IN-SITU ZIRCONIA OXYGEN ANALYZER < ZIRCOMAT-C >
DATA SHEET
ZFK, ZRY
This oxygen analyzer is used to continuously measure
oxygen concentration in combustion exhaust gas of industrial
boilers or furnaces, and is ideally suited for combustion moni-
toring and control.
The detector (ZFK) used with the analyzer is directly
inserted into the objects measured, eliminating the need for a
sampling device and provides quick response.
The converter (ZRY) features 3 measuring ranges, one
touch calibration and NEMA4 housing.
General-use detector
Comparing to the current converter (ZRM) , this ZRY is
recognized as economical type with simple functions.
FEATURES
1. Sampling device is unnecessary
Gas sampling devices such as a gas aspirator, a dehu-
midifier, etc. are unnecessary because of use of direct-
insertion type detector.
2. High speed response
The adoption of a flow guide tube utilizing the flow of the
measured gas assures quick response (less than 7sec).
3. Dust-tight and water-proof housing
Converter
Aluminum die casting housing of converter satisfies
NEMA4 and IP65 standard.
4. 3-measuring ranges and one touch calibration
The converter has 0-5, 10, 25 vol% 3 ranges easily
changed by pin connection and zero and span are simply
calibrated by key operation.
Oxygen detector (ZFK2, 5)
Measuring detector:
For general-use: ZFK2
For corrosive gas: ZFK5
Measured gas temperature:
Flow guide tube system; –20 to +600°C
Measured gas pressure:
–3 to +3kPa
Flow guide tube: Flange; JIS5K 65A FF
SPECIFICATIONS
(JIS5K-80AFF for high particulate gas)
Insertion length; 0.3, 0.5, 0.75, 1m
(0.8m for high particulate gas)
General
Measuring object: Oxygen contained in noncombustible
Ambient temperature:
gas
–20 to +60°C for cable section
125°C or less at detector flange surface
with power applied
Dust/rain-proof structure(IEC IP55
equivalent)
Measuring principle:
Direct-insertion zirconia system
Measuring range: 0 to 5, 10 or 25 vol% O2 (Changed by
internal set pin)
Structure:
Filter:
Repeatability:
Linearity:
Within ±1.0% of full scale
±2% of full scale
Alumina(filtering accuracy 50µm) and
quartz paper
Response time:
Within 7sec for 90% response (from
calibration gas inlet)
Main materials of gas-contacting parts:
General-use detector(ZFK 2); Zirconia,
Power supply:
100, 115, 220 or 230V AC, 50/60Hz
SUS316, platinum, SUS304
Power consumption:
Anticorrosive detector(ZFK 5); Zirconia,
tita-nium, platinum, SUS316
Flow guide tube; SUS304 or SUS316
(approx.)
15 + 50VA (at steady state)
15 + 200VA (at start)
Approx. 15min
Warmup time:
EDSX3-119b
Date
Mer. 6, 2007
ZFK, ZRY
Calibration gas inlet:
Exclusine cable (ZRZP)
Brass joint for φ1/4 inch tube.
Reference air inlet (option):
Rc1/8 or NPT1/8
Detector mounting:
Cable:
4-cores sealed wire (O2 signal and R-
themo couple signal) and 2-cores cable
(power)
Flexible type
Refer to code symbols
Cable conduit:
Length:
Horizontal plane ± 45°, ambient sur-
rounding air should be clean.
Outer dimensions: (L x max. dia.) 210mm x 100mm
(detector)
Mass (approx.) {weight}:
SCOPE OF DELIVERY
Detector; 1.6kg
Flow guide tube (general-use, 1m); 5kg
Detector:
Detector main unit x 1, Viton O ring x 1,
Finish color:
Silver and SUS metallic color
mounting screw (M5mm x 12) x 6,
thermal sticker x 1, flow guide tube (as
specified) x 1, ceramic filter x 1, cover (as
specified) x 1, reference air inlet (as
specified) x 1
Converter main unit x 1
Accessories (AC250V 500mA T fuse x 1,
AC250V 3.15A T fuse x 1)
Oxygen converter (ZRY)
Measuring range: 0-5, 10, 25 vol% O2
Changeable by internal set pin.
±1.0% of full scale
±2.0% of full scale
Repeatability:
Lineality:
Converter:
Indication:
Oxygen concentration; 3-digit LED
Oxygen concentration output signal:
4 to 20mA DC (allowable load resistance:
500Ω or less)
Isolated output, linear
Fault contact output:
Items to be prepared separately:
(1) Standard gas for calibration
Type ZBMMNSK4-01
(2) Reduction valve for standard gas (type ZBD61003)
(3) Flowmeter
250V AC, 2A rating (close contact or
open contact for fault should be specified
when you place an order.)
Type; ZBD52203, 0.2 to 2L/min (for calibrating gas)
Self-diagnoses:
Fault of sensor temperature, zero
calibration error, span calibration error
CAUTIONS
Calibration method:
Manual calibration with key operation
Calibration gas: • Recommended calibration gas concen-
•
If combustible gas (CO, H2 etc.) exists in the measured
gas, error will occur due to burning at the sensor section.
The inclusion of corrosive gas (Si vapor, alkaline metal, P,
Pb etc.) will shorten the life of the sensor.
When the measured gas temperature is high (+300°C or
higher), the flange should be separated from the furnace
wall in order to bring the detector flange surface
temperature below the specified value +125°C). The flow
guide should be attached in the direction in which the gas
flow to the detector decreases.
tration
Zero gas; 1.0 to 2.0% O2
Span gas; 20.6 to 21.0% O2
(oxygen concentration in the
air)
•
Ambient temperature:
–10 to +50°C
When dust more than 1g/Nm3 is included in the gas,
manual blow down is necessary.
Ambient humidity: 90% RH or less
•
Power supply:
Construction:
90 to 220 or 230 V AC, 50/60Hz
Dust-tight, waterproof construction,
NEMA4 (corresponding to IP65 of IEC)
Aluminum die casting
Material:
Outer dimensions (H x W x D):
220 X 230 X 95mm
Mass {weight}:
Approx. 4.5kg (excluding cable and de-
tector)
Finish color:
Munsell 6PB 3.5/10.5 (blue): cover,
silver; case
Mounting method: Mounted flush on panel
The product conforms to the requirements of the
Electromagnetic compatibility Directive 89/336/EEC as
detailed within the technical construction file number
TZ737041. The applicable standards used to demonstrate
compliance are :
EN 55011 : 1992 CLASSA Conducted and Radiated emissions
EN 50082-1 : 1992
Radiated immunity, ESD and FBT
ZFK, ZRY
2
DEVICE CONFIGURATION
The device to be combined differ according to the conditions of the gas to be measured. Please select the devices to be
combined with reference to the following table.
Measured gas
Device configuration
Detector type Converter
Temperature Gas Flow
Application
Protection
cover
Note
DUST
Ejector
type
type
Less than 0.2g/m3 [nor]
—
—
—
600°C or
5 to 20m/s
5
6
General-use
(boiler)
Fuel; gas, oil
ZFK
ZFK
R
R
4-
4-
ZRY
ZRY
Less than 10g/m3 [nor] no
less
8
Fuel: coal
D
E
Anti-corrosive
use
with manual blow down
Less than 25g/m3 [nor]
—
6
8
Included high moisture ZFK
with manual blow down
R
4-
ZRY
yes
(incinerator)
Note (1) Dust volume is approximate value.
(2) Instrument quality air or bottled air is available as reference air by selecting detector with reference air inlet, when oxygen concentration in air
around sensor changes.
CODE SYMBOLS
(Converter)
(Detector)
4
5
6
7
8
9 10 11 12 13
1
2
3
4
5
6
7
8
9 10 11 12
Description
ZFK
R 2
4
Description
Z R Y 1 Y 1 2 - 1 R
0
Application
General use.
For corrosive gas (refuse incinerator)
Output signal and fault output
4 to 20mA DC, close contact
4 to 20mA DC, open contact
2
5
B
C
Power supply
Cal. gas inlet
1
90 to 230V AC 50/60Hz
(CE marking approved)
2
For φ1/4 inch tube
Power supply
Mounting method
Panel mounting
1
3
5
100/115VAC 50/60Hz
200/220VAC 50/60Hz
230VAC 50/60Hz (CE-marking approved)
1
Instruction manual
NO
YES (English)
Y
E
Flow guide tube
flange
application
None
length
0 Y 0
5 A 3
5 A 5
5 A 7
5 A 1
5 B 3
5 B 5
5 B 7
5 B 1
6 D 8
6 E 8
SUS304 general use
SUS304 general use
SUS304 general use
SUS304 general use
SUS316 for corrosive gas
SUS316 for corrosive gas
SUS316 for corrosive gas
SUS316 for corrosive gas
SUS316 for high particulate
300mm
500mm
750mm
1000mm
300mm
500mm
750mm
1000mm
800mm
(Exclusive-special cable)
1
2
3
4
5
6
7
8
9
Description
Connectable devices
Z R Z P R
1 -
P
For ZRY
Types
For R thermocouple
SUS316 for high particulate with 800mm
cover
R
Conduit length
None
None
None
None
None
None
None
None
None
None
None
None
6m
Cable length
6m
Protection cover
Without
With
YA
YB
YC
YD
YE
YF
YG
YH
YJ
Y
A
10m
15m
20m
30m
40m
50m
60m
70m
80m
90m
100m
6m
Reference air inlet
Non
Rc1/8
NPT1/8
Y
A
B
YK
YL
( Replacement Detector element)
YM
AA
BB
CC
DD
4
5
6
7
8
9 10 11 12 13
Description
ZFK
R 2
4
10m
15m
20m
10m
15m
20m
Application
General use.
For corrosive gas (refuse in cinerator)
2
5
Cable end treatment
None
One side (detector side)
Both sides
Cal. gas inlet
For φ1/4 inch tube
0
1
2
2
Power supply
1
3
5
100/115VAC 50/60Hz
200/220VAC 50/60Hz
230VAC 50/60Hz
Note: For connection between detector and
converter, the conduit to be used should
be rainproof flexible type.
3
ZFK, ZRY
CONFIGURATION
Flow guide tube system
Cover for retaining warmth
Flow guide tube
Status of fault contact output
5 digit of
model
Power off
Power on and
normal
Power on and
fault
Detector
ZFK2
B
C
open
open
open
close
close
open
Gas temperature.
600°C max.
*3
1
2
3
4
5
6
Ball valve
Rainproof flexible conduit.
20m max.
Bule
Black
Red Yellow
White White
Sample gas outlet
4-core
cable
Converter (ZRY)
2-core
cable
Fault
contact
output
Thermocouple Zirconia
Analog
output
Sensor
heater
Gas temperature.
600°C max.
signal
signal
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Flow guide tube
for high particulate
Red White Blue Yellow
Black White
DC4 to
20mA
6mm or 4 inch tube
for calibration gas
(not supplied)
*3
AC power
supply
Cell/thermocouple output
Heater power supply
Ball valve
Sample gas outlet
Gas temperature.
600°C max.
Flow guide for high
particulate with cover
Reduction
valve (ZBD6)
Change valve
Standrad gas
for zero point
(ZBM)
Flowmeter
(ZBD4)
(1 to 1.5L/min)
Reduction valve (unnecessary when supply air pressure is
50 to 100kPa
*2
Ref. air
*1 Air supply
to Detector
Reduction vale
(ZBD6)
(0.2 to 0.5L/min)
*1 Standard gas for
span point (ZBM)
*1 Sandard gas or instrumentation air can be used in place of span gas.
Note
*2 Instrument quality air or bottled air is available as reference air instead of ambient air,
when oxygen concentration in air around sensor changes.
*3 Blow down air inlet pressure in 200 to 300kpa.
4
OUTLINE DIAGRAM (Unit:mm)
Detector (ZFK2)
80
67
Proof cover (to order)
Approx. 65
Approx. 130
Approx. 118
45.4
36
5
26
Filter
EXTERNAL CONNECTION DIAGRAM
Element output
Heater
Thermocouple
A
1
3
2
4
5
6
Black / White/ Red / White/ Yellow / Blue
2-core wire
4-core wire
Exclusive-special cable
6-
6
Ref. air inlet (to order)
NPT1/8 or Rc1/8
Calibration gas inlet
for 1/4 inch tube.
Exclusive
cable gland
Flow guide tube
155
130
Approx. L
12
Approx. 20
67
Gas inlet
Z F K 2 R
4 - 5 A 3
5
7
1
6-M5
Gas outlet
detector side
3
1
Cord 11th
5
7
Z
8–15
MTG. holes
1.0
0.5 0.75
0.3
L (m)
Oxygen detector
L=
(to order)
MASS
Approx (kg)
4.1
2.7 3.3
4.8
5
ZFK, ZRY
Detector (ZFK5)
80
67
Proof cover (to order)
Approx. 65
Approx. 130
Approx. 118
45.4
36
5
26
Filter
EXTERNAL CONNECTION DIAGRAM
Element output
Heater
Thermocouple
1
3
2
4
5
6
Black / White/ Red / White/ Yellow / Blue
2-core wire
4-core wire
Exclusive-special cable
Ref. air inlet (to order)
NPT1/8 or Rc1/8
6-
6
Exclusive cable gland
Calibration gas inlet
for 1/4 inch tube.
Flow guide tube
155
130
Approx. 40
12
3
Approx. L
67
Gas inlet
Z F K 5 R
4 - 5 B 3
5
7
1
6-M5
detctor side
Gas outlet
3
1
Cord 11th
L (m)
5
7
Z
8- 15
MTG. holes
1.0
0.3 0.5
0.75
Oxygen detector
L=
MASS
Approx (kg)
(to order)
3.3
4.5 6.1
7.6
6
Flow guide tube (for high particulate)
4 - Rc 1/4
with plug int thd.
Blow down air inlet
Select an upper port to
avoid the condensation
in the piping
8 - 19
MTG. hole
Flange: JIS 5K80A FF
ZFK MTG. position
Packing
Z F K
R
4 - 6 D8
Gas outlet
30
Gas flow
Tube (50A SCH40)
Flow guide tube (for high particulate with cover)
4 - Rc 1/4
with plug int thd.
Blow down air inlet
Select an upper port to
avoid the condensation
in the piping
8 - 19
MTG. holes
Flange: JIS 5K80A FF
ZFK MTG. position
Packing
Z F K
R
4 - 6 E8
Gas outlet
30
Gas flow
(38)
Tube (50A SCH40)
Protection tube (65A SCH40)
7
ZFK, ZRY
Converter (ZRY)
Panel flush mounted
2-
72
φ
9
3.2
95
14
230
15
φ
CABLE GRAND
5-G1/2
40
40
40
40
CONNECTION DIAGRAM
1
2
3
4
5
6
7
8
9 10 11 12 13 14
+
–
+
–
+
–
L
N
THERMO- ELEMENT OUTPUT
FAULT
HEATER
POWER
SUPPLY
CONTACT POWER
COUPLE
INPUT
SIGNAL
OUTPUT
FOR
SENSOR
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
Head Office
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fesys.co.jp/eng
Instrumentation Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http://www.fic-net.jp/eng
Printed in Japan
Information in this catalog is subject to change without notice.
相关型号:
©2020 ICPDF网 联系我们和版权申明