EDS6-126 [FUJI]
OPTICAL DIFFERENTIAL PRESSURE TRANSMITTER REMOTE SEAL TYPE; 光差压变送器远传密封式型号: | EDS6-126 |
厂家: | FUJI ELECTRIC |
描述: | OPTICAL DIFFERENTIAL PRESSURE TRANSMITTER REMOTE SEAL TYPE |
文件: | 总8页 (文件大小:118K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
FFI SYSTEM
OPTICAL DIFFERENTIAL PRESSURE TRANSMITTER
REMOTE SEAL TYPE
DATA SHEET
FFM…3
The Model FFM 3 Optical Differential Pressure Transmit-
ter Remote Seal Type measures pressures of various fluids
accurately, converts them into optical digital signals and
outputs them. This is an intelligent transmitter providing
excellent performance and functions due to incorporation
of electrostatic capacitance type silicon sensor and micro-
processor.
A fiber optic cable used for the signal transmission line
forms an optical field instrumentation system together with
an optical star coupler and a master station.
FEATURES
1. Resistive to noise and lightning
Optical signal ensures a reliable signal transmission, be-
cause it is not affected by external noise and inductive
lightning. Use of a nonmetallic optical (fiber) cable pre-
vents propagation of inductive lightning through the cable,
so a signal transmission immune to lightning can be real-
ized.
2. Reliability due to redundant configuration
Host system can be duplicated by using two optical cable
trunk lines (between an optical star coupler and host sys-
tem). This enhances reliability of users’ systems.
3. Intrinsic safety type explosion-proof
Process temperature, Allowable pressure limit:
13th
code
digit
Process
temperature pressure limit
Allowable
Fill-fluid
Fluorolube oil
Silicon oil
Silicon oil
Silicon oil
Silicon oil
Silicon oil
W, A, D
-
-
20 to 120˚C Atmospheric pressure
or higher
H
15 to 250˚C
J
ꢀ85 to 300˚C
Y, G
S
-
-
40 to 120˚C 2.7 kPa abs or higher
See Fig. 1.
15 to 250˚C
Each equipment with a built-in battery can be constructed
so as to be an intrinsic safety type individually (intrinsic
safety type barrier unnecessary).
T
ꢀ85 to 300˚C
0.13 kPa abs or higher
See Fig. 2.
Silicon oil
K
-15 to 200˚C
For small bores 40A, 50A, 1-1/2B, 2B:
13th
code
digit
Process
temperature pressure limit
Allowable
SPECIFICATIONS
Fill-fluid
-
-
20 to 120˚C
0 to 250˚C
40 to 120˚C
0 to 250˚C
W, A, D
Fluorolube oil
Silicon oil
Atmospheric pressure
or higher
Functional specifications
H
Fluids measured: Liquid, gas or steam
Measuring range:
Y, G
S
Silicon oil
2.7 kPa abs or higher
See Fig. 1.
Silicon oil
Span [kPa]
Range limit [kPa]
Self-diagnosis: Displayed on indication unit (option) and
Maximum
Lower
Upper
Minimum
value
Type
transmitted to master station.
value
range limit range limit
Item
Host system Indication unit
FFM□□3
FFM□□5
FFM□□6
0.8
32
-
32
32
Measuring range abnormal
Detecting unit failure
Amplifier abnormal
Battery voltage
○
○
○
○
○
○
○
○
-
○
3.25
12.5
130
500
-130
130
500
-500
Operating pressure:
Up to the maximum value of
operating pressure of flange.
Battery voltage low alarm
Remote control function:
See Table 1.
Output signal: Optical digital signal
Power supply: Built-in lithium battery
(expected life about 4 years)
EDS6-126
Date May 20, 2004
FFM···3
Optical cable:
Code set type, silica fiber ... core/clad di-
ameter 100/140 µm
Remarks:
(1) This is an output change when the process
pressure receiving unit and the transmitter
body are at the same height and temperature.
Optical connector:
FC connector
Transmission distance:
1.5 km max. (when transmission loss of
(2) Error is larger when there is a temperature
difference among the process pressure
receiving unit, capillary and transmitter body.
optical cable is 4 dB/km)
Variable from 0.2 to 32 sec (time constant)
Ambient temperature effect:
Damping:
Zero elevation and suppression:
For small bores 40A, 50A, 1-1/2B, 2B:
Zero shift: ±0.7% / 28°C (x equal to 1/2
URL or more)
Possible within ± 100% of maximum
URL
2x
span.
Zero shift: ±0.7
% / 28°C (x less
Explosion-proof: Intrinsic safety type, JIS ib IIC T3
Ambient temperature:
than 1/2 URL)
Overall shift: ±0.9% / 28°C (x equal to
1/2 URL or more)
Overall shift: ±(0.4 + 0.5
28°C (x less than 1/2 URL)
Note 1: Condition; Capillary length is
limited to max. 3 m. With a
capillary of 5 m long, the effect
is 1.5 times as large as the
above.
Note 2: The effect is 2.5 times as large
as the above when the 7th
code digit (material) is other
than V, A, B, C and D.
–30 to +70°C
–10 to +60°C for intrinsic safety explo-
sion-proof type
–20 to +70°C when provided with indica-
URL
)% /
2x
tor
–10 to +60°C when filled with fluorolube
oil
–10 to +70°C for silicone oil codes H, S
and K
+20 to +70°C for silicone oil codes J and
T
For small bores 40A, 50A, 1-1/2B, 2B:
–15 to +65°C
–10 to +60°C for intrinsic safety explo-
sion-proof type
–15 to +65°C when provided with indica-
Overrange effect:
Zero shift at maximum span
±0.1% / nominal pressure of flange
2 times as large as above when
the 7th digit (material) is other
than V, A, B, C and D. (2.5 times
as large as above for small bores
40A, 50A, 1-1/2B and 2B)
tor
–10 to +60°C when filled with fluorolube
oil
–10 to +60°C for silicone oil codes H and
S.
Static pressure effect:
Zero shift at maximum span
±0.2% /1MP:
2 times as large as above when
Storage temperature:
–40 to +80°C
the 7th digit (material) is other
than V, A, B, C and D.
(2.5 times as large as above for
small bores 40A, 50A, 1-1/2B and
2B)
Performance specifications
±0.2% when measuring span is 1/10 or more
of maximum span
Accuracy rating(Note)
max. span
measuring span
Change of measurement span:
±(0.1 + 0.01
)% when
-
0.2±0.2% / 1MPa
measuring span is less than 1/10 of maximum
span.
±0.2% / 1MPa for small bore 40A,
50A, 1-1/2B and 2B specifications
Note: Percent value with respect to measuring span (including
Measurement period: 0.2 sec
linearity, hysteresis and repeatability in standard 23°C status)
Dead time
[sec]
Time constant
*[sec]
For small bores 40A, 50A, 1-1/2B, 2B:
Type
Response time:
FFM□□3
5
FFM□□
6
2
±0.25% when measuring span is 1/10 or more
Accuracy rating(Note)
About 0.2
of maximum span
max. span
measuring span
1.7
±(0.17 + 0.008
)% when
(Note *) Value at capillary length 1.5 m and 23°C
measuring span is less than 1/10 of maximum
span.
Physical specifications
Flange material: SUS304
Material:
Ambient temperature effect:
Zero shift: ±(0.5
URL
x
)% / 28°C
Other wetted parts
SUS316
Hastelloy-C�
Monel�
Tantalum
Titanium�
Zirconium
URL
Seal diaphragm
SUS316L
Material code
V, A, B, C, D
Overall shift: ±(0.7
Where; URL: Maximum span
x : Measuring span
2 times as large as above when the 7th
digit (material) is other than V, A, B, C and
D.
)% / 28°C
x
H, F, G, K, L
Hastelloy-C�
Monel�
Tantalum
M
T
P
R
Titanium�
Zirconium
Note 1: Selected according to combination of type codes.
Refer to CODE SYMBOLS.
2
Capillary:
Finish:
Stainless steel pipe coated with PVC
Epoxy-polyurethane double coat,
Color: silver (blue for amplifier case cover)
Fill-fluid: Silicon oil (Code: Y,G)
Fill-fluid: Silicon oil (Code: S)
[kPa abs]
Fill-fuid: Silicon oil (Code: T)
Environmental protection:
Meets JIS C0920, immersion-proof
(equivalent to IEC IP67 or NEMA 6/6P)
External dimensions:
With
small bore
101
Operating area
53.3
See OUTLINE DIAGRAM.
13 to 30 kg
Mass:
Optical cable connection:
G1/2 or 1/2 -14NPT (whichever selected
by code symbol)
Process connection:
JIS specifications;
10K, 20K, 30K-40A, 50A
13.3
2.7
Non-operating area
10K, 30K-80A, 100A
ANSI/JPI specifications;
-40
-15
0
60 85 120
Process temperature
200 230250
300
C]
[
°
150LB, 300LB, 1-1/2B, 2B, 3B, 4B
Diaphragm extension:
0, 50, 100, 150 or 200 mm (as specified)
Mounting method:
Note: For use at a vacuum level, the transmitter body should be
mounted below the flange section mounting position.
Fig. 1 Relation between process temperature and
operating pressure
U-bolt mounting to a 50A (2B) pipe. De-
tection unit is mounted with flange or
between flanges (wafer type).
Fill-fluid: Silicon oil (Code: K)
[kPa abs]
101
Optional specifications
Indication unit: 5-digit LCD indication, % or real scale in-
dication (as specified by code symbol)
Operating temperature range: –20 to
+70°C
Operating area
Oxygen oil-proof processing:
Varies with
material.
Refer to
CODE
Fluorolube filled.
Wetted parts degreased
and cleaned.
Non-
operating
area
Chlorine service:Fluorolube oil filled.
0.27
0.13
SYMBOLS.
-
15
100
200
[°C]
Process temperature
Table 1 Remote Control Function
(Items readable and setting from hand-held
communicator)
Fig. 2 Relation between process temperature and
operating pressure
Item
Reading Setting
Description
Maximum range
Maximum measuring range
of equipment
Measuring range
Damping
Real scale
Actual measuring range
Variable within 0.2 to 32 sec.
Indication in industrial value
indication
Battery voltage
Error indication
Battery voltage of equipment
Errors of detection unit and
amplifier
Measured value
Adjustment
Measured data
Zero and span adjustment
Note: For operation of the “3” type transmitter (“3” at the 8th digit
of product code), a hand-held communicator is required to
have a version 1.6 or higher, but a communicator before
version 1.6 can be operated with memory data updated.
(Refer to the instruction manual of transmitter.)
3
FFM···3
SYSTEM BLOCK DIAGRAM
1:1 system
Host system
1 to 5V DC
4 to 20mA DC
Master station
Optical fiber type
electric converter
(Type; PSN)
Control room
Field
Star coupler
Transmitter
Transmitter
Field equipment
(max. 8 units)
Transmission distance: 1.5 km max.
(between master station and field equipment)
Transmission distance: 4 km max.
4
CODE SYMBOLS
1
2
3
4
5
6
7
8
9 10111213
Description
F F
M
3
F
Cable lead-in port (4th digit)
G1/2
1/2-14NPT
S
T
Flange (5th digit)
Code usable
Code usable
at 7th digit
Rating
Material
at 6th digit
3,5,6
3,5,6
3,5,6
3,5,6
3,5,6
3,5,6
3,5,6
3,5,6
5,6
0
1
2
3
4
5
6
7
JIS 10K 80A
JIS 10K 100A
JIS 30K 80A
JIS 30K 100A
ANSI/JPI 150LB 3B
ANSI/JPI 150LB 4B
ANSI/JPI 300LB 3B
ANSI/JPI 300LB 4B
JIS 10K 40A
All codes usable
P and R unusable
All codes usable
P and R unusable
All codes usable
P and R unusable
All codes usable
P and R unusable
V,H,M,T
SUS304
A
B
C
D
E
F
JIS 10K 50A
JIS 20K 40A
JIS 20K 50A
JIS 30K 40A
5,6
5,6
5,6
5,6
5,6
5,6
5,6
5,6
V,A to D,H,M,T
V,H,M,T
V,A to D,H,M,T
V,H,M,T
V,A to D,H,M,T
V,H,M,T
V,A to D,H,M,T
V,H,M,T
V,A to D,H,M,T
All codes usable
P and R unusable
V,H,M,T
JIS 30K 50A
1
J
L
ANSI/JPI 150LB 1 /
2
B
B
ANSI/JPI 150LB 2B
1
M
N
P
Q
S
R
ANSI/JPI 300LB 1 /
2
ANSI/JPI 300LB 2B
For JIS 80A, ANSI/JPI 3B 3,5,6
5,6
Wafer type
(without flange)
For JIS100A, ANSI/JPI 4B 3,5,6
1
For JIS 40A, ANSI/JPI 1 /
2
B
5,6
For JIS 50A, ANSI/JPI 2B 5,6
V,A to D,H,M,T
Measuring span (6th digit)
0.8......... 32kPa
3.25...... 130kPa
3
5
6
12.5........ 500kPa
Material and diaphragm extension (7th digit)
Seal diaphragm
SUS316L
SUS316L
SUS316L
SUS316L
Other wetted parts Diaphragm extension (mm)
V
A
B
C
D
H
F
G
K
L
SUS316
0
SUS316
SUS316
SUS316
SUS316
Hastelloy-C
Hastelloy-C
Hastelloy-C
Hastelloy-C
Hastelloy-C
Monel
50
100
150
200
0
50
100
150
200
0
Only 5th codes 1, 3, 5, 7 and Q are
selectable when these codes
are used in combination
SUS316L
with 13th codes S, T or K.
Hastelloy-C
Hastelloy-C
Hastelloy-C
Hastelloy-C
Hastelloy-C
Monel
Tantalum
Titanium
Zirconium
M
T
P
R
Tantalum
Titanium
Zirconium
0
0
0
Selectable when 5th digit is
0, 2, 4, 6 or P.
Indicator (9th digit)
Not provided
Digital, % indication
Digital, real scale indication
A
L
P
Explosion-proof (10th digit)
Non-explosion proof
Intrinsic safety, JIS
A
G
5
FFM···3
1
2
3
4
5
6
7
8
9 10111213
Description
F FM
3
F
Capillary length (11th digit)
1.5m
3m
5m
A
B
C
G
H
J
6m.....Selectable when 5th digit is 0 to 7, P or Q.
7m
Selectable when 5th digit is 0 to 7, P or Q
8m
and 13th digit is Y, W, G, A or D.
10m
K
Treatment and fill-fluid (13th digit)
Treatment
None
None
Fill-fluid
Silicon oil
Fluorolube oil
Silicon oil
Y
W
G
Degreasing
Oxygen oil-proof processing Fluorolube oil.....ꢀwhen 7th digit is V, A, B, C or D
A
D
H
Chlorine service
None
Fluorolube oil.....ꢀwhen 7th digit is H, F, G, K, L or T
Silicon oil (for high temperature)
None
None
None
None
Silicon oil (for high temperature)
when 7th digit is
V, A, B, C or D
Note 1 J
S
Note 1 T
Note 1 K
Silicon oil (for high temperature and vacuum)
Silicon oil (for high temperature and vacuum)
Silicon oil (for high temperature and high vacuum)
Note 1: Inapplicable for small bores 40A, 50A, 1-1/2B and 2B.
6
OUTLINE DIAGRAM (Unit : mm)
0 6 E
1 7 F
2 A J
3 B L
4 CM
5 D N
Low pressure side
S
FFM
3
S: Capillary length
Mounting pipe
S
High pressure side
φF
241
JIS 50A (2B)
Inside diameter
of gasket seat
80
108
70
φK
φJ
φH
U-bolt
(M8)
306
54
N-φh
15
hole
Indicator (option)
B
28 27
Optical cable lead-in port
φG
47
φ15
E
Type
FFMS
FFMT
G1/2
1/2-14NPT
Type
φF φG φH φJ φK
t
P
N-φh
FLANGE
E
FFM*0**3 185 150 126 100 73 38 116 8-19 JIS-10K-80A
FFM*1**3 210 175 151 103 96 38 141 8-19 JIS-10K-100A
FFM*2**3 210 170 126 100 73 48 116 8-23 JIS-30K-80A
FFM*3**3 240 195 151 103 96 52 141 8-25 JIS-30K-100A
FFM*4**3 191 152.5 126 100 73 44 116 4-20 ANSI/JPI 150LB 3B
FFM*5**3 229 190.5 151 103 96 44 141 8-20 ANSI/JPI 150LB 4B
FFM*6**3 210 168 126 100 73 49 116 8-23 ANSI/JPI 300LB 3B
FFM*7**3 254 200 151 103 96 52 141 8-23 ANSI/JPI 300LB 4B
FFM*A**3 140 105 84 49 48 36 71 4-19 JIS-10K-40A
FFM*B**3 155 120 84 49 48 36 71 4-19 JIS-10K-50A
FFM*C**3 140 105 84 49 48 38 71 4-19 JIS-20K-40A
FFM*D**3 155 120 84 49 48 38 71 8-19 JIS-20K-50A
FFM*E**3 160 120 84 49 48 42 71 4-23 JIS-30K-40A
Details of "B"
Type
Weight (kg)
15 to 20.5
L
0
FFM***V3
FFM***H3
FFM***M3
FFM***T3
FFM***P3
FFM***R3
FFM***A3
50
16 to 31.5
16.5 to 32
17 to 32.5
17.5 to 33
FFM***F3
FFM***B3
FFM***G3
100
150
200
FFM*F**3 165 130 84 49 48 42 71 8-19 JIS-30K-50A
1
FFM***C3
FFM*J**3 127 98.4 84 49 48 37.5 71 4-16 ANSI/JPI 150LB 1 /
2
B
B
FFM***K3
FFM*L**3 152 120.6 84 49 48 39.5 71 4-20 ANSI/JPI 150LB 2B
1
FFM***D3
FFM***L3
FFM*M**3 156 114.3 84 49 48 41 71 4-23 ANSI/JPI 300LB 1 /
FFM*N**3 165 127 84 49 48 42.5 71 8-20 ANSI/JPI 300LB 2B
Note) For S (capillary length), refer to CODE SYMBOLS.
2
7
FFM···3
P
Q
S
R
FFM
3
241
80
108
70
306
20
54
15
Indicator (option)
28 27
47
B
Optical cable lead-in port
2-M6, depth 9
Type
FFM*P**3
FFM*Q**3
FFM*R**3
FFM*S**3
φH φJ
P
FLANGE
126 100 116 For JIS-80A, ANSI/JPI 3B
151 103 141 For JIS-100A, ANSI/JPI 4B
84 49 71 For JIS-50A, ANSI/JPI 2B
1
84 49 71 For JIS-40A, ANSI/JPI 1 /
2
B
SCOPE OF DELIVERY
Instrument body and pipe fixture (as specified)
ORDERING INFORMATION
1. Model type
2. Measuring range
3. Indication scale for real scale specification
4. Others
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
Head Office
6-17, Sanbancho, Chiyoda-ku, Tokyo 102-0075, Japan
http://www.fesys.co.jp/eng
Sales Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http: //www.fic-net.jp/eng
Printed in Japan
Information in this catalog is subject to change without notice.
相关型号: