TPMBU150-108 [TOTAL-POWER]
150W Open Frame Type Medical Power Supplies;型号: | TPMBU150-108 |
厂家: | TOTAL POWER INTERNATIONAL |
描述: | 150W Open Frame Type Medical Power Supplies |
文件: | 总2页 (文件大小:327K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
Total Power International, Inc.
150W Open Frame Type
Medical Power Supplies
TPMBU150 SERIES
Features:
Wide Operating Voltage 90 to 260 VAC,47 to 63 Hz
Internal EMI filter
Single Output
Active Power Factor Correction
Size: 3"x5"x1.44”
Class I
3 year warranty
Electrical Characteristics:
Safety Approvals Input Voltage Range
100~240VAC
Vin
Application:
Operate Voltage Range
90~260VAC
47~63Hz
- Medical Touch Panel PC
- Patient Monitoring System
- Ultrasound System
Input Frequency
fin
Power Factor Correction
Output Power Range
Io=Full load, Vin=240VAC
0.95~1
PF
Po
Vo
Io
See rating chart
See rating chart
See rating chart
Output Voltage Range
Output Current Range
Input Current (Low Line)
Input Current (High Line)
Io=Full load, Vin=100VAC
Io=Full load, Vin=240VAC
Io=Full load, 25°C, Cool start, Vin=115VAC
Io=Full load, 25°C, Cool start, Vin=230VAC
Io=Full Load, Vin=230VAC
Io=Full Load
2.0A
Iil
1.0A
Iih
Safety Approvals:
54A (max.)
108A (max.)
Low Line Inrush Current
High Line Inrush Current
Efficiency
Ir
82~87%
Eff
RoHS
COMPLIANT
EN 60601-1
IEC 60601-1
2002/95/EC
Line Regulation
0.5~1%
REG-i
REG-o
OVP
Load Regulation
Vin=230VAC
3~5%
Over Voltage Protection
112~132%
110~150%
4mS (max.)
16mS (min.)
2S (max.)
1% (max.)
0.15mA (max.)
OCP Over Current Protection
Time of Transient Response
Hold-Up Time
Io=Full Load to Half Load, Vin=100VAC
Io=Full Load, Vin=110VAC
Io=Full Load, Vin=100VAC
Full Load, Vin=90VAC
Ttr
Th
Ts
Start Up Time
Vp-p Ripple & Noise(Peak to Peak)
Safety Ground Leakage Current
Ilk
Vin=240VAC/60Hz
All output
Temperature Coefficient
TC
0.04%/°C
Operating Altitude
Up to 3000m
See rating chart
5656VDC (min.)
OA
Pno
Vps
No-Load Power Consumption
Dielectric Withstanding Voltage
No load, Vin=230VAC
Primary to secondary
Dielectric Withstanding Voltage
for Primary to PE
Vpg
Ris
Primary to PE
2121VDC (min.)
Test Voltage=500VDC
(min.)
50M
Isolation Resistance
Environmental
Operating Temperature
See derating curve
-40~85°C
To
Ts
Storage Temperature
Operating Humidity
Storage Humidity
0~95%
Ho
0~95%
Hr
Operating Temperature at 25°C, Calculated per MIL-HDBK-217F
0.1M Hrs (min.)
MTBF
Pd
Derate linearly from 100% load at 50°C to 50% load at 70°C
Total Power International, Inc
Toll Free: 877-646-0900
www.total-power.com
Total Power International, Inc.
150W Open Frame Type
Medical Power Supplies
TPMBU150 SERIES
Output Voltage And Current Rating Chart ( Single Output ) :
Total
Regulation
Max.
Output Power
Output Voltage
Output Current
Pno (max.)
Model Number
12 VDC
24 VDC
12.50 A
6.25 A
5%
3%
150W
150W
1.5W
1.5W
TPMBU150-105
TPMBU150-108
PIN CHART
Mechanical Specifications:
PIN
1
2
3
4
5
6
7
8
MODEL
TPMBU150-1XX
Vout Vout Vout Vout RTN RTN RTN RTN
Note:
1. Dimensions are shown in inches or mm.
2. Weight: 420gs approx.
3. Input connector mates with Molex housing
09-50-3051and Molex 2478 series crimp terminal.
4. Output connector mates with Molex housing
09-50-3081 and Molex 2478 series crimp terminal
LED
1
8
P2
L
N
G
P1
Derating Curve :
120%
1. Operating Temperature: 0 to 70°C
2. Derate linearly from 100% load at 50°C to 50%
load at 70°C
100%
80%
60%
40%
20%
0%
0
10
20
30
40
50
60
70
80
Temperature (deg c)
Total Power International, Inc
Toll Free: 877-646-0900
www.total-power.com