GAS ANALYZER FOR STACK GAS  
ZSQ  
This instrument is equipped with a non-dispersive infrared  
CO sensor and a zirconia O2 sensor. It is used for measur-  
ing CO and O2 in the gas emitted from a garbage incinera-  
tor to meet Japanese regulation on emission control of  
dioxins (tetrachlorodibenzoparadioxins).  
CO instantaneous value, O2 instantaneous value, CO instan-  
taneous value after O2 correction and CO average values  
after O2 correction (1-hour and time-variable averagings) can  
be output, and a peak count/alarm function is incorporated.  
FEATURES  
1. Suitable functions for emission control of dioxins.  
• Output includes CO instantaneous value, O2 instantaneous  
value, CO instantaneous value after O2 correction and  
CO average values after O2 correction (1-hour and time-  
variable averagings).  
• Count/alarm function against excess over specified value  
2. Space-saving configuration by unitizing the analyzer and  
sampling device into 19-inch types.  
3. Easy to maintain with auto calibration and with “From  
the front maintenance construction”.  
4. A microjet recorder (Fuji Electric’s type PHC or PHE) can  
be built in for recording measured gas concentration and  
temperature values.  
• Output signals: Each signal within range from 4 to 20 mA  
DC  
Non-isolated (isolated output available at  
option)  
CO instantaneous value  
O2 instantaneous value  
CO instantaneous value after O2 correc-  
tion  
1-hour moving average CO value after O2  
correction  
Moving average CO value after O2 cor-  
rection (averaging time variable)  
Time setting in 1 to 59 minutes (1-minute  
increment) or 1 to 4 hours (1-hour incre-  
ment)  
SPECIFICATIONS  
Allowable load resistance 550 or less  
• Contact output: 1a no-voltage contact (contact capacity  
250 V AC, 2 A)  
• Measuring system:  
Non-dispersive infrared absorption (NDIR)  
method for CO  
Zirconia method for O2  
• Measurable component and range:  
CO ; 0~200/500ppm  
Peak count alarm, CO instantaneous  
value upper limit alarm, auto calibration  
and maintenance statuses, calibration er-  
ror, analyzing block error and range dis-  
crimination  
0~200/1000ppm  
• Contact input: No-voltage contact (ON/0V, OFF/5 V DC,  
DC 5mA at ON)  
0~500/1000ppm  
0~500/2000ppm  
Auto calibration remote start (for 500  
msec or longer), remote range  
changeover  
Remote average value resetting (for 500  
msec or longer)  
LCD with back light for indicating CO in-  
stantaneous value, O2 instantaneous  
value, CO instantaneous value after O2  
correction and CO average values after  
O2 correction (1-hour and time-variable  
averagings), peak count value and param-  
eter assignment  
O2 ; 0~25vol%  
• Repeatability: ±0.5% of full scale  
• Zero drift:  
Max. ±2.0% of full scale/week for CO  
Max. ±2.0% of full scale/month for O2  
Max. ±2.0% of full scale/week  
Max. ±1.0% of full scale  
• Span drift:  
• Linearity:  
• Indication:  
• Sample gas extracting rate:  
Approx. 2R/min  
• Response time: Within 90 seconds for 90% indication (af-  
ter extracting sample gas through the  
inlet)  
• Warm-up time: Within 4 hours after power-on  
EDS3-116e  
Date Jun. 30, 2011  
ZSQ  
• Standard requirements for sample gas:  
• Recorder (option):  
• Temperature; 60 to 800˚C  
100 mm-width recorder (Fuji Electric’s  
type PHC or PHE) built in, max. 6 points  
recordable  
• Dust;  
100 mg/Nm3 or less  
• Pressure;  
–2.94 to +2.94kPa  
{–300 to +300mmH2O}  
• Components; SO2 500 ppm or less  
NOx 1000 ppm or less  
• Gas extractor: Electrical heating type (filter built in)  
• Filter mesh; 40µm mesh of SUS 316  
stainless steel  
CO2 0 to 15%  
• Probe; SUS 316 stainless steel of 300,  
400, 600 or 800 mm length  
• Flange; JIS 5K65AFF  
CO  
O2  
0 to 2000 ppm  
0 to 21%  
HCR 1000 ppm or less  
N2, H2O Remaining per-  
cent  
• Mass; Approx. 9 kg  
• Sample inlet tube:  
φ10/φ8 Teflon tube or heating tube (max.  
30 m)  
• Rated operating conditions:  
• Ambient temperature; –5 to +40˚C  
• Ambient humidity; 90% RH or less  
• Power supply voltage; 100 V AC ±10 V  
• Power supply frequency;  
50 or 60 Hz ±0.5 Hz  
• Power consumption; Max. 400 VA  
(without gas extractor)  
• Functions:  
1) O2 correction  
Conversion of measured CO gas concentration into a value  
at standard O2 concentration  
CS(21ON)  
Calculating equation: C =  
21OS  
C; CO concentration after O2 correction  
CS; Measured CO concentration  
OS; Measured O2 concentration (%)  
ON; Standard O2 concentration (12% for  
garbage incinerator)  
• Installation requirements:  
1) Selection of a place which does not  
receive direct sunlight or radiation  
from hot substances  
2) Avoidance of a place under heavy vi-  
bration  
2) Auto calibration  
3) Clean atmospheric air  
• Auto calibration cycle settable range: 1 to 99 hours (1-  
hour step) or 1 to 40 days (1-day step)  
• Auto calibration gas flow time settable range: 60 to 599  
seconds (1-sec step)  
• Auto/manual calibration error contact output: Provided  
when calibration quantity exceeds 50% of full scale.  
• Contact output during auto calibration and maintenance:  
Provided during calibration gas flow and replacement.  
Also provided during maintenance.  
• External dimensions (H x W x D):  
Indoor type  
Outdoor type  
1550 x 735 x 650 mm  
1640 x 745 x 765 mm  
• Mass:  
• Cubicle finish color:  
Munsell 5Y7/1 semi-gloss  
Approx. 140 kg (excluding standard gas)  
• Other:  
Three standard gas (3.4R) cylinders  
accommodable  
• Auto calibration remote start contact input: Calibration  
starts by opening and closing the contact input of 500  
msec or longer in duration.  
3) Output holding function  
Output signal is held during auto calibration and mainte-  
nance.  
4) Alarm  
SCOPE OF DELIVERY  
1 CO instantaneous value upper limit alarm  
2 CO instantaneous value peak count  
Alarm is issued and indicated when CO instantaneous  
value has exceeded the set upper limit by the set number  
of times. (Settable number of times: 1 to 99, settable  
value range: 0 to 1000 ppm)  
• Gas analyzer system  
• Specified external drain separator  
• Specified gas extractor/probe set  
• Specified gas inlet tube set  
• Specified standard gas set (with pressure reducing  
valve)  
5) Other functions  
• Remote range changeover input: Low range is selected  
when closing the contact input, and high range is se-  
lected when opening the contact input  
• Average value resetting input: The output and indication  
of CO average value after O2 correction is reset by open-  
ing the contact after the contact is closed for 1.5 sec or  
longer.  
• Specified recorder to be accommodated in cubicle  
• Standard accessory set  
ORDERING INFORMATION  
• Range discrimination contact output: Contact closed for  
low range  
• Analyzing block error contact output:  
• Temperature input signal: K thermocouple input x 2 (in-  
put for recorder)  
1. Code symbols  
2. Necessity of spares for 1-year measurement  
3. Type of recorder (option)  
4. Anchor bolt  
2
CODE SYMBOLS  
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17  
-
Description  
Z S Q 1  
4 -  
Measurement range (5th code)  
0~200 / 500 ppmCO 0~25% O  
0~200 / 1000 ppmCO 0~25% O  
0~500 / 1000 ppmCO 0~25% O  
0~500 / 2000 ppmCO 0~25% O  
A
B
C
D
2
2
2
2
Power supply (6th code)  
AC 100V 50Hz  
AC 100V 60Hz  
AC 115V 60Hz  
AC 230V 50Hz  
3
4
6
7
Cubicle structure (7th code)  
Indoor type  
Outdoor type  
A
B
C
D
Indoor type  
Outdoor type  
Specify when installation place is at –5˚C  
or lower (–10˚C min.)  
Gas pressure at extraction point (9th code)  
1
2
3
– 0.98~+4.90kPa {– 100~+500mmH  
– 2.94~+2.94kPa {– 300~+300mmH  
– 4.90~+0.98kPa {– 500~+100mmH  
2O}  
2O}  
2
O}  
External drain separator (10th code)  
Not provided  
Provided (Note 1)  
Y
A
Recorder (11th code)  
Y
C
E
Not provided  
With 6-point recorder PHC  
With 2-point recorder PHE  
(Note 2) (Recorder hardware needs to be  
prepared separately.)  
Isolated output (12th code)  
Non-isolated output  
Isolated output (5 outputs)  
0
5
Jpapn quality assurance organization test (13th code)  
0
1
No  
Yes (Test of individual products needs to be ordered separately.)  
Standard gas (14th code)  
1
2
3
4
5
Not provided (space for accommodating three 3.4Rcylinders of air + O  
2
/N  
+ CO/N  
(3.4R, 3 cylinders)(incorporated) with official organization's test  
(3.4R, 2 cylinders)(incorporated)  
+ CO/N (3.4R, 3 cylinders)(incorporated)  
2 + CO/N2)  
2)  
Not provided (space for accommodating two 3.4Rcylinders of O  
Air + O /N + CO/N  
/N + CO/N  
Air + O /N  
2/N  
2
2
2
2
O2  
2
2
2
2
2
Gas extractor flange size (15th code)  
Without gas extractor  
JIS 5K 65A  
Y
A
Gas extractor insertion length (16th code)  
Y
A
B
C
D
Without gas extractor  
300mm  
400mm  
600mm  
800mm  
Sample inlet tube type  
Not provided  
Length (17th code)  
Not provided  
5m  
10m  
15m  
20m  
25m  
30m  
Y
A
B
C
D
E
F
G
H
J
φ10/ φ8 Teflon tube  
φ10/ φ8 Teflon tube  
φ10/ φ8 Teflon tube  
φ10/ φ8 Teflon tube  
φ10/ φ8 Teflon tube  
φ10/ φ8 Teflon tube  
Heating tube  
Heating tube  
Heating tube  
Heating tube  
Heating tube  
10m  
15m  
20m  
25m  
30m  
Specify when the ambient temperature  
may fall to –5˚C or lower (–10˚C at the  
lowest level).  
K
L
(Note 1) Specify this code when the downward inclination of the sample inlet  
tube from the gas extraction point to the analyzer gas inlet is less than 15˚,  
or when the water contents at the gas extraction point exceed 30 vol%.  
(Note 2) Recorder assignment is as follows.  
In 6-point recording:  
In 2-point recording:  
CH1; CO instantaneous value  
CH1; CO instantaneous value  
CH2; O  
2
instantaneous value  
CH2; O instantaneous value  
2
CH3; CO instantaneous value after O  
CH4; 1-hour average CO value after O  
CH5; K thermocouple 1  
2
correction  
correction  
2
CH6; K thermocouple 2  
3
ZSQ  
Gas extractor  
Sampling module  
Gas inlet tube  
15˚ min.  
Membrane filter  
Solenoid  
valve  
CO  
Gas aspirator  
Flowmeter  
Infrared gas  
sensor  
Sample gas inlet,  
φ10/φ8 joint  
(Note 1)  
(Rc 3/8)  
Electronic  
gas cooler  
O
2
Zirconia  
O2 sensor  
Solenoid  
valve  
Gas conditioner  
Gas outlet  
ZERO  
SPAN 1 SPAN 2  
Pressure  
reducing valve  
CO  
/N  
O
/N  
2
Air  
2
2
Standard gas (3.4r)  
Drain  
(Note 2)  
Gas Sampling System Diagram  
Note 1: Solenoid valve for using atmospheric air as zero gas  
Note 2: Unnecessary when atmospheric air is used as zero gas  
Functions of Individual Components  
• Gas extractor: Gas extraction, with heating type stain-  
less steel filter having a standard diam-  
eter of 40µm  
• Standard gas: Reference gas used for calibrating zero  
and span of the analyzer; air and CO for  
span, and O2 for zero.  
• Gas conditioner:  
• Zirconia O2 sensor:  
For separation of drain, prevention of  
drain from being sucked through second-  
ary filter and composite operation of con-  
stant-pressure bubbler.  
Zirconia oxygen sensor used for measur-  
ing the oxygen concentration (0 to 25%)  
in sample gas.  
• Solenoid valve for atmospheric air:  
Can be built in for using the atmospheric  
air instead of standard air.  
• Gas aspirator: For aspiration of sample gas (sample gas  
flow rate approx. 2R/min).  
• Sampling module:  
Incorporates electronic gas cooler, sole-  
noid valve, membrane filter and flowme-  
ter.  
• Electronic gas cooler;  
Dries the moisture in sample gas to a  
dew point of approx. 2˚C.  
• Solenoid valve;  
Used for introducing calibration gas.  
• Membrane filter;  
Glass fiber and PTFE filter used to elimi-  
nate fine dust particles and permit  
monitoring of dust adhering condition  
on the front panel of the gas analyzer.  
• Flowmeter;  
Adjusts and monitors the flow rate of  
sample gas.  
4
Standard Accessories  
No.  
Name  
Type/part No.  
Quantity  
1
2
3
4
5
Filter paper for membrane filter/as spare TK7H6750P1  
Filter paper for membrane filter/as spare TK700735P3  
2 sheets  
2 sheets  
2 fuses  
2 fuses  
3 joints  
Fuse (2A)/as spare  
75718N9  
Fuse (3.2A)/as spare  
Standard gas joint R1/4–φ6  
75718N10  
TK7M6047P1  
(2 pieces for using only 2 standard gases)  
6
Hose band for fixing standard gas  
cylinder  
TK7M2005P2  
12bands  
(8 pieces for using only 2 standard gases)  
7
8
9
Toalon tube for standard gas connection,  
0.45 m and φ9/φ5  
TK727528P1  
415966P4  
1 tube  
1 tube  
1
Polyethylene tube for standard gas  
connection, 5 m and φ6/φ4  
Water bottle for injection into gas  
conditioner  
TK7H3493P1  
TK7H4240P1  
10 Gas extractor flange packing  
11 Gas extractor fastening bolt and nut  
12 O-ring (G50) for gas extractor/as spare  
13 Heating tube support  
1 sheet  
1 set  
1
TK7N0820C1  
INZ-TN4ZSQ  
1 set  
(only when the heating tube has been specified)  
14 Instruction manual  
1 copy  
Spare Parts for 1-Year Measurement  
Type: ZBNISQ12 (without gas extractor)  
Type: ZBNISQ42 (with gas extractor)  
Contents  
• 25-sheet filter paper for membrane filter × 1  
• 2-sheet filter paper for membrane filter × 1  
• Membrane filter O-ring (P49) × 2  
• Membrane filter O-ring (P3) × 2  
• Fuse (2 A) × 3  
• Fuse (3.2 A) × 4  
• Filter for gas conditioner × 2  
• O-ring (G65) for gas conditioner × 2  
• Diaphragm for gas aspirator × 1  
• Valve for gas aspirator × 1  
• Fixed restrictor (φ1) × 1  
• Fixed restrictor (φ1.4) × 1  
• O-ring (G50) for gas extractor × 1  
• O-ring (G45) for gas extractor × 1  
• Packing for gas extractor mesh filter × 1  
• Mesh filter for gas extractor × 1  
when equipped  
with gas extractor  
5
ZSQ  
CONNECTION DIAGRAM  
Terminal block 3  
<TN3>  
Terminal block 2  
<TN2>  
Contact output,  
for atmospheric a  
ir solenoid valve  
Contact output,  
range discrimination  
signal (CO)  
1
2
11  
12  
13  
14  
15  
16  
17  
18  
19  
20  
1
2
11  
12  
13  
14  
15  
16  
17  
18  
19  
20  
+
+
+
For internal Analog input,  
connection  
sensor  
Contact input,  
average value resetting  
O
2
+
3
3
Contact output,  
analyzing block error  
Contact input,  
auto calibration remote start  
4
4
5
5
Contact output,  
calibration error  
(Output holding during  
maintenance)  
For internal connection  
6
6
7
7
Contact output,  
upper limit alarm (CO)  
Contact output,  
pump ON/OFF  
For internal connection  
8
8
9
9
Contact output,  
peak count alarm  
Contact output,  
auto calibration status  
and maintenance status  
Contact input,  
remote range (CO)  
10  
10  
+
(M3.5 screw)  
(M3.5 screw)  
Terminal block 1  
<TN1>  
Terminal block  
TB1  
Terminal block  
TB2 (option for isolated output)  
Analog output,  
Analog output (1-hour  
average CO value after  
1
2
11  
12  
13  
14  
15  
16  
17  
18  
19  
20  
1
2
3
4
5
6
7
8
1
+
+
+
+
+
100 V AC  
main power supply  
1-hour average CO value  
after O  
2
correction  
+
2
O
2
correction)  
Analog output, average CO  
value after O correction  
Analog output,  
3
3
CO average value after  
Ground  
2
O correction  
2
4
4
+
(averaging time variable)  
(averaging time variable)  
Analog output, CO  
instantaneous value after  
5
5
Gas extractor  
power supply  
Analog output,  
CO instantaneous value  
6
6
+
O
2
correction  
after O correction  
2
7
7
Heating tube  
power supply  
Analog output, O  
2
Analog output,  
instantaneous value  
instantaneous value  
O
2
8
8
+
9
+
+
9
9
10  
Temperature input 1  
(option)  
Analog output, CO  
instantaneous value  
Analog output,  
CO instantaneous value  
+
10  
10  
11  
(M3.5 screw)  
(M4 screw)  
Temperature input 2  
(option)  
12  
(M4 screw)  
Notes: 1. Connection to the parenthesized terminals has already been made.  
2. Avoid connection to the terminals for internal connection.  
OUTLINE DIAGRAM (Unit: mm)  
765 (for outdoor installation)  
Sample gas inlet  
(with joint for  
φ10/φ8 tube)  
615  
Anchor plan and door  
opening diagram  
(Rc 3/  
8)  
420  
For outdoor  
installation  
Gas conditioner  
Rear side  
Front side  
3 wiring ports  
with φ34  
grommet  
Door  
Anchor bolt  
(4-M12 × 160 × 50) (option)  
Drain, gas outlet  
(Rc 1/  
4)  
420  
600  
610  
650  
4-φ15  
(135)  
Left side view  
Front view  
6
OUTLINE DIAGRAM (Unit: mm)  
<Gas extractor>  
Rc1/2 Gas outlet with joint for φ10/φ8  
Handle exchange filter  
UP  
Heating tube support  
(added for heating  
tube specifications)  
PL name plate  
f155  
Cover  
Cable hole  
G1/2  
Name plate  
Terminal box  
Rc3/4  
Flange  
JIS 5K 65A FF  
Sampling pipe 20A  
4-φ15  
φ130  
φ155  
Gas inlet  
φ27  
7
ZSQ  
Caution on Safety  
*Before using this product, be sure to read its instruction manual in advance.  
International Sales Div  
Sales Group  
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,  
Shinagawa-ku, Tokyo 141-0032, Japan  
http://www.fujielectric.com  
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425  
http://www.fjielectric.com/products/instruments/  
Information in this catalog is subject to change without notice.  
Printed in Japan  

相关型号:

EDS3-117

PORTABLE GAS ANALYZER
FUJI

EDS3-122I

INFRARED GAS ANALYZER
FUJI

EDS3-123M

NDIR TYPE INFRARED GAS ANALYZER(5-COMPONENT ANALYZER)
FUJI

EDS3-124C

IN-SITU ZIRCONIA OXYGEN ANALYZER
FUJI

EDS3-125B

PARAMAGNETIC OXYGEN ANALYZER
FUJI

EDS3-126A

NDIR TYPE INFRARED GAS ANALYZER
FUJI

EDS3-127C

THERMAL CONDUCTIVITY GAS ANALYZER
FUJI

EDS3-128

CO/O2 GAS ANALYZER FOR STACK GAS
FUJI

EDS3-130F

COMPACT TYPE GAS ANALYZER
FUJI

EDS3-132C

COMPACT TYPE GAS ANALYZER
FUJI

EDS3-133F

INFRARED GAS ANALYZER
FUJI

EDS3-134A

INFRARED GAS ANALYZER FOR HEAT TREATMENT FURNACES
FUJI