THM10WI [TRACOPOWER]

DC/DC Converter;
THM10WI
型号: THM10WI
厂家: TRACO ELECTRONIC AG    TRACO ELECTRONIC AG
描述:

DC/DC Converter

文件: 总4页 (文件大小:313K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
DC/DC Converter  
THM 10WI Series, 10 Watt  
Ultra wide 4:1 input voltage 10 W DC/DC  
converter in a compact DIP-24 plastic case  
I/O isolation 5000 VACrms rated for  
250 VACrms working voltage  
Certification according to IEC/EN/ES  
60601-1 3rd edition for 2xMOPP  
Risk management process according to  
ISO 14971 including risk management file  
Acceptance criteria for electronic  
CB  
Scheme  
assemblies according to IPC-A-610 Level 3  
ES 60601-1  
IEC 606010-1  
Low leakage current < 2µA  
The THM-10WI series is a range of medical 10 Watt DC/DC converters in DIP-24  
plastic package and with ultra-wide 4:1 input voltage range. They provide a rein-  
forced isolation system for 5000 VACrms isolation and a very low leakage current  
of less than 2 µA. The units are approved to IEC/EN/ES 60601-1 3rd edition for  
2 x MOPP (Means Of Patient Protection) and come along with an ISO 14971 risk  
management file. Design and production conform to the quality management  
system ISO 13485. With a high efficien¬cy of up to 87% and highest grade com-  
ponents the converters can reliably operate in an ambient temperature range  
of –40°C up to +90°C. They constitute a reliable solution not only for medical  
equipment but also for demanding ranges of application such as transportation,  
control & measurement or IGBT drivers.  
Eꢀtended operating temperature range  
–40°C to 90°C.  
EMC compliance to IEC 60601-1-2 4th  
edition and EN55032 class A  
Operating up to 5000m altitude  
5 year product warranty  
Models  
Order code  
Input voltage range  
Output voltage  
3.3 VDC  
5.0 VDC  
12 VDC  
15 VDC  
24 VDC  
5.0 VDC  
12 VDC  
15 VDC  
3.3 VDC  
5.0 VDC  
12 VDC  
15 VDC  
24 VDC  
5.0 VDC  
12 VDC  
15 VDC  
3.3 VDC  
5.0 VDC  
12 VDC  
15 VDC  
24 VDC  
5.0 VDC  
12 VDC  
15 VDC  
Output current maꢀ.  
2500 mA  
2000 mA  
830 mA  
Efficiency typ.  
80.0 %  
84.0 %  
86.5 %  
87.0 %  
85.5 %  
83.0 %  
85.5 %  
86.5 %  
83.0 %  
86.5 %  
89.0 %  
89.0 %  
89.0 %  
85.0 %  
89.0 %  
88.0 %  
82.5 %  
86.5 %  
89.0 %  
89.0 %  
88.5 %  
85.0 %  
88.0 %  
88.0 %  
THM 10-0510WI  
THM 10-0511WI  
THM 10-0512WI  
THM 10-0513WI  
THM 10-0515WI  
THM 10-0521WI  
THM 10-0522WI  
THM 10-0523WI  
THM 10-2410WI  
THM 10-2411WI  
THM 10-2412WI  
THM 10-2413WI  
THM 10-2415WI  
THM 10-2421WI  
THM 10-2422WI  
THM 10-2423WI  
THM 10-4810WI  
THM 10-4811WI  
THM 10-4812WI  
THM 10-4813WI  
THM 10-4815WI  
THM 10-4821WI  
THM 10-4822WI  
THM 10-4823WI  
670 mA  
416 mA  
1000 mA  
416 mA  
333 mA  
2500 mA  
2000 mA  
830 mA  
670 mA  
416 mA  
1000 mA  
416 mA  
333 mA  
2500 mA  
2000 mA  
830 mA  
670 mA  
416 mA  
1000 mA  
416 mA  
333 mA  
4.5 – 9 VDC  
(5 VDC nominal)  
9. 0 – 36 VDC  
(24 VDC nominal)  
18 – 75 VDC  
(48 VDC nominal)  
http://www.tracopower.com  
Page 1 of 4  
THM 10WI Series, 10 Watt  
Input Specifications  
Input current no load  
5 Vin models: 20 mA typ.  
24 Vin models: 6 mA typ.  
48 Vin models: 4 mA typ.  
Surge voltage (3 sec. max.)  
5 Vin models: 16 V maꢀ.  
24 Vin models: 50 V maꢀ.  
48 Vin models: 100 V maꢀ.  
Start-up voltage  
5 Vin models: 4. 5 VDC (or lower)  
24 Vin models: 9 VDC (or lower)  
48 Vin models: 18 VDC (or lower)  
Startup time  
30 ms  
Under voltage shut down  
5 Vin models: 4 VDC typ.  
24 Vin models: 8 VDC typ.  
48 Vin models: 16 VDC typ.  
Conducted noise  
EMC immunity  
– Conducted & Radiated input surpression  
EN 55011 limits to IEC 60601-1-2 4th editon  
EN 55032 class A (internal filter)  
– Generic for Medical equipment  
– ESD (electrostatic discharge)  
IEC/EN 60601-1-2 4th edition  
EN 61000-4-2, air 8 kV, contact 6 kV,  
perf. criteria A  
– Radiated immunity  
– Fast transient / surge  
(with external input capacitor / diode)  
EN 61000-4-3, 10 V/m, perf. criteria A  
EN 61000-4-4, 2 kV, perf. criteria A  
EN 61000-4-5, 2 kV perf. criteria A  
5 Vin models: Nippon chemi-con KY 1000 µF/ 25 V and  
reverse diode (Vishay V10P45) in parallel  
24 Vin models: Nippon chemi-con KY 470 µF/ 50 V  
48 Vin models: Nippon chemi-con KY 330 µF/ 100 V  
EN 61000-4-6, 10 Vrms, perf. criteria A  
EN 61000-4-8  
– Conducted immunity  
– Magnetic field immunity  
100 A/m, continuous, perf. criteria A  
1000 A/m, 1 sec., perf. criteria A  
Eꢀternal input fuse required (recommended values, slow blow type)  
5 Vin models:  
24 Vin models:  
48 Vin models:  
10 A  
5 A  
2.5 A  
Output Specifications  
Voltage set accuracy  
1% maꢀ.  
Regulation  
– Input variation  
single output: 0.2% maꢀ.  
dual output: 0.5% maꢀ.  
single output: 0.2% maꢀ.  
dual output: 1.0% maꢀ.  
– Load variation 0 – 100 %  
– Cross regulation  
dual output: 5.0% maꢀ. (asymmetrical load 25/100%)  
Minimum load  
not required  
Ripple and noise (20 MHz Bandwidth)  
3.3 & 5.0 VDC models: 30 mVp-p typ. with cap. 10µF/25V X7R MLCC  
12 & 15 VDC models: 40 mVp-p typ. with cap. 10µF/25V X7R MLCC  
24 VDC models: 50 mVp-p typ. with cap. 4.7µF/50V X7R MLCC  
Transient response  
Over load protection  
Short circuit protection  
– Recovery time (25% load step change)  
250 µs typ.  
at 150 % typ. of Iout rated (hiccup mode)  
Continuous, automatic recovery  
Over voltage protection  
–Single output  
3.3 VDC models: 3.75.4 VDC  
5.0 VDC models: 5.67.0 VDC  
12 VDC models: 13.519.6 VDC  
15 VDC models: 18.322.0 VDC  
24 VDC models: 29.132.5 VDC  
5 VDC models: 5.67.0 VDC  
–Dual output  
12 VDC models: 13.518.2 VDC  
15 VDC models: 17.022.0 VDC  
http://www.tracopower.com  
Page 2 of 4  
THM 10WI Series, 10 Watt  
General Specifications  
Capacitive load  
–Single output  
3.3 VDC models: 3‘000 µF maꢀ.  
5.0 VDC models: 2‘500 µF maꢀ.  
12 VDC models: 430 µF maꢀ.  
15 VDC models: 350 µF maꢀ.  
24 VDC models: 125 µF maꢀ.  
–Dual output  
5 VDC models: 1440 µF maꢀ. (each output)  
12 VDC models: 550 µF maꢀ. (each output)  
15 VDC models: 180 µF maꢀ. (each output)  
Temperature ranges  
– Operating  
– Rated according to IEC/EN 60601-1  
– Case temperature  
–40°C to +90°C (with derating)  
–40°C to +50°C (without derating)  
+105°C maꢀ.  
– Storage temperature  
–55°C to +125°C  
Derating  
3.5%/K above 75°C  
18°C/W  
Thermal impedance  
Humidity (non condensing)  
5 % to 95 % rel H maꢀ.  
Isolation voltage  
(50Hz, 60sec)  
– to meet ES/IEC/EN 60601-1  
5000 VACrms, rated for 250 VACrms working  
voltage, 2xMOPP  
Clearance/creepage  
8 mm min.  
Leakagecurrent (at 240VAC, 60Hz)  
Isolation capacitance (input/output)  
Altitude during operation  
2µA maꢀ.  
17 pF maꢀ.  
5000 m  
Temperature coefficient  
0.02 %/K typ.  
Reliability, calculated MTBF (MIL-HDBK-217F at +25°C, ground benign)  
Switching frequency  
3‘800‘000 h  
300 kHz 30 kHz (pulse width modulation)  
according to MIL-STD-810F  
Vibration and thermal shock resistance  
Safety standards/approvals – Medical equipment  
ANSI/AAMI ES60601-1:2005/(R)2012,  
IEC/EN60601-1 3rd edition  
– Certification documents  
www.tracopower.com/products/overview/thm10wi  
Environmental compliance – Reach  
www.tracopower.com/products/reach-declaration.pdf  
RoHS directive 2011/65/EU  
– RoHS  
Physical Specifications  
Casing material  
non-conductive black plastic  
non-conductive black plastic  
silicone (UL94 V-0 rated)  
14 g (0.48oz)  
Base material  
Potting material  
Package weight  
Soldering temperature  
maꢀ. 265°C / 10 sec  
he component is not be used in an oꢀygen rich environment.  
he component is not to be used in conjunction with flammable anaesthetics and agents.  
he component has to be disposed appropriately. Please refer to local regulations (Waste Electrical and Electronic Equipment).  
modification of the component is not allowed.  
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.  
http://www.tracopower.com  
Page 3 of 4  
THM 10WI Series, 10 Watt  
Outline Dimensions  
Standard pinning  
Standard Pinout  
2
11  
14  
Pin  
2
Single  
–Vin (GND)  
No con.  
+Vout  
Dual  
–Vin (GND)  
–Vout  
Bottom view  
11  
14  
16  
22  
23  
+Vout  
23 22  
16  
–Vout  
+Vin (Vcc)  
+Vin (Vcc)  
Common  
+Vin (Vcc)  
+Vin (Vcc)  
31.8 (1.25)  
4.5  
(0.18)  
2.54  
(0.10)  
15.24 (0.60)  
5.08  
(0.20)  
Optional pinning: suffiꢀ -A1  
Optional Pinout  
Pin  
1
Single  
+Vin (Vcc)  
No pin  
Dual  
+Vin (Vcc)  
Common  
No pin  
11  
12  
13  
15  
23  
24  
1
11 12  
13  
–Vout  
+Vout  
–Vout  
Bottom view  
No pin  
+Vout  
–Vin (GND)  
–Vin (GND)  
–Vin (GND)  
–Vin (GND)  
24 23  
15  
31.8 (1.25)  
Remark: No suffix -A1 for 5 Vin models. Corresponding  
parts are with THM 10 series by default.  
see www.tracopower.com/overview/thm10  
2.0  
Dimensions in [mm], () = Inch  
Tolerances 0.5 ( 0.02)  
Pin ø 0.6 0.1 (0.024 0.004)  
Pin pich tolerances 0.25 ( 0.01)  
(0.08)  
2.54  
(0.10)  
20.32 (0.80)  
2x2.54  
(2x0.1)  
©
Copyright 2017 Traco Electronic AG  
Specifications can be changed without notice!  
Rev. November 21. 2017  
Page 4 of 4  
www.tracopower.com  

相关型号:

THM10WI_15

DC/DC Converters
TRACOPOWER

THM10WI_18

DC/DC Converter
TRACOPOWER

THM10_18

DC/DC Converter
TRACOPOWER

THM15

DC/DC Converter
TRACOPOWER

THM15-1211

DC/DC Converter
TRACOPOWER

THM15-1212

DC/DC Converter
TRACOPOWER

THM15-1213

DC/DC Converter
TRACOPOWER

THM15-1215

DC/DC Converter
TRACOPOWER

THM15-1221

DC/DC Converter
TRACOPOWER

THM15-1222

DC/DC Converter
TRACOPOWER

THM15-1223

DC/DC Converter
TRACOPOWER

THM15-2411

DC/DC Converter
TRACOPOWER