HIH-4010-003 [HONEYWELL]

Humidity Sensors;
HIH-4010-003
型号: HIH-4010-003
厂家: Honeywell    Honeywell
描述:

Humidity Sensors

文件: 总8页 (文件大小:441K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
HIH-4010/4020/4021  
Series  
Humidity Sensors  
DESCRIPTION  
The HIH-4010 is an uncovered integrated humidity sensor, the  
HIH-4020 is a covered integrated circuit humidity sensor, and  
the HIH-4021 is a covered, condensation-resistant, integrated  
circuit humidity sensor. All three products are available in two  
lead space configurations.  
The HIH-40104020/4021 Series Humidity Sensors are  
designed specifically for high volume OEM (Original Equipment  
Manufacturer) users.  
Direct input to a controller or other device is made possible by  
this sensor’s near linear voltage output. With a typical current  
draw of only 200 µA, the HIH-4010/4020/4021 Series is often  
ideally suited for low drain, battery operated systems.  
The RH sensor is a laser trimmed, thermoset polymer  
capacitive sensing element with on-chip integrated signal  
conditioning.  
Tight sensor interchangeability reduces or eliminates OEM  
production calibration costs. Individual sensor calibration data  
is available.  
The sensing element's multilayer construction provides  
excellent resistance to most application hazards such as  
wetting, dust, dirt, oils and common environmental chemicals.  
Both products are available in two lead spacing configurations,  
as well as with or without calibration and data printouts.  
The HIH-4010/4020/4021 Series delivers instrumentation-  
quality RH (Relative Humidity) sensing performance in a  
competitively priced, solderable SIP (Single In-line Package).  
FEATURES  
POTENTIAL APPLICATIONS  
Molded thermoset plastic housing  
Near linear voltage output vs %RH  
Laser trimmed interchangeability  
Low power design  
Refrigeration equipment  
HVAC (Heating, Ventilation and Air Conditioning)  
equipment  
Medical equipment  
Drying  
Enhanced accuracy  
Fast response time  
Metrology  
Stable, low drift performance  
Chemically resistant  
Battery-powered systems  
OEM assemblies  
HIH-4010/4020/4021 Series  
TABLE 1. PERFORMANCE SPECIFICATIONS (At 5 Vdc supply and 25 ºC [77 ºF] unless otherwise noted.)  
Parameter  
Minimum  
Typical  
Maximum  
Unit  
Specific  
Note  
Interchangeability (first order curve)  
0% RH to 59% RH  
60% RH to 100% RH  
Accuracy (best fit straight line)  
Hysterisis  
Repeatability  
Settling time  
Response time (1/e in slow moving air)  
Stability (at 50 %RH in 1 year)  
Stability (at 50 %RH in 1 year)  
Voltage supply  
-5  
-8  
-3.5  
5
8
+3.5  
% RH  
% RH  
% RH  
% RH  
% RH  
ms  
1
3
0.5  
70  
5
s
1.2  
0.5  
% RH  
% RH  
Vdc  
2
3
4
4
5.8  
500  
Current supply  
200  
µA  
Voltage output (1st order curve fit)  
Temperature compensation  
Output voltage temp. coefficient at 50% RH, 5 V  
Operating temperature  
Operating humidity (HIH-4010)  
Operating humidity (HIH-4020)  
Operating humidity (HIH-4021)  
Storage temperature  
VOUT=(VSUPPLY)(0.0062(sensor RH) + 0.16), typical at 25 ºC  
True RH = (Sensor RH)/(1.0546 – 0.00216T), T in ºC  
-4  
85[185]  
100  
100  
100  
mV/ºC  
ºC[ºF]  
% RH  
% RH  
% RH  
ºC[ºF]  
% RH  
5
5
5
-40[-40]  
See Figure 1.  
See Figure 1.  
See Figure 1.  
See Figure 1.  
0
0
0
-50[-58]  
125[257]  
Storage humidity  
Specific Notes:  
See Figure 2.  
General Notes:  
1. For HIH-4010/20/21-003/004 catalog listings only.  
Sensor is ratiometric to supply voltage.  
2. Includes testing outside of recommended operating zone.  
3. Includes testing for recommended operating zone only.  
4. Device is calibrated at 5 Vdc and 25 ºC.  
5. Non-condensing environment. When liquid water falls on the  
humidity sensor die, output goes to a low rail condition  
indicating no humidity.  
Extended exposure to >90% RH causes a reversible  
shift of 3% RH.  
Sensor is light sensitive. For best performance, shield  
sensor from bright light.  
FACTORY CALIBRATION DATA  
For HIH-4010-001/002/003/004 catalog listings only.  
HIH-4010/4020/4021 Sensors may be ordered with a  
calibration and data printout. See Table 2 and the order guide  
on the back page.  
TABLE 2. EXAMPLE DATA PRINTOUT  
Model  
Channel  
HIH-4010-003  
92  
Wafer  
MRP  
030996M  
337313  
For HIH-4020-001/002/003/004 and HIH-4021-001/002/003/  
004 catalog listings only.  
Calculated values at 5 V  
V
at 0% RH  
0.958 V  
3.268 V  
VOOUUTT at 75.3% RH  
Linear output for 3.5% RH  
accuracy at 25 °C  
Zero offset  
Slope  
Sensor RH  
0.958 V  
30.680 mV/%RH  
(VOUT - zero offset)/slope  
(VOUT - 0.958)/0.0307  
Ratiometric response for  
0% RH to 100% RH  
VOUT  
VSUPPLY (0.1915 to 0.8130)  
2
www.honeywell.com/sensing  
Humidity Sensors  
FIGURE 1. OPERATING ENVIRONMENT (Non-condensing environment for HIH-4010 and HIH-4020 catalog listings only.)  
FIGURE 2. STORAGE ENVIRONMENT (Non-condensing environment for HIH-4010 and HIH-4020 catalog listings only.)  
Honeywell Sensing and Control  
3
HIH-4010/4020/4021 Series  
FIGURE 3. TYPICAL OUTPUT VOLTAGE VS RELATIVE HUMIDITY (At 25 ºC and 5 V.)  
4
3.5  
3
2.5  
2
Sensor Response  
Sensor Response  
Best Linear Fit  
1.5  
1
0.5  
0
20  
40  
60  
80  
100  
Relative Humidity (%RH)  
FIGURE 4. TYPICAL OUTPUT VOLTAGE (BFSL) VS RELATIVE HUMIDITY (At 0 ºC, 70 ºC and 5 V.)  
4.5  
4
3.5  
3
2.5  
2
0 C  
70 C  
1.5  
1
0.5  
0
0
10  
20  
30  
40  
50  
60  
70  
80  
90  
100  
Relative Humidity (%)  
4
www.honeywell.com/sensing  
Humidity Sensors  
FIGURE 5. HIH-4010 MOUNTING DIMENSIONS (For reference only. mm/[in])  
FIGURE 6. HIH-4020 MOUNTING DIMENSIONS (For reference only. mm/[in])  
Honeywell Sensing and Control  
5
HIH-4010/4020/4021 Series  
FIGURE 7. HIH-4021 MOUNTING DIMENSIONS (For reference only. mm/[in])  
6
www.honeywell.com/sensing  
Humidity Sensors  
FIGURE 8. TYPICAL APPLICATION CIRCUIT  
ORDER GUIDE  
Catalog  
Description  
Listing  
HIH-4010-001  
HIH-4010-002  
HIH-4010-003  
HIH-4010-004  
HIH-4020-001  
HIH-4020-002  
HIH-4020-003  
HIH-4020-004  
HIH-4021-001  
HIH-4021-002  
HIH-4021-003  
Integrated circuit humidity sensor, 2,45 mm [0.100 in] lead pitch SIP  
Integrated circuit humidity sensor, 1.27 mm [0.050 in] lead pitch SIP  
Integrated circuit humidity sensor, 2,45 mm [0.100 in] lead pitch SIP, calibration and data printout  
Integrated circuit humidity sensor, 1.27 mm [0.050 in] lead pitch SIP, calibration and data printout  
Covered integrated circuit humidity sensor, 2,45 mm [0.100 in] lead pitch SIP  
Covered integrated circuit humidity sensor, 1.27 mm [0.050 in] lead pitch SIP  
Covered integrated circuit humidity sensor, 2,45 mm [0.100 in] lead pitch SIP, calibration and data printout  
Covered integrated circuit humidity sensor, 1.27 mm [0.050 in] lead pitch SIP, calibration and data printout  
Covered, filtered integrated circuit humidity sensor, 2,45 mm [0.100 in]lead pitch SIP  
Covered, filtered integrated circuit humidity sensor, 1.27 mm [0.050 in] lead pitch SIP  
Covered, filtered integrated circuit humidity sensor, 2,45 mm [0.100 in] lead pitch SIP, calibration and data  
printout  
HIH-4021-004  
Covered, filtered integrated circuit humidity sensor, 1.27 mm [0.050 in] lead pitch SIP, calibration and data  
printout  
FURTHER HUMIDITY SENSOR INFORMATION  
See the following associated literature at www.honeywell.com/sensing:  
Product installation instructions  
Application sheets:  
Humidity Sensor Performance Characteristics  
Humidity Sensor Theory and Behavior  
Humidity Sensor Moisture and Psychrometrics  
Thermoset Polymer-based Capacitive Sensors  
Honeywell Sensing and Control  
7
WARNING  
MISUSE OF DOCUMENTATION  
WARNING  
PERSONAL INJURY  
The information presented in this product sheet is for  
reference only. Do not use this document as a product  
installation guide.  
DO NOT USE these products as safety or emergency stop  
devices or in any other application where failure of the  
product could result in personal injury.  
Complete installation, operation, and maintenance  
information is provided in the instructions supplied with  
each product.  
Failure to comply with these instructions could result  
in death or serious injury.  
Failure to comply with these instructions could result  
in death or serious injury.  
SALES AND SERVICE  
Honeywell serves its customers through a worldwide network  
of sales offices, representatives and distributors. For  
application assistance, current specifications, pricing or name  
of the nearest Authorized Distributor, contact your local sales  
office or:  
WARRANTY/REMEDY  
Honeywell warrants goods of its manufacture as being free of  
defective materials and faulty workmanship. Honeywell’s  
standard product warranty applies unless agreed to otherwise  
by Honeywell in writing; please refer to your order  
acknowledgement or consult your local sales office for specific  
warranty details. If warranted goods are returned to Honeywell  
during the period of coverage, Honeywell will repair or replace,  
at its option, without charge those items it finds defective. The  
foregoing is buyer’s sole remedy and is in lieu of all other  
warranties, expressed or implied, including those of  
merchantability and fitness for a particular purpose. In no  
event shall Honeywell be liable for consequential, special,  
or indirect damages.  
E-mail: info.sc@honeywell.com  
Internet: www.honeywell.com/sensing  
Phone and Fax:  
Asia Pacific  
+65 6355-2828  
+65 6445-3033 Fax  
+44 (0) 1698 481481  
+44 (0) 1698 481676 Fax  
+1-305-805-8188  
+1-305-883-8257 Fax  
+1-800-537-6945  
Europe  
Latin America  
USA/Canada  
While we provide application assistance personally, through  
our literature and the Honeywell web site, it is up to the  
customer to determine the suitability of the product in the  
application.  
+1-815-235-6847  
+1-815-235-6545 Fax  
Specifications may change without notice. The information we  
supply is believed to be accurate and reliable as of this  
printing. However, we assume no responsibility for its use.  
Automation and Control Solutions  
Sensing and Control  
Honeywell  
1985 Douglas Drive North  
009020-1-EN IL50 GLO Printed in USA  
January 2007  
© 2007 Honeywell International Inc. All rights reserved.  
Minneapolis, MN 55422  
www.honeywell.com/sensing  

相关型号:

HIH-4010-004

Interactive Catalog Replaces Catalog Pages
HONEYWELL

HIH-4020

Humidity Sensors
HONEYWELL

HIH-4020-002

THERMOPLASTIC RH PACKAGE HIH-4020 SERIES
HONEYWELL

HIH-4021

Humidity Sensors
HONEYWELL

HIH-4030

Humidity Sensors
HONEYWELL

HIH-4030-001

HIH-4030/31 Series Humidity Sensors
HONEYWELL

HIH-4030-001S

HIH-4030/31 Series Humidity Sensors
HONEYWELL

HIH-4030-003

HIH-4030/31 Series Humidity Sensors
HONEYWELL

HIH-4030-003S

HIH-4030/31 Series Humidity Sensors
HONEYWELL

HIH-4031

Humidity Sensors
HONEYWELL

HIH-4031-001

HIH-4030/31 Series Humidity Sensors
HONEYWELL

HIH-4031-001S

HIH-4030/31 Series Humidity Sensors
HONEYWELL