MSCU01-24S15M [MINMAX]
DC/DC CONVERTER 1W;型号: | MSCU01-24S15M |
厂家: | MINMAX TECHNOLOGY CO., LTD. |
描述: | DC/DC CONVERTER 1W |
文件: | 总5页 (文件大小:711K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
MSCU01M SERIES
DC/DC CONVERTER 1W, Reinforced Insulation, Medical Safety
FEATURES
►Industrial Standard SMD Package
►I/O Isolation 4000VAC with Reinforced Insulation, rated for 250Vrms
Working Voltage
►Low Leakage Current < 2μA
►Operating Ambient Temp. Range -40℃ to 95℃
►Short Circuit Protection
►Medical EMC Standard meets 4th Edition of EMI EN55011 and
EMS EN60601-1-2
►Medical Safety meets 2xMOPP per 3rd Edition of IEC/EN 60601-1 &
ANSI/AAMI ES60601-1 with CE Marking
PRODUCT OVERVIEW
The MINMAX MSCU01M series is a new range of medical approved 1W isolated dc-dc converter within enclosed SMD-14 package which specifically
design for medical applications. There are 15 models available for input voltage of 5, 12, 24VDC and 5, 12, 15, ±12, ±15VDC output. The I/O isolation is
specified for 4000VAC with reinforced insulation, which rated for 250Vrms working voltage. Further features include short circuit protection, low leakage
current 2μA max. and operating ambient temp. range by -40℃ to 95℃ without derating. MSCU01M series conform to 4th edition medical EMC standard,
medical safety approval meets 2xMOPP (Means Of Patient Protection) per 3rd edition of IEC/EN 60601-1 & ANSI/AAMI ES60601-1.
The MSCU01M series offer a economical solution for demanding application in medcial instrument requesting a certified supplementary and reinforced
insulation system to comply with latest medcial safety approval for 2xMOPP requirement.
Model Selection Guide
Model
Number
Input
Voltage
(Range)
Output
Voltage
Output Current
Input Current
Max. capacitive
Load
Efficiency
(typ.)
Max.
Min.
mA
@Max. Load
@No Load
mA(typ.)
@Max. Load
VDC
VDC
5
mA
200
84
mA(typ.)
263
252
246
252
236
110
106
106
106
103
55
μF
%
MSCU01-05S05M
MSCU01-05S12M
MSCU01-05S15M
MSCU01-05D12M
MSCU01-05D15M
MSCU01-12S05M
MSCU01-12S12M
MSCU01-12S15M
MSCU01-12D12M
MSCU01-12D15M
MSCU01-24S05M
MSCU01-24S12M
MSCU01-24S15M
MSCU01-24D12M
MSCU01-24D15M
4
76
80
83
80
84
76
79
80
79
80
76
80
80
80
80
12
1.68
1.36
±0.84
±0.66
4
220
5
15
68
50
35
20
(4.5 ~ 5.5)
±12
±15
5
±42
±33
200
84
100#
220
12
1.68
1.36
±0.84
±0.66
4
12
15
68
(10.8 ~ 13.2)
±12
±15
5
±42
±33
200
84
100#
220
12
1.68
1.36
±0.84
±0.66
53
24
15
68
53
(21.6 ~ 26.4)
±12
±15
±42
±33
53
100#
52
# For each output
Input Specifications
Parameter
Model
Min.
Typ.
5
Max.
Unit
5V Input Models
12V Input Models
24V Input Models
5V Input Models
12V Input Models
24V Input Models
All Models
4.5
10.8
21.6
-0.7
-0.7
-0.7
5.5
13.2
26.4
9
Input Voltage Range
12
24
---
VDC
Input Surge Voltage (1 sec. max.)
Input Filter
---
18
---
30
Internal Capacitor
E-mail:sales@minmax.com.tw Tel:886-6-2923150
2017/12/18 REV:2 Page 1 of 5
DC/DC CONVERTER 1W, Reinforced Insulation, Medical Safety
Output Specifications
Parameter
Conditions
Min.
---
Typ.
±1.0
±0.1
±1.2
---
Max.
±3.0
±1.0
±1.5
±10
Unit
%Vnom.
%
Output Voltage Setting Accuracy
Output Voltage Balance
Line Regulation
Dual Output, Balanced Loads
For Vin Change of 1%
Io=10% to 100%
---
---
%
Load Regulation
---
---
---
%
Ripple & Noise
0-20 MHz Bandwidth
---
100
mV P-P
%/℃
Temperature Coefficient
Short Circuit Protection
±0.01
±0.02
Continuous, Automatic Recovery
Isolation, Safety Standards
Parameter
Conditions
Min.
Typ.
---
Max.
---
Unit
60 Seconds
I/O Isolation Voltage
4000
VACrms
Reinforced insulation, rated for 250Vrms working voltage
Leakage Current
240VAC, 60Hz
500 VDC
---
10
---
---
---
20
2
μA
GΩ
pF
I/O Isolation Resistance
I/O Isolation Capacitance
---
---
100KHz, 1V
ANSI/AAMI ES60601-1, CAN/CSA-C22.2 No. 60601-1
IEC/EN 60601-1 3rd Edition 2xMOPP
ANSI/AAMI ES60601-1 2xMOPP recognition (UL certificate), IEC/EN 60601-1 3rd Edition (CB-report)
Safety Standards
Safety Approvals
General Specifications
Parameter
Conditions
Min.
---
Typ.
55
Max.
---
Unit
kHz
Switching Frequency
MTBF (calculated)
MIL-HDBK-217F@25℃, Ground Benign
4,771,507
---
---
Hours
Moisture Sensitivity Level (MSL)
IPC/JEDEC J-STD-020D.1
Level 2
Environmental Specifications
Parameter
Conditions
Min.
-40
Max.
+95
Unit
Operating Ambient Temperature Range
(See Power Derating Curve)
Case Temperature
Natural Convection
℃
---
-50
---
+105
+125
95
℃
℃
Storage Temperature Range
Humidity (non condensing)
Cooling
% rel. H
Natural Convection
IPC/JEDEC J-STD-020D.1
Lead-free Reflow Solder Process
EMC Specifications
Parameter
Standards & Level
Performance
Class A(5)
EMI
Conduction
EN 60601-1-2 4th
ESD
EN 55011, FCC part 15
EN 61000-4-2 Air ± 15kV , Contact ± 8kV
EN 61000-4-3 10V/m
A
A
A
A
A
A
Radiated immunity
Fast transient (6)
Surge (6)
EMS
EN 61000-4-4 ±2kV
EN 61000-4-5 ±1kV
Conducted immunity
PFMF
EN 61000-4-6 10Vrms
EN 61000-4-8 30A/m
E-mail:sales@minmax.com.tw Tel:886-6-2923150
2017/12/18 REV:2 Page 2 of 5
DC/DC CONVERTER 1W, Reinforced Insulation, Medical Safety
Power Derating Curve
Notes
1
Specifications typical at Ta=+25℃, resistive load, nominal input voltage and rated output current unless otherwise noted.
2
These power converters require a minimum output loading to maintain specified regulation, operation under no-load conditions will not damage these modules;
however they may not meet all specifications listed.
3
4
5
6
7
8
We recommend to protect the converter by a slow blow fuse in the input supply line.
Other input and output voltage may be available, please contact factory.
To meet EN 55011 Class A an external filter, please contact MINMAX.
To meet EN 61000-4-4 & EN61000-4-5 an external capacitor across the input pins is required, please contact MINMAX.
That “natural convection” is about 20LFM but is not equal to still air (0 LFM).
Specifications are subject to change without notice.
Package Specifications
Mechanical Dimensions
Connecting Pin Patterns
18.9 [0.74]
12.7 [0.5]
1.83
2.54
[0.07]
[0.10]
15.24 [0.60]
9
6
8
7
14
1
Top View
1.00 [0.04]
S
SEATING
PLANE
1.6 [0.06]
2.54 [0.10]
1.1 [0.04]
15.0 [0.6]
0.15
S
4.0 [0.16]
1.5 [
0.06]
►All dimensions in mm (inches)
►Tolerance: X.X±0.5 (X.XX±0.02)
X.XX±0.25 ( X.XXX±0.01)
Bottom View
►Pins ±0.05 (±0.002)
Pin Connections
Physical Characteristics
Case Size
Pin
1
Single Output
-Vin
Dual Output
-Vin
:
:
:
:
18.9x13.7x10.2 mm (0.74x0.54x0.40 inches)
6
NC
Common
-Vout
Case Material
Pin Material
Weight
Non-Conductive Black Plastic (flammability to UL 94V-0 rated)
7
NC
8
+Vout
-Vout
+Vin
+Vout
Phosphor Bronze
4.1g
9
Common
+Vin
14
NC: No Connection
E-mail:sales@minmax.com.tw Tel:886-6-2923150
2017/12/18 REV:2 Page 3 of 5
DC/DC CONVERTER 1W, Reinforced Insulation, Medical Safety
Order Code Table
Standard
For water-washable process
MSCU01-05S05M-W
MSCU01-05S12M-W
MSCU01-05S15M-W
MSCU01-05D12M-W
MSCU01-05D15M-W
MSCU01-12S05M-W
MSCU01-12S12M-W
MSCU01-12S15M-W
MSCU01-12D12M-W
MSCU01-12D15M-W
MSCU01-24S05M-W
MSCU01-24S12M-W
MSCU01-24S15M-W
MSCU01-24D12M-W
MSCU01-24D15M-W
MSCU01-05S05M
MSCU01-05S12M
MSCU01-05S15M
MSCU01-05D12M
MSCU01-05D15M
MSCU01-12S05M
MSCU01-12S12M
MSCU01-12S15M
MSCU01-12D12M
MSCU01-12D15M
MSCU01-24S05M
MSCU01-24S12M
MSCU01-24S15M
MSCU01-24D12M
MSCU01-24D15M
E-mail:sales@minmax.com.tw Tel:886-6-2923150
2017/12/18 REV:2 Page 4 of 5
DC/DC CONVERTER 1W, Reinforced Insulation, Medical Safety
Test Setup
Peak-to-Peak Output Noise Measurement Test
Refer to the output specifications or add 4.7μF capacitor if the output specifications undefine Cout.. Scope measurement should be made by using a BNC socket,
measurement bandwidth is 0-20 MHz. Position the load between 50 mm and 75 mm from the DC/DC Converter.
Copper Strip
Cout
+Vin
+Out
-Out
Copper Strip
+Vin
+Out
Com.
-Out
Scope
Scope
Single Output
DC / DC
Converter
Dual Output
DC / DC
Converter
Resistive
Load
Resistive
Load
Cout
Scope
Copper Strip
Cout
-Vin
Copper Strip
-Vin
Copper Strip
Technical Notes
Maximum Capacitive Load
The MSCU01M series has limitation of maximum connected capacitance at the output. The power module may be operated in current limiting mode during start-up,
affecting the ramp-up and the startup time. For optimum performance we recommend 100μF maximum capacitive load for dual outputs and 220μF capacitive load for
single outputs. The maximum capacitance can be found in the data sheet.
Input Source Impedance
The power module should be connected to a low ac-impedance input source. Highly inductive source impedances can affect the stability of the power module. In
applications where power is supplied over long lines and output loading is high, it may be necessary to use a capacitor at the input to ensure startup.
Capacitor mounted close to the power module helps ensure stability of the unit, it is recommended to use a good quality low Equivalent Series Resistance (ESR <
1.0Ω at 100 KHz) capacitor of a 2.2μF for the 5V input devices, a 1.0μF for the 12V input devices and a 0.47μF for the 24V input devices.
+
+Vin
+Out
+
DC / DC
Converter
DC Power
Source
Load
Cin
-
-Vin
-Out
Output Ripple Reduction
A good quality low ESR capacitor placed as close as practicable across the load will give the best ripple and noise performance. To reduce output ripple, it is
recommended to use 3.3μF capacitors at the output.
+
+Vin
-Vin
+Out
+
+Vin
-Vin
+Out
Single Output
DC / DC
Converter
Load
Load
Dual Output
DC Power
Source
Cout
Cout
DC Power
Source
DC / DC Com.
Converter
Cout
Load
-
-Out
-
-Out
Thermal Considerations
Many conditions affect the thermal performance of the power module, such as orientation, airflow over the module and board spacing. To avoid exceeding the
maximum temperature rating of the components inside the power module, the case temperature must be kept below 105℃. The derating curves are determined from
measurements obtained in a test setup.
Position of air velocity
probe and thermocouple
Air Flow
50mm / 2in
15mm / 0.6in
DUT
Minmax Technology Co., Ltd.
18, Sin Sin Road, An-Ping Industrial District, Tainan 702, Taiwan
Tel: 886-6-2923150 Fax: 886-6-2923149 E-mail: sales@minmax.com.tw
2017/12/18 REV:2
Page 5 of 5
相关型号:
SI9130DB
5- and 3.3-V Step-Down Synchronous ConvertersWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9135LG-T1
SMBus Multi-Output Power-Supply ControllerWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9135LG-T1-E3
SMBus Multi-Output Power-Supply ControllerWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9135_11
SMBus Multi-Output Power-Supply ControllerWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9136_11
Multi-Output Power-Supply ControllerWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9130CG-T1-E3
Pin-Programmable Dual Controller - Portable PCsWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9130LG-T1-E3
Pin-Programmable Dual Controller - Portable PCsWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9130_11
Pin-Programmable Dual Controller - Portable PCsWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9137
Multi-Output, Sequence Selectable Power-Supply Controller for Mobile ApplicationsWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9137DB
Multi-Output, Sequence Selectable Power-Supply Controller for Mobile ApplicationsWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9137LG
Multi-Output, Sequence Selectable Power-Supply Controller for Mobile ApplicationsWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
SI9122E
500-kHz Half-Bridge DC/DC Controller with Integrated Secondary Synchronous Rectification DriversWarning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
VISHAY
©2020 ICPDF网 联系我们和版权申明