CO/O
DATA SHEET
2
GAS ANALYZER FOR STACK GAS
ZSQ
This instrument is equipped with a non-dispersive infrared
CO sensor and a zirconia O2 sensor. It is used for measur-
ing CO and O2 in the gas emitted from a garbage incinera-
tor to meet Japanese regulation on emission control of
dioxins (tetrachlorodibenzoparadioxins).
CO instantaneous value, O2 instantaneous value, CO instan-
taneous value after O2 correction and CO average values
after O2 correction (1-hour and time-variable averagings) can
be output, and a peak count/alarm function is incorporated.
FEATURES
1. Suitable functions for emission control of dioxins.
• Output includes CO instantaneous value, O2 instantaneous
value, CO instantaneous value after O2 correction and
CO average values after O2 correction (1-hour and time-
variable averagings).
• Count/alarm function against excess over specified value
2. Space-saving configuration by unitizing the analyzer and
sampling device into 19-inch types.
3. Easy to maintain with auto calibration and with “From
the front maintenance construction”.
4. A microjet recorder (Fuji Electric’s type PHC or PHE) can
be built in for recording measured gas concentration and
temperature values.
• Output signals: Each signal within range from 4 to 20 mA
DC
Non-isolated (isolated output available at
option)
CO instantaneous value
O2 instantaneous value
CO instantaneous value after O2 correc-
tion
1-hour moving average CO value after O2
correction
Moving average CO value after O2 cor-
rection (averaging time variable)
Time setting in 1 to 59 minutes (1-minute
increment) or 1 to 4 hours (1-hour incre-
ment)
SPECIFICATIONS
Allowable load resistance 550 Ω or less
• Contact output: 1a no-voltage contact (contact capacity
250 V AC, 2 A)
• Measuring system:
Non-dispersive infrared absorption (NDIR)
method for CO
Zirconia method for O2
• Measurable component and range:
CO ; 0~200/500ppm
Peak count alarm, CO instantaneous
value upper limit alarm, auto calibration
and maintenance statuses, calibration er-
ror, analyzing block error and range dis-
crimination
0~200/1000ppm
• Contact input: No-voltage contact (ON/0V, OFF/5 V DC,
DC 5mA at ON)
0~500/1000ppm
0~500/2000ppm
Auto calibration remote start (for 500
msec or longer), remote range
changeover
Remote average value resetting (for 500
msec or longer)
LCD with back light for indicating CO in-
stantaneous value, O2 instantaneous
value, CO instantaneous value after O2
correction and CO average values after
O2 correction (1-hour and time-variable
averagings), peak count value and param-
eter assignment
O2 ; 0~25vol%
• Repeatability: ±0.5% of full scale
• Zero drift:
Max. ±2.0% of full scale/week for CO
Max. ±2.0% of full scale/month for O2
Max. ±2.0% of full scale/week
Max. ±1.0% of full scale
• Span drift:
• Linearity:
• Indication:
• Sample gas extracting rate:
Approx. 2R/min
• Response time: Within 90 seconds for 90% indication (af-
ter extracting sample gas through the
inlet)
• Warm-up time: Within 4 hours after power-on
EDS3-116e
Date Jun. 30, 2011
ZSQ
• Standard requirements for sample gas:
• Recorder (option):
• Temperature; 60 to 800˚C
100 mm-width recorder (Fuji Electric’s
type PHC or PHE) built in, max. 6 points
recordable
• Dust;
100 mg/Nm3 or less
• Pressure;
–2.94 to +2.94kPa
{–300 to +300mmH2O}
• Components; SO2 500 ppm or less
NOx 1000 ppm or less
• Gas extractor: Electrical heating type (filter built in)
• Filter mesh; 40µm mesh of SUS 316
stainless steel
CO2 0 to 15%
• Probe; SUS 316 stainless steel of 300,
400, 600 or 800 mm length
• Flange; JIS 5K65AFF
CO
O2
0 to 2000 ppm
0 to 21%
HCR 1000 ppm or less
N2, H2O Remaining per-
cent
• Mass; Approx. 9 kg
• Sample inlet tube:
φ10/φ8 Teflon tube or heating tube (max.
30 m)
• Rated operating conditions:
• Ambient temperature; –5 to +40˚C
• Ambient humidity; 90% RH or less
• Power supply voltage; 100 V AC ±10 V
• Power supply frequency;
50 or 60 Hz ±0.5 Hz
• Power consumption; Max. 400 VA
(without gas extractor)
• Functions:
1) O2 correction
Conversion of measured CO gas concentration into a value
at standard O2 concentration
CS(21−ON)
Calculating equation: C =
21−OS
C; CO concentration after O2 correction
CS; Measured CO concentration
OS; Measured O2 concentration (%)
ON; Standard O2 concentration (12% for
garbage incinerator)
• Installation requirements:
1) Selection of a place which does not
receive direct sunlight or radiation
from hot substances
2) Avoidance of a place under heavy vi-
bration
2) Auto calibration
3) Clean atmospheric air
• Auto calibration cycle settable range: 1 to 99 hours (1-
hour step) or 1 to 40 days (1-day step)
• Auto calibration gas flow time settable range: 60 to 599
seconds (1-sec step)
• Auto/manual calibration error contact output: Provided
when calibration quantity exceeds 50% of full scale.
• Contact output during auto calibration and maintenance:
Provided during calibration gas flow and replacement.
Also provided during maintenance.
• External dimensions (H x W x D):
Indoor type
Outdoor type
1550 x 735 x 650 mm
1640 x 745 x 765 mm
• Mass:
• Cubicle finish color:
Munsell 5Y7/1 semi-gloss
Approx. 140 kg (excluding standard gas)
• Other:
Three standard gas (3.4R) cylinders
accommodable
• Auto calibration remote start contact input: Calibration
starts by opening and closing the contact input of 500
msec or longer in duration.
3) Output holding function
Output signal is held during auto calibration and mainte-
nance.
4) Alarm
SCOPE OF DELIVERY
1 CO instantaneous value upper limit alarm
2 CO instantaneous value peak count
Alarm is issued and indicated when CO instantaneous
value has exceeded the set upper limit by the set number
of times. (Settable number of times: 1 to 99, settable
value range: 0 to 1000 ppm)
• Gas analyzer system
• Specified external drain separator
• Specified gas extractor/probe set
• Specified gas inlet tube set
• Specified standard gas set (with pressure reducing
valve)
5) Other functions
• Remote range changeover input: Low range is selected
when closing the contact input, and high range is se-
lected when opening the contact input
• Average value resetting input: The output and indication
of CO average value after O2 correction is reset by open-
ing the contact after the contact is closed for 1.5 sec or
longer.
• Specified recorder to be accommodated in cubicle
• Standard accessory set
ORDERING INFORMATION
• Range discrimination contact output: Contact closed for
low range
• Analyzing block error contact output:
• Temperature input signal: K thermocouple input x 2 (in-
put for recorder)
1. Code symbols
2. Necessity of spares for 1-year measurement
3. Type of recorder (option)
4. Anchor bolt
2
CODE SYMBOLS
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17
-
Description
Z S Q 1
4 -
Measurement range (5th code)
0~200 / 500 ppmCO 0~25% O
0~200 / 1000 ppmCO 0~25% O
0~500 / 1000 ppmCO 0~25% O
0~500 / 2000 ppmCO 0~25% O
A
B
C
D
2
2
2
2
Power supply (6th code)
AC 100V 50Hz
AC 100V 60Hz
AC 115V 60Hz
AC 230V 50Hz
3
4
6
7
Cubicle structure (7th code)
Indoor type
Outdoor type
A
B
C
D
Indoor type
Outdoor type
Specify when installation place is at –5˚C
or lower (–10˚C min.)
Gas pressure at extraction point (9th code)
1
2
3
– 0.98~+4.90kPa {– 100~+500mmH
– 2.94~+2.94kPa {– 300~+300mmH
– 4.90~+0.98kPa {– 500~+100mmH
2O}
2O}
2
O}
External drain separator (10th code)
Not provided
Provided (Note 1)
Y
A
Recorder (11th code)
Y
C
E
Not provided
With 6-point recorder PHC
With 2-point recorder PHE
(Note 2) (Recorder hardware needs to be
prepared separately.)
Isolated output (12th code)
Non-isolated output
Isolated output (5 outputs)
0
5
Jpapn quality assurance organization test (13th code)
0
1
No
Yes (Test of individual products needs to be ordered separately.)
Standard gas (14th code)
1
2
3
4
5
Not provided (space for accommodating three 3.4Rcylinders of air + O
2
/N
+ CO/N
(3.4R, 3 cylinders)(incorporated) with official organization's test
(3.4R, 2 cylinders)(incorporated)
+ CO/N (3.4R, 3 cylinders)(incorporated)
2 + CO/N2)
2)
Not provided (space for accommodating two 3.4Rcylinders of O
Air + O /N + CO/N
/N + CO/N
Air + O /N
2/N
2
2
2
2
O2
2
2
2
2
2
Gas extractor flange size (15th code)
Without gas extractor
JIS 5K 65A
Y
A
Gas extractor insertion length (16th code)
Y
A
B
C
D
Without gas extractor
300mm
400mm
600mm
800mm
Sample inlet tube type
Not provided
Length (17th code)
Not provided
5m
10m
15m
20m
25m
30m
Y
A
B
C
D
E
F
G
H
J
φ10/ φ8 Teflon tube
φ10/ φ8 Teflon tube
φ10/ φ8 Teflon tube
φ10/ φ8 Teflon tube
φ10/ φ8 Teflon tube
φ10/ φ8 Teflon tube
Heating tube
Heating tube
Heating tube
Heating tube
Heating tube
10m
15m
20m
25m
30m
Specify when the ambient temperature
may fall to –5˚C or lower (–10˚C at the
lowest level).
K
L
(Note 1) Specify this code when the downward inclination of the sample inlet
tube from the gas extraction point to the analyzer gas inlet is less than 15˚,
or when the water contents at the gas extraction point exceed 30 vol%.
(Note 2) Recorder assignment is as follows.
In 6-point recording:
In 2-point recording:
CH1; CO instantaneous value
CH1; CO instantaneous value
CH2; O
2
instantaneous value
CH2; O instantaneous value
2
CH3; CO instantaneous value after O
CH4; 1-hour average CO value after O
CH5; K thermocouple 1
2
correction
correction
2
CH6; K thermocouple 2
3
ZSQ
Gas extractor
Sampling module
Gas inlet tube
15˚ min.
Membrane filter
Solenoid
valve
CO
Gas aspirator
Flowmeter
Infrared gas
sensor
Sample gas inlet,
φ10/φ8 joint
(Note 1)
(Rc 3/8)
Electronic
gas cooler
O
2
Zirconia
O2 sensor
Solenoid
valve
Gas conditioner
Gas outlet
ZERO
SPAN 1 SPAN 2
Pressure
reducing valve
CO
/N
O
/N
2
Air
2
2
Standard gas (3.4r)
Drain
(Note 2)
Gas Sampling System Diagram
Note 1: Solenoid valve for using atmospheric air as zero gas
Note 2: Unnecessary when atmospheric air is used as zero gas
Functions of Individual Components
• Gas extractor: Gas extraction, with heating type stain-
less steel filter having a standard diam-
eter of 40µm
• Standard gas: Reference gas used for calibrating zero
and span of the analyzer; air and CO for
span, and O2 for zero.
• Gas conditioner:
• Zirconia O2 sensor:
For separation of drain, prevention of
drain from being sucked through second-
ary filter and composite operation of con-
stant-pressure bubbler.
Zirconia oxygen sensor used for measur-
ing the oxygen concentration (0 to 25%)
in sample gas.
• Solenoid valve for atmospheric air:
Can be built in for using the atmospheric
air instead of standard air.
• Gas aspirator: For aspiration of sample gas (sample gas
flow rate approx. 2R/min).
• Sampling module:
Incorporates electronic gas cooler, sole-
noid valve, membrane filter and flowme-
ter.
• Electronic gas cooler;
Dries the moisture in sample gas to a
dew point of approx. 2˚C.
• Solenoid valve;
Used for introducing calibration gas.
• Membrane filter;
Glass fiber and PTFE filter used to elimi-
nate fine dust particles and permit
monitoring of dust adhering condition
on the front panel of the gas analyzer.
• Flowmeter;
Adjusts and monitors the flow rate of
sample gas.
4
Standard Accessories
No.
Name
Type/part No.
Quantity
1
2
3
4
5
Filter paper for membrane filter/as spare TK7H6750P1
Filter paper for membrane filter/as spare TK700735P3
2 sheets
2 sheets
2 fuses
2 fuses
3 joints
Fuse (2A)/as spare
75718N9
Fuse (3.2A)/as spare
Standard gas joint R1/4–φ6
75718N10
TK7M6047P1
(2 pieces for using only 2 standard gases)
6
Hose band for fixing standard gas
cylinder
TK7M2005P2
12bands
(8 pieces for using only 2 standard gases)
7
8
9
Toalon tube for standard gas connection,
0.45 m and φ9/φ5
TK727528P1
415966P4
1 tube
1 tube
1
Polyethylene tube for standard gas
connection, 5 m and φ6/φ4
Water bottle for injection into gas
conditioner
TK7H3493P1
TK7H4240P1
10 Gas extractor flange packing
11 Gas extractor fastening bolt and nut
12 O-ring (G50) for gas extractor/as spare
13 Heating tube support
1 sheet
1 set
1
TK7N0820C1
INZ-TN4ZSQ
1 set
(only when the heating tube has been specified)
14 Instruction manual
1 copy
Spare Parts for 1-Year Measurement
Type: ZBNISQ12 (without gas extractor)
Type: ZBNISQ42 (with gas extractor)
Contents
• 25-sheet filter paper for membrane filter × 1
• 2-sheet filter paper for membrane filter × 1
• Membrane filter O-ring (P49) × 2
• Membrane filter O-ring (P3) × 2
• Fuse (2 A) × 3
• Fuse (3.2 A) × 4
• Filter for gas conditioner × 2
• O-ring (G65) for gas conditioner × 2
• Diaphragm for gas aspirator × 1
• Valve for gas aspirator × 1
• Fixed restrictor (φ1) × 1
• Fixed restrictor (φ1.4) × 1
• O-ring (G50) for gas extractor × 1
• O-ring (G45) for gas extractor × 1
• Packing for gas extractor mesh filter × 1
• Mesh filter for gas extractor × 1
when equipped
with gas extractor
5
ZSQ
CONNECTION DIAGRAM
Terminal block 3
<TN3>
Terminal block 2
<TN2>
Contact output,
for atmospheric a
ir solenoid valve
Contact output,
range discrimination
signal (CO)
1
2
11
12
13
14
15
16
17
18
19
20
–
1
2
11
12
13
14
15
16
17
18
19
20
–
+
–
+
–
+
For internal Analog input,
connection
sensor
Contact input,
average value resetting
O
2
+
3
3
Contact output,
analyzing block error
Contact input,
auto calibration remote start
4
4
5
5
Contact output,
calibration error
(Output holding during
maintenance)
For internal connection
6
6
7
7
Contact output,
upper limit alarm (CO)
Contact output,
pump ON/OFF
For internal connection
8
8
9
9
–
Contact output,
peak count alarm
Contact output,
auto calibration status
and maintenance status
Contact input,
remote range (CO)
10
10
+
(M3.5 screw)
(M3.5 screw)
Terminal block 1
<TN1>
Terminal block
TB1
Terminal block
TB2 (option for isolated output)
Analog output,
Analog output (1-hour
average CO value after
1
2
11
12
13
14
15
16
17
18
19
20
1
2
3
4
5
6
7
8
1
–
–
+
–
+
–
+
–
+
–
+
100 V AC
main power supply
1-hour average CO value
after O
2
correction
+
2
O
2
correction)
Analog output, average CO
value after O correction
Analog output,
3
3
–
CO average value after
Ground
2
O correction
2
4
4
+
–
(averaging time variable)
(averaging time variable)
Analog output, CO
instantaneous value after
5
5
Gas extractor
power supply
Analog output,
CO instantaneous value
6
6
+
–
O
2
correction
after O correction
2
7
7
Heating tube
power supply
Analog output, O
2
Analog output,
instantaneous value
instantaneous value
O
2
8
8
+
–
9
+
–
+
–
9
9
10
Temperature input 1
(option)
Analog output, CO
instantaneous value
Analog output,
CO instantaneous value
+
10
10
11
(M3.5 screw)
(M4 screw)
Temperature input 2
(option)
12
(M4 screw)
Notes: 1. Connection to the parenthesized terminals has already been made.
2. Avoid connection to the terminals for internal connection.
OUTLINE DIAGRAM (Unit: mm)
765 (for outdoor installation)
Sample gas inlet
(with joint for
φ10/φ8 tube)
615
Anchor plan and door
opening diagram
(Rc 3/
8)
420
For outdoor
installation
Gas conditioner
Rear side
Front side
3 wiring ports
with φ34
grommet
Door
Anchor bolt
(4-M12 × 160 × 50) (option)
Drain, gas outlet
(Rc 1/
4)
420
600
610
650
4-φ15
(135)
Left side view
Front view
6
OUTLINE DIAGRAM (Unit: mm)
<Gas extractor>
Rc1/2 Gas outlet with joint for φ10/φ8
Handle exchange filter
UP
Heating tube support
(added for heating
tube specifications)
PL name plate
f155
Cover
Cable hole
G1/2
Name plate
Terminal box
Rc3/4
Flange
JIS 5K 65A FF
Sampling pipe 20A
4-φ15
φ130
φ155
Gas inlet
φ27
7
ZSQ
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fjielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan