INFRARED GAS ANALYZER  
DATAꢀSHEET  
ZPA  
This gas analyzer (ZPA) is capable of measuring the con-  
centration of NO,SO2,CO2,CO,CH4 and O2 components in  
sample gas. NO, SO2, CO2, CO and CH4 are measured by  
non-dispersion infrared method (NDIR), while O2 is mea-  
sured by fuel cell or paramagnetic and zirconia method.  
Optimum use as an analyzer unit for measurement of  
combustion exhaust gas from refuse incinerator, boiler  
exhaust or gas from various industrial furnaces.  
FEATURES  
1.ꢀHighꢀsensitivity  
Equipped with the newest upgraded mass flow sensor.  
2.ꢀEasyꢀmaintenance  
Maintenance is easy due to simple measurement unit of  
single-beam system adapted.  
3.ꢀSmallꢀandꢀlight  
The size is small 133x483x382mm (HxWxD). (This is  
for 19inch rack mounting type) and light (9kg). Unit is  
capable max. 5 component measurement in one analyzer.  
4.ꢀExtensiveꢀfunctions  
Various optional functions are available such as auto  
calibration control, atmospheric pressure correction, high  
and low concentration alarm, remote range switch, and  
range identification signal,etc.  
Measurableꢀgasꢀcomponentsꢀandꢀmeasuringꢀrange:  
Minimum range  
0 - 200ppm  
0 - 200ppm  
0 - 100ppm  
0 - 200ppm  
0 - 500ppm  
Maximum range  
0 - 5000ppm  
0 - 10vol%  
NO  
SO2  
CO2  
CO  
0 - 100vol%  
0 - 100vol%  
0 - 100vol%  
CH4  
O2  
Built in  
fuel cell  
0 - 10vol%  
0 - 25vol%  
O2  
Built-in  
Paramagnetic  
0 - 5vol%  
None  
0 - 100vol%  
100 - 95vol%  
O2  
External  
Zirconia  
0 - 5vol%  
0 - 25vol%  
ꢀꢀꢀꢀꢀꢀꢀꢀꢀꢀ  
• Max. 5 components measurement includ-  
ing O2.  
For reverse range O2 measurement, in-  
frared gas measurement is not available  
(single range O2 only).  
• Measuring range ratio max. 1:10 (except  
O2)  
• Measuring ranges are changeable be-  
tween the specified minimum and maxi-  
mum range  
SPECIFICATIONS  
Standard Specifications  
Principleꢀofꢀmeasurement:ꢀ  
NO, SO2, CO2, CO, CH4 ;  
Non-dispersion infrared-ray absorption  
method  
Settable one range or two ranges  
* In measurement range low range is  
called first range, high range is called  
second range.  
• For possible combinations of components  
and ranges, refer to Table1.  
Single light source and single beams  
(single beam system)  
;Fuel cell O2 sensor (build-in) or paramag-  
netic O2 sensor (build-in) or zirconia O2  
sensor (externally installed TYPE: ZFK7)  
O2  
Measuredꢀvalueꢀindication:ꢀ  
Digital indication in 4 digits  
(LCD panel with LED back light)  
• Instantaneous value of each component  
• Instantaneous value after O2 correction  
(only in NO, SO2, CO measurement with  
O2)  
• Average value after O2 correction  
(only in NO, SO2, CO measurement with  
O2)  
• O2 average value  
EDS3-143  
Date  
Nov.ꢀ2,ꢀ2011  
ZPA  
Analogꢀoutputꢀsignals:ꢀ  
Gasꢀinlet/outlet:ꢀRc1/4 or NPT1/4 internal thread  
Purgeꢀgasꢀflowꢀrate: 1L/min ( when required)  
LifeꢀtimeꢀofꢀfuelꢀcellꢀO2ꢀsensor: 2 years  
4 to 20mA DC or 0 to 1V DC, isolated  
internally from circuit and ground. Output  
lines are non-isolated each other.; 12  
outputs max.  
Allowable load 550Ω for 4 to 20mA DC  
Allowable load 100KΩ for 0 to 1V DC  
* Refer to Table2 for the channel No. of  
displayed values and analog output  
signals.  
Standard Functions  
Outputꢀsignalꢀholding:ꢀ  
Output signals are held unchanged during  
manual and auto calibrations by activation  
of holding (turning “ON” its setting).  
The values held are those just before start  
calibration mode or setting value.  
It is selectable.  
Analogꢀinputꢀsignal:  
For signal input from externally installed  
O2 sensor.  
Signal requirement;  
(1) Signal from Fuji’s Zirconia O2 sensor  
(TYPE: ZFK7)  
(2) 0 to 1V DC from an O2 sensor  
Input section is not isolated. This  
feature is effective when an O2 sensor  
is not built in.  
Indication of instantaneous values will  
not be held.  
Switchꢀranges:ꢀ The switch ranges function is available in  
manual, auto, and remote modes. Only  
preset switch method is effective.  
Manual: Allows range to switch by key operation.  
Auto:  
Automatically switched from first range to  
second range when the measured value  
exceeds to 90%FS of first range.  
* Externally installed O2 sensor should  
be purchased separately.  
Digitalꢀoutput:ꢀ(Option)ꢀ  
Automatically switched from second range  
to first range when the measured value  
drops to 80% or less first range.  
1c contact (24V DC/1A, resistive load)  
max.15 outputs  
Instrument error, calibration error,  
range identification, auto calibration  
status, solenoid valve drive for auto  
calibration, High/Low limit alarm con-  
tact output.  
Remote: Voltage contact input (for measurable  
(Option) components)  
Allows range to switch via an external  
signal when remote range switch input  
is received.  
When the contact input terminals for  
each component are input voltage, the  
first range is selected, and it is switched  
to the second range when the terminals  
are open.  
* All relay contacts are isolated mutually  
and from the internal circuit.  
Digitalꢀinput:ꢀ(Option)  
Voltage contact (supply 12-24VDC (15mA  
Max.)) Max. 9 inputs  
Remote range change over, auto cali-  
bration remote start, remote hold,  
average value reset, Isolated from  
the internal circuit with photocoupler.  
* These range value are settable between original  
first range and second range.  
Powerꢀsupply:ꢀ Voltage rating  
; 100V to 240V AC  
; 85V to 264V AC  
; 50Hz/60Hz  
Optional Functions  
Allowable range  
Frequency  
Remoteꢀoutputꢀholding:ꢀ  
Power consumption ; 100VA max.  
Operationꢀconditions:ꢀ  
Output signal is held at the last value or  
preset value by voltage input the remote  
output holding input terminals.  
Ambient temperature ;  
-
5˚C to 45˚C  
Holding is maintained while the voltage  
input the terminals. Indication of instan-  
taneous values are not held.  
(40˚C max. when 2 optical sys-  
tem at 200V AC power source)  
Ambient humidity  
; 90% RH max.,  
non-condensing  
Rangeꢀidentificationꢀsignal:ꢀ  
The present measuring range is identified  
by a contact position.  
Storageꢀconditions:  
Ambient temperature ; -20˚C to 60˚C  
The contact output terminals for each  
component close when the first range  
is selected, and when the second range  
is selected, the terminals are open.  
Autoꢀcalibration:ꢀ  
Ambient humidity  
; 100% RH max.,  
non-condensing  
Dimensionsꢀ(Hꢀ×ꢀWꢀ×ꢀD):  
133 x 483 x 382mm  
Approx. 9 kg  
Front panel; Cool gray (PANTON 1C-F)  
Steel casing, for indoor use  
Mass:ꢀ  
Finishꢀcolor:ꢀꢀ  
Enclosure:  
Auto calibration is carried out periodically  
at the preset cycle.  
When a standard gas cylinder for calibra-  
tion and a solenoid valve for opening/  
closing the gas flow line are prepared  
externally by the customer, calibration  
will be carried out with the solenoid  
valve drive contacts for zero calibration  
and each span calibration turned on/off  
sequentially at the set auto calibration  
timing.  
Materialꢀofꢀgas-contactingꢀparts:  
Gas inlet/outlet; SUS304  
Sample cell; SUS304,chloroprene rubber  
Infrared-ray transmitting window; CaF2  
Internal piping;Toaron,Teflon, Polypropyl-  
ene  
Paramagnetic O2 sensor cell : SUS316  
Fuel cell O2 sensor cell : ABS resin  
2
Auto calibration cycle setting:  
Auto calibration cycle is set.  
Correction formula:  
21-On  
21-Os  
Setting is variable within 1 to 99 hours  
(in increments of 1 hour) or 1 to 40 days  
(in increments of 1 day).  
C =  
C
x Cs  
: Sample gas concentration after O2  
correction  
Gas flow time setting:  
The time for flowing each calibration gas  
in auto calibration is set.  
CS : Measured concentration of sample  
gas  
Settable within 60 to 900 seconds (in  
increments of 1 second)  
OS : Measured O2 concentration (Limit  
setting: 1 to 20% O2)  
Autoꢀcalibrationꢀremoteꢀstart:ꢀ  
On : Reference O2 concentration  
(value changeable by setting.0 to  
19% O2)  
Auto calibration starts by opening the auto  
calibration remote start input terminal  
after short circuiting for 1.5 sec or longer.  
Auto calibration starts when contacts  
open.  
AverageꢀvalueꢀafterꢀO  
calculation:ꢀ  
2
ꢀcorrectionꢀandꢀO ꢀaverageꢀvalue  
2
The result of O2 correction or instan-  
taneous O2 value can be output as an  
average value over the preset period of  
time.  
Autoꢀzeroꢀcalibration:  
Auto zero calibration is carried out periodi-  
cally at the preset cycle.  
This cycle is independent on “Auto cali-  
bration” cycle.  
Used for averaging is the moving average  
method in which sampling is carried out  
at intervals of 30 seconds.  
(Output is updated every 30 seconds. It  
is the average value in the determined  
period of time just before the latest up-  
dating.)  
Averaging time is settable within 1 to 59  
minutes (in increments of 1 minute) or 1  
to 4 hours (in increments of 1 hour).  
When zero calibration gas and solenoid  
valve for opening/closing the calibration  
gas flow line are prepared externally by  
the customer, zero calibration will be  
carried out with the solenoid valve drive  
contact for zero calibration turned on/off  
at the set auto zero calibration timing.  
Auto zero calibration cycle setting:  
Auto zero calibration cycle is set.  
Setting is variable within 1 to 99 hours  
(in increments of 1 hour) or 1 to 40 days  
(in increments of 1 day)  
Averageꢀvalueꢀresetting:ꢀ  
The above-mentioned output of average  
value is started from the initial state by  
opening the average value resetting input  
terminals after short circuiting for 1.5 sec  
or longer.  
Output is reset by input voltage and  
restarted by opening  
Gas flow time setting:  
The timing for flowing zero gas in auto  
zero calibration is set.  
Settable within 60 to 900 seconds (in  
increments of 1 second)  
Communicationꢀꢀfunction:  
RS-485 (9pins D-sub connector)  
High/lowꢀlimitꢀalarm:ꢀ  
Alarm contact output turns on when  
Half-duplex bit serial  
measurement value reaches the preset  
high or low limit alarm value.  
Start-stop synchronization  
ModbusTM protocol  
Contacts close when the instantaneous  
value of each channel exceeds the high  
alarm limit value or falls below the low  
alarm limit value.  
Contents : Read/Write parameters  
Read measurement concen-  
tration and instrument status  
Remark : When connecting via RS-  
232C interface, an RS-232C  
RS-485 converter should  
be used.  
Instrumentꢀerrorꢀcontactꢀoutput:ꢀ  
Contacts turn on at occurrence of analyzer  
error No. 1, 2, 3 or 10.  
Calibrationꢀerrorꢀcontactꢀoutput:ꢀ  
Contacts turn on at occurrence of manual  
Atmosphericꢀpressureꢀcorrection:  
Measure atmospheric pressure and cal-  
or auto calibration error (any of errors No.  
4 to 9).  
Autoꢀcalibrationꢀstatusꢀcontactꢀoutputs:  
Contacts turn on during auto calibration.  
culate compensation (for use, be sure to  
relieve the exhaust gas from analyzer to  
the atmosphere)  
After atmospheric pressure correction;  
Zero point : No influenced  
O2ꢀcorrection: Correction of measured NO, SO2 and CO  
gas concentrations into values at refer-  
ence O2 concentration  
Span point: The change is 0.5% mea-  
sured value or less relating  
to the change of the atmo-  
spheric pressure 1%.  
Correction range: 700hPa-1050hPa  
3
ZPA  
Corrosiveꢀcomponent:ꢀ1 ppm or less  
Standardꢀgasꢀforꢀcalibration:  
Performance  
Repeatability:ꢀ  
Linearity:ꢀ  
0.5% of full scaleꢀ  
1% of full scale  
prior to atmospheric pressure correction  
(option)  
2% of full scale/week  
In the case of Auto zero calibration use  
for 500ppm or less range.  
2% of full scale/week  
1) Infrared-ray measurable component, standard O2  
Zero gas ; Dry N2  
Span gas ; Each sample gas having con-  
centration 90 to 100% of its  
measuring range (recom-  
mended).  
In case a zirconia O2 analyzer is installed  
externally and calibration is carried out on  
the same calibration gas line:  
Zero gas ; Dry air or atmospheric air (Do  
not use with CO2 measure-  
ment)  
Zeroꢀdrift:ꢀ  
Spanꢀdrift:ꢀ  
Responseꢀtimeꢀ(forꢀ90%ꢀFSꢀresponse)ꢀ:ꢀ  
1 to 15 sec electrical response. Within  
10-30 seconds including replacement time  
of sampling gas.  
Gas replacement time depends on the  
number of measuring components,and  
measuring range.  
Span gas ; For other than O2 measure-  
ment, each sample gas hav-  
ing concentration 90 to 100%  
of its measuring range  
Interferenceꢀfromꢀotherꢀgases:ꢀ  
For O2 measurement, O2 gas  
of 1 to 2 vol%/remains N2 gas  
Interference  
component  
CO2  
CO  
CH4  
SO2  
NO  
analyzer analyzer analyzer analyzer analyzer  
2) Reverse range O2 measurement  
CO  
1.0%FS  
––  
1.0%FS  
1.0%FS 1.0%FS  
Zero gas ; 100% O2  
1000ppm  
*1)  
*2)  
Span gas ; For O2 measurement, O2 gas  
of 95.0%~95.5%/remains N2  
gas  
CO2 15%  
1.0%FS 1.0%FS 1.0%FS 1.0%FS  
––  
*2)  
H2O saturation  
at 20°C  
1.0%FS  
1.0%FS  
1.0%FS  
––  
––  
*For reverse range O2 measurement,  
infrared measurable component is not  
measurable.  
H2O saturation  
––  
2.0%FS  
––  
2.0%FS 2.0%FS  
at 2°C  
CH4  
1000ppm  
1.0%FS 1.0%FS  
––  
20ppm  
––  
*1) 0-200ppm range  
*2) 0-500ppm range or less  
2.0%FS  
2.0%FS  
Installation Requirements  
• Indoor use (Select a place where the equipment does  
not receive direct sunlight, draft/rain or radiation from hot  
substances. If such a place cannot be found, a roof or  
cover should be prepared for protection.)  
• Avoid a place where unit receives heavy vibration  
• Select a place where atmospheric air is clean  
EC Directive Compliance  
The product conforms to the requirements of the Low Volt-  
age Directive 2006/95/EC and EMC directive 89/336/EEC  
(as amended by Directive 92/31/EEC), both as amended by  
Directive 93/68/EEC.  
It conforms to following standards for product safety and  
electromagnetic compatibility ;  
EN61010-1 : 2010 Safety requirements for electrical equip-  
ment for measurement, control and laboratory  
use.  
“Installation Category II”  
“Pollution Degree 2”  
EN61326-1 : 2006  
Electrical equipment for measurement, control  
and laboratory use — EMC requirements.  
Standard Requirements for Sample Gas  
Flowꢀrate:ꢀ  
Temperature:ꢀ  
Pressure:ꢀ  
0.5L / min 0.2L / min  
0 to 50˚C  
10 kPa or less (Gas outlet side should be  
open to the atmospheric air.)  
100 µg/Nm3 or less in particle size of 0.3  
µm or less  
Dust:ꢀ  
Mist:ꢀ  
Unallowable  
Moisture:ꢀ  
Below a level where saturation occurs at  
room temperature (condensation unallow-  
able).  
Below the level where saturation occurs  
at 2˚C for CO measurement in 0 to 200  
ppm range, NO measurement, and SO2  
measurement.  
4
Principle diagram of NDIR type measurement (For NO, SO2, CO2, CO, CH4)  
Front expansion chamber  
Gas inlet  
Gas outlet  
Infrared-ray  
light source  
Rear  
expansion  
chamber  
Detector  
Mass flow  
sensor  
Motor  
Measuring cell  
Chopper  
Preamplifier  
Signal  
Indication  
processing  
and calculation  
block  
Output  
Principle diagram of fuel cell type measurment (For O2)  
Resistor  
Thermistor  
Electrolyte  
Negative  
electrode  
Diaphragm  
Positive  
electrode  
Principle diagram of paramagnetic type measurment (For O2)  
Mirror  
Permanent magnet  
Dumb-bell  
Gas inlet  
Magnetic field  
Measuring cell  
Gas outlet  
Preamplifier  
Permanent  
magnet  
Photo diode  
Light emitting diode  
Signal processing and  
calculation block  
Indication  
Output  
5
ZPA  
Examples of sampling system configuration including gas analyzer  
To measure low moisture content (saturated at room temperature or lower) sample gas (CO, CO2, CH4)  
9
Flow meter  
10  
Membrane  
filter  
1
3
Pump  
Mist filter  
Sample  
gas inlet  
7
Two-way  
Solenoid  
valve  
Gas  
analyzer  
(ZPA)  
7
7
7
Solenoid  
valve  
5
Drain pot  
11  
Demister  
O2 analyzer  
6
Ball valve  
(In the case of zirconia type)  
Exhaust  
6
Ball valve  
(atmospheric pressure)  
Drain  
Reference gas  
for zero  
calibration  
Reference gas for  
span calibration  
8
To measure high moisture content sample gas, NO, SO2, or CO (0 to 200 ppm range)  
13  
Solenoid  
valve  
9
Flow meter  
10  
Membrane  
filter  
1
3
Pump  
Mist filter  
Sample  
gas inlet  
12 NOX/NO converter  
7
Two-way  
Solenoid  
valve  
3
Pump  
(for NOX measurement)  
Atmospheric  
(Air)  
7
7
7
Solenoid  
valve  
1)  
4
Electronic  
cooler  
*
Gas  
analyzer  
(ZPA)  
2
Safety  
drain trap  
5
Drain  
pot  
O2 analyzer  
6
(In the case of  
Ball  
valve  
zirconia type)  
Reference gas  
for zero  
calibration  
Reference gas for  
span calibration  
Drain  
*
Exhaust  
(atmospheric  
pressure)  
8
1) Be sure to use a dehumidifier such as electronic cooler for NO, SO2, and CO  
analyzers of 0 to 200 ppm range (=2ºC saturation or lower).  
List of sampling devices (example)  
No.  
Device name  
Fujis type  
1
2
3
4
5
Mist filter  
ZBBK1V03-0  
ZBH51603  
Safety drain trap  
Pump  
ZBG80  
Electoric cooler  
Drain pot  
ZBC91004  
ZBH13003 (Length 255mm)  
ZBFB1  
Ball valve  
6
7
8
Two-way solenoid valve  
Standard gas for calibration  
Flow meter  
ZBM Y04-0 (Codes in to be selected depending on application)  
ZBD42203  
ZBBM2V03-0  
ZBH35003  
ZDL02001  
9
10  
11  
12  
13  
Membrane filter  
Demister  
NO2/NO converter  
Three-way solenoid valve  
Note) The above is a typical configuration example. As configuration may differ depending on measuring objects,  
please consult us.  
6
CODE SYMBOLS  
standard multi components  
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25  
Digit  
note  
Z P A  
B
1
Digit  
4
Description  
<Specification/Structure>  
Horizontal type (Terminal block for power supply)  
Horizontal type (Power inlet, with lock)  
<Mounting>  
19inch rack mounting type EIA conformity  
<Measurable component (NDIR)>  
1st component 2nd component 3rd component 4th component  
A
D
note1  
5
6
B
note2  
Y
P
A
D
B
E
F
G
J
K
L
None  
NO  
SO2  
CO2  
CO  
CH4  
NO  
NO  
CO2  
CH4  
CO2  
NO  
CO2  
NO  
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
SO2  
CO  
CO  
CO  
CH4  
SO2  
CO  
SO2  
-
-
-
N
T
V
Z
CO  
CH4  
CO2  
-
CO  
Others  
7
<Measurable component (O2)>  
None  
External O2 analyzer  
External zirconia O2 sensor (ZFK7)  
Built-in fuel cell O2 sensor  
Built-in paramagnetic O2 sensor  
<Revision code>  
Y
1
2
3
4
note3  
8
9
1
note4  
note4  
note4  
note4  
note4  
note4  
note4  
note4  
<Measuring range (NDIR)>1st component, 1st range  
10 <Measuring range (NDIR)>1st component, 2nd range  
11 <Measuring range (NDIR)>2nd component, 1st range  
12 <Measuring range (NDIR)>2nd component, 2nd range  
13 <Measuring range (NDIR)>3rd component, 1st range  
14 <Measuring range (NDIR)>3rd component, 2nd range  
15 <Measuring range (NDIR)>4th component, 1st range  
16 <Measuring range (NDIR)>4th component, 2nd range  
17 <Measuring range (O2)>  
None  
0-5/10%  
0-5/25%  
0-10/25%  
Y
A
B
C
L
0-5%  
0-10%  
0-25%  
0-50%  
M
V
P
0-100%  
100-95%  
Others  
R
S
Z
note5  
18 <Gas connection>  
1
2
Rc1/4  
NPT1/4  
19 <Output>  
A
B
C
D
DC0-1V  
DC4-20mA  
DC0-1V+Communication function  
DC4-20mA+Communication function  
20 <Indication/power supply cable>  
Japanese (PSE)  
note6  
note6  
note6  
note6  
note7  
J
E
U
C
English (UL)  
English (CEE)  
Chinese (CCC)  
21 <O2 correction and O2 correction average output>  
None  
O2 correction  
Y
A
B
C
O2 correction average  
O2 correction and O2 correction average  
22 <Optional function (DIO)>  
FAULT A. Cal. H/L Alarm RangeID/Remote range  
None  
Y
A
B
C
D
E
F
G
H
note8  
7
ZPA  
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25  
Digit  
note  
Z R E  
1
Description  
<Pressure compensation>  
None  
Digit  
23  
Y
1
Pressure compensation  
<Un>  
24  
25  
A
B
ppm, Vol%  
mg/m3, g/m3  
note9  
note10  
<Adjustment>  
For standard  
For heat treatment furnace  
For converter  
Others  
A
C
D
Z
RANGE CODE  
O measurement range  
2
Range  
None  
Code  
Y
B
C
D
S
E
F
Range  
Code  
J
K
Q
L
M
N
V
W
P
X
R
Z
Measurement Range Fuel cell O  
2
Paramagnet Zirconia O  
ic O sensor sensor  
(built - in) (built - in) (external)  
2
range  
code sensor  
2
0~1%  
0~2%  
0~3%  
0~5%  
0~100ppm  
0~200ppm  
0~250ppm  
0~300ppm  
0~500ppm  
0~1000ppm  
0~2000ppm  
0~2500ppm  
0~3000ppm  
0~5000ppm  
0~5/10 vol%  
0~5/25 vol%  
0~10/25 vol%  
0~5 vol%  
A
B
C
L
0~10%  
0~20%  
0~25%  
0~40%  
0~50%  
0~70%  
0~100%  
Others  
0~10 vol%  
0~25 vol%  
0~50 vol%  
0~100 vol%  
100~95 vol%  
M
V
P
R
S
G
U
T
H
note1)When “D” is specified at 4th digit, Power cable is supplied in the scope of supply. Cable specification should be  
specified at the 20th digit.  
note2)When only O2 measurement is necessary, "Y" should be specified at 6th digit.  
note3)When "1" is specified at 7th digit, O2 pt sensor signal has to be set as 0-1V DC linear corresponding to full scale.  
External zirconia O2 sensor and external O2 analyzer are not included in the scope of supply, and has to be separately  
ordered.  
note4)Refer toTable 1 for possible combination of measuring components and ranges in the data sheet. When "Y" is specified  
at 6th digit, "Y" should be specified at 9th to 16th digit. For fuel cell O2 sensor, range is 0-10% or more.  
note5)Select the type of voltage rating, plug type and applicable standard of the power cable by 20th digit. Select a power  
cable for using at the location of end-user. When A” is specified at 4th digit, Power cable will not be included in scope  
of delivery. When English is specified for display language, "E" should be selected at 20th digit.  
note6)O2 correction is calculated only for NO, SO2 and CO.  
note7)When 5 components measurement is specified, "H" must not be specified at 22nd digit.  
When 4 components measurement is specified and “H” is specified at 22nd digit, 3 points is maximum for alarm  
output function.  
note8) When "B" is specified at 24th digit, measuring range should be specified by ppm range code. In this case NO, SO2 and  
3
CO measuring range are corresponding range in mg/m . Please refer to the table shown below for the corresponding  
3
range code based on "mg/m ".  
note9) When A to D is specified at 25th digit, the analyzer will be adjusted and delivered with the following gasses. Standard  
"A": balance gas N2.  
For heat treatment furnace "C": balance gas 30% H2/remamins N2.  
For converter "D": balance gas CO, CO2.  
When other adjustment is required, please specify "Z".  
When "Z" is specified, please attach a list of gas composition contained in the measuring gas.  
Corresponding mg/m3  
Corresponding range in mg/m3  
Range code  
Unit : ppm  
0-200ppm  
0-250ppm  
0-300ppm  
0-500ppm  
0-1,000ppm  
0-2,000ppm  
0-2,500ppm  
0-3,000ppm  
0-5,000ppm  
NO  
SO2  
CO  
C
D
S
E
0-260mg/m3  
0-325mg/m3  
0-400mg/m3  
0-650mg/m3  
0-1,300mg/m3  
0-2,600mg/m3  
0-3,300mg/m3  
0-4,000mg/m3  
0-6,600mg/m3  
0-570mg/m3  
0-700mg/m3  
0-850mg/m3  
0-1,400mg/m3  
0-2,800mg/m3  
0-5,600mg/m3  
0-7,100mg/m3  
0-8,500mg/m3  
0-14.00g/m3  
0-250mg/m3  
0-300mg/m3  
0-375mg/m3  
0-600mg/m3  
0-1,250mg/m3  
0-2,500mg/m3  
0-3,000mg/m3  
0-3,750mg/m3  
0-6,250mg/m3  
F
The conversion formula "ppm" unit  
into "mg/m3" unit.  
G
U
T
NO (mg/m3) = 1.34 × NO (ppm)  
SO2 (mg/m3) = 2.86 × SO2 (ppm)  
CO (mg/m3) = 1.25 × CO (ppm)  
H
8
Table 1 Measurable component and range - availability check table -  
Procedure of range selection  
On one component analyzer:  
First determine 1st range, then select 2nd range from the right of your determination range in following tables.  
More than two components:  
The procedure is same as one component. Difference is 2nd range in the tables.  
The 2nd range in the tables for two and more components is maximum available range.  
2nd range is selectable from 1st range to 2nd range(max) on the table.  
1-component analyzer : CO  
1st range  
1-component analyzer : NO  
1st range  
2nd range  
2nd range  
0 - 200ppm None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm  
0 - 250ppm None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm  
0 - 300ppm None, 0 - 500ppm,1000ppm,2000ppm,2500ppm  
0 - 500ppm None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm  
0 - 1000ppm None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1%  
0 - 2000ppm None, 0 - 2500ppm,3000ppm,5000ppm,1%,2%  
0 - 2500ppm None, 0 - 3000ppm,5000ppm,1%,2%  
0 - 3000ppm None, 0 - 5000ppm,1%,2%  
0 - 200ppm None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm  
0 - 250ppm None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm  
0 - 300ppm None, 0 - 500ppm,1000ppm,2000ppm,2500ppm  
0 - 500ppm None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm  
0 - 1000ppm None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm  
0 - 2000ppm None, 0 - 2500ppm,3000ppm,5000ppm  
0 - 2500ppm None, 0 - 3000ppm,5000ppm  
0 - 3000ppm None, 0 - 5000ppm  
0 - 5000ppm None, 0 - 1%,2%,3%,5%  
0 - 5000ppm None  
0 - 1% None, 0 - 2%,3%,5%,10%  
0 - 2% None, 0 - 3%,5%,10%,20%  
0 - 3% None, 0 - 5%,10%,20%,25%  
1-component analyzer : SO2  
1st range  
2nd range  
0 - 200ppm None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm  
0 - 250ppm None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm  
0 - 300ppm None, 0 - 500ppm,1000ppm,2000ppm,2500ppm  
0 - 500ppm None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm  
0 - 1000ppm None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1%  
0 - 2000ppm None, 0 - 2500ppm,3000ppm,5000ppm,1%,2%  
0 - 2500ppm None, 0 - 3000ppm,5000ppm,1%,2%  
0 - 3000ppm None, 0 - 5000ppm,1%,2%  
0 - 5% None, 0 - 10%,20%,25%,40%,50%  
0 - 10% None, 0 - 20%,25%,40%,50%,70%,100%  
0 - 20% None, 0 - 25%,40%,50%,70%,100%  
0 - 25% None, 0 - 40%,50%,70%,100%  
0 - 40% None, 0 - 50%,70%,100%  
0 - 50% None, 0 - 70%,100%  
0 - 70% None, 0 - 100%  
0 - 100% None  
0 - 5000ppm None, 0 - 1%,2%,3%,5%  
0 - 1% None, 0 - 2%,3%,5%,10%  
1-component analyzer : CO2  
0 - 2% None, 0 - 3%,5%,10%  
0 - 3% None, 0 - 10%  
1st range  
2nd range  
0 - 5% None, 0 - 10%  
0 - 10% None  
1-component analyzer : CH4  
1st range  
0 - 500ppm None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm  
0 - 1000ppm None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1%  
0 - 2000ppm None, 0 - 2500ppm,3000ppm,5000ppm,1%,2%  
0 - 2500ppm None, 0 - 3000ppm,5000ppm,1%,2%  
0 - 3000ppm None, 0 - 5000ppm,1%,2%  
0 - 5000ppm None, 0 - 1%,2%,3%,5%  
0 - 1% None, 0 - 2%,3%,5%,10%  
0 - 100ppm None, 0 - 200ppm,250ppm,300ppm,500ppm,1000ppm  
0 - 200ppm None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm  
0 - 250ppm None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm  
0 - 300ppm None, 0 - 500ppm,1000ppm,2000ppm,2500ppm  
0 - 500ppm None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm  
0 - 1000ppm None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1%  
0 - 2000ppm None, 0 - 2500ppm,3000ppm,5000ppm,1%,2%  
0 - 2500ppm None, 0 - 3000ppm,5000ppm,1%,2%  
0 - 3000ppm None, 0 - 5000ppm,1%,2%  
2nd range  
0 - 5000ppm None, 0 - 1%,2%,3%,5%  
0 - 1% None, 0 - 2%,3%,5%,10%  
0 - 2% None, 0 - 3%,5%,10%,20%  
0 - 2% None, 0 - 3%,5%,10%,20%  
0 - 3% None, 0 - 5%,10%,20%,25%  
0 - 3% None, 0 - 5%,10%,20%,25%  
0 - 5% None, 0 - 10%,20%,25%,40%,50%  
0 - 5% None, 0 - 10%,20%,25%,40%,50%  
0 - 10% None, 0 - 20%,25%,40%,50%,70%,100%  
0 - 20% None, 0 - 25%,40%,50%,70%,100%  
0 - 25% None, 0 - 40%,50%,70%,100%  
0 - 40% None, 0 - 50%,70%,100%  
0 - 10% None, 0 - 20%,25%,40%,50%,70%,100%  
0 - 20% None, 0 - 25%,40%,50%,70%,100%  
0 - 25% None, 0 - 40%,50%,70%,100%  
0 - 40% None, 0 - 50%,70%,100%  
0 - 50% None, 0 - 70%,100%  
0 - 50% None, 0 - 70%,100%  
0 - 70% None, 0 - 100%  
0 - 70% None, 0 - 100%  
0 - 100% None  
0 - 100% None  
2-component analyzer : NO/SO2  
2-component analyzer : NO/CO  
1-component : NO  
2-component : SO2  
1-component : NO  
2-component : CO  
1st range 2nd range (max.)  
0 - 200ppm 0 - 2000ppm  
0 - 250ppm 0 - 2500ppm  
0 - 300ppm 0 - 2500ppm  
0 - 500ppm 0 - 5000ppm  
0 - 1000ppm 0 - 5000ppm  
0 - 2000ppm 0 - 5000ppm  
0 - 2500ppm 0 - 5000ppm  
0 - 3000ppm 0 - 5000ppm  
0 - 5000ppm None  
1st range 2nd range (max.)  
0 - 200ppm 0 - 2000ppm  
0 - 250ppm 0 - 2500ppm  
0 - 300ppm 0 - 2500ppm  
0 - 500ppm 0 - 5000ppm  
0 - 1000ppm 0 - 5000ppm  
0 - 2000ppm 0 - 5000ppm  
0 - 2500ppm 0 - 5000ppm  
0 - 3000ppm 0 - 5000ppm  
0 - 5000ppm None  
1st range 2nd range (max.)  
0 - 200ppm 0 - 2000ppm  
0 - 250ppm 0 - 2500ppm  
0 - 300ppm 0 - 2500ppm  
0 - 500ppm 0 - 5000ppm  
0 - 1000ppm 0 - 5000ppm  
0 - 2000ppm 0 - 5000ppm  
0 - 2500ppm 0 - 5000ppm  
0 - 3000ppm 0 - 5000ppm  
0 - 5000ppm None  
1st range 2nd range (max.)  
0 - 200ppm 0 - 2000ppm  
0 - 250ppm 0 - 2500ppm  
0 - 300ppm 0 - 2500ppm  
0 - 500ppm 0 - 5000ppm  
0 - 1000ppm 0 - 5000ppm  
0 - 2000ppm 0 - 5000ppm  
0 - 2500ppm 0 - 5000ppm  
0 - 3000ppm 0 - 5000ppm  
0 - 5000ppm None  
• The 2nd component should  
• The 2nd component should  
be selected as shown in the  
right table.  
be selected as shown in the  
right table.  
9
ZPA  
2-component analyzer: CO2/CO  
1-component: CO2  
2-component: CO  
1st range 2nd range (max.) 1st range/2nd range (max.)  
0-100ppm  
0-200ppm  
0-250ppm  
0-300ppm  
0-500ppm  
0-500ppm  
0-1000ppm  
0-2000ppm  
0-1000ppm  
0-2000ppm  
0-2000ppm  
0-2500ppm  
0-1000ppm  
0-2000ppm  
0-2500ppm  
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm  
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/3%,  
0-3%  
0-5000ppm  
0-5000ppm  
0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/3%, 0-3%  
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%,  
0-3/10%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%  
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%  
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%,  
0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%  
0-1%  
0-1%  
0-2%  
0-1%  
0-2500ppm  
0-3000ppm  
0-3000ppm  
0-2%  
0-1%  
0-2%  
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%,  
0-20/50%, 0-25/50%, 0-40/50%, 0-50%  
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%,  
0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/25%,  
0-10/100%,0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-5000ppm  
0-1%  
0-2%  
0-3%  
0-5%  
0-5%  
0-5%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,  
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-5000ppm  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%,  
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-1%  
0-2%  
0-10%  
0-20%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%,  
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%,  
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-2%  
0-3%  
0-5%  
0-10%  
0-20%  
0-25%  
0-40%  
0-50%  
0-70%  
0-100%  
0-10%  
0-25%  
0-50%  
0-100%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,  
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%,  
0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
None  
2-component analyzer: CH4/CO  
1-component: CH4  
2-component: CO  
1st range 2nd range (max.) 1st range/2nd range (max.)  
0-500ppm  
0-1000ppm  
0-1000ppm  
0-2000ppm  
0-5000ppm  
0-5000ppm  
0-1%  
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm  
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm  
0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm  
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%,  
0-3/20%, 0-5/20%, 0-10/20%  
0-5000ppm  
0-2500ppm  
0-3000ppm  
0-2000ppm  
0-2500ppm  
0-3000ppm  
0-2000ppm  
0-2500ppm  
0-3000ppm  
0-5000ppm  
0-5000ppm  
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1% 0-3000ppm/1%,  
0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/25%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/25%  
0-1%  
0-1%  
0-2%  
0-2%  
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%  
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/25%  
0-1%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%,  
0-25/50%, 0-40/50%, 0-50%  
0-5000ppm  
0-5000ppm  
0-1%  
0-3%  
0-5%  
0-5%  
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%  
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5% 0-2/5%, 0-3/25%, 0-5/25%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1% ,0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,  
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-1%  
0-2%  
0-2%  
0-3%  
0-3%  
0-5%  
0-5%  
0-10%  
0-10%  
0-10%  
0-20%  
0-10%  
0-25%  
0-25%  
0-50%  
0-50%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1% ,0-3000ppm/1%, 0-5000ppm/1%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/50%, 0-20/100%,  
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/100%, 0-20/100%,  
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/100%, 0-20/100%,  
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/100%, 0-20/100%,  
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%,  
0-50/100%, 0-70/100%, 0-100%  
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%,  
0-50/100%, 0-70/100%, 0-100%  
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/5%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%,  
0-50/100%, 0-70/100%, 0-100%  
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%,  
0-50/100%, 0-70/100%, 0-100%  
0-10%  
0-20%  
0-25%  
0-40%  
0-20%  
0-25%  
0-40%  
0-50%  
0-70%  
0-100%  
0-100%  
0-50%  
0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%,  
0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-100%  
None  
0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
10  
2-component analyzer: CO2/CH4  
1-component: CO2  
2-component: CH4  
1st range 2nd range (max.) 1st range/2nd range (max.)  
0-100ppm  
0-200ppm  
0-250ppm  
0-300ppm  
0-500ppm  
0-500ppm  
0-1000ppm  
0-1000ppm  
0-1000ppm  
0-2000ppm  
0-2000ppm  
0-2000ppm  
0-2500ppm  
0-2500ppm  
0-3000ppm  
0-5000ppm  
0-1000ppm  
0-2000ppm  
0-2500ppm  
0-2500ppm  
0-2500ppm  
0-5000ppm  
0-2500ppm  
0-5000ppm  
0-1%  
0-2500ppm  
0-5000ppm  
0-2%  
0-5000ppm  
0-2%  
0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm, 0-5000ppm  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/1%, 0-1%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/2%, 0-2/10%, 0-3/10%, 0-5/10%, 0-10%  
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/2%, 0-2/10%, 0-3/10%, 0-5/10%, 0-10%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%  
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3% 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%  
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%  
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%  
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%  
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/25%, 0-10/25%, 0-20/25%, 0-25%  
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/25%, 0-10/25%, 0-20/25%, 0-25%  
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5% 0-2/20%, 0-3/20%, 0-5/25%, 0-10/25%, 0-20/25%, 0-25%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%,  
0-40/50%, 0-50%  
0-2%  
0-3%  
0-5000ppm  
0-1%  
0-5%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/50%, 0-20/50%, 0-25/50%,  
0-40/50%, 0-50%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%,  
0-40/50%, 0-50%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/25%, 0-10/100%, 0-20/100%,  
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,  
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%,  
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,  
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%,  
0-50/100%, 0-70/100%, 0-100%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,  
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-10%  
0-20%  
0-10%  
0-25%  
0-20%  
0-50%  
0-20%  
0-50%  
0-2%  
0-2%  
0-3%  
0-3%  
0-5%  
0-5%  
0-10%  
0-10%  
0-20%  
0-25%  
0-40%  
0-10%  
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%,  
0-50/100%, 0-70/100%, 0-100%  
0-100%  
0-100%  
0-100%  
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%,  
0-70/100%, 0-100%  
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%,  
0-70/100%, 0-100%  
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%,  
0-70/100%, 0-100%  
0-20%  
0-25%  
0-40%  
0-50%  
0-70%  
0-100%  
None  
3-component analyzer: NO/SO2/CO >>> Combination of 1st component NO and 2nd component SO2 / 3rd component CO  
1-component: NO  
2-component: SO2  
3-component: CO  
1st range 2nd range (max.)  
1st range 2nd range (max.) 1st range/2nd range (max.)  
0-200ppm  
0-250ppm  
0-300ppm  
0-500ppm  
0-1000ppm  
0-2000ppm  
0-2500ppm  
0-3000ppm  
0-5000ppm  
0-2000ppm  
0-2500ppm  
0-2500ppm  
0-5000ppm  
0-5000ppm  
0-5000ppm  
0-5000ppm  
0-5000ppm  
None  
0-200ppm  
0-250ppm  
0-300ppm  
0-500ppm  
0-1000ppm  
0-2500ppm  
0-1000ppm  
0-2000ppm  
0-2500ppm  
0-3000ppm  
0-5000ppm  
0-2000ppm  
0-2500ppm  
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm  
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-5000ppm  
0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm, 0-5000ppm  
0-2500ppm  
None  
0-5000ppm  
None  
3-component analyzer: CO2/CO/CH4 >>>  
Combination of 1st component CO2 / 2nd component CO and 3rd component CH4  
1-component: CO2  
2-component: CO  
3-component: CH4  
1st range 2nd range (max.) 1st range/2nd range (max.)  
1st range 2nd range (max.) Availability of product  
0-5000ppm  
0-1%  
0-2%  
0-3%  
0-5%  
0-5%  
0-5%  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%,  
0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-5000ppm  
0-5%  
Product available only  
when CO analyzer max.  
measuring range is  
50% or less  
0-5000ppm  
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%,  
0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-500/5000ppm , 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%,  
0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-500/5000ppm , 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%,  
0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-500/5000ppm , 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%,  
0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-1%  
0-2%  
0-10%  
0-20%  
Product available  
0-1%  
0-2%  
0-10%  
0-20%  
Product available only  
when CO analyzer  
measuring range is 0 to  
1000ppm or more.  
Product available only  
when CO analyzer  
measuring range is 0 to  
5000ppm or more.  
Product available only  
when CO analyzer  
measuring range is more  
than 5000ppm or CO2  
analyzer range is more  
than 2%.  
0-3%  
0-25%  
0-10%  
0-20%  
0-25%  
0-40%  
0-50%  
0-5%  
0-10%  
0-20%  
0-25%  
0-40%  
0-2%  
0-3%  
0-5%  
0-10%  
0-20%  
0-25%  
0-40%  
0-50%  
0-70%  
0-100%  
0-10%  
0-25%  
0-50%  
0-100%  
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%,  
0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%,  
0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%  
0-50%  
0-70%  
0-100%  
0-70%  
0-100%  
None  
None  
11  
ZPA  
4-component analyzer: NO/SO2/CO2/CO >>> Combination of 1st component NO /4th component CO and component 2nd component SO2/3rd component CO2  
1-component: NO  
4-component: CO  
1st range 2nd range (max.) 1st range/2nd range (max.)  
0-200ppm  
0-250ppm  
0-300ppm  
0-500ppm  
0-1000ppm  
0-2000ppm  
0-500ppm  
0-1000ppm  
0-2000ppm  
0-2500ppm  
0-3000ppm  
0-5000ppm  
0-2000ppm  
0-2500ppm  
0-2500ppm  
0-2000ppm  
0-2000ppm  
None  
0-5000ppm  
0-5000ppm  
0-5000ppm  
0-5000ppm  
0-5000ppm  
None  
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm, None  
0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm, None  
2-component analyzer: SO2 3-component analyzer: CO2  
1st range 2nd range (max.) 1st range/2nd range (max.)  
0-200ppm  
0-250ppm  
0-300ppm  
0-500ppm  
0-1000ppm  
0-2000ppm  
0-2500ppm  
0-3000ppm  
0-5000ppm  
0-2000ppm  
0-2500ppm  
0-2500ppm  
0-5000ppm  
0-5000ppm  
0-5000ppm  
0-5000ppm  
0-5000ppm  
None  
0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%/None  
Table 2 Channel (Ch) No. and display/output contents comparison table  
Code symbol  
6th digit 7th digit 21st digit Display/output contents  
Y
P
A
D
B
E
F
1 to 3  
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Ch1:O2  
Ch1:NO  
Ch1:SO2  
Ch1:CO2  
Ch1:CO  
Ch1:CH4  
Ch1:NO, Ch2:SO2  
Y
Y
G
J
Y
Y
Y
Y
Ch1:NO, Ch2:CO  
Ch1:CO2, Ch2:CO  
K
L
Y
Y
Y
Y
Ch1:CH4, Ch2:CO  
Ch1:CO2, Ch2:CH4  
N
T
V
P
A
D
B
E
F
G
J
K
L
N
T
V
P
A
B
F
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
A *  
A *  
A *  
A *  
A *  
A *  
A
A
C
C
C
C
Ch1:NO, Ch2:SO2, Ch3:CO  
Ch1:CO2, Ch2:CO, Ch3:CH4  
Ch1:NO, Ch2:SO2, Ch3:CO2, Ch4:CO  
Ch1:NO, Ch2:O2  
Ch1:SO2, Ch2:O2  
Ch1:CO2, Ch2:O2  
Ch1:CO, Ch2:O2  
Ch1:CH4, Ch2:O2  
Ch1:NO, Ch2:SO2, Ch3:O2  
Ch1:NO, Ch2:CO, Ch3:O2  
Ch1:CO2, Ch2:CO, Ch3:O2  
Ch1:CH4, Ch2:CO, Ch3:O2  
Ch1:CO2, Ch2:CH4, Ch3:O2  
Ch1:NO, Ch2:SO2, Ch3:CO, Ch4:O2  
Ch1:CO2, Ch2:CO, Ch3:CH4, Ch4:O2  
Ch1:NO, Ch2:SO2, Ch3:CO2, Ch4:CO, Ch5:O2  
Ch1:NOx, Ch2:O2, Ch3:corrected NOx  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
1 to 3  
Ch1:SO2, Ch2:O2, Ch3:corrected SO2  
Ch1:CO, Ch2:O2, Ch3:corrected CO  
Ch1:NOx, Ch2:SO2, Ch3:O2, Ch4:corrected NOx, Ch5:corrected SO2  
Ch1:NOx, Ch2:CO, Ch3:O2, Ch4:corrected NOx, Ch5:corrected CO  
Ch1:CO2, Ch2:CO, Ch3:O2, Ch4:corrected CO  
Ch1:NOx, Ch2:SO2, Ch3:CO, Ch4:O2, Ch5:corrected NOx, Ch6:corrected SO2, Ch7:corrected CO  
Ch1:NOx, Ch2:SO2, Ch3:CO2, Ch4:CO, Ch5:O2, Ch6:corrected NOx, Ch7:corrected SO2, Ch8:corrected CO  
Ch1:NOx, Ch2:O2, Ch3:corrected NOx, Ch4:corrected NOx average  
Ch1:SO2, Ch2:O2, Ch3:corrected SO2, Ch4:corrected SO2 average  
Ch1:CO, Ch2:O2, Ch3:corrected CO, Ch4corrected CO average  
Ch1:NOx, Ch2:SO2, Ch3:O2, Ch4:corrected NOx, Ch5:corrected SO2, Ch6:corrected NOx average,  
Ch7:corrected SO2 average  
G
J
*
*
*
*
*
*
N
V
P
A
B
F
*
G
1 to 3  
C
Ch1:NOx, Ch2:CO, Ch3:O2, Ch4:corrected NOx, Ch5:corrected CO, Ch6:corrected NOx average,  
Ch7:corrected CO average  
*
*
J
N
1 to 3  
1 to 3  
C
C
Ch1:CO2, Ch2:CO, Ch3:O2, Ch4:corrected CO, Ch5:corrected CO average  
Ch1:NOx, Ch2:SO2, Ch3:CO, Ch4:O2, Ch5:corrected NOx, Ch6:corrected SO2, Ch7:corrected CO,  
Ch8:corrected NOx average, Ch9:corrected SO2 average, Ch10:corrected CO average  
Ch1:NOx, Ch2:SO2, Ch3:CO2, Ch4:CO, Ch5:O2, Ch6:corrected NOx, Ch7:corrected SO2, Ch8:corrected CO,  
Ch9:corrected NOx average, Ch10:corrected SO2 average2, Ch11:corrected CO average  
*
V
1 to 3  
C
* When the 21st digit code is A or C, the component of the NO analyzer is displayed as NOx.  
12  
OUTLINE DIAGRAMS (Unit : mm)  
Refer to appendix. (TC drawing)  
429  
<TOP VIEW>  
<FRONT VIEW>  
415  
463  
483  
Power Switch  
Digital Input/output Connector (DIO1 to 3)  
Analog Output Connctor (A/O)  
Communication Connector (RS485)  
Analog Input Connector (A/I)  
Purge Gas Inlet Rc1/4 or NPT1/4  
Sample Gas Inlet Rc1/4 or NPT1/4  
Sample Gas Outlet Rc1/4 or NPT1/4  
Fuse  
<REAR VIEW>  
<SIDE VIEW>  
(M4Terminal BlockType)  
Power Inlet(100V to 240V AC 50/60Hz)  
25  
2
380  
396  
24  
16  
M4Terminal BlockType)  
(Power Inlet Type(100V to 240V AC 50/60Hz))  
13  
ZPA  
Mounting method  
The analyzer weight should be supported at the bottom of the case.  
19-inch rack mounting type  
<Rack dimensions>  
450 or more  
465  
Mounting diagram  
Guide rails  
Guide rails  
EXTERNAL CONNECTION  
<Analog output> A/O connector  
<RS485 communication signal>  
1
2
1
2
3
4
5
13  
1
5
1
AO1+  
AO1  
AO2+  
AO2  
AO3+  
AO3  
AO4+  
AO4  
AO5+  
AO5  
AO6+  
AO6  
AO7+  
AO7  
AO8+  
AO8  
AO9+  
AO9  
AO10+  
AO10  
AO11+  
AO11  
AO12+  
(GND)  
14  
15  
16  
17  
18  
19  
20  
21  
22  
23  
24  
25  
6
7
8
9
-
RT×D+  
25  
14  
9
6
-
3
RT×D  
-
D-sub 25pins female  
-
D-sub 9pins female  
4
* In standard, displayed Channel No.  
and Analog Output No. are same.  
-
5
-
6
<Screw terminal (M4)>  
-
7
P
N
L
-
8
<Analog input> A/I connector (O2 signal input)  
-
100 to 240V AC  
9
-
AI  
-
10  
11  
12  
13  
-
<Power inlet>  
-
AI+  
POWERꢀINLET  
100-240VꢀAC  
AO12  
-
NC  
14  
<Digital I/O> DIO 1 to 3 connector (option)  
DIO1  
connector connector connector  
DI1+ DI4+ DI7+  
DI1 DI4 DI7  
DI2+ DI5+ DI8+  
DI2 DI5 DI8  
DI3+ DI6+ DI9+  
DI3 DI6 DI9  
DIO2  
DIO3  
13  
1
1
2
14  
15  
16  
17  
18  
19  
20  
21  
22  
23  
24  
25  
-
-
-
Digital input  
OFF : 0V  
ON : 12 to 24V DC  
25  
14  
-
-
-
3
D-sub 25pins female  
-
-
-
4
NC  
* DIO 1 to 3 are all as same connector.  
com  
NO  
NC  
com  
NO  
NC  
com  
NO  
NC  
com  
NO  
NC  
DO1  
DO2  
DO3  
DO4  
DO5  
DO6  
DO7  
DO8  
DO9  
DO10  
DO11  
DO12  
DO13  
DO14  
DO15  
5
6
Digital output  
max. contact load  
rating 24V DC/1A  
7
Contents of digital input  
signal  
8
9
DI1  
DI2  
DI3  
DI4  
DI5  
DI6  
DI7  
DI8  
DI9  
Remote hold  
Average value reset  
A. cal. start  
10  
11  
12  
13  
com  
NO  
A. zero. cal. start  
Remote range Ch1  
Remote range Ch2  
Remote range Ch3  
Remote range Ch4  
Remote range Ch5  
Contents of digital output signal  
Independent on the  
number of component  
1-component analyzer  
2-component analyzer 3-component analyzer  
22th digit  
DO1  
A, C  
B, E  
D, F, G, H  
B, D, E, F, G, H  
Instrument error  
Calibration error  
(A.cal.status)  
B, D, E, F, G, H  
Instrument error  
Calibration error  
(A.cal.status)  
Instrument error  
Calibration error  
Instrument error  
Calibration error  
A.cal.status  
For zero gas  
For span gas Ch1  
(Alarm1)  
Instrument error  
Calibration error  
(A.cal.status)  
(For zero gas)  
DO2  
The items in the parentheses  
may not be available depend-  
ing on the selected type on  
22th digit.  
DO3  
DO4  
(For zero gas)  
(For zero gas)  
(For span gas Ch1)  
(For span gas Ch2)  
(For span gas Ch3)  
(Range identification Ch1)  
(Range identification Ch2)  
(Range identification Ch3)  
(Alarm1)  
DO5  
(For span gas Ch1) (For span gas Ch1)  
(For span gas Ch2)  
DO6  
(Alarm1)  
(Alarm2)  
(Alarm3)  
(Alarm4)  
(Alarm5)  
DO7  
(Alarm2)  
The normal open side (NO) of  
digital output is close when  
the function is active without  
range ID.  
DO8  
(Alarm3)  
DO9  
(Alarm4)  
(Range identification Ch1)  
DO10  
DO11  
DO12  
DO13  
DO14  
DO15  
(Alarm5)  
Range identification Ch1 (Range identification Ch2)  
(Alarm1)  
(Alarm2)  
(Alarm3)  
(Alarm4)  
(Alarm5)  
(Alarm1)  
(Alarm2)  
(Alarm3)  
(Alarm4)  
(Alarm5)  
In case of range ID, normal  
open (NO) side is close with  
First range.  
The normal close (NC) side is  
close with Second range.  
(Alarm2)  
(Alarm3)  
(Alarm4)  
(Alarm5)  
4-component analyzer  
5-component analyzer  
B, E D, F  
22th digit  
DO1  
B, E  
D, F  
G
H
G
Instrument error Instrument error  
Calibration error Calibration error  
A.cal.status  
Instrument error  
Calibration error  
A.cal.status  
Instrument error  
Calibration error  
A.cal.status  
Instrument error Instrument error  
Calibration error Calibration error  
A.cal.status  
Instrument error  
Calibration error  
A.cal.status  
DO2  
DO3  
DO4  
For zero gas  
For zero gas  
For zero gas  
For zero gas  
For zero gas  
DO5  
For span gas Ch1  
For span gas Ch1  
For span gas Ch2  
For span gas Ch1  
For span gas Ch2  
For span gas Ch3  
For span gas Ch4  
For span gas Ch1  
For span gas Ch1  
DO6  
For span gas Ch2  
For span gas Ch2 Range identification Ch1 For span gas Ch2  
For span gas Ch3 Range identification Ch2 For span gas Ch3  
For span gas Ch4 Range identification Ch3 For span gas Ch4  
DO7  
For span gas Ch3 Range identification Ch1 For span gas Ch3  
For span gas Ch4 Range identification Ch2 For span gas Ch4  
Range identification Ch3  
DO8  
DO9  
Range identification Ch1 For span gas Ch5 Range identification Ch4 For span gas Ch5  
DO10  
DO11  
DO12  
DO13  
DO14  
DO15  
Range identification Ch4  
Range identification Ch2  
(Alarm1)  
Range identification Ch5  
(Alarm1)  
(Alarm1)  
(Alarm2)  
(Alarm3)  
(Alarm4)  
(Alarm5)  
(Alarm1)  
(Alarm2)  
(Alarm3)  
(Alarm4)  
(Alarm5)  
(Alarm1)  
(Alarm2)  
(Alarm3)  
Range identification Ch1  
Range identification Ch2  
Range identification Ch3  
Range identification Ch4  
Range identification Ch5  
Range identification Ch1 (Alarm2)  
Range identification Ch2 (Alarm3)  
(Alarm2)  
(Alarm3)  
Range identification Ch3 Range identification Ch3 (Alarm4)  
Range identification Ch4 Range identification Ch4 (Alarm5)  
(Alarm4)  
(Alarm5)  
SCOPE OF DELIVERY  
ORDERING INFORMATION  
•ꢀGasꢀanalyzerꢀ…ꢀ1ꢀunit  
1.Codeꢀsymbols  
2.ꢀApplicationꢀandꢀcompositionꢀofꢀsampleꢀgas  
•ꢀReplacementꢀfuseꢀ(250V,ꢀ2AꢀAC,ꢀdelayꢀtype)ꢀ…ꢀ2ꢀpcs  
•ꢀInstructionꢀmanualꢀ…ꢀ1ꢀcopy  
•ꢀConnectorꢀforꢀI/Oꢀconnectionꢀ…ꢀ1ꢀset  
•ꢀPowerꢀcable(standardꢀinletꢀtypeꢀ2m)ꢀ…ꢀ1ꢀpc  
15  
ZPA  
Gasꢀinlet/outletꢀsize:  
Rc /4 or NPT /4  
Powerꢀsupply:ꢀ Rated voltage  
1
1
Exclusive Zirconia O2 Sensor  
(to be purchased separately)  
For O2 correction, the gas analyzer ZPA can accept linearized  
0 to 1V DC signal coming from analyzer calibrated 0 to 25%  
O2 full scale. If the analyzer is not available, Fuji can supply  
exclusive Zirconia O2 sensor Model ZFK.  
Measuringꢀmethod:  
; 100 to 115V AC or  
200 to 240V AC  
Rated frequency ; 50Hz/60Hz  
Max. rated power; 215VA (during power  
ON)  
65VA (during steady-  
state operation)  
Enclosure:ꢀ  
Indication:ꢀ  
Steel casing, for indoor application  
Temperature indication (LED)  
Zirconia system  
Measurableꢀcomponentꢀandꢀmeasuringꢀrange:  
Temperatureꢀalarmꢀoutput:  
Contact output 1a contact,  
Measurable component  
O2 Oxygen  
Range  
0 to 25vol%  
Contact capacity 220V, 1A AC (resistive  
load)  
Outerꢀdimensionsꢀ(HꢀxꢀWꢀxꢀD):  
Repeatability:ꢀ Within 0.5% of full scale  
Linearity:ꢀ  
Zeroꢀdrift:ꢀ  
Spanꢀdrift:ꢀ  
Within 1% of full scale  
Within 1% of full scale/week  
Within 2% of full scale/week  
141 x 170 x 190mm  
Massꢀ{weight}:ꢀ Approx. 3kg  
Finishꢀcolor:ꢀ  
Munsell 5Y 7/1  
Responseꢀtime:ꢀ Approx. 20 seconds (for 90% response)  
Measuredꢀgasꢀflowꢀrate:  
0.5 0.25L / min  
CODE SYMBOLS  
Remark: The Zirconia system, due to its principle,  
may produce a measuring error due to rela-  
tive concentration versus the combustible  
1
2
3
4
5
6
7
8
9
10 11 12 13  
Y 0 Y Y  
Description  
Measuring method  
Zirconia method  
Z F K 7 Y Y  
4
-
O
2
gas concentration. Also, a corrosive  
gas (SO of 250 ppm or more, etc.) may  
affect the life of the sensor.  
7YY  
2
Power supply  
9
B
C
100 to 115V AC 50/60Hz(Standard)  
200 to 240V AC 50/60Hz(Standard)  
200 to 240V AC 50/60Hz(CE mark)  
OUTLINE DIAGRAM (Unit:mm)  
Gas inlet/outlet size  
Rc 1/4  
NPT 1/4  
1
8
170  
152  
OXYGEN ANALYZER  
TEMP.  
1
2
4
5
AC250V  
T3.15A  
3
6
7
L
N
AC  
OUT  
ALM  
INPUT  
9TH DIGIT  
1
1: RC /4  
8: NPT1/4  
ø5  
OUTLET  
INLET  
EXTERNAL CONNECTION DIAGRAM  
1
2
3
4
5
6
7
+
-
AC power supply  
E
Output  
Temperature  
to analyzer alarm output  
Caution on Safety  
*Before using this product, be sure to read its instruction manual in advance.  
International Sales Div  
Sales Group  
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,  
Shinagawa-ku, Tokyo 141-0032, Japan  
http://www.fujielectric.com  
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425  
http://www.fujielectric.com/products/instruments/  
Information in this catalog is subject to change without notice.  
Printed in Japan  

相关型号:

EDS3-144

INFRARED GAS ANALYZER(HIGH PERFORMANCE MODEL)
FUJI

EDS3-145

INFRARED GAS ANALYZER(LOW-CONCENTRATION MEASUREMENT TYPE)
FUJI

EDS3-1I

SAMPLING DEVICE SERIES FOR GAS ANALYZER
FUJI

EDS3-212A

MULTI-CHANNEL ZIRCONIA OXYGEN ANALYZER
FUJI

EDS3-61H

THERMAL CONDUCTIVITY GAS ANALYZER THERMOMAT
FUJI

EDS3-62H

PARAMAGNETIC OXYGEN ANALYZER
FUJI

EDS3-69F

PARAMAGNETIC OXYGEN ANALYZER
FUJI

EDS5-101A

FC SERIES PRESSURE TRANSMITTER
FUJI

EDS5-102B

FC SERIES ABSOLUTE PRESSURE TRANSMITTER
FUJI

EDS5-103A

FC SERIES PRESSURE TRANSMITTER WITH REMOTE SEALING
FUJI

EDS5-14D

FCX SERIES ABSOLUTE PRESSURE TRANSMITTER
FUJI

EDS5-15F

FCX SERIES REMOTE SEAL TYPE PRESSURE TRANSMITTER
FUJI