi-Output Power-Supply Controller

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY

SI9135LG-T1-E3

SMBus Multi-Output Power-Supply Controller

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY

SI9135_11

SMBus Multi-Output Power-Supply Controller

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY

SI9136_11

Multi-Output Power-Supply Controller

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY

SI9130CG-T1-E3

Pin-Programmable Dual Controller - Portable PCs

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY

SI9130LG-T1-E3

Pin-Programmable Dual Controller - Portable PCs

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY

SI9130_11

Pin-Programmable Dual Controller - Portable PCs

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY

SI9137

Multi-Output, Sequence Selectable Power-Supply Controller for Mobile Applications

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY

SI9137DB

Multi-Output, Sequence Selectable Power-Supply Controller for Mobile Applications

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY

SI9137LG

Multi-Output, Sequence Selectable Power-Supply Controller for Mobile Applications

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY

SI9122E

500-kHz Half-Bridge DC/DC Controller with Integrated Secondary Synchronous Rectification Drivers

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY