TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
030AAC
Highlights & Features
–
–
–
–
–
–
–
–
–
–
–
MDS-030AAC05, 12, 15, 24 Meet Efficiency Level VI
MDS-030AAC07 Meet Efficiency Level V
Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1
Compliant with IEC 60601-1-2 4th Ed. Requirements
Meets Limited Power Source (LPS) requirements
Low touch current. Suitable for type BF applications
Detachable AC plug with multiple country options
IP22 ingress protection rating
2 × MOPP isolation
500K hours MTBF
3 years warranty
Safety Standards
CB Certified for worldwide use
Model Number:
Unit Weight:
MDS-030AAC☐
150 grams
Dimensions (W × L × H): 53.5 × 88.0 × 27.5 mm
2.11 × 3.46 × 1.08 inch
General Description
The MDS series of external power supply comes with universal AC input at 85Vac to 264Vac. Other features include low touch current,
risk management report available and the electric shock protection comply with 2 × MOPP. The MDS series is certified for EMC
standards according to EN 55011 for industrial, scientific and medical (ISM) radio-frequency equipment and EN 55022 for Information
Technology Equipment (ITE) radio-frequency equipment. In addition, only recognized Japanese capacitors are used.
The MDS series come with both medical and ITE safety approvals including UL/cUL and IEC-60950-1 CB certification as a Limited
Power Source (LPS), and CCC approval. Designs are fully compliant with RoHS Directive 2011/65/EU for environmental protection.
Model Information
Medical AC-DC Adapter
Model Number
MDS-030AAC05
MDS-030AAC07
MDS-030AAC12
MDS-030AAC15
MDS-030AAC24
Input Voltage Range Efficiency Level
Rated Output Voltage
Rated Output Current
85-264Vac
Level VI
Level V
Level VI
Level VI
Level VI
5Vdc
3A
7Vdc
3A
12Vdc
15Vdc
24Vdc
2A
2A
1.25A
Model Numbering
MDS –
030
AAC
Delta Medical
Power Supply
Max wattage in the
product series
Family Code
Output Voltage (Single
Output)
Revision Code1
Revision Code2
DC plug type and output cable length
Country duck-head type
- 05 for 5V
- 07 for 7V
- 12 for 12V
- 15 for 15V
- 24 for 24V
A:2.1 x 5.5 x10mm DC plug
B:2.1 x 5.5 x12mm DC plug
A:China ; B:United States;
C:European; D:United Kingdom;
E:Australia; G:Korea; H: India,
J :Argentina ; K:Brazil;
Both A & B are 1200mm cable length
M:South Africa
* MDS-030AAC15 BB is not available.
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (May 2017, Rev.07)
1
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
Specifications
MDS-030AAC12
MDS-030AAC15
MDS-030AAC24
Model Number
MDS-030AAC05
MDS-030AAC07
Input Ratings / Characteristics
Nominal Input Voltage
Input Voltage Range
100-240Vac
85-264Vac
50-60Hz
Nominal Input Frequency
Input Frequency Range
47-63Hz
Input Current (max)
115Vac 0.5A
230Vac 0.3A
0.6A
0.8A
0.8A
0.8A
0.4A
0.6A
0.6A
0.6A
Average Efficiency (min)
Standby Power (max)
115Vac
230Vac
115Vac
230Vac
81.4%
0.1W
81.4%
86.21%
87.0%
87.0%
0.1W
0.1W
0.1W
0.1W
Inrush Current (typ.)
Touch Current (max)
60A @ 115Vac and 100A @ 230Vac
0.1mA @ 264Vac NC1), 0.3mA @ 264Vac SFC2)
0.1mA @ 264Vac NC1), 0.5mA @ 264Vac SFC2)
Output-PE (protective earth)
leakage current for Type BF
application (max)
1)
2)
NC: Normal condition
SFC: Single fault condition
MDS-030AAC12
MDS-030AAC15
15Vdc
MDS-030AAC24
24Vdc
Model Number
MDS-030AAC05
5Vdc
MDS-030AAC07
Output Ratings / Characteristics
Nominal Output Voltage
Output Total Regulation
Output Current
7Vdc
12Vdc
±6%
2A
3A
3A
2A
1.25A
30W
Output Power
15W
±1%
±5%
21W
24W
30W
Line Regulation
Load Regulation
Ripple & Noise (max)
100mVpk-pk @
Rated load
100mVpk-pk @
Rated load
150mVpk-pk @
Rated load
200mVpk-pk @
Rated load
200mVpk-pk
@ Rated load
Start-up Time (max)
Hold-up Time (min)
3000ms @ 115Vac
10ms @ 115Vac
Dynamic Response (Overshoot & ±10% @ 50-100% load
Undershoot O/P Voltage)
Capacitive load (max)
4700uF
*Periodic and Random Deviation.
Ripple & Noise measurement circuit with 20MHz BW
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (May 2017, Rev.07)
2
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
MDS-030AAC12
MDS-030AAC15
MDS-030AAC24
Model Number
MDS-030AAC05
MDS-030AAC07
Mechanical
Case Chassis
Case Cover
PC
PC
Dimensions (W × L × H)
Unit Weight
53.5 × 88.0 × 27.5 mm (2.11 × 3.46 × 1.08 in)
150 grams (5.3 ounces)
Indicator
NA
Cooling System
Terminal
NA
Input Detachable type AC plug
Output Barrel type. Dimensions 2.1 × 5.5 × 10 mm (see page 5)
Environment
Surrounding Air
Temperature
Operating 0°C to +40°C
Storage -40°C to +85°C
Operating Humidity
Operating Altitude
5 - 95% RH (Non-Condensing)
5,000 meters (16400 feet)
Shock Test
50G, 11ms, 3 shocks for each direction
(Non-Operating)
Vibration
5-500Hz, 2.09Grms, 20 minute for each three axis
(Operating)
Protections
Overvoltage (max)
150%, Latch Mode
Overload / Overcurrent (max)
Over Temperature
Short Circuit
250% of rated load current, Hiccup Mode,
(Non-Latching, Auto-Recovery)
Hiccup Mode,
(Non-Latching, Auto-Recovery)
Hiccup Mode,
(Non-Latching, Auto-Recovery)
Degree of Protection
IP22
Protection Against Shock
Class II
Reliability Data
MTBF (min)
500K Hours based on Telecordia SR-332 (at 100Vac, Max. load and 25°C Ambient)
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (May 2017, Rev.07)
3
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
MDS-030AAC12
MDS-030AAC15
MDS-030AAC24
Model Number
MDS-030AAC05
MDS-030AAC07
Safety Standards / Directives
Medical Safety
IEC60601-1 Edition 3.1 (2012), EN60601-1 (2006) + A11 + A1 + A12, CAN/CSA-C22.2 NO.
60601-1:14, ANSI/AAMI ES60601-1:2005/(R)2012
UL60601-1 (File No. E356265)
ITE Safety
IEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-2008, GB17625.1-2003
UL60950-1 (File No. E131881), LPS (Limited Power Source)
MDD Directive 93/42/EEC
CE
Material and Parts
RoHS Directive 2011/65/EU Compliant
Galvanic Isolation
Input to 4000 Vac
Output
EMC (Compliant with IEC 60601-1-2 4th Ed. Requirements)
EMC / Emissions
EN55011/EN55022,FCC Title 47:Class B
Harmonic Current Emissions
Immunity to
IEC61000-3-2
Voltage Flicker
IEC61000-3-3
Electrostatic Discharge
IEC61000-4-2 Level 4 Criteria A1)
Air Discharge: 15kV
Contact Discharge: 8kV
Criteria A1)
Radiated Field
IEC61000-4-3
80MHz-2700MHz, 10V/m AM modulation
385MHz-5785MHz, 28V/m Pulse mode and other
modulation
Electrical Fast Transient / Burst
Surge
IEC61000-4-4 Level 3 Criteria A1): 2kV
IEC61000-4-5 Level 3 Criteria A1)
Differential Mode3): 1kV
Level 2 Criteria A1)
IEC61000-4-6
Conducted
150kHz-80MHz, 3Vrms, 6Vrms at ISM bands and
Amateur radio bands
IEC61000-4-8 Criteria A1)
Power Frequency Magnetic Fields
Voltage Dips
Magnetic field strength 30A/m
Criteria A1)
IEC61000-4-11
0% UT, 0.5 cycle (10ms)
, 0o/45o/90o/135o/180o/225o/270o/315o/360o
Criteria B2)
0% UT, 1 cycle (20ms), 0o
Criteria B2)
70% UT, 25 cycle (500ms), 0o
Criteria B2)
0% UT, 250 cycle (5000ms), 0o
1) Criteria A: Normal performance within the specification limits
2) Criteria B: Output out of regulation, or shuts down during test. Automatically restore to normal operation after test.
3) Symmetrical: Differential mode (Line to line)
Block Diagram
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (May 2017, Rev.07)
4
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
Dimensions
W × H × H: 53.5 × 88.0 × 27.5 mm
Note:
The plug’s polarity is
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (May 2017, Rev.07)
5
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
Duck Head Type:
CN type
US/JP type
EU type
UK type
AU
KR type
IN type
AR type
BZ type
SA type
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (May 2017, Rev.07)
6
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
Functions
Start-up Time
The time required for the output voltage (Vo) to reach 90% of its set value, after the input AC voltage is applied.
Rise Time
The time required for the output voltage (Vo) to change from 10% to 90% of its steady state value.
Hold-up Time
Hold up time is the time when the AC input collapses and output voltage retains regulation for a certain period of time. The time
required for the output to reach 95% of its set value, after the input voltage is removed.
Graph illustrating the Start-up Time, Rise Time, and Hold-up Time
Inrush Current
Dynamic Response
Inrush current is the input current that occurs when the input
voltage is first applied. For AC input voltages, the maximum
peak value of inrush current will occur during the first half cycle
of the applied AC voltage. This peak value decreases
exponentially during subsequent cycles of AC voltage.
The power supply output voltage will remain within ±3% of its
steady state value, when subjected to a dynamic load change
from 50 to 100% of its rated current.
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (May 2017, Rev.07)
7
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
Overload & Over current Protections
Overvoltage Protection
The power supply’s Overload (OLP) and Over current (OCP)
Protections will be activated when output current is between
110% and 250% of IO (Max load). Upon such an occurrence, VO
will start to drop. Once the power supply has reached its
maximum power limit, the protection will be activated and the
power supply will go into “Hiccup mode” (Auto-Recovery). The
power supply will recover once the fault condition causing the
OLP and OCP is removed and IO is back within the specified limit.
The power supply’s overvoltage circuit will be activated when its
internal feedback circuit fails. The output voltage shall not exceed
its specifications defined on Page 3 under “Protections”. Power
supply will latch off, and require removal/re-application of input
AC voltage in order to restart.
Over Temperature Protection
As mentioned above, the power supply also has Over
Temperature Protection (OTP). This is activated when the
overload condition persists for an extended duration and the
output current is below the overload trigger point but >100% load.
In the event of a higher operating condition at 100% load, the
power supply will run into OTP when the surrounding air
temperature is higher than the operating temperature. When
activated, the output voltage will go into hiccup mode until the
input voltage is removed; then, reapplied, and the surrounding air
temperature drops to its normal operating temperature.
Additionally, if the IO is <250% but >110% for a prolong period of
time (depending on the load), the Over Temperature Protection
(OTP) will be activated due to high temperature on critical
components. The power supply will then go into hiccup mode until
the fault is removed; and, the input voltage is removed, then
reapplied.
Short Circuit Protection (Auto-Recovery)
The power supply’s output OLP/OCP function also provides
protection against short circuits. When a short circuit is applied,
the output current will operate in “Hiccup mode”, as shown in the
illustration in the OLP/OCP section on this page. The power
supply will return to normal operation after the short circuit is
removed.
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (May 2017, Rev.07)
8
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
Certificate
All Delta Medical Power products conform to the European directive 2011/65/EU. RoHS is the abbreviation for
“Restriction of the use of certain hazardous substances.
Typically, the input current waveform is not sinusoidal due to the periodical peak charging of the input capacitor.
In industrial environment, complying with EN 61000-3-2 is only necessary under special conditions. Complying
to this standard can have some technical drawbacks, such as lower efficiency as well as some commercial
aspects such as higher purchasing costs. Frequently, the user does not profit from fulfilling this standard,
therefore, it is important to know whether it is mandatory to meet this standard for a specific application.
In addition to a UL Total Certification Program (TCP) approved client laboratory for IEC60950 and IEC60065.
Delta also has participated UL Client Test Data Program (CDTP) for IEC 60601.
Meet Level V Efficiency Requirement for MDS-030AAC07
Meet Level VI Efficiency Requirement for MDS-030AAC05, 12, 15, 24
Meets Limited Power Source (LPS) requirement
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (May 2017, Rev.07)
9