RPS-120S-15 [MEANWELL]
120W 3"Ã2" Reliable Green Medical Power Supply;型号: | RPS-120S-15 |
厂家: | MEAN WELL ENTERPRISES CO., LTD. |
描述: | 120W 3"Ã2" Reliable Green Medical Power Supply |
文件: | 总4页 (文件大小:116K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
120W 3"×2" Reliable Green Medical Power Supply
RPS-120S series
EN 60601-1
IEC 60601-1
+ +
8
ANSI/AAMI ES60601-1
EN60601-1
IEC60601-1
■ Features
■ Applications
ꢀOral irrigator
ꢀ3"×2" compact size
ꢀ120W convention, 150W peak (10sec.)
ꢀHemodialysis machine
ꢀMedical monitors
ꢀSleep apnea devices
ꢀPumps machine
ꢀMedical safety approved (2 x MOPP) accroding to
ANSI/AAMI ES60601-1 and IEC/EN60601-1
ꢀSuitable for BF application with appropriate system consideration
ꢀEMI for both ClassⅠ& Class Ⅱ configuration
ꢀ-30~+85℃ wide range operating temperature
ꢀNo load power consumption<0.3W
ꢀExtremely low leakage current
ꢀProtections: Short circuit / Overload / Over voltage / Over temperature
ꢀOperating altitude up to 4000 meters (Note.6)
ꢀ3 years warranty
■ Description
RPS-120S is a 120W highly reliable green PCB type medical power supply with a high power density on
a 3 by 2 footprint. It accepts 80~264VAC input and offers various models with the output voltages between
12V and 48V. The working efficiency is up to 94% and the extremely low no load power consumption is down
below 0.3W. RPS-120S is able to be used for both ClassⅠ(with FG) & ClassⅡ(no FG) system design. The
extremely low leakage current is less than 150 A. In addition, it conforms to the international medical regulations
μ
(2*MOPP)
equipment.
and EMC EN55011, perfectly fitting all kinds of BF rated patient contact medical system
■ Model Encoding
RPS -
- 12
120S
Output voltage
Rated wattage
Series name
File Name:RPS-120S-SPEC 2019-09-02
120W 3"×2" Reliable Green Medical Power Supply
RPS-120S series
SPECIFICATION
MODEL
RPS-120S-12
12V
RPS-120S-15
15V
RPS-120S-24
24V
RPS-120S-27
27V
RPS-120S-48
48V
DC VOLTAGE
Peak(10 sec.)
Convection
11.8A
9.5A
6.25A
5.55A
3.125A
2.5A
CURRENT
9.5A
7.6A
5A
4.44A
Peak(10 sec.) 141.6W
114W
142.5W
114W
150W
149.8W
119.9W
150mVp-p
25.6 ~ 28.4V
1.0%
150W
RATED
POWER
120W
120W
Convection
RIPPLE & NOISE (max.) Note.2 100mVp-p
VOLTAGE ADJ. RANGE 11.4~12.6V
120mVp-p
14.3~15.8V
2.0%
150mVp-p
22.8~25.2V
1.0%
200mVp-p
45.6 ~50.4V
1.0%
OUTPUT
VOLTAGE TOLERANCE Note.3
LINE REGULATION
LOAD REGULATION
SETUP, RISE TIME
2.0%
0.5%
0.5%
0.5%
0.5%
0.5%
1.0%
1.0%
1.0%
1.0%
1.0%
600ms, 30ms/230VAC
600ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)
15ms/230VAC
15ms/115VAC at full load
113 ~ 370VDC
VOLTAGE RANGE Note.4 80 ~ 264VAC
FREQUENCY RANGE
POWER FACTOR
47 ~ 63Hz
PF>0.94/230VAC
91%
PF>0.98/115VAC at full load
EFFICIENCY (Typ.)
92%
93%
94%
93.5%
INPUT
2.3A/115VAC
1.1A/230VAC
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
LEAKAGE CURRENT(max.) Note.5
COLD START 30A/115VAC
60A/230VAC
Earth leakage current < 150 A/264VAC , touch current < 80 A/264VAC
μ μ
130~160% rated output power
OVERLOAD
Protection type : Hiccup mode, recovers automatically after fault condition is removed
16.5 ~ 19.5V 26.4 ~ 31.2V 29.7 ~ 35V
Protection type : Shut down o/p voltage, re-power on to recover
52.8 ~ 62.4V
13.2 ~ 15.6V
PROTECTION
OVER VOLTAGE
OVER TEMPERATURE
WORKING TEMP.
Protection type : Shut down o/p voltage, recovers automatically after temperature goes down
-30 ~ +85℃ (Refer to "Derating Curve")
20 ~ 90% RH non-condensing
WORKING HUMIDITY
STORAGE TEMP.
-40 ~ +85℃
ENVIRONMENT
TEMP. COEFFICIENT
VIBRATION
0.03%/℃ (0 ~ 50℃)
10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes
4000 meters
OPERATING ALTITUDE Note.6
IEC60601-1, TUV EN60601-1, EAC TP TC 004, UL ANSI / AAMI ES60601-1 (3.1 version),
SAFETY STANDARDS
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved; Design refer to EN60335-1 (By request)
Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP, Secondary-Earth:1xMOPP
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
ISOLATION RESISTANCE
WITHSTAND VOLTAGE
ISOLATION RESISTANCE I/P-O/P, I/P-FG, O/P-FG: 100M Ohms / 500VDC / 25℃/ 70% RH
Parameter
Standard
Test Level / Note
EN55011 (CISPR11)
EN55011 (CISPR11)
EN61000-3-2
Class B
Conducted emission
Radiated emission
Harmonic current
Voltage flicker
EN60601-1-2
Parameter
EMC EMISSION
ClassⅠ: Class B , Class Ⅱ: Class A
Class A
SAFETY &
EN61000-3-3
-----
EMC
(Note 7)
Standard
Test Level / Note
ESD
EN61000-4-2
Level 4, 15KV air ; Level 4, 8KV contact
Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
Level 3, 2KV
RF field susceptibility
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EFT bursts
EMC IMMUNITY
Surge susceptibility
Conducted susceptibility
Magnetic field immunity
Level 4, 4KV/Line-FG; 2KV/Line-Line
Level 3, 10V
Level 4, 30A/m
95% dip 1 periods, 30% dip 25 periods,
95% interruptions 250 periods
Voltage dip, interruption
EN61000-4-11
MTBF
468Khrs min. MIL-HDBK-217F (25℃)
76.2*50.8*28mm or 3" * 2" *1.1" inch
0.13Kg; 100pcs/14Kg/1.13CUFT
OTHERS DIMENSION (L*W*H)
PACKING
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
μ
μ
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1 f & 47 f parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. Derating may be needed under low input voltages. Please check the derating curve for more details.
5. Touch current was measured from primary input to DC output.
NOTE
6. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
7. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by
mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies.
(as available on http://www.meanwell.com)
File Name:RPS-120S-SPEC 2019-09-02
120W 3"×2" Reliable Green Medical Power Supply
RPS-120S series
Block Diagram
fosc : 85KHz
RECTIFIERS
RECTIFIERS
POWER
EMI
FILTER
+V
-V
I/P
FG
&
PFC
&
SWITCHING
FILTER
DETECTION
CIRCUIT
O.T.P.
O.L.P.
CONTROL
O.V.P.
Derating Curve
Convection
100
80
70
60
50
30
(HORIZONTAL)
70 85
-30
0
10
20
30
40
50
AMBIENT TEMPERATURE (℃)
Output Derating VS Input Voltage
(for 24V/27V/48V)
120W
114W
(for 12V/15V)
100W
80W
80 90 110 120 130 140 160 180 200 220 240 264
INPUT VOLTAGE (VAC) 60Hz
File Name:RPS-120S-SPEC 2019-09-02
120W 3"×2" Reliable Green Medical Power Supply
RPS-120S series
Mechanical Specification
Unit:mm
SVR1
C100
HS1
1
2
3
4
T1
FS1
CN100
C5
3
2
1
C105
C119
C1
T3.15A
/250V
FS2
69.9
3.175
76.2
AC Input Connector (CN1) : JST B3P-VH or equivalent
DC Output Connector (CN2) : JST B4P-VH or equivalent
Pin No. Assignment
Mating Housing
Terminal
Pin No. Assignment
Mating Housing
Terminal
1
2
3
AC/L
No Pin
AC/N
1,2
3,4
+V
-V
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
1.HS1 must have safety isolation distance
with system case.
※Note :
1.RPS-120S model delivers EMI Class B for both conducted emission and radiated emission for the power supply,
when configured into Class Ⅰ(with FG) system.
2.RPS-120S model delivers EMI Class B conducted emission and Class A radiated emission with King Core K5B RC (12*15*7)
in output cable for the power supply when configured into Class Ⅱ (no FG) system.
Installation Manual
Please refer to : http://www.meanwell.com/manual.html
File Name:RPS-120S-SPEC 2019-09-02