WMIHPU15-104x [WALL]

Up to 15 Watts, Medical AC - DC Wall Mount Power Supplies;
WMIHPU15-104x
型号: WMIHPU15-104x
厂家: WALL INDUSTRIES,INC.    WALL INDUSTRIES,INC.
描述:

Up to 15 Watts, Medical AC - DC Wall Mount Power Supplies

文件: 总4页 (文件大小:531K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
WMIHPU15 SERIES  
Up to 15 Watts, Medical AC/DC Wall Mount Power Supplies  
Rev A  
Wall Industries, Inc.  
FEATURES  
Class II System  
RoHS Compliant  
Single Outputs Ranging from 3VDC to 55VDC  
-10°C to +70°C Operating Temperature Range  
Up to 15 Watts Output Power  
86% High Efficiency  
Meets FCC Part-18 Class B & CISPR-11 EN55011  
Class B Emission Limits  
Energy Star 2.0, Efficiency Level V  
80-275VAC Input Voltage Range  
Over Voltage & Over Load Protection  
Input to Output: 2MOPP  
UL/cUL 3rd Edition Medical Approvals  
Interchangeable Plug Options: United States,  
Europe, Australia, & United Kingdom Types  
Optional Output Connectors Available  
SAFETY APPROVALS  
Size:  
Weight:  
5.82oz (165g)  
2.36 x 1.71 x 1.58 inches  
60.0 x 43.5 x 40.2 mm  
DESCRIPTION  
The WMIHPU15 series of Class II medical AC/DC wall mount power supplies offers up to 15  
watts of output power in a 2.36” x 1.71” x 1.58” package. This series consists of single output  
models ranging from 3VDC to 55VDC with an 80~275VAC input voltage range and a -10°C  
to +70°C operating temperature. This series meets FCC Part-18 Class B and CISPR-11  
EN55011 Class B Emission Limits and has ANSI/AAMI ES 60601-1: 2005 (UL/cUL 3rd edition)  
and CE safety approvals. All units are RoHS and Energy Star Level V compliant. Plugs come  
in United States (“U” suffix), Europe (“E” suffix), Australia (“A” suffix), and United Kingdom  
(“K” suffix) types. Plugs are sold separately so please contact factory for ordering details.  
AC Plug Types:  
Applications:  
Home Healthcare Equipment  
Patient Monitors  
Unites States (“U” Suffix)  
Europe (“E” Suffix)  
Australia (“A” Suffix”)  
United Kingdom (“K” Suffix) Portable Medical Devices  
ECG Machines  
Blood Pressure Systems  
MODEL SELECTION TABLE  
Adjustable Voltage Range Adjustable Current Range  
Maximum  
Output Power  
Total  
Regulation  
No Load  
Consumption  
Model Number (1) Input Voltage  
Ripple & Noise  
Efficiency  
Min  
Max  
Min  
Max  
WMIHPU15-101x  
WMIHPU15-102x  
WMIHPU15-103x  
WMIHPU15-104x  
WMIHPU15-105x  
3.0 VDC  
4.99 VDC  
1.44 A  
2.40 A  
7.2W  
12W  
12W  
13.5W  
15W  
15W  
15W  
15W  
15W  
15W  
15W  
15W  
66mVp-p  
50mVp-p  
70mVp-p  
90mVp-p  
100mVp-p  
100mVp-p  
100mVp-p  
100mVp-p  
100mVp-p  
100mVp-p  
100mVp-p  
100mVp-p  
5%  
5%  
5%  
5%  
5%  
5%  
5%  
5%  
3%  
3%  
3%  
3%  
65%  
80%  
82%  
83%  
84%  
85%  
85%  
86%  
86%  
86%  
86%  
86%  
0.2W  
0.3W  
0.3W  
0.3W  
0.3W  
0.3W  
0.3W  
0.3W  
0.3W  
0.3W  
0.3W  
0.3W  
5.0 VDC  
6.5 VDC  
5.99 VDC  
8.0 VDC  
2.00 A  
1.50 A  
1.22 A  
1.15 A  
0.94 A  
0.72 A  
0.55 A  
0.45 A  
0.37 A  
0.32 A  
0.28 A  
2.40 A  
1.84 A  
1.68 A  
1.36 A  
1.15 A  
0.94 A  
0.72 A  
0.55 A  
0.45 A  
0.37 A  
0.32 A  
8.0 VDC  
11.0 VDC  
13.0 VDC  
16.0 VDC  
21.0 VDC  
27.0 VDC  
33.0 VDC  
40.0 VDC  
48.0 VDC  
55.0 VDC  
11.0 VDC  
13.0 VDC  
16.0 VDC  
21.0 VDC  
27.0 VDC  
33.0 VDC  
40.0 VDC  
48.0 VDC  
WMIHPU15-106x  
80 ~ 275VAC  
WMIHPU15-107x  
WMIHPU15-108x  
WMIHPU15-109x  
WMIHPU15-110x  
WMIHPU15-111x  
WMIHPU15-112x  
NOTES  
1. The “x” in the model number can be “U” for United States type plug; “E” for Europe type plug, “A” for Australia type plug, or “K” for United Kingdom type  
plug. Plugs are sold separately so please contact factory for ordering details.  
2. Output can provide up to peak load when power supply starts up. Continuous staying in more than rated load is not allowed.  
3. At factory, in 60% rated load condition, each output is checked to be within voltage accuracy.  
4. Line regulation is defined by changing 10% of input voltage from nominal line at rated load.  
5. Load regulation is defined by changing 40% of measured output load from 60% rated load.  
6. Ripple & noise is measured by using 20MHz limited bandwidth and with 0.47F capacitor in parallel across the output at nominal line and rated load.  
7. Hold-up time is measured from the end of the last charging pulse to the time which the main output drops down to low limit of main output at rated  
load and nominal line.  
8. Optional output connectors are available for this series. Please call factory for ordering details.  
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797  
08/04/2014  
Page 1 of 4  
website: www.wallindustries.com • e-mail: sales@wallindustries.com  
WMIHPU15 SERIES  
Up to 15 Watts, Medical AC/DC Wall Mount Power Supplies  
Rev A  
Wall Industries, Inc.  
TECHNICAL SPECIFICATIONS: WMIHPU15 SERIES  
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.  
We reserve the right to change specifications based on technological advances.  
SPECIFICATION  
TEST CONDITIONS  
Min  
Typ  
Max  
Unit  
INPUT SPECIFICATIONS  
Safety Approvals Input Voltage Range  
Derating linearly from 100% load at 90VAC to 80% load at 80VAC  
Sine wave  
100VAC, full load  
240VAC, full load  
100VAC, full load, 25°C, cold start  
240VAC, full load, 25°C, cold start  
230VAC, no load  
100  
80  
47  
240  
275  
63  
0.6  
0.4  
35  
Input Voltage  
Input Frequency  
Input Current  
VAC  
Hz  
A
Inrush Current  
A
85  
No Load Power Consumption  
OUTPUT SPECIFICATIONS  
Output Voltage  
See Table  
See Table  
Line Regulation  
Total Regulation  
Output Power  
Output Current  
See Note 4  
See Note 5  
1
%
See Table  
See Table  
See Table  
See Table  
Ripple & Noise (peak to peak)  
Hold-up Time  
Start-up Time  
Transient Response Time  
Temperature Coefficient  
PROTECTION  
Short Circuit Protection  
Over Load Protection  
GENERAL SPECIFICATIONS  
Efficiency  
See Note 6  
See Note 7  
100~240VAC, full load  
110VAC, Full load  
12  
0.3  
ms  
s
ms  
%/°C  
2
4
-0.04  
110  
+0.04  
Auto-recovery  
150  
Recovers automatically after fault condition is removed  
%
Nominal line and rated load  
See Table  
Dielectric Withstanding Voltage  
Protection Classes  
ENVIRONMENTAL SPECIFICATIONS  
Operating Temperature  
Storage Temperature  
Operating Humidity  
Storage Humidity  
Primary to Secondary, limit current <10mA  
5656  
VDC  
Double insulated, Class II  
Derating linearly from 100% Load at 40°C to 50% load at 70°C  
10~95% RH  
Non-condensing  
-10  
-40  
0
+70  
+85  
95  
95  
°C  
°C  
%RH  
%RH  
0
Cooling  
Flammability Rating  
Vibration  
Operating Altitude (Elevation)  
MTBF  
Free air convection  
UL94V-1  
10~500Hz, 10min./1cycle, 60 min each along X, Y, Z axes  
MIL-HDBK-217F, 25°C  
5
G
m
hours  
3000  
100,000  
PHYSICAL SPECIFICATIONS  
Weight  
5.82oz (165g)  
Dimensions (L x W x H)  
2.36 x 1.71 x 1.58 inches (60.0 x 43.5 x 40.2 mm)  
“U” suffix  
“K” suffix  
“E” suffix  
“A” suffix  
United States  
United Kingdom  
Europe  
AC Plug  
Australia  
Output Connector  
Several options available  
SAFETY, EMC, & COMPLIANCE  
ANSI/AAMI ES 60601-1:2005 (UL/cUL 3rd ed.)  
Safety Approvals  
IEC60601-1 3rd edition  
CE  
EMC Emission  
Surge Voltage  
Compliance to EN55011 (CISPR11), EN61000-3-2,-3  
B
2
8
6
Class  
kV  
Air Discharge, IEC61000-4-2  
Contact Discharge, IEC61000-4-2  
Electro Static Discharge  
kV  
Compliance  
CEC & Energy Star  
RoHS Compliant  
CEC and Energy Star 2.0, Efficiency Level V  
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797  
website: www.wallindustries.com • e-mail: sales@wallindustries.com  
08/04/2014  
Page 2 of 4  
WMIHPU15 SERIES  
Up to 15 Watts, Medical AC/DC Wall Mount Power Supplies  
Rev A  
Wall Industries, Inc.  
DERATING  
Input Voltage Derating Curve  
Temperature Derating Curve  
110  
100  
90  
110  
100  
90  
80  
70  
60  
50  
40  
30  
20  
10  
0
80  
70  
60  
50  
80  
100  
120  
140  
160  
180  
200  
220  
240  
260  
280  
-10  
0
10  
20  
30  
40  
50  
60  
70  
80  
Input Voltage (VAC)  
Temperature (°C)  
MECHANICAL DRAWING  
1.71±±.±0 [03.5±1.±]  
United States Type  
(SuffixU”)  
OPTIONAL  
1.58±±.±0 [0±.2±1.±]  
United Kingdom Type  
(SuffixK”)  
Europe Type  
(SuffixE”)  
NOTES:  
1. Unit: inches [mm]  
2.Tolerance: ±±.±0 [±1.±]  
3.Weight: 5.82oz (165g)  
AustraliaType  
(SuffixA”)  
0. Models WMIHPU15-1±1~1±0 need to use AWG#18/0FT output cable in order to meet the total regulation specified.  
Models WMIHPU15-1±5~1±9 need to use AWG#2±/0FT output cable in order to meet the total regulation specified.  
Models WMIHPU15-11±~112 need to use AWG#22/0FT output cable in order to meet the total regulation specified.  
The technical specifications will change if a different output cable is used.  
5. Plugs are sold separately. Please call factory for ordering details.  
6. Optional output connectors available. Please call factory for ordering details.  
7. All dimensions are for reference only.  
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797  
website: www.wallindustries.com • e-mail: sales@wallindustries.com  
08/04/2014  
Page 3 of 4  
WMIHPU15 SERIES  
Up to 15 Watts, Medical AC/DC Wall Mount Power Supplies  
Rev A  
Wall Industries, Inc.  
COMPANY INFORMATION  
Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development engineers will  
provide a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008 certification is just one  
example of our commitment to producing a high quality, well-documented product for our customers.  
Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working closely with its  
customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to provide ongoing  
support for your project above and beyond the design and production phases. Give us a call today to discuss your future projects.  
Contact Wall Industries for further information:  
Phone:  
Toll Free:  
Fax:  
(603)778-2300  
(888)597-9255  
(603)778-9797  
E-mail:  
Web:  
Address:  
sales@wallindustries.com  
www.wallindustries.com  
37 Industrial Drive  
Exeter, NH 03833  
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797  
website: www.wallindustries.com • e-mail: sales@wallindustries.com  
08/04/2014  
Page 4 of 4  

相关型号:

WMIHPU15-105x

Up to 15 Watts, Medical AC - DC Wall Mount Power Supplies
WALL

WMIHPU15-106x

Up to 15 Watts, Medical AC - DC Wall Mount Power Supplies
WALL

WMIHPU15-107x

Up to 15 Watts, Medical AC - DC Wall Mount Power Supplies
WALL

WMIHPU15-108x

Up to 15 Watts, Medical AC - DC Wall Mount Power Supplies
WALL

WMIHPU15-109x

Up to 15 Watts, Medical AC - DC Wall Mount Power Supplies
WALL

WMIHPU15-110x

Up to 15 Watts, Medical AC - DC Wall Mount Power Supplies
WALL

WMIHPU15-111x

Up to 15 Watts, Medical AC - DC Wall Mount Power Supplies
WALL

WMIHPU15-112x

Up to 15 Watts, Medical AC - DC Wall Mount Power Supplies
WALL

WMIHPU25

Up to 25 Watts, Medical AC - DC Wall Mount Power Supplies
WALL

WMIHPU25-101x

Up to 25 Watts, Medical AC - DC Wall Mount Power Supplies
WALL

WMIHPU25-102x

Up to 25 Watts, Medical AC - DC Wall Mount Power Supplies
WALL

WMIHPU25-103x

Up to 25 Watts, Medical AC - DC Wall Mount Power Supplies