MODEL: 47LLC
output to power
INPUT SPECIFICATIONS
■ Strain Gauge Input
Excitation: 2.5 V, 5 V, 7.5 V or 10 V selectable (front)
Tare adjustment: Via the front button control or by the
external contact input
• Strain Gauge: Refer to the next table
• Excitation: 2.5 V, 5 V, 7.5 V or 10 V selectable
Maximum current: 120 mA
Setting: (Front button)
• Scaled range
Load resistance: ≥ 85 Ω
• Alarm setpoint
Default setting: 2.5 V
• Hysteresis (deadband)
■ Tare Command Input: TTL level (5V-CMOS level), open
collector or dry contact
• Moving average
• Others
(saturation voltage ≤ 1 V, sink current 0.5 mA)
(Refer to the instruction manual for details)
Read rate: 40 times/sec. (25 msec.)
Averaging: None or moving average
Lockout setting: Prohibiting certain operations; protecting
settings
TYPE
S1
STRAIN GAUGE RATING
0.0 – 1.0mV/V
ZERO ADJUSTMENT
-10.00 – +10.00mV
-30.00 – +30.00mV
-100.0 – +100.0mV
-300.0 – +300.0mV
S2
0.0 – 3.0mV/V
S3
0.0 – 10.0mV/V
0.0 – 30.0mV/V
S4
SPAN
INPUT VOLTAGE
-10.00 – +20.00mV
-30.00 – +60.00mV
-100.0 – +200.0mV
-300.0 – +600.0mV
TYPE
S1
1.00 – 10.00mV
3.00 – 30.00mV
10.0 – 100.0mV
30.0 – 300.0mV
DISPLAY
S2
Display: 16 mm (.63) high, 4 ½ digits, 7-segment LED
S3
Display range: -19999 to 19999
S4
Decimal point position: 10–1, 10–2, 10–3, 10–4 or none
Zero indication: Higher-digit zeros are suppressed.
Over-range indication: ‘-19999’ or ‘19999’ blinking for
display values out of the scaled range.
‘S.ERR’ and ‘Max’ blinking when the input signal is out of
the usable range.
Note 1: Choose Type S2 for the strain gauge rating 2.0mV/V with the
excitation voltage 10V.
For detailed information, refer to “Input Scaling Values A/B v.s.
Tare Adjustment” under “PARAMETER LIST” on the instruction manual.
Note 2: Consult factory for use with a compression/tension load cells.
Alarm status indication
OUTPUT SPECIFICATIONS
■ DC Output
LL indicator: Green LED turns on when the LL alarm is
tripped.
• DC Current
L indicator: Green LED turns on when the L alarm is
tripped.
Operational range: -5 – +105 %
• DC Voltage
H indicator: Red LED turns on when the H alarm is tripped.
HH indicator: Red LED turns on when the HH alarm is
tripped.
Operational range: -5 – +105 %
■ Alarm Output
• Relay contact
P indicator: Amber LED turns on when none of the other
alarms is tripped.
Rated load: 250 V AC @ 3 A (cos ø = 1)
30 V DC @ 3 A (resistive load)
Maximum switching voltage: 250 V AC, 30 V DC
Maximum switching power: 750 VA, 90 W (resistive load)
Minimum load: 5 V DC @ 10 mA
Only ‘P’ turns on with no-alarm-output type. ‘LL’ or ‘HH’
does not turn on with dual-alarm-output type.
All setpoints can be independently set either for Hi or Lo
alarm trip.
Mechanical life: ≥ 5 × 10
6
cycles (rate 180 cycles/min.)
Function indicators:
• Photo MOSFET Relay
Zro, Spn, D/P, Tch, Fnc, Min, Max
Rating: 120 V AC/DC @ 80 mA (resistive load)
ON resistance: ≤ 30 Ω
Display mode status and operation status, amber ON or
blink
Permissible: 250 mW
Engineering unit indication: Sticker label attached
DC, AC, mV, V, kV, μA, mA, A, kA, mW, W,
kW, var, kvar, Mvar, VA, Hz, Ω, kΩ, MΩ,
cm, mm, m, m/sec, mm/min, cm/min, m/min,
INSTALLATION
Power consumption
•AC:
m/h, m/s2, inch, ℓ, ℓ/s, ℓ/min, ℓ/h, m3, m3/sec,
Approx. 9 VA at 100 V
Approx. 12 VA at 200 V
Approx. 13 VA at 264 V
•DC: Approx. 5 W
m
3
/min, m
3
/h, Nm /h, N·m, N/m , g, kg, kg/h,
3
2
N, kN, Pa, kPa, MPa, t, t/h, ℃, °F, %RH, J,
kJ, MJ, rpm, sec, min, min-1, pH, %, ppm, etc.