Agilent U1270 Series Handheld  
Digital Multimeters  
Shaped to fit, tailored to perform, built to last  
Data Sheet  
Features  
• Intelligent features:  
ZLOW*, Smart Ω*, Low Pass Filter,  
Qik-V**  
• Visual (Backlight Alert) and  
audible continuity indication in  
noisy environments  
• Ergonomic shape for better grip  
• IP 54 certified – water and dust  
resistant  
• Easy-to-operate large knob and  
buttons  
Continuity detection in noisy and dark places  
Get a better grip on your  
• 30,000-count dual display  
• AC + DC capability*  
is made easy with the U1270 series’ loud  
beeper and Backlight Alert function which  
DMM  
flashes the backlight to indicate continuity  
• CAT III 1000 V, CAT IV 600 V  
safety rating  
and improve safety.  
The U1270 series is shaped perfectly to fit in  
your hand, with or without gloves on. Its non-  
slip ergonomic shape enables you to carry  
your DMM and perform measurements on the  
go easily. Additionally, the controls are easy to  
operate due to the large knob and buttons.  
When it comes to Variable Frequency Drive  
(VFD) troubleshooting, the U1270 series has  
Low Pass Filter to handle the job with ease.  
• Easy maintenance – convenient  
fuse access  
• Easy connectivity to PC with  
optional IR-USB cable  
Water and dust resistant  
The series’ tightly sealed design helps protect  
against water, dust and damage. Each DMM  
is IP 54 certified so that you can carry out test  
and measurement with confidence, even in  
harsh working conditions.  
• Internal memory for data logging  
Improve productivity with  
intelligent features  
Designed for maximum efficiency and  
productivity in industrial settings, these  
DMMs offer convenient functions such as  
* U1272A only  
** U1271A only  
Z
LOW to eliminate stray voltages, Smart Ω  
to minimize false readings due to leakage  
current and Qik-V to determine existence  
of AC and/or DC voltages.  
Key functions  
Low Impedance (ZLOW  
)
The U1272A is a dual input impedance digital multimeter. The DMM’s high input impedance is preferred in most electrical  
measurements because it would not load the circuit under test. However, to obtain accurate measurements on circuits  
that may contain stray voltages, the U1272A’s 2 kΩ low impedance function comes in handy. Stray voltages are usually  
found in non-energized electrical wiring adjacent to powered wires due to capacitive or inductive coupling between these  
wires. When a pair of test leads is placed between the open circuit and neutral conductor, the circuit is then complete  
and forms a voltage divider in conjunction with the input impedance of the multimeter. High input impedance multimeter  
is sensitive enough to measure voltage coupled into the disconnected conductor, thus giving an inaccurate indication of  
a live conductor. The low impedance function serves to eliminate false readings by dissipating the stray voltages, thus  
improves safety and measurement efficiency during voltage.  
Figure 1: U1272A helps you identify the presence of stray voltage on a disconnected wire running parallel with the wire  
powering up the VFD to an industrial motor. The image on the right shows the U1272A in low impedance mode.  
Low Pass Filter (LPF)  
The U1270 series offers a 1 kHz LPF or Low Pass Filter to provide accurate Variable Frequency Drive (VFD) output  
measurement. This function eliminates high frequency noise and harmonics. This ensures the efficiency of your motor  
filter as well.  
Figure 2: Comparison of voltage output from industrial motor VFD without and with Low Pass Filter functionality.  
Smart Ω  
The U1272A provides an additional 30 Ohm range for low resistance measurement. This Smart Ω function is available  
for ranges of 30 Ohm to 300 kOhm. It enhances measurement accuracy with offset compensation by removing residual  
DC voltages of up to 1000 mV induced by ground current and thermal EMF. This function also enables ‘live’ resistance  
measurement without isolating the measurement circuit. With this, you will be able to obtain leakage current using the  
secondary display.  
2
Front and back panel description  
Dual display  
Backlight Alert flashes the  
backlight to provide visual  
indication in continuity test  
30,000 counts resolution  
Large buttons  
Peak detect records  
transients as fast as 250µs  
Smart Ω** removes residual  
voltage of up to 1000 mV  
Low Pass Filter (LPF)  
removes unwanted high  
frequency signals  
Auto diode** automatically  
determines diode polarity  
Temperature measurement:  
J type thermocouple  
-200 to 1200 °C**, K type  
thermocouple -200 to 1732 °C  
Z
LOW** provides both high  
and low impedance modes  
to eliminate stray voltages  
AC + DC** capability  
Large knob for ease of use  
Qik-V* allows AC+DC  
measurement to quickly  
check for energized voltage  
* U1271A only  
** U1272A only  
Front panel  
3
Hang hole  
Probe holder/storage  
IR-USB connectivity with  
optional IR-USB cable  
Slim, ergonomic design for  
better grip  
Easy fuse and battery  
access from battery cover  
beneath the stand  
Back panel  
4
Choose between these two models  
Di  
U1271A  
U1272A  
Basic Features  
Display resolution  
Auto/manual ranging  
Analog bar graph  
Backlight  
30,000  
Yes  
Yes  
30,000  
Yes  
Yes  
Yes  
Yes  
AC bandwidth  
True RMS  
20 kHz  
AC  
100 kHz  
AC + DC  
Measurements  
Range  
300 mV to 1000 V  
0.05% + 2 cnts  
300 mV to 1000 V  
0.7% + 20 cnts  
45 Hz to 20 kHz  
300 µA to 10 A  
0.2% + 5 cnts  
300 µA to 10 A  
0.9% + 25 cnts  
45 Hz to 2 kHz  
30 mV to 1000 V  
0.05% + 2 cnts  
30 mV to 1000 V  
0.6% + 20 cnts  
45 Hz to 100 kHz  
300 µA to 10 A  
0.2% + 5 cnts  
300 µA to 10 A  
0.6% + 25 cnts  
45 Hz to 2 kHz  
Voltage DC  
Voltage AC  
Current DC  
Current AC  
Accuracy  
Range  
Accuracy  
Bandwidth  
Range  
Accuracy  
Range  
Accuracy  
Bandwidth  
Range  
300 Ω to 100 MΩ  
0.2 % + 5 cnts  
99.999 Hz to 999.99 kHz  
0.005% + 5 cnts  
30 Ω to 300 MΩ  
0.2% + 5 cnts  
99.999 Hz to 999.99 kHz  
0.005% + 5 cnts  
Resistance  
Frequency  
Capacitance  
Accuracy  
Range  
Accuracy  
Range  
Accuracy  
10 nF to 10 mF  
1% + 2 cnts  
10 nF to 10 mF  
1% + 2 cnts  
Range  
Accuracy  
K: -200 to 1372 °C  
1% + 1°C  
K: -200 to 1372 °C  
J: -200 to 1200 °C  
1% + 1 °C  
Temperature  
Continuity with beeper  
Diode test  
Yes  
Yes  
Yes  
Yes  
Data Management  
Min/Max Recording  
Display Hold  
Yes  
Yes  
Yes  
Yes  
Peak Hold  
Yes  
Yes  
Manual Datalogging  
Null  
Yes  
Yes  
Yes  
Yes  
PC Connectivity  
% scale of 4-20 mA  
IR-USB  
Yes  
IR-USB  
Yes  
5
Di  
U1271A  
U1272A  
Special Features  
Beep + Backlight Alert  
Low Pass Filter (LPF)  
ZLOW - Low impedance mode  
Smart Ω  
Yes  
Yes  
-
Yes  
Yes  
Yes  
Yes  
-
-
Qik-V  
Yes  
Safety and Regulatory  
Over-voltage safety protection  
EN/IEC 61010-1:2001 compliance  
General  
CAT III 1000 V, CAT IV 600 V  
Yes  
CAT III 1000 V, CAT IV 600 V  
Yes  
Battery  
4x AAA  
4x AAA  
Operating temperature  
-20 °C to 55 °C, 0 to 80% R.H  
-20 °C to 55 °C, 0 to 80% R.H  
Standard test leads, test probes with 19-  
mm and 4-mm tips, K-type thermocouple  
Standard test leads, test probes with 19-  
mm and 4-mm tips, K-type thermocouple  
Standard Accessories  
and adapter, 4x AAA batteries, Certificate and adapter, 4x AAA batteries, Certificate  
of Calibration, test report, Quick Start  
Guide  
of Calibration, test report, Quick Start  
Guide  
6
General Specifications  
Liquid crystal display (LCD) (with maximum reading of 33000 counts)  
460 mVA maximum (with backlight enabled)  
Display  
Power consumption  
Battery Type  
4 × 1.5 V Alkaline battery (ANSI/NEDA 24A or IEC LR03), or  
4 × 1.5 V Zinc Chloride battery (ANSI/NEDA 24D or IEC R03)  
300 hours typical (based on new Alkaline batteries for dc voltage measurement)  
Low battery indicator will flash when the battery voltage drops below 4.4 V (approximately)  
10 × 35 mm 440 mA/1000 V 30 kA fast-acting fuse  
10 × 38 mm 11 A/1000 V 30 kA fast-acting fuse  
Operating temperature from –20 to 55 °C, 0 to 80% RH  
Full accuracy up to 80% RH for temperatures up to 30 °C, decreasing linearly to 50%  
RH at 55 °C  
Battery Life  
Fuse  
Operating Environment  
Altitude up to 2000 meters  
Pollution degree II  
–40 to 70 °C, 0 to 80% RH  
Storage Compliance  
Safety Compliance  
CAN/CSA-C22.2 No. 61010-1-04  
EN/IEC 61010-1:2001  
ANSI/UL 61010-1:2004  
CAT III 1000 V/ CAT IV 600 V  
Measurement Category  
Commercial limits compliance with EN61326-1  
Electromagnetic Compatibility  
(EMC)  
IP-54  
Ingress Protection Rating  
Temperature Coefficient  
0.05 × (specified accuracy) / °C (from –20 to 18 °C, or 28 to 55 °C)  
>120 dB at DC, 50/60 Hz 0.1% (1 kΩ unbalanced)  
Common Mode Rejection Ratio  
(CMRR)  
>60 dB at 50/60 Hz 0.1%  
Normal Mode Rejection Ration  
(NMRR)  
Dimensions (W x H x D)  
Weight  
92 × 207 × 59 mm  
U1271A: 518 grams (with batteries)  
U1272A: 520 grams (with batteries)  
Three years for product  
Warranty  
Three months for product’s accessories  
One year  
Calibration Cycle  
Specification Assumptions  
Accuracy is given as (% of reading + counts of least significant digit) at 23 °C 5 °C, with relative humidity less than 80% RH.  
AC V and AC μA/mA/A specifications are ac coupled, true RMS and are valid from 5% of range to 100% of range.  
The crest factor may be up to 3.0 at full- scale except for the 1000 V range where it is 1.5 at full scale.  
For non- sinusoidal waveforms, add (2% reading + 2% full scale) typical, for crest factors up to 3.  
After ZLOW voltage measurements, wait at least 20 minutes for thermal impact to cool before proceeding with any other measurement.  
7
Electrical specifications  
DC specifications for U1271A and U1272A  
Accuracy (% of reading +  
Test current/  
counts of least significant digit)  
Function  
Range  
Resolution  
Burden voltage  
U1271A  
U1272A  
0.05 + 20  
0.05 + 5  
0.05 + 5  
0.05 + 2  
0.05 + 2  
0.05 + 2  
30 mV  
300 mV  
3 V  
30 V  
300 V  
1000 V  
0.001 mV  
0.01 mV  
0.0001 V  
0.001 V  
0.01 V  
-
0.05 + 5  
0.05 + 5  
0.05 + 2  
0.05 + 2  
0.05 + 2  
-
-
-
-
-
Voltage [1]  
0.1 V  
ZLOW (low input impedance)  
enabled, applicable for 1000 V  
range and resolution only.  
-
1 + 20  
30 Ω  
300 Ω  
3 kΩ  
30 kΩ  
300 kΩ  
3 MΩ  
0.001 Ω  
0.01 Ω  
-
0.2 + 10  
0.2 + 5  
0.2 + 5  
0.2 + 5  
0.5 + 5  
0.6 + 5  
1.2 + 5  
-
0.65 mA  
0.65 mA  
65 mA  
6.5 mA  
0.65 mA  
0.2 + 5  
0.2 + 5  
0.2 + 5  
0.5 + 5  
0.6 + 5  
1.2 + 5  
2.0 +10  
0.0001 kΩ  
0.001 kΩ  
0.01 kΩ  
0.0001 MΩ  
0.001 MΩ  
-
Resistance [2]  
93 nA//10 MΩ  
93 nA//10 MΩ  
93 nA//10 MΩ  
30 MΩ  
100 MΩ  
2.0 % + 10 @ <100 MΩ  
8.0 % + 10 @ >100 MΩ  
300 MΩ  
0.01 MΩ  
-
93 nA//10 MΩ  
300 nS  
300 µA  
3000 µA  
30 mA  
300 mA  
3 A  
10 A  
3 V  
Auto  
0.01 nS  
0.01 µA  
0.1 µA  
0.001 mA  
0.01 mA  
0.0001 A  
0.001 A  
0.0001 V  
0.0001 V  
1 + 10  
0.2 + 5  
0.2 + 5  
0.2 + 5  
0.2 + 5  
0.3 +10  
0.3 +10  
0.5 + 5  
-
1 + 10  
0.2 + 3  
0.2 + 3  
0.2 + 3  
0.2 + 3  
0.3 +10  
0.3 +10  
0.5 + 5  
0.5 + 5  
93 nA//10 MΩ  
<0.04 V/ 100 Ω  
<0.4 V/ 100 Ω  
<0.08 V/ 1 Ω  
<1.00 V/ 1 Ω  
Current [3]  
<0.1 V/ 0.01 Ω  
<0.3 V/ 0.01 Ω  
Approximately 1 to 2 mA  
Approximately 0.1 to 0.3 mA  
Diode Test [4]  
1. Notes for voltage specifications:  
The accuracy of the 30 to 300 mV range is specified after the Null function is used to subtract the thermal effect (by shorting the test leads).  
For ZLOW measurements, autoranging is disabled and the multimeter’s range is set to 1000 volts in the manual ranging mode.  
2. Notes for resistance specifications:  
Overload protection: 1000 Vrms for short circuits with <0.3 A current.  
Maximum open voltage is <+3.3 V  
Built-in buzzer beeps when the resistance measured is less than 25 Ω 10 Ω. The multimeter can capture intermittent measurements longer  
than 1 ms.  
The accuracy of the 300 Ω to 3 kΩ range is specified after the Null function is used to subtract the test lead resistance and thermal effect (by  
shorting the test leads).  
For the ranges of 30 MΩ and 100 MΩ, the RH is specified for <60%.  
The accuracy for ranges <50 nS is specified after the Null function is used on an open test lead.  
The temperature coefficient of the 100 MΩ and 300 MΩ range is 0.1 × (specified accuracy)/°C (from -20 °C to 18 °C or 28 °C to 55 °C)  
8
3. Notes for current specifications:  
Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse  
Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse  
Specification for 300 mA range: 440 mA continuous.  
Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 to 20 A for 30 seconds  
maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the mea-sured time before proceeding with low  
current measurements.  
4. Notes for diode specifications:  
Overload protection: 1000 Vrms for short circuits with <0.3 A current.  
Built-in buzzer beeps continuously when the voltage measured is less than 50 mV and beeps once for forward-biased di-ode or semiconductor  
junctions measured between 0.3 V and 0.8 V (0.3 V ≤ reading ≤ 0.8 V).  
Open voltage for diode: <+3.3 V DC  
Open voltage for Auto diode: <+2.5 V DC and > -1.0 V DC  
AC specifications for U1271A  
Accuracy  
Function  
Range Resolution  
45 Hz to  
65 Hz  
1 kHz to 5 kHz to  
30 Hz to 1 kHz  
5 kHz  
2.0 + 25  
2.0 + 25  
2.0 + 25  
2.0 + 25  
-
20 kHz  
2.0 + 40  
2.0 + 40  
2.0 + 40  
-
300 mV  
3 V  
30 V  
300 V  
1000 V  
0.01 mV  
0.0001 V  
0.001 V  
0.01 V  
0.7 + 20  
0.7 + 20  
0.7 + 20  
0.7 + 20  
0.7 + 20  
0.7 + 20  
1.0 + 25  
1.0 + 25  
1.0 + 25  
1.0 + 25  
1.0 + 25  
True RMS AC  
Voltage [1]  
0.1 V  
-
LPF (low pass filter)  
enabled, applicable for  
all voltage ranges and  
resolution  
1.0 + 25 @ <200 Hz  
5.0 + 25 @ <440 Hz  
-
-
Accuracy  
45 Hz to 2 kHz  
Function  
Range  
Resolution  
Burden voltage/Shunt  
300 µA  
3000 µA  
30 mA  
300 mA  
3A  
0.01 µA  
0.1 µA  
0.001 mA  
0.01 mA  
0.0001 A  
0.001 A  
0.9 + 25  
0.9 + 25  
0.9 + 25  
0.9 + 25  
1.0 + 25  
1.0 + 25  
<0.04 V/ 100 Ω  
<0.4 V/ 100 Ω  
<0.08 V/ 1 Ω  
<1.00 V/ 1 Ω  
<0.1 V/ 0.01 Ω  
<0.3 V/ 0.01 Ω  
True RMS AC  
Current [2]  
10 A  
1. Notes for voltage specifications:  
Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with <0.3 A current.  
Input impedance: 10 MΩ (nominal) in parallel with <100 pF.  
2. Notes for current specifications:  
Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse  
Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse  
Specification for 300 mA range: 440 mA continuous.  
Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 to 20 A for 30 seconds  
maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the measured time before proceeding with low  
current measurements.  
9
AC specifications for U1272A  
Accuracy  
20 kHz  
to 100  
kHz  
Function  
Range Resolution  
45 to 20 Hz to 1 1 kHz to 5 kHz to  
65 Hz  
kHz  
5 kHz  
20 kHz  
30 mV  
300 mV  
3 V  
30 V  
300 V  
1000 V  
0.001 mV  
0.01 mV  
0.0001 V  
0.001 V  
0.01 V  
0.6 + 20  
0.6 + 20  
0.6 + 20  
0.6 + 20  
0.6 + 20  
0.6 + 20  
0.7 + 25  
0.7 + 25  
1.0 + 25  
1.0 + 25  
1.0 + 25  
1.0 + 25  
1.0 + 25  
1.0 + 25  
1.5 + 25  
1.5 + 25  
1.5 + 25  
1.5 + 25  
1.0 + 40 3.5 + 40  
1.0 + 40 3.5 + 40  
2.0 + 40 3.5 + 40  
2.0 + 40 3.5 + 40  
2.0 + 40  
-
-
-
True RMS AC  
Voltage [1]  
0.1 V  
LPF (low pass filter)  
enabled, applicable for  
all voltage ranges and  
resolution  
1.0 + 25 @  
<200 Hz  
0.6 + 20  
2.0 + 40  
-
-
-
-
-
-
5.0 + 25 @  
<440 Hz  
2 + 40 @  
<440 Hz  
ZLOW 1000 V  
Accuracy  
Burden  
20 Hz  
Function Range Resolution  
voltage/  
Shunt  
45 to  
to  
65 Hz  
2 kHz  
300 µA  
3000 µA  
30 mA  
300 mA  
3A  
0.01 µA  
0.1 µA  
0.001 mA  
0.01 mA  
0.0001 A  
0.001 A  
0.6 + 25 0.9 + 25 <0.04 V/ 100 Ω  
0.6 + 25 0.9 + 25 <0.4 V/ 100 Ω  
True RMS  
0.6 + 25 0.9 + 25  
0.6 + 25 0.9 + 25  
0.8 + 25 1.0 + 25 <0.1 V/ 0.01 Ω  
0.8 + 25 1.0 + 25 <0.3 V/ 0.01 Ω  
<0.08 V/ 1 Ω  
<1.00 V/ 1 Ω  
AC Current  
[2]  
10 A  
1. Notes for voltage specifications  
Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with <0.3 A current.  
Input impedance: 10 MΩ (nominal) in parallel with <100 pF.  
ZLOW impedance: 2 kΩ (nominal)  
The input signal is lower than the product of 20,000,000 V×Hz.  
For 20 to 100 kHz accuracy: Three counts of the LSD per kHz of additional error is to be added for frequencies >20 kHz and signal inputs <10%  
of range.  
2. Notes for current specifications  
Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse  
Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse  
Specification for 300 mA range: 440 mA continuous.  
Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 to 20 A for 30 seconds  
maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the measured time before proceeding with low  
current measurements.  
10  
AC + DC specifications for U1272A  
Accuracy  
20 kHz  
to 100  
kHz  
Function  
Range Resolution  
45 Hz to 20 Hz to 1 kHz to 5 kHz to  
65 Hz  
2 kHz  
5 kHz  
20 kHz  
30 mV  
300 mV  
3 V  
30 V  
300 V  
1000 V  
0.001 mV  
0.01 mV  
0.0001 V  
0.001 V  
0.01 V  
0.7 + 40  
0.7 + 25  
0.7 + 25  
0.7 + 25  
0.7 + 25  
0.7 + 25  
0.8 + 45  
0.8 + 30  
1.1 + 30  
1.1 + 30  
1.1 + 30  
1.1 + 30  
1.1 + 45  
1.1 + 30  
1.6 + 30  
1.6 + 30  
1.6 + 30  
1.6 + 30  
1.1 + 60  
1.1 + 45  
2.1 + 45  
2.1 + 45  
2.1 + 45  
-
3.6 + 60  
3.6 + 45  
3.6 + 45  
3.6 + 45  
-
True RMS  
AC + DC  
Voltage [1]  
0.1 V  
-
Accuracy  
Range Resolution  
45 Hz to 20 Hz to  
Burden  
voltage/  
Shunt  
<0.04 V/100 Ω  
<0.4 V/100 Ω  
<0.08 V/1 Ω  
<1.00 V/1 Ω  
<0.1 V/0.01 Ω  
<0.3 V/0.01 Ω  
Function  
65 Hz  
0.8 + 30  
0.8 + 30  
0.8 + 30  
0.8 + 30  
0.9 + 35  
0.9 + 35  
2 kHz  
1.1 + 30  
1.1 + 30  
1.1 + 30  
1.1 + 30  
1.3 + 35  
1.3 + 35  
300 µA  
3000 µA  
30 mA  
300 mA  
3A  
0.01 µA  
0.1 µA  
0.001 mA  
0.01 mA  
0.0001 A  
0.001 A  
True RMS  
AC + DC  
[2]  
Current  
10 A  
1. Notes for voltage specifications:  
Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with <0.3 A current.  
Input impedance: 10 MΩ (nominal) in parallel with <100 pF.  
For 20 to 100 kHz accuracy: Three counts of the LSD per kHz of additional error is to be added for frequencies >20 kHz and signal inputs <10%  
of range.  
2. Notes for current specifications:  
Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse  
Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse  
Specification for 300 mA range: 440 mA continuous.  
Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 to 20 A for 30 seconds  
maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the measured time before proceeding with low  
current measurements.  
11  
Temperature specifications [1] - [6]  
Thermocouple type  
Accuracy  
Range  
Resolution  
U1271A  
U1272A  
1% + 1 °C  
1% + 1.8 °F  
1% + 1 °C  
1% + 1.8 °F  
-200 to 1372 °C  
-328 to 2502 °F  
-200 to 1200 °C  
-328 to 2192 °F  
0.1 °C  
0.1 °F  
0.1 °C  
0.1 °F  
1% + 1 °C  
K
J
1% + 1.8 °F  
-
-
1. The specifications above is specified after 60 minutes of warm-up time.  
2. The accuracy does not include the tolerance of the thermocouple probe.  
3. Do not allow the temperature sensor to contact a surface that is energized above 30 Vrms or 60 V DC. Such voltages poses a shock hazard.  
4. Ensure that the ambient temperature is stable within 1 ꢀC and that the Null function is used to reduce the test lead’s thermal effect and temperature  
offset. Before using Null function, set the multimeter to measure temperature without ambient compensation (°C) and keep the thermocouple probe  
as close to the multimeter as possible (avoid contact with any surface that has a different temperature from the ambient temperature).  
5. When measuring temperature with respect to any temperature calibrator, try to set both the calibrator and multimeter with an external reference  
(without internal ambient compensation). If both the calibrator and multimeter are set with internal reference (with internal ambient compensation),  
some deviations may show between the readings of the calibrator and multimeter, due to differences in ambient compensation between the  
calibrator and multimeter. Keeping the multimeter close to the output terminal of calibrator will help reduce the deviation.  
6. The temperature calculation is specified according to the safety standards of EN/IEC-60548-1 and NIST175.  
Capacitance specifications [7][8]  
Accuracy  
Range  
Resolution  
U1271A  
1% + 5  
1% + 2  
1% + 2  
1% + 2  
1% + 2  
1% + 2  
1% + 2  
U1272A  
1% + 5  
1% + 2  
1% + 2  
1% + 2  
1% + 2  
1% + 2  
1% + 2  
10 nF  
100 nF  
1000 nF  
10 µF  
100 µF  
1000 µF  
10 mF  
0.001 nF  
0.01 nF  
0.1 nF  
0.001 µF  
0.01 µF  
0.1 µF  
0.001 mF  
7. Overload protection: 1000 Vrms for short circuits with <0.3 A current.  
8. The accuracy for all ranges is specified based on a film capacitor or better, and after the Null function is used to subtract the test  
lead resistance and thermal effect (by shorting the test leads).  
12  
Frequency specifications [1][2]  
Range  
99.999 Hz  
999.99 Hz  
9.9999 kHz  
99.999 kHz  
999.99 kHz  
>1 MHz  
Resolution  
0.001 Hz  
0.01 Hz  
0.1 Hz  
Accuracy  
0.02% + 5  
0.005% + 5  
0.005% + 5  
0.005% + 5  
Minimum input frequency  
0.5 Hz  
1 Hz  
0.01 kHz  
0.1 kHz  
0.005% + 5  
0.005% + 5 @ <1 MHz  
1. Overload protection: 1000 V; input signal is <20,000,000 V × Hz (product of voltage and frequency).  
2. The frequency measurement is susceptible to error when measuring low-voltage, low-frequency signals. Shielding  
inputs from external noise pickup is critical for minimizing measurement errors. Turning on the low pass filter may help  
you to filter out the noise and achieve a stable reading.  
Duty Cycle [3]  
Mode  
Range  
99.99%  
99.99%  
Accuracy at full scale  
0.3 % per kHz + 0.3 %  
DC Coupling  
AC Coupling  
0.3 % per kHz + 0.3 %  
Pulse Width [4]  
Range  
999.99 ms  
2000.0 ms  
Resolution  
0.01 ms  
Accuracy at full scale  
(duty cycle accuracy/frequency) + 0.01 ms  
(duty cycle accuracy/frequency) + 0.1 ms  
0.1 ms  
3. Notes for duty cycle specifications:  
The accuracy for duty cycle and pulse width measurements is based on a 3 V square wave input to the dc 3  
V range. For ac couplings, the duty cycle range can be measured within the range of 10% to 90% for signal  
frequencies >20 Hz.  
The range of the duty cycle is determined by the frequency of the signal: {10 μs × frequency × 100%} to {[1 – (10  
μs × frequency)] × 100%}.  
The pulse width (positive or negative) must be >10 μs. The range of the pulse width is determined by the  
frequency of the signal.  
4. Notes for pulse width specifications:  
The accuracy for duty cycle and pulse width measurements is based on a 3 V square wave input to the dc 3 V range.  
The pulse width (positive or negative) must be >10 μs. The range of the pulse width is determined by the  
frequency of the signal.  
13  
Frequency sensitivity for voltage measurements [1][2][3]  
Minimum sensitivity (RMS sine wave)  
15 Hz to 100 kHz 0.5 Hz to 200 kHz Up to 1 MHz  
Trigger level for dc coupling  
0.5 Hz to 200 kHz  
Input range  
U1271A  
U1272A  
30 mV  
300 mV  
3 V  
30 V  
300 V  
1000 V  
3 mV  
6 mV  
0.12 V  
0.6 V  
6 V  
3 mV  
8 mV  
0.2 V  
-
40 mV  
0.4 V  
2.6 V  
-
-
5 mV  
10 mV  
15 mV  
0.15 V  
0.15 V  
0.8 V  
1.5 V  
1.5 V  
8 V @ <100 kHz  
50 V @ <100 kHz  
9 V @ <100 kHz  
90 V @ <100 kHz  
9 V @ <100 kHz  
90 V @ <100 kHz  
50 V  
-
1. Maximum input for specified accuracy, refer to “AC specifications” on page 9.  
2. 30 mV range applicable for U1272A only.  
3. 200 kHz to 1 MHz range applicable for U1272A only.  
Frequency sensitivity for current measurements [4]  
Minimum sensitivity (RMS sine wave)  
Input range  
2 Hz to 30 kHz  
300 µA  
3000 µA  
30 mA  
300 mA  
3 A  
100 µA  
70 µA  
1.2 mA  
12 mA  
0.12 A  
10 A  
1.2 A  
4. Maximum input for specified accuracy, refer to “AC specifications” on page 9.  
Peak hold  
Signal width  
Single event >1 ms  
Repetitive >250 µs  
Accuracy for DC Voltage and Current  
Specified accuracy + 400  
Specified accuracy + 1000  
14  
Decibel (dB) for U1272A [1] - [3]  
dB base  
1 mW (dBm)  
1 V(dBV)  
Reference  
1 to 9999 Ω  
1 V  
Default reference  
50 Ω  
1 V  
1. The reading of dBm is indicated in decibels of power above or below 1 mW, or decibels of voltage above or below 1 V. The  
formula is calculated according to the voltage measurement and specified reference impedance. Its accuracy is  
depended on the accuracy of the voltage measurement. See Decibel (dBV) accuracy table below.  
2. Auto-ranging mode is used.  
3. The bandwidth is according to voltage measurement.  
Decibel (dBV) accuracy  
dBV range  
Minimum Maximum 45 to 65 Hz  
Accuracy  
Range  
20 Hz to 1 Hz to 5 kHz to 20 Hz to  
1 kHz  
0.07  
0.07  
0.09  
0.09  
5 kHz 20 kHz 100 kHz  
30 mV  
300 mV  
3 V  
30 V  
300 V  
1000 V  
-56.48  
-36.48  
-16.48  
+3.52  
-30.46  
-10.46  
+9.54  
+29.54  
+49.54  
+60  
0.06  
0.06  
0.06  
0.06  
0.06  
0.06  
0.09  
0.09  
0.14  
0.14  
0.14  
0.14  
0.1  
0.1  
0.19  
0.19  
0.19  
-
0.32  
0.32  
0.32  
0.32  
-
+23.52  
+33.98  
0.09  
0.09  
-
Measurement rate (approximate)  
Function  
Times/second  
U1271A  
U1272A  
ACV  
7
7
DCV  
7
14  
7
14  
Ω with offset compensation  
Diode  
-
14  
3
14  
Auto Diode  
Capacitance  
DCI  
-
3
4 (<100 µF)  
4 (<100 µF)  
7
7
ACI  
7
7
Temperature  
Frequency  
Duty cycle  
Pulse width  
7
7
2 (>10 Hz)  
1 (>10 Hz)  
1 (>10 Hz)  
2 (>10 Hz)  
1 (>10 Hz)  
1 (>10 Hz)  
15  
Ordering Information  
U1271A  
U1272A  
Standard Shipped Accessories  
Standard test leads, test probes with 19-mm and 4-mm tips, K-type thermocouple and adapter, 4x AAA batteries, Certificate of  
Calibration, test report, Quick Start Guide  
Optional Accessories  
Measuring Accessories (non-temperature)  
U1160A Standard test lead kit  
Includes two test leads (red and black), alligator clips,  
fine-tip test probes, SMT grabbers and mini grabber  
(black).  
U1161A Extended test lead kit  
Includes two test leads (red and black), two test  
probes, medium-sized alligator clips and 4-mm banana  
plugs.  
U1162A Alligator clips  
• One pair of insulated alligator clips (red and black).  
Recommended for use with Agilent standard test  
leads.  
• Rated CAT III 1000 V, 10 A.  
• Test leads: CAT III 1000 V, 15 A  
• Alligator clips: CAT III 1000 V, 10 A  
• Fine-tip test probes: CAT II 300 V, 3 A  
• SMT grabbers: CAT II 300 V, 3 A  
• Mini grabber: CAT II 300 V, 3 A  
• Test leads: CAT III 1000 V, CAT IV 600 V, 15 A  
• Test probes: CAT III 1000 V, 15 A  
• Medium-sized alligator clips: CAT III 600 V, 10 A  
• 4-mm banana plugs: CAT II 600 V, 10 A  
U1163A SMT grabbers  
U1164A Fine-tip test probes  
U1165A Test probe leads  
• Rated CAT III 1000 V, 15 A  
• One pair of SMT grabbers (red and black).  
Recommended for use with Agilent standard  
test leads.  
• One pair of fine-tip test probes (red and black).  
Recommended for use with Agilent standard  
test leads.  
• Rated CAT II 300 V, 3 A.  
• Rated CAT II 300 V, 3 A.  
16  
Measuring Accessories (non-temperature)  
U1169A Test probe leads  
U1583B AC current clamp  
U1168A Standard test lead kit  
Includes two test leads (red and black), and a pair each  
• Dual range: 40 A and 400 A  
• Rated CAT III 600 V  
• BNC-to-banana-plug adapter provided for use  
with DMMs  
Includes two test leads (red and black), 19-mm and  
4-mm test probes, alligator clips, fine-tip test probes,  
SMT grabbers and mini grabber (black).  
of 19-mm and 4-mm test probes.  
Test leads: CAT III 1000 V, CAT IV 600 V, 15 A  
Test probes (19-mm tip): CAT III 1000 V, CAT IV 600 V,  
15 A  
Test probes (4-mm tip): CAT III 1000 V, CAT IV 600 V,  
15 A (highly recommended for CAT IV environment)  
Test leads: CAT III 1000 V, CAT IV 600 V, 15 A  
Test probes (19-mm tip): CAT III 1000 V, CAT IV 600 V,  
15 A  
Test probes (4-mm tip): CAT III 1000 V, CAT IV 600 V,  
15 A (highly recommended for CAT IV environment)  
Alligator clips: CAT III 1000 V, 10 A  
Fine-tip test probes: CAT II 300 V, 3 A  
SMT grabber: CAT II 300 V, 3 A  
Mini grabber: CAT II 300 V, 3 A  
Measuring Accessories (temperature)  
U1180A Thermocouple adapter+lead kit, J and  
K types  
U1181A Immersion temperature probe  
Type-K T/C for use in oil and other liquids  
Measurement range: –50 °C to 700 ºC  
Includes adapter U1184A for connection to DMM  
U1182A Industrial surface temperature probe  
Type-K T/C for use on still surfaces  
Measurement range: –50 ºC to 400 ºC  
Includes adapter U1184A for connection to DMM  
Includes thermocouple adapter, thermocouple bead  
J-type and thermocouple bead K-type.  
• T/C adapter J/K-type  
• T/C bead J-type: –20 °C to 200 °C  
• T/C bead K-type: –20 °C to 200 °C  
U1183A Air temperature probe  
U1184A Temperature probe adapter  
U1185A J-type thermocouple and adapter  
Type-K T/C for use in air and non-caustic gas  
Measurement range: –50 ºC to 800 ºC  
Includes adapter U1184A for connection to DMM  
• Mini-connector-to-banana-plug adapter for use  
• T/C adapter J/K-type  
• T/C bead J-type: –20 °C to 200 °C  
with DMM  
U1186A K-type thermocouple and adapter  
• T/C adapter J/K-type  
• T/C bead J-type: –20 °C to 200 °C  
17  
Cable  
U1173A IR-to-USB cable  
U1174A Soft carrying case  
• For remote control and data logging to PC  
• Max. baud rate: 19,200 bits per second  
The convenient way to carry your DMM and  
essential accessories  
• Dimension: 9” (H) x 5” (W) x 3” (D)  
Hanging Kit  
U1171A Magnetic hanging kit  
For fastening of DMM to a steel surface so both hands  
are free.  
Probe Clip Light  
U1176A LED Probe Clip Light  
• 3 inches in length  
• To be clipped onto test probes to increase visibility  
• Comes with one AAA battery  
18  
www.agilent.com  
www.agilent.com/find/  
handhelddmm  
For more information on Agilent  
Agilent Email Updates  
Technologies’ products, applications or  
services, please contact your local Agilent  
office. The complete list is available at:  
www.agilent.com/find/emailupdates  
Get the latest information on the  
products and applications you select.  
www.agilent.com/find/contactus  
Agilent Advantage Services is committed  
to your success throughout your equip-  
ment’s lifetime. We share measurement  
and service expertise to help you create  
the products that change our world. To  
keep you competitive, we continually  
invest in tools and processes that speed  
up calibration and repair, reduce your cost  
of ownership, and move us ahead of your  
development curve.  
Americas  
Canada  
(877) 894 4414  
(11) 4197 3500  
305 269 7500  
Brazil  
Latin America  
Mexico  
www.lxistandard.org  
01800 5064 800  
(800) 829 4444  
LXI is the LAN-based successor to  
GPIB, providing faster, more efficient  
connectivity. Agilent is a founding  
member of the LXI consortium.  
United States  
Asia Pacific  
Australia  
China  
1 800 629 485  
800 810 0189  
800 938 693  
Agilent Channel Partners  
Hong Kong  
India  
www.agilent.com/find/advantageservices  
www.agilent.com/quality  
1 800 112 929  
0120 (421) 345  
080 769 0800  
1 800 888 848  
1 800 375 8100  
0800 047 866  
1 800 226 008  
www.agilent.com/find/channelpartners  
Get the best of both worlds: Agilent’s  
measurement expertise and product  
breadth, combined with channel  
partner convenience.  
Japan  
Korea  
Malaysia  
Singapore  
Taiwan  
Thailand  
Europe & Middle East  
Austria  
Belgium  
Denmark  
Finland  
France  
43 (0) 1 360 277 1571  
32 (0) 2 404 93 40  
45 70 13 15 15  
358 (0) 10 855 2100  
0825 010 700*  
*0.125 €/minute  
Germany  
Ireland  
49 (0) 7031 464 6333  
1890 924 204  
Israel  
972-3-9288-504/544  
39 02 92 60 8484  
31 (0) 20 547 2111  
34 (91) 631 3300  
0200-88 22 55  
Italy  
Netherlands  
Spain  
Sweden  
Switzerland  
0800 80 53 53  
United Kingdom 44 (0) 118 9276201  
Other European Countries:  
www.agilent.com/find/contactus  
Revised: July 8, 2010  
Product specifications and descriptions  
in this document subject to change  
without notice.  
© Agilent Technologies, Inc. 2010  
Printed in USA, October 25, 2010  
5990-6425EN  

相关型号:

U1273A

U1270 Series Handheld Digital Multimeters

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
KEYSIGHT

U1273AX

U1270 Series Handheld Digital Multimeters

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
KEYSIGHT

U12C05

POWER RECTIFIERS(12A,50-200V)

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
MOSPEC

U12C05_05

Switchmode Dual Ultrafast Power Rectifiers

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
MOSPEC

U12C10

POWER RECTIFIERS(12A,50-200V)

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
MOSPEC

U12C15

POWER RECTIFIERS(12A,50-200V)

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
MOSPEC

U12C20

POWER RECTIFIERS(12A,50-200V)

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
MOSPEC

U12C20A

12 Ampere Common Anode Ultra Fast Recovery Rectifier Diodes

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
THINKISEMI

U12C20A_18

12.0 Ampere Dual Common Anode Ultra Fast Recovery Rectifiers

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
THINKISEMI

U12C20C

12 Ampere Common Cathode Ultra Fast Recovery Rectifiers

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
THINKISEMI

U12C20C_18

12.0 Ampere Dual Common Cathode Ultra Fast Recovery Rectifiers

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
THINKISEMI

U12C20D

12.0 Ampere Dual Common Cathode Ultra Fast Recovery Rectifiers

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
THINKISEMI