Keysight Technologies  
E4980A Precision LCR Meter  
20 Hz to 2 MHz  
E4980AL Precision LCR Meter  
20 Hz to 300 kHz/500 kHz/1 MHz  
Data Sheet  
mpliant to  
LXClass C specification  
02 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
All specifications apply to the conditions of a 0 to 55 °C temperature range, unless  
otherwise stated, and 30 minutes after the instrument has been turned on.  
Definitions  
Specifications (spec.): Warranted performance. Specifications include guardbands  
to account for the expected statistical performance distribution, measurement  
uncertainties, and changes in performance due to environmental conditions.  
Supplemental information is provided as information that is useful in operating the  
instrument, but is not covered by the product warranty. This information is classified  
as either typical or nominal.  
Typical (typ.): Expected performance of an average unit without taking guardbands  
into account.  
Nominal (nom.): A general descriptive term that does not imply a level of performance.  
When measurement conditions fall under multiple categories in a table, apply the  
best value.  
How to Use Tables  
For example, basic accuracy Ab is 0.10% under the following conditions;  
Measurement time mode SHORT  
Test frequency  
125 Hz  
Test signal voltage  
0.3 Vrms  
The E4980A is the model number of the 20 Hz to 2 MHz frequency range LCR meter.  
The E4980AL is the model number of the 20 Hz to 300 kHz, 500 kHz or 1 MHz frequency  
range LCR meter. See the E4980A/E4980AL Configuration Guide (5989-8321EN) for  
more details.  
E4980A/E4980AL  
Frequency range  
20 Hz to 2 MHz  
20 Hz to 1 MHz  
20 Hz to 500 kHz  
20 Hz to 300 kHz  
Model number and option  
E4980A  
E4980AL-102  
E4980AL-052  
E4980AL-032  
03 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Measurement functions  
Basic Specifications  
Measurement parameters  
– Cp-D, Cp-Q, Cp-G, Cp-Rp  
– Cs-D, Cs-Q, Cs-Rs  
– Lp-D, Lp-Q, Lp-G, Lp-Rp, Lp-Rdc  
– Ls-D, Ls-Q, Ls-Rs, Ls-Rdc  
– R-X  
– Z-θd, Z-θr  
– G-B  
– Y-θd, Y-θr  
– Vdc-Idc1  
Definitions  
Cp  
Cs  
Lp  
Ls  
D
Capacitance value measured with parallel-equivalent circuit model  
Capacitance value measured with series-equivalent circuit model  
Inductance value measured with parallel-equivalent circuit model  
Inductance value measured with series-equivalent circuit model  
Dissipation factor  
Q
Quality factor (inverse of D)  
G
Rp  
Rs  
Equivalent parallel conductance measured with parallel-equivalent circuit model  
Equivalent parallel resistance measured with parallel-equivalent circuit model  
Equivalent series resistance measured with series-equivalent circuit model  
Rdc Direct-current resistance  
R
Resistance  
X
Reactance  
Z
Impedance  
Y
Admittance  
qd  
qr  
B
Phase angle of impedance/admittance (degree)  
Phase angle of impedance/admittance (radian)  
Susceptance  
Vdc Direct-current voltage  
Idc Direct-current electricity  
Deviation measurement function: Deviation from reference value and percentage of  
deviation from reference value can be output as the result.  
Equivalent circuits for measurement: Parallel, Series  
Impedance range selection: Auto (auto range mode), manual (hold range mode)  
Trigger mode: Internal trigger (INT), manual trigger (MAN), external trigger (EXT),  
GPIB trigger (BUS)  
1. E4980A-001 is required.  
04 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Table 1. Trigger delay time  
Range  
0 s - 999 s  
Resolution  
100 µs (0 s - 100 s)  
1 ms (100 s - 999 s)  
Table 2. Step delay time  
Range  
0 s - 999 s  
Resolution  
100 µs (0 s - 100 s)  
1 ms (100 s - 999 s)  
Measurement terminal: Four-terminal pair  
Test cable length: 0 m, 1 m, 2 m, 4 m  
Measurement time modes: Short mode, medium mode, long mode.  
Table 3. Averaging  
Range  
1 - 256 measurements  
1
Resolution  
Test signal  
Table 4. Test frequencies  
Test frequencies  
20 Hz - 2 MHz (E4980A)  
20 Hz - 1 MHz (E4980AL-102)  
20 Hz - 500 kHz (E4980AL-052)  
20 Hz - 300 kHz (E4980AL-032)  
Resolution  
0.01 Hz (20 Hz - 99.99 Hz)  
0.1 Hz (100 Hz - 999.9 Hz)  
1 Hz (1 kHz - 9.999 kHz)  
10 Hz (10 kHz - 99.99 kHz)  
100 Hz (100 kHz - 999.9 kHz)  
1 kHz (1 MHz - 2 MHz)  
Measurement accuracy ± 0.01%  
Table 5. Test signal modes  
Normal  
Program selected voltage or current at the measurement terminals when  
they are opened or short-circuited, respectively.  
Constant  
Maintains selected voltage or current at the device under test (DUT)  
independently of changes in impedance of DUT.  
05 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Signal level  
Table 6. Test signal voltage  
Range  
0 Vrms - 2.0 Vrms  
Resolution  
100 µVrms (0 Vrms - 0.2 Vrms)  
200 µVrms (0.2 Vrms - 0.5 Vrms)  
500 µVrms (0.5 Vrms - 1 Vrms)  
1 mVrms (1 Vrms - 2 Vrms)  
Accuracy  
Normal  
±(10% + 1 mVrms)  
Test frequency ≤ 1 MHz: spec.  
Test frequency > 1 MHz: typ.  
Constant1  
±(6% + 1 mVrms)  
Test frequency ≤ 1 MHz: spec.  
Test frequency > 1 MHz: typ.  
Table 7. Test signal current  
Range  
0 Arms - 20 mArms  
Resolution  
1 µArms (0 Arms - 2 mArms)  
2 µArms (2 mArms - 5 mArms)  
5 µArms (5 mArms - 10 mArms)  
10 µArms (10 mArms - 20 mArms)  
Accuracy  
Normal  
±(10% + 10 µArms)  
Test frequency ≤ 1 MHz: spec.  
Test frequency > 1 MHz: typ.  
Constant1  
±(6% + 10 µArms)  
Test frequency < = 1 MHz: spec.  
Test frequency > 1 MHz: typ.  
Output impedance: 100 Ω (nominal)  
Test signal level monitor function  
– Test signal voltage and test signal current can be monitored.  
– Level monitor accuracy:  
Table 8. Test signal voltage monitor accuracy (Vac)  
Test signal voltage2  
Test frequency  
≤ 1 MHz  
Specification  
5 mVrms - 2 Vrms  
± (3% of reading value + 0.5 mVrms)  
± (6% of reading value + 1 mVrms)  
> 1 MHz  
Table 9. Test signal current monitor accuracy (lac)  
Test signal current2  
Test frequency  
Specification  
50 µArms - 20 mArms  
≤ 1 MHz  
> 1 MHz  
± (3% of reading value + 5 µArms)  
± (6% of reading value + 10 µArms)  
1. When auto level control function is on.  
2. This is not an output value but rather a  
displayed test signal level.  
06 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Measurement display ranges  
Table 10 shows the range of measured value that can be displayed on the screen.  
For the effective measurement ranges, refer to Figure 1 impedance measurement  
accuracy example.  
Table 10. Allowable display ranges for measured values  
Parameter  
Measurement display range  
± 1.000000 aF to 999.9999 EF  
± 1.000000 aH to 999.9999 EH  
± 0.000001 to 9.999999  
Cs, Cp  
Ls, Lp  
D
Q
± 0.01 to 99999.99  
R, Rs, Rp, X, Z, Rdc  
± 1.000000 aΩ to 999.9999 EΩ  
± 1.000000 aS to 999.9999 ES  
± 1.000000 aV to 999.9999 EV  
± 1.000000 aA to 999.9999 EA  
± 1.000000 arad to 3.141593 rad  
± 0.0001 deg to 180.0000 deg  
± 0.0001 % to 999.9999 %  
G, B, Y  
Vdc  
Idc  
θr  
θd  
Δ%  
a: 1 x 10-18, E: 1 x 1018  
07 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Absolute measurement accuracy  
The following equations are used to calculate absolute accuracy.  
Absolute accuracy Aa of |Z|, |Y|, L, C, R, X, G, B (L, C, X, and B accuracies apply when  
Dx ≤ 0.1, R and G accuracies apply when Qx ≤ 0.1 )  
Equation 1.  
Aa = Ae + Acal  
Aa  
Ae  
Acal  
Absolute accuracy (% of reading value)  
Relative accuracy (% of reading value)  
Calibration accuracy (%)  
where G accuracy is applied only to G-B measurements.  
D accuracy (when Dx ≤ 0.1)  
Equation 2.  
De + θcal  
Dx  
Measured D value  
De  
θcal  
Relative accuracy of D  
Calibration accuracy of θ (radian)  
Q accuracy (When Qx × Da < 1)  
Equation 3.  
(Qx2 × Da)  
± ———————————  
(1 Qx × Da)  
Qx  
Da  
Measured Q value  
Absolute accuracy of D  
θ accuracy  
Equation 4.  
θe + θcal  
θe  
θcal  
Relative accuracy of θ (degree)  
Calibration accuracy of θ (degree)  
08 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
G accuracy (when Dx ≤ 0.1)  
Equation 5.  
Bx + Da (S)  
Bxꢀ=ꢀ2πfCxꢀ=ꢀꢀ—————  
ꢀꢀꢀꢀꢀꢀꢀꢀꢀꢀ2πfLxꢀ  
1
Dx  
Bx  
Da  
f
Cx  
Lx  
Measured D value  
Measured B value (S)  
Absolute accuracy of D  
Test frequency (Hz)  
Measured C value (F)  
Measured L value (H)  
where the accuracy of G is applied to Cp-G measurements.  
Absolute accuracy of Rp (when Dx ≤ 0.1)  
Equation 6.  
Rpx × Da  
± ꢀ ——————— ꢀ ꢀ ꢀ ꢀ ꢀ ꢀ ( Ω )  
Dx  
Da  
Rpx  
Dx  
Measured Rp value (Ω)  
Measured D value  
Da  
Absolute accuracy of D  
Absolute accuracy of Rs (when Dx ≤ 0.1)  
Equationꢀ7.ꢀꢀꢀ  
Xxꢀ×ꢀDaꢀ ꢀꢀꢀꢀꢀꢀꢀ(Ω)  
1
Xxꢀ=ꢀꢀ———— ꢀ=ꢀ2πfLx  
ꢀꢀꢀꢀꢀꢀꢀꢀꢀꢀꢀ2πfCx  
Dx  
Xx  
Da  
f
Cx  
Lx  
Measured D value  
Measured X value (Ω)  
Absolute accuracy of D  
Test frequency (Hz)  
Measured C value (F)  
Measured L value (H)  
09 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Relative accuracy  
Relative accuracy includes stability, temperature coefficient, linearity, repeatability, and  
calibration interpolation error. Relative accuracy is specified when all of the following  
conditions are satisfied:  
– Warm-up time: 30 minutes  
– Test cable length: 0 m, 1 m, 2 m, or 4 m (Keysight Technologies, Inc. 16048A/D/E)  
– A “Signal Source Overload” warning does not appear.  
When the test signal current exceeds a value in table 11 below, a “Signal Source  
Overload” warning appears.  
Table 11.  
Test signal voltage Test frequency  
Condition1  
≤ 2 Vrms  
> 2 Vrms  
≤ 1 MHz  
The smaller value of either 110 mA or  
130 mA - 0.0015 × Vac × (Fm / 1 MHz) × (L_cable + 0.5)  
70 mA - 0.0015 × Vac × (Fm / 1 MHz) × (L_cable + 0.5)  
> 1 MHz  
Vac [V]  
Fm [Hz]  
L_cable [m]  
Test signal voltage  
Test frequency  
Cable length  
– OPEN and SHORT corrections have been performed.  
– Bias current isolation: Off  
– The DC bias current does not exceed a set value within each range of the DC bias  
current  
– The optimum impedance range is selected by matching the impedance of DUT to the  
effective measuring range.  
– Under an AC magnetic field, the following equation is applied to the measurement  
accuracy.  
A x (1 + B x ( 2 + 0.5 / Vs))  
Where  
A:  
Absolute accuracy  
B:  
Vs:  
Magnetic flux density [Gauss]  
Test signal voltage level [Volts]  
|Z|, |Y|, L, C, R, X, G, and B accuracy (L, C, X, and B accuracies apply when Dx ≤ 0.1,  
R and G accuracies apply Qx ≤ 0.1)  
Relative accuracy Ae is given as:  
Equation 8.  
Ae = [Ab + Zs /|Zm| × 100 + Yo × |Zm| × 100 ] × Kt  
Impedance of DUT  
Basic accuracy  
Zm  
Ab  
Zs  
Yo  
Kt  
Short offset  
Open offset  
Temperature coefficient  
D accuracy  
D accuracy De is given as  
– when Dx ≤ 0.1  
Equation 9.  
De = ±Ae/100  
Measured D value  
Relative accuracies of |Z|, |Y|, L, C, R, X, G, and B  
Dx  
Ae  
1. When the calculation result is a negative  
value, 0 A is applied.  
– when Dx > 0.1, multiply De by (1 + Dx)  
10 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Q accuracy (when Q x De < 1)  
Q accuracy Qe is given as:  
Equation 10.  
(Qx2 × De)  
Q e = ± ————————————  
(1 Qx × De)  
Qx  
De  
Measured Q value  
Relative D accuracy  
θ accuracy  
θ accuracy θe is given as:  
Equation 11.  
180 × Ae  
θe = ————————— ( d e g )  
ꢀꢀꢀꢀꢀꢀꢀꢀꢀꢀꢀπꢀ×ꢀ100  
Ae  
Relative accuracies of |Z|, |Y|, L, C, R, X, G, and B  
G accuracy (when Dx ≤ 0.1)  
G accuracy Ge is given as:  
Equation 12.  
Ge = Bx × De  
(S)  
1
Bxꢀ=ꢀ2πfCxꢀ=ꢀꢀ————ꢀꢀꢀꢀ  
ꢀꢀꢀꢀꢀꢀꢀꢀꢀꢀ2πfLxꢀ  
Ge  
Dx  
Bx  
De  
f
Relative G accuracy  
Measured D value  
Measured B value  
Relative D accuracy  
Test frequency (Hz)  
Measured C value (F)  
Measured L value (H)  
Cx  
Lx  
Rp accuracy (when Dx ≤ 0.1)  
Rp accuracy Rpe is given as:  
Equationꢀ13.ꢀꢀ  
ꢀꢀRpxꢀ×ꢀDeꢀꢀꢀꢀꢀꢀꢀꢀ(Ω)  
R p e = ± ————————  
Dx  
De  
Rpe  
Rpx  
Dx  
Relative Rp accuracy  
Measured Rp value (Ω)  
Measured D value  
De  
Relative D accuracy  
Rs accuracy (when Dx ≤ 0.1)  
Rs accuracy Rse is given as:  
Equationꢀ14.ꢀꢀ  
Rseꢀ=ꢀXxꢀ×ꢀDeꢀ  
ꢀꢀꢀꢀꢀꢀꢀꢀꢀ(Ω)  
1
Xxꢀ=ꢀꢀ————– ꢀ=ꢀ2πfLx  
ꢀꢀꢀꢀꢀꢀꢀꢀꢀꢀꢀ2πfCx  
Rse  
Dx  
Xx  
De  
f
Relative Rs accuracy  
Measured D value  
Measured X value (Ω)  
Relative D accuracy  
Test frequency (Hz)  
Measured C value (F)  
Measured L value (H)  
Cx  
Lx  
11 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Example of C-D accuracy calculation  
Measurement conditions  
Test Frequency:  
1 kHz  
Measured C value:  
Test signal voltage:  
Measurement time mode:  
100 nF  
1 Vrms  
Medium  
Measurement temperature: 23 °C  
Ab = 0.05%  
|Zm| = 1 / (2π × 1 × 103 × 100 × 10-9) = 1590 Ω  
Zs = 0.6 mΩ × (1 + 0.400/1) × (1 + (1000/1000) = 1.68 mΩ  
Yo = 0.5 nS × (1 + 0.100/1) × (1 + (100/1000) = 0.72 nS  
C accuracy: Ae = [0.05 + 1.68 m/1590 × 100 + 0.72 n × 1590 × 100] × 1 = 0.05%  
D accuracy: De = 0.05/100 = 0.0005  
Basic accuracy  
Basic accuracy Ab is given below.  
Table 12. Measurement time mode = SHORT  
Test  
Test signal voltage  
frequency [Hz]  
5 mVrms -  
50 mVrms  
50 mVrms - 0.3 Vrms -  
1 Vrms -  
10 Vrms  
10 Vrms -  
20 Vrms  
0.3 Vrms  
1 Vrms  
20 - 125  
125 - 1 M  
1 M - 2 M  
(0.6%) ×  
(50 mVrms/Vs)  
0.60%  
0.30%  
0.30%  
0.15%  
0.30%  
0.30%  
0.15%  
0.30%  
(0.2%) ×  
(50 mVrms/Vs)  
0.20%  
0.40%  
0.10%  
0.20%  
(0.4%) ×  
(50 mVrms/Vs)  
Table 13. Measurement time mode = MED, LONG  
Test  
Test signal voltage  
frequency [Hz]  
5 mVrms - 50  
mVrms  
50 mVrms - 0.3 Vrms -  
1 Vrms -  
10 Vrms  
10 Vrms -  
20 Vrms  
0.3 Vrms  
1 Vrms  
20 - 100  
100 - 1 M  
1 M - 2 M  
(0.25%) ×  
(30 mVrms/Vs)  
0.25%  
0.10%  
0.15%  
0.10%  
0.20%  
0.15%  
0.15%  
0.30%  
(0.1%) ×  
(30 mVrms/Vs)  
0.10%  
0.20%  
0.05%  
0.10%  
(0.2%) ×  
(30 mVrms/Vs)  
Vs [Vrms]  
Test signal voltage  
12 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Effect by impedance of DUT  
Table 14. For impedance of DUT below 30 Ω, the following value is added.  
Test frequency [Hz]  
Impedance of DUT  
1.08 Ω ≤ |Zx| < 30 Ω  
0.05%  
|Zx| < 1.08 Ω  
0.10%  
20 - 1 M  
1 M - 2 M  
0.10%  
0.20%  
Table 15. For impedance of DUT over 9.2 k Ω, the following value is added.  
Test frequency [Hz]  
Impedance of DUT  
9.2 kΩ < |Zx| ≤ 92 kΩ  
92 kΩ < |Zx|  
0.05%  
10 k - 100 k  
100 k - 1 M  
1 M - 2 M  
0%  
0.05%  
0.10%  
0.05%  
0.10%  
Effect of cable extension  
When the cable is extended, the following element is added per one meter.  
0.015 % × (Fm/1 MHz)2 × (L_cable)2  
Fm [Hz]  
L_cable [m]  
Test frequency  
Cable length  
13 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Short offset Zs  
Table 16. Impedance of DUT > 1.08 Ω  
Test  
frequency [Hz]  
Measurement time mode  
SHORT  
MED, LONG  
20 - 2 M  
2.5 mΩ × (1 + 0.400/Vs) ×  
0.6 mΩ × (1 + 0.400/Vs) ×  
(1 + (1000/Fm))  
(1 + (1000/Fm))  
Table 17. Impedance of DUT ≤ 1.08 Ω  
Test  
frequency [Hz]  
Measurement time mode  
SHORT  
MED, LONG  
0.2 mΩ × (1 + 1/Vs) × (1 + (1000/Fm))  
20 - 2 M  
1 mΩ × (1 + 1/Vs) × (1 + (1000/Fm))  
Vs [Vrms]  
Fm [Hz]  
Test signal voltage  
Test frequency  
Effect of cable extension (Short offset)  
Table 18. When the cable is extended, the following value is added to Zs  
(independent of the measurement time mode).  
Test  
Cable length  
frequency [Hz]  
0 m  
0
1 m  
2 m  
4 m  
20 - 1 M  
0.25 mΩ  
1 mΩ  
0.5 mΩ  
2 mΩ  
1 mΩ  
4 mΩ  
1 M - 2 M  
0
Open offset Yo  
Table 19. Test signal voltage ≤ 2.0 Vrms  
Test  
frequency [Hz]  
Measurement time mode  
SHORT  
MED, LONG  
√ √  
2 nS × (1 + 0.100/Vs) × (1 + (100/Fm)) 0.5 nS × (1 + 0.100/Vs) × (1 + (100/Fm))  
20 - 100 k  
100 k - 1 M  
1 M - 2 M  
20 nS × (1 + 0.100/Vs)  
40 nS × (1 + 0.100/Vs)  
5 nS × (1 + 0.100/Vs)  
10 nS × (1 + 0.100/Vs)  
Table 20. Test signal voltage > 2.0 Vrms  
Test  
frequency [Hz]  
Measurement time mode  
SHORT  
MED, LONG  
20 - 100 k  
100 k - 1 M  
1 M - 2 M  
2 nS × (1 + 2/Vs) × (1 + (100/Fm))  
0.5 nS × (1 + 2/Vs) × (1 + (100/Fm))  
5 nS × (1 + 2/Vs)  
20 nS × (1 + 2/Vs)  
40 nS × (1 + 2/Vs)  
10 nS × (1 + 2/Vs)  
Vs [Vrms]  
Fm [Hz]  
Test signal voltage  
Test frequency  
Note  
The Open Offset may become three times greater  
in the ranges of 40 to 70 kHz and 80 to 100 kHz  
due to residual response.  
14 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Effect of cable length  
Table 21. When the cable is extended, multiply Yo by the following factor.  
Test  
Cable length  
frequency [Hz]  
0 m  
1
1 m  
2 m  
4 m  
100 - 100 k  
100 k - 1 M  
1 M - 2 M  
1 + 5 × Fm/1 MHz  
1 + 0.5 × Fm/1 MHz  
1 + 1 × Fm/1 MHz  
1 + 10 × Fm/1 MHz  
1 + 1 × Fm/1 MHz  
1 + 2 × Fm/1 MHz  
1 + 20 × Fm/1 MHz  
1 + 2 × Fm/1 MHz  
1 + 4 × Fm/1 MHz  
1
1
Fm [Hz] Test frequency  
Temperature factor Kt  
Table 22. The temperature factor Kt is given below.  
Temperature [°C]  
0 - 18  
Kt  
4
18 - 28  
1
28 - 55  
4
15 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Calibration accuracy Acal  
Calibration accuracy Acal is given below.  
For impedance of DUT on the boundary line, apply the smaller value.  
Table 23. Impedance range = 0.1, 1, 10 Ω  
Test frequency [Hz]  
20 - 1 k  
1 k - 10 k  
10 k -100 k  
100 k - 300 k 300 k - 1 M 1 M - 2 M  
0.05 + 0.05 + 0.1 +  
5 × 10-5 Fm 5 × 10-5 Fm 1 × 10-4 Fm  
3 × 10-4 3 × 10-4 6 × 10-4+  
|Z| [%]  
0.03  
0.05  
0.05  
θ [radian] 1 × 10-4  
2 × 10-4  
3 × 10-4  
+
+
2 × 10-7 Fm 2 × 10-7 Fm 4 × 10-7 Fm  
Table 24. Impedance range = 100 Ω  
Test frequency [Hz]  
20 - 1 k  
1 k - 10 k  
10 k -100 k  
100 k - 300 k 300 k - 1 M 1 M - 2 M  
|Z| [%]  
0.03  
0.05  
0.05  
0.05 +  
5 × 10-5 Fm 5 × 10-5 Fm 1 × 10-4 Fm  
3 × 10-4 3 × 10-4 6 × 10-4  
0.05 +  
0.1 +  
θ [radian] 1 × 10-4  
2 × 10-4  
3 × 10-4  
Table 25. Impedance range = 300, 1 kΩ  
Test frequency [Hz]  
20 - 1 k  
1 k - 10 k  
0.03  
10 k -100 k  
0.05  
100 k - 300 k 300 k - 1 M 1 M - 2 M  
|Z| [%]  
0.03  
0.05  
0.05  
0.1  
θ [radian] 1 × 10-4  
1 × 10-4  
3 × 10-4  
3 × 10-4  
3 × 10-4  
6 × 10-4  
Table 26. Impedance range = 3 k, 10 kΩ  
Test frequency [Hz]  
20 - 1 k  
1 k - 10 k  
10 k -100 k  
100 k - 300 k 300 k - 1 M 1 M - 2 M  
0.03 + 0.03 + 0.06 +  
|Z| [%]  
0.03 +  
0.03 +  
0.03 +  
1 × 10-4 Fm 1 × 10-4 Fm 1 × 10-4 Fm 1 × 10-4 Fm 1 × 10-4 Fm 2 × 10-4 Fm  
θ [radian] (100 + 2.5 Fm) (100 + 2.5 Fm) (100 + 2.5 Fm) (100 + 2.5 Fm) (100 + 2.5 Fm) (200 + 5 Fm)  
× 10-6 × 10-6 × 10-6 × 10-6 × 10-6 × 10-6  
Table 27. Impedance range = 30 k, 100 kΩ  
Test frequency [Hz]  
20 - 1 k  
1 k - 10 k  
10 k -100 k  
100 k - 300 k 300 k - 1 M 1 M - 2 M  
0.03 + 0.03 + 0.06 +  
|Z| [%]  
0.03 +  
0.03 +  
0.03 +  
1 × 10-3 Fm 1 × 10-3 Fm 1 × 10-3 Fm 1 × 10-3 Fm 1 × 10-4 Fm 2 × 10-4 Fm  
θ [radian] (100 + 20 Fm) (100 + 20 Fm) (100 + 20 Fm) (100 + 20 Fm) (100 + 2.5 Fm) (200 + 5 Fm)  
× 10-6 × 10-6 × 10-6 × 10-6 × 10-6 × 10-6  
Fm[kHz] Test frequency  
16 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Measurement accuracy  
The impedance measurement calculation example below is the result of absolute  
measurement accuracy.  
100aF  
100fF  
10fF  
1pF  
10pF  
1fF  
1MH  
100kH  
10kH  
1kH  
100H  
1n  
10n  
100n  
1µ  
1G  
100M  
10M  
1M  
100pF  
10.0%  
10H  
1H  
1nF  
10nF  
1.0%  
0.3%  
100mH  
10mH  
1mH  
0.1%  
100nF  
10µ  
100µ  
1m  
100k  
10k  
1k  
1
µ
F
H
µ
10  
100  
µ
F
H
µ
100  
10  
µ
F
10m  
100m  
1
100  
10  
0.1%  
H
µ
1mF  
10mF  
1
100nH  
10nH  
1nH  
1
0.3%  
1.0%  
100mF  
10  
100m  
10m  
1m  
1F  
100  
10  
10.0%  
100pH  
20  
100  
1k  
10k  
100k  
1M 2M  
Frequency [ Hz ]  
Figure 1. Impedance measurement accuracy (Test signal voltage = 1 Vrms, cable length=0 m, measurement time  
mode = MED)  
17 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Compensation function  
Table 28. The E4980A provides three types of compensation functions:  
OPEN compensation, SHORT compensation, and LOAD compensation.  
Type of compensation  
Description  
OPEN compensation  
Compensates errors caused by the stray admittance (C, G) of the test  
fixture.  
SHORT compensation  
LOAD compensation  
Compensates errors caused by the residual impedance (L, R) of the test  
fixture.  
Compensates errors between the actual measured value and a known  
standard value under the measurement conditions desired by the user.  
List sweep  
Points: There is a maximum of 201 points.  
First sweep parameter (primary parameter): Test frequency, test signal voltage, test  
signal current, test signal voltage of DC bias signal, test signal current of DC bias  
signal, DC source voltage.  
Note  
A parameter selected for one of the two  
parameters cannot be selected for the other  
parameter. It is not possible to set up a  
combination of test signal voltage and test  
signal current or one of test signal voltage of DC  
bias signal and test signal current of DC bias.  
Second sweep parameter (secondary parameter): None, impedance range, test  
frequency, test signal voltage, test signal current, test signal voltage of DC bias signal,  
test signal current of DC bias signal, DC source voltage  
The secondary parameter can be set only with  
SCPI commands.  
Trigger mode  
Sequential mode: When the E4980A is triggered once, the device is measured at all  
sweep points. /EOM/INDEX is output only once.  
Step mode: The sweep point is incremented each time the E4980A is triggered.  
/EOM/INDEX is output at each point, but the result of the comparator function of the  
list sweep is available only after the last /EOM is output.  
18 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Comparator function of list sweep: The comparator function enables setting one pair of  
lower and upper limits for each measurement point.  
You can select from: Judge with the first sweep parameter/Judge with the second  
parameter/Not used for each pair of limits.  
Time stamp function: In the sequential mode, it is possible to record the measurement  
starting time at each measurement point by defining the time when FW detects a trigger  
as 0 and obtain it later with the SCPI command.  
Comparator function  
Bin sort: The primary parameter can be sorted into 9 BINs, OUT_OF_BINS, AUX_BIN,  
and LOW_C_REJECT. The secondary parameter can be sorted into HIGH, IN, and LOW.  
The sequential mode and tolerance mode can be selected as the sorting mode.  
Limit setup: Absolute value, deviation value, and % deviation value can be used for setup.  
BIN count: Countable from 0 to 999999.  
DC bias signal  
Table 29. Test signal voltage  
Range  
0 V to +2 V  
Resolution  
Accuracy  
0 V / 1.5 V / 2 V only  
0.1% + 2 mV (23 °C ± 5 °C)  
(0.1% + 2 mV) × 4  
(0 to 18 °C or 28 to 55 °C)  
Output impedance:  
100 Ω (nominal)  
Measurement assistance functions  
Note  
Data buffer function: Up to 201 measurement results can be read out in a batch.  
The following USB memory can be used.  
Complies with USB 1.1; mass storage class,  
FAT16/FAT32 format; maximum consumption  
current is below 500 mA.  
Save/Recall function:  
– Up to 10 setup conditions can be written to/read from the built-in non-volatile  
memory.  
– Up to 10 setup conditions can be written to/read from the USB memory.  
– Auto recall function can be performed when the setting conditions are written to  
Register 10 of the USB memory.  
Recommended USB memory: 4 GB USB Flash  
memory (Keysight PN 1819-0637).  
Use the recommended USB memory device  
exclusively for the E4980A, otherwise,  
previously saved data may be cleared.  
If you use a USB memory other than the  
recommended device, data may not be  
saved or recalled normally.  
Key lock function: The front panel keys can be locked.  
GPIB: 24-pin D-Sub (Type D-24), female; complies with IEEE488.1, 2 and SCPI  
Keysight will NOT be responsible for data loss in  
the USB memory caused by using the E4980A.  
USB host port: Universal serial bus jack, type-A (4 contact positions, contact 1 is on  
your left), female (for connection to USB memory only).  
USB interface port: Universal serial bus jack, type mini-B (4 contact positions); complies  
with USBTMC-USB488 and USB 2.0; female; for connection to the external controller.  
USBTMC: Abbreviation for USB Test & Measurement Class  
LAN: 10/100 BaseT Ethernet, 8 pins (two speed options)  
LXI Compliance: Class C (only applies to units with firmware revision A.02.00 or later)  
19 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Frequency options  
Options  
E4980A  
20 Hz to 2 MHz  
20 Hz to 300 kHz  
20 Hz to 500 kHz  
20 Hz to 1 MHz  
Note  
E4980AL-032  
E4980AL-052  
E4980AL-102  
Option xxx is described as E4980A-xxx in the  
order information  
Table 30. Installable options  
Options  
E4980A  
E4980AL  
Power and DC bias enhancement (001)  
Bias Current Interface (002)  
Entry model (005)  
Installable  
Installable  
Installable  
Installable  
Installable1  
Installable  
Installable  
Not installable  
Not installable  
Not installable  
Not installable  
Not installable2  
Not installable  
Not installable  
Standard model (007)  
DCR measurement (200)  
Handler interface (201)  
Scanner interface (301)  
Interface options  
Option 002 (Bias current interface)  
Adds a digital interface to allow the E4980A to control the 42841A’s bias current source.  
Option 201 (Handler interface)  
Adds handler interface.  
Option 301 (Scanner interface)  
Adds scanner interface.  
Option 710 (No interface)  
An option with no interface.  
Up to 2 interface options can be installed in the interface connector on the rear panel.  
When no interface is installed, two of the option 710 are installed. When one interface is  
installed, the option number of its interface and one option 710 are installed.  
Other options  
Option 001 (Power and DC Bias enhancement)  
Increases test signal voltage and adds the variable DC bias voltage.  
Option 005 (Entry model)  
Low-priced option that requires a longer measurement time. The level of measurement  
accuracy is the same as the standard.  
Option 007 (Standard model)  
Note  
Upgrades the entry model to the standard.  
Option 007 can be installed only in the E4980A  
with option 005.  
Option 200 (DCR measurement)  
Adds DCR measurement  
Note  
E4980A-200/001 and E4980AL-032/052/102  
supports DCR measurement function.  
1. Mandatory option  
2. DCR measurement function is equipped  
by default.  
20 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Power and DC bias enhancement specification  
Increases test signal voltage and adds the variable DC bias voltage function.  
The Vdc-Idc measurement function is available when the option 001 is installed.  
Measurement parameters  
The following parameters can be used.  
– Lp-Rdc  
– Ls-Rdc  
– Vdc-Idc  
where  
Rdc  
Vdc  
Idc  
Direct-current resistance (DCR)  
Direct-current voltage  
Direct-current electricity  
Test signal  
Signal level  
Table 31. Test signal voltage  
Range  
0 Vrms to 20 Vrms (test frequency ≤ 1 MHz)  
0 Vrms to 15 Vrms (test frequency > 1 MHz)  
Resolution  
100 µVrms (0 Vrms - 0.2 Vrms)  
200 µVrms (0.2 Vrms - 0.5 Vrms)  
500 µVrms (0.5 Vrms - 1 Vrms)  
1 mVrms (1 Vrms - 2 Vrms)  
2 mVrms (2 Vrms - 5 Vrms)  
5 mVrms (5 Vrms - 10 Vrms)  
10 mVrms (10 Vrms - 20 Vrms)  
Setup accuracy  
normal ± (10% + 1 mVrms) (test signal voltage ≤ 2 Vrms)  
(test frequency ≤ 1 MHz : spec., test frequency > 1 MHz : typ.)  
± (10% + 10 mVrms) (Test frequency ≤ 300 kHz,  
test signal voltage > 2 Vrms) (spec.)  
± (15% + 20 mVrms) (test frequency > 300 kHz,  
test signal voltage > 2 Vrms) (test frequency ≤ 1 MHz : spec., test  
frequency > 1 MHz : typ.)  
Constant1 ± (6% + 1 mVrms) (test signal voltage ≤ 2 Vrms)  
(test frequency ≤ 1 MHz : spec., test frequency > 1 MHz : typ.)  
± (6% + 10 mVrms) (test frequency ≤ 300 kHz,  
test signal voltage > 2 Vrms) (spec.)  
± (12% + 20 mVrms) (test frequency > 300 kHz,  
test signal voltage > 2 Vrms) (test frequency ≤ 1 MHz : spec.,  
test frequency > 1 MHz : typ.)  
1. When auto level control function is on.  
21 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Table 32. Test signal current  
Range  
0 Arms - 100 mArms  
Resolution  
1 µArms (0 Arms - 2 mArms)  
2 µArms (2 mArms - 5 mArms)  
5 µArms (5 mArms - 10 mArms)  
10 µArms (10 mArms - 20 mArms)  
20 µArms (20 mArms - 50 mArms)  
50 µArms (50 mArms - 100 mArms)  
Setup accuracy  
normal  
±(10% + 10 µArms) (test signal voltage ≤ 20 mArms)  
(test frequency ≤ 1 MHz : spec., test frequency > 1 MHz : typ.)  
±(10% + 100 µArms) (test frequency ≤ 300 kHz,  
test signal current > 20 mArms) (spec.)  
±(15% + 200 µArms) (test frequency > 300 kHz,  
test signal voltage > 20 mArms) (test frequency ≤ 1 MHz : spec.,  
test frequency > 1 MHz : typ.)  
Constant1  
±(6% + 10 µArms) (test signal voltage ≤ 20 mArms)  
(test frequency ≤ 1 MHz : spec., test frequency > 1 MHz : typ.)  
±(6% + 100 µArms) (test frequency ≤ 300 kHz,  
test signal voltage > 20 mArms) (spec.)  
±(12% + 200 µArms) (test frequency > 300 kHz,  
test signal voltage > 20 mArms) (test frequency ≤ 1 MHz : spec.,  
test frequency > 1 MHz : typ.)  
Test signal level monitor function  
– Test signal voltage and test signal current can be monitored.  
– Level monitor accuracy:  
Table 33. Test signal voltage monitor accuracy (Vac)  
Test signal voltage2  
Test frequency  
≤ 1 MHz  
Specification  
5 mVrms to 2 Vrms  
±(3% of reading value + 0.5 mVrms)  
±(6% of reading value + 1 mVrms)  
±(3% of reading value + 5 mVrms)  
±(6% of reading value + 10 mVrms)3  
> 1MHz  
> 2 Vrms  
≤ 300 kHz  
> 300 kHz  
Table 34. Test signal current monitor accuracy (Iac)  
Test signal current2  
Test frequency  
≤ 1 MHz  
Specification  
50 µArms to 20 mArms  
±(3% of reading value + 5 µArms)  
±(6% of reading value + 10 µArms)  
±(3% of reading value + 50 µArms)  
±(6% of reading value + 100 µArms)  
> 1MHz  
> 20 mArms  
≤ 300 kHz  
> 300 kHz  
1. When auto level control function is on.  
2. This is not an output value but a displayed test  
signal level  
3. Typ. when test frequency is > 1 MHz with test  
signal voltage > 10 Vrms.  
22 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
DC bias signal  
Table 35. Test signal voltage  
Range  
–40 V to +40 V  
Resolution  
Setup resolution: 100 µV, effective resolution:  
330 µV ±(0 V - 5 V)  
1 mV ±(5 V - 10 V)  
2 mV ±(10 V - 20 V)  
5 mV ±(20 V - 40 V)  
Accuracy  
test signal voltage ≤ 2 Vrms  
0.1% + 2 mV (23 °C ± 5 °C)  
(0.1% + 2 mV) x 4  
(0 to 18 °C or 28 to 55 °C)  
0.1 % + 4 mV (23 °C ± 5 °C)  
(0.1% + 4 mV) x 4  
test signal voltage > 2 Vrms  
(0 to 18 °C or 28 to 55 °C)  
Table 36. Test signal current  
Range  
–100 mA - 100 mA  
Resolution  
Setup resolution: 1 µA, effective resolution:  
3.3 µA ±(0 A - 50 mA)  
10 µA ±(50 mA - 100 mA)  
DC bias voltage level monitor Vdc  
(0.5% of reading value + 60 mV) × Kt  
When using Vdc-Idc measurement: (spec.)  
When using level monitor: (typ.)  
Kt  
Temperature coefficient  
DC bias current level monitor Idc  
(A [%] of the measurement value + B [A]) × Kt  
When using Vdc-Idc measurement: (spec.)  
When using level monitor: (typ.)  
A [%]  
When the measurement time mode is SHORT: 2%  
When the measurement time mode is MED or LONG: 1%  
B [A]  
Kt  
given below  
Temperature coefficient  
When the measurement mode is SHORT, double the following value.  
23 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Table 37. Test signal voltage ≤ 0.2 Vrms (measurement time mode = MED, LONG)  
DC bias Impedance range [Ω]  
current range  
< 100  
100  
300, 1 k  
3 µA  
3 k, 10 k  
300 nA  
300 nA  
3 µA  
30k, 100 k  
45 nA  
20 µA  
200 µA  
2 mA  
150 µA  
150 µA  
150 µA  
150 µA  
150 µA  
30 µA  
30 µA  
30 µA  
30 µA  
150 µA  
3 µA  
300 nA  
3 µA  
3 µA  
20 mA  
100 mA  
30 µA  
150 µA  
30 µA  
30 µA  
150 µA  
150 µA  
Table 38. 0.2 Vrms < test signal voltage ≤ 2 Vrms (measurement time mode = MED, LONG)  
DC bias  
Impedance range [Ω]  
current range  
< 100  
100, 300  
1k, 3 k  
3 µA  
10k, 30 k  
300 nA  
300 nA  
3 µA  
100 k  
45 nA  
300 nA  
3 µA  
20 µA  
200 µA  
2 mA  
150 µA  
150 µA  
150 µA  
150 µA  
150 µA  
30 µA  
30 µA  
30 µA  
30 µA  
150 µA  
3 µA  
3 µA  
20 mA  
100 mA  
30 µA  
150 µA  
30 µA  
30 µA  
150 µA  
150 µA  
Table 39. Test signal voltage > 2 Vrms (measurement time mode = MED, LONG)  
DC bias  
Impedance range [Ω]  
current range  
≤ 300  
150 µA  
150 µA  
150 µA  
150 µA  
150 µA  
1 k, 3 k  
10k, 30 k  
3 µA  
100 k  
20 µA  
30 µA  
30 µA  
30 µA  
30 µA  
150 µA  
300 nA  
300 nA  
3 µA  
200 µA  
2 mA  
3 µA  
3 µA  
20 mA  
100 mA  
30 µA  
150 µA  
30 µA  
150 µA  
Table 40. Input impedance (nominal)  
Input impedance  
Conditions  
Other than conditions below.  
0 Ω  
20 Ω  
Test signal voltage ≤ 0.2 Vrms, Impedance range ≥ 3 k Ω,  
DC bias current range ≤ 200 µA  
Test signal voltage ≤ 2 Vrms, Impedance range ≥ 10 kΩ,  
DC bias current range ≤ 200 µA  
Test signal voltage > 2 Vrms, Impedance range = 100 kΩ,  
DC bias current range ≤ 200 µA  
DC source signal  
Table 41. Test signal voltage  
Range  
–10 V to 10 V  
Resolution  
Accuracy  
1 mV  
0.1% + 3 mV (23 °C ± 5 °C)  
(0.1% + 3 mV) x 4  
(0 to 18 °C or 28 to 55 °C)  
Table 42. Test signal current  
Range  
–45 mA to 45 mA (nominal)  
Output impedance  
100 Ω (nominal)  
24 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
DCR measurement specification  
DC resistance (Rdc) measurement function is available when either E4980A-001/200 or  
E4980AL-032/052/102 is installed.  
DC resistance (Rdc) accuracy  
Absolute measurement accuracy Aa  
Absolute measurement accuracy Aa is given as  
Equation 15.  
Aa = Ae + Acal  
Aa  
Ae  
Acal  
Absolute accuracy (% of reading value)  
Relative accuracy (% of reading value)  
Calibration accuracy  
Relative measurement accuracy Ae  
Relative measurement accuracy Ae is given as  
Equation 16.  
Ae = [Ab + (Rs /|Rm|+ Go × |Rm|) × 100 ] × Kt  
Rm  
Ab  
Rs  
Go  
Kt  
Measurement value  
Basic accuracy  
Short offset [Ω]  
Open offset [S]  
Temperature coefficient  
Calibration accuracy Acal  
Calibration accuracy Acal is 0.03%.  
Basic accuracy Ab  
Table 43. Basic accuracy Ab is given below.  
Measurement time mode  
Test signal voltage  
≤ 2 Vrms  
1.00%  
> 2 Vrms  
2.00%  
SHORT  
MED  
0.30%  
0.60%  
Open offset Go  
Table 44. Open offset Go is given below.  
Measurement time mode  
Test signal voltage  
≤ 2 Vrms  
50 nS  
> 2 Vrms  
500 nS  
100 nS  
SHORT  
MED  
10 nS  
Short offset Rs  
Table 45. Short offset Rs is given below.  
Measurement time mode  
Test signal voltage  
≤ 2 Vrms  
25 mΩ  
5 mΩ  
> 2 Vrms  
250 mΩ  
50 mΩ  
SHORT  
MED  
25 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Effect of cable length (Short offset)  
Table 46. The following value is added to Rs when the cable is extended.  
Cable length  
1 m  
2 m  
4 m  
0.25 mΩ  
0.5 mΩ  
1 mΩ  
Temperature coefficient Kt  
Table 47. Temperature coefficient Kt is given below.  
Temperature [°C]  
0 - 18  
Kt  
4
18 - 28  
1
28 - 55  
4
26 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Table 48. Power source  
General Specifications  
Voltage  
90 VAC - 264 VAC  
47 Hz - 63 Hz  
Max. 150 VA  
Frequency  
Power consumption  
Table 49. Operating environment  
Temperature  
0 - 55 °C  
Humidity (≤ 40 °C, no condensation)  
Altitude  
15% - 85% RH  
0 m - 2000 m  
Table 50. Storage environment  
Temperature  
–20 - 70 °C  
Humidity ( ≤ 60 °C, no condensation)  
Altitude  
0% - 90% RH  
0 m - 4572 m  
Outer dimensions: 375 (width) x 105 (height) × 390 (depth) mm (nominal)  
367.4  
14.4  
338.6  
14.4  
41.8  
E4980A  
20 Hz - 2 MHz  
Preset  
Trigger  
Precision LCR Meter  
7
4
1
0
8
5
9
6
3
DC  
DC  
Source  
USB  
Bias  
2
.
UNKNOWN  
DC Bias  
Discharge test device before connecting  
42V Peak Max Output CAT  
Display  
Format  
Meas  
Setup  
I
DC  
Source  
L
CUR  
L
POT  
H
POT  
HCUR  
DC Source  
(Option 001)  
Return  
10VDC Max  
Save/  
Recall  
Local/  
Lock  
Recall  
A
Recall  
B
System  
22  
27  
22  
22  
40.1  
55.2  
Figure 2. Dimensions (front view, with handle and bumper, in millimeters, nominal)  
319.1  
32.0  
E4980A  
20 Hz - 2 MHz  
Precision LCR Meter  
Preset  
Trigger  
7
4
1
0
8
5
9
6
3
DC  
DC  
Source  
Bias  
USB  
2
.
UNKNOWN  
DC Bias  
Discharge test device before connecting  
42V Peak Max Output CAT  
Display  
Format  
Meas  
Setup  
I
DC  
Source  
L
CUR  
L
POT  
H
POT  
H
CUR  
DC Source  
(Option 001)  
Return  
10VDC Max  
Save/  
Recall  
Local/  
Lock  
Recall  
A
Recall  
B
System  
55.2  
27  
22  
22  
22  
30.3  
Figure 3. Dimensions (front view, without handle and bumper, in millimeters, nominal)  
27 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
367.4  
332.2  
17.6  
17.6  
118.1  
41.6  
23.2  
34.7  
Serial Label  
E4980A  
LAN  
Trigger  
GPIB  
115V  
-230V  
50/60Hz  
150VA MAX  
Fuse  
Option 710: No Interface  
Option 710: No Interface  
T3A , 250V  
113.9  
72.3  
72.3  
31.1  
0.4  
Option 002: DC Current Control Interface  
Option 301: Scanner Interface  
Option 201: Handler Interface  
36.2  
36.2  
72.3  
36.2  
72.3  
72.3  
Figure 4. Dimensions (rear view, with handle and bumper, in millimeters, nominal)  
317.8  
110.9  
41.6  
23.2  
27.5  
Serial Label  
E4980A  
LAN  
Trigger  
GPIB  
115V  
-230V  
50/60Hz  
150VA MAX  
Fuse  
Option 710: No Interface  
Option 710: No Interface  
T3A , 250V  
106.7  
72.3  
72.3  
24.0  
0.4  
Option 002: DC Current Control Interface  
Option 301: Scanner Interface  
Option 201: Handler Interface  
Figure 4. Dimensions (rear view, with handle and bumper, in millimeters, nominal)  
28 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
388.7  
66.6  
141.4  
Figure 6. Dimensions (side view, with handle and bumper, in millimeters, nominal)  
374.0  
347.9  
15.7  
10.5  
19.7  
50.9  
Figure 7. Dimensions (side view, without handle and bumper, in millimeters, nominal)  
Weight: 5.3 kg (nominal)  
Display: LCD, 320 × 240 (pixels), RGB color  
Note  
Effective pixels are more than 99.99%. There  
may be 0.01% (approx. 7 pixels) or smaller  
missing pixels or constantly lit pixels, but this  
is not a malfunction.  
The following items can be displayed:  
– Measurement value  
– Measurement conditions  
– Limit value and judgment result of comparator  
– List sweep table  
– Self-test message  
29 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Description  
EMC  
Supplemental Information  
European Council Directive 2004/108/EC  
IEC 61326-1:2012  
EN 61326-1:2013  
CISPR 11:2009 +A1:2010  
EN 55011: 2009 +A1:2010  
Group 1, Class A  
IEC 61000-4-2:2008  
EN 61000-4-2:2009  
4 kV CD / 8 kV AD  
IEC 61000-4-3:2006 +A1:2007 +A2:2010  
EN 61000-4-3:2006 +A1:2008 +A2:2010  
3 V/m, 80-1000 MHz, 1.4 - 2.0 GHz / 1V/m, 2.0 - 2.7 GHz, 80% AM  
IEC 61000-4-4:2004 +A1:2010  
EN 61000-4-4:2004 +A1:2010  
1 kV power lines / 0.5 kV signal lines  
IEC 61000-4-5:2005  
EN 61000-4-5:2006  
0.5 kV line-line / 1 kV line-ground  
IEC 61000-4-6:2008  
EN 61000-4-6:2009  
3 V, 0.15-80 MHz, 80% AM  
IEC 61000-4-8:2009  
EN 61000-4-8:2010  
30A/m, 50/60Hz  
IEC 61000-4-11:2004  
EN 61000-4-11:2004  
0.5-300 cycle, 0% / 70%  
ICES/NMB-001  
ICES-001:2006 Group 1, Class A  
AS/NZS CISPR11:2004  
Group 1, Class A  
KN11, KN61000-6-1 and KN61000-6-2  
Group 1, Class A  
Safety  
European Council Directive 2006/95/EC  
IEC 61010-1:2001/EN 61010-1:2001  
Measurement Category I, Pollution Degree 2, Indoor Use  
IEC60825-1:1994 Class 1 LED  
CAN/CSA C22.2 61010-1-04  
Measurement Category I, Pollution Degree 2, Indoor Use  
Environment  
This product complies with the WEEE Directive (2002/96/EC) marking  
requirements. The afixed label indicates that you must not discard this elec-  
trical/electronic product in domestic house hold waste.  
Product Category: With reference to the equipment types in the WEEE  
Directive Annex I, this product is classed as a “Monitoring and Control  
instrumentation” product.  
30 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Settling time  
Supplemental Information  
Table 51. Test frequency setting time  
Test frequency setting time  
Test frequency (Fm)  
Fm ≥ 1 kHz  
5 ms  
12 ms  
22 ms  
42 ms  
1 kHz > Fm ≥ 250 Hz  
250 Hz > Fm ≥ 60 Hz  
60 Hz > Fm  
Table 52. Test signal voltage setting time  
Test signal voltage setting time  
Test frequency (Fm)  
Fm ≥ 1 kHz  
11 ms  
18 ms  
26 ms  
48 ms  
1 kHz > Fm ≥ 250 Hz  
250 Hz > Fm ≥ 60 Hz  
60 Hz > Fm  
Switching of the impedance range is as follows:  
≤ 5 ms/ range switching  
Measurement circuit protection  
The maximum discharge withstand voltage, where the internal circuit remains protected  
if a charged capacitor is connected to the UNKNOWN terminal, is given below.  
Note  
Discharge capacitors before connecting them  
to the UNKNOWN terminal or a test fixture to  
avoid damages to the instrument.  
Table 53. Maximum discharge withstand voltage  
Maximum discharge withstand voltage  
Range of capacitance value C of DUT  
1000 V  
C < 2 µF  
———  
2 µF ≤ C  
2/C V  
1200  
1000  
800  
600  
400  
200  
0
1.E–15  
1.E–13  
1.E–11  
1.E–09  
1.E–07  
1.E–05  
1.E–03  
Capacitance [F]  
Figure 8. Maximum discharge withstand voltage  
31 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Measurement time  
Definition  
This is the time between the trigger and the end of measurement (EOM) output on the  
handler interface.  
Conditions  
Table 54 shows the measurement time when the following conditions are satisfied:  
– Normal impedance measurement other than Ls-Rdc, Lp-Rdc, Vdc-Idc  
– Impedance range mode: hold range mode  
– DC bias voltage level monitor: OFF  
– DC bias current level monitor: OFF  
– Trigger delay: 0 s  
– Step delay: 0 s  
– Calibration data: OFF  
– Display mode: blank  
Table 54. E4980A measurement time [ms](DC bias:OFF)  
Measurement  
time mode  
Test frequency  
20 Hz  
480  
100 Hz  
1 kHz  
240  
110  
10 kHz  
230  
92  
100 kHz 1 MHz  
2 MHz  
220  
88  
1
2
3
LONG  
MED  
300  
180  
100  
220  
89  
220  
88  
380  
SHORT  
330  
20  
7.7  
5.7  
5.6  
5.6  
10  
1
1. LONG  
2. MED  
3. SHORT  
0.1  
0.01  
0.001  
20  
100  
1k  
10k  
Test frequency [Hz]  
100k  
1M 2M  
Figure 9. Measurement time (E4980A, DC bias: OFF)  
32 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Table 55. E4980A-005 measurement time [ms] (DC bias: OFF)  
Measurement  
time mode  
Test frequency  
20 Hz  
1190  
1150  
1040  
100 Hz 1 kHz  
10 kHz 100 kHz 1 MHz  
2 MHz  
570  
180  
1
2
3
LONG  
MED  
650  
380  
240  
590  
200  
37  
580  
180  
25  
570  
180  
23  
570  
180  
23  
SHORT  
23  
10  
1
1. LONG  
2. MED  
3. SHORT  
0.1  
0.01  
0.001  
20  
100  
1k  
10k  
Test frequency [Hz]  
100k  
1M 2M  
Figure 10. Measurement time (DC bias: OFF, E4980A-005)  
Table 56. E4980AL measurement time [ms]  
Measurement  
time mode  
Test frequency  
20 Hz  
729  
100 Hz  
1 kHz  
10 kHz  
100 kHz  
343  
1 MHz  
1
2
3
LONG  
MED  
423  
250  
149  
363  
140  
26  
353  
122  
14  
343  
118  
12  
650  
119  
SHORT  
579  
12  
10  
1
1. LONG  
2. MED  
3. SHORT  
0.1  
0.01  
0.001  
20  
100  
1k  
10k  
Test frequency [Hz]  
100k  
1M 2M  
Figure 11. Measurement time (E4980AL)  
33 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
When DC bias is ON, the following time is added:  
Table 57. Additional time when DC bias is ON [ms]  
Test frequency  
20 Hz  
100 Hz  
1 kHz  
10 kHz  
100 kHz  
1 MHz  
2 MHz  
30  
30  
10  
13  
2
0.5  
0.5  
When the number of averaging increases, the measurement time is given as  
Equation 17.  
MeasTime + (Ave – 1) × AveTime  
MeasTime  
Ave  
Measurement time calculated based on Table 53 and Table 54  
Number of averaging  
AveTime  
Refer to Table 56  
Table 58. Additional time per averaging [ms]  
Measurement  
time mode  
Test frequency  
20 Hz  
51  
100 Hz  
1 kHz  
2.4  
10 kHz  
2.3  
100 kHz 1 MHz  
2 MHz  
2.2  
SHORT  
MED  
11  
2.3  
85  
2.2  
84  
110  
81  
88  
87  
84  
LONG  
210  
210  
220  
220  
220  
210  
210  
Table 59. Measurement time when Vdc-Idc is selected [ms]  
Measurement time Test frequency  
mode  
20 Hz  
210  
100 Hz  
46  
1 kHz  
14  
10 kHz  
14  
100 kHz 1 MHz  
2 MHz  
14  
SHORT  
MED  
14  
14  
210  
170  
170  
410  
170  
170  
410  
170  
410  
170  
410  
LONG  
410  
410  
410  
Add the same measurement time per 1 additional average  
Additional Measurement time when the Vdc and Idc monitor function is ON.  
Add SHORT mode of Table 59. When using only Vdc or Idc, add a half of SHORT mode of  
Table 59.  
Table 60. Measurement time when Ls-Rdc or Lp-Rdc is selected [ms]  
Measurement time Test frequency  
mode  
20 Hz  
910  
100 Hz  
230  
1 kHz  
43  
10 kHz  
24  
100 kHz 1 MHz  
2 MHz  
22  
SHORT  
MED  
22  
22  
1100  
1400  
450  
300  
700  
280  
670  
270  
660  
270  
650  
270  
LONG  
820  
650  
Add the three times of measurement time per 1 additional average number  
34 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Display time  
Except for the case of the DISPLAY BLANK page, the time required to update the display  
on each page (display time) is as follows. When a screen is changed, drawing time and  
switching time are added. The measurement display is updated about every 100 ms.  
Table 61. Display time  
Item  
When Vdc, Idc  
monitor is OFF  
When Vdc, Idc  
monitor is ON  
MEAS DISPLAY page drawing time  
10 ms  
10 ms  
10 ms  
10 ms  
40 ms  
35 ms  
13 ms  
13 ms  
13 ms  
13 ms  
MEAS DISPLAY page (large) drawing time  
BIN No. DISPLAY page drawing time  
BIN COUNT DISPLAY page drawing time  
LIST SWEEP DISPLAY page drawing time  
Measurement display switching time  
Measurement data transfer time  
This table shows the measurement data transfer time under the following conditions.  
The measurement data transfer time varies depending on measurement conditions and  
computers.  
Table 62. Measurement transfer time under the following conditions:  
Host computer:  
DELL OPTIPLEX GX260 Pentium 4 2.6 GHz  
Display:  
ON  
Impedance range mode:  
OPEN/SHORT/LOAD compensation:  
Test signal voltage monitor:  
AUTO (The overload has not been generated.)  
OFF  
OFF  
Table 63. Measurement data transfer time [ms]  
Interface Data  
transfer  
Using :FETC? command  
(one point measurement)  
Using data buffer memory  
(list sweep measurement)  
format  
Comparator Comparator 10  
51  
128  
201  
ON  
OFF  
points  
points  
points  
points  
GPIB  
USB  
LAN  
ASCII  
2
2
4
5
4
3
4
3
5
5
5
13  
15  
10  
8
28  
34  
21  
16  
19  
9
43  
53  
32  
23  
28  
13  
36  
44  
26  
ASCII Long  
Binary  
2
2
2
2
ASCII  
2
2
ASCII Long  
Binary  
2
2
9
2
2
5
ASCII  
3
4
12  
13  
9
24  
29  
18  
ASCII Long  
Binary  
3
3
3
3
35 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
DC bias test signal current (1.5 V/2.0 V): Output current: Max. 20 mA  
Option 001 (Power and DC Bias enhance):  
DC bias voltage: DC bias voltage applied to DUT is given as:  
Equation 18.  
Vdut = Vb – 100 × Ib  
Vdut [V] DC bias voltage  
Vb [V] DC bias setting voltage  
Ib [A]  
DC bias current: DC bias current applied to DUT is given as:  
Equation 19. Idut = Vb/(100 + Rdc)  
DC bias current  
Idut [A] DC bias current  
Vb [V] DC bias setting current  
Rdc [Ω] DUT’s DC resistance  
Maximum DC bias current  
Table 64. Maximum DC bias current when the normal measurement can be performed.  
Impedance Bias current isolation  
range [Ω]  
ON  
OFF  
Test signal voltage ≤ 2 Vrms  
Test signal voltage > 2 Vrms  
0.1  
Auto range mode: 20 mA  
100 mA  
100 mA  
100 mA  
100 mA  
100 mA  
20 mA  
20 mA  
2 mA  
100 mA  
1
20 mA  
20 mA  
20 mA  
2 mA  
Hold range mode:  
its values for  
the range.  
10  
100  
300  
1 k  
2 mA  
3 k  
200 µA  
200 µA  
20 µA  
20 µA  
10 k  
30 k  
100 k  
2 mA  
200 µA  
When DC bias is applied to DUT  
When DC bias is applied to the DUT, add the following value to the absolute accuracy Ab.  
Table 65. Only when Fm < 10 kHz and |Vdc| > 5 V  
SHORT  
MED, LONG  
0.05% × (100 mV/Vs) × (1 + √(100/Fm))  
0.01% × (100 mV/Vs) × (1 + √(100/Fm))  
Fm [Hz] Test frequency  
Vs [V]  
Test signal voltage  
36 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Relative measurement accuracy with bias current isolation  
When DC bias Isolation is set to ON, add the following value to the open offset Yo.  
Equationꢀ20.ꢀ  
Yo_DCI1ꢀ×ꢀ(1ꢀ+ꢀ1/(Vs))ꢀ×ꢀ(1ꢀ+ꢀꢀ (500/Fm))ꢀ+ꢀYo_DCI2  
Zm [Ω]  
Impedance of DUT  
Fm [Hz]  
Vs [V]  
Test frequency  
Test signal voltage  
Yo_DCI1,2 [S]  
Idc [A]  
Calculate this by using Table 61 and 62  
DC bias isolation current  
Table 66. Yo_DCI1 value  
DC bias current range  
Measurement time mode  
SHORT  
0 S  
0.25 nS  
2.5 nS  
25 nS  
250 nS  
MED, LONG  
0 S  
0.05 nS  
0.5 nS  
5 nS  
20 µA  
200 µA  
2 mA  
20 mA  
100 mA  
50 nS  
Table 67. Yo_DCI2 value  
DC bias current  
range  
20 µA  
Measurement time mode  
≤ 100 Ω  
0 S  
300 Ω, 1 k Ω  
0 S  
3 k Ω, 10 k Ω  
0 S  
30 k Ω, 100 k Ω  
0 S  
200 µA  
2 mA  
20 mA  
0 S  
0 S  
0 S  
0 S  
0 S  
0 S  
0 S  
300 nS  
0 S  
0 S  
30 nS  
300 nS  
0 S  
3 nS  
30 nS  
300 nS  
100 mA  
DC bias settling time  
When DC bias is set to ON, add the following value to the settling time:  
Table 68. DC bias settling time  
Bias  
Settling time  
1
2
Standard  
Option 001  
Capacitance of DUT × 100 × loge (2/1.8 m) + 3 m  
Capacitance of DUT × 100 × loge (40/1.8 m) + 3 m  
100 sec  
10 sec  
1 sec  
100 msec  
10 msec  
2.  
1.  
1 mF  
1 µF  
10 µF  
100 µF  
10 mF  
100 mF  
DUT capacitance  
Figure 12. DC bias settling time  
37 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Web Resources  
Visit our Web sites for additional product information and literature.  
E4980A Precision LCR Meter  
www.keysight.com/find/e4980a  
E4980AL Precision LCR Meter  
www.keysight.com/find/e4980al  
LCR Meters & Impedance Measurement Products  
www.keysight.com/find/impedance  
RF & MW test accessories  
www.keysight.com/find/mta  
38 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
For more information on Keysight  
Evolving  
Technologies’ products, applications or  
services, please contact your local Keysight  
office. The complete list is available at:  
www.keysight.com/find/contactus  
Our unique combination of hardware, software, support, and people can help  
you reach your next breakthrough. We are unlocking the future of technology.  
Americas  
Canada  
Brazil  
Mexico  
United States  
(877) 894 4414  
55 11 3351 7010  
001 800 254 2440  
(800) 829 4444  
Asia Pacific  
Australia  
China  
Hong Kong  
India  
Japan  
Korea  
Malaysia  
Singapore  
Taiwan  
1 800 629 485  
800 810 0189  
800 938 693  
1 800 11 2626  
0120 (421) 345  
080 769 0800  
1 800 888 848  
1 800 375 8100  
0800 047 866  
(65) 6375 8100  
From Hewlett-Packard to Agilent to Keysight  
myKeysight  
www.keysight.com/find/mykeysight  
A personalized view into the information most relevant to you.  
Other AP Countries  
Keysight Infoline  
Europe & Middle East  
Austria  
Belgium  
Finland  
www.keysight.com/find/Infoline  
Keysight’s insight to best in class information management. Free access to  
your Keysight equipment company reports and e-library.  
0800 001122  
0800 58580  
0800 523252  
0805 980333  
0800 6270999  
1800 832700  
1 809 343051  
800 599100  
+32 800 58580  
0800 0233200  
8800 5009286  
800 000154  
0200 882255  
0800 805353  
Opt. 1 (DE)  
France  
Germany  
Ireland  
Israel  
Italy  
Luxembourg  
Netherlands  
Russia  
Spain  
Sweden  
Keysight Services  
www.keysight.com/find/service  
Our deep offering in design, test, and measurement services deploys an  
industry-leading array of people, processes, and tools. The result? We help  
you implement new technologies and engineer improved processes that  
lower costs.  
KEYSIGHT  
SERVICES  
Three-Year Warranty  
www.keysight.com/find/ThreeYearWarranty  
Switzerland  
Keysight’s committed to superior product quality and lower total cost  
of ownership. Keysight is the only test and measurement company with  
three-year warranty standard on all instruments, worldwide. And, we provide  
a one-year warranty on many accessories, calibration devices, systems and  
custom products.  
Opt. 2 (FR)  
Opt. 3 (IT)  
0800 0260637  
United Kingdom  
For other unlisted countries:  
Keysight Assurance Plans  
www.keysight.com/find/contactus  
(BP-06-08-16)  
www.keysight.com/find/AssurancePlans  
Up to ten years of protection and no budgetary surprises to ensure your  
instruments are operating to specification, so you can rely on accurate  
measurements.  
DEKRA Certified  
Keysight Channel Partners  
www.keysight.com/find/channelpartners  
ISO9001 Quality Management System  
Get the best of both worlds: Keysight’s measurement expertise and product  
breadth, combined with channel partner convenience.  
www.keysight.com/go/quality  
Keysight Technologies, Inc.  
DEKRA Certified ISO 9001:2015  
Quality Management System  
www.keysight.com/find/e4980a  
This information is subject to change without notice.  
© Keysight Technologies, 2013 - 2014, 2016  
Published in USA, August 17, 2016  
5989-4435EN  
www.keysight.com  

相关型号:

E4980AU-007

Precision LCR Meter
KEYSIGHT

E4990A

Impedance Analyzer
KEYSIGHT

E4990A-001

Impedance Analyzer
KEYSIGHT

E4991A

RF Impedance/Material Analyzer
AGILENT

E4991A-001

RF Impedance/Material Analyzer
HP

E4991A-002

RF Impedance/Material Analyzer
AGILENT

E4991A-007

RF Impedance/Material Analyzer
AGILENT

E4991A-010

RF Impedance/Material Analyzer
AGILENT

E4991A-1D5

RF Impedance/Material Analyzer
AGILENT

E4991B

Impedance Analyzer
KEYSIGHT

E4991B-002

Impedance Analyzer
KEYSIGHT

E4991B-007

Impedance Analyzer
KEYSIGHT