EDS3-135E [FUJI]
cross stacK laser Gas analYZer; 交叉叠激光气体分析仪型号: | EDS3-135E |
厂家: | FUJI ELECTRIC |
描述: | cross stacK laser Gas analYZer |
文件: | 总12页 (文件大小:676K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
cross stacK laser Gas analYZer
DATA SHEET
ZSS
Cross stack laser gas analyzer (ZSS) provides continuous
measurement of density of HCl contained in flue gas in an
incineration system, NH
3
in denitration equipment, com-
with quick response speed.
bustion gas, CO, CO and O
2
2
Due to in-situ method, Measured gas does not require the
preparation by sampling system. This analyzer can be
used under high dust conditions, so it is possible to install
upstream of the bagfilter where gas sampling is normally
difficult.
Control unit
Receiver unit
Features
1. Maintenance is minimized thanks to no need of gas sam-
pling devices.
Transmitter unit
2. Running cost is negligible due to having few expendable
parts.
3.Zero drift is within ±2.0%FS, and an excellent stability is
ensured for a long period of time.
Measurable component and range:
4.Almost no interference by other gas component be-
cause an infrared semiconductor laser, which has exact
absorption wavelength as the measured component, is
used.
5.High speed response is available since the gas sampling
device is unnecessary.
6.High vibration resistance because the optical adjust-
ment mechanism is separated from the airtight compart-
ment, and it is easy to make a fine adjustment and large
adjustment angle.
Min. measuring Max. measuring
Component
range
range
HCl
10 ppm
5000 ppm
HCl+H2O (*1)
NH3
50 ppm (HCl)
15 ppm
1000 ppm (HCl)
5000 ppm
NH3+H2O (*1)
O2 (Class 1 Laser)
50 ppm (NH3)
4 vol%
1000 ppm (NH3)
100 vol%
O2 (For use in high
dust)
4 vol%
50 vol%
50 vol%
50 vol%
50 vol%
50 vol%
50 vol%
50 vol%
50 vol%
CO
2.0 vol%
10 vol%
2.0 vol%
10 vol%
2.5 vol%
10 vol%
100 ppm
CO (For use in high
temperature)
specIFIcatIons
CO2
Measurement principle:
CO2 (For use in high
temperature)
Non-dispersive infrared absorbance sys-
tem (NDIR)
Measuring method:
Cross-stack system (path system)
Use application: Incineration facilities, denitration
equipment, etc.
CO+CO2
CO+CO2 (For use in
high temperature)
CH4
(*1) Range for H2O is fixed at 50vol%·m.
(Range in the case stack (optical path length) diameter is 1m)
Light source:
Laser class:
Near-infrared laser
Class 1 (excluding O2 meter for use in
high dust)
Max. and min. ranges for each stack diameter
Outline (D × W × H) mm:
Receiver unit (180 × 400 × 200 mm)
= (Max. and min. ranges when stack diameter is 1m)
÷ Stack diameter
Transmitter unit (240 × 400 (432) ×
200 mm)
Control unit (135 × 240 × 320 mm)
Receiver/Transmitter unit: Approx.
10kg
Control unit: Approx. 8kg
Outdoor use type, dustproofing/rainproof-
ing structure (IP65)
* Stack diameter = Optical path length
(
Example: In case stack diameter is 2m for HCl measurement,
Max.range: 2500ppm and min. range: 5ppm).
Mass:
Material:
Receiver unit / Transmitter unit:
Aluminum, SUS316
Structure:
Control unit: Aluminum
Materials of gas-contacting parts:
SUS316, BK7, FKM, PTFE, glass-cloth,
silicone
Box finish color: Receiver/Transmitter box: gray
Control unit cover: blue
Control unit case: silver
EDS3-135e
Date
Apr. 1, 2011
ZSS
Air purge connection diameter:
contact output contents (1ꢀ contact)
RC1/4 (tube φ8 × 6)
Low light transmission:
Power supply: Rated voltage 100 to 240V AC ±10%
Rated frequency 50/60 Hz
Power consumption:
Max. rated power: Approx. 75VA or
less
Calibration interval:
Contact output is performed (close) when
the amount of light transmission is insuf-
ficient.
Outside the range of upper/lower limits:
According to the preset upper or lower
limit alarm value, contact output is per-
formed (close) when it becomes lower
than alarm upper/lower limit.
Once every six months (Maintenance
cycle may vary depending on the operat-
ing environment.)
Device failure: Contact output is performed (close) when
laser failure, laser temperature control
failure, exceeding the amount of light
transmission or communication error are
occurred.
During hold/during calibration:
While AO output is held by the hold set-
ting, the values, output (close) of which is
held, is the ones just before the holding
or arbitrary set values.
Indicator (control unit): LCD with back light
Communication functions:
USB (for loader), RS-485 (Modbus® pro-
tocol)
Cable length:
Receiver unit to Transmitter unit:
Standard 2m (Maximum 25m)
Receiver unit to control unit:
Standard 5m (Maximum 100m)
Analog output: 4 to 20mADC or 0 to 1VDC x 2 (4)
Isolated output indicated in the paren-
thesis is optional.
Output (close) during calibration
Power supply off:
(1 to 5V DC, 0 to 5V or 0 to 10V is avail-
able.)
Output (close) while power supply is
off
Allowable load: 4 to 20mA DC 550Ω or
less, 0 to 1V DC 100kΩ or more
(Output measurement value and O2 cor-
responding value. Average value and
instantaneous value are switchable by
the settings.)
contact Input contents (optIon)
Average value reset signal:
Output of converted average value is
started from the initial state by applying
rectangular-wave voltage (pulse width
2 sec or more) to the input terminal of
average value resetting. Output is reset
by inputting and restarted by opening.
Switching instantaneous value/moving average value:
Switching to and from the instantaneous
value and the average value of the analog
output is performed by applying rectan-
gular-wave voltage (pulse width 2 sec or
more) to the input terminal for switching
between the instantaneous value and the
moving average values.
Analog input:
4 to 20mA DC × 6
Measured gas pressure, measured gas
temperature, measured gas velocity, O2
gas concentration, water concentration,
air purge pressure
(Measurement concentration correction,
O2 correction or alarm output is performed
according to the input signal.)
Contact output: Relay contact output (contact capacity
24V DC 1A 1a or 1b × 5 (Standard: 1a,
1b: for power supply off only)
Low light transmission, outside the range
of upper/lower limits, device failure, dur-
ing hold/during calibration, power supply
off.
Remote hold:
The analog output is held by applying
rectangular-wave voltage (pulse width
2 sec or more) to the remote hold input
terminal, and restarted by applying it
again.
Contact input (option):
Photo coupler receiver contact input
(operating voltage 12 to 24V DC / 5 to
20mA) × 3
Average value reset signal, switching in-
stantaneous value/moving average value
and remote hold
FunctIons
O2 correction:
Conversion of measured HCl gas con-
centrations into values at standard O2
concentration
Alarm output (screen-displayed):
LD failure, LD temperature error, high
gas temperature, air purge (low pres-
sure), box temperature warning, low light
transmission, PD over range, connection
error, AI under, Range (H-Limit) or Range
(L-Limit)
Correction formula:
21 – On
21 – Os
C =
× Cs
C: Converted concentration
Cs: Measured concentration of sample
gas
Os: Measured O2 concentration (Upper
limit settable 1 to 20% O2)
On: Standard O2 concentration (value
changeable by setting; 0 to 19%
O2)
Display contents:
Component, concentration (instanta-
neous value, average value O2 correction
instantaneous value and O2 correction
average value), alarm (fault status)
The result of calculation is indicated and
output in an analog output signal.
2
Warm up time: 90 minutes or less
(* Excluding influenced range with absorptive gases such as HCl
InstallatIon envIronment
Ambient temperature:
and NH )
3
–20 to 55°C (Receiver unit/Transmitter
unit), –5 to 45°C (control unit)
Ambient humidity: 90% R.H. or less
Measurable optical path length (stack diameter):
0.5 to 10m
ec dIrectIve complIance
Low voltage safety:EN61010-1(2001)
Electromagnetic: "Installation Category II"
Standard flange: JIS10K 50A flange (JIS B 2212)
compatibility
EN61326-1(2006)
Air purge:
Instrument air (compressor must be
installed when power supply cannot
be provided.) N2 for an O2 meter less
than 20 vol% Pressure 0.5 to 0.7MPa
or more
EN61326-2-3(2006)
EN61000-3-2(2006)
EN6100-3-3(1995), A1(2001), A2(2005)
Air purge flow rate:
20L/min or more (depending on measured
gas temperature, velocity, pressure, mois-
ture and dust)
One-side air purge flow rate: (L/min) = Gas flow rate (m/s)×10
Measured gas condition:
BasIc prIncIple
An infrared semiconductor laser is adopted for the light
source and photo-diode for photo-detection. Each mea-
sured gas component has its own wavelength range where
the gas absorbs the light (see the figure below).
Temperature: 1200°C or less (O2, CO, CO2,
CO+CO2)(Note)
The wavelength range consists of a set of many absorption
lines, and one of these lines is used for measurement.
130°C or more, and 450°C
As measurement is performed in this very narrow wave-
length range, no other gas does not have an interference
in principle.
or less (NH3+H2O)
130°C or more, and 450°C
or less (HCl+H2O)
300°C or more (CH4)
Pressure: ±10kPa
Moisture:50vol% or less (Should not be
saturated water vapor.)
To detect the gas concentration, used is not a variation
of light intensity, but intensity of modulated signal ampli-
tude.
Velocity: 25m/s or less (Consult us for
use in high dust or moisture
environment.)
N2O
N2O
NO
Note) When measuring high-temperature gas at 500°C or
higher, install a pressure sensor on the air-purge unit to
monitor air purge. If measurement is taken while purge
is not performed, the device may be damaged.
H2S
H2S
O2
CH4
HCl
HF
HF
Dust:
5 to 30g/Nm3 (depending on environmental
conditions such as measuring component,
equipment specification, light path length
and particle diameter).
If dust exceeds 30g/Nm3-m, consult with
Fuji.
NH3
NH3
CO2
NH3
CO
750nm 1000nm 1250nm 1500nm 1750nm 2000nm 2250nm
Wavelength
Vibration:
0.5G or less (0.2G or less when the fre-
quency range is 20 to 40Hz) (when optical
path length is 1m)
Absorbing spectrum map of
measuring gases
Absorption line used for
measurement
Mounting angle: Horizontally ±5 degrees or less (No dew
condensation should accumulate on the
window.)
1720nm 1740nm 1760nm 1780nm 1800nm 1820nm 1840nm
perFormance (except For H2o)
Wavelength
Absorption lines of HCl
Repeatability:
±1.0%FS (depending on measuring
component and measuring range)
±1.0%FS (depending on measuring
component and measuring range)
±2.0% FS
Linearity:
Zero drift:
(NH3 ±3.0%, FS when range is 20ppm
or less)
Interference from other gas components:
±2.0%FS (The gasses which have no
absorption within the wavelength range
of measured gas (several tens pm) do
not interfere in principle).
Minimum detectable limit:
1% of minimum detectable limit
Response time (90% FS response): 1 to 5 seconds
3
ZSS
code sYmBols
When ordering, be sure to submit the order sheet on the last page of this Data Sheet.
Note 1
4
5
6
7
8
9
10 11 12 13 14 15 16 17 18 19 20 21
ZSS
Note
Note 1
Y 4
0
Digit
4
Specification
CO
Measurable components
A
B
C
F
G
H
K
CO (For use in high temperature) Note 1, 5
HCl
HCl+H2O
CO2
Note 1
Note 1, 9
CO2 (For use in high temperature) Note 5
CO+CO2
CO+CO2 (For use in high temperature) Note 5
O2 (Class 1 Laser)
L
P
O2 (For use in high dust)
CH4
NH3
Q
R
W
X
Note 1
Note 1, 9
NH3+H2O
ppm
1
3
5
5
6
Unit
mg/m3
vol%
Note 2
K
Q
S
L
Measuring range
0 to 2
0 to 2.5
0 to 4
0 to 5
V
0
1
0 to 10
0 to 15
0 to 20
0 to 25
0 to 50
0 to 100
0 to 200
0 to 250
0 to 400
0 to 500
0 to 1000
0 to 2000
0 to 5000
0 to 6000
Others
T
A
B
C
D
J
E
F
G
H
M
X
Y
4
7
8
9
Modification No.
Flange rating
A
B
C
D
10K 50A (JIS B 2212)
10K 100A
DN50/PN10
ANSI #150 2B
0
1
A
B
1
2
3
4
5
10 Number of analog output points 2 points
4 points
11 Number of analog input points
2 points
6 points
4 to 20mA DC
12 Analog output
0 to 20mA DC
0 to 1V DC
0 to 5V DC
1 to 5V DC
13 Contact output/input
5 output points, No input
0
5output points, 3 input points
5m
10m
20m
30m
40m
50m
1
A
B
C
D
E
F
14 Cable length between receiver
and control unit
Note 3
Note 6
Note 1) When O2 conversion is required,
specify the reference O2
G
H
X
A
B
C
D
E
F
80m
100m
Others
2m
5m
10m
15m
20m
25m
Others
Japanese
English
Chinese
0m
1m
2m
3m
4m
5m
6m
7m
8m
9m
0.0m
0.1m
0.2m
0.3m
0.4m
0.5m
0.6m
0.7m
0.8m
concentration (settable within
0 to 19vol%, O2: Integer).
(HCl meter, NH3 meter, CO meter)
Note 2) Specify the range within the
max/min range calculated from
path length.
Note 4
Note 7
15 Cable length between receiver
and transmitter
If the measuring range x stack
length (optical path length)
exceeds 1000ppm/m,
X
16 Display and operation manual
J
E
C
0
1
2
3
4
5
6
7
8
9
0
1
2
3
4
5
6
7
8
9
0
5
9
N
D
consult Fuji.
Note 3) Cable length between receiver
and control unit: Max. 100m
Note 4) Cable length between receiver
and transmitter: Max. 25m
Note 5) Specified to use in high
temperature gas: 500°C or more,
and 1200°C or less
Note 6) Others (specified with X) are
possible only for 10 m or more.
Note 7) Others (specified with X) are
possible only for 5 m or more.
Note 8) Specify ‘D’ when dust exceeds
5g/Nm3-m.
Note 9) If H2O is contained in measured
component, be sure to write the
measured gas temperature into
“Specification sheet” at the last
page of this catalog, or contact
directly to Fuji’s service
18 Measuring optical path length
(unit: 1m)
19 Measuring optical path length
(unit: 0.1m)
department.
0.9m
0.00m
0.05m
How to specify measuring optical path
length
Example: 2.25m 2 for 18th digit, 2 for
19th digit and 5 for 20th digit
10m 9 for 18th digit, 9 for
19th digit and 9 for 20th digit
20 Measuring optical path length
(unit: 0.01m)
(Used only when 10m is specified)
Standard
Dust proof
21 Dust proof specification
Note 8
4
lIst For comBInatIons oF measuraBle components, unIts and measurement ranGes
Measurable range 6th digit
(Unit: When 5th digit is 1 or 3)
Not feasible
Measurable range 6th digit
(Unit: When 5th digit is 5)
K, Q, S, L, V, 0, 1, T, A, X
V, 0, 1, T, A, X
Not feasible
Not feasible
Measurable component
4th digit
CO
A
B
C
F
G
H
K
CO (For use in high temperature)
HC
HC+H2O
CO2
CO2 (For use in high temperature)
CO+CO2
Not feasible
V, 0, 1, T, A, B, C, D, J, E, F, G, H, Z
A, B, C, D, J, E, F
Not feasible
Not feasible
Not feasible
Not feasible
Not feasible
Not feasible
Not feasible
K, Q, S, L, V, 0, 1, T, A, X
V, 0, 1, T, A, X
Q, S, L, V, 0, 1, T, A, X (CO)
Q, S, L, V, 0, 1, T, A, X (CO2)
V, 0, 1, T, A, X (CO)
V, 0, 1, T, A, X (CO2)
S, L, V, 0, 1, T, A, B, X
S, L, V, 0, 1, T, A, X
K, Q, S, L, V, 0, 1, T, A, X
Not feasible
CO+CO2
L
(For use in high temperature)
O2(Class 1 Laser)
O2(For use in high dust)
CH4
NH3
NH3+H2O
P
Q
R
W
X
Not feasible
B, C, D, J, E, F, G, H, M, Z
0, 1, T, A, B, C, D, J, E, F, G, H, Z
A, B, C, D, J, E, F
Not feasible
scope oF delIverY
• Receiver unit
• Transmitter unit
• Control unit
•
Cable between receiver unit and transmitter unit (specified
length)
•
Cable between receiver unit and control unit (specified length)
• Standard accessory set, instruction manual
optIonal Items
• Spare parts for one year (ZBN1SS12)
• Calibration gas cell (*1) (*2)
*1: One set of the cables and calibration gas cell are neces-
sary for installation and annual maintenance.
*2: Standard length 1m (200mm or 500mm when the mea-
suring range is low concentration)
• Cable between receiver unit and transmitter unit (for
calibration) (*1)
• Cable between receiver unit and control unit (for calibra-
tion) (*1)
• Standard gas (ZBM), pressure regulator (ZBD)
• Recorder (when necessary, Fuji's product type PHL/PHR, etc.)
• Others
standard accessorIes
Name
Quantity
8 (16)
Specification
M16×55 (70) SUS (*)
M16 SUS (*)
Bolt
8 (16)
Nut
8 (16)
M16 SUS (*)
Spring washer
Flat washer
8 (16)
M16 SUS (*)
Companion flange packing or
flange packing specified for
use in high temperature
2
According to flange specifications
Hex socket bolt M8 × 70
6
2
1
Bolt for angle adjustment
Power supply fuse
Instruction manual
(*When "B" or "C" is specified in the 9th digit in a code symbol, quantity is 16 pieces. 8 pieces are attached in other cases.)
(*When "B", "C" or "D" is specified in the 9th digit, Bolt length is 70mm. It is 55mm in other cases. Inch-sized bolts are not
applicable.)
spare parts For one Year (ZBn1ss12)
Name
Quantity
2 pieces For bellows (ZZP*ZSSTK7N3508P1)
2 pieces (ZZP*ZSSTK7P2530P3)
Remarks
Silicon packing A
O-ring
5
ZSS
calIBratIon/InstallatIon FIxtures lIst (optIon)
Parts name
Quantity Type
(3) Calibration gas cell / about 1000mm
Cable between receiver unit and
control unit (for calibration)
Cable between receiver unit and
transmitter unit (for calibration)
Calibration gas cell (*3)
1
1
1
ZZP ZSSTK4J1271C2
(1)
(2)
*
ZZP ZSSTK4J0641C3
*
ZZP ZSSTK4J3676C1
*
(HCl, NH3, CO,CO2, CO+CO2,CH4)
Calibration gas cell (*3)
(3)
(4)
ZZP ZSSTK4J5026C1
*
(HCl+H2O, NH3+H2O, O2)
Optical axis adjusting tool (laser
pointer, target)
(2) Cable between receiver and
transmitter (approx. 1600mm)
1
ZZP ZSSTK4J1274C1
*
(5) IR card (for NH3)
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
ZZP ZSSTK7N4505P1
*
(6) Check cell
ZZP ZSSTK4J2605C1
*
(7) Filter regulator
ZZP ZSSTK7F9554P2
*
(8) Mist separator
ZZP ZSSTK7H8049P1
*
(9) R1/4 stopper (plug) for mist separator
Flow meter (20 to 100L/min) (*4)
(10) Flow meter (4 to 50L/min) (*4)
Flow meter (30 to 300L/min) (*4)
Air-set box (20 to 100L/min)
ZZP ZSSTK738114P7
*
ZZP ZSSTK7N4624P1
*
Parts used for calibration
(1) Cable between receiver unit and control unit
ZZP ZSSTK7N4624P2
*
ZZP ZSSTK7N8849P1
*
ZZP ZSSTK7N6685C1
*
(11) Air-set box (4 to 50L/min)
Air-set box (30 to 300L/min)
ZZP ZSSTK7N6685C2
*
ZZP ZSSTK7N6685C3
*
Stack
Air purge mechanism (20 to 100L/min)
(12) Air purge mechanism (4 to 50L/min)
Air purge mechanism (30 to 300L/min)
(13) BNC cable for optical axis adjustment
ZZP ZSSTK7P1433C1
*
ZZP ZSSTK7P1433C2
*
ZZP ZSSTK7P1433C3
*
(4) Optical axis adjusting tool
(laser pointer)
ZZP ZSSTK7P2524C1
*
(4) Optical axis adjusting tool (target)
* 3: Standard length 1m (200mm or 500mm when the
measuring range is low concentration)
* 4: Usually, 2 units are required to adjust each air purge
for receiver unit and transmitter unit.
Parts for adjusting the optical axis
Transmitter unit
(5) IR card
IR CARD
Laser beam (visible)
Invisible light
Invisible light / visible fixtures(for NH3)
(6) Check cell
Transmitter unit
Receiver unit
Stack
Control unit
Laser gas analyzer with check cell (ZSS)
SV3
SV2
Atmospheric
release
SV1
(AO)
Sequencer
Remote check input (AI)
Supplies for checking span value on site
N2 cylinder Standard gas
cylinder
6
(11) Air set box
4-φ11
(Mounting hole)
Parts for air purge
(7) Filter regulator
Box
(8) Mist separator
Flow
Mist separator
Filter regulator
meter
Pipe with
external
diameter φ10
(out of supply
scope)
(IN)
(OUT)
(OUT)
(IN)
(OUT)
(IN)
(OUT)
(IN)
measuring
instrument
air inlet
Receiver
purge outlet
Drain outlet
(φ10 piping)
(not supplied by us)
Transmitter
purge outlet
Piping system diagram
(10) Flow meter
420
500
150
110
65
43
Measuring instrument
air inlet
Rc1/4
Receiver
purge outlet
Rc1/4
Transmitter
purge outlet
Rc1/4
Drain outlet
Piping retainer for φ10
(12) Air purge mechanical part
1m
total 20m
Piping material: φ8/6 Teflon pipe
Mist separator
to Transmitter
purge inlet
Filter regulator
for φ8 tube
(OUT)
(OUT)
(IN)
(metallic tube
recommendable)
(IN)
(OUT)
to Receiver
purge inlet
(IN)
Measuring
instrument air inlet
0.5m
0.5m
to Receiver
purge inlet
(Coupling joint)
Piping system diagram
(13) BNC cable for optical axis adjustment
7
ZSS
outlIne dIaGram (uꢁiꢂ : ꢃꢃ)
Or air-set box
Copper or SUS pipe (not supplied)
Instrumentation airline joint (not supplied)
2-Flowmeter
(option)
Filter regulator (option)
20 to 100L/min
φ8/6 Teflon pipe
Instrumentation
air inlet
(not supplied)
Mist separator
(option)
150 to 200
Approx. 400
Flange of angle
adjustment
150 to 200
Note 1
Approx. 400 (432)
With bolt and nut
(M16×55)
Transmitter box
(160×240×100)
Flange of angle
adjustment
Purging gas inlet
Flange pipe to
Receiver box
(with bell-and-spigot
(120×180×100)
mount stack
joint for
φ
8/6)
(not suppried)
Purging gas inlet
100
16
100
Stack
0.5 to 10m
JIS10K50A FF
(with packing)
φ34
(Connector)
Receiver unit
130
Transmitter unit
φ34
(Connector)
74
2-φ12
Cable between receiver unit
and transmitter unit (standard : 2m)
2-φ12
74
Note 1: Be sure to set the R1/4 sealing
plug (ZZP*ZSSTK738114P7)
on the drain outlet of the mist
separator.
240
φ10.5
Control unit
LASER GAS ANALYZER
ZSS
134
3
Power
supply
AI
DI/O
AO
7-Waterproof grand
74
(Internal diameter φ12)
12
Cable length
(200mm)
Mounting diagram
Connector
Be sure to prepare instrument air (N2 for an O2 meter).
Note) Air purge should be performed if the equipment is
installed in a flue whether or not it is operated.
Cable between receiver unit and control unit
(standard : 5m)
8
external connectIon dIaGram (WHen numBer oF analoG output Is 2 poInts)
There are two types of output/input terminal.
Power supply terminal
1
2
3
Screw diameter : M4
100 to 240 V AC
(50/60Hz)
FG
AO terminal
21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17 18 19 20
Screw diameter : M2 or the equivalent
1
AO1+
AO1-
21 AO2+
22 AO2-
Analog output 1
Analog output 2
2
Al terminal
21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17 18 19 20
Screw diameter : M2 or the equivalent
3
AI1+
AI1-
23 AI4+
24 AI4-
25 AI5+
26 AI5-
27 AI6+
28 AI6-
Analog input 1
Analog input 2
Analog input 4 (AI extension board is required)
Analog input 5 (AI extension board is required)
Analog input 6 (AI extension board is required)
4
5
6
7
8
AI2+
AI2-
AI3+
AI3-
Analog input 3 (AI extension board is required)
DI/DO terminal
21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17 18 19 20
Screw diameter : M2 or the equivalent
9
DO1
29 DO7
30 DO7
31 DO8
32 DO8
33 DI1
34 DI1
35 DI2
36 DI2
37 DI3
38 DI3
39 DI4
40 DI4
Low light transmission
Device failure
10 DO1
11 DO2
12 DO2
13 DO3
14 DO3
15 DO4
16 DO4
17 DO5
18 DO5
19 DO6
20 DO6
Power supply off
During hold / during calibration
Average value reset signal (option)
Switching instantaneous value/moving average value (option)
Remote hold (option)
Outside the range of upper /
lower limits (component 1)
Range identification (component 1)
Note) Unassigned terminal may be connected to internal circuit. So they should not be used as repeating terminals.
9
ZSS
external connectIon dIaGram (WHen numBer oF analoG output Is 4 poInts)
Power supply terminals
1
2
3
AC100 to 240V
FG
50/60Hz
Screws diameter: M4
AO terminal
AO4 Unassigned Unassigned
1
2
3
4
AO1+
AO1-
AO2+
AO2-
5
6
7
8
AO3+
AO3-
AO4+
AO4-
+
-
Analog output 1
Analog output 2
Analog output 3
Analog output 4
7
8
9
3
10 11 12
1
2
4
5
6
+
-
+
-
+
-
AO1
AO2
AO3
AI terminal
AI4
AI5
AI6
+
1
2
3
4
5
6
AI1+
AI1-
7
8
9
AI4+
AI4-
Analog input 4
(AI extension board is required)
Analog input 1
Analog input 2
+
-
+
-
-
7
8
9
10 11 12
AI2+
AI2-
AI5+
Analog input 5
(AI extension board is required)
10 AI5-
11 AI6+
12 AI6-
1
2
3
4
5
6
Analog input 3
(AI extension board
is required)
AI3+
AI3-
Analog input 6
(AI extension board is required)
+
-
+
-
+
-
AI1
AI2
AI3
DI/DO terminal
25 26 27 28 29 30 31 32 33 34 35 36
13 14 15 16 17 18 19 20 21 22 23 24
1
2
3
4
5
6
7
8
9 10 11 12
1
2
3
4
5
6
7
8
9
DI1
DI1
DI2
DI2
DI3
DI3
DI4
DI4
DI5
19 DO4
Average value reset signal (option)
Outside the range of upper/lowerlimits (component 1)
Range identification (component 1)
20 DO4
21 DO5
22 DO5
23 DO6
24 DO6
25 DO7
26 DO7
27 DO8
28 DO8
29 DO9
30 DO9
31 DO10
32 DO10
33 DO11
34 DO11
Switching instantaneous value/moving
average value (option)
Remote hold (option)
Power supply off
10 DI5
13 DO1
14 DO1
15 DO2
16 DO2
17 DO3
18 DO3
Low light transmission
Device failure
During hold/during callibration
Note) Unassigned terminals may be connected to internal circuit. So they should not be used as repeating terminals.
10
mountInG dImensIons (uꢁiꢂ: ꢃꢃ)
Do not install the device in a stack if air purge is not performed.
It may become unusable because the window will get dirty.
Stack
c
d
a
a: Inner diameter of φ50mm or more
b: 20 to 70mm (*5)
b
c: Approximately 200 to 250mm (*6)
d: 400mm or more (*7)
Flange pipe to mount stack (*9)
e
e: 500 to 1500mm (*7)
Scaffold (*8)
*5: When used in an enviroment with high dust, reserve at least 50mm.
*6: When internal gas temperature is high, reserve the distance of 250mm or more. Note that, adjustable angle
range narrows down, so mounting angle of flange pipe to mount stack should be adjusted accurately.
*7: Sufficiently secure a space for installation which permits easy operation.
*8: Scaffold is required for the both side of receiver unit and transmitter unit.
*9: Use a flange provided with a valve where there is a risk of gas injection or the operator may be exposed to
dangerous situation.
*10: Tighten the flange with a torque of 118±114N-m. When tightening the flange, apply grease to the bolts
(recommendable to use grease containing molybdenum).
rouGH GuIde to mount tHe companIon FlanGe
When is larger than 5°, mount it within the angle (≈5°) inside the circle with the A radius.
θ
γ
Y
X
A
θ
Z
X: Inner diameter of flange of receiver unit and transmitter unit
Y: Length of flange of receiver uit and transmitter unit
Z: Length of stack
X
γ, θ = ATAN
2 (Y + 125)
A ≈ 0.087 × (Z + Y + 125)
11
ZSS
Please specify the following items when you contact us for inquiry.
1.Code symbol ZSS
2. Parameter of measured gas
Item
Measured gas concentration
Temperature (°C)
Pressure (kPa)
Min. value Typical value Max. value
Flow (m/s)
Moisture (vol%)
Dust (mg/Nm3)
Other gas component (vol%/ppm)
Other gas component (vol%/ppm)
Other gas component (vol%/ppm)
3.Stack diameter (Inner diameter) (measured optical path length)
A
m
m
m
4.Cable between receiver unit and control unit.
B
5.Cable between receiver unit and transmitter unit.
C
A
Instrumentation air
Instrumentation air
Stack
Receiver unit
Transmitter unit
C
Control unit
B
6.Spares for 1-year measurement
7. Optional items
no need necessary
no need necessary
8. Output of O2 correction value (for only HCl and NH3 meters)
no need necessary
9.Oxygen correction reference value (%) (When "necessary" is selected in 8)
%
10.Vibration
no need necessary (
G)
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fjielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan