EDS6-126 [FUJI]

OPTICAL DIFFERENTIAL PRESSURE TRANSMITTER REMOTE SEAL TYPE; 光差压变送器远传密封式
EDS6-126
型号: EDS6-126
厂家: FUJI ELECTRIC    FUJI ELECTRIC
描述:

OPTICAL DIFFERENTIAL PRESSURE TRANSMITTER REMOTE SEAL TYPE
光差压变送器远传密封式

文件: 总8页 (文件大小:118K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
FFI SYSTEM  
OPTICAL DIFFERENTIAL PRESSURE TRANSMITTER  
REMOTE SEAL TYPE  
DATA SHEET  
FFM…3  
The Model FFM 3 Optical Differential Pressure Transmit-  
ter Remote Seal Type measures pressures of various fluids  
accurately, converts them into optical digital signals and  
outputs them. This is an intelligent transmitter providing  
excellent performance and functions due to incorporation  
of electrostatic capacitance type silicon sensor and micro-  
processor.  
A fiber optic cable used for the signal transmission line  
forms an optical field instrumentation system together with  
an optical star coupler and a master station.  
FEATURES  
1. Resistive to noise and lightning  
Optical signal ensures a reliable signal transmission, be-  
cause it is not affected by external noise and inductive  
lightning. Use of a nonmetallic optical (fiber) cable pre-  
vents propagation of inductive lightning through the cable,  
so a signal transmission immune to lightning can be real-  
ized.  
2. Reliability due to redundant configuration  
Host system can be duplicated by using two optical cable  
trunk lines (between an optical star coupler and host sys-  
tem). This enhances reliability of users’ systems.  
3. Intrinsic safety type explosion-proof  
Process temperature, Allowable pressure limit:  
13th  
code  
digit  
Process  
temperature pressure limit  
Allowable  
Fill-fluid  
Fluorolube oil  
Silicon oil  
Silicon oil  
Silicon oil  
Silicon oil  
Silicon oil  
W, A, D  
-
-
20 to 120˚C Atmospheric pressure  
or higher  
H
15 to 250˚C  
J
85 to 300˚C  
Y, G  
S
-
-
40 to 120˚C 2.7 kPa abs or higher  
See Fig. 1.  
15 to 250˚C  
Each equipment with a built-in battery can be constructed  
so as to be an intrinsic safety type individually (intrinsic  
safety type barrier unnecessary).  
T
85 to 300˚C  
0.13 kPa abs or higher  
See Fig. 2.  
Silicon oil  
K
-15 to 200˚C  
For small bores 40A, 50A, 1-1/2B, 2B:  
13th  
code  
digit  
Process  
temperature pressure limit  
Allowable  
SPECIFICATIONS  
Fill-fluid  
-
-
20 to 120˚C  
0 to 250˚C  
40 to 120˚C  
0 to 250˚C  
W, A, D  
Fluorolube oil  
Silicon oil  
Atmospheric pressure  
or higher  
Functional specifications  
H
Fluids measured: Liquid, gas or steam  
Measuring range:  
Y, G  
S
Silicon oil  
2.7 kPa abs or higher  
See Fig. 1.  
Silicon oil  
Span [kPa]  
Range limit [kPa]  
Self-diagnosis: Displayed on indication unit (option) and  
Maximum  
Lower  
Upper  
Minimum  
value  
Type  
transmitted to master station.  
value  
range limit range limit  
Item  
Host system Indication unit  
FFM□□3  
FFM□□5  
FFM□□6  
0.8  
32  
-
32  
32  
Measuring range abnormal  
Detecting unit failure  
Amplifier abnormal  
Battery voltage  
3.25  
12.5  
130  
500  
-130  
130  
500  
-500  
Operating pressure:  
Up to the maximum value of  
operating pressure of flange.  
Battery voltage low alarm  
Remote control function:  
See Table 1.  
Output signal: Optical digital signal  
Power supply: Built-in lithium battery  
(expected life about 4 years)  
EDS6-126  
Date May 20, 2004  
FFM···3  
Optical cable:  
Code set type, silica fiber ... core/clad di-  
ameter 100/140 µm  
Remarks:  
(1) This is an output change when the process  
pressure receiving unit and the transmitter  
body are at the same height and temperature.  
Optical connector:  
FC connector  
Transmission distance:  
1.5 km max. (when transmission loss of  
(2) Error is larger when there is a temperature  
difference among the process pressure  
receiving unit, capillary and transmitter body.  
optical cable is 4 dB/km)  
Variable from 0.2 to 32 sec (time constant)  
Ambient temperature effect:  
Damping:  
Zero elevation and suppression:  
For small bores 40A, 50A, 1-1/2B, 2B:  
Zero shift: ±0.7% / 28°C (x equal to 1/2  
URL or more)  
Possible within ± 100% of maximum  
URL  
2x  
span.  
Zero shift: ±0.7  
% / 28°C (x less  
Explosion-proof: Intrinsic safety type, JIS ib IIC T3  
Ambient temperature:  
than 1/2 URL)  
Overall shift: ±0.9% / 28°C (x equal to  
1/2 URL or more)  
Overall shift: ±(0.4 + 0.5  
28°C (x less than 1/2 URL)  
Note 1: Condition; Capillary length is  
limited to max. 3 m. With a  
capillary of 5 m long, the effect  
is 1.5 times as large as the  
above.  
Note 2: The effect is 2.5 times as large  
as the above when the 7th  
code digit (material) is other  
than V, A, B, C and D.  
30 to +70°C  
10 to +60°C for intrinsic safety explo-  
sion-proof type  
20 to +70°C when provided with indica-  
URL  
)% /  
2x  
tor  
10 to +60°C when filled with fluorolube  
oil  
10 to +70°C for silicone oil codes H, S  
and K  
+20 to +70°C for silicone oil codes J and  
T
For small bores 40A, 50A, 1-1/2B, 2B:  
15 to +65°C  
10 to +60°C for intrinsic safety explo-  
sion-proof type  
15 to +65°C when provided with indica-  
Overrange effect:  
Zero shift at maximum span  
±0.1% / nominal pressure of flange  
2 times as large as above when  
the 7th digit (material) is other  
than V, A, B, C and D. (2.5 times  
as large as above for small bores  
40A, 50A, 1-1/2B and 2B)  
tor  
10 to +60°C when filled with fluorolube  
oil  
10 to +60°C for silicone oil codes H and  
S.  
Static pressure effect:  
Zero shift at maximum span  
±0.2% /1MP:  
2 times as large as above when  
Storage temperature:  
40 to +80°C  
the 7th digit (material) is other  
than V, A, B, C and D.  
(2.5 times as large as above for  
small bores 40A, 50A, 1-1/2B and  
2B)  
Performance specifications  
±0.2% when measuring span is 1/10 or more  
of maximum span  
Accuracy rating(Note)  
max. span  
measuring span  
Change of measurement span:  
±(0.1 + 0.01  
)% when  
-
0.2±0.2% / 1MPa  
measuring span is less than 1/10 of maximum  
span.  
±0.2% / 1MPa for small bore 40A,  
50A, 1-1/2B and 2B specifications  
Note: Percent value with respect to measuring span (including  
Measurement period: 0.2 sec  
linearity, hysteresis and repeatability in standard 23°C status)  
Dead time  
[sec]  
Time constant  
*[sec]  
For small bores 40A, 50A, 1-1/2B, 2B:  
Type  
Response time:  
FFM□□3  
5
FFM□□  
6
2
±0.25% when measuring span is 1/10 or more  
Accuracy rating(Note)  
About 0.2  
of maximum span  
max. span  
measuring span  
1.7  
±(0.17 + 0.008  
)% when  
(Note *) Value at capillary length 1.5 m and 23°C  
measuring span is less than 1/10 of maximum  
span.  
Physical specifications  
Flange material: SUS304  
Material:  
Ambient temperature effect:  
Zero shift: ±(0.5  
URL  
x
)% / 28°C  
Other wetted parts  
SUS316  
Hastelloy-C�  
Monel�  
Tantalum  
Titanium�  
Zirconium  
URL  
Seal diaphragm  
SUS316L  
Material code  
V, A, B, C, D  
Overall shift: ±(0.7  
Where; URL: Maximum span  
x : Measuring span  
2 times as large as above when the 7th  
digit (material) is other than V, A, B, C and  
D.  
)% / 28°C  
x
H, F, G, K, L  
Hastelloy-C�  
Monel�  
Tantalum  
M
T
P
R
Titanium�  
Zirconium  
Note 1: Selected according to combination of type codes.  
Refer to CODE SYMBOLS.  
2
Capillary:  
Finish:  
Stainless steel pipe coated with PVC  
Epoxy-polyurethane double coat,  
Color: silver (blue for amplifier case cover)  
Fill-fluid: Silicon oil (Code: Y,G)  
Fill-fluid: Silicon oil (Code: S)  
[kPa abs]  
Fill-fuid: Silicon oil (Code: T)  
Environmental protection:  
Meets JIS C0920, immersion-proof  
(equivalent to IEC IP67 or NEMA 6/6P)  
External dimensions:  
With  
small bore  
101  
Operating area  
53.3  
See OUTLINE DIAGRAM.  
13 to 30 kg  
Mass:  
Optical cable connection:  
G1/2 or 1/2 -14NPT (whichever selected  
by code symbol)  
Process connection:  
JIS specifications;  
10K, 20K, 30K-40A, 50A  
13.3  
2.7  
Non-operating area  
10K, 30K-80A, 100A  
ANSI/JPI specifications;  
-40  
-15  
0
60 85 120  
Process temperature  
200 230250  
300  
C]  
[
°
150LB, 300LB, 1-1/2B, 2B, 3B, 4B  
Diaphragm extension:  
0, 50, 100, 150 or 200 mm (as specified)  
Mounting method:  
Note: For use at a vacuum level, the transmitter body should be  
mounted below the flange section mounting position.  
Fig. 1 Relation between process temperature and  
operating pressure  
U-bolt mounting to a 50A (2B) pipe. De-  
tection unit is mounted with flange or  
between flanges (wafer type).  
Fill-fluid: Silicon oil (Code: K)  
[kPa abs]  
101  
Optional specifications  
Indication unit: 5-digit LCD indication, % or real scale in-  
dication (as specified by code symbol)  
Operating temperature range: 20 to  
+70°C  
Operating area  
Oxygen oil-proof processing:  
Varies with  
material.  
Refer to  
CODE  
Fluorolube filled.  
Wetted parts degreased  
and cleaned.  
Non-  
operating  
area  
Chlorine service:Fluorolube oil filled.  
0.27  
0.13  
SYMBOLS.  
-
15  
100  
200  
[°C]  
Process temperature  
Table 1 Remote Control Function  
(Items readable and setting from hand-held  
communicator)  
Fig. 2 Relation between process temperature and  
operating pressure  
Item  
Reading Setting  
Description  
Maximum range  
Maximum measuring range  
of equipment  
Measuring range  
Damping  
Real scale  
Actual measuring range  
Variable within 0.2 to 32 sec.  
Indication in industrial value  
indication  
Battery voltage  
Error indication  
Battery voltage of equipment  
Errors of detection unit and  
amplifier  
Measured value  
Adjustment  
Measured data  
Zero and span adjustment  
Note: For operation of the 3type transmitter (3at the 8th digit  
of product code), a hand-held communicator is required to  
have a version 1.6 or higher, but a communicator before  
version 1.6 can be operated with memory data updated.  
(Refer to the instruction manual of transmitter.)  
3
FFM···3  
SYSTEM BLOCK DIAGRAM  
1:1 system  
Host system  
1 to 5V DC  
4 to 20mA DC  
Master station  
Optical fiber type  
electric converter  
(Type; PSN)  
Control room  
Field  
Star coupler  
Transmitter  
Transmitter  
Field equipment  
(max. 8 units)  
Transmission distance: 1.5 km max.  
(between master station and field equipment)  
Transmission distance: 4 km max.  
4
CODE SYMBOLS  
1
2
3
4
5
6
7
8
9 10111213  
Description  
F F  
M
3
F
Cable lead-in port (4th digit)  
G1/2  
1/2-14NPT  
S
T
Flange (5th digit)  
Code usable  
Code usable  
at 7th digit  
Rating  
Material  
at 6th digit  
3,5,6  
3,5,6  
3,5,6  
3,5,6  
3,5,6  
3,5,6  
3,5,6  
3,5,6  
5,6  
0
1
2
3
4
5
6
7
JIS 10K 80A  
JIS 10K 100A  
JIS 30K 80A  
JIS 30K 100A  
ANSI/JPI 150LB 3B  
ANSI/JPI 150LB 4B  
ANSI/JPI 300LB 3B  
ANSI/JPI 300LB 4B  
JIS 10K 40A  
All codes usable  
P and R unusable  
All codes usable  
P and R unusable  
All codes usable  
P and R unusable  
All codes usable  
P and R unusable  
V,H,M,T  
SUS304  
A
B
C
D
E
F
JIS 10K 50A  
JIS 20K 40A  
JIS 20K 50A  
JIS 30K 40A  
5,6  
5,6  
5,6  
5,6  
5,6  
5,6  
5,6  
5,6  
V,A to D,H,M,T  
V,H,M,T  
V,A to D,H,M,T  
V,H,M,T  
V,A to D,H,M,T  
V,H,M,T  
V,A to D,H,M,T  
V,H,M,T  
V,A to D,H,M,T  
All codes usable  
P and R unusable  
V,H,M,T  
JIS 30K 50A  
1
J
L
ANSI/JPI 150LB 1 /  
2
B
B
ANSI/JPI 150LB 2B  
1
M
N
P
Q
S
R
ANSI/JPI 300LB 1 /  
2
ANSI/JPI 300LB 2B  
For JIS 80A, ANSI/JPI 3B 3,5,6  
5,6  
Wafer type  
(without flange)  
For JIS100A, ANSI/JPI 4B 3,5,6  
1
For JIS 40A, ANSI/JPI 1 /  
2
B
5,6  
For JIS 50A, ANSI/JPI 2B 5,6  
V,A to D,H,M,T  
Measuring span (6th digit)  
0.8......... 32kPa  
3.25...... 130kPa  
3
5
6
12.5........ 500kPa  
Material and diaphragm extension (7th digit)  
Seal diaphragm  
SUS316L  
SUS316L  
SUS316L  
SUS316L  
Other wetted parts Diaphragm extension (mm)  
V
A
B
C
D
H
F
G
K
L
SUS316  
0
SUS316  
SUS316  
SUS316  
SUS316  
Hastelloy-C  
Hastelloy-C  
Hastelloy-C  
Hastelloy-C  
Hastelloy-C  
Monel  
50  
100  
150  
200  
0
50  
100  
150  
200  
0
Only 5th codes 1, 3, 5, 7 and Q are  
selectable when these codes  
are used in combination  
SUS316L  
with 13th codes S, T or K.  
Hastelloy-C  
Hastelloy-C  
Hastelloy-C  
Hastelloy-C  
Hastelloy-C  
Monel  
Tantalum  
Titanium  
Zirconium  
M
T
P
R
Tantalum  
Titanium  
Zirconium  
0
0
0
Selectable when 5th digit is  
0, 2, 4, 6 or P.  
Indicator (9th digit)  
Not provided  
Digital, % indication  
Digital, real scale indication  
A
L
P
Explosion-proof (10th digit)  
Non-explosion proof  
Intrinsic safety, JIS  
A
G
5
FFM···3  
1
2
3
4
5
6
7
8
9 10111213  
Description  
F FM  
3
F
Capillary length (11th digit)  
1.5m  
3m  
5m  
A
B
C
G
H
J
6m.....Selectable when 5th digit is 0 to 7, P or Q.  
7m  
Selectable when 5th digit is 0 to 7, P or Q  
8m  
and 13th digit is Y, W, G, A or D.  
10m  
K
Treatment and fill-fluid (13th digit)  
Treatment  
None  
None  
Fill-fluid  
Silicon oil  
Fluorolube oil  
Silicon oil  
Y
W
G
Degreasing  
Oxygen oil-proof processing Fluorolube oil.....when 7th digit is V, A, B, C or D  
A
D
H
Chlorine service  
None  
Fluorolube oil.....when 7th digit is H, F, G, K, L or T  
Silicon oil (for high temperature)  
None  
None  
None  
None  
Silicon oil (for high temperature)  
when 7th digit is  
V, A, B, C or D  
Note 1 J  
S
Note 1 T  
Note 1 K  
Silicon oil (for high temperature and vacuum)  
Silicon oil (for high temperature and vacuum)  
Silicon oil (for high temperature and high vacuum)  
Note 1: Inapplicable for small bores 40A, 50A, 1-1/2B and 2B.  
6
OUTLINE DIAGRAM (Unit : mm)  
0 6 E  
1 7 F  
2 A J  
3 B L  
4 CM  
5 D N  
Low pressure side  
S
FFM  
3
S: Capillary length  
Mounting pipe  
S
High pressure side  
φF  
241  
JIS 50A (2B)  
Inside diameter  
of gasket seat  
80  
108  
70  
φK  
φJ  
φH  
U-bolt  
(M8)  
306  
54  
N-φh  
15  
hole  
Indicator (option)  
B
28 27  
Optical cable lead-in port  
φG  
47  
φ15  
E
Type  
FFMS  
FFMT  
G1/2  
1/2-14NPT  
Type  
φF φG φH φJ φK  
t
P
N-φh  
FLANGE  
E
FFM*0**3 185 150 126 100 73 38 116 8-19 JIS-10K-80A  
FFM*1**3 210 175 151 103 96 38 141 8-19 JIS-10K-100A  
FFM*2**3 210 170 126 100 73 48 116 8-23 JIS-30K-80A  
FFM*3**3 240 195 151 103 96 52 141 8-25 JIS-30K-100A  
FFM*4**3 191 152.5 126 100 73 44 116 4-20 ANSI/JPI 150LB 3B  
FFM*5**3 229 190.5 151 103 96 44 141 8-20 ANSI/JPI 150LB 4B  
FFM*6**3 210 168 126 100 73 49 116 8-23 ANSI/JPI 300LB 3B  
FFM*7**3 254 200 151 103 96 52 141 8-23 ANSI/JPI 300LB 4B  
FFM*A**3 140 105 84 49 48 36 71 4-19 JIS-10K-40A  
FFM*B**3 155 120 84 49 48 36 71 4-19 JIS-10K-50A  
FFM*C**3 140 105 84 49 48 38 71 4-19 JIS-20K-40A  
FFM*D**3 155 120 84 49 48 38 71 8-19 JIS-20K-50A  
FFM*E**3 160 120 84 49 48 42 71 4-23 JIS-30K-40A  
Details of "B"  
Type  
Weight (kg)  
15 to 20.5  
L
0
FFM***V3  
FFM***H3  
FFM***M3  
FFM***T3  
FFM***P3  
FFM***R3  
FFM***A3  
50  
16 to 31.5  
16.5 to 32  
17 to 32.5  
17.5 to 33  
FFM***F3  
FFM***B3  
FFM***G3  
100  
150  
200  
FFM*F**3 165 130 84 49 48 42 71 8-19 JIS-30K-50A  
1
FFM***C3  
FFM*J**3 127 98.4 84 49 48 37.5 71 4-16 ANSI/JPI 150LB 1 /  
2
B
B
FFM***K3  
FFM*L**3 152 120.6 84 49 48 39.5 71 4-20 ANSI/JPI 150LB 2B  
1
FFM***D3  
FFM***L3  
FFM*M**3 156 114.3 84 49 48 41 71 4-23 ANSI/JPI 300LB 1 /  
FFM*N**3 165 127 84 49 48 42.5 71 8-20 ANSI/JPI 300LB 2B  
Note) For S (capillary length), refer to CODE SYMBOLS.  
2
7
FFM···3  
P
Q
S
R
FFM  
3
241  
80  
108  
70  
306  
20  
54  
15  
Indicator (option)  
28 27  
47  
B
Optical cable lead-in port  
2-M6, depth 9  
Type  
FFM*P**3  
FFM*Q**3  
FFM*R**3  
FFM*S**3  
φH φJ  
P
FLANGE  
126 100 116 For JIS-80A, ANSI/JPI 3B  
151 103 141 For JIS-100A, ANSI/JPI 4B  
84 49 71 For JIS-50A, ANSI/JPI 2B  
1
84 49 71 For JIS-40A, ANSI/JPI 1 /  
2
B
SCOPE OF DELIVERY  
Instrument body and pipe fixture (as specified)  
ORDERING INFORMATION  
1. Model type  
2. Measuring range  
3. Indication scale for real scale specification  
4. Others  
Caution on Safety  
*Before using this product, be sure to read its instruction manual in advance.  
Head Office  
6-17, Sanbancho, Chiyoda-ku, Tokyo 102-0075, Japan  
http://www.fesys.co.jp/eng  
Sales Div.  
International Sales Dept.  
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan  
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187  
http: //www.fic-net.jp/eng  
Printed in Japan  
Information in this catalog is subject to change without notice.  

相关型号:

EDS6-128C

EQUALIZING VALVE, OVAL FLANGE
FUJI

EDS6-132F

HYBRID ULTRASONIC FLOWMETER (PULSE">(Pulse Doppler method Transit Time method)
FUJI

EDS6-141D

ULTRASONIC FLOWMETER(2-Path System type)
FUJI

EDS6-33B

FCX SERIES DIFFERENTIAL PRESSURE TRANSMITTER
FUJI

EDS6-34F

FCX SERIES REMOTE SEAL TYPE DIFFERENTIAL PRESSURE TRANSMITTER
FUJI

EDS6-35D

FCX SERIES FLOW TRANSMITTER
FUJI

EDS6-50I

DIFFERENTIAL PRESSURE DEVICE ORIFICE PLATE WITH CARRIER RINGS
FUJI

EDS6-51E

DIFFERENTIAL PRESSURE DEVICE ORIFICE PLATES WITHOUT CARRIER RINGS
FUJI

EDS6-52G

DIFFERENTIAL PRESSURE DEVICE ORIFICE PLATES WITH FLANGE TAPPINGS
FUJI

EDS6-54E

SINGLE-PART DIFFERENTIAL PRESSURE DEVICE ORIFICE PLATE
FUJI

EDS6-58F

PORTABLE TYPE ULTRASONIC FLOWMETER(PORTAFLOW)
FUJI

EDS6-61H

COMPACT ELECTROMAGNETIC FLOWMETER(COMPACT FLOW)
FUJI