3EH1 [TE]

Power Inlet Line Filter for Medical Equipment; 医疗设备电源输入线路滤波器
3EH1
型号: 3EH1
厂家: TE CONNECTIVITY    TE CONNECTIVITY
描述:

Power Inlet Line Filter for Medical Equipment
医疗设备电源输入线路滤波器

连接器 电源连接器 医疗
文件: 总3页 (文件大小:265K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
Catalog: 1654001  
Issue Date: 06.2011  
Corcom Product Guide  
Power Inlet Line Filter for Medical Equipment  
H Series  
UL Recognized  
CSA Certified  
VDE Approved*  
EH8  
EH9  
EH4  
EH1  
EH5  
H Series  
Specifications  
• Minimal leakage current suitable for medical  
equipment  
Maximum leakage current each Line to Ground:  
@ 120 VAC 60 Hz:  
2 A  
5 A  
@250 VAC 50 Hz:  
• Two element circuit provides basic EMI  
attenuation above 1 MHz  
Hipot rating (one minute):  
Line to Ground:  
Line to Line:  
2250 VDC  
1450 VDC  
• Available with an internal ground circuit inductor  
(C suffix versions) to isolate equipment chassis  
from power line ground at radio frequencies  
3
Rated Voltage (max.):  
250 VAC  
50/60 Hz  
3 to 15A*  
• Flanged mounting the same as the EC, ED and EF  
Series  
Operating Frequency:  
Rated Current:  
• Capacitive output (see EAH, EBH and EJH Series  
for capacitive input)  
Operating Ambient Temperature Range  
(at rated current I ):  
-10°C to +40°C  
In an ambient temperature (T ) higher than +40°C  
r
Ordering Information  
3 E H 1 C  
a
the maximum operating current (I ) is calculated as  
o
follows: I = I (85-Ta)/45  
o
r
Ground Circuit Inductor  
Omit if none  
Electrical Schematic  
LINE  
LOAD  
Output Styles  
1
Chassis Mount Style  
1 – .250 [6.3] spade terminals  
3 – Wire leads  
Power Inlet Style *  
4 – .250 [6.3] spade terminals  
Available Part Numbers  
(back)  
5 – Wire leads (bottom)  
3EH1  
3EH3  
6EH1  
6EH3  
6EH4  
6EH5  
6EH8  
8 – Wire leads (back)  
9 – .250 [6.3] spade terminals  
(bottom)  
6EH9  
10EH1  
10EH3  
10EH4  
15EH4  
H Series  
Ground Circuit Inductor Versions  
10EH4C  
Leakage current designation  
E – Low Leakage (<0.5 mA)  
Current Rating  
3, 6, 10 or 15A  
*15A versions are tested by Underwriters Laboratories  
to US and Canadian requirements and are  
VDE approved at 10A, 250VAC  
*IEC 60320-1 C14 inlet mates with C13 connector  
169  
Dimensions are in inches and millimeters unless otherwise specified. Values in italics  
are metric equivalents. Dimensions are shown for reference purposes only.  
Specifications subject to change.  
For email, phone or live chat, please go to  
te.com/help  
corcom.com  
Catalog: 1654001  
Issue Date: 06.2011  
Corcom Product Guide  
Power Inlet Line Filter for Medical Equipment (continued)  
H Series  
Case Styles  
H5  
H1 (Chassis Mount)  
C
D
D
E B  
E
A
B
F
C
A
  
  
F
  
  
     
   
   
  
  
  
   
H3 (Chassis Mount)  
H8  
F
Blue  
Green/Yellow  
E
D
B
D
E
B
Brown  
F
A
A
C
  
  
   
  
   
C
  
   
   
H4 & H4C  
  
   
C
D
E
A
B
Recommended Panel Cutouts  
  
   
   
  
  
  
  
Front Mount  
Back Mount  
.820  
.900  
22.9  
20.8  
H9  
.140  
3.56  
.140  
3.56  
1.200  
30.48  
1.200  
30.48  
1.575  
40.01  
1.575  
40.01  
C
.062  
1.58  
.234  
5.94  
R.  
Typ.  
R.  
Typ.  
.65  
16.5  
Tolerances  
.005 [0.13] unless otherwise noted  
Max.  
D
E
  
  
  
  
A
B
  
   
  
   
  
  
  
170  
Dimensions are in inches and millimeters unless otherwise specified. Values in italics  
are metric equivalents. Dimensions are shown for reference purposes only.  
Specifications subject to change.  
For email, phone or live chat, please go to  
te.com/help  
corcom.com  
Catalog: 1654001  
Issue Date: 06.2011  
Corcom Product Guide  
Power Inlet Line Filter for Medical Equipment (continued)  
H Series  
Case Dimensions  
Performance Data  
A
  
B
  
C
  
D
E
  
F
  
Typical Insertion Loss  
Measured in closed 50 Ohm system  
Part No.  
  
  
      
3EH  
6EH  
H1  
db  
db  
      
100  
100  
90  
80  
70  
60  
50  
40  
30  
20  
10  
90  
80  
70  
60  
50  
40  
30  
20  
10  
  
     
H3  
      
      
      
      
      
      
      
6EH4  
10EH4,  
10EH4C  
0
0
0.01  
0.1  
1
10  
30  
0.1  
1
10  
30  
Frequency in MHz  
Frequency in MHz  
15EH4  
H5  
10EH  
15EH  
db  
db  
       
       
       
100  
100  
3
90  
80  
70  
60  
50  
40  
30  
20  
10  
90  
80  
70  
60  
50  
40  
30  
20  
10  
H8  
      
      
      
  
H9  
0
0
0.1  
1
10  
30  
0.01  
0.1  
1
10  
30  
Frequency in MHz  
Frequency in MHz  
Common Mode / Asymmetrical (L-G)  
Differential Mode / Symmetrical (L-L)  
Accessories  
GA400: NEMA 5-15P to IEC 60320-1 C-13 line cord  
Minimum Insertion Loss  
Measured in closed 50 Ohm system  
Common Mode / Asymmetrical (Line to Ground)  
   
  
    
  
  
  
  
FA601  
: Insulating Shroud  
.413  
10.5  
.271  
6.9  
  
  
  
  
  
  
  
  
Dia.  
Dia.  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
.118  
3.0  
1.39  
35.4  
1.17  
29.7  
.767  
19.5  
171  
Dimensions are in inches and millimeters unless otherwise specified. Values in italics  
are metric equivalents. Dimensions are shown for reference purposes only.  
Specifications subject to change.  
For email, phone or live chat, please go to  
te.com/help  
corcom.com  

相关型号:

3EH3

Power Inlet Line Filter for Medical Equipment
TE

3EHG1-2

Smallest Filtered Power Entry Modules with Metric Fuseholders
TE

3EHG1-2

Smallest Power Entry Module with Metric Fuse Holders
MACOM

3EHG8-2

Smallest Power Entry Module with Metric Fuse Holders
TE

3EHGS1-2

Smallest Filtered Power Entry Modules with Metric Fuseholders
TE

3EHQ1

Highest Performance RFI Filters for Medical Equipment
TE

3EHQ1_11

Highest Performance RFI Filters for Medical Equipment
TE

3EHQ3

Highest Performance RFI Filters for Medical Equipment
TE

3EHQ8

Highest Performance RFI Filters for Medical Equipment
TE

3EHQ8M

Highest Performance RFI Filters for Medical Equipment
TE

3EHT1

High Performance RFI Power Line Filters for Medical Equipment
TE

3EHT1

High Performance RFI Power Line Filters for Medical Equipment
MACOM