Open frame 100 W power supply with
JST connection in 2.0” x 3.0” package
•
Certification according to IEC/EN/ES
60601-1 3rd edition for 2 x MOPP
•
Low leakage current <75 µA rated for
BF applications
•
Risk management process according to
ISO 14971 incl. risk management file
•
Acceptance criteria for electronic
assemblies acc. to IPC-A-610 Level 3
•
Active power factor correction >0.95
•
Protection class l and ll prepared
•
•
ES 60601-1 IEC 60601-1
UL 60950-1 IEC 60950-1
Operating up to 5000 m altitude
The TPP 100A-J series of 100 Watt AC/DC open frame power supplies feature
a reinforced double I/O isolation system according to latest medical safety
standards (60601-1 3rd edition, 2 x MOPP). The earth leakage current is below
75 µA which makes the units suitable for BF (body floating) applications. The
excellent efficiency of up to 92% allows a high power density for the standard
2.0” x 3.0” packaging format. The full load operating temperature range is
–25°C to +55°C while it goes up to 80°C with 50% load derating. The EMC
characteristic is dedicated for applications in industrial and medical fields. High
reliability is provided by the use of industrial quality grade components and an
excellent thermal management. It makes the products an ideal solution for
medical devices and for demanding safety and space critical applications.
Ready to meet ErP directive, <0.3 W no
•
load power consumption
5-year product warranty
•
Models
Order Code
Output Power
max.
Output Voltage
nom. (adjustable)
Output Current
max.
Efficiency
typ.
TPP 100-112A-J
TPP 100-115A-J
TPP 100-124A-J
TPP 100-128A-J
TPP 100-136A-J
TPP 100-148A-J
12 VDC (10.8 - 13.2 VDC)
15 VDC (13.5 - 16.5 VDC)
24 VDC (21.6 - 26.4 VDC)
28 VDC (25.2 - 30.8 VDC)
36 VDC (32.4 - 39.6 VDC)
48 VDC (43.2 - 52.8 VDC)
8'340 mA
6'670 mA
4'170 mA
3'580 mA
2'780 mA
2'090 mA
91 %
92 %
92 %
92 %
91 %
91 %
100 W
Page 1 / 4
www.tracopower.com
June 4, 2019
Input Specifications
Input Voltage
- AC Range
- DC Range
85 - 264 VAC (Full Range)
120 - 370 VDC (Designed for, no certification)
Input Frequency
Input Current
47 - 63 Hz
550 mA max.
1'150 mA max.
300 mW max. (Ready to meet ErP directive)
60 A max.
0.95 min. (Active Power Factor Correction)
0.95 min. (Active Power Factor Correction)
- Full Load & Vin = 230 VAC
- Full Load & Vin = 115 VAC
- at no Load
- at 230 VAC
- at 230 VAC
Power Consumption
Input Inrush Current
Power Factor
- at 115 VAC
Input Protection
T 3.15 A / 250 VAC (Internal Fuse in L & N)
Output Specifications
Output Voltage Adjustment
10% (By trim potentiometer)
Output power must not exceed rated power!
Voltage Set Accuracy
Regulation
1% max.
0.2% max.
0.5% max.
- Input Variation (Vmin - Vmax)
- Load Variation (0 - 100%)
Ripple and Noise
12 VDC model: 120 mVp-p typ. (with 10 µF X7R)
(20 MHz Bandwidth)
15 VDC model: 150 mVp-p typ. (with 10 µF X7R)
24 VDC model: 160 mVp-p typ. (with 1 µF X7R)
28 VDC model: 180 mVp-p typ. (with 1 µF X7R)
36 VDC model: 190 mVp-p typ. (with 1 µF X7R)
48 VDC model: 340 mVp-p typ. (with 0.1 µF X7R)
12 VDC model: 6'950 µF max.
15 VDC model: 4'450 µF max.
24 VDC model: 1'750 µF max.
28 VDC model: 1'280 µF max.
36 VDC model: 770 µF max.
48 VDC model: 430 µF max.
Not required
Capacitive Load
Minimum Load
Temperature Coefficient
Hold-up Time
0.02 %/K max.
16 ms min.
16 ms min.
1'000 ms max.
- at 230 VAC
- at 115 VAC
- at 230 VAC
- at 115 VAC
Start-up Time
1'000 ms max.
Short Circuit Protection
Output Current Limitation
Overvoltage Protection
Transient Response
Continuous, Automatic recovery
115 - 150% of Iout max.
115 - 135% of Vout nom.
3% max. (50% to 75% Load Step)
500 µs typ. (50% to 75% Load Step)
- Response Deviation
- Response Time
Safety Specifications
Safety Standards
- IT / Multimedia Equipment
IEC 60950-1
EN 60950-1
UL 60950-1
- Medical Equipment
EN 60601-1
IEC 60601-1
ANSI/AAMI ES 60601-1
2 x MOPP (Means Of Patient Protection)
www.tracopower.com/overview/tpp100a-j
Class I Prepared: Connection to PE
Class II Prepared: Reinforced Insulation
PD 2
- Certification Documents
Protection Class
Pollution Degree
Over Voltage Category
OVC II
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
www.tracopower.com June 4, 2019
Page 2 / 4
EMC Specifications
EMC Emissions
EN 60601-1-2 edition 4 (Medical Devices)
EN 55011 class B (internal filter)
EN 55032 class B (internal filter)
FCC Part 15, class B
- Conducted Emissions
- Radiated Emissions
FCC Part 18, class B
EN 55011 class A (internal filter)
EN 55032 class A (internal filter)
FCC Part 15, class A
FCC Part 18, class A
- Harmonic Current Emissions
- Voltage Fluctuations & Flicker
EN 61000-3-2, class A
EN 61000-3-2, class D
EN 61000-3-3
EMC Immunity
EN 55024 (IT Equipment)
EN 60601-1-2 edition 4 (Medical Devices)
Air: EN 61000-4-2, 15 kV, perf. criteria A
Contact: EN 61000-4-2, 8 kV, perf. criteria A
EN 61000-4-3, 20 V/m, perf. criteria A
EN 61000-4-4, 2 kV, perf. criteria A
L to L: EN 61000-4-5, 1 kV, perf. criteria A
L to PE: EN 61000-4-5, 2 kV, perf. criteria A
EN 61000-4-6, 20 Vrms, perf. criteria A
EN 61000-4-8, 10 A/m, perf. criteria A
- Electrostatic Discharge
- RF Electromagnetic Field
- EFT (Burst)
- Surge
- Conducted RF Disturbances
- PF Magnetic Field
- Voltage Dips & Interruptions
230 VAC / 50 Hz: EN 61000-4-11
30%, 25 periods, perf. criteria A
>95%, 0.5 periods, perf. criteria A
>95%, 1 period, perf. criteria A
>95%, 250 periods, perf. criteria B
115 VAC / 60 Hz: EN 61000-4-11
30%, 25 periods, perf. criteria A
>95%, 0.5 periods, perf. criteria A
>95%, 1 period, perf. criteria A
>95%, 250 periods, perf. criteria B
General Specifications
Relative Humidity
95% max. (non condensing)
–25°C to +85°C
–40°C to +85°C
2 %/K above 55°C
1.33 %/V below 100 VAC
Natural convection (20 LFM)
5'000 m max.
45 - 75 kHz (PWM QR)
Reinforced Insulation
250 VAC
5'657 VDC
2'121 VDC
2'121 VDC
100 MOhm min.
75 µA max.
Temperature Ranges
- Operating Temperature
- Storage Temperature
- High Temperature
- Low Input Voltage
Power Derating
Cooling System
Altitude During Operation
Switching Frequency
Insulation System
Working Voltage (rated)
Isolation Test Voltage
- Input to Output, 60 s
- Input to Case or PE, 60 s
- Output to Case or PE, 60 s
- Input to Output, 500 VDC
- Touch Current
Isolation Resistance
Leakage Current
(264 VAC)
Reliability
Environment
- Calculated MTBF
- Vibration
- Mechanical Shock
790'000 h (MIL-HDBK-217F, ground benign)
IEC 60068-2-6
IEC 60068-2-27
JST
Connection Type
Weight
156 g
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
www.tracopower.com June 4, 2019
Page 3 / 4
www.tracopower.com/info/reach-declaration.pdf
www.tracopower.com/info/rohs-declaration.pdf
Environmental Compliance - Reach
- RoHS
Supporting Documents
Overview Link (for additional Documents)
www.tracopower.com/overview/tpp100a-j
Outline Dimensions
Each one of the 4 screw holes can be used
as a PE connection for CLASS I application.
Pin connectors
Input (CON1)
Output (CON2)
76.2 (3.00)
Pin Function Pin* Function
1
3
1,2
3,4
Line
–Vout
+Vout
69.8 (2.748)
Neutral
*Terminal rated for 7 A max.
(at higher current connection has to be split)
CON1: JST series
1
3
mates with JST crimp terminal: SVH-21T-P1.1
and terminal housing: VHR-3N
4
1
CON2: JST series
mates with JST crimp terminal: SVH-21T-P1.1
and terminal housing: VHR-4N
Voltage adjustment
TOP VIEW
Dimension in mm, () = inch
Tolerances: x.x 0.50 ( 0.02)
x.xx 0.25 ( 0.01)
2.5 (0.10) max.
FRONT VIEW
©
Copyright 2019 Traco Electronic AG
Specifications can be changed without notice.
Rev. June 4, 2019
www.tracopower.com
Page 4 / 4
相关型号: