EDS3-135E [FUJI]

cross stacK laser Gas analYZer; 交叉叠激光气体分析仪
EDS3-135E
型号: EDS3-135E
厂家: FUJI ELECTRIC    FUJI ELECTRIC
描述:

cross stacK laser Gas analYZer
交叉叠激光气体分析仪

文件: 总12页 (文件大小:676K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
cross stacK laser Gas analYZer  
DATA SHEET  
ZSS  
Cross stack laser gas analyzer (ZSS) provides continuous  
measurement of density of HCl contained in flue gas in an  
incineration system, NH  
3
in denitration equipment, com-  
with quick response speed.  
bustion gas, CO, CO and O  
2
2
Due to in-situ method, Measured gas does not require the  
preparation by sampling system. This analyzer can be  
used under high dust conditions, so it is possible to install  
upstream of the bagfilter where gas sampling is normally  
difficult.  
Control unit  
Receiver unit  
Features  
1. Maintenance is minimized thanks to no need of gas sam-  
pling devices.  
Transmitter unit  
2. Running cost is negligible due to having few expendable  
parts.  
3.Zero drift is within ±2.0%FS, and an excellent stability is  
ensured for a long period of time.  
Measurable component and range:  
4.Almost no interference by other gas component be-  
cause an infrared semiconductor laser, which has exact  
absorption wavelength as the measured component, is  
used.  
5.High speed response is available since the gas sampling  
device is unnecessary.  
6.High vibration resistance because the optical adjust-  
ment mechanism is separated from the airtight compart-  
ment, and it is easy to make a fine adjustment and large  
adjustment angle.  
Min. measuring Max. measuring  
Component  
range  
range  
HCl  
10 ppm  
5000 ppm  
HCl+H2O (*1)  
NH3  
50 ppm (HCl)  
15 ppm  
1000 ppm (HCl)  
5000 ppm  
NH3+H2O (*1)  
O2 (Class 1 Laser)  
50 ppm (NH3)  
4 vol%  
1000 ppm (NH3)  
100 vol%  
O2 (For use in high  
dust)  
4 vol%  
50 vol%  
50 vol%  
50 vol%  
50 vol%  
50 vol%  
50 vol%  
50 vol%  
50 vol%  
CO  
2.0 vol%  
10 vol%  
2.0 vol%  
10 vol%  
2.5 vol%  
10 vol%  
100 ppm  
CO (For use in high  
temperature)  
specIFIcatIons  
CO2  
Measurement principle:  
CO2 (For use in high  
temperature)  
Non-dispersive infrared absorbance sys-  
tem (NDIR)  
Measuring method:  
Cross-stack system (path system)  
Use application: Incineration facilities, denitration  
equipment, etc.  
CO+CO2  
CO+CO2 (For use in  
high temperature)  
CH4  
(*1) Range for H2O is fixed at 50vol%·m.  
(Range in the case stack (optical path length) diameter is 1m)  
Light source:  
Laser class:  
Near-infrared laser  
Class 1 (excluding O2 meter for use in  
high dust)  
Max. and min. ranges for each stack diameter  
Outline (D × W × H) mm:  
Receiver unit (180 × 400 × 200 mm)  
= (Max. and min. ranges when stack diameter is 1m)  
÷ Stack diameter  
Transmitter unit (240 × 400 (432) ×  
200 mm)  
Control unit (135 × 240 × 320 mm)  
Receiver/Transmitter unit: Approx.  
10kg  
Control unit: Approx. 8kg  
Outdoor use type, dustproofing/rainproof-  
ing structure (IP65)  
* Stack diameter = Optical path length  
(
Example: In case stack diameter is 2m for HCl measurement,  
Max.range: 2500ppm and min. range: 5ppm).  
Mass:  
Material:  
Receiver unit / Transmitter unit:  
Aluminum, SUS316  
Structure:  
Control unit: Aluminum  
Materials of gas-contacting parts:  
SUS316, BK7, FKM, PTFE, glass-cloth,  
silicone  
Box finish color: Receiver/Transmitter box: gray  
Control unit cover: blue  
Control unit case: silver  
EDS3-135e  
Date  
Apr. 1, 2011  
ZSS  
Air purge connection diameter:  
contact output contents (1ꢀ contact)  
RC1/4 (tube φ8 × 6)  
Low light transmission:  
Power supply: Rated voltage 100 to 240V AC ±10%  
Rated frequency 50/60 Hz  
Power consumption:  
Max. rated power: Approx. 75VA or  
less  
Calibration interval:  
Contact output is performed (close) when  
the amount of light transmission is insuf-  
ficient.  
Outside the range of upper/lower limits:  
According to the preset upper or lower  
limit alarm value, contact output is per-  
formed (close) when it becomes lower  
than alarm upper/lower limit.  
Once every six months (Maintenance  
cycle may vary depending on the operat-  
ing environment.)  
Device failure: Contact output is performed (close) when  
laser failure, laser temperature control  
failure, exceeding the amount of light  
transmission or communication error are  
occurred.  
During hold/during calibration:  
While AO output is held by the hold set-  
ting, the values, output (close) of which is  
held, is the ones just before the holding  
or arbitrary set values.  
Indicator (control unit): LCD with back light  
Communication functions:  
USB (for loader), RS-485 (Modbus® pro-  
tocol)  
Cable length:  
Receiver unit to Transmitter unit:  
Standard 2m (Maximum 25m)  
Receiver unit to control unit:  
Standard 5m (Maximum 100m)  
Analog output: 4 to 20mADC or 0 to 1VDC x 2 (4)  
Isolated output indicated in the paren-  
thesis is optional.  
Output (close) during calibration  
Power supply off:  
(1 to 5V DC, 0 to 5V or 0 to 10V is avail-  
able.)  
Output (close) while power supply is  
off  
Allowable load: 4 to 20mA DC 550or  
less, 0 to 1V DC 100kor more  
(Output measurement value and O2 cor-  
responding value. Average value and  
instantaneous value are switchable by  
the settings.)  
contact Input contents (optIon)  
Average value reset signal:  
Output of converted average value is  
started from the initial state by applying  
rectangular-wave voltage (pulse width  
2 sec or more) to the input terminal of  
average value resetting. Output is reset  
by inputting and restarted by opening.  
Switching instantaneous value/moving average value:  
Switching to and from the instantaneous  
value and the average value of the analog  
output is performed by applying rectan-  
gular-wave voltage (pulse width 2 sec or  
more) to the input terminal for switching  
between the instantaneous value and the  
moving average values.  
Analog input:  
4 to 20mA DC × 6  
Measured gas pressure, measured gas  
temperature, measured gas velocity, O2  
gas concentration, water concentration,  
air purge pressure  
(Measurement concentration correction,  
O2 correction or alarm output is performed  
according to the input signal.)  
Contact output: Relay contact output (contact capacity  
24V DC 1A 1a or 1b × 5 (Standard: 1a,  
1b: for power supply off only)  
Low light transmission, outside the range  
of upper/lower limits, device failure, dur-  
ing hold/during calibration, power supply  
off.  
Remote hold:  
The analog output is held by applying  
rectangular-wave voltage (pulse width  
2 sec or more) to the remote hold input  
terminal, and restarted by applying it  
again.  
Contact input (option):  
Photo coupler receiver contact input  
(operating voltage 12 to 24V DC / 5 to  
20mA) × 3  
Average value reset signal, switching in-  
stantaneous value/moving average value  
and remote hold  
FunctIons  
O2 correction:  
Conversion of measured HCl gas con-  
centrations into values at standard O2  
concentration  
Alarm output (screen-displayed):  
LD failure, LD temperature error, high  
gas temperature, air purge (low pres-  
sure), box temperature warning, low light  
transmission, PD over range, connection  
error, AI under, Range (H-Limit) or Range  
(L-Limit)  
Correction formula:  
21 – On  
21 – Os  
C =  
× Cs  
C: Converted concentration  
Cs: Measured concentration of sample  
gas  
Os: Measured O2 concentration (Upper  
limit settable 1 to 20% O2)  
On: Standard O2 concentration (value  
changeable by setting; 0 to 19%  
O2)  
Display contents:  
Component, concentration (instanta-  
neous value, average value O2 correction  
instantaneous value and O2 correction  
average value), alarm (fault status)  
The result of calculation is indicated and  
output in an analog output signal.  
2
Warm up time: 90 minutes or less  
(* Excluding influenced range with absorptive gases such as HCl  
InstallatIon envIronment  
Ambient temperature:  
and NH )  
3
–20 to 55°C (Receiver unit/Transmitter  
unit), –5 to 45°C (control unit)  
Ambient humidity: 90% R.H. or less  
Measurable optical path length (stack diameter):  
0.5 to 10m  
ec dIrectIve complIance  
Low voltage safety:EN61010-1(2001)  
Electromagnetic: "Installation Category II"  
Standard flange: JIS10K 50A flange (JIS B 2212)  
compatibility  
EN61326-1(2006)  
Air purge:  
Instrument air (compressor must be  
installed when power supply cannot  
be provided.) N2 for an O2 meter less  
than 20 vol% Pressure 0.5 to 0.7MPa  
or more  
EN61326-2-3(2006)  
EN61000-3-2(2006)  
EN6100-3-3(1995), A1(2001), A2(2005)  
Air purge flow rate:  
20L/min or more (depending on measured  
gas temperature, velocity, pressure, mois-  
ture and dust)  
One-side air purge flow rate: (L/min) = Gas flow rate (m/s)×10  
Measured gas condition:  
BasIc prIncIple  
An infrared semiconductor laser is adopted for the light  
source and photo-diode for photo-detection. Each mea-  
sured gas component has its own wavelength range where  
the gas absorbs the light (see the figure below).  
Temperature: 1200°C or less (O2, CO, CO2,  
CO+CO2)(Note)  
The wavelength range consists of a set of many absorption  
lines, and one of these lines is used for measurement.  
130°C or more, and 450°C  
As measurement is performed in this very narrow wave-  
length range, no other gas does not have an interference  
in principle.  
or less (NH3+H2O)  
130°C or more, and 450°C  
or less (HCl+H2O)  
300°C or more (CH4)  
Pressure: ±10kPa  
Moisture:50vol% or less (Should not be  
saturated water vapor.)  
To detect the gas concentration, used is not a variation  
of light intensity, but intensity of modulated signal ampli-  
tude.  
Velocity: 25m/s or less (Consult us for  
use in high dust or moisture  
environment.)  
N2O  
N2O  
NO  
Note) When measuring high-temperature gas at 500°C or  
higher, install a pressure sensor on the air-purge unit to  
monitor air purge. If measurement is taken while purge  
is not performed, the device may be damaged.  
H2S  
H2S  
O2  
CH4  
HCl  
HF  
HF  
Dust:  
5 to 30g/Nm3 (depending on environmental  
conditions such as measuring component,  
equipment specification, light path length  
and particle diameter).  
If dust exceeds 30g/Nm3-m, consult with  
Fuji.  
NH3  
NH3  
CO2  
NH3  
CO  
750nm 1000nm 1250nm 1500nm 1750nm 2000nm 2250nm  
Wavelength  
Vibration:  
0.5G or less (0.2G or less when the fre-  
quency range is 20 to 40Hz) (when optical  
path length is 1m)  
Absorbing spectrum map of  
measuring gases  
Absorption line used for  
measurement  
Mounting angle: Horizontally ±5 degrees or less (No dew  
condensation should accumulate on the  
window.)  
1720nm 1740nm 1760nm 1780nm 1800nm 1820nm 1840nm  
perFormance (except For H2o)  
Wavelength  
Absorption lines of HCl  
Repeatability:  
±1.0%FS (depending on measuring  
component and measuring range)  
±1.0%FS (depending on measuring  
component and measuring range)  
±2.0% FS  
Linearity:  
Zero drift:  
(NH3 ±3.0%, FS when range is 20ppm  
or less)  
Interference from other gas components:  
±2.0%FS (The gasses which have no  
absorption within the wavelength range  
of measured gas (several tens pm) do  
not interfere in principle).  
Minimum detectable limit:  
1% of minimum detectable limit  
Response time (90% FS response): 1 to 5 seconds  
3
ZSS  
code sYmBols  
When ordering, be sure to submit the order sheet on the last page of this Data Sheet.  
Note 1  
4
5
6
7
8
9
10 11 12 13 14 15 16 17 18 19 20 21  
ZSS  
Note  
Note 1  
Y 4  
0
Digit  
4
Specification  
CO  
Measurable components  
A
B
C
F
G
H
K
CO (For use in high temperature) Note 1, 5  
HCl  
HCl+H2O  
CO2  
Note 1  
Note 1, 9  
CO2 (For use in high temperature) Note 5  
CO+CO2  
CO+CO2 (For use in high temperature) Note 5  
O2 (Class 1 Laser)  
L
P
O2 (For use in high dust)  
CH4  
NH3  
Q
R
W
X
Note 1  
Note 1, 9  
NH3+H2O  
ppm  
1
3
5
5
6
Unit  
mg/m3  
vol%  
Note 2  
K
Q
S
L
Measuring range  
0 to 2  
0 to 2.5  
0 to 4  
0 to 5  
V
0
1
0 to 10  
0 to 15  
0 to 20  
0 to 25  
0 to 50  
0 to 100  
0 to 200  
0 to 250  
0 to 400  
0 to 500  
0 to 1000  
0 to 2000  
0 to 5000  
0 to 6000  
Others  
T
A
B
C
D
J
E
F
G
H
M
X
Y
4
7
8
9
Modification No.  
Flange rating  
A
B
C
D
10K 50A (JIS B 2212)  
10K 100A  
DN50/PN10  
ANSI #150 2B  
0
1
A
B
1
2
3
4
5
10 Number of analog output points 2 points  
4 points  
11 Number of analog input points  
2 points  
6 points  
4 to 20mA DC  
12 Analog output  
0 to 20mA DC  
0 to 1V DC  
0 to 5V DC  
1 to 5V DC  
13 Contact output/input  
5 output points, No input  
0
5output points, 3 input points  
5m  
10m  
20m  
30m  
40m  
50m  
1
A
B
C
D
E
F
14 Cable length between receiver  
and control unit  
Note 3  
Note 6  
Note 1) When O2 conversion is required,  
specify the reference O2  
G
H
X
A
B
C
D
E
F
80m  
100m  
Others  
2m  
5m  
10m  
15m  
20m  
25m  
Others  
Japanese  
English  
Chinese  
0m  
1m  
2m  
3m  
4m  
5m  
6m  
7m  
8m  
9m  
0.0m  
0.1m  
0.2m  
0.3m  
0.4m  
0.5m  
0.6m  
0.7m  
0.8m  
concentration (settable within  
0 to 19vol%, O2: Integer).  
(HCl meter, NH3 meter, CO meter)  
Note 2) Specify the range within the  
max/min range calculated from  
path length.  
Note 4  
Note 7  
15 Cable length between receiver  
and transmitter  
If the measuring range x stack  
length (optical path length)  
exceeds 1000ppm/m,  
X
16 Display and operation manual  
J
E
C
0
1
2
3
4
5
6
7
8
9
0
1
2
3
4
5
6
7
8
9
0
5
9
N
D
consult Fuji.  
Note 3) Cable length between receiver  
and control unit: Max. 100m  
Note 4) Cable length between receiver  
and transmitter: Max. 25m  
Note 5) Specified to use in high  
temperature gas: 500°C or more,  
and 1200°C or less  
Note 6) Others (specified with X) are  
possible only for 10 m or more.  
Note 7) Others (specified with X) are  
possible only for 5 m or more.  
Note 8) Specify ‘D’ when dust exceeds  
5g/Nm3-m.  
Note 9) If H2O is contained in measured  
component, be sure to write the  
measured gas temperature into  
“Specification sheet” at the last  
page of this catalog, or contact  
directly to Fuji’s service  
18 Measuring optical path length  
(unit: 1m)  
19 Measuring optical path length  
(unit: 0.1m)  
department.  
0.9m  
0.00m  
0.05m  
How to specify measuring optical path  
length  
Example: 2.25m 2 for 18th digit, 2 for  
19th digit and 5 for 20th digit  
10m 9 for 18th digit, 9 for  
19th digit and 9 for 20th digit  
20 Measuring optical path length  
(unit: 0.01m)  
(Used only when 10m is specified)  
Standard  
Dust proof  
21 Dust proof specification  
Note 8  
4
lIst For comBInatIons oF measuraBle components, unIts and measurement ranGes  
Measurable range 6th digit  
(Unit: When 5th digit is 1 or 3)  
Not feasible  
Measurable range 6th digit  
(Unit: When 5th digit is 5)  
K, Q, S, L, V, 0, 1, T, A, X  
V, 0, 1, T, A, X  
Not feasible  
Not feasible  
Measurable component  
4th digit  
CO  
A
B
C
F
G
H
K
CO (For use in high temperature)  
HC  
HC+H2O  
CO2  
CO2 (For use in high temperature)  
CO+CO2  
Not feasible  
V, 0, 1, T, A, B, C, D, J, E, F, G, H, Z  
A, B, C, D, J, E, F  
Not feasible  
Not feasible  
Not feasible  
Not feasible  
Not feasible  
Not feasible  
Not feasible  
K, Q, S, L, V, 0, 1, T, A, X  
V, 0, 1, T, A, X  
Q, S, L, V, 0, 1, T, A, X (CO)  
Q, S, L, V, 0, 1, T, A, X (CO2)  
V, 0, 1, T, A, X (CO)  
V, 0, 1, T, A, X (CO2)  
S, L, V, 0, 1, T, A, B, X  
S, L, V, 0, 1, T, A, X  
K, Q, S, L, V, 0, 1, T, A, X  
Not feasible  
CO+CO2  
L
(For use in high temperature)  
O2(Class 1 Laser)  
O2(For use in high dust)  
CH4  
NH3  
NH3+H2O  
P
Q
R
W
X
Not feasible  
B, C, D, J, E, F, G, H, M, Z  
0, 1, T, A, B, C, D, J, E, F, G, H, Z  
A, B, C, D, J, E, F  
Not feasible  
scope oF delIverY  
• Receiver unit  
• Transmitter unit  
• Control unit  
Cable between receiver unit and transmitter unit (specified  
length)  
Cable between receiver unit and control unit (specified length)  
• Standard accessory set, instruction manual  
optIonal Items  
Spare parts for one year (ZBN1SS12)  
Calibration gas cell (*1) (*2)  
*1: One set of the cables and calibration gas cell are neces-  
sary for installation and annual maintenance.  
*2: Standard length 1m (200mm or 500mm when the mea-  
suring range is low concentration)  
Cable between receiver unit and transmitter unit (for  
calibration) (*1)  
Cable between receiver unit and control unit (for calibra-  
tion) (*1)  
Standard gas (ZBM), pressure regulator (ZBD)  
Recorder (when necessary, Fuji's product type PHL/PHR, etc.)  
Others  
standard accessorIes  
Name  
Quantity  
8 (16)  
Specification  
M16×55 (70) SUS (*)  
M16 SUS (*)  
Bolt  
8 (16)  
Nut  
8 (16)  
M16 SUS (*)  
Spring washer  
Flat washer  
8 (16)  
M16 SUS (*)  
Companion flange packing or  
flange packing specified for  
use in high temperature  
2
According to flange specifications  
Hex socket bolt M8 × 70  
6
2
1
Bolt for angle adjustment  
Power supply fuse  
Instruction manual  
(*When "B" or "C" is specified in the 9th digit in a code symbol, quantity is 16 pieces. 8 pieces are attached in other cases.)  
(*When "B", "C" or "D" is specified in the 9th digit, Bolt length is 70mm. It is 55mm in other cases. Inch-sized bolts are not  
applicable.)  
spare parts For one Year (ZBn1ss12)  
Name  
Quantity  
2 pieces For bellows (ZZP*ZSSTK7N3508P1)  
2 pieces (ZZP*ZSSTK7P2530P3)  
Remarks  
Silicon packing A  
O-ring  
5
ZSS  
calIBratIon/InstallatIon FIxtures lIst (optIon)  
Parts name  
Quantity Type  
(3) Calibration gas cell / about 1000mm  
Cable between receiver unit and  
control unit (for calibration)  
Cable between receiver unit and  
transmitter unit (for calibration)  
Calibration gas cell (*3)  
1
1
1
ZZP ZSSTK4J1271C2  
(1)  
(2)  
*
ZZP ZSSTK4J0641C3  
*
ZZP ZSSTK4J3676C1  
*
(HCl, NH3, CO,CO2, CO+CO2,CH4)  
Calibration gas cell (*3)  
(3)  
(4)  
ZZP ZSSTK4J5026C1  
*
(HCl+H2O, NH3+H2O, O2)  
Optical axis adjusting tool (laser  
pointer, target)  
(2) Cable between receiver and  
transmitter (approx. 1600mm)  
1
ZZP ZSSTK4J1274C1  
*
(5) IR card (for NH3)  
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
ZZP ZSSTK7N4505P1  
*
(6) Check cell  
ZZP ZSSTK4J2605C1  
*
(7) Filter regulator  
ZZP ZSSTK7F9554P2  
*
(8) Mist separator  
ZZP ZSSTK7H8049P1  
*
(9) R1/4 stopper (plug) for mist separator  
Flow meter (20 to 100L/min) (*4)  
(10) Flow meter (4 to 50L/min) (*4)  
Flow meter (30 to 300L/min) (*4)  
Air-set box (20 to 100L/min)  
ZZP ZSSTK738114P7  
*
ZZP ZSSTK7N4624P1  
*
Parts used for calibration  
(1) Cable between receiver unit and control unit  
ZZP ZSSTK7N4624P2  
*
ZZP ZSSTK7N8849P1  
*
ZZP ZSSTK7N6685C1  
*
(11) Air-set box (4 to 50L/min)  
Air-set box (30 to 300L/min)  
ZZP ZSSTK7N6685C2  
*
ZZP ZSSTK7N6685C3  
*
Stack  
Air purge mechanism (20 to 100L/min)  
(12) Air purge mechanism (4 to 50L/min)  
Air purge mechanism (30 to 300L/min)  
(13) BNC cable for optical axis adjustment  
ZZP ZSSTK7P1433C1  
*
ZZP ZSSTK7P1433C2  
*
ZZP ZSSTK7P1433C3  
*
(4) Optical axis adjusting tool  
(laser pointer)  
ZZP ZSSTK7P2524C1  
*
(4) Optical axis adjusting tool (target)  
* 3: Standard length 1m (200mm or 500mm when the  
measuring range is low concentration)  
* 4: Usually, 2 units are required to adjust each air purge  
for receiver unit and transmitter unit.  
Parts for adjusting the optical axis  
Transmitter unit  
(5) IR card  
IR CARD  
Laser beam (visible)  
Invisible light  
Invisible light / visible fixtures(for NH3)  
(6) Check cell  
Transmitter unit  
Receiver unit  
Stack  
Control unit  
Laser gas analyzer with check cell (ZSS)  
SV3  
SV2  
Atmospheric  
release  
SV1  
(AO)  
Sequencer  
Remote check input (AI)  
Supplies for checking span value on site  
N2 cylinder Standard gas  
cylinder  
6
(11) Air set box  
4-φ11  
(Mounting hole)  
Parts for air purge  
(7) Filter regulator  
Box  
(8) Mist separator  
Flow  
Mist separator  
Filter regulator  
meter  
Pipe with  
external  
diameter φ10  
(out of supply  
scope)  
(IN)  
(OUT)  
(OUT)  
(IN)  
(OUT)  
(IN)  
(OUT)  
(IN)  
measuring  
instrument  
air inlet  
Receiver  
purge outlet  
Drain outlet  
(φ10 piping)  
(not supplied by us)  
Transmitter  
purge outlet  
Piping system diagram  
(10) Flow meter  
420  
500  
150  
110  
65  
43  
Measuring instrument  
air inlet  
Rc1/4  
Receiver  
purge outlet  
Rc1/4  
Transmitter  
purge outlet  
Rc1/4  
Drain outlet  
Piping retainer for φ10  
(12) Air purge mechanical part  
1m  
total 20m  
Piping material: φ8/6 Teflon pipe  
Mist separator  
to Transmitter  
purge inlet  
Filter regulator  
for φ8 tube  
(OUT)  
(OUT)  
(IN)  
(metallic tube  
recommendable)  
(IN)  
(OUT)  
to Receiver  
purge inlet  
(IN)  
Measuring  
instrument air inlet  
0.5m  
0.5m  
to Receiver  
purge inlet  
(Coupling joint)  
Piping system diagram  
(13) BNC cable for optical axis adjustment  
7
ZSS  
outlIne dIaGram (uꢁiꢂ : ꢃꢃ)  
Or air-set box  
Copper or SUS pipe (not supplied)  
Instrumentation airline joint (not supplied)  
2-Flowmeter  
(option)  
Filter regulator (option)  
20 to 100L/min  
φ8/6 Teflon pipe  
Instrumentation  
air inlet  
(not supplied)  
Mist separator  
(option)  
150 to 200  
Approx. 400  
Flange of angle  
adjustment  
150 to 200  
Note 1  
Approx. 400 (432)  
With bolt and nut  
(M16×55)  
Transmitter box  
(160×240×100)  
Flange of angle  
adjustment  
Purging gas inlet  
Flange pipe to  
Receiver box  
(with bell-and-spigot  
(120×180×100)  
mount stack  
joint for  
φ
8/6)  
(not suppried)  
Purging gas inlet  
100  
16  
100  
Stack  
0.5 to 10m  
JIS10K50A FF  
(with packing)  
φ34  
(Connector)  
Receiver unit  
130  
Transmitter unit  
φ34  
(Connector)  
74  
2-φ12  
Cable between receiver unit  
and transmitter unit (standard : 2m)  
2-φ12  
74  
Note 1: Be sure to set the R1/4 sealing  
plug (ZZP*ZSSTK738114P7)  
on the drain outlet of the mist  
separator.  
240  
φ10.5  
Control unit  
LASER GAS ANALYZER  
ZSS  
134  
3
Power  
supply  
AI  
DI/O  
AO  
7-Waterproof grand  
74  
(Internal diameter φ12)  
12  
Cable length  
(200mm)  
Mounting diagram  
Connector  
Be sure to prepare instrument air (N2 for an O2 meter).  
Note) Air purge should be performed if the equipment is  
installed in a flue whether or not it is operated.  
Cable between receiver unit and control unit  
(standard : 5m)  
8
external connectIon dIaGram (WHen numBer oF analoG output Is 2 poInts)  
There are two types of output/input terminal.  
Power supply terminal  
1
2
3
Screw diameter : M4  
100 to 240 V AC  
(50/60Hz)  
FG  
AO terminal  
21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40  
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17 18 19 20  
Screw diameter : M2 or the equivalent  
1
AO1+  
AO1-  
21 AO2+  
22 AO2-  
Analog output 1  
Analog output 2  
2
Al terminal  
21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40  
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17 18 19 20  
Screw diameter : M2 or the equivalent  
3
AI1+  
AI1-  
23 AI4+  
24 AI4-  
25 AI5+  
26 AI5-  
27 AI6+  
28 AI6-  
Analog input 1  
Analog input 2  
Analog input 4 (AI extension board is required)  
Analog input 5 (AI extension board is required)  
Analog input 6 (AI extension board is required)  
4
5
6
7
8
AI2+  
AI2-  
AI3+  
AI3-  
Analog input 3 (AI extension board is required)  
DI/DO terminal  
21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40  
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17 18 19 20  
Screw diameter : M2 or the equivalent  
9
DO1  
29 DO7  
30 DO7  
31 DO8  
32 DO8  
33 DI1  
34 DI1  
35 DI2  
36 DI2  
37 DI3  
38 DI3  
39 DI4  
40 DI4  
Low light transmission  
Device failure  
10 DO1  
11 DO2  
12 DO2  
13 DO3  
14 DO3  
15 DO4  
16 DO4  
17 DO5  
18 DO5  
19 DO6  
20 DO6  
Power supply off  
During hold / during calibration  
Average value reset signal (option)  
Switching instantaneous value/moving average value (option)  
Remote hold (option)  
Outside the range of upper /  
lower limits (component 1)  
Range identification (component 1)  
Note) Unassigned terminal may be connected to internal circuit. So they should not be used as repeating terminals.  
9
ZSS  
external connectIon dIaGram (WHen numBer oF analoG output Is 4 poInts)  
Power supply terminals  
1
2
3
AC100 to 240V  
FG  
50/60Hz  
Screws diameter: M4  
AO terminal  
AO4 Unassigned Unassigned  
1
2
3
4
AO1+  
AO1-  
AO2+  
AO2-  
5
6
7
8
AO3+  
AO3-  
AO4+  
AO4-  
+
-
Analog output 1  
Analog output 2  
Analog output 3  
Analog output 4  
7
8
9
3
10 11 12  
1
2
4
5
6
+
-
+
-
+
-
AO1  
AO2  
AO3  
AI terminal  
AI4  
AI5  
AI6  
+
1
2
3
4
5
6
AI1+  
AI1-  
7
8
9
AI4+  
AI4-  
Analog input 4  
(AI extension board is required)  
Analog input 1  
Analog input 2  
+
-
+
-
-
7
8
9
10 11 12  
AI2+  
AI2-  
AI5+  
Analog input 5  
(AI extension board is required)  
10 AI5-  
11 AI6+  
12 AI6-  
1
2
3
4
5
6
Analog input 3  
(AI extension board  
is required)  
AI3+  
AI3-  
Analog input 6  
(AI extension board is required)  
+
-
+
-
+
-
AI1  
AI2  
AI3  
DI/DO terminal  
25 26 27 28 29 30 31 32 33 34 35 36  
13 14 15 16 17 18 19 20 21 22 23 24  
1
2
3
4
5
6
7
8
9 10 11 12  
1
2
3
4
5
6
7
8
9
DI1  
DI1  
DI2  
DI2  
DI3  
DI3  
DI4  
DI4  
DI5  
19 DO4  
Average value reset signal (option)  
Outside the range of upper/lowerlimits (component 1)  
Range identification (component 1)  
20 DO4  
21 DO5  
22 DO5  
23 DO6  
24 DO6  
25 DO7  
26 DO7  
27 DO8  
28 DO8  
29 DO9  
30 DO9  
31 DO10  
32 DO10  
33 DO11  
34 DO11  
Switching instantaneous value/moving  
average value (option)  
Remote hold (option)  
Power supply off  
10 DI5  
13 DO1  
14 DO1  
15 DO2  
16 DO2  
17 DO3  
18 DO3  
Low light transmission  
Device failure  
During hold/during callibration  
Note) Unassigned terminals may be connected to internal circuit. So they should not be used as repeating terminals.  
10  
mountInG dImensIons (uꢁiꢂ: ꢃꢃ)  
Do not install the device in a stack if air purge is not performed.  
It may become unusable because the window will get dirty.  
Stack  
c
d
a
a: Inner diameter of φ50mm or more  
b: 20 to 70mm (*5)  
b
c: Approximately 200 to 250mm (*6)  
d: 400mm or more (*7)  
Flange pipe to mount stack (*9)  
e
e: 500 to 1500mm (*7)  
Scaffold (*8)  
*5: When used in an enviroment with high dust, reserve at least 50mm.  
*6: When internal gas temperature is high, reserve the distance of 250mm or more. Note that, adjustable angle  
range narrows down, so mounting angle of flange pipe to mount stack should be adjusted accurately.  
*7: Sufficiently secure a space for installation which permits easy operation.  
*8: Scaffold is required for the both side of receiver unit and transmitter unit.  
*9: Use a flange provided with a valve where there is a risk of gas injection or the operator may be exposed to  
dangerous situation.  
*10: Tighten the flange with a torque of 118±114N-m. When tightening the flange, apply grease to the bolts  
(recommendable to use grease containing molybdenum).  
rouGH GuIde to mount tHe companIon FlanGe  
When is larger than 5°, mount it within the angle (5°) inside the circle with the A radius.  
θ
γ
Y
X
A
θ
Z
X: Inner diameter of flange of receiver unit and transmitter unit  
Y: Length of flange of receiver uit and transmitter unit  
Z: Length of stack  
X
γ, θ = ATAN  
2 (Y + 125)  
A 0.087 × (Z + Y + 125)  
11  
ZSS  
Please specify the following items when you contact us for inquiry.  
1.Code symbol ZSS  
2. Parameter of measured gas  
Item  
Measured gas concentration  
Temperature (°C)  
Pressure (kPa)  
Min. value Typical value Max. value  
Flow (m/s)  
Moisture (vol%)  
Dust (mg/Nm3)  
Other gas component (vol%/ppm)  
Other gas component (vol%/ppm)  
Other gas component (vol%/ppm)  
3.Stack diameter (Inner diameter) (measured optical path length)  
A
m
m
m
4.Cable between receiver unit and control unit.  
B
5.Cable between receiver unit and transmitter unit.  
C
A
Instrumentation air  
Instrumentation air  
Stack  
Receiver unit  
Transmitter unit  
C
Control unit  
B
6.Spares for 1-year measurement  
7. Optional items  
no need necessary  
no need necessary  
8. Output of O2 correction value (for only HCl and NH3 meters)  
no need necessary  
9.Oxygen correction reference value (%) (When "necessary" is selected in 8)  
%
10.Vibration  
no need necessary (  
G)  
Caution on Safety  
*Before using this product, be sure to read its instruction manual in advance.  
International Sales Div  
Sales Group  
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,  
Shinagawa-ku, Tokyo 141-0032, Japan  
http://www.fujielectric.com  
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425  
http://www.fjielectric.com/products/instruments/  
Information in this catalog is subject to change without notice.  
Printed in Japan  

相关型号:

EDS3-136

INFRARED GAS ANALYZER FOR STIC GAS
FUJI

EDS3-138B

INFRARED GAS ANALYZER FOR STACK GAS
FUJI

EDS3-139B

IN-SITU ZIRCONIA OXYGEN ANALYZER FOR HAZARDOUS LOCATION
FUJI

EDS3-141A

NDIR TYPE INFRARED GAS ANALYZER(3-COMPONENT ANALYZER)
FUJI

EDS3-143

INFRARED GAS ANALYZER
FUJI

EDS3-144

INFRARED GAS ANALYZER(HIGH PERFORMANCE MODEL)
FUJI

EDS3-145

INFRARED GAS ANALYZER(LOW-CONCENTRATION MEASUREMENT TYPE)
FUJI

EDS3-1I

SAMPLING DEVICE SERIES FOR GAS ANALYZER
FUJI

EDS3-212A

MULTI-CHANNEL ZIRCONIA OXYGEN ANALYZER
FUJI

EDS3-61H

THERMAL CONDUCTIVITY GAS ANALYZER THERMOMAT
FUJI

EDS3-62H

PARAMAGNETIC OXYGEN ANALYZER
FUJI

EDS3-69F

PARAMAGNETIC OXYGEN ANALYZER
FUJI