INFRARED GAS ANALYZER
DATAꢀSHEET
ZPA
This gas analyzer (ZPA) is capable of measuring the con-
centration of NO,SO2,CO2,CO,CH4 and O2 components in
sample gas. NO, SO2, CO2, CO and CH4 are measured by
non-dispersion infrared method (NDIR), while O2 is mea-
sured by fuel cell or paramagnetic and zirconia method.
Optimum use as an analyzer unit for measurement of
combustion exhaust gas from refuse incinerator, boiler
exhaust or gas from various industrial furnaces.
FEATURES
1.ꢀHighꢀsensitivity
Equipped with the newest upgraded mass flow sensor.
2.ꢀEasyꢀmaintenance
Maintenance is easy due to simple measurement unit of
single-beam system adapted.
3.ꢀSmallꢀandꢀlight
The size is small 133x483x382mm (HxWxD). (This is
for 19inch rack mounting type) and light (9kg). Unit is
capable max. 5 component measurement in one analyzer.
4.ꢀExtensiveꢀfunctions
Various optional functions are available such as auto
calibration control, atmospheric pressure correction, high
and low concentration alarm, remote range switch, and
range identification signal,etc.
Measurableꢀgasꢀcomponentsꢀandꢀmeasuringꢀrange:
Minimum range
0 - 200ppm
0 - 200ppm
0 - 100ppm
0 - 200ppm
0 - 500ppm
Maximum range
0 - 5000ppm
0 - 10vol%
NO
SO2
CO2
CO
0 - 100vol%
0 - 100vol%
0 - 100vol%
CH4
O2
Built in
fuel cell
0 - 10vol%
0 - 25vol%
O2
Built-in
Paramagnetic
0 - 5vol%
None
0 - 100vol%
100 - 95vol%
O2
External
Zirconia
0 - 5vol%
0 - 25vol%
ꢀꢀꢀꢀꢀꢀꢀꢀꢀꢀ
• Max. 5 components measurement includ-
ing O2.
For reverse range O2 measurement, in-
frared gas measurement is not available
(single range O2 only).
• Measuring range ratio max. 1:10 (except
O2)
• Measuring ranges are changeable be-
tween the specified minimum and maxi-
mum range
SPECIFICATIONS
Standard Specifications
Principleꢀofꢀmeasurement:ꢀ
NO, SO2, CO2, CO, CH4 ;
Non-dispersion infrared-ray absorption
method
Settable one range or two ranges
* In measurement range low range is
called first range, high range is called
second range.
• For possible combinations of components
and ranges, refer to Table1.
Single light source and single beams
(single beam system)
;Fuel cell O2 sensor (build-in) or paramag-
netic O2 sensor (build-in) or zirconia O2
sensor (externally installed TYPE: ZFK7)
O2
Measuredꢀvalueꢀindication:ꢀ
Digital indication in 4 digits
(LCD panel with LED back light)
• Instantaneous value of each component
• Instantaneous value after O2 correction
(only in NO, SO2, CO measurement with
O2)
• Average value after O2 correction
(only in NO, SO2, CO measurement with
O2)
• O2 average value
EDS3-143
Date
Nov.ꢀ2,ꢀ2011
ZPA
Analogꢀoutputꢀsignals:ꢀ
Gasꢀinlet/outlet:ꢀRc1/4 or NPT1/4 internal thread
Purgeꢀgasꢀflowꢀrate: 1L/min ( when required)
LifeꢀtimeꢀofꢀfuelꢀcellꢀO2ꢀsensor: 2 years
4 to 20mA DC or 0 to 1V DC, isolated
internally from circuit and ground. Output
lines are non-isolated each other.; 12
outputs max.
Allowable load 550Ω for 4 to 20mA DC
Allowable load 100KΩ for 0 to 1V DC
* Refer to Table2 for the channel No. of
displayed values and analog output
signals.
Standard Functions
Outputꢀsignalꢀholding:ꢀ
Output signals are held unchanged during
manual and auto calibrations by activation
of holding (turning “ON” its setting).
The values held are those just before start
calibration mode or setting value.
It is selectable.
Analogꢀinputꢀsignal:
For signal input from externally installed
O2 sensor.
Signal requirement;
(1) Signal from Fuji’s Zirconia O2 sensor
(TYPE: ZFK7)
(2) 0 to 1V DC from an O2 sensor
Input section is not isolated. This
feature is effective when an O2 sensor
is not built in.
Indication of instantaneous values will
not be held.
Switchꢀranges:ꢀ The switch ranges function is available in
manual, auto, and remote modes. Only
preset switch method is effective.
Manual: Allows range to switch by key operation.
Auto:
Automatically switched from first range to
second range when the measured value
exceeds to 90%FS of first range.
* Externally installed O2 sensor should
be purchased separately.
Digitalꢀoutput:ꢀ(Option)ꢀ
Automatically switched from second range
to first range when the measured value
drops to 80% or less first range.
1c contact (24V DC/1A, resistive load)
max.15 outputs
Instrument error, calibration error,
range identification, auto calibration
status, solenoid valve drive for auto
calibration, High/Low limit alarm con-
tact output.
Remote: Voltage contact input (for measurable
(Option) components)
Allows range to switch via an external
signal when remote range switch input
is received.
When the contact input terminals for
each component are input voltage, the
first range is selected, and it is switched
to the second range when the terminals
are open.
ꢀ
* All relay contacts are isolated mutually
and from the internal circuit.
Digitalꢀinput:ꢀ(Option)
Voltage contact (supply 12-24VDC (15mA
Max.)) Max. 9 inputs
Remote range change over, auto cali-
bration remote start, remote hold,
average value reset, Isolated from
the internal circuit with photocoupler.
* These range value are settable between original
first range and second range.
Powerꢀsupply:ꢀ Voltage rating
; 100V to 240V AC
; 85V to 264V AC
; 50Hz/60Hz
Optional Functions
ꢀ
Allowable range
Frequency
Remoteꢀoutputꢀholding:ꢀ
Power consumption ; 100VA max.
Operationꢀconditions:ꢀ
Output signal is held at the last value or
preset value by voltage input the remote
output holding input terminals.
Ambient temperature ;
-
5˚C to 45˚C
Holding is maintained while the voltage
input the terminals. Indication of instan-
taneous values are not held.
(40˚C max. when 2 optical sys-
tem at 200V AC power source)
Ambient humidity
; 90% RH max.,
non-condensing
Rangeꢀidentificationꢀsignal:ꢀ
The present measuring range is identified
by a contact position.
Storageꢀconditions:
Ambient temperature ; -20˚C to 60˚C
The contact output terminals for each
component close when the first range
is selected, and when the second range
is selected, the terminals are open.
Autoꢀcalibration:ꢀ
Ambient humidity
; 100% RH max.,
non-condensing
Dimensionsꢀ(Hꢀ×ꢀWꢀ×ꢀD):
133 x 483 x 382mm
Approx. 9 kg
Front panel; Cool gray (PANTON 1C-F)
Steel casing, for indoor use
Mass:ꢀ
Finishꢀcolor:ꢀꢀ
Enclosure:
Auto calibration is carried out periodically
at the preset cycle.
When a standard gas cylinder for calibra-
tion and a solenoid valve for opening/
closing the gas flow line are prepared
externally by the customer, calibration
will be carried out with the solenoid
valve drive contacts for zero calibration
and each span calibration turned on/off
sequentially at the set auto calibration
timing.
Materialꢀofꢀgas-contactingꢀparts:
Gas inlet/outlet; SUS304
Sample cell; SUS304,chloroprene rubber
Infrared-ray transmitting window; CaF2
Internal piping;Toaron,Teflon, Polypropyl-
ene
Paramagnetic O2 sensor cell : SUS316
Fuel cell O2 sensor cell : ABS resin
2
Auto calibration cycle setting:
Auto calibration cycle is set.
Correction formula:
21-On
21-Os
Setting is variable within 1 to 99 hours
(in increments of 1 hour) or 1 to 40 days
(in increments of 1 day).
C =
C
x Cs
: Sample gas concentration after O2
correction
Gas flow time setting:
The time for flowing each calibration gas
in auto calibration is set.
CS : Measured concentration of sample
gas
Settable within 60 to 900 seconds (in
increments of 1 second)
OS : Measured O2 concentration (Limit
setting: 1 to 20% O2)
Autoꢀcalibrationꢀremoteꢀstart:ꢀ
On : Reference O2 concentration
(value changeable by setting.0 to
19% O2)
Auto calibration starts by opening the auto
calibration remote start input terminal
after short circuiting for 1.5 sec or longer.
Auto calibration starts when contacts
open.
AverageꢀvalueꢀafterꢀO
calculation:ꢀ
2
ꢀcorrectionꢀandꢀO ꢀaverageꢀvalue
2
ꢀ
The result of O2 correction or instan-
taneous O2 value can be output as an
average value over the preset period of
time.
Autoꢀzeroꢀcalibration:
ꢀ
Auto zero calibration is carried out periodi-
cally at the preset cycle.
This cycle is independent on “Auto cali-
bration” cycle.
Used for averaging is the moving average
method in which sampling is carried out
at intervals of 30 seconds.
(Output is updated every 30 seconds. It
is the average value in the determined
period of time just before the latest up-
dating.)
Averaging time is settable within 1 to 59
minutes (in increments of 1 minute) or 1
to 4 hours (in increments of 1 hour).
When zero calibration gas and solenoid
valve for opening/closing the calibration
gas flow line are prepared externally by
the customer, zero calibration will be
carried out with the solenoid valve drive
contact for zero calibration turned on/off
at the set auto zero calibration timing.
Auto zero calibration cycle setting:
Auto zero calibration cycle is set.
Setting is variable within 1 to 99 hours
(in increments of 1 hour) or 1 to 40 days
(in increments of 1 day)
Averageꢀvalueꢀresetting:ꢀ
The above-mentioned output of average
ꢀ
value is started from the initial state by
opening the average value resetting input
terminals after short circuiting for 1.5 sec
or longer.
Output is reset by input voltage and
restarted by opening
Gas flow time setting:
The timing for flowing zero gas in auto
zero calibration is set.
Settable within 60 to 900 seconds (in
increments of 1 second)
Communicationꢀꢀfunction:
RS-485 (9pins D-sub connector)
High/lowꢀlimitꢀalarm:ꢀ
Alarm contact output turns on when
ꢀ
ꢀ
Half-duplex bit serial
measurement value reaches the preset
high or low limit alarm value.
Start-stop synchronization
ModbusTM protocol
Contacts close when the instantaneous
value of each channel exceeds the high
alarm limit value or falls below the low
alarm limit value.
Contents : Read/Write parameters
Read measurement concen-
tration and instrument status
Remark : When connecting via RS-
232C interface, an RS-232C
↔ RS-485 converter should
be used.
Instrumentꢀerrorꢀcontactꢀoutput:ꢀ
Contacts turn on at occurrence of analyzer
error No. 1, 2, 3 or 10.
Calibrationꢀerrorꢀcontactꢀoutput:ꢀ
Contacts turn on at occurrence of manual
ꢀ
Atmosphericꢀpressureꢀcorrection:
ꢀ Measure atmospheric pressure and cal-
ꢀ
or auto calibration error (any of errors No.
4 to 9).
Autoꢀcalibrationꢀstatusꢀcontactꢀoutputs:
Contacts turn on during auto calibration.
culate compensation (for use, be sure to
relieve the exhaust gas from analyzer to
the atmosphere)
After atmospheric pressure correction;
Zero point : No influenced
O2ꢀcorrection: Correction of measured NO, SO2 and CO
gas concentrations into values at refer-
ence O2 concentration
Span point: The change is 0.5% mea-
sured value or less relating
to the change of the atmo-
spheric pressure 1%.
Correction range: 700hPa-1050hPa
3
ZPA
Corrosiveꢀcomponent:ꢀ1 ppm or less
Standardꢀgasꢀforꢀcalibration:
Performance
Repeatability:ꢀ
Linearity:ꢀ
0.5% of full scaleꢀ
1% of full scale
prior to atmospheric pressure correction
(option)
2% of full scale/week
In the case of Auto zero calibration use
for 500ppm or less range.
2% of full scale/week
1) Infrared-ray measurable component, standard O2
ꢀ
Zero gas ; Dry N2
Span gas ; Each sample gas having con-
centration 90 to 100% of its
measuring range (recom-
mended).
In case a zirconia O2 analyzer is installed
externally and calibration is carried out on
the same calibration gas line:
Zero gas ; Dry air or atmospheric air (Do
not use with CO2 measure-
ment)
Zeroꢀdrift:ꢀ
Spanꢀdrift:ꢀ
Responseꢀtimeꢀ(forꢀ90%ꢀFSꢀresponse)ꢀ:ꢀ
1 to 15 sec electrical response. Within
10-30 seconds including replacement time
of sampling gas.
Gas replacement time depends on the
number of measuring components,and
measuring range.
Span gas ; For other than O2 measure-
ment, each sample gas hav-
ing concentration 90 to 100%
of its measuring range
Interferenceꢀfromꢀotherꢀgases:ꢀ
For O2 measurement, O2 gas
of 1 to 2 vol%/remains N2 gas
Interference
component
CO2
CO
CH4
SO2
NO
analyzer analyzer analyzer analyzer analyzer
2) Reverse range O2 measurement
CO
≤1.0%FS
––
≤1.0%FS
≤
1.0%FS ≤1.0%FS
ꢀ
Zero gas ; 100% O2
1000ppm
*1)
*2)
Span gas ; For O2 measurement, O2 gas
of 95.0%~95.5%/remains N2
gas
CO2 15%
≤1.0%FS ≤1.0%FS ≤1.0%FS ≤1.0%FS
––
*2)
H2O saturation
at 20°C
≤1.0%FS
≤1.0%FS
≤1.0%FS
––
––
*For reverse range O2 measurement,
infrared measurable component is not
measurable.
H2O saturation
––
≤2.0%FS
––
≤2.0%FS ≤2.0%FS
at 2°C
CH4
1000ppm
≤1.0%FS ≤1.0%FS
––
≤20ppm
––
*1) 0-200ppm range
*2) 0-500ppm range or less
≤
2.0%FS
2.0%FS
≤
Installation Requirements
• Indoor use (Select a place where the equipment does
not receive direct sunlight, draft/rain or radiation from hot
substances. If such a place cannot be found, a roof or
cover should be prepared for protection.)
• Avoid a place where unit receives heavy vibration
• Select a place where atmospheric air is clean
EC Directive Compliance
The product conforms to the requirements of the Low Volt-
age Directive 2006/95/EC and EMC directive 89/336/EEC
(as amended by Directive 92/31/EEC), both as amended by
Directive 93/68/EEC.
It conforms to following standards for product safety and
electromagnetic compatibility ;
EN61010-1 : 2010 Safety requirements for electrical equip-
ment for measurement, control and laboratory
use.
“Installation Category II”
“Pollution Degree 2”
EN61326-1 : 2006
Electrical equipment for measurement, control
and laboratory use — EMC requirements.
Standard Requirements for Sample Gas
Flowꢀrate:ꢀ
Temperature:ꢀ
Pressure:ꢀ
0.5L / min 0.2L / min
0 to 50˚C
10 kPa or less (Gas outlet side should be
open to the atmospheric air.)
100 µg/Nm3 or less in particle size of 0.3
µm or less
Dust:ꢀ
Mist:ꢀ
Unallowable
Moisture:ꢀ
Below a level where saturation occurs at
room temperature (condensation unallow-
able).
Below the level where saturation occurs
at 2˚C for CO measurement in 0 to 200
ppm range, NO measurement, and SO2
measurement.
4
Principle diagram of NDIR type measurement (For NO, SO2, CO2, CO, CH4)
Front expansion chamber
Gas inlet
Gas outlet
Infrared-ray
light source
Rear
expansion
chamber
Detector
Mass flow
sensor
Motor
Measuring cell
Chopper
Preamplifier
Signal
Indication
processing
and calculation
block
Output
Principle diagram of fuel cell type measurment (For O2)
Resistor
Thermistor
Electrolyte
Negative
electrode
Diaphragm
Positive
electrode
Principle diagram of paramagnetic type measurment (For O2)
Mirror
Permanent magnet
Dumb-bell
Gas inlet
Magnetic field
Measuring cell
Gas outlet
Preamplifier
Permanent
magnet
Photo diode
Light emitting diode
Signal processing and
calculation block
Indication
Output
5
ZPA
Examples of sampling system configuration including gas analyzer
To measure low moisture content (saturated at room temperature or lower) sample gas (CO, CO2, CH4)
9
Flow meter
10
Membrane
filter
1
3
Pump
Mist filter
Sample
gas inlet
7
Two-way
Solenoid
valve
Gas
analyzer
(ZPA)
7
7
7
Solenoid
valve
5
Drain pot
11
Demister
O2 analyzer
6
Ball valve
(In the case of zirconia type)
Exhaust
6
Ball valve
(atmospheric pressure)
Drain
Reference gas
for zero
calibration
Reference gas for
span calibration
8
To measure high moisture content sample gas, NO, SO2, or CO (0 to 200 ppm range)
13
Solenoid
valve
9
Flow meter
10
Membrane
filter
1
3
Pump
Mist filter
Sample
gas inlet
12 NOX/NO converter
7
Two-way
Solenoid
valve
3
Pump
(for NOX measurement)
Atmospheric
(Air)
7
7
7
Solenoid
valve
1)
4
Electronic
cooler
*
Gas
analyzer
(ZPA)
2
Safety
drain trap
5
Drain
pot
O2 analyzer
6
(In the case of
Ball
valve
zirconia type)
Reference gas
for zero
calibration
Reference gas for
span calibration
Drain
*
Exhaust
(atmospheric
pressure)
8
1) Be sure to use a dehumidifier such as electronic cooler for NO, SO2, and CO
analyzers of 0 to 200 ppm range (=2ºC saturation or lower).
List of sampling devices (example)
No.
Device name
Fuji’s type
1
2
3
4
5
Mist filter
ZBBK1V03-0
ZBH51603
Safety drain trap
Pump
ZBG80
Electoric cooler
Drain pot
ZBC91004
ZBH13003 (Length 255mm)
ZBFB1
Ball valve
6
7
8
Two-way solenoid valve
Standard gas for calibration
Flow meter
ZBM Y04-0 (Codes in to be selected depending on application)
ZBD42203
ZBBM2V03-0
ZBH35003
ZDL02001
9
10
11
12
13
Membrane filter
Demister
NO2/NO converter
Three-way solenoid valve
Note) The above is a typical configuration example. As configuration may differ depending on measuring objects,
please consult us.
6
CODE SYMBOLS
standard multi components
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25
Digit
note
Z P A
B
1
Digit
4
Description
<Specification/Structure>
Horizontal type (Terminal block for power supply)
Horizontal type (Power inlet, with lock)
<Mounting>
19inch rack mounting type EIA conformity
<Measurable component (NDIR)>
1st component 2nd component 3rd component 4th component
A
D
note1
5
6
B
note2
Y
P
A
D
B
E
F
G
J
K
L
None
NO
SO2
CO2
CO
CH4
NO
NO
CO2
CH4
CO2
NO
CO2
NO
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
SO2
CO
CO
CO
CH4
SO2
CO
SO2
-
-
-
N
T
V
Z
CO
CH4
CO2
-
CO
Others
7
<Measurable component (O2)>
None
External O2 analyzer
External zirconia O2 sensor (ZFK7)
Built-in fuel cell O2 sensor
Built-in paramagnetic O2 sensor
<Revision code>
Y
1
2
3
4
note3
8
9
1
note4
note4
note4
note4
note4
note4
note4
note4
<Measuring range (NDIR)>1st component, 1st range
10 <Measuring range (NDIR)>1st component, 2nd range
11 <Measuring range (NDIR)>2nd component, 1st range
12 <Measuring range (NDIR)>2nd component, 2nd range
13 <Measuring range (NDIR)>3rd component, 1st range
14 <Measuring range (NDIR)>3rd component, 2nd range
15 <Measuring range (NDIR)>4th component, 1st range
16 <Measuring range (NDIR)>4th component, 2nd range
17 <Measuring range (O2)>
None
0-5/10%
0-5/25%
0-10/25%
Y
A
B
C
L
0-5%
0-10%
0-25%
0-50%
M
V
P
0-100%
100-95%
Others
R
S
Z
note5
18 <Gas connection>
1
2
Rc1/4
NPT1/4
19 <Output>
A
B
C
D
DC0-1V
DC4-20mA
DC0-1V+Communication function
DC4-20mA+Communication function
20 <Indication/power supply cable>
Japanese (PSE)
note6
note6
note6
note6
note7
J
E
U
C
English (UL)
English (CEE)
Chinese (CCC)
21 <O2 correction and O2 correction average output>
None
O2 correction
Y
A
B
C
O2 correction average
O2 correction and O2 correction average
22 <Optional function (DIO)>
FAULT A. Cal. H/L Alarm RangeID/Remote range
None
Y
A
B
C
D
E
F
G
H
note8
7
ZPA
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25
Digit
note
Z R E
1
Description
<Pressure compensation>
None
Digit
23
Y
1
Pressure compensation
<Un>
24
25
A
B
ppm, Vol%
mg/m3, g/m3
note9
note10
<Adjustment>
For standard
For heat treatment furnace
For converter
Others
A
C
D
Z
RANGE CODE
O measurement range
2
Range
None
Code
Y
B
C
D
S
E
F
Range
Code
J
K
Q
L
M
N
V
W
P
X
R
Z
Measurement Range Fuel cell O
2
Paramagnet Zirconia O
ic O sensor sensor
(built - in) (built - in) (external)
2
range
code sensor
2
0~1%
0~2%
0~3%
0~5%
0~100ppm
0~200ppm
0~250ppm
0~300ppm
0~500ppm
0~1000ppm
0~2000ppm
0~2500ppm
0~3000ppm
0~5000ppm
0~5/10 vol%
0~5/25 vol%
0~10/25 vol%
0~5 vol%
A
B
C
L
0~10%
0~20%
0~25%
0~40%
0~50%
0~70%
0~100%
Others
0~10 vol%
0~25 vol%
0~50 vol%
0~100 vol%
100~95 vol%
M
V
P
R
S
G
U
T
H
note1)When “D” is specified at 4th digit, Power cable is supplied in the scope of supply. Cable specification should be
specified at the 20th digit.
note2)When only O2 measurement is necessary, "Y" should be specified at 6th digit.
note3)When "1" is specified at 7th digit, O2 pt sensor signal has to be set as 0-1V DC linear corresponding to full scale.
External zirconia O2 sensor and external O2 analyzer are not included in the scope of supply, and has to be separately
ordered.
note4)Refer toTable 1 for possible combination of measuring components and ranges in the data sheet. When "Y" is specified
at 6th digit, "Y" should be specified at 9th to 16th digit. For fuel cell O2 sensor, range is 0-10% or more.
note5)Select the type of voltage rating, plug type and applicable standard of the power cable by 20th digit. Select a power
cable for using at the location of end-user. When “A” is specified at 4th digit, Power cable will not be included in scope
of delivery. When English is specified for display language, "E" should be selected at 20th digit.
note6)O2 correction is calculated only for NO, SO2 and CO.
note7)When 5 components measurement is specified, "H" must not be specified at 22nd digit.
When 4 components measurement is specified and “H” is specified at 22nd digit, 3 points is maximum for alarm
output function.
note8) When "B" is specified at 24th digit, measuring range should be specified by ppm range code. In this case NO, SO2 and
3
CO measuring range are corresponding range in mg/m . Please refer to the table shown below for the corresponding
3
range code based on "mg/m ".
note9) When A to D is specified at 25th digit, the analyzer will be adjusted and delivered with the following gasses. Standard
"A": balance gas N2.
For heat treatment furnace "C": balance gas 30% H2/remamins N2.
For converter "D": balance gas CO, CO2.
When other adjustment is required, please specify "Z".
When "Z" is specified, please attach a list of gas composition contained in the measuring gas.
Corresponding mg/m3
Corresponding range in mg/m3
Range code
Unit : ppm
0-200ppm
0-250ppm
0-300ppm
0-500ppm
0-1,000ppm
0-2,000ppm
0-2,500ppm
0-3,000ppm
0-5,000ppm
NO
SO2
CO
C
D
S
E
0-260mg/m3
0-325mg/m3
0-400mg/m3
0-650mg/m3
0-1,300mg/m3
0-2,600mg/m3
0-3,300mg/m3
0-4,000mg/m3
0-6,600mg/m3
0-570mg/m3
0-700mg/m3
0-850mg/m3
0-1,400mg/m3
0-2,800mg/m3
0-5,600mg/m3
0-7,100mg/m3
0-8,500mg/m3
0-14.00g/m3
0-250mg/m3
0-300mg/m3
0-375mg/m3
0-600mg/m3
0-1,250mg/m3
0-2,500mg/m3
0-3,000mg/m3
0-3,750mg/m3
0-6,250mg/m3
F
The conversion formula "ppm" unit
into "mg/m3" unit.
G
U
T
NO (mg/m3) = 1.34 × NO (ppm)
SO2 (mg/m3) = 2.86 × SO2 (ppm)
CO (mg/m3) = 1.25 × CO (ppm)
H
8
Table 1 Measurable component and range - availability check table -
Procedure of range selection
On one component analyzer:
First determine 1st range, then select 2nd range from the right of your determination range in following tables.
More than two components:
The procedure is same as one component. Difference is 2nd range in the tables.
The 2nd range in the tables for two and more components is maximum available range.
2nd range is selectable from 1st range to 2nd range(max) on the table.
1-component analyzer : CO
1st range
1-component analyzer : NO
1st range
2nd range
2nd range
0 - 200ppm None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm
0 - 250ppm None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm
0 - 300ppm None, 0 - 500ppm,1000ppm,2000ppm,2500ppm
0 - 500ppm None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm
0 - 1000ppm None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1%
0 - 2000ppm None, 0 - 2500ppm,3000ppm,5000ppm,1%,2%
0 - 2500ppm None, 0 - 3000ppm,5000ppm,1%,2%
0 - 3000ppm None, 0 - 5000ppm,1%,2%
0 - 200ppm None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm
0 - 250ppm None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm
0 - 300ppm None, 0 - 500ppm,1000ppm,2000ppm,2500ppm
0 - 500ppm None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm
0 - 1000ppm None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm
0 - 2000ppm None, 0 - 2500ppm,3000ppm,5000ppm
0 - 2500ppm None, 0 - 3000ppm,5000ppm
0 - 3000ppm None, 0 - 5000ppm
0 - 5000ppm None, 0 - 1%,2%,3%,5%
0 - 5000ppm None
0 - 1% None, 0 - 2%,3%,5%,10%
0 - 2% None, 0 - 3%,5%,10%,20%
0 - 3% None, 0 - 5%,10%,20%,25%
1-component analyzer : SO2
1st range
2nd range
0 - 200ppm None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm
0 - 250ppm None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm
0 - 300ppm None, 0 - 500ppm,1000ppm,2000ppm,2500ppm
0 - 500ppm None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm
0 - 1000ppm None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1%
0 - 2000ppm None, 0 - 2500ppm,3000ppm,5000ppm,1%,2%
0 - 2500ppm None, 0 - 3000ppm,5000ppm,1%,2%
0 - 3000ppm None, 0 - 5000ppm,1%,2%
0 - 5% None, 0 - 10%,20%,25%,40%,50%
0 - 10% None, 0 - 20%,25%,40%,50%,70%,100%
0 - 20% None, 0 - 25%,40%,50%,70%,100%
0 - 25% None, 0 - 40%,50%,70%,100%
0 - 40% None, 0 - 50%,70%,100%
0 - 50% None, 0 - 70%,100%
0 - 70% None, 0 - 100%
0 - 100% None
0 - 5000ppm None, 0 - 1%,2%,3%,5%
0 - 1% None, 0 - 2%,3%,5%,10%
1-component analyzer : CO2
0 - 2% None, 0 - 3%,5%,10%
0 - 3% None, 0 - 10%
1st range
2nd range
0 - 5% None, 0 - 10%
0 - 10% None
1-component analyzer : CH4
1st range
0 - 500ppm None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm
0 - 1000ppm None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1%
0 - 2000ppm None, 0 - 2500ppm,3000ppm,5000ppm,1%,2%
0 - 2500ppm None, 0 - 3000ppm,5000ppm,1%,2%
0 - 3000ppm None, 0 - 5000ppm,1%,2%
0 - 5000ppm None, 0 - 1%,2%,3%,5%
0 - 1% None, 0 - 2%,3%,5%,10%
0 - 100ppm None, 0 - 200ppm,250ppm,300ppm,500ppm,1000ppm
0 - 200ppm None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm
0 - 250ppm None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm
0 - 300ppm None, 0 - 500ppm,1000ppm,2000ppm,2500ppm
0 - 500ppm None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm
0 - 1000ppm None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1%
0 - 2000ppm None, 0 - 2500ppm,3000ppm,5000ppm,1%,2%
0 - 2500ppm None, 0 - 3000ppm,5000ppm,1%,2%
0 - 3000ppm None, 0 - 5000ppm,1%,2%
2nd range
0 - 5000ppm None, 0 - 1%,2%,3%,5%
0 - 1% None, 0 - 2%,3%,5%,10%
0 - 2% None, 0 - 3%,5%,10%,20%
0 - 2% None, 0 - 3%,5%,10%,20%
0 - 3% None, 0 - 5%,10%,20%,25%
0 - 3% None, 0 - 5%,10%,20%,25%
0 - 5% None, 0 - 10%,20%,25%,40%,50%
0 - 5% None, 0 - 10%,20%,25%,40%,50%
0 - 10% None, 0 - 20%,25%,40%,50%,70%,100%
0 - 20% None, 0 - 25%,40%,50%,70%,100%
0 - 25% None, 0 - 40%,50%,70%,100%
0 - 40% None, 0 - 50%,70%,100%
0 - 10% None, 0 - 20%,25%,40%,50%,70%,100%
0 - 20% None, 0 - 25%,40%,50%,70%,100%
0 - 25% None, 0 - 40%,50%,70%,100%
0 - 40% None, 0 - 50%,70%,100%
0 - 50% None, 0 - 70%,100%
0 - 50% None, 0 - 70%,100%
0 - 70% None, 0 - 100%
0 - 70% None, 0 - 100%
0 - 100% None
0 - 100% None
2-component analyzer : NO/SO2
2-component analyzer : NO/CO
1-component : NO
2-component : SO2
1-component : NO
2-component : CO
1st range 2nd range (max.)
0 - 200ppm 0 - 2000ppm
0 - 250ppm 0 - 2500ppm
0 - 300ppm 0 - 2500ppm
0 - 500ppm 0 - 5000ppm
0 - 1000ppm 0 - 5000ppm
0 - 2000ppm 0 - 5000ppm
0 - 2500ppm 0 - 5000ppm
0 - 3000ppm 0 - 5000ppm
0 - 5000ppm None
1st range 2nd range (max.)
0 - 200ppm 0 - 2000ppm
0 - 250ppm 0 - 2500ppm
0 - 300ppm 0 - 2500ppm
0 - 500ppm 0 - 5000ppm
0 - 1000ppm 0 - 5000ppm
0 - 2000ppm 0 - 5000ppm
0 - 2500ppm 0 - 5000ppm
0 - 3000ppm 0 - 5000ppm
0 - 5000ppm None
1st range 2nd range (max.)
0 - 200ppm 0 - 2000ppm
0 - 250ppm 0 - 2500ppm
0 - 300ppm 0 - 2500ppm
0 - 500ppm 0 - 5000ppm
0 - 1000ppm 0 - 5000ppm
0 - 2000ppm 0 - 5000ppm
0 - 2500ppm 0 - 5000ppm
0 - 3000ppm 0 - 5000ppm
0 - 5000ppm None
1st range 2nd range (max.)
0 - 200ppm 0 - 2000ppm
0 - 250ppm 0 - 2500ppm
0 - 300ppm 0 - 2500ppm
0 - 500ppm 0 - 5000ppm
0 - 1000ppm 0 - 5000ppm
0 - 2000ppm 0 - 5000ppm
0 - 2500ppm 0 - 5000ppm
0 - 3000ppm 0 - 5000ppm
0 - 5000ppm None
• The 2nd component should
• The 2nd component should
be selected as shown in the
right table.
be selected as shown in the
right table.
9
ZPA
2-component analyzer: CO2/CO
1-component: CO2
2-component: CO
1st range 2nd range (max.) 1st range/2nd range (max.)
0-100ppm
0-200ppm
0-250ppm
0-300ppm
0-500ppm
0-500ppm
0-1000ppm
0-2000ppm
0-1000ppm
0-2000ppm
0-2000ppm
0-2500ppm
0-1000ppm
0-2000ppm
0-2500ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/3%,
0-3%
0-5000ppm
0-5000ppm
0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/3%, 0-3%
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%,
0-3/10%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%,
0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-1%
0-1%
0-2%
0-1%
0-2500ppm
0-3000ppm
0-3000ppm
0-2%
0-1%
0-2%
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%,
0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%,
0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/25%,
0-10/100%,0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-5000ppm
0-1%
0-2%
0-3%
0-5%
0-5%
0-5%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-5000ppm
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-1%
0-2%
0-10%
0-20%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-2%
0-3%
0-5%
0-10%
0-20%
0-25%
0-40%
0-50%
0-70%
0-100%
0-10%
0-25%
0-50%
0-100%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%,
0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
None
2-component analyzer: CH4/CO
1-component: CH4
2-component: CO
1st range 2nd range (max.) 1st range/2nd range (max.)
0-500ppm
0-1000ppm
0-1000ppm
0-2000ppm
0-5000ppm
0-5000ppm
0-1%
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm
0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%,
0-3/20%, 0-5/20%, 0-10/20%
0-5000ppm
0-2500ppm
0-3000ppm
0-2000ppm
0-2500ppm
0-3000ppm
0-2000ppm
0-2500ppm
0-3000ppm
0-5000ppm
0-5000ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1% 0-3000ppm/1%,
0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/25%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/25%
0-1%
0-1%
0-2%
0-2%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/25%
0-1%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%,
0-25/50%, 0-40/50%, 0-50%
0-5000ppm
0-5000ppm
0-1%
0-3%
0-5%
0-5%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5% 0-2/5%, 0-3/25%, 0-5/25%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1% ,0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-1%
0-2%
0-2%
0-3%
0-3%
0-5%
0-5%
0-10%
0-10%
0-10%
0-20%
0-10%
0-25%
0-25%
0-50%
0-50%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1% ,0-3000ppm/1%, 0-5000ppm/1%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/50%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%,
0-50/100%, 0-70/100%, 0-100%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%,
0-50/100%, 0-70/100%, 0-100%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/5%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%,
0-50/100%, 0-70/100%, 0-100%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%,
0-50/100%, 0-70/100%, 0-100%
0-10%
0-20%
0-25%
0-40%
0-20%
0-25%
0-40%
0-50%
0-70%
0-100%
0-100%
0-50%
0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%,
0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-100%
None
0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
10
2-component analyzer: CO2/CH4
1-component: CO2
2-component: CH4
1st range 2nd range (max.) 1st range/2nd range (max.)
0-100ppm
0-200ppm
0-250ppm
0-300ppm
0-500ppm
0-500ppm
0-1000ppm
0-1000ppm
0-1000ppm
0-2000ppm
0-2000ppm
0-2000ppm
0-2500ppm
0-2500ppm
0-3000ppm
0-5000ppm
0-1000ppm
0-2000ppm
0-2500ppm
0-2500ppm
0-2500ppm
0-5000ppm
0-2500ppm
0-5000ppm
0-1%
0-2500ppm
0-5000ppm
0-2%
0-5000ppm
0-2%
0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm, 0-5000ppm
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/1%, 0-1%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/2%, 0-2/10%, 0-3/10%, 0-5/10%, 0-10%
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/2%, 0-2/10%, 0-3/10%, 0-5/10%, 0-10%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3% 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/25%, 0-10/25%, 0-20/25%, 0-25%
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/25%, 0-10/25%, 0-20/25%, 0-25%
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5% 0-2/20%, 0-3/20%, 0-5/25%, 0-10/25%, 0-20/25%, 0-25%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%,
0-40/50%, 0-50%
0-2%
0-3%
0-5000ppm
0-1%
0-5%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/50%, 0-20/50%, 0-25/50%,
0-40/50%, 0-50%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%,
0-40/50%, 0-50%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/25%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%,
0-50/100%, 0-70/100%, 0-100%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-10%
0-20%
0-10%
0-25%
0-20%
0-50%
0-20%
0-50%
0-2%
0-2%
0-3%
0-3%
0-5%
0-5%
0-10%
0-10%
0-20%
0-25%
0-40%
0-10%
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%,
0-50/100%, 0-70/100%, 0-100%
0-100%
0-100%
0-100%
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%,
0-70/100%, 0-100%
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%,
0-70/100%, 0-100%
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%,
0-70/100%, 0-100%
0-20%
0-25%
0-40%
0-50%
0-70%
0-100%
None
3-component analyzer: NO/SO2/CO >>> Combination of 1st component NO and 2nd component SO2 / 3rd component CO
1-component: NO
2-component: SO2
3-component: CO
1st range 2nd range (max.)
1st range 2nd range (max.) 1st range/2nd range (max.)
0-200ppm
0-250ppm
0-300ppm
0-500ppm
0-1000ppm
0-2000ppm
0-2500ppm
0-3000ppm
0-5000ppm
0-2000ppm
0-2500ppm
0-2500ppm
0-5000ppm
0-5000ppm
0-5000ppm
0-5000ppm
0-5000ppm
None
0-200ppm
0-250ppm
0-300ppm
0-500ppm
0-1000ppm
0-2500ppm
0-1000ppm
0-2000ppm
0-2500ppm
0-3000ppm
0-5000ppm
0-2000ppm
0-2500ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-5000ppm
0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm, 0-5000ppm
0-2500ppm
None
0-5000ppm
None
3-component analyzer: CO2/CO/CH4 >>>
Combination of 1st component CO2 / 2nd component CO and 3rd component CH4
1-component: CO2
2-component: CO
3-component: CH4
1st range 2nd range (max.) 1st range/2nd range (max.)
1st range 2nd range (max.) Availability of product
0-5000ppm
0-1%
0-2%
0-3%
0-5%
0-5%
0-5%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%,
0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-5000ppm
0-5%
Product available only
when CO analyzer max.
measuring range is
50% or less
0-5000ppm
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%,
0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-500/5000ppm , 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%,
0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-500/5000ppm , 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%,
0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-500/5000ppm , 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%,
0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-1%
0-2%
0-10%
0-20%
Product available
0-1%
0-2%
0-10%
0-20%
Product available only
when CO analyzer
measuring range is 0 to
1000ppm or more.
Product available only
when CO analyzer
measuring range is 0 to
5000ppm or more.
Product available only
when CO analyzer
measuring range is more
than 5000ppm or CO2
analyzer range is more
than 2%.
0-3%
0-25%
0-10%
0-20%
0-25%
0-40%
0-50%
0-5%
0-10%
0-20%
0-25%
0-40%
0-2%
0-3%
0-5%
0-10%
0-20%
0-25%
0-40%
0-50%
0-70%
0-100%
0-10%
0-25%
0-50%
0-100%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%,
0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%,
0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-50%
0-70%
0-100%
0-70%
0-100%
None
None
11
ZPA
4-component analyzer: NO/SO2/CO2/CO >>> Combination of 1st component NO /4th component CO and component 2nd component SO2/3rd component CO2
1-component: NO
4-component: CO
1st range 2nd range (max.) 1st range/2nd range (max.)
0-200ppm
0-250ppm
0-300ppm
0-500ppm
0-1000ppm
0-2000ppm
0-500ppm
0-1000ppm
0-2000ppm
0-2500ppm
0-3000ppm
0-5000ppm
0-2000ppm
0-2500ppm
0-2500ppm
0-2000ppm
0-2000ppm
None
0-5000ppm
0-5000ppm
0-5000ppm
0-5000ppm
0-5000ppm
None
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm, None
0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm, None
2-component analyzer: SO2 3-component analyzer: CO2
1st range 2nd range (max.) 1st range/2nd range (max.)
0-200ppm
0-250ppm
0-300ppm
0-500ppm
0-1000ppm
0-2000ppm
0-2500ppm
0-3000ppm
0-5000ppm
0-2000ppm
0-2500ppm
0-2500ppm
0-5000ppm
0-5000ppm
0-5000ppm
0-5000ppm
0-5000ppm
None
0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%/None
Table 2 Channel (Ch) No. and display/output contents comparison table
Code symbol
6th digit 7th digit 21st digit Display/output contents
Y
P
A
D
B
E
F
1 to 3
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Ch1:O2
Ch1:NO
Ch1:SO2
Ch1:CO2
Ch1:CO
Ch1:CH4
Ch1:NO, Ch2:SO2
Y
Y
G
J
Y
Y
Y
Y
Ch1:NO, Ch2:CO
Ch1:CO2, Ch2:CO
K
L
Y
Y
Y
Y
Ch1:CH4, Ch2:CO
Ch1:CO2, Ch2:CH4
N
T
V
P
A
D
B
E
F
G
J
K
L
N
T
V
P
A
B
F
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
A *
A *
A *
A *
A *
A *
A
A
C
C
C
C
Ch1:NO, Ch2:SO2, Ch3:CO
Ch1:CO2, Ch2:CO, Ch3:CH4
Ch1:NO, Ch2:SO2, Ch3:CO2, Ch4:CO
Ch1:NO, Ch2:O2
Ch1:SO2, Ch2:O2
Ch1:CO2, Ch2:O2
Ch1:CO, Ch2:O2
Ch1:CH4, Ch2:O2
Ch1:NO, Ch2:SO2, Ch3:O2
Ch1:NO, Ch2:CO, Ch3:O2
Ch1:CO2, Ch2:CO, Ch3:O2
Ch1:CH4, Ch2:CO, Ch3:O2
Ch1:CO2, Ch2:CH4, Ch3:O2
Ch1:NO, Ch2:SO2, Ch3:CO, Ch4:O2
Ch1:CO2, Ch2:CO, Ch3:CH4, Ch4:O2
Ch1:NO, Ch2:SO2, Ch3:CO2, Ch4:CO, Ch5:O2
Ch1:NOx, Ch2:O2, Ch3:corrected NOx
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
1 to 3
Ch1:SO2, Ch2:O2, Ch3:corrected SO2
Ch1:CO, Ch2:O2, Ch3:corrected CO
Ch1:NOx, Ch2:SO2, Ch3:O2, Ch4:corrected NOx, Ch5:corrected SO2
Ch1:NOx, Ch2:CO, Ch3:O2, Ch4:corrected NOx, Ch5:corrected CO
Ch1:CO2, Ch2:CO, Ch3:O2, Ch4:corrected CO
Ch1:NOx, Ch2:SO2, Ch3:CO, Ch4:O2, Ch5:corrected NOx, Ch6:corrected SO2, Ch7:corrected CO
Ch1:NOx, Ch2:SO2, Ch3:CO2, Ch4:CO, Ch5:O2, Ch6:corrected NOx, Ch7:corrected SO2, Ch8:corrected CO
Ch1:NOx, Ch2:O2, Ch3:corrected NOx, Ch4:corrected NOx average
Ch1:SO2, Ch2:O2, Ch3:corrected SO2, Ch4:corrected SO2 average
Ch1:CO, Ch2:O2, Ch3:corrected CO, Ch4corrected CO average
Ch1:NOx, Ch2:SO2, Ch3:O2, Ch4:corrected NOx, Ch5:corrected SO2, Ch6:corrected NOx average,
Ch7:corrected SO2 average
G
J
*
*
*
*
*
*
N
V
P
A
B
F
*
G
1 to 3
C
Ch1:NOx, Ch2:CO, Ch3:O2, Ch4:corrected NOx, Ch5:corrected CO, Ch6:corrected NOx average,
Ch7:corrected CO average
*
*
J
N
1 to 3
1 to 3
C
C
Ch1:CO2, Ch2:CO, Ch3:O2, Ch4:corrected CO, Ch5:corrected CO average
Ch1:NOx, Ch2:SO2, Ch3:CO, Ch4:O2, Ch5:corrected NOx, Ch6:corrected SO2, Ch7:corrected CO,
Ch8:corrected NOx average, Ch9:corrected SO2 average, Ch10:corrected CO average
Ch1:NOx, Ch2:SO2, Ch3:CO2, Ch4:CO, Ch5:O2, Ch6:corrected NOx, Ch7:corrected SO2, Ch8:corrected CO,
Ch9:corrected NOx average, Ch10:corrected SO2 average2, Ch11:corrected CO average
*
V
1 to 3
C
* When the 21st digit code is A or C, the component of the NO analyzer is displayed as NOx.
12
OUTLINE DIAGRAMS (Unit : mm)
Refer to appendix. (TC drawing)
429
<TOP VIEW>
<FRONT VIEW>
415
463
483
Power Switch
Digital Input/output Connector (DIO1 to 3)
Analog Output Connctor (A/O)
Communication Connector (RS485)
Analog Input Connector (A/I)
Purge Gas Inlet Rc1/4 or NPT1/4
Sample Gas Inlet Rc1/4 or NPT1/4
Sample Gas Outlet Rc1/4 or NPT1/4
Fuse
<REAR VIEW>
<SIDE VIEW>
(M4Terminal BlockType)
Power Inlet(100V to 240V AC 50/60Hz)
25
2
380
396
24
16
(M4Terminal BlockType)
(Power Inlet Type(100V to 240V AC 50/60Hz))
13
ZPA
Mounting method
The analyzer weight should be supported at the bottom of the case.
19-inch rack mounting type
<Rack dimensions>
450 or more
465
Mounting diagram
Guide rails
Guide rails
EXTERNAL CONNECTION
<Analog output> A/O connector
<RS485 communication signal>
1
2
1
2
3
4
5
13
1
5
1
AO1+
AO1
AO2+
AO2
AO3+
AO3
AO4+
AO4
AO5+
AO5
AO6+
AO6
AO7+
AO7
AO8+
AO8
AO9+
AO9
AO10+
AO10
AO11+
AO11
AO12+
(GND)
14
15
16
17
18
19
20
21
22
23
24
25
6
7
8
9
-
RT×D+
25
14
9
6
-
3
RT×D
-
D-sub 25pins female
-
D-sub 9pins female
4
* In standard, displayed Channel No.
and Analog Output No. are same.
-
5
-
6
<Screw terminal (M4)>
-
7
P
N
L
-
8
<Analog input> A/I connector (O2 signal input)
-
100 to 240V AC
9
-
AI
-
10
11
12
13
-
<Power inlet>
-
AI+
POWERꢀINLET
100-240VꢀAC
AO12
-
NC
14
<Digital I/O> DIO 1 to 3 connector (option)
DIO1
connector connector connector
DI1+ DI4+ DI7+
DI1 DI4 DI7
DI2+ DI5+ DI8+
DI2 DI5 DI8
DI3+ DI6+ DI9+
DI3 DI6 DI9
DIO2
DIO3
13
1
1
2
14
15
16
17
18
19
20
21
22
23
24
25
-
-
-
Digital input
OFF : 0V
ON : 12 to 24V DC
25
14
-
-
-
3
D-sub 25pins female
-
-
-
4
NC
* DIO 1 to 3 are all as same connector.
com
NO
NC
com
NO
NC
com
NO
NC
com
NO
NC
DO1
DO2
DO3
DO4
DO5
DO6
DO7
DO8
DO9
DO10
DO11
DO12
DO13
DO14
DO15
5
6
Digital output
max. contact load
rating 24V DC/1A
7
Contents of digital input
signal
8
9
DI1
DI2
DI3
DI4
DI5
DI6
DI7
DI8
DI9
Remote hold
Average value reset
A. cal. start
10
11
12
13
com
NO
A. zero. cal. start
Remote range Ch1
Remote range Ch2
Remote range Ch3
Remote range Ch4
Remote range Ch5
Contents of digital output signal
Independent on the
number of component
1-component analyzer
2-component analyzer 3-component analyzer
22th digit
DO1
A, C
B, E
D, F, G, H
B, D, E, F, G, H
Instrument error
Calibration error
(A.cal.status)
B, D, E, F, G, H
Instrument error
Calibration error
(A.cal.status)
Instrument error
Calibration error
Instrument error
Calibration error
A.cal.status
For zero gas
For span gas Ch1
(Alarm1)
Instrument error
Calibration error
(A.cal.status)
(For zero gas)
DO2
The items in the parentheses
may not be available depend-
ing on the selected type on
22th digit.
DO3
DO4
(For zero gas)
(For zero gas)
(For span gas Ch1)
(For span gas Ch2)
(For span gas Ch3)
(Range identification Ch1)
(Range identification Ch2)
(Range identification Ch3)
(Alarm1)
DO5
(For span gas Ch1) (For span gas Ch1)
(For span gas Ch2)
DO6
(Alarm1)
(Alarm2)
(Alarm3)
(Alarm4)
(Alarm5)
DO7
(Alarm2)
The normal open side (NO) of
digital output is close when
the function is active without
range ID.
DO8
(Alarm3)
DO9
(Alarm4)
(Range identification Ch1)
DO10
DO11
DO12
DO13
DO14
DO15
(Alarm5)
Range identification Ch1 (Range identification Ch2)
(Alarm1)
(Alarm2)
(Alarm3)
(Alarm4)
(Alarm5)
(Alarm1)
(Alarm2)
(Alarm3)
(Alarm4)
(Alarm5)
In case of range ID, normal
open (NO) side is close with
First range.
The normal close (NC) side is
close with Second range.
(Alarm2)
(Alarm3)
(Alarm4)
(Alarm5)
4-component analyzer
5-component analyzer
B, E D, F
22th digit
DO1
B, E
D, F
G
H
G
Instrument error Instrument error
Calibration error Calibration error
A.cal.status
Instrument error
Calibration error
A.cal.status
Instrument error
Calibration error
A.cal.status
Instrument error Instrument error
Calibration error Calibration error
A.cal.status
Instrument error
Calibration error
A.cal.status
DO2
DO3
DO4
For zero gas
For zero gas
For zero gas
For zero gas
For zero gas
DO5
For span gas Ch1
For span gas Ch1
For span gas Ch2
For span gas Ch1
For span gas Ch2
For span gas Ch3
For span gas Ch4
For span gas Ch1
For span gas Ch1
DO6
For span gas Ch2
For span gas Ch2 Range identification Ch1 For span gas Ch2
For span gas Ch3 Range identification Ch2 For span gas Ch3
For span gas Ch4 Range identification Ch3 For span gas Ch4
DO7
For span gas Ch3 Range identification Ch1 For span gas Ch3
For span gas Ch4 Range identification Ch2 For span gas Ch4
Range identification Ch3
DO8
DO9
Range identification Ch1 For span gas Ch5 Range identification Ch4 For span gas Ch5
DO10
DO11
DO12
DO13
DO14
DO15
Range identification Ch4
Range identification Ch2
(Alarm1)
Range identification Ch5
(Alarm1)
(Alarm1)
(Alarm2)
(Alarm3)
(Alarm4)
(Alarm5)
(Alarm1)
(Alarm2)
(Alarm3)
(Alarm4)
(Alarm5)
(Alarm1)
(Alarm2)
(Alarm3)
Range identification Ch1
Range identification Ch2
Range identification Ch3
Range identification Ch4
Range identification Ch5
Range identification Ch1 (Alarm2)
Range identification Ch2 (Alarm3)
(Alarm2)
(Alarm3)
Range identification Ch3 Range identification Ch3 (Alarm4)
Range identification Ch4 Range identification Ch4 (Alarm5)
(Alarm4)
(Alarm5)
SCOPE OF DELIVERY
ORDERING INFORMATION
•ꢀGasꢀanalyzerꢀ…ꢀ1ꢀunit
1.ꢀCodeꢀsymbols
2.ꢀApplicationꢀandꢀcompositionꢀofꢀsampleꢀgas
•ꢀReplacementꢀfuseꢀ(250V,ꢀ2AꢀAC,ꢀdelayꢀtype)ꢀ…ꢀ2ꢀpcs
•ꢀInstructionꢀmanualꢀ…ꢀ1ꢀcopy
•ꢀConnectorꢀforꢀI/Oꢀconnectionꢀ…ꢀ1ꢀset
•ꢀPowerꢀcable(standardꢀinletꢀtypeꢀ2m)ꢀ…ꢀ1ꢀpc
15
ZPA
Gasꢀinlet/outletꢀsize:
Rc /4 or NPT /4
Powerꢀsupply:ꢀ Rated voltage
1
1
Exclusive Zirconia O2 Sensor
(to be purchased separately)
For O2 correction, the gas analyzer ZPA can accept linearized
0 to 1V DC signal coming from analyzer calibrated 0 to 25%
O2 full scale. If the analyzer is not available, Fuji can supply
exclusive Zirconia O2 sensor Model ZFK.
Measuringꢀmethod:
; 100 to 115V AC or
200 to 240V AC
Rated frequency ; 50Hz/60Hz
Max. rated power; 215VA (during power
ON)
65VA (during steady-
state operation)
Enclosure:ꢀ
Indication:ꢀ
Steel casing, for indoor application
Temperature indication (LED)
Zirconia system
Measurableꢀcomponentꢀandꢀmeasuringꢀrange:
Temperatureꢀalarmꢀoutput:
Contact output 1a contact,
Measurable component
O2 Oxygen
Range
ꢀ
0 to 25vol%
Contact capacity 220V, 1A AC (resistive
load)
Outerꢀdimensionsꢀ(HꢀxꢀWꢀxꢀD):
Repeatability:ꢀ Within 0.5% of full scale
Linearity:ꢀ
Zeroꢀdrift:ꢀ
Spanꢀdrift:ꢀ
Within 1% of full scale
Within 1% of full scale/week
Within 2% of full scale/week
ꢀ
141 x 170 x 190mm
Massꢀ{weight}:ꢀ Approx. 3kg
Finishꢀcolor:ꢀ
Munsell 5Y 7/1
Responseꢀtime:ꢀ Approx. 20 seconds (for 90% response)
Measuredꢀgasꢀflowꢀrate:
ꢀ
0.5 0.25L / min
CODE SYMBOLS
Remark: The Zirconia system, due to its principle,
may produce a measuring error due to rela-
tive concentration versus the combustible
1
2
3
4
5
6
7
8
9
10 11 12 13
Y 0 Y Y
Description
Measuring method
Zirconia method
Z F K 7 Y Y
4
-
O
2
gas concentration. Also, a corrosive
gas (SO of 250 ppm or more, etc.) may
affect the life of the sensor.
7YY
2
Power supply
9
B
C
100 to 115V AC 50/60Hz(Standard)
200 to 240V AC 50/60Hz(Standard)
200 to 240V AC 50/60Hz(CE mark)
OUTLINE DIAGRAM (Unit:mm)
Gas inlet/outlet size
Rc 1/4
NPT 1/4
1
8
170
152
OXYGEN ANALYZER
TEMP.
1
2
4
5
AC250V
T3.15A
3
6
7
L
N
AC
OUT
ALM
INPUT
9TH DIGIT
1
1: RC /4
8: NPT1/4
ø5
OUTLET
INLET
EXTERNAL CONNECTION DIAGRAM
1
2
3
4
5
6
7
+
-
AC power supply
E
Output
Temperature
to analyzer alarm output
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fujielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan