NV1-1G000-M [TDK]

AC-DC电源;
NV1-1G000-M
型号: NV1-1G000-M
厂家: TDK ELECTRONICS    TDK ELECTRONICS
描述:

AC-DC电源

输出元件
文件: 总4页 (文件大小:1097K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
NV-Power NV-175M  
https://emea.tdk-lambda.com/nv  
https://product.tdk.com/en/power/nv  
180-200W Configurable  
Medical power supply.  
Medical  
Industrial  
Test  
Broadcast  
Renewable  
Comms  
Features  
Benefits  
Highꢀefficiency  
Minimisesꢀheatꢀinꢀsystem  
Lowꢀprofile  
Fitsꢀ1Uꢀapplications  
LowꢀearthꢀleakageꢀandꢀClassꢀBꢀEMC  
2ꢀxꢀMOPPsꢀisolation  
Simplifiesꢀsystemꢀdesign,ꢀreducesꢀcost  
Simplifiesꢀsystemꢀdesign  
Lowꢀcostꢀofꢀownership  
3ꢀyearꢀwarranty  
Input  
90-264Vac  
Input Voltage  
Input Frequency  
45 - 63Hz  
(100 - 240Vac nominal)  
EN61000-3-2 compliant  
Fast acting (not user accessible)  
123µA at 120Vac (60Hz), 257µA max at 240Vac (60Hz)  
Input Harmonics  
Input Fuse  
Inrush Current  
<40A at 25°C and 230Vac (cold start)  
Earth Leakage  
Current  
Worst case leakage current is less than 300µA at 264Vac, 63Hz (normal condition, 0.5mA Single Fault Condition)  
Lower leakage versions available, contact sales office for details  
Isolation  
2 x MOPPs (3rd edition 60601)  
Input to Output  
type tested to 4kVac (equivalent to 5.7kVdc), production tested to 4.3kVdc  
4kVac  
1 x MOOP (3rd edition 60601)  
1.5kVac, 2.3kVdc  
Input to Earth  
Output to Earth  
200Vdc  
QUICK SELECTOR - example configurations  
Additionalꢀvariantsꢀavailable,ꢀseeꢀ“HowꢀtoꢀCreateꢀaꢀProductꢀDescription”  
Model  
Ch1  
Ch3  
Ch4  
NV1-1T000-M  
NV1-1G000-M  
NV1-3G0TT-M  
NV1-3G0FF-M  
12V / 15A  
24V / 7.5A  
24V / 7.5A  
24V / 7.5A  
-
-
-
-
12V / 5A  
15V / 5A  
-12V / 1A  
-15V / 1A  
How To Create A Product Description  
Confirm availability of created product with TDK-Lambda  
Blank standard, vertical connector  
Case  
Option  
Connector  
Option  
Right angled connector  
#o/p  
Ch1  
0
Ch32  
Ch43  
NV1-  
-M  
R
(see handbook for ‘R’ connection  
and mechanical details)  
Medical with 4kVac (2 x MOPPs)  
input to output isolation  
Number of outputs  
Letter / number from ‘Available  
Outputs’ table to represent  
output voltage  
Blank no case  
-C U chassis + cover  
-U U chassis  
Ch1, Ch3,  
Ch4  
-F End fan + case1  
-I End fan + case + IEC inlet1  
1. Needs 0H, 3H, 5H, TH or FH type channel 4.  
The fan speed is temperature dependent, ensuring optimum cooling and lowest audible noise.  
2. For negative output channel 3, follow chosen letter by ‘Y’. For example, TY channel 3 = -12V / 5A.  
3. For positive output channel 4, follow chosen letter by ‘P’. For example, TP channel 4 = +12V / 1A.  
NV175M_Datasheet_69924 - v18.0  
1
NV-Power NV-175M  
Available Outputs  
Otherꢀoutputꢀoptionsꢀareꢀavailable,ꢀpleaseꢀcontactꢀsalesꢀofficeꢀwithꢀyourꢀrequirements  
Adjustment  
Range  
Adjustment  
Range  
Adjustment  
Range  
Channel 1  
Channel 2  
Channel 31  
Channel 42  
Fixed  
Fixed  
Fixed  
Fixed  
Fixed  
Fixed  
T
F
-12V / 1A  
-15V / 1A  
T
F
12V / 15A  
15V / 12A  
12 - 15V3  
12 - 15V4  
T
F
G
0
12V / 5A  
15V / 5A  
24V / 2.5A  
Omit  
12 - 15V  
12 - 15V  
18 - 24V  
3HP +3.3V / 2A6  
5HP +5V / 2A6  
TH  
FH  
0H  
0
Not available  
-12V / 2A6  
-15V / 2A6  
Fan supply only  
Omit  
G
24V / 7.5A  
24 - 28V5  
1. Followꢀlettersꢀinꢀredꢀbyꢀ‘Y’ꢀforꢀnegativeꢀoutputꢀchannelꢀ3.  
2. Followꢀlettersꢀinꢀredꢀbyꢀ‘P’ꢀforꢀpositiveꢀoutputꢀchannelꢀ4.  
3. 12ꢀ-ꢀ12.5Vꢀifꢀ24Vꢀchannelꢀ3ꢀfitted.  
4. 14.5ꢀ-ꢀ15Vꢀifꢀ24Vꢀchannelꢀ3ꢀfitted  
5. 24ꢀ-ꢀ26Vꢀifꢀ24Vꢀchannelꢀ3ꢀfitted.  
6. 1.5Aꢀmaxꢀwithꢀ‘-F’ꢀorꢀ‘-I’option.  
Output Specification  
Turn on time  
1.5s max  
at 90Vac and 100% rated output power  
configuration dependent  
at 90Vac  
Efficiency  
up to 90%  
16ms min  
<1%  
Hold up  
Ripple and Noise  
Voltage Accuracy  
Remote Sense  
Minimum Load  
(or 50mV if higher) pk-pk, using EIAJ test method & 20MHz bandwidth  
±1%  
±4% for Channel 4 with ‘T’ or ‘F’ type outputs, +4/-3% for all other channel 4.  
Channel 1. Max 0.5V total line drop  
Yes  
No  
on any output  
Including Line (for 90-264Vac input change), Load (for 0-100% load change) and  
Cross (for 0-100% load change on any other output) regulation.  
1%  
Total Regulation  
(or 50mV if greater)  
of set voltage for 50% load change  
(in 50µs within the range 25-100% load)  
Transient Response  
<4%  
Recovery  
500µs  
Yes  
for recovery to 1% of set voltage  
See Application Notes for details  
Over Voltage Protection  
Short Circuit Protection  
Over Temperature Protection  
Yes  
Yes  
Single output units with 12V, 15V or 24V (T, F or G). Average output power must not  
exceed 180W over any 5 minute period  
Peak Output Power  
Ch1 Good Signal  
200W  
Yes  
Provides a Logic ‘Low’ signal after Channel 1 output is within 90% (±5%) of nominal  
Environment  
0°C to 50°C operational, -40°C to 70°C storage (max 12 months).  
Full load, with either ‘-F’ option fitted or 2m/s air blown from input to output (approximately 10CFM)  
Temperature  
Derating  
50°C to 65°C derate each output by 2.5% per °C  
-20°C  
Low Temp Startup  
Humidity  
5 - 95% RH non condensing  
±3 x 30g shocks in each plane, total 18 shocks  
30g shock = 11ms (+/-0.5msec), half sine  
Conforms to EN60068-2-27, EN60068-2-47, IEC68-2-27, IEC68-2-47, JIS C0041-1987.  
Conforms to MIL-STD-810E/F, Method 516.5, Pro I, IV, VI  
Shock  
Single axis 10 - 500 Hz at 2g (sweep and endurance at resonance) in all 3 planes  
Conforms to EN60068-2-6, IEC68-2-6  
Vibration  
Conforms to MIL-STD-810E, Method 514.4, Pro I, Cat 1,9  
Altitude  
3000 metres operational (4000m for NV1-1G000-M for 60601-1 3rd edition)  
Degree 2, Material group IIIb  
Pollution  
Immunity EN61000-6-2:2005, EN60601-1-2:2007  
Criteria  
Air discharge 15kV, Contact discharge 8kV. Not applicable to open  
frame units  
A
A
A
Electrostatic Discharge  
Electromagnetic Field  
Fast / Burst Transient  
EN61000-4-2  
EN61000-4-3  
EN61000-4-4  
Level 4  
Level 3  
Level 4  
12V/m  
ac input tested to 4.4kV  
dc output tested to 2.2kV  
Common mode - 2.2kV, Differential - 1.1kV  
Surge Immunity  
EN61000-4-5  
EN61000-4-6  
EN61000-4-8  
Level 3  
Level 3  
Level 4  
A
A
A
A
12V  
Conducted RF Immunity  
Power Frequency Magnetic Field  
30A/m  
Criteria B for 5 sec interruption and dips to 40% for 5 cycles below  
154Vac nominal input  
Voltage Dips, Variations, Interruptions EN61000-4-11 Class 3  
Emissions EN61000-6-3:2007, EN60601-1-2:2007  
(as per CISPR.11/22) Class B, FCC47 part 15 subpart B  
see application note for details. Additional filtering required for IEC inlet version.  
Radiated Electric Field  
EN55011, EN55032  
Conducted Emissions  
Conducted Harmonics  
Flicker  
EN55011, EN55032  
EN61000-3-2  
(as per CISPR.11/22) Class B, FCC47 part 15 subpart B  
Class A  
EN61000-3-3  
Compliant - dmax only  
NV175M_Datasheet_69924 - v18.0  
NV-Power NV-175M  
2
Approvals / Accreditations  
Notes  
IEC/EN 62368-1, UL62368-1 / CSA 22.2 No 62368-1  
File E135494  
File E135494  
IEC/EN 60950-1, UL60950-1 / CSA 22.2 No 60950-1  
IEC/EN 60601-1, UL/CSA 60601-1, ANSI/AAMI ES60601-1  
CAN/CSA-C22.2 No 60601-1-08  
File E349607  
IEC/EN61010  
Designed to meet  
Low Voltage Directive (LVD), electromagnetic compatibility  
(EMC) and Restriction of Hazardous Substances (RoHS)  
CE Mark (EN62368-1)  
CB certificate and Report available on request  
Please check with technical sales for status of approvals  
Designed and manufactured under the control of ISO9001 and ISO13485 (including risk management).  
Mating Parts (Molex or Equivalent)  
Conn  
Housing  
Pins  
J1  
39-01-2245  
09-50-8051  
45750-3112  
08-52-0113  
J2  
  
Inputꢀandꢀoutputꢀconnectorsꢀ  
areꢀnotꢀincludedꢀwithꢀtheꢀ  
product.ꢀTheyꢀareꢀavailableꢀ  
fromꢀTDK-Lambda  
  
  
  
  
  
  
Top View  
  
NV175 models without  
cover or chassis  
  
  
  
Part  
Contents  
No  
  
  
  
  
  
  
  
  
94910 1ꢀoffꢀinputꢀconnectorꢀ  
  
  
andꢀ3ꢀcrimps  
  
94668 1ꢀoffꢀoutputꢀ  
connectorꢀandꢀ20ꢀ  
crimps  
  
  
  
  
128.20  
Side View  
End View  
Top View  
NV175-F models  
Top View  
NV175-F models  
Side View  
NV175-C & NV175-U  
models  
Side View  
NV175-F models  
Bottom View  
NV175-C & NV175-U  
models  
Bottom View  
NV175-F models  
Notes: 1. All customer fixings M3  
NV175M_Datasheet_69924 - v18.0  
2. Maximum thread penetration 4.5mm  
3. Maximum torque 0.9Nm  
4. All tolerances +/-0.5mm  
3
NV-Power NV-175M  
TDK-Lambda France SAS  
TDK-Lambda Americas  
Tel: +1 800-LAMBDA-4 or 1-800-526-2324  
powersolutions@us.tdk-lambda.com  
www.us.lambda.tdk.com  
Tel: +33 1 60 12 71 65  
france@fr.tdk-lambda.com  
www.emea.lambda.tdk.com/fr  
Italy Sales Office  
Tel: +39 02 61 29 38 63  
info.italia@it.tdk-lambda.com  
www.emea.lambda.tdk.com/it  
TDK Electronics do Brasil Ltda  
Tel: +55 11 3289-9599  
sales.br@tdk-electronics.tdk.com  
www.tdk-electronics.tdk.com/en  
Netherlands  
info@nl.tdk-lambda.com  
www.emea.lambda.tdk.com/nl  
TDK-Lambda Germany GmbH  
Tel: +49 7841 666 0  
info.germany@de.tdk-lambda.com  
TDK-Lambda Corporation  
Tel: +81-3-6778-1113  
www.jp.lambda.tdk.com  
www.emea.lambda.tdk.com/de  
TDK-Lambda (China) Electronics Co. Ltd.  
Tel: +86 21 6485-0777  
powersolutions@cn.tdk-lambda.com  
www.lambda.tdk.com.cn  
TDK-Lambda Singapore Pte Ltd.  
Tel: +65 6251 7211  
tls.mkt@sg.tdk-lambda.com  
www.sg.lambda.tdk.com  
Austria Sales Office  
Tel: +43 2256 655 84  
info@at.tdk-lambda.com  
www.emea.lambda.tdk.com/at  
TDK India Private Limited, Power Supply Division  
Switzerland Sales Office  
Tel: +41 44 850 53 53  
info@ch.tdk-lambda.com  
www.emea.lambda.tdk.com/ch  
Tel: +91 80 4039-0660  
mathew.philip@in.tdk-lambda.com  
www.sg.lambda.tdk.com  
Nordic Sales Office  
Tel: +45 8853 8086  
info@dk.tdk-lambda.com  
www.emea.lambda.tdk.com/dk  
TDK-Lambda UK Ltd.  
Tel: +44 (0) 12 71 85 66 66  
powersolutions@uk.tdk-lambda.com  
www.emea.lambda.tdk.com/uk  
TDK-Lambda Ltd.  
Tel: +9 723 902 4333  
info@tdk-lambda.co.il  
www.emea.lambda.tdk.com/il  
C.I.S.  
Commercial Support:  
Tel: +7 (495) 665 2627  
Technical Support:  
Tel: +7 (812) 658 0463  
info@tdk-lambda.ru  
www.emea.lambda.tdk.com/ru  
Forꢀadditionalꢀinformation,ꢀpleaseꢀvisitꢀ  
https://product.tdk.com/en/power  
Errorsꢀandꢀomissionsꢀexcepted  
NV175M_Datasheet_69924 - v18.0  
NV-Power NV-175M  
4

相关型号:

NV1-1T000

AC-DC电源
TDK

NV1-1T000-M

AC-DC电源
TDK

NV1-2

AC-DC电源
TDK

NV1-3

AC-DC电源
TDK

NV1-3G0FF-M

AC-DC电源
TDK

NV1-3G0TT-M

AC-DC电源
TDK

NV1-4

AC-DC电源
TDK

NV100D07UCTB2

Varistor, 85V, 6J, Through Hole Mount, RADIAL LEADED
KOA

NV100D07UTB2

Varistor, 85V, 6J, Through Hole Mount, RADIAL LEADED
KOA

NV1100D07UTB2

Varistor, 895V, Through Hole Mount, RADIAL LEADED
KOA

NV1100D14UTB2

Varistor, 895V, Through Hole Mount, RADIAL LEADED
KOA

NV13M08Y

Voltage-Controlled Crystal Oscillator (VCXO)
NDK