EDS3-132C [FUJI]
COMPACT TYPE GAS ANALYZER; 紧凑型气体分析仪型号: | EDS3-132C |
厂家: | FUJI ELECTRIC |
描述: | COMPACT TYPE GAS ANALYZER |
文件: | 总12页 (文件大小:135K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
COMPACT TYPE GAS ANALYZER
DATA SHEET
ZSVS
OVERVIEW
The compact type gas analyzer with built-in pump and filter
is intended for heat treat furnace, plant cultivation, and re-
search-purpose chemical analysis. With the gas extractor,
either simplified measurement probe (non-fixed type) or con-
tinuous measurement probe (fixed type) is selectable at
option. Since a high-sensitivity single-beam mass flow con-
troller is adopted for the infrared sensor, long-term stability
and maintainability are excellent.
FEATURES
1. Gas concentrations of 4 components measurable simul-
taneously and continuously (Note).Concentrations of max.
3 gas components among CO2, CO and CH4 are continu-
ously measurable with the infrared sensor, and that of
O2 is continuously measurable with galvanic cell type
oxygen sensor.
• Measurable component and min./max. measuring range:
CO2; 0 to 200 ppm / 0 to 100 %
CO; 0 to 200 ppm / 0 to 100 %
CH4; 0 to 1000 ppm / 0 to 100 %
O2; 0 to 5 % / 0 to 25 %
2. Standard with 3-range analyzer
Max. 4 components measurable includ-
ing O2
Measurement can be performed over a broad range us-
ing the range selector key.
• Number of measuring ranges:
3 ranges
3. Compact type small and easy to use.
The compact body with built-in pump and membrane filter
is easy to carry and install.
• Max. range ratio 1:5
• Warm-up time: 30 min after power-on
Provided with count-down timer indicat-
ing function.
• Analog output: In up to 8 channels.
4 to 20 mA DC or 0 to 1 V DC (linear)
Non-isolated output
Operation is easier because operation keys and display
unit are all gathered in the analyzing block.
4. A rich variety of standard functions incorporated.
Max. 8-channel outputs are allowed including instanta-
neous concentration value, O2-corrected value, O2-cor-
rected moving average value and CP calculation value
outputs.
Allowable load; 4 to 20 mA DC, 550Ω or
less
0 to 1 V DC, 100 kΩ or
more
• Instantaneous value output of each gas
component
(Note) For continuous measurement, the standard require-
ments for sample gas (shown on page 3) need to be
met, and zero/span calibrations and membrane filter
replacement are required periodically.
• Instantaneous value output after O2 cor-
rection (when provided with O2 analyzer)
• Average value output after O2 correc-
tion (when provided with O2 analyzer)
• CP calculation value output (when pro-
vided with CO2 analyzer)
*The channel numbers of indicated value
and output value correspond to each
other one by one.
SPECIFICATIONS
Standard Specifications
• Measuring system:
CO2, CO and CH4; Non-dispersive infra-
red absorption method with single light
source and single beam (single beam
method)
*An exclusive 25-pin cable is standard-
equipped.
O2; Galvanic cell method
• Communication output:
RS-232C Modbus protocol
*Use a commercially available product
(D-sub 9-pin cable).
This product is not explosion-proof. When handling dangerous gas, adequate attention shall be paid.
EDS3-132c
Date
Apr. 1, 2011
ZSVS
• Indicated values:
• Instrument/calibration error indication:
When the instrument or calibration is
Digital 4-digit indication (by LCD with back
light)
abnormal, an error number is indicated
to help analysis of the error.
• O2 correction: Conversion of measured NOx, SO2 and
CO gas concentrations into values at stan-
dard O2 concentration
• Instantaneous values of respective gas
components
• Instantaneous values after O2 correction
(when provided with O2 analyzer)
• Average value after O2 correction (when
provided with O2 analyzer)
Calculating equation;
21–On
21–Os
× Cs
C=
• CP calculation value display (when pro-
vided with CO2 analyzer)
C; Sample gas concentration after O2
correction
Cs; Measured concentration of sample
gas
*The channel numbers of indicated value
and output value correspond to each
other one by one.
• Power supply: Rated voltage;
Os; Measured O2 concentration
On; Standard O2 concentration for conver-
sion (settable within 0 to 19%)
The result of conversion is indicated and
output in a signal simultaneously.
*An Os value of 20% or more is taken as
20% for calculation.
100 to 115 V AC or 200 to 240 V AC
Working voltage;
85 to 132 V AC or 180 to 264 V AC
*Depending on customer’s code selection.
Rated frequency; 50/60 Hz
Max. rated power; 150 VA
Inlet; Class 1 type conforming with EN60320
• Averaging after O2 correction;
• Operating conditions:
The result of O2 correction is subjected
to moving average for the determined
period of time. And the result of averag-
ing is indicated and output in a signal si-
multaneously.
Average value will be taken at a cycle of
30 sec. (Indication and output are updated
every 30 sec.)
Ambient temperature; 0 to 40˚C
Ambient humidity; 90% RH or less
*Condensation unallowable
• Storage conditions:
Ambient temperature; -20 to 60˚C
Ambient humidity; 95% RH or less
*Condensation unallowable.
• External dimensions (H × W × D mm):
• Resetting of output average value:
211 × 365 × 527
Indication and output of average value are
cleared in response to resetting.
• Weight:
• Finish color:
Approx. 12 kg
Cover; White pearl mica
Base; Medium gray metallic
*Effective only when average value se-
lection is specified in CODE SYMBOLS.
• CP calculation: The carbon potential of carburizing fur-
nace and conversion furnace are calcu-
lated using furnace temperature (fixed
input value) and CO concentration value
(fixed or measured value) while referring
to CO2 measured value.
• Enclosure design:
Casing made of steel plates for indoor
installation.
• Gas-contacting part materials:
Gas inlet/outlet; Polypropylene
Sample cell; SUS304/neoprene rubber
Transparent window: CaF2
Internal pipes: Toalon tube/Teflon tube
CPS×(PCO)2
K1×PCO2
CP=
Calculation equation;
• Gas inlet/outlet:φ6/φ3 hose end
• Purge gas flow rate:
1 L/min (to be purged as required)
where,
CPS ; Saturated carbon concentration
(partial pressure)
Standard Functions
0.0028t–1.30 (800˚C
0.0030t–1.47 (850˚C
0.0034t–1.85 (950˚C
; Furnace temperature
850˚C)
950˚C)
1000˚C)
• Zero gas flow time:
Z180 to 999 sec (settable in 1-sec step)
• Auto indication off:
Indication automatically turns off when
t
PCO ; CO concentration value (partial
pressure)
PCO2 ; CO2 concentration value (partial
pressure)
no key is operated for the determined
period of time in the standby status.
Light off time; OFF/ON (1 to 30 min)
(settable in 1-min step)
K1 ; Constant K1=10 (9.06–15966/T)
• Output holding:At calibration during measurement, out-
put holds the value just before the cali-
bration according to hold setting. In the
standby status, output will not be held.
Indication will not be held either.
T
; Rankine temperature
(tx9/5+32+46˚)
Hold setting; OFF/ON
• Key lock:
None of the set values can be changed
when key lock is turned ON.
This is helpful for reducing operation er-
rors and wrong inputs.
2
Options
Performance
• Gas extractor: Used for aspirating sample gas.
Non-fixed type; Since this type is used
for intermittent measure-
• Repeatability: Within ±0.5% of full scale
• Linearity:
• Zero drift:
• Span drift:
Within ±2% of full scale
Within ±1% of full scale/day
Within ±1% of full scale/day
ment, it cannot be fixed.
Material;
SUS304/polypropylene
• Response time: 90% response time: Within 50 sec Gal-
vanic cell type O2 analyzer: Within 3 min
• Other gases' influence:
Fixed type; Used for continuous measure-
ment. Flange 5K25A FF
Sampling pipe length select-
able among 300, 400, 600 and
800mm
Sample
CO
2
analyzer
CO analyzer
O
2
analyzer CH4 analyzer
Interfer-
ence compo-
nent concentration
component/
range
200ppm max 500ppm min 200ppm max 500ppm min All ranges All ranges
Within ±2%
Within ±2%
Within ±2%
Within ±2%
Within ±2% Within ±2%
Within ±2% Within ±2%
NO
1000ppm
Material; SUS316
SO
CO
CO
2
1000ppm
15%
• Sample inlet tube:
Used for delivering gas from the extrac-
tor to sampling block.
φ6/φ4 × 5 m or φ6/φ4 × 10 m
Material; Teflon
- - Within ±3% Within ±3% Within ±2% Within ±5%
2
Within ±2%
Within ±2%
Within ±2%
- - Within ±2% Within ±2%
1000ppm
1000ppm
50ppm
Shape;
Within ±2%
Within ±2%
Within ±2%
-
CH4
Within ±2% Within ±2%
NH3
Installation Requirements
Within ±3% Within ±2% Within ±3% Within ±2% Within ±2% Within ±2%
H2O 2˚C saturatio
• Selection of a place which does not receive direct sun-
light, rain, wind nor radiation from hot substances.
If such a place cannot be found, a roof or cover should be
prepared for protection.
• Avoidance of a place under heavy vibration
• Selection of a place where atmospheric air is clean
• Discharge of exhaust gas into atmospheric air at a safe
location
Standard Requirements for Sample Gas
• Flow rate:
0.5 L/min ±0.2 L/min
• Temperature: 0 to 40˚C at inlet of sampling gas
10 to 70˚C at tip of non-fixed type probe
(available at option)
70 to 400˚C at tip of fixed type probe (avail-
able at option)
• Avoidance of use in an explosion-proof area
• Pressure:
0 to 2 kPa (Gas shall be discharged into
atmospheric air.)
Scope of Delivery
• Dust:
• Mist:
10 mg/Nm3 or less
• Gas analyzer system
• Standard accessories (Refer to the table at top Following
Unallowable
table.)
• Instruction manual
• Corrosive gas: HCl 10 ppm or less
Others Unallowable
• Standard gas for calibration:
Zero gas; N2 or clean air
Items to be Prepared Separately
However, clean air cannot be
• Standard gas (ZBM) and pressure regulator (ZBD)
• Recorder (when necessary, Fuji's product type PHR)
used if CO2 and O2 are included
in sample gas components.
Span gas; Concentration limited within 90
to 100% of the range of each
sample gas component.
Standard Accessories
Unusable at concentrations
beyond 100%.
Name
Quantity
2 pcs
Tubular fuse (2A)(for analyzing block)
Either one
pair
Power cord (for domestic use, for 100/115V AC) (2m) × 2
Power cord (for North American use, for 100/115V AC) (2m) × 2
Power cord (for European use, for 200/220V AC) (2m) × 2
*
Depending on
customer's code
selection.
Grounding cable (5m)
1 cables
1 cable
5 sheets
1 tube
Output signal cable (1m)
Filter paper (glass fiber) for membrane filter
Connection tube (5m)
Instruction manual (in Japanese or English)
1 copy
Note) Standard accessories include consumables for 6 months.
3
ZSVS
Spare Parts for 1-Year Measurement
Name
Quantity Ordering No.
Filter paper (glass fiber) for membrane
filter
1 pc
(25 sheets) TK700735P2
Large O-ring for membrane filter
Small O-ring for membrane filter
Diaphragm unit for pump
1 pc 8553765
1 pc TK733572P1
1 unit TK713248P1
Specify the ordering No. and the quantity when placing an order.
Other
• A galvanic cell type oxygen sensor has a service life of
about 18 months from the date of its delivery. Periodic
replacement is recommended.
Replacement part ordering No. : TK7M3502C1
Principle Diagram of Infrared Type Measurement (CO
2
, CO, CH )
4
Front expansion chamber
Gas inlet
Gas outlet
Infrared-ray
light source
Rear expansion
chamber
Detector
Mass flow
sensor
Motor
Measurement cell
Chopper
Preamplifier
Indication
Signal processing
block
Output
Principle Diagram of Galvanic Cell Type Measurement (O )
2
Resistor
Thermistor
Electrolyte
Negative electrode
Diaphragm
Positive electrode
4
CODE SYMBOLS
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19 20
Digit No.
of code
Digit
4
Description
note
Z S V S
Y 1
Y
Y
< Specification >
Analyzing block
S
5
< Sample components (CO
1-component analyzer
CO
2, CO, CH4) >
B
D
E
CO
CH
2-component analyzer (1st component + 2nd component)
2
4
CO
CH
CO
2
+CO
G
H
J
4
+CO
2
+CH4
3-component analyzer (1st component + 2nd component + 3rd component
)
CH
With out
Other
4
+(CO
2
+CO)
S
Y
Z
note 2
6
< Sample component (O2) and measuring range >
Galvanic cell type oxygen analyzer/0 to 5%/10%/25%
With out
1
Y
1
8
9
< Revision code >
< Power supply >
For domestic use 100 to 115V AC, 50/60Hz
For European use 200 to 240V AC, 50/60Hz
For North American use 100 to 115V AC, 50/60Hz
note 3
note 4
1
2
3
10 < Measuring range (1st component) >
0 to 200ppm/500ppm/1000ppm
0 to 500ppm/1000ppm/2000ppm
0 to 1000ppm/2000ppm/5000ppm
0 to 2000ppm/5000ppm/1%
0 to 5000ppm/1%/2%
A
B
C
D
E
F
0 to 1%/2%/5%
G
H
J
0 to 2%/5%/10%
0 to 5%/10%/20%
0 to 10%/20%/50%
K
Y
0 to 20%/50%/100%
With out
note 2
note 4
11 < Measuring range (2nd component) >
0 to 200ppm/500ppm/1000ppm
0 to 500ppm/1000ppm/2000ppm
0 to 1000ppm/2000ppm/5000ppm
0 to 2000ppm/5000ppm/1%
0 to 5000ppm/1%/2%
A
B
C
D
E
F
G
H
J
0 to 1%/2%/5%
0 to 2%/5%/10%
0 to 5%/10%/20%
0 to 10%/20%/50%
0 to 20%/50%/100%
With out
K
Y
note 2
note 4
12 < Measuring range (3rd component) >
0 to 200ppm/500ppm/1000ppm
0 to 500ppm/1000ppm/2000ppm
0 to 1000ppm/2000ppm/5000ppm
0 to 2000ppm/5000ppm/1%
0 to 5000ppm/1%/2%
A
B
C
D
E
F
G
H
J
0 to 1%/2%/5%
0 to 2%/5%/10%
0 to 5%/10%/20%
0 to 10%/20%/50%
0 to 20%/50%/100%
With out
K
Y
note 2
5
ZSVS
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19 20
Digit No.
of code
Digit
Description
note
Z S V
Y 1
Y
Y
14 < Output >
0 to 1 V DC, non-isolated
4 to 20 mA DC, non-isolated
15 < Output type >
Instantaneous value after O
1
2
note 5,7
2
correction
0
1
2
Y
Average value after O
CP calculation value
With out
2 correction
note 10
note 6
17 < Language >
Japanese
English
1
2
18 < Gas extractor >
Non-fixed type (for intermittent measurement)
Fixed type (for continuous measurement), flange 5K25A, L = 300 mm
Fixed type (for continuous measurement), flange 5K25A, L = 400 mm
Fixed type (for continuous measurement), flange 5K25A, L = 600 mm
Fixed type (for continuous measurement), flange 5K25A, L = 800 mm
With out
1
2
3
4
5
Y
19 < Sample inlet tube >
note 8
A
B
C
Y
5m×φ6/φ4, Teflon
10m×φ6/φ4, Teflon
20m×φ6/φ4, Teflon
With out
20 < Adjustment >
note 9
Standard adjustment
Adjustment for heat treatment furnace
Other
A
B
Z
Note 1) A parenthesized sample component stands for the 2nd optical system.
Note 2) Specify code Y when only O analyzer is needed.
2
Note 3) Between "1", "2" and "3" of the 9th digit, the rated voltage and plug shape of the attached power cord are different.
“1”: For domestic use, rated voltage 125V AC (PSE), plug shape North American type
“2”: For European use, rated voltage 250V AC (ECC), plug shape European type
“3”: For North American use, rated voltage 125V AC (UL), plug shape North American type
Note 4) For possible combinations of sample component and measuring range, refer to the following tables
(on pages 8 and 9).
Note 5) Specify this code when "1" or "2" is specified at the 6th digit.
Note 6) When "Y" is specified at the 6th digit, "Y" should also be specified at the 15th digit.
Note 7) The kind of output after O
SO and CO is specified.
2 correction will be added to all target components only when an analyzer for NOx,
2
Note 8) Sample inlet tube should be connected within 20 m.
Note 9) Calibration curve varies with gas components contained in sample gas.
"A ; standard adjustment" stands for adjustment in N
"B ; adjustment for heat treatment furnace" is applied to CO analyzer and CO
CO analyzer: CO range gas + 25% CO + 31% H /N
CO analyzer: CO range gas + 5% CO + 31% H /N
When "Z; other" is specified, a gas composition table should be attached.
Note 10) Can be manufactured only when “ CO analyzer” is selected for the 5th digit.
2 balance.
2
analyzer.
2
2
2
2
2
2
2
2
6
Tables of Sample Component and Measuring Range - Availability Check Tables -
Table 1: 1-Component Analyzer (CO2, CO, CH4)
Sample component
CO2 analyzer CO analyzer CH4 analyzer
D
B
E
Range
A
0 to 200/500/1000ppm
0 to 500/1000/2000ppm
0 to 1000/2000/5000ppm
0 to 2000/5000ppm/1%
0 to 5000ppm/1/2%
0 to 1/2/5%
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
-
-
○
○
○
○
○
○
○
○
B
C
D
E
F
G
H
J
0 to 2/5/10%
0 to 5/10/20%
0 to 10/20/50%
K
0 to 20/50/100%
○:Product available
Table 2: 2-Component Analyzer (CO2 analyzer + CO analyzer)
CO analyzer range
Range values are the same as those of CO2 analyzer.
A
B
C
D
E
F
G
H
J
K
CO
A
2
analyzer range
0 to 200/500/1000ppm
○
○
-
-
-
○
○
○
-
-
○
○
-
○
○
○
○
○
○
○
-
○
○
○
○
○
○
○
○
○
-
-
○
○
○
○
○
○
○
○
-
-
-
○
○
○
○
○
○
○
-
-
-
-
○
○
○
○
○
○
-
-
-
-
○
○
○
○
○
○
-
-
-
-
○
○
○
○
○
○
-
-
-
-
-
○
○
○
○
○
-
-
-
-
-
-
○
○
○
○
B
C
D
E
F
0 to 500/1000/2000ppm
0 to 1000/2000/5000ppm
0 to 2000/5000ppm/1%
0 to 5000ppm/1/2%
0 to 1/2/5%
G
H
J
0 to 2/5/10%
0 to 5/10/20%
0 to 10/20/50%
K
0 to 20/50/100%
○:Product available
Table 3: 2-Component Analyzer (CH4 analyzer + CO analyzer)
CO analyzer range
Range values are the same as those of CH4 analyzer.
A
B
C
D
E
F
G
H
J
K
CH
A
4
analyzer range
0 to 200/500/1000ppm
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
○
○
○
○
-
-
-
-
○
○
○
○
○
○
-
-
-
-
○
○
○
○
○
○
-
-
-
-
○
○
○
○
○
○
-
-
-
-
○
○
○
○
○
○
-
-
-
-
-
○
○
○
○
○
-
-
-
-
-
○
○
○
○
○
-
-
-
-
-
-
○
○
○
○
-
-
-
-
-
-
○
○
○
○
B
C
D
E
F
0 to 500/1000/2000ppm
0 to 1000/2000/5000ppm
0 to 2000/5000ppm/1%
0 to 5000ppm/1/2%
0 to 1/2/5%
G
H
J
0 to 2/5/10%
0 to 5/10/20%
0 to 10/20/50%
K
0 to 20/50/100%
○:Product available
Table 4: 2-Component Analyzer (CO2 analyzer + CH4 analyzer)
CH4 analyzer
range
Range values are the same as those of CO2 analyzer.
A
B
C
D
E
F
G
H
J
K
CO2
analyzer range
A
0~200/500/1000ppm
0~500/1000/2000ppm
0~1000/2000/5000ppm
0~2000/5000ppm/1%
0~5000ppm/1/2%
0~1/2/5%
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
○
○
○
○
○
○
○
-
-
○
○
○
○
○
○
○
○
-
-
-
○
○
○
○
○
○
○
-
-
-
-
○
○
○
○
○
○
-
-
-
-
-
○
○
○
○
○
-
-
-
-
-
-
○
○
○
○
-
-
-
-
-
-
-
○
○
○
B
C
D
E
F
G
H
J
0~2/5/10%
0~5/10/20%
0~10/20/50%
K
0~20/50/100%
○:Product available
• 3-component analyzer (CH4 analyzer + CO2 analyzer + CO analyzer);
Possible range in combination of Table 1 (CH4 analyzer) and Table 2 (CO2 analyzer + CO analyzer)
7
ZSVS
One optical line (1 to 3-component gas sampling system diagram): Type ZSVS
Gas extractor
Exhaust
gas port
FL
O2
analyzer
Measurement
cell
Calibration
gas inlet
Purge gas
port
Filter
M
P
Sample gas
inlet
Membrane
filter
Pump
Two optical lines (3 to 4-component gas sampling system diagram): Type ZSVS
Gas extractor
Exhaust
gas port
FL
O
2
analyzer
Measurement
cell
Measurement
cell
Calibration
gas inlet
Purge gas
port
Filter
M
P
Sample gas
inlet
Membrane
filter
Pump
8
OUTLINE DIAGRAM (Unit : mm)
(1) Analyzing block
<Top view>
<Side view>
211
HANDLE
194
PORTABLE ANALYZER
365
<Front view>
<Rear view>
HEAT EXHAUST PORT
OUTPUT CONNECTOR
FLOW CHECKER
DISPLAY
OPERATING UNIT
MEMBRNE FILTER
PORTABLE ANALYZER
SYSTEM 1
EARTHING TERMINAL
POWER SOCKET
RUBBUR FEEY
POWER SWITCH
COMMUNICATION CONNECTOR
FUSE
PURGE GAS INLET φ6 HOSE END
EXHAUST GAS INLET φ6 HOSE END
CALIBRATION GAS INLET φ6 HOSE END
Weight Approx. 12kg
SAMPLE GAS INLET φ6 HOSE END
9
ZSVS
(2) External connection diagrams
Caution) Between male (P) and female (S) connectors, pin numbers are different.
Connect them properly with utmost care.
13
25
1
<Analog output>
On the analyzer side, a female connector (DB-25S-T-NR
made by Japan Aviation Electronics Industry) is attached.
For connection, the furnished cables (1 m)(DB-25P) should
be used.
14
Color of furnished cable
Orange Red Brown Black White Gray Purple Blue Green Yellow Orange Red Brown
CH7+ -CH6+ -CH5+ -CH4+ -CH3+ -CH2+ -CH1+
Pin name
Pin No.
13 12 11 10
25 24 23 22 21 20 19 18 17 16 15 14
NC
9
8
7
6
5
4
3
2
1
Pin No.
Pin name
-CH8+ -CH7
Blue Green Yellow
Color of furnished cable
<Communication output>
1
6
5
9
On the analyzer side, a male connector (DE-9P-T-NR made
by Japan Aviation Electronics Industry) is attached.
For connection, commercially available cross cables (DE-
9S) should be used.
Pin name
Pin No.
NC
RXD TXD
NC GND
1
2
3
4
5
Pin No.
6
7
8
9
Pin name
NC NC NC NC
10
OUTLINE DIAGRAM (Unit : mm)
(3) Power cord and signal cable
• Power cord for domestic and North American use (North American type), rated voltage 125V AC.
Note: The standards for domestic and North American use are different, but the shape is the same.
(2000)
• Power cord for European use (European type), rated voltage 250 V AC
(2000)
• Output cable
Output channel
Connector
+CH4-
+CH5-
+CH6-
+CH7-
+CH8-
+CH1-
+CH2-
+CH3-
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16
Cable color Brown Red Orange Yellow Green Blue Purple Gray White Black Brown Red Orange Yellow Green Blue
Details of area A
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Area A
• Control input/output cable
11
ZSVS
OUTLINE DIAGRAM (Unit : mm)
(4) Non-fixed type gas extractor
(640)
Gas inlet
(
145)
150
°
Sampling pipe
Probe grip
Probe cap
Specifications
Main materials of exhaust gas block: SUS304,
polypropylene, chloroprene rubber
Weight: 0.5 kg
φ6.5/φ4 hose band
(5) Fixed type gas extractor
Attached joint
Sampling pipe 15A
(φ21.7)
1
RC /2
4-φ12
Gas inlet
Attached joint (for φ6 pipe connection)
Flange
JIS 5K 25A FF
95
1
R /2
φ4
25
Specifications
Material : Teflon
Main materials of gas-contacting parts: SUS316, Teflon
Weight: 1 kg
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fjielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan
相关型号:
©2020 ICPDF网 联系我们和版权申明