EDS5-88 [FUJI]
OPTICAL ABSOLUTE PRESSURE TRANSMITTER; 光绝对压力变送器型号: | EDS5-88 |
厂家: | FUJI ELECTRIC |
描述: | OPTICAL ABSOLUTE PRESSURE TRANSMITTER |
文件: | 总8页 (文件大小:123K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
FFI SYSTEM
OPTICAL ABSOLUTE PRESSURE TRANSMITTER
DATA SHEET
FBK···3
The Model FBK 3 Optical Absolute Pressure Transmitter
measures absolute pressures of various fluids accurately,
converts them into optical digital signals and outputs them.
This is an intelligent transmitter providing excellent perfor-
mance and functions due to incorporation of electrostatic
capacitance type silicon sensor and microprocessor.
A fiber optical cable used for the signal transmission line
forms an optical field instrumentation system together with
an optical star coupler and a master station.
FEATURES
1. Resistive to noise and lightning
Optical signal ensures a reliable signal transmission, be-
cause it is not affected by external noise and inductive
lightning. Use of a nonmetallic optical (fiber) cable pre-
vents propagation of inductive lightning through the cable,
so a signal transmission immune to lightning can be re-
alized.
2. Reliability due to redundant configuration
Host system can be duplicated by using two optical cable
trunk lines (between an optical star coupler and host
system). This enhances reliability of users’ systems.
3. Intrinsic safety type explosion-proof
SPECIFICATIONS
Functional specifications
Each equipment with a built-in battery can be constructed
so as to be an intrinsic safety type individually (intrinsic
safety type barrier unnecessary).
• Fluids measured: Liquid, gas or steam
• Measuring range, operating pressure and maximum
allowable over-pressure:
Span
[kPa abs]
Range limits
[kPa abs]
Max.
allowable
over-
pressure
[MPa abs]
Operating
pressure
[kPa abs]
Type
Mini-
mum
value
Maxi-
mum
value
Lower
Upper
range
limit
range
limit
FBK 01 0 to 16
FBK 02 0 to 130
FBK 03 0 to 500
FBK 04 0 to 3000
1.6
3.25
12.5
75
16
130
500
3000
0
0
0
0
16
130
500
3000
0.5
0.5
1.5
9
• Process temperature: –40 to +85°C
EDS5-88
Date April 15, 2005
FBK···3
• Self-diagnosis: Displayed on indication unit (option) and
Performance specifications
transmitted to master station.
±0.2% when measuring span is 1/10 or more
of maximum span
Maximum span
Accuracy rating(Note)
Item
Host system Indication unit
Measuring range abnormal
Detecting unit failure
Amplifier abnormal
Battery voltage
± 0.1+ 0.01
%
Measuring span
when measuring span is less than
1
/
10 of maximum span
Battery voltage low alarm
Note: Percent value with respect to measuring span (including
linearity, hysteresis and repeatability in standard 23°C status)
• Remote control function:
See Table 1.
• Ambient temperature effect:
Zero shift: ±(0.125 + 0.2 × URL/x )%/ 28°C
Overall shift: ±(0.175 + 0.2 × URL/x )%/
28°C
• Output signal: Optical digital signal
• Power supply: Built-in lithium battery (expected life about
4 years)
• Optical cable: Code set type, silica fiber ... core/clad di-
ameter 100/140 µm
Where; URL: Maximum span
x :
Measuring span
• Optical connector:
2 times as large as above when the 7th
digit (material) is other than V.
FC connector
• Transmission distance:
• Maximum allowable pressure effect:
1.5 km max. (when transmission loss of
optical cable is 4 dB/km)
Zero shift at maximum span
±0.2% to maximum allowable pressure
• Damping:
• Zero elevation and suppression:
Possible within a range from 0 kPa abs
to maximum span.
• Explosion-proof:
Intrinsic safety type, JIS ib IIC T3
• Ambient temperature:
Variable from 0.2 to 32 sec (time constant)
• Inclination effect:
0.1 kPa/10°
• Measurement period:
0.2 sec
• Response time:Time constant 0.2 sec (at 23°C)
Dead time; About 0.2 sec
–30 to +70°C
–10 to +60°C for intrinsic safety explo-
Physical specifications
sion-proof type
Material:
–20 to +70°C when provided with indica-
tor
• Storage temperature:
–40 to +80°C
Detecting unit
Material
code
Process cover
Seal diaphragm Other wetted parts
V
H
M
T
SCS14
SCS14
SCS14
SCS14
SUS316L
Hastelloy-C
Monel
SUS316
Hastelloy-C
Monel
Tantalum
Tantalum
• Environmental protection:
Meets JIS C0920, immersion-proof
(equivalent to IEC IP67 or NEMA 6/6P)
• Process connection:
Rc1/4 or 1/4-18NPT (whichever selected
by code symbol)
Oval flange threads 7/16-20UNF
• Optical cable connection:
G1/2 or 1/2 -14NPT (whichever selected
by code symbol)
• Mounting method:
Mounted on the 50A (2B) pipe with U-
bolt
• Finish:
Epoxy-polyurethane double coat,
Color: silver (blue for amplifier case cover)
• External dimensions:
See OUTLINE DIAGRAM.
4.5 kg
• Orientation of transmission unit:
Indicator unit turnable 90° upward/down-
ward relative to detection unit.
• Mass:
2
Optional specifications
• Indication unit: 5-digit LCD indication, % or real scale in-
dication (as specified by code symbol)
Operating temperature range: –20 to
+70°C
• NACE specification:
H2S-proof treatment in
accordance with NACE
specification
Table 1 Remote Control Function
(Items readable and setting from hand-held
communicator)
Item
Reading Setting
Description
Maximum range
Maximum measuring range
of equipment
Measuring range
Damping
Real scale
Actual measuring range
Variable within 0.2 to 32 sec.
Indication in industrial value
indication
Battery voltage
Error indication
Battery voltage of equipment
Errors of detection unit and
amplifier
Measured value
Adjustment
Measured data
Zero and span adjustment
Note: For operation of the “3” type transmitter (“3” at the 8th digit
of product code), a hand-held communicator is required to
have a version 1.6 or higher, but a communicator before
version 1.6 can be operated with memory data updated.
(Refer to the instruction manual of transmitter.)
3
FBK···3
SYSTEM BLOCK DIAGRAM
Host system
Master station
Control room
Field
Star coupler
Transmitter
Field equipment
(max. 8 units)
Transmission distance: 1.5 km max.
(between master station and field equipment)
1:1 system
1 to 5V DC
4 to 20mA DC
Optical fiber type
electric converter
Control room
(Type; PSN)
Field
Transmitter
Transmission distance: 4 km max.
4
CODE SYMBOLS
1
2
3
4
5
6
7
8
9 10111213 1415
F B K
0
3
F
Description
Process cover (4th digit)
Process connection Cable lead-in port
Rc1/4
G1/2
S
T
1/4-18NPT
1/2-14NPT
Measuring span (6th digit)
1.6...........16 kPa abs
3.25.......130 kPa abs
12.5.......500 kPa abs
75........3000 kPa abs
1
2
3
4
Material (7th digit)
Process cover Seal diaphragm Other wetted parts
SCS14
SCS14
SCS14
SCS14
SUS316L
Hastelloy-C
Monel
SUS316
Hastelloy-C
Monel
Tantalum Note) Note)
V
H
M
T
Tantalum
Not available
when 6th digit is “1”.
Indicator (9th digit)
Not provided
Digital, % indication
Digital, real scale indication
A
L
P
Explosion-proof (10th digit)
Non-explosion proof
Intrinsic safety, JIS
A
G
Side vent/drain and fixture (11th digit)
Side vent/drain
Fixture
None
Yes (stainless steel)
None
Yes (stainless steel)
A
C
D
F
None
None
Yes
Yes
Treatment and Fill-fluid (13th digit)
Treatment
None
Degreasing
Fill-fluid
Silicon oil
Silicon oil
Y
G
N
NACE specification Silicon oil Note)
Note)
Unavailable when 7th digit is “T”.
O-ring material (for process cover)(14th digit)
Viton
A
Bolt/nut (15th digit)
NACE bolt/nut (ASTM A193 B7M/A194 2HM)
NACE bolt/nut (ASTM A320 L7M/A194 2HM)
SUS304/SUS304.... (For general use)
C
D
E
5
FBK···3
OUTLINE DIAGRAM (Unit : mm)
P
241
V
H
M
T
Mounting pipe
JIS 50A (2B)
80
S
T
A
C
FBK
0
3–
108
70
U-bolt
(M8)
315
54
15
Indicator (option)
B
Gas vent, drain port
37 27
47
Optical cable lead-in port
φ7
φ15
A
Pressure connection
D
E
φ18.5
Details of “B” Details of “A”
4–7/16-20UNF
Thread depth 13
(for oval flange)
27
View from arrow P
Type
FBKS
FBKT
D
E
Rc1/4
G1/2
1/4–18NPT
1/2–14NPT
6
SCOPE OF DELIVERY
Instrument body and pipe fixture (as specified)
ITEM TO BE PREPARED SEPARATELY
Oval flange:
To be used as a flange of connecting pipe
port.
For details, refer to the DATA SHEET of
oval flange (EDS6-10).
ORDERING INFORMATION
1. Model type
2. Measuring range
3. Indication scale for real scale specification
4. Others
7
FBK···3
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
Head Office
6-17, Sanbancho, Chiyoda-ku, Tokyo 102-0075, Japan
http://www.fesys.co.jp/eng
Sales Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http: //www.fic-net.jp/eng
Printed in Japan
Information in this catalog is subject to change without notice.